ReleaseNote OmniPC 4.1
ReleaseNote OmniPC 4.1
1 Release Notes
New features
- New features were limited to the OmniScan unit acquisition software (MXU) for this release.
New features
- Support of the files generated with the 8ch UT module.
- Support of the RollerFORM Phased Array wheel probe.
New features
- Support of the files generated with a PA2 32:128 module.
- AdBr reading functionality.
o AdBr gives the difference in dB between the signal maximum amplitude in Gate A and the
Reference level.
o Once the Reference Gain is set, the adjustment gain will not impact the AdBr value. Example:
Gain = 20.0dB, A% = 40%, AdBr = -6dB
Set the Reference Gain and add 6dB of adjustment gain
Gain = 20.0dB + 6.0dB, A% = 80%, AdBr = -6dB
o A%r gives the % equivalent to the AdBr reading.
o The signal amplitude has to be lower than 100% for the reading to return a value.
o The same readings (AdBr and A%r) exist for Gate B.
OmniPC 4.1R7
New features
- Improvements to the Technique function
o The parameters defined by this function allow a better usability
- The technique “Uncorrected” was added to the list of existing techniques
o This button allows the S-Scan to be displayed in the form of stacked A-Scans
o The only layouts available are: A-S Scan and A-C-S Scan
- The A-Scan readings are displayed by default with TOFD configuration. This allows monitoring the
amplitude of the signal at the cursor’s position.
OmniPC 4.1R6
New features
- The TOFD calibration shortcut is now available at all time through the interactive menu
o The user no longer has to go through the Wizard first
- Improvements to the different TOFD calibrations
o Better LW synchronisation
o The cursors position remains the one defined by the user prior to the Wizard
OmniPC 4.1R3
New features
The A-Scan was added to the Multiple PA C-S layout, which now becomes PA A-C-S
OmniPC 4.1R2
New features
- New report templates to reflect the nature of the application
- High quality sample data files are provided through the data file manager
- Interactive menus giving direct access to parameters that are linked to the click area
- Improved TOFD calibration, and Lateral Wave Synchronization
- 0% reference line is added to the TOFD A-Scan
- New Interactive title bar for direct access to the most common features
- The axis zoom bar is strictly used to Pan
o Double-tap to zoom out one axis
o Tap and hold, then select Zoom out for all axis at the same time
- Cursors are moved by selecting it, and then scrolling the rotary knob
- Gates are adjusted by selecting a zone of the gate, and then scrolling the rotary knob
o Left end = Start
o Middle = Threshold
o Right end = Width
- Additional layouts from MXU 4.0