0% found this document useful (0 votes)
89 views

A 1069 Technical Specification

The document provides the technical specifications for the A1069 current clamp from Metrel, including: 1) It has a dielectric strength of 3000 Volts AC for 1 minute, is for indoor use only below 2000m altitude, and operates from -10°C to +50°C temperature. 2) Measurement features include a nominal current of 100 Amps, maximum of 200 Amps, minimum of 50mA, with 1V output equal to 100 Amps and accuracy of 1% up to 2000Hz frequencies. 3) A graph shows the typical accuracy is within 1% for currents between 5-85 Amps and phase error is less than 3 degrees for currents above 10 Amps.

Uploaded by

ruben conde
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
89 views

A 1069 Technical Specification

The document provides the technical specifications for the A1069 current clamp from Metrel, including: 1) It has a dielectric strength of 3000 Volts AC for 1 minute, is for indoor use only below 2000m altitude, and operates from -10°C to +50°C temperature. 2) Measurement features include a nominal current of 100 Amps, maximum of 200 Amps, minimum of 50mA, with 1V output equal to 100 Amps and accuracy of 1% up to 2000Hz frequencies. 3) A graph shows the typical accuracy is within 1% for currents between 5-85 Amps and phase error is less than 3 degrees for currents above 10 Amps.

Uploaded by

ruben conde
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

METREL d.d.

Ljubljanska cesta 77, 1354 HORJUL


SLOVENIJA
Measuring and Regulation Equipment Manufacturer Phone.: +386 (1) 7558-200
Fax.: +386 (1) 7549-095, 7549-226
http://www.metrel.si Swift Code: LJBASI2X 02923-0018420925
E-mail: [email protected] VAT No: SI 56387512

A1069 - TECHNICAL SPECIFICATION


General:
Dielectric Strength 3000 Volt 50Hz for 1 minute
Environment Indoor use only, altitude less than 2000m
Temperature Range -10°C to +50°C (-20°C to +70°C storage)
Humidity Max 80% RH at 30°C, reducing to 40% RH at 50°C
Safety Class Class III per EN61010-1
Max Working Voltage 30 Volt rms Bare Conductor
600 Volt rms Insulated Conductor
Protection Degree IP20
Jaw aperture 15 x 17 mm
Dimensions 90 x 43 x 23 mm
Weight 115 grams
Output Terminals 4 mm Safety Sockets
Clamps output impedance 5

Measurement:
Nominal Input Current INom 100 Amp
Maximum Input Current IMax 200 Amp
Minimal Current IMin 50mA
Output Signal UOut 1 V = 100 A
Typical accuracy err 1%
Phase Error ( I > 10 A ) Φerr < 3°
Frequency Range (err < 3%) f3% 40 to 2000Hz

Typical Measurement Results (Measured at 50 Hz)

3,5
3
2,5
2
1,5
Ierr (% of m.v.)
1
Err (º)
0,5
0
-0,5 5 25 45 65 85
-1
-1,5
Current (A)

You might also like