SCND Mdle PDF
SCND Mdle PDF
MODULE II
Prepared by: Muhammedali Shafeeque K
+91 9744144505
[email protected]
onlinemas.weebly.com
References:
1. A.K. Sawhney, A course in Electrical and Electronics Measurements and
Instrumentation, Dhanpat Rai and sons
2. Dhogal ,Basic Electrical Engineering, Volume 2
3. U.A.Bakshi, A.V.Bakshi, Electrical Instrumentation
Construction: Figure illustrates the main pans and connection of this instrument.
It consists of two permanent magnets having north and south poles and an aluminium or copper
disc attached to a spindle. This disc is kept under the poles of magnets in an air-light chamber
formed by placing a fiber ring in between to round brass disc B and B’. This airtight chamber
contains mercury and on its left side a conductor C passes through fiber ring. This conductor
makes contact with the mercury. The aluminium disc floats over the mercury and lower end of
its spindle rests on an adjustable screw A. The mercury in the chamber reduces the friction of the
disc at A and it alio provides path for the current to flow through the disc. For holding the
magnets in position, there are two iron bars in the instrument, one above the chamber and other
at the bottom of the chamber. On the lower iron bar, a coil of few turns is wound which is known
as a friction compensating coil. A recording device is attached at the upper end of the spindle.
When the instrument is in use, the current is passed in the disc through the conductor C and
mercury and from where it goes to the current coil and pressure coil, which interacts with the
aluminium disc placed near the coils between bearings. This field induces eddy currents in the
disc and hence the disc rotates.
Single Phase Induction Type Meters:. The construction and principle of operation of single
Phase Energy explained below :
Construction of Induction Type Energy Meters. The construction varies in details from one
product to the next. However, the differences are very minor in nature. There are four main parts
of the operating mechanism:
Copper shading bands are provided in the central limb.The position of these bands are
adjustable.The function of these bands is to bring the flux produced by the shunt magnet exactly
to quadrature with the applied voltage.
Moving System: This consists of an aluminium disc mounted on a light alloy shaft.This disc
is positioned in the air gap between series and shunt magnets. The upper bearing of
toe rotor (moving system) is a steel pin located in a bole in the bearing cap fixed to the top of
the shaft. The rotor runs on a hardened steel pivot, screwed to the foot of the shaft. The pivot is
supported by a jewel bearing. A pinion engages the shaft with the counting or registering
mechanism.
A unique design for the suspension of the disc is used in the floating-shaft energy meter.
Here the rotating shaft has a small magnet at each end, where the upper magnet of the shaft is
attracted to a magnet in the upper bearing and the lower magnet of the shaft is attracted to a
magnet in the lower bearing. The moving system thus floats without touching either bearing
surface, and the only contact with the movement is that of the gear connecting the shaft with the
gear of the train, thus the friction is drastically reduced.
Braking system : Permanent magnet positioned near the edge of the aluminium disc forms the
braking system. The aluminium disc moves in the field of this magnet and thus provides a
braking torque. The position of the permanent magnet is adjustable, and therefore, braking torque
can be adjusted by shifting the permanent magnet to different radial positions as explained
earlier.
Registering (counting) Mechanism: The function of a registering or counting mechanism is to
record continuously a number which is proportional to the revolutions made by the moving
system. By suitable system,a train of reduction gears the pinion on the rotor shaft drives a series
of five or six pointers. These rotate on round dials which are marked with ten equal divisions.
The pointer type of register to shown in Fig. Cyclo-Meter register as shown in fig can also be
used.
Theory of Operation of Single Phase Energy Meters. A simple functional diagram of the driving
system of the meter is shown in Fig. The corresponding phasor diagram is also shown.
The supply voltage is applied across the pressure coil. The pressure coil winding is highly
inductive as it has very large number of turns and the reluctance of its magnetic circuit is very
small owing to presence of air gaps of very small length. Thus the current Ip through the pressure
coil is proportional to the supply voltage and lags it by a few degrees less than 90°. This is
because the winding has a small resistance and there are iron losses in the magnetic circuit.
Current Ip produces a flux Φpt.This flux divides itself into two parts Φg and ΦP.The major
portion Φg flows across the side gaps as reluctance of this path is small. The reluctance to the
path of flux ΦP is large and hence its magnitude is small. This flux ΦP goes across aluminium
disc and hence is responsible for production of driving torque. Flux ΦP is in phase with current Ip
and is proportional to it.
Therefore flux ΦP is proportional to voltage V and lags it by an angle a few degrees less than 90°.
Since flux ΦP is alternating in nature, it induces an eddy emf in the disc which in turn produces
eddy current,.Iep.
The load current I flows through the current coil and produces a flux Φs. This flux is
proportional to the load current and is in phase with it.This flux produces eddy current Ies in the
disc. Now the eddy current,Ies interacts withflux ΦP to produce a torque and eddy current Iep
interacts with Φs to produce another torque. These two torques are in the opposite direction (as
shown, in Fig) and the net torque is the difference of these.
Driving torque is given as
The required mmf is obtained from a 'lag coil 'which is located on the central limb of the shunt
magnet close to the disc gap and links with the flux that cuts the disc.The pressure coil is excited
by voltage V and carries a current Ip which produces an mmf ATpt which in turn produces a fiux
ΦP lagging the voltage by an angle 90°-Δ. The flux ΦPt divides itself into two parts Φg and ΦP.
Flux ΦP cuts the disc and also links with the "lag coil". A voltage EL is induced in this coil
lagging ΦP by 90deg. This voltage circulates a current i L through the lag coil. iL lags behind EL
by an angle γ which depends upon the reactance and resistance of the lag coil. The lag coil
current produces an mmf ATL. The flux ΦP in the disc air gap will be created by the combined
action of the main rnmf ATpt in phase with IP and the lag coil mmf ATL in phase with IL. Thus
the flux ΦP will be in phase with the resultant mmf ATp. Hence it is clear that the phase of
flux ΦP can be adjusted by varying the mmf of the shading coil either in magnitude or in phase or
both.
The arrangements for adjusting the mmf of the lag coil are:
1.Adjustable resistance: A few turns of fairly thick wire are placed around the central limb of the
shunt magnet and the circuit is closed through a low adjustable resistance (Fig. ).
The resistance of this circuit is altered to adjust the lag angles of flux ΦP. An increase in
resistance decreases the current and mmf in the lag coil and therefore the value of lag angle θ is
decreased. The value of lag angle θ is increased by decreasing the resistance of the lag coil
circuit. The resistance of the lag coil is so adjusted that A becomes equal to 90deg
2. Shading bands: In this arrangement copper shading bands L1 are placed around the central
limb of shunt magnet instead of a lagcoil with adjustable resistance (Fig.).
The adjustment can be done by moving the shading bands along the axis of the limb. As the
shading bands are moved up the limb, they embrace more flux. This results in greater values for
induced emf, current and the mmf ATL produced by the shading bands and therefore the value of
lag angle θ increases. When the shading bands are moved down limb, mmf AT L decreases and
the angle is reduced. The adjustment is so done that θ is such that it makes Δ=90 deg.
Creep:
In some meters a slow but continuous rotation is obtained even when there is no current flowing
through the current coil and only pressure coil is energized. This is called creeping, the major
cause, for creeping is over-compensation, for friction. If the friction compensating device is
adjusted to give a driving torque to compensate for starting friction which is bigger than the
running friction, there is a tendency for the disc to run even when there is no current through the
current coils Because the friction compensating torque is independent of the load current as the
compensating device is voltage actuated. The other causes of creeping are excessive voltage
across the potential coil (which is responsible for production of excessive torque by the friction
compensating device), vibrations, and stray magnetic fields.
In order to prevent this creeping two diametrically opposite holes are drilled in the disc ; the disc
will come to rest with one of the holes under the edge of a pole of the shunt magnet, the rotation
being thus limited to a maximum of half a revolution. The action may be understood by
reference to Fig. If a hole is under the edge of a pole, the circular eddy current paths in the disc
will be distorted as shown. The effective centre of the eddy-current paths is then at A 'and this
point is the central point of the equivalent magnetic pole produced by the currents. Examination
of polarities, shows that there is a resulting force on the disc, lending to move A' away from the
pole axis A. Thus the disc may creep until the hole reaches a position near the edge of a pole, but
further movement is opposed by the torque produced in the manner just described. The
magnitude of this torque is not sufficient to affect the action of the meter on load.
In some cases a small piece of iron is attached to the edge of the disc. The force of attraction
exerted by the brake magnet prevents creeping of disc.
One unit of three clement three phase clement Is always cheaper than three units of single phase energy
meter. But due to interaction between eddy currents produced by one element with Ihe flux produced by
another clement, there may be errors in the measurement by three phase energy meter. Such errors may be
reduced by suitable adjustments.
The conventional mechanical energy meter is based on the phenomenon of Magnetic Induction.
It has a rotating aluminium Wheel and many toothed wheels. Based on the flow of current, the
wheel rotates which makes rotation of other wheels. This will be converted into corresponding
measurements in the display section. Since many mechanical parts are involved, mechanical
defects and breakdown are common. More over chances of manipulation and current theft will
be higher.
In energy measure, the power information varying with time is calculated by a direct
multiplication of the voltage signal and the current signal. Assume that the current signal and the
voltage signal are cosine functions; Umax, Imax are the peak values of the voltage signal and the
current signal;
ω is the angluar frequency of the input signals; the phase difference between the current signal
and the voltage signal is expressed as φ
Then the power is given as follows:
The ideal p(t) consists of the dc component (first term, independent of t) and ac component
whose frequency is 2ω. The dc component is called as the average active power, that is:
The two ADCs digitize the voltage signal s from the current and voltage transducers. These
ADCs convert the sampled instantaneous voltage and current signals to digital. A high pass filter
in the current and voltage channel removes any dc component from the current signal. This
eliminates any inaccuracies in the real power calculation due to offsets in the voltage or current
signals.
The real power calculation is derived from the instantaneous power signal. The instantaneous
power signal is generated by a direct multiplication of the current and voltage signals.We have
derived already the dc component of this instatneous power is the average power. In order to
extract the real power component (i.e., the dc component), the instantaneous power signal is low-
pass filtered.
This scheme correctly calculates real power for both sinusoidal and nonsinusoidal current and
voltage waveforms at all power factors. All signal processing is carried out in the digital domain
for superior stability overtemperature and time.
This average power is integrated to get the energy ( E= 𝑃 𝑡 𝑑𝑡 ). Finally this signal is
converted to pulses by a didgital to frequency converter. The frequency of this output pulses is
therefore proportional to the average real power. This average real power information can, in
turn, be accumulated (e.g., by a counter) to generate real energy information.
Differential Opamp circuit with potential divider arrangement (to down the line voltage) is used
to measure the instantaneous voltage as shown below which is fed to Voltage ADC.
(Note that Static Wattmeter and Energy meter block diagram and theory only differ in averaging
in case of wattmeter and integral in case of energymeter)
Static Wattmeter:
The power is measured by a direct multiplication of the voltage signal and the current signal.
Assume that the current signal and the voltage signal are cosine functions; Umax, Imax are the
peak values of the voltage signal and the current signal;
ω is the angluar frequency of the input signals; the phase difference between the current signal
and the voltage signal is expressed as φ
Then the power is given as follows:
The ideal p(t) consists of the dc component (first term, independent of t) and ac component
whose frequency is 2ω. The dc component is called as the average active power, that is:
The two ADCs digitize the voltage signal s from the current and voltage transducers. These
ADCs convert the sampled instantaneous voltage and current signals to digital. A high pass filter
in the current and voltage channel removes any dc component from the current signal. This
eliminates any inaccuracies in the real power calculation due to offsets in the voltage or current
signals.
The real power calculation is derived from the instantaneous power signal. The instantaneous
power signal is generated by a direct multiplication of the current and voltage signals.We have
derived already the dc component of this instatneous power is the average power. In order to
extract the real power component (i.e., the dc component), the instantaneous power signal is low-
pass filtered.
This scheme correctly calculates real power for both sinusoidal and nonsinusoidal current and
voltage waveforms at all power factors. All signal processing is carried out in the digital domain
for superior stability overtemperature and time.
Theory:In order to simplify the problem, we assume that the current through coil B lags the
voltage by exactly 90 deg. Also that the angle between planes of coils is exactly 90deg, (i.e., Δ
=90°).
Now, there will be two deflecting torques, one acting on coil A and the other on coil B. The coil
windings are so arranged that the torques due to the two coils are opposite in direction. Therefore
the pointer will take up a position where these two torques are equal.
Let us consider the case of a lagging power factor of cos Φ.
This torque acts in the anticlockwise direction. The value of Mmax is the same in the two
expressions, owing to similar constructions of the coils.
The coils will take up such a,position that the two torques are equal.
Hence at equilibrium TA=TB
or θ=Φ
Therefore the deflection of the instrument is a measure of phase angle of the circuit, The scale of
the instrument can be calibrated in directly in terms of power factor.
Now V1=V2=V3=V.
TA= KV12IMmax cos (30°+Φ) sin (60°+θ) = 3 KVIMmax cos (30°+Φ) sin (60°+θ)
Torques TA and TB act in the opposite directions and the moving system takes up a position
where TA = TB.
Thus the angular deflection of the pointer from the plane of reference is equal to the phase angle
of the circuit to which the meter is connected.
The three phase power factor meter gives indications which are independent of waveform and
frequency of supply, since the currents in the two moving coils are equally affected by any
change of frequency.
For measurement of power factor in 3 phase unbalanced systems a two element power factor
meter (where 2 sets of fixed coils and 2 sets of moving coils mounted on the spindle) has to be
used.
to a magnet in the upper bearing and the lower magnet of the shaft is attracted to a magnet in the
lower bearing. The moving system thus floats without touching either bearing surface, and the
only contact with the movement is that of the gear connecting the shaft with the gear of the train,
thus the friction is drastically reduces.
Braking System: A permanent magnetpositioned near the edge of the aluminium disc formsthe
braking system. The aluminium disc moves in the field of this magnet and thus provides a
braking torque. The position of the permanent magnet is adjustable, and therefore, braking
torque can be adjusted by shifting the permanent magnet to different radial positions as explained
earlier.
Registering (counting) Mechanism: The function of a registering or counting mechanism is to
record continuously a number which is proportional to the revolutions made by the moving
system.
MEASUREMENT OF RESISTANCE
Ohmmeters:
Ohmmeter is a convenient direct reading device for measurement of resistance. These
instruments have a low degree of accuracy. The statement regarding accuracy is not intended in
an unfavorable sense there is a wide field of application for this instrument in determining the
approximate value of resistance. An ohmmeter is useful for getting the approximate resistance of
circuit components such as heater elements or machine field coils, measuring and sorting
resistors used in electronic circuits and for checking continuity of circuits. It is also- useful in
laboratories as an aid to a precision bridge, for it can help to know the approximate value of
resistance which can save time in balancing the bridge.
Series-type Ohmmeter: A circuit of a series-type ohmmeter is shown in Fig. It consists of
basic d'Arsonval movement connected in parallel with a shunting resistor,R2 . This parallel
circuit is in series with resistance R1 and a battery of emf E. The series circuit is connected to the
terminals A and B of the unknown resistance Rx.
It is observed that when the unknown resistance Rx=0 (terminals A and B shorted) maximum
current flows through the meter. Under this condition resistor R2 is adjusted until (he basic
movement (meter) indicates full scale current In. The full-scale current position of the pointer , is
marked "0 ohm" on the scale. Similarly when Rx is removed from circuit, Rx=infinity (that is
when terminals A and B are open), the current in the meter drops to zero and the movement
indicates zero current which is then marked "Infinity". Thus the meter will read infinite
resistance at the zero current position and zero resistance at full scale current position. Since
zero resistance is indicated when the current in the meter is maximum and hence the pointer goes
to the top mark. When the unknown resistance is inserted at terminals A, B the current through
the meter is reduced and hence pointer drops lower on the scale. Therefore the meter has "0" at
extreme right and "infinity" at the extreme left. Intermediate scale markings may be placed on
the scale by different known values of resistance Rx to the instrument. The accuracy of these
scale markings depends on the repeating accuracy of the movement and the tolerances of the
calibrating resistors. Fig. 8'17 {a) shows the shape of scale of series type ohmmeter.
A convenient quantity to use in the design of a series ohmmeter is the value of Rx which causes
the half scale deflection of the meter. At this position, the resistance across terminals A and B is
defined as the half scale position resistance Rh. The design can be approached by recognizing the
fact that when Rh is connected across terminals A and B the meter current reduces to 1/2 of its
full scalevalue.
where Im=current through the meter, Ifs-current through the meter for full scale deflection. This
clearly means that Rh is equal to the internal resistance of the ohmmeter looking into terminals A
and B.
(1)
The battery current at half scale deflection ,
In order to produce full scale deflection the battery current must be doubled
(2)
Current through the shunt
(3)
(4)
From (1) ,(4)
Shunt type Ohmmeters: The circuit diagram of a shunt type ohmmeter is shown in Fig It
consists of a battery in series with an adjustable resistor R1 and a basic d'Arsoival movement
(meter). The unknown resistance is connected across terminals A and B parallel with the meter.
In this circuit it is necessary to have an "off-on"switch to disconnect the battery from the circuit
when the instrument is not in use. When the unknown resistor Rx=0 (A and B are shorted), the
meter current is zero. If the unknown resistance Rx=inf (A and B are open), the current finds
path only through the meter and selecting a proper value for resistance R1, the pointer may be
made to read full scale. This ohmmeter therefore has "zero" mark on the left hand side of the
scale (no current) and infinite mark on the right hand side of the scale (full scale deflection
current).
Fall of Potential Method. Fig shows the circuit for measurement of earth resistance with fall of
potential method.
A current is passed through earth electrode E to an auxiliary electrode B (which is usually an iron
spike) inserted in earth at a distance away from the earth electrode. A second auxiliary electrode
A is inserted in earth between E and B. The potential difference V between E and A is measured
for a given current I. The flow of ground currents is shown in Fig The lines of the first electrode
current diverge and those of the second electrode current converge. As a result the current
density is much greater in the vicinity of the electrodes than at a distance from them. The
potential distribution between the electrodes is shown in Fig
It is apparent from this curve that the potential rises in the proximity of electrodes E and B and is
constant along the middle section. The resistance of earth, therefore, is RE = V/I or VEA/I. The
position of electrodes E and B is fixed and the position of electrode A is changed and resistance
measurements are done for various positions of electrode A.
A graph is plotted between earth resistance against the distance between electrode E and A. This
graph is shown in Fig.
From Fig, it is clear that the measured value of earth resistance depends upon the position of the
auxiliary electrode A. The earth resistance rises rapidly initially, When the distance between
earth electrode B and auxiliary electrode A is increased, it then becomes constant, and when [ the
auxiliary electrode A approaches the, auxiliary electrode B,the resistance rises again. The placing
of electrodes is thus very important and serious error may be caused by incorrect placing of the
electrodes. The correct value of resistance of earth,RE, is when the auxiliary electrode A is at
such a distance that the resistance lies on the flat part of curve of Fig. The spacing between the
earth electrode E and the auxiliary electrodes A, B should be large so as to get proper results. The
distance may be a few hundred metres in case the earth resistance is low.
Earth Tester.
The resistance of earth can be measured by an earth tester shown in Fig.
The "Earth Tester" is a specialtype of Megger and it has some additional constructional features
and they are :
(i) a rotating current reverser, and (ii) a rectifier.
Both these additional features consist of simple commutators made up of 'L' shaped segments.
They are mounted on the shaft of the hand driven generator. Each commutator has four fixed
brushes. One pair of each set of brushes is so positioned that they, make contact, alternately ,
with one segment and then with the other as the commutator rotates. The second pair, of each of
set of brushes is positioned on the commutator so that continuous contact is made with one
segment whatever the position of the commutator. The earth tester has four terminals P1, P2 and
C1, C2. Two terminals P1 and C1 are shorted to form a common point to be connected to the
earth electrode. The other two terminals P2 and C2 are connected to auxiliary electrodes P and C
respectively. The indication of the earth tester depends upon the ratio of the voltage. across the
pressure coil and the current through the coil. The deflection of its pointer indicates the
resistance of earth directly. Although the "Earth Tester", which is a permanent magnet moving
coil instrument and can operate on d.c. only, yet by including the reverser and the rectifying
device it is possible to make measurements with a.c. flowing in the soil.
The sending of a.c, current through the soil has many advantages and therefore this system is
used. The use of a.c. passing through the soil eliminates unwanted effects due to production of a
back emf in the soil on account of electrolytic action. Also the instrument is free from effects of
alternating or direct currents presents in the soil.
Megger:
. The essential parts of a Megger are shown in Fig.. The current coil is similar to that of the
permanent magnet moving coil instrument. There are two voltage (potential) coils V1 and V2
The voltage coil V1 embraces (threads over) the annular magnetic core. It is clear from Fig. 8'25
that voltage coil V1 is in a weak magnetic field when the pointer is at 'inf' position and hence this
coil can exert very little torque.
The torque exerted by the voltage coil increases as it moves into a stronger field and this torque
is maximum when it is under the pole face and under this condition the pointer is at its zero end
of the resistance scale. In order to modify further the torque in the voltage circuit, another
voltage coil V2 is used. This coil is also so located that it moves into stronger field as the pointer
moves from the 'inf' position towards the zero position of the resistance scale. The coil finally
embraces (threads around) the extension H of the pole piece.
The combined action of the two voltage coils V1 and V2 may be considered as though the coils
constituted a spring of variable stiffness, being very stiff near the zero end of the scale where the
current in the current coil is very large (on account of unknown resistance Rx being small), and
very weak near the 'inf' end of the scale where the current in the current coil is very small (on
account of unknown resistance Rx being very large).
Thus this effect compresses the low resistance portion of the scale and opens up the high
resistance portion of the scale This is a great advantage since this instrument is meant to be used
as ''insulation tester" and the insulation resistances are quite high.
The voltage range of the instrument can be controlled by a voltage selector switch. This can be
done by varying resistance 'R' connected in series with the current coil. The test voltages, usually
500, 1000 or 2500 V are generated by a hard cranked generator G. A centrifugal clutch is incor-
porated in the generator drive mechanism which slips at a predetermined speed so that a constant
voltage is applied to the insulation under test. This voltage provides a test on strength of low
voltage insulation as well as a measure of its insulation resistance, since it is sufficient to cause
breakdown at faults. Such breakdowns are indicated by sudden motion of the pointer off scale at.
zero end. As the same magnet system supplies magnetic fields for both instrument and generator,
and as current and voltage coils move in a common magnetic field, the instrument indications are
independent of the strength of the magnet.
The galvanometer G, measures the current IR between the conductor and the metal sheath. The
leakage current IL, over the insulating material is carried by the guard wire wound on the
insulation and therefore does not flow through the galvanometer.Cables without metal sheaths
can be tested in a similar way if the cable, except the end of ends on which connections are
made, is immersed in water in a tank. The water and the tank then form the return path for the
current. The cable is immersed in slightly saline water for about 24 hours and the temperature is
kept constant (at about 20°C) and then the measurement is taken as in Fig.
The insulation resistance of the cable R=V/IR
In some cases, the deflection of the galvanometer is observed and its scale is afterwards
calibrated by replacing the insulation by a standard high resistance (usually 1 Mohm), the
galvanometer shunt being varied, as required to give a deflection of the same order as before.
In tests on cables the,galvanometer should be short-circuited before applying the voltage. The
short circuiting connection is removed, only after sufficient time has elapsed so that charging and
absorption currents cease to flow. The galvanometer should be well shunted during the early
stages of measurement, and it is normally desirable to include a protective series resistance (of
several megohm) in the galvanometer circuit. The value of this resistance should be subtracted
from the observed resistance value in order to determine the true resistance. A high voltage
battery of 500 V emf is required and its emf should remain constant throughout the test.
Murray Loop Test: The connection for this test are shown in Fig a relates to the ground fault and
Fig. b relates to the short circuit fault.
In both cases, the loop circuit formed-by the cable conductors is essentially a Wheatstone bridge
consisting of resistances P, Q, R and X. G is a galvanometer for indication of balance.
The resistors P, Q forming the ratio arms may be decade resistance boxes or slide wires.
where (R+X) is total loop resistance formed by the sound cable and the faulty cable. When the
conductors have the same cross-sectional area and the same resistivity, the resistances are
proportional to lengths. If l1 represents the length of the fault from the test end and l the length
of each cable. Then,
The above relation shows that the position of the fault may be located when the length of the
cable is known. Also, the fault resistance does not alter the balance condition because it
(resistance) enters the battery circuit However, if the magnitude of the fault resistance is high,
difficulty may be experienced in obtaining the balance condition on account of decrease in
sensitivity and hence accurate determination of the position of the fault may not be possible. In
such a case, the resistance of the fault may be reduced by applying a high direct or alternating
voltage-in consistence with the insulation rating of the cable on the line so as to carbonize the
insulation at the point of the fault.
A single pole double throw switch K is used in this circuit. Switch K is first thrown to position '1'
and the resistance S is varied and balance obtained.
Let the value of S for balance be S1. The four arms of the Wheatstone bridge arc P, Q. At
balance ;
This determines R+X i.e., the total loop resistance as P, Q and S1 are known.
Then switch K is then thrown to position '2' and the bridge is rebalanced. Let the new value of S
for balance be S2. The four arms of the bridge now are P, Q, R,S2.
At balance,
Hence X is known from the known values of P, Q, S2 from this equation and R+X (the total
resistance of 2 cables) as determined . Knowing the value of X, the position of the fault is
determined as :
AC BRIDGES:
General form of an A.C Bridge: As an example let us consider the bridge circuit of Fig. R3 and
R4 are non-inductive resistances L1 and L2 are inductances of the negligible resistance and R\
and R2 are non-inductive resistors.
At balance condition,
Then, if L1 and R1 are unknown, the above bridge may be used measure these quantity in terms
of R2, R3,R4 and L2. Two balance equations are always obtained for an a.c. bridge circuii.
This follows from the fact that for balance in an a.c. bridge, both magnitude and phase
relationships must be satisfied. This requires that real and imaginary terms must be separated,
which gives two equations to be satisfied for balance.
(R1+jωL1)R4=(r2+R2+ jωL2)R3
and
Then,
Schering’s Bridge: Used to measure capacitance. The connections and the phasor diagrams for
balance conditions are shows in Fig.
MAGNETIC MEASUREMENTS
Measurement of Flux and Flux Density. The measurement of flux density inside a specimen
can be done by winding a search coil over the specimen. This search coil is known as a "B coil".
Ibis search coil is then connected to a ballistic galvanometer or a flux meter. Let us consider
that we have to measure the flux density in a ring specimen shown in Fig. The ring specimen is
wound with a magnetizing winding which carries a current I. A search coil of convenient number
of turns is wound on the specimen and connected through a resistance and calibrating coil, to a
ballistic galvanometer as shown.
The current through the magnetizing coil is reversed and therefore the flux linkages of the search
coil change inducing an emf in it. This emf sends a current through the ballistic galvanometer
causing it to deflect.
Flux density,
The search coil, as positioned in Fig. , measures the value of flux density, Bo, in air. This
search coil is called an "H coil". While for the flux densities encountered in iron testing, there is
usually no trouble in getting a good sensitivity by using B coil of sufficient turns but there is
some difficulty in securing adequate sensitivity in the 'H' coil placed at the surface, In the first
instance, its cross-sectional area is much smaller than the coil surrounding the specimen (i.e. B
coil) and then H is not constant across the section Secondly, the permeability of iron is very large
as compared to air-may be some thousand times and therefore Mux density B0, in search coil, is
very smalt compared to that in specimen. The value of flux density B0 in H coil is measured in a
similar manner as described above for determination of B in the specimen.
Flux Meter
The flux meter is a special type of ballistic galvanometer in which the controlling torque is
very small and the electromagnetic damping is heavy.
Construction. :The construction of a flux meter is shown in Fig. . In general the construction is
similar to that of a moving coil milli-ammeter. A coil of small cross-section is suspended from a
spring support by means of a single silk thread. The coil moves in the narrow gap of a permanent
magnet. There are no control springs. The current is led into the coil with the
help of a very loose helices of very thin, annealed silver strips. The controlling torque is thus
reduced to minimum.
The coil is formerless and the air friction damping is negligible.
Operation. The terminals of the flux meter are connected to a search coil as shown in Fig. The
flux linking with the search coil is changed either by removing the coil from the magnetic field
or by reversing the field. Due to the change in the value of flux linking with the search coil an
emf is induced in it. This emf sends a current through the flux meter which deflects through
anangle depending upon the change in the value of flux linkages. The instrument coil deflects
during the period the flux linkages change but as soon as the change ceases the coil stops, owing
to high electromagnetic damping in the coil circuit. This high electromagnetic damping is
obtained by having a low resistance of the circuit comprising the flux meter and the search coil.
Measurement of flux Density with Gauss Meter: The general electric Company makes a small
and simple "Gauss meter". (This meter may safely be called a tesla meter now as the unit of
flux density prevalent these days is tesla i.e. Wb/ma). It can be usefully employed in testing field
strength in places such as the gap between the poles of a magnet.
It consists of a tiny permanent magnet mounted at the end of a long, thin shaft (Fig) which is
supported and protected by a bronze tube. The assembly is used as a probe and has an outside
diameter less than one-tenth of an inch. The head of the meter consists of a bearing, scale,
pointer, ,and spring and has no electric or magnetic elements in it. The small probe magnet tends
to align itself with any magnetic field around it. The person using the meter turns the head until
the maximum indication is given on the scale; this gives the strength of the field. This meter is
made possible by a new magnetic material called Silmanal, which has a very high coercive force
(about ten times that of Alnieo). This instrument has an accuracy of i per cent when special
calibration methods are used, but commercial tolerances are of the order of 5 per cent.