Instruction Manual
Instruction Manual
Confidential Information
09.01.2007 Z
LEGAL WARNING
RALCO SRL WILL PURSUE ANY AND ALL LEGAL REMEDIES AVAILABLE UNDER
THE LAW TO RECOVER DAMAGES INCURRED BASED UPON THE ABOVE
VIOLATION(S).
Model Certifications
Revision Level:
The English version of this manual is a translation of the Italian text which
prevails in case of doubts.
Rev. N Date N
MRM/009/08
A 08.04.2008
MRM/010/08
B 10.04.2008 MRM/011/08
C 06.11.2008 MRD/015/08
D 09.01.2009 MRE/01/09
E 11.12.2009 MRE/46/09
F 17.12.2009 MRE/47/09
G 19.01.2010 VVS/04/10
H 28.01.2010 VVS/08/10
I 09.03.2010 MRE/09/10
J 22.03.2010 MRD/003/10
K 08.04.2010 MRE/13/10
L 10.05.2010 VVS/29/10
M 31.05.2010 VVS/36/10
N 23.09.2010 MRD/011/10
VVS/007/11
O 11.02.2011
VVS/008/11
P 21.04.2011 MRD/005/11
Q 23.06.2011 VVS/024/11
14.07.2011 MRD/013/11
R
24.10.2011 MRD/018/11
S 09.03.2012 MRD/006/12
T 15.05.2012 VVS/15/2012
U 14.11.2012 MRD/012/12
V 12.03.2013 MRM/031/13
28.02.2013 MRE/012/13
W 16.01.2014 MRD/001/14
29.01.2014 MRE/006/14
24.03.2014 MRE/015/14
10.09.2014 ACD/020/14
X
10.09.2014 ACD/026/14
13.11.2014 MRE/041/14
Y 19.10.2015 VVS/028/15
Z 15.09.2016 MRD/031/16
THIS INSTRUCTION MANUAL PROVIDES THE SPECIFICATIONS, DIMENSIONS, AND FUNCTIONS FOR A STANDARD
COLLIMATOR. PERSONALIZATIONS ARE AVAILABLE UPON REQUEST. WITH PERSONALIZED COLLIMATORS, THE CUSTOMER
MUST ENSURE THE FOLLOWING HAS BEEN PROVIDED:
• YOUR SPECIFIC CODE IS CLEARLY VISIBLE ON THE BACK LABEL AFTER “CUSTOMIZATION”, ABOVE THE SERIAL
NUMBER, TO ENSURE THE CORRECT PERSONALIZED COLLIMATOR HAS BEEN PROVIDED (SEE ABOVE).
• THE CHAPTER ENTITLED “PERSONALIZATIONS,” WHICH INCLUDES RELEVANT INFORMATION REGARDING YOUR
PERSONALIZATION.
• OTHER PROVIDED ATTACHMENTS WHICH REFER TO YOUR PERSONALIZATIONS.
CONTENTS
THE COLLIMATOR ELECTRONICS SUPPLIES A SIGNAL DEFINED AS READY. THE X-RAY EQUIPMENT MANUFACTURER
WILL MANAGE THIS SIGNAL AS THE X-RAY CONSENT. IN THE EVENT OF A COLLIMATOR FAILURE, THE LIABILITY
FOR THE CORRECT BYPASS OF THIS SIGNAL, IN ORDER TO ALWAYS ENSURE THE X-RAY CONSENT, DEVOLVES
UPON THE X-RAY EQUIPMENT MANUFACTURER.
REPORTING
R302DACS/A - R302DACS/A DHHS - Advisory
INFORMATION REGARDING ACCIDENTS THAT HAVE OCCURRED WHILE USING THIS COLLIMATOR MUST BE REPORTED
IMMEDIATELY TO RALCO, SRL.
LIMITATION OF LIABILITY
RALCO IS NOT LIABLE IF THE PROVIDED INSTRUCTIONS ARE NOT COMPLIED WITH. FURTHERMORE, RALCO
IS NOT LIABLE IF ONE OR SEVERAL OF THE FOLLOWING INSTANCES APPLY:
IF THE UNIT IS SPECIFICALLY DESIGNED PER CLIENT SPECIFICATIONS AND THE CERTIFICATION WAS THE DUTY
OF THE CLIENT;
THE COLLIMATOR DESCRIBED HERE, IS USED ON RADIOLOGICAL SYSTEMS AND IS CLASSIFIED AS A TYPE IIB
ACCORDING TO ATTACHMENT IX. THE COLLIMATOR HAS BEEN DESIGNED AND MANUFACTURED IN COMPLIANCE
TO ATTACHMENT II OF LEGISLATIVE DECREE, 2 FEBRUARY 1997, N. 46, IMPLEMENTATION OF DIRECTIVE
93/42/CEE OF 14 JUNE 1993 AND SUCCESSIVE MODIFICATIONS APPLYING DIRECTIVE 2007/47/CE.
LIST APPLICABLE STANDARDS AND TO WHICH RALCO ADHERES TO:
EN 60601-1:2006
EN 60601-1-2:2007
EN 60601-1-3:2008
EN 60601-2-45:2012
EN 60601-2-54:2009
EN 60825-1:2007
ISO 9001:2008
UNI ESO 13485:2012
21 CFR SUBCHAPTER J (FDA)
8750 01 & 8750 81 (CSA)
CE
*RALCO CAN PROVIDE DOCUMENTATION REGARDING ITS ADHERENCE TO ANY OF THE ABOVE STANDARDS.
EN 60601-1 par. 5
• PROTECTION AGAINST ELECTRIC HAZARDS: "CLASS I" EQUIPMENT.
• PROTECTION AGAINST DIRECT AND INDIRECT CONTACTS: TYPE B EQUIPMENT WITH APPLIED PARTS.
• PROTECTION AGAINST WATER SEE PAGE: "COMMON EQUIPMENT".
• SAFETY OF OPERATION IN THE PRESENCE OF INFLAMMABLE ANAESTHETICS WITH AIR OR OXYGEN OR NITROUS
OXIDE: EQUIPMENT NOT SUITED TO APPLICATION IN THE PRESENCE OF INFLAMMABLE ANAESTHETIC MIXTURES
CONTAINING AIR OR OXYGEN OR NITROUS OXIDE.
• OPERATION CONDITIONS: EQUIPMENT FOR CONTINUOUS OPERATION AT INTERMITTENT LOADS - SEE CHAPTER
OPERATION INSTRUCTIONS.
R302DACS/A - R302DACS/A DHHS - Advisory
• SHOULD LABEL DATA ON THE COLLIMATOR NOT CORRESPOND TO THE SPECIFICATIONS HEREIN, INFORM RALCO OF
THE NON CONFORMITY.
• VERIFICATIONS OF THE SPECIFICATIONS ARE TO BE PERFORMED ACCORDING TO THE INDICATED
EQUIPMENT STANDARDS.
THE INHERENT RISK OF USING COLLIMATORS IN RADIOLOGICAL SYSTEMS IS DEEMED REASONABLE AS DETERMINED
BY APPLICABLE STANDARDS. THE USE OF COLLIMATORS FOR RADIOLOGICAL SYSTEMS IS STRICTLY REGULATED.
RALCO FOLLOWS ALL APPLICABLE STANDARDS. IT IS UP TO THE INSTALLER AND OPERATOR TO ENSURE THAT
ALL POSSIBLE STEPS ARE FOLLOWED TO ENSURE THE HEALTH AND SAFETY OF THE PATIENT AND OPERATOR.
THE CONTACT WITH THE MOVING PARTS CAN BE CONSIDERED IMPOSSIBLE AND MAY OCCUR ONLY DURING
INSTALLATION OR MAINTENANCE (LOW HAZARD).THE INTERNAL PARTS ARE PROTECTED BY A MECHANICAL SAFEGUARD
IN THE FORM OF A METAL OR PLASTIC SHEATH WHICH IMPEDES ACCESS. ALSO MOTOR UNITS ARE NOT ACCESSIBLE
AS KNOBS ARE FIXED USING THE ALLEN SCREWS AND CAN BE REMOVED ONLY BY USING A TOOL, IN ACCORDANCE
WITH PARAGRAPH 5.92.3 OF EN 60601-1 AND EN 60601-2-54 PAR. 201.9.2.2.5. IN ADDITION THE COLLIMATOR
HAS BEEN TESTED IN ACCORDANCE WITH EN 60601-1, PARAGRAPH 4.8.2
SYMBOLS
IN THIS MANUAL THE FOLLOWING SYMBOLS ARE USED:
THE TRIANGLE EMPHASIZES WARNINGS AND CAUTIONARY MESSAGES WHICH ARE IMPORTANT FOR YOUR SAFETY
AND/OR EFFICIENT OPERATION OF THE EQUIPMENT. THESE WARNINGS MUST BE STRICTLY COMPLIED WITH.
INFORMATION PROVIDED WITH THE LAMP IS ADDITIONAL ADVICE FOR THE PROPER CONTROL AND PRACTICAL
USE OF THE UNIT.
RALCO INSTRUCTION MANUALS ARE AVAILABLE IN TWO VERSIONS: ITALIAN AND ENGLISH.
THE ENGLISH VERSION OF THIS MANUAL IS A TRANSLATION OF THE ITALIAN TEXT WHICH PREVAILS IN CASE OF DOUBTS.
R302DACS/A - R302DACS/A DHHS - Advisory
417-iec
Alternating Current 01-14
503
417-iec
Direct Current 01-18
5031
417-iec
Continuous and Alternating Current 01-19
5033
417-iec
Protective Ground 01-20
5019
417-iec
Plus; Positive Polarity 01-27
5019
417-iec
Minus; Negative Polarity 01-28
5006
417-iec
Input 01-36
5006
417-iec
Output 01-37
5034
417-iec
Iris Diaphragm: Closed 01-70
5324
417-iec
Radiation Filter or Filtration 04-51
5381
417-iec
Light Indicator of Radiation Field 04-51
5381
417-iec
Beam Limiting Device: Open 04-55
5385
417-iec
Beam Limiting Device: Closed 04-56
5386
Electrostatic-Sensitive Device
A Open Cross
B Open Long
C Closed Cross
D Closed Long
E Filter (optional)
R302DACS/A - R302DACS/A DHHS - Description
F Light
RO DESCRIPTION
Final control testing (light intensity, light field/radiation field alignment, radiation
RO 063
leakage, light field border contrast ratio, general funcitonality and electronic)
RO 240 Focus-skin spacers, preventing X-ray exposure too close to the patient
RO 271 Halogen lamp 150 W for high luminosity with cooling system
Auto-centering top plate with coupling for plastic flange allowing rotation of the
RO 318
collimator, the plastic flange is included
RO 323 Additional variable filtration through both automatic and manual selection
Plastic flange with a notch rotating +/- 50°, 20 mm thickness without stop
RO 441
(available only for collimators assembled with RO 318)
Plastic flange with a notch rotating +/- 90°, 20 mm thickness without stop
RO 442
(available only for collimators assembled with RO 318)
RO 444 Software that controls the collimator functions via CANopen protocol
Specifications of the basic version of the collimator are described here below.
24 V AC - 6 A 50/60 Hz
Power Supply 24 V DC
Operation Environment
10°- 40°C
• Ambient Temperature
0%- 75%
• Relative Humidity
700 - 1060 hPa
• Atmospheric Pressure
Storage
-40° - 70°C
• Temperature
10% - 95%
• Humidity
500 - 1060 hPa
• Atmospheric Pressure
Weight 10.5 Kg
L: 268 mm
Dimensions W: 195.5 mm
H: 206.5 mm
R302DACS/A - R302DACS/A DHHS - Specifications
• Use the X-ray tube housing literature to determine the distance from the focal spot to the
tube port face.
• Subtract the resulting distance from 80 mm (3.14") and determine the number of 1.5 mm
(0.06") spacers that, combined with the thickness of the mounting flange, will make up the
difference: (18 mm) Allowable tolerance is 1 mm. (0.04").
• Different flange models with their dimensions are listed in the Chapter DESCRIPTION, para-
graph Optional Items.
IMPORTANT
TO SAFEGUARD THE OPERATOR AND PATIENT AGAINST THE HAZARD OF
A FALLING COLLIMATOR, THE FOLLOWING INDICATIONS MUST BE RESPECTED.
• Unscrew the four mounting and centering adjustment Allen screws on the collimator until
the four tabs are withdrawn from the collimator top opening, see Fig. Mounting Flange, in
this Chapter.
MAKE SURE TO TIGHTEN THE ALLEN SCREWS SECURING THE CONTROL TABS.
APPROPRIATE TIGHTENING OF THE 4 ALLEN SCREWS ENSURES SECURE MOUNTING
OF THE COLLIMATOR. TIGHTENING FORCE USED MUST BE GREATER THAN 0.50 NM.
R302DACS/A - R302DACS/A DHHS - Installation
• Check to see that the distance between the collimator housing and the mounting flange is
equal in all directions and that the collimator face is parallel to the axis of the table.
• Loosen the screws and adjust if required.
THE WIRING DIAGRAM INCLUDED IN THIS CHAPTER REFERS TO THE STANDARD PRODUCT.
IT IS THE RESPONSIBILITY OF THE CUSTOMER WHO HAS REQUESTED AN ELECTRIC
CUSTOMIZATION, TO ENSURE THAT AN ELECTRIC DIAGRAM RELATING TO THE CUSTOMIZATION
HAS BEEN PROVIDED WITH THE DOCUMENTATION.
THE DEVICE MUST BE EXCLUSIVELY CONNECTED TO POWER NETWORK, WITH EARTH PROTEC-
TION, IN ORDER TO AVOID A RISK OF ELECTRICAL SHOCK.
Supply and signals to the collimator must be to 2007/47/CE standards. Devices that supply the
collimator must therefore feature double or reinforced insulation as provided by the General
Standard on Electro-medical Equipment CEI62-5 +A2 (EN 60601-1).
The device is electrostatic sensitive, consequently all the relating safety standards must
be complied with.
CABLES AND TERMINALS USED FOR THE INTERNAL CONNECTION OF THE COLLIMATOR MUST
BE SUITABLE FOR OPERATION AT TEMPERATURES OF 70°C AND COLLIMATOR CURRENT
ABSORPTION.
R302DACS/A - R302DACS/A DHHS - Installation
WARNING:
THE SYSTEM MUST BE SUPPLIED AT 24 V AC +/-10%.
VOLTAGE IN EXCESS WILL DAMAGE THE STEPPER MOTOR CONTROL ELECTRONICS.
• Use cable RO 146 to connect the collimator. The AMP 201298-1 at one end is to be
plugged into the connector at the back of the collimator, see Fig. RO 146 in this chapter.
• The end of the cable with the DB9 connector and three wires (yellow, purple, yellow-green)
is to be connected to the CanBus interface and supply.
DB9 Pin 1
DB9 Pin 4
DB9 Pin 5
DB9 Pin 6
DB9 Pin 8
DB9 Pin 9
• The SID in use is too short for the cassette in use; max. open-
ing of shutters: 43x43 cm at 90 cm - i.e. the shutters are at
Red LED end run but the motors are still running to attain a correct
Exposure HOLD position.
• Increase the SID.
• Motors move to their correct position.
Increase SID • The collimator will automatically adjust the field, etc.
Green LED • The system is ready.
READY • All pre-set functions are allowed.
• Cassette is not inserted.
• Cassette tray not inserted correctly.
• The system has no SID - or SID signal below or over the min-
Yellow LED
imum and maximum set value.
MANUAL
• Collimator inclination exceeds 3°.
• Key switch (back of collimator) is turned to the Manual Mode.
• Manual Mode selected.
• The collimator will automatically adjust the field to the size of
Insert a cassette
the cassette inserted.
• The collimator will automatically read SID and cassette size
inserted.
Readings • The two knobs on the front panel reduce field size. Note:
R302DACS/A - R302DACS/A DHHS - Operation Instructions
(Cassette sizes, FFD shutters cannot be opened to a field larger than that of the
(SID): cassette size inserted.
If the system is in the manual mode, the display will show the last
memorised SID and the last memorised field size.
Prior to operating the collimator, check the settings related to correct X-ray beam functions.
Manual Mode
The collimator is in the manual mode if:
• The collimator is inclined at an angle exceeding 3° and less than 87°.
• Manual control is selected.
• The cassette-IN microswitch is open.
• The Bucky provides no information regarding cassette format.
• No SID signal is present.
When the system is in the Manual Mode, the yellow LED on the front panel is lit and the output
contact is activated.
The X-ray contact is always closed.
The operator can:
• Turn the light field ON by pressing the related button on the front panel.
• Reduce field directly with the knobs or push-buttons related to shutter opening.
Check on opening with the light field.
The collimator is normally operated by activating the push-button on the front panel to switch
the light field ON.
R302DACS/A - R302DACS/A DHHS - Operation Instructions
IF THE DISTANCE OF 100 CM (40") SID CANNOT BE DETERMINED, USE THE SID
VALUE CLOSEST TO 100 CM (40") AND CALCULATE THE MEASUREMENT TOLERANCES
AS THE APPROPRIATE PERCENTAGES OF THE DISTANCE.
IF THE X-RAY BEAM CANNOT BE POSITIONED VERTICALLY USE CLAMPS, MASKING TAPE,
OR OTHER MATERIAL AS REQUIRED TO POSITION THE X-RAY TUBE AND IMAGE RECEPTOR.
THE X-RAY TUBE AND THE IMAGE RECEPTOR MUST BE SET OUT AT THE SPECIFIED
SID DISTANCE PERPENDICULARLY TO THE X-RAY TUBE AS DESCRIBED IN THE FOLLOWING
PROCEDURE.
NOTE
A DIGITAL SYSTEM.
THE HEEL EFFECT WILL CAUSE THE FIELD TOWARD THE CATHODE TO BE SLIGHTLY
LESS SHARP THAN ON THE OTHER THREE SIDES. THIS IS NORMAL AND CANNOT BE
CORRECTED BY ADJUSTMENT. IN ADDITION, AN X-RAY TUBE OF 12° OR LESS
TARGET ANGLE WILL PRODUCE AN ASYMMETRICALLY SHAPED FIELD WHEN A LARGE
FIELD SIZE IS USED AT SHORT SID, BECAUSE OF ANODE CUT-OFF EFFECT. THIS IS
NORMAL AND MAY NOT BE CORRECTED BY ADJUSTMENT.
• Place the test film on the face of the cassette over the white paper or repeat exposure on
the flat panel.
• Place the cassette in the position originally marked.
• Check the correct position of the film or the X-ray field on the flat panel by the shadows
cast by the copper strips.
• Using the images of the collimator shutters as the references for the shape and size of the
X-ray field, adjust the light field to match.
• If adjustment is necessary, adjust the travel of the light source. All procedures can be found
in LIGHT FIELD ADJUSTMENTS in this chapter.
• Remove the part of the cover necessary to access the screws, see Chapter - COVER
REMOVAL
• Remove the light source protection heatsink by unscrewing the fixing screws . This allows
you to access the light source.
• If the light-field needs to be moved laterally, loosen (not remove) the fixing screws A.
• Adjust through screw B.
• When calibration is terminated, lock the screws A.
See the following Fig. Light Field Adjustment
Vertical Alignment
• If the light field is smaller than the X-ray field, lower the lamp by adjusting screw C.
• If the light field is bigger than the X-ray field, raise the lamp by adjusting screw C.
See Fig. Light Field Adjustment.
via CanBus interface including power-up and power down data for possible accessories.
Description of Connections
J1: DC Supply for ionization chamber (+12 V 1 A)
J2: DC Supply for board ASR002 (+31 V 1.5 A)
J3: Terminals for AC supply of the board (from 24 to 32 V AC/DC) and for connection to the
lamp (24 V AC/DC - max 150 V A)
J4: 10-pin programming connector
J8: Connector for laser and fan outputs, cycle start input and temperature sensor
J9: CanBus connector, + 24VDC supply
D7: Red LED, lights up briefly at each CanBus control
D8: Green LED, cyclic flickering signifies correct board operation
W1...W4: Programming jumpers to preset lamp On and ramp times
W1 and W2 Serve to program light source ON time:
W3 and W4 Serve to program ramp time:
Power supply
closed, closed 5 closed, closed 1
closed, open 30 closed, open 2
open, closed 45 open, closed 3
open, open 60 open, open 4
+24 V DC
Ramp Time (s)
W1, W2 Time W3, W4
Power supply
closed, open 30 open, open 2
collimatori\acs\scheda_1.jpg
R302DACS/A - R302DACS/A DHHS - Electronic System
Supply M1 - M3
Earth M2
Iris Photocell J1
CanBus J2
Filter Photocell J3
Cross Motor J4
Long Pot J5
Cross Pot J6
Key Terminal J7
Long Motor J8
CanBus J12
Serial J13
Programming Connector J9
Serial Controls
The collimator accepts the following serial controls:
C0 L0
SID 0
English language:
C0 L0
SID 0
Following is the readout in the case of inches rather than centimetres:
C 0.0 L 0.0
SID 0.0
Filtering level will be shown if a four-position radiological filter is in place:
C 0 L0
SID 0 0Al
C0 L0
-90°SID 0 0Al
R302DACS/A - R302DACS/A DHHS - System Start-Up
b7 b6 b5 b4 b3 b2 b1 b0
Where b7 is the most significant (MSB) bit and b0 the least significant (LSB).
QUICK START
SID, Cross, Long Values
SID, Cross and Long values are set by sending message 7A0 via CanBus.
Regulate:
• SID at 100 cm
• Cross at 430 mm
• Long at 130 mm
A message of the following type must be sent via CanBus:
ID DLC D0 D1 D2 D3 D4 D5 D6 D7
7A0 8 40 00 00 64 01 AE 00 82
length automatic SID values set at 100 Cross value set at 430 Long value set at 130
mode cm mm mm
ID DLC D0 D1 D2 D3 D4 D5 D6 D7
7A0 8 40 00 00 64 01 AE 00 82
length automatic SID values set at 100 Cross value set at 430 Long value set at 130
mode cm mm mm
A message of the following type must be sent via CanBus to enable the manual mode:
ID DLC D0 D1 D2 D3 D4 D5 D6 D7
7A0 8 80 00 00 64 01 AE 00 82
length manual SID values set at 100 Cross value set at 430 Long value set at 130
mode cm mm mm
Status Messages
The status messages feature must be activated to received collimator status data. To do this,
set the time value in milliseconds of the message return frequency.
Status messages are activated with two messages:
• 7A7 - enables message transmission; 7F0 (SID, Cross, Long status and position); 7F1 (Fil-
ter and Flag)
To enable status messages 7F0 and 7F1 with a frequency of 100 ms.
ID DLC D0 D1 D2 D3 D4 D5 D6 D7
7A7 8 00 64 00 00 00 00 00 00
length Time in milliseconds
R302DACS/A - R302DACS/A DHHS - CanBus Messages
Cross/Long
Once a movement message is sent, the status messages provide data related to collimator
movement, current position and possible movement errors.
Status message 7F0 give the current Cross/Long value.
ID D0 D1 D2 D3 D4 D5 D6 D7
7F0 FLAGS xx yy zz CrossH CrossL LongH LongL
Byte Cross position Byte Long position
DLC=8:
D0
b5 If 1: Cross closed
b4 If 1: Cross open
b3 If 1: Long closed
b2 If 1: Long open
b1 If 1: Iris closed
b0 If 1: Iris open
D1
b6 Not used
b5 ATS version:
If 1: Inserts filter position 1; if used with bit 7, inserts the filter position 3
b4 Not used
b3 Not used
b2 Not used
Byte D2 is used exclusively to communicate with the ATS control panel. If the vertical
receptor has been given a different designation, D2 will always be 0.
D2
b7 Iris Freeze
b6 Not used
b5 Not used
b4 Not used
b3 If 1: System is aligned with the colimator to perform right lateral rays (-90 ) °
b2 °
R302DACS/A - R302DACS/A DHHS - CanBus Messages
If 1: System is aglined with the collimator to perform upward vertical rays (180 ); not used in this application
b1 If 1 : System is aligned with the collimator to perform left lateral rays (+90)
D3 ...D7:
D3 Vertical SID in cm
0x7A1
ID Type DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7A1 Input per ASR002 8
0x7A1
0x7A2
R302DACS/A - R302DACS/A DHHS - CanBus Messages
SID Measurement
ID Type DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7A2 Input per ASR002 8
0x7A2
0x7A3
ID Type DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7A3 Input per ASR002 8
0x7A4,
These messages are not used.
0x7A5
Long and Cross Shutter Values.
ID Type DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7A5 Input per ASR002 8
D0 = 00h Not Flash stored (the frequency values will be used until ASR002 powers down).
D0 = 01h Flash stores (the frequency values will be used at subsequent ASR002 power-ups).
D1, D2 Show a frequency value for a manual shutter movement, the valid values are from 100
to 5000 Hz (0064 h...1388 h).
D3, D4 show a value of factor that multiplies frequency of shutters automatic movement, that
R302DACS/A - R302DACS/A DHHS - CanBus Messages
0x7A6
Software version readout.
ID Type DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7A6 Input per ASR002 8
ID Type DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7A7 Input per ASR002 8
0x7A7
If set time is shorter than 100 msec, the messages will be sent once only.
If the value is the same or greater, the messages will be repeated at a value equal to the set
time value.
In the case of an ATS receptor, default time of the message 7F0 is set at 200 ms. In all other
cases the messages are disabled at power-up.
0x7A8
0x7A8
If the set time is shorter than 100 msec, the messages will be sent once only. If the value is
the same or greater, the messages will be repeated at a value equal to the set time value.
Messages 0x7F8, 0x7FA and 0x7FB are not implemented in this software version.
0x7A9
Set the Light Source Time
ID Type DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7A9 Input per ASR002 8
0x7A9
0x7AA
Switches the system in configuration mode.
ID Type DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7AA Input per ASR002 8
Output per ASR002
D0
b7 Not used
b6 Not used
b5 Not used
b4 Not used
b3 Not used
b2 Not used
b1 Not used
b0 Not used
D1
b6 Not used
b5 Not used
b4 Not used
b3 Not used
b2 Not used
b1 Not used
b0 Not used
D2
b7 Not used
b6 If 1: Cross closed
b4 If 1: Long closed
b3 If 1: Long open
D3
b2 Not used
b0 Not used
D4
b7 Not used
b6 If 1: Inclinometer excluded
b5 Not used
b0 Not used
D5
b7 Not used
b6 Not used
b5 Not used
b1 Not used
b0 Not used
D6
b7 Not used
b6 Not used
b5 Not used
b3 Not used
b2 Not used
b1 Not used
b0 Not used
D7
b6 Not used
b5 Not used
b2 Not used
b1 Not used
R302DACS/A - R302DACS/A DHHS - CanBus Messages
b0 Not used
STATUS Messages
0x7F0
Collimator responds:
ID Type DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7F0 Output per ASR002 8
D0
b6 If 1: Lamp is ON
b5 Not used
b4 Not used
b3 Not used
b2 Not used
b1 Not used
D1
b7 Not used
b6 Not used
b5 Not used
b4 Not used
b3 Not used
b2 Not used
0x7F0
D1 Not used
D2 Not used
R302DACS/A - R302DACS/A DHHS - CanBus Messages
D3 Vertical SID in cm
0x7F1
Collimator responds:
ID Type DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7F1 Output per ASR002 8
0x7F1
D2 Not used
D3 Current filter selection (0: 0 Al, 1: 1 Al 0.1 Cu, 2: 1 Al 0.2 Cu, 3: 1 Al)
D6 Not used
D7 Not used
0x7F2
Indicates the serial number.
ID Type DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7F2 Input per ASR002 8
0x7F2
D0 ‘S’
D1 ‘/’
D2 ‘N’
0x7F8
Not used; transmission time may be used for future applications.
0x7F9
Indicates the light source status.
ID Tipo/Type DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7F9 Output per ASR002 8
or ASR003
D0
b6 Not used
b5 If 1: Incates that collimator is in the automatic or manual mode (red LED off)
b4 Not used
b3 Not used
b2 Not used
b1 Not used
b0 If 1: Light ON
D1
b7 Not used
b6 Not used
b5 Not used
b4 Not used
D2
b6 Not used
R302DACS/A - R302DACS/A DHHS - CanBus Messages
b3 If 1: Filter present
b2 If 1: Iris present
b1 If 1: Long present
b0 If 1: Cross pressent
D3
b7 Not used
b6 Not used
b5 Not used
b1 If 1: Vertical Bucky
D4
b7 Not used
b5 Not used
D5 Not used
D6
b7 Not used
b6 Not used
R302DACS/A - R302DACS/A DHHS - CanBus Messages
b5 Not used
b4 Not used
b3 Not used
D7 Not used
The above HVL requirements can be met if it is demonstrated that the aluminium equivalent in
the primary beam is not less than that shown in the following table:
30 0.3
Below 50 40 0.4
50 0.5
51 1.2
From 51 a70 60 1.3
70 1.5
71 2.1
80 2.3
R302DACS/A - R302DACS/A DHHS - Compliance Verification
90 2.5
100 2.7
Above 71 110 3.0
120 3.2
130 3.5
140 3.8
150 4.1
The information contained in the above table was extracted from the Code of Federal
Regulations FDA 21 1020.30 (m).
Type 100 Aluminium Alloy (as given in “ALUMINUM STANDARDS AND DATA” verification of
compliance).
Below 50 0.5
From 51 to 70 1.5
Over 70 2.2
The Aluminium equivalence of each component in the primary beam (X-ray tube and housing,
beam limiting device and any additional filtration in the system) is specified on the component,
in the technical data attached to the component or can be measured. Determine the total
aluminium equivalence in the primary beam and make sure that it is equal or greater than those
specified in the above Table Total Filtration of Primary Beam in Aluminium Equivalence.
the collimator. Make an exposure using the same technique factors; record the reading.
c) Verify that the radiation read with the 2.5 mm Al in the beam is greater or equal to 50% of
the radiation read with no filtration in the beam.
Standard Absorber Method
The HVL determination obtained from the following procedures are to be compared with those
illustrated in the Table - Minimum Filtration Requirement - Beam Quality (HVL). The HVL
in millimetres of aluminium obtained during the test must be greater or equal than the values
listed in the above mentioned table.
a) Direct the central X-ray beam perpendicular and in the center of a RAD-Check instrument.
Determine the exact distance from the X-ray tube focal spot to the window of the collimator
(273 mm - 10.75"). Place the input area of the RAD-CHECK at an equal distance from the
collimator window. Collimate the beam to an area slightly larger than the detector.
b) Select a tube potential of 100 kVp and appropriate mA and seconds, with no added filtration
in the beam make an exposure and record the reading. Using a set of several sheets of 1100
Aluminium Alloy, each having a thickness of 0.5 or 1.0 mm, tape the filtration to the window
of the collimator. Make an exposure for each increments of filtration and record the reading.
generale\12_verif_conf_graph.gif
CROSSHAIR ALIGNMENT
CHAPTER- ADJUSTMENTS, PARAGRAPH CROSSHAIR ALIGNMENT.
EMC COMPLIANCE
Electromagnetic Emissions
The R108F DHHS colimator is suitable for use in the specified electromagnetic environment. The purchaser or user of the
R108F DHHS should assure that it is used in an electromagnetic environment as described below:
RF emissions Class [A] This R108F DHHS is suitable for use in all establishments other
CISPR 11 than domestic and those directly connected to the low voltage
power supply network which supplies buildings used for domestic
Harmonic emissions [Not applicable] purposes.
IEC 61000-3-2 Warning
This equipment is intended for use by healthcare
Voltage fluctuations/flicker [Not applicable]
professionals only.
emissions This equipment may cause radio interference or may
R302DACS/A - R302DACS/A DHHS - Compliance Verification
CISPR 14 N.A. This Collimator is not suitable for interconnection with other
equipment.
CISPR 15 N.A. This Collimator is not suitable for interconnection with other
equipment.
The R108F DHHS collimator is intended for use in the electromagnetic environment specified below. The customer or the
user of the collimator should assure that it is used in such an environment.
Collimator R108F DHHS is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the collimator can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the collimator as recommended below,
according to the maximum output power of the communications equipment.
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
R302DACS/A - R302DACS/A DHHS - Compliance Verification
(1).
(2).
(3).
PHOTOGRAPH.
DISCONNECT THE
CONNECTOR.
(5).
IN COLLIMATORS WITH
MEASURING TAPES, REMOVE
THE TAPE STOP BY LIFTING IT
OFF WITH A SCREW DRIVER.
GENTLY EASE THE TAPE INTO
ITS CONTAINER WITHIN THE
COLLIMATOR.
(6).
(8).
(9).
ACCESS THE COLLIMATOR BOARD ASR001, TAPE MEASURE AND LASER BY REMOVING:
• the knobs
• the collimator front panel
• ASR002 board
inconveniences:
• Breakage of the tape or,
• Distortion of the tape or,
• Impossibility of retracting the tape into its lodging because the grip of the
spring has been forced and hook-up is consequently distorted.
NOTE: DO NOT EXTEND THE TAPE BEYOND THE INDICATION OF STOP.
ACCESS THE COLLIMATOR BOARD ASR002 BY REMOVING THE KNOBS AND/OR THE COLLIMATOR
FRONT PANEL.
Power Supply
ACCESS THE CROSS POTENTIOMETER BY REMOVING THE WHOLE COVER, SEE CHAPTER- COVER
REMOVAL.
Longitudinal Adjustment
WARNING:
THIS IS PROVIDED AS AN AID TO THE END USER. RALCO IS NOT
LIABLE FOR ANY DAMAGES RESULTING FROM THE ALTERING OF PRESET
FACTORY CONFIGURATIONS.
CROSSHAIR ADJUSTMENT
• Activate the light field.
• Adjust the light field to a narrow line for each pair of shutters
by turning the two knobs alternately.
• Check that the project cross line is exact halfway between
the edges of the shutters, see Fig. Cross Line.
• If adjustment is required, remove the cover from the sides
and bottom of the collimator, see Chapter - COVER
REMOVAL.
• Loosen the four screws securing the plastic panel and
adjust the cross lines to coincide with the light lines, see Fig. Cross Line
Chapter - COVER REMOVAL.
• Tighten the screws.
MECHANICAL MOTION STOP ADJUSTMENT
The purpose of these stops is to prevent stressing the control shafts of the shutters at the two
limits of travel. This is done by limiting rotation of the control shafts.
• Adjustment of these stops will not be required unless:
• The shutters can not be completely closed or opened to the largest size.
• Torsion is felt at one or both ends of the travel.
• The shutters themselves appear to limit the motion rather than the stops.
Fig. Close Stop Long Shutters Fig. Close Stop Cross Shutters
• Tighten the screw on the left side of the collimator, see Fig. Cross Shutters.
• Activate the light field and open the shutters to 43 cm (17”)projection at 100 cm (40").
• Loosen the screw until the head touches the shutter support lever.
• Open and close the shutters several times and make sure that both levers touch the
screws.
• Tighten the fixing nut on the screw.
Potentiometer Adjustment
Potentiometers are located as follows:
Cross - right lateral wall - remove the cover to access, see Chapter - COVER REMOVAL.
Long - back wall, upper left side - remove the rear cover to access, see Chapter - COVER
REMOVAL.
Adjust by loosening the Allen screw on the gear and turning the potentiometer shaft, see Fig.
Potentiometer Long Shutters and Fig. Potentiometer Cross Shutters that follows.
Potentiometer Long Shutters Fig. Potentiometer Cross Shutters
R302DACS/A - R302DACS/A DHHS - Adjustments
• Shift the laser system by loosening the two screws holding the laser system base to the
beam limiting device front plate.
• Move the base until the laser beam falls over the perpendicular bisector line on the
anti-dust panel.
• Tighten the two screws.
Centring
Problem Cause Solution
The light field of the The mirror or light See Chapter- CALIBRATION,
collimator is not source are not paragraph Light Field to X-ray Field
centered positioned correctly Alignment
Shutters
Problem Cause Solution
Adjustment procedures,
Photocells are faulty see Chapter- ADJUSTMENTS
Shutters do not hold
or incorrectly
position or substitute them,
calibrated
see Chapter- SUBSTITUTIONS
External electronics Consult the manual supplied with the
R302DACS/A - R302DACS/A DHHS - Troubleshooting
Additional Filtration
Problem Cause Solution
Consult the manual supplied with the
Faulty electronics
Additional Filtration: unit that controls the collimator
WARNING: DO NOT TOUCH THE LAMP, THE SOCKET, OR THE LAMP BRACKET
WITH YOUR FINGER. THEY CAN BE VERY HOT AND CAUSE SEVERE BURNS.
WARNING: DO NOT TOUCH THE LAMP WITH YOUR FINGERS, EVEN WHEN
IT IS COLD.OIL FROM YOUR SKIN WILL CAUSE THE LAMP TO CRACK AND
POSSIBLY EXPLODE. IF YOU HAVE TOUCHED THE LAMP, WIPE THE SURFACE
WITH ALCOHOL, THEN HANDLE THE LAMP WITH A PIECE OF PAPER.
THE SPARE PARTS LISTED BELOW REFER TO THE STANDARD MODEL ONLY.
Labels
LABEL 1C
R302DACS/A - R302DACS/A DHHS - Spare Parts
THE HEEL EFFECT WILL CAUSE THE FIELD TOWARD THE CATHODE TO BE SLIGHTLY LESS SHARP
THAN ON THE OTHER THREE SIDES. THIS IS NORMAL AND CANNOT BE CORRECTED BY
ADJUSTMENT. IN ADDITION, AN X-RAY TUBE OF 12° OR LESS TARGET ANGLE WILL PRODUCE AN
ASYMMETRICALLY SHAPED FIELD WHEN A LARGE FIELD SIZE IS USED AT SHORT SID, BECAUSE OF
ANODE CUT-OFF EFFECT. THIS IS NORMAL AND MAY NOT BE CORRECTED BY ADJUSTMENT.
R302DACS/A - R302DACS/A DHHS - Optional Items
RO 082 Mirror
Glass mirror, 0.8 mm thickness, inherent filtration 1 mm Al equivalent. With this mirror, the
collimator equivalent filtration value of 2 mm Al is reduced to 1 mm Al.
NOTE
RALCO RECOMMENDS FOLLOWING THE BELOW PROCEDURE, DURING THE INSTALLATION OF THE COLLIMATOR
EQUIPPED WITH THE SIEMENS TOP COVER TO THE X-RAY TUBE. THE CORRECTLY PERFORMED PROCEDURE
ALLOWS THE COLLIMATOR TO ROTATE FREELY.
PROCEDURE:
1. REMOVE THE COLLIMATOR FROM ITS PACKAGING AND PLACE IT ON A FLAT SURFACE,
2. UNSCREW ALL 3 FIXING SCREWS OF THE COLLIMATOR, SEE FIG. 1,
3. INSTALL THE COLLIMATOR ONTO THE FLANGE PREVIOUSLY MOUNTED TO
THE X-RAY TUBE.
4. TIGHTEN ALL 3 SCREWS COMPLETELY, SO THAT THE TABS WILL TOUCH
THE FLANGE SLIGHTLY,
5. ONCE THE SCREWS ARE COMPLETELY TIGHTENED LOOSEN ONE BY ONE ABOUT
1/2 TURN (180°) TO ALLOW THE COLLIMATOR TO ROTATE FREELY. SEE FIG. 2.
Fig. 1 Fig. 2
R302DACS/A - R302DACS/A DHHS - Optional Items
Description
• Board ASR003 is an extension of the radiological system, providing an interface for the col-
limator to other system devices with analogical signals.
• It operates via CAN BUS with a Ralco proprietary protocol.
• The user end features 32 digital inputs, 8 analogue inputs and 10 relay outputs.
• Digital inputs are optoisolated to allow connection to devices that supply output control volt-
ages (typical range 12-32 V DC).
• Supply: 24 V AC 50/60 Hz 6.5 A; 24 V DC.
• Fuse: n.a. (not supplied by Ralco).
Digital Input
Since the input signal sometimes comes from a voltage-free contact, the board features a
number of pins to which an external supply may be connected. As an alternative, these pins
may be connect to a voltage incoming from board ASR003 itself by simply connecting pin M79
to pin M80 and, pin M81 to pin M82.
The available voltage is not taken upstream of the voltage regulator but downstream of the fuse
and rectifier bridge.
The type of software applied determines the function of each input.
.
Analogue Input
In this case the input range will be reduced with respect to the digitalization capacity of the
microprocessor digital/analogue convertor. However, the board will have better protection
against wrong Potentiometer connections.
The analogue inputs are designed primarily for connection to 1 Kohm potentiometers.
Additional safeguards have been added because the user could supply an analogue signal
directly without using potentiometers.
R302DACS/A - R302DACS/A DHHS - Optional Items
The main purpose of these inputs is to furnish the collimator with the cassette size data and
information related to focus-film/focus-floor distances (depending on the configuration).
In general, non linearity introduced by the safeguards is contained. However, should a greater
degree of precision be required (such as in the case of focus-film/focus-floor signals) we
suggest that a voltage between 0 and 6 V DC be supplied so as to make the most of the
convertor range.
Higher voltages up to 12 V, which could be caused by incorrect connections, will not damage
the input.
The 10 output relays each have an NO/NC contact and may consequently be adapted to a
number of different configurations. Their use depends on the software applied.
This board also supplies the go-ahead to program some of the system parameters.
• Connect the collimator using the RO 330 cable. The cable end with the AMP 201298-1
connector is to be inserted into the connector at the back of the collimator.
• The other end of the cable, with the DB9 connector and 3 wires, must be connected to the
CanBus interface and power supply, see Fig. 1.
• Consult the electric drawing to perform a correct connection of inputs.
Fig. 1
R302DACS/A - R302DACS/A DHHS - Optional Items
SUPPLY CABLES (24 V AC 50/60 HZ 6.5 A; 24 V DC) USED FOR COLLIMATOR CONNECTION
MUST BE SUITABLE FOR THE PURPOSE AND COLLIMATOR CURRENT
ABSORPTION.(MIN. SECTION 1.5 MM2).
THE SYSTEM REQUIRES SCREENED CABLES WITH THE ADDITION OF FERRITE CORES (NOT
SUPPLIED) TO AVOID ELECTROMAGNETIC INTERFERENCE WITH THE SERVER ELECTRONICS.
CABLES FROM THE SYSTEM TO THE EXTERNAL BOARD MUST BE SCREENED.
THE FERRITE CORES ARE TO BE MOUNTED ON THE CABLES INSIDE THE BOX AS SHOWN
IN THE DRAWING BELOW. FERRITE CORES ARE USED TO ATTENUATE
INDUCTION AND IRRADIATION OF ELECTROMAGNETIC INTERFERENCE.
WE SUGGEST A 279 FERRITE, ATTENUATION RANGE 100 MHZ TO 200 MHZ, SUITED TO
YOUR CABLES; IE: MSFC-10 FERRITE BY RICHCO.
THE WIRING DIAGRAM INCLUDED IN THIS CHAPTER REFERS TO THE STANDARD PRODUCT.
R302DACS/A - R302DACS/A DHHS - Optional Items
THE DEVICE MUST BE EXCLUSIVELY CONNECTED TO POWER NETWORK, WITH EARTH PROTEC-
TION, IN ORDER TO AVOID A RISK OF ELECTRICAL SHOCK.
CABLES AND TERMINALS USED FOR THE INTERNAL CONNECTION OF THE COLLIMATOR MUST
BE SUITABLE FOR OPERATION AT TEMPERATURES OF 70°C AND COLLIMATOR CURRENT
ABSORPTION.
TAKE CARE THAT ELECTRICAL RESIDUES HOWEVER SMALL (BITS OF COPPER LEADS FROM
WIRES, ETC.) ARE NOT LEFT TO CLUTTER THE CIRCUIT - THESE COULD CAUSE BOARD MALFUNCTION
OR ANOMALIES.
Earthing of the collimator circuit is used to disperse noise or extra voltage coming from the
collimator supply system.
The device that disperses noise is reversible, consequently noise instead of being eliminated
is fed into the system power supply circuit. Therefore, it is essential that the earth outlet to
R302DACS/A - R302DACS/A DHHS - Optional Items
which the system is connect be effectively the unipotential protection point of the system or
the room.
POWER SUPPLY
OUTPUTS
All outputs are simple relay contacts (1 A 24 V DC) configured as normally open.Output relays can switch 1 A to 110 V
AC or 30 V DC; do not exceed these values and provide fuses on load supply circuits.
If relays are used to switch inductive loads (coils of other relays), provide these loads with appropriate snubber R-C
circuits.
Should an electric signal be produced by a multi-contact relay, ground the contacts surrounding the signal or set them
at minimum potential (max. 24 V) by using a snubber circuit for the relating load.
OUT1 With closed contact the system is in "Ready " mode COM M1
N.C. M2
N.O. M3
N.C. M5
N.O. M6
N.C. M8
N.O. M9
N.C. M11
N.O. M12
OUT5 With closed contact the system allows the X-ray “Exposure” COM M13
N.C. M14
N.O. M15
IN M31 Common +V DC
R302DACS/A - R302DACS/A DHHS - Optional Items
IN M34 Common +V DC
IN M37 Common +V DC
IN M40 Common +V DC
IN M43 Common +V DC
IN M46 Common +V DC
IN M49 Common +V DC
IN M57 Fluoro 1
IN M61 Common contact and full size left lateral (24 V DC)
IN M64 Common contacts Bucky and full size right lateral (+V DC)
IN M67 Common contacts Bucky and full size right lateral (+V DC)
IN M69 Fluoro 2
IN M74 Fluoro 3
IN M75 Fluoro 4
IN M76 Fluoro 5
IN-AN M104 Cross potentiometer/ left lateral Bucky terminal (low-level 0 V DC)
IN-AN M112 Cross potentiometer/ right lateral Bucky terminal (low-level 0 V DC)
• The analogue signal (V DC) from the table Bucky potentiometer sliders must increase with
the increase of cassette format. Use a voltmeter to check this condition between contacts
M99 (+) and M100 (-) for the Long function and, between M95 (+) and M96 (-) for the Cross
function.
• The analogue signal (V DC) from the right lateral Bucky potentiometer sliders must
increase with the increase of cassette format. Use a voltmeter to check this condition
between contacts M115 (+) and M116 (-) for the Long function and between M111 (+) and,
M112 (-) for the Cross function.
• The analogue signal (V DC) from the left lateral Bucky potentiometer sliders must increase
with the increase of cassette format. Use a voltmeter to check this condition between con-
tacts M107 (+) and M108 (-) for the Long function and between M103 (+) and M104 (-) for
the Cross function.
• The analogue signal (V DC) from the table translator potentiometer slider must increase
with the decrease of the SID (as the table rises). Use a voltmeter to check this condition
between contacts M91 (+) and M92 (-).
Using the ASR003 board, the operator is able to calibrate the collimator to the components,
see Fig. Radiology System.
R302DACS/A - R302DACS/A DHHS - Optional Items
System calibration is the process by which the collimator is programmed to operate within the
system on which it is installed.
Calibration should be performed by qualified personnel when installing the collimator.
BEFORE STARTING THE CALIBRATION PROCEDURE, TURN THE KEY TO MANUAL MODE (OPEN
PADLOCK IS DISPLAYED).
DATA WILL NOT BE STORED IF THE KEY IS IN THE AUTOMATIC MODE (CLOSED PADLOCK
DISPLAYED).
ON SUBSEQUENT POWER-UPS, THE COLLIMATOR WILL USE THE DATA PREVIOUSLY SAVED.
ID DLC D0 D1 D2 D3 D4 D5 D6 D7
0x7AA 8 00 00 00 00 00 00 10 00
The display will indicate that calibration has started by showing the current language setting
(Italian or English).
Calibration will be performed with the collimator control push-buttons. Display information will
always indicate the key status.
The following section, Flow Chart of Calibration Process, depicts a flow chart of the
calibration process including reference numbers succeeded by the section Calibration
Procedure, where each reference number is explained in detail.
Language
ENGLISH
REF. 1
Measuring Units
CENTIMETERS
REF. 2
CAN DISCRETE
ATS ATS
Vertical BUCKY NO Vertical
Receptor Receptor
NO BUCKY
REF. 6 REF. 6
CAN CAN
Left Lateral
Left Lateral Receptor
Receptor REF. 7 BUCKY=DIFFERENTIAL
REF.7 NO
NO BUCKY
Move Table Down
CAN CAN xxx pts.
REF. 9
Set Table at 25 cm
xxx pts.
REF. 10
A0 B0 C0 D0 F0
E0
Lower 50 cm
xxx pts.
REF. 12
Set Stand at 1 m
xxx pts.
REF. 13
Horiz. SID
Meas # 1
REF. 14
Horiz. SID
Meas # 2
REF. 15
Horiz. SID
Meas # 3
REF. 16
Horiz. SID
Meas # 4
REF. 17
Horiz. SID
Meas # 5
REF. 18
Panoramic
SID # 1
REF. 19
R302DACS/A - R302DACS/A DHHS - Optional Items
Panoramic
SID # 2
REF. 20
Vertical SID
Min.
REF. 21
Vertical SID
Max.
REF. 22
A1
Exp. Hold
Mode
REF. 23
Film Table
Distance
Cross 1 T cm 13: xxx REF. 24
Cross 1 T cm: xxx
Long 1 T cm 13: xxx Long 1 T cm: xxx
REF. 28 REF. 28
SID
Visualization
Cross 2 T cm 43: xxx REF. 25 Cross 2 T cm: xxx
Long 2 T cm 43: xxx Long 2 T cm: xxx
REF. 28 REF. 28
Key
Status Visual. Cross 3 T cm: xxx
REF. 26 Long 3 T cm: xxx
REF. 28
A2
REF. 29
A3
End of Calibration
REF. 32
Lingua Language
Italiano English
Measuring unit
centimetres
Press LAMP to confirm the setting; the next selection will be displayed.
REF. 3 Vertical SID Type
Push-buttons “+” and “-” will allow to switch between SINGLE SID, DIFFERENTIAL, FIXED
and CAN.
SINGLE SID: uses one analogue channel (*) for SID calculation
DIFFERENTIAL: uses two analogue inputs (*) (one for the table and one for the stand)
FIXED: the fixed value is set during calibration
CAN: SID value is transmitted to the collimator via CanBus
Press LAMP to confirm the setting; the next selection will be displayed:
• If FIXED has been selected, vertical SID is set (see REF. 4)
• Lateral SID setting in all the other cases (see REF. 5)
REF. 4 Fixed Vertical SID Value
Push-buttons “+” and “-” will allow to increase or decrease the SID value from 25 cm
minimum to 250 cm maximum.
Press LAMP to confirm the setting; the next selection will be displayed.
Press LAMP to confirm the setting; the next selection will be displayed.
REF. 6 Vertical Receptor
Push-buttons “+” and “-” will allow to switch between BUCKY, CAN, ATS or NO.
• NO: no receptor is connected
• BUCKY: the receptor is a classical Bucky
• CAN: formats are CanBus related to the collimator
• ATS: formats are CanBus related to the collimator (ATS detector mode)
Vertical Receptor
CAN
Press LAMP to confirm the setting; the next selection will be displayed.
REF. 7 Right Lateral Receptor
The type of lateral (right) receptor installed with the system must be selected.
Push-buttons “+” and “-” will allow to switch between BUCKY, CAN, or NO (disabled).
• NO: no receptor is connected
• BUCKY: classical Bucky receptor
• CAN: formats are transmitted to the collimator via CanBus.
Press LAMP to confirm the setting; the next selection will be displayed.
R302DACS/A - R302DACS/A DHHS - Optional Items
Press LAMP to confirm the setting; depending on the vertical SID value entered.
The second line of the display shows the digitized voltage value generated by the table
potentiometer. The operator will lower the table. When reading is stabilized, press LAMP to
complete the setting. Should the operator wish to keep the value previously stored and
maintain the last calibration, this setting may be avoided by pressing FILTER.
Both instances will lead to the next measurement.
REF. 10 Calibration of Vertical SID - Raise Table 25 cm
Set Table at + 25 cm
xxx pts
The operator is required to place the table at the stated height. When the setting is stabilized,
press LAMP to confirm the setting.
Should the operator wish to keep the value previously stored and maintain the last
calibration, this setting may be avoided by pressing FILTER.
REF. 11 Calibration of Vertical SID - Stand All Up
Move Stand Full Up
xxx pts
The display will provide calibration indications for the stand analogue signal, requiring that
the stand be raised to a higher position.
The second line of the display shows the digitized voltage value generated by the stand
potentiometer.
The operator is required to raise the stand. When the value is stabilized, press LAMP to
confirm the setting.
R302DACS/A - R302DACS/A DHHS - Optional Items
Should the operator wish to keep the value previously stored and maintain the last
calibration, this setting may be avoided by pressing FILTER.
REF. 12 Calibration of Vertical SID - Lower Stand 50 cm
Lower 50 cm
xxx pts
The operator is required to place the table at the assigned height. When the value is
stabilized, press LAMP to complete the setting.
Should the operator wish to keep the value previously stored and maintain the last
calibration, this setting may be avoided by pressing FILTER.
The operator is required to set the stand at 1 m (100cm) from the table top.
When the value is stabilized press LAMP to complete the setting
Should the operator wish to keep the value previously stored and maintain the last
calibration, this setting may be avoided by pressing FILTER. At this point, if DISCRETE
lateral SID setting is selected, the lateral SID will need to be entered. Lateral SID is the
distance from the focus to the wall Bucky.
Should the lateral SID value be select via Can, the minimum vertical SID value will need to
be entered.
REF. 14 Setting of SID Horizontal # 1
The display will require the fixed values for the lateral SID to be entered.
Horiz. SID Meas. #1
xxx cm
Push-buttons “+” and “-” will allow to enter the measurements from 25 cm minimum to 250
cm maximum.
Values will be given in inches if this was the initial setting.
Press LAMP to confirm the setting and move forward to the to the next Lateral SID value.
REF. 15 Setting of SID Horizontal # 2
Horiz. SID Meas. #2
xxx cm
Push-buttons “+” and “-” will allow to enter the measurements from 25 cm minimum to 250
cm maximum.
Press LAMP to confirm the setting.
R302DACS/A - R302DACS/A DHHS - Optional Items
Push-buttons “+” and “-” will allow to enter the measurements from 25 cm minimum to 250
cm maximum.
Press LAMP to confirm the setting.
Push-buttons “+” and “-” will allow to enter the measurements from 25 cm minimum to 250
cm maximum.
Press LAMP to confirm the setting.
REF. 18 Setting of SID Horizontal # 5
Horiz. SID Meas. # 5
xxx cm
Push-buttons “+” and “-” will allow to enter the measurements from 25 cm minimum to 250
cm maximum.
Press LAMP to confirm the setting.
REF. 19 Setting of FULL-SIZE SID # 1
Panoramic SID #1
xxx cm
Push-buttons “+” and “-” will allow to enter the measurements from 25 cm minimum to 250
cm maximum.
LAMP confirms the selection and passes to the next SID Panoramic value.
REF. 20 Setting of FULL-SIZE SID # 2
Panoramic SID #2
xxx cm
Push-buttons “+” and “-” will allow to enter the measurements from 25 cm minimum to 250
cm maximum.
LAMP confirms the selection and passes to the next SID Panoramic value.
R302DACS/A - R302DACS/A DHHS - Optional Items
Push-buttons “+” and “-” will allow to enter the measurements from 25 cm minimum to 250
cm maximum.
Press LAMP to confirm the setting.
Push-buttons “+” and “-” will allow to enter the measurements from 25 cm minimum to 250
cm maximum.
Press LAMP to confirm the setting; maximum Vertical SID data will be displayed.
REF. 23 Exposure Hold Mode
This requests the distance from the table top to the image detector. Input this distance.
The following information will be displayed:
Exp. Hold Inval. SID
Enabled
Push-buttons “+” and “-” will allow to enter values from 0 cm minimum to 8 cm maximum
(equal to 0 and 3.1 inches).
Press LAMP to confirm the setting.
REF. 25 SID Visualization
SID Visualization
Yes
R302DACS/A - R302DACS/A DHHS - Optional Items
Push-buttons “+” and “-” activate or deactivate the display of the SID value on the main
window.
Press LAMP to confirm the setting.
REF. 26 Key Status Visualization
Key Status Visual.
Yes
Push-buttons “+” and “-” activate or deactivate the display of the key status on the main
window (open or closed padlock).
Press LAMP to confirm the setting.
Push-buttons “+” and “-” activate or deactivate the display of the collimator tilt angle in
degrees.
Press LAMP to confirm the setting.
Calibrate the vertical Bucky cassettes.
The system will, otherwise, analyze the settings related to the two lateral receptors: if at least
one of the two has been programmed as analogue Bucky, the lateral left and/or right
cassettes will be calibrated.
If no Bucky was selected, the system will display the end of calibration. Cassette Calibration
Information.
The first thing the system requires is to select a quick calibration procedure or a full
calibration procedure for all types of cassettes (vertical, lateral left and lateral right).
The differences between the two calibrations are:
• Quick Cassette Calibration: The system requires to enter two cassette sizes only (one
small and one large).
• Full Cassette Calibration: The system requires to enter all possible cassette values
(max. 10 Buckys).
ie. If 3 cassette sizes are stored as 13x18, 24x30 and 30x40 cassettes, the system will
also be able to identify cassette sizes such as 13x30, 13x40, 24x18, 24x40, 30x18,
30x30.
REF. 28 Table Bucky Cassettes, Calibration
Table Bucky Cal.
Quick
Push-buttons “+” and “-” will allow to switch from Quick to Full calibration.
LAMP confirms the selection.
Press FILTER to cancel the procedure. This keeps last calibration made and moves on to
the next phase.
Cross 1 S cm 13: XXX
Long 1 S cm 13: XXX
Push-buttons “+” and “-” will allow to increase or decrease the cassette number from 1 min
to 10 max.
At this point the operator may:
• Increase Cross up to 43 cm (17”) max with push-button OPEN CROSS.
• Decrease Cross to 13 cm (8”) min with push-button CLOSE CROSS up.
• Increase Long up to 43 cm (17”) max using push-button OPEN LONG.
• Decrease Long up to 13 cm (8”) min using push-button CLOSED LONG.
When values are stabilized, confirm the setting by pressing LAMP.
Push-buttons “+” and “-” will allow to switch from Quick to Full calibration.
LAMP confirms the selection.
Press FILTER to cancel the procedure. This keeps last calibration made and moves on to
the next phase.
The number of cassette sizes to be processed must be entered in the event of a Full
Calibration.
Push-buttons “+” and “-” will allow to increase or decrease the cassette number from 1 min
to 10 max.
LAMP confirms the selection.
Press FILTER to cancel the procedure. This keeps last calibration made and moves on to
the next phase.
Both types of calibration procedures will require the prepossessing of signals from Bucky
potentiometers.
At this point the operator may:
• Increase Cross up to 43 cm (17”) max with push-button OPEN CROSS.
• Decrease Cross to 13 cm (8”) min with push-button CLOSE CROSS up.
• Increase Long up to 43 cm (17”) max using push-button OPEN LONG.
R302DACS/A - R302DACS/A DHHS - Optional Items
The operator may set the format openings by using the push-buttons:
• Cross open, increases Cross opening [max. 43 cm (17”)].
• Cross closed, decreases Cross opening [min. 13 cm (8”)].
• Long opening, increase Long opening [max. 43 cm (17”)].
• Long closed, decreases Long opening [min. 13 cm (8”)].
• Lamp acquires the set value and skips to the next input.
Once the enabled inputs have been entered, calibration is finished.
REF. 32 Saving Calibration Data
The system informs the operator that calibration has terminated.
End of Calibration
For the collimator to rotate to position "Ø", two of the four tabs retract to allow to rotation. Adjust
and tighten the four Allen screws as follows:
screws 1 and 3 = 7 turns until the tongue touches slightly the flange;
screws 2 and 4 = 5½ turns exactly;
See Fig. Mounting Flange in this chapter. Please Note: If Allen screws "2" and "4" are
tightened more than 5 ½ turns, the collimator will not rotate.
Check that the distance between the collimator housing and the mounting flange is equal in all
directions and, that the collimator face is parallel to the table axis.
Loosen the screws and adjust if required.
DO NOT TOUCH THE DISSIPATER WITH YOUR HANDS; IT COULD BE HOT AND
CAUSE SEVERE BURNS.
• If the light-field needs to be moved laterally, loosen (not remove) the fixing screws A.
• Use screw B to adjust transversely.
• When calibration is terminated, lock the screws A.
See Fig. Light Field Adjustment
CONNECTORS
J2 - SUPPLY AND OUTPUTS 1. 20-30 V AC or 12-45 V DC
2. 20-30 V AC or 0 V DC
3. +5 V Fan
4. GND Fan
5. +5 V Laser
6. GND Laser
7. + Power-up/Power down external push-button
8. GND Power-up external push-button
J1 - LED OUTPUT + LED 5 A positive output
- GND Led 5A
J3 - PROGRAMMING 1. Vpp
CONNECTORS 2. +5 V
3. GND
4. PGD
R302DACS/A - R302DACS/A DHHS - Optional Items
5. PGC
6. NC
J4 - HOME SENSOR INPUT + EXTER- 1. + External LED
NAL LED 2. - External LED
3. - Photosensor LED Cathode
4. + Photosensor LED Anode
5. Input signal (Photosensor output collector)
6. GND (Photosensor output emitter)
Jumpers OFF: without jumper
ON: with jumper
JP1, JP2 - ALTERNATE/DIRECT OFF: alternate power supply
POWER SUPPLY ON: direct power supply (pay attention to polarity)
TR1- CURRENT CONTROL Adjust the output current intensity to the LED.
Step-less current adjustment from 0 to 5.1 A. Adjusting
current to approximately 4.7 A (+/- 5%) will increase LED
life without a significant loss of luminosity.
TR2 - TIME ADJUSTMENT Adjusts the time from 30s to 120s up to 90% of run,
maximum light ON time is 15 min. for safety reasons.
R302DACS/A - R302DACS/A DHHS - Optional Items
CLEANING RECOMMENDATIONS
• The collimator housing must be cleaned as prescribed by the sanitary regulations followed
by the operator.
• Disconnect supply.
• Use non abrasive cleaning products.
Care must be taken to prevent liquid from entering the collimator. Please Note: The colli-
mator cover is not watertight.
• Do not reapply power if inflammable liquids have leaked into the collimator.
See the following Maintenance Instructions.
• Clean the varnished and aluminium surfaces with a damp cloth only, using a neutral
cleansing agent then dry the surfaces with a soft cloth.
• Clean chrome surfaces with a dry soft cloth.
Do not spray water or detergent directly over the collimator. The unit’s
liquid protection level is IPx0.
DISINFECTION
The disinfection method use must conform with the currently applicable norms and directives
covering disinfection and protection against explosion hazards.
.
Never use caustic substances, solvents or abrasive detergents.
If products that could form explosive gas mixtures are used, allow the
gas to evaporate before starting the system.
• Disconnect supply.
R302DACS/A - R302DACS/A DHHS - Maintenance
• Disinfect the unit including accessories and cables with a dampened cloth.
• Do not spray the unit with the disinfectant because it could leak into the collimator.
MAKE SURE TO TIGHTEN THE ALLEN SCREWS SECURING THE COLLIMATOR OR THE CONTROL TABS.
APPROPRIATE TIGHTENING OF THE 4 ALLEN SCREWS ENSURES SECURE MOUNTING OF THE
COLLIMATOR. TIGHTENING FORCE USED MUST NOT EXCEED 0.50 NM.
WARRANTY
This product has been manufactured and tested to the highest quality standards by Ralco, srl.
Ralco undertakes to replace and repair any collimator during a period of 24 months for
mechanical and 12 months for electrical parts (motors, potentiometers, electrical boards,
lasers) from the date of invoice (shipment date from Ralco).
The warranty applies provided the product has been handled properly in accordance with its
operating instructions and its intended use.
Warranty covers cost of all components and labor involved, unless:
• Product documents have been altered in any way or made illegible;
• The model or production number on the product has been altered, deleted, removed or
made illegible;
• Repairs or product modifications and alterations have been performed by unauthorized and
unqualified persons;
• Unauthorized repairs and/or modifications have been performed;
• Damage caused by misuse or neglect, incorrect installation or accidental damage;
• Damage occurred during transit due to shipping company, or incorrect packing by cus-
tomer;
• Unoriginal spare parts and accessories have been used.
In-warranty spares will be available only upon return to Ralco, at the customer's expense, of
the parts considered to be faulty to allow Ralco to assess the cause of the fault.
Components Not Covered by this Warranty:
• Consumable items such as lamp bulbs, lexan panels and mirrors;
• Cosmetic damage such as scratches;
• Any missing components when product arrives for repair.
REPAIRS
In the event the customer finds any Non Conformity in the product, please contact Ralco via e-
mail at the address: [email protected] (Ralco Repair Assistance).
To successfully resolve any Non Conformity, the following information must be provided:
• The model and serial number of the collimator found on the label;
• A detailed description of the problem (in Italian or English);
• Whether you want a repair, refurbishment, or model upgrade of the product.
Ralco and the customer will work together to resolve the problem by either providing
R302DACS/A - R302DACS/A DHHS - General
SAFETY/RESPONSIBILITY
Ralco adheres to the directives governing manufacturers of electro-medical equipment:
Directive 2007/47/CE para.10 -Legislative Decree n° 46 para.10
Ralco shall not be held responsible when instructions provided in the present manual are not
complied with. Ralco shall not be held responsible if the collimator relates to one or several of
the following instances:
• The unit is of Ralco construction, built to client specifications with no CE marking.
• The unit has been modified by the OEM or end user.
• The unit has been installed without respecting the instructions provided in this manual.
• The unit is used without respecting the instructions provided in this manual.
• The unit has not been subject to routine functional inspection.
• The unit has not been subject to routine maintenance.
• The unit has been repaired with unoriginal spare parts.
• Ralco shall decline all responsibility for any damage, direct or indirect, caused to persons
or things by inappropriate accessories.
INFORMATION REGARDING ACCIDENTS THAT HAVE OCCURRED WHILE USING THE RADIOLOGICAL
COLLIMATOR MUST BE REPORTED IMMEDIATELY TO RALCO SRL.
RESIDUAL RISKS
The collimator has been constructed to current standards to meet the safety requisites of
directive 2007/47/CE. However, due to the presence of x-rays, the type of application implies
R302DACS/A - R302DACS/A DHHS - General
a residual risk derived from possible faults that could occur during operation of the unit.
The Instructions contained in the this Manual will ensure the correct use of the device and
reduce the causes of possible hazards.
The residual risks of the device are reasonable; they have been assessed and approved in the
related Risk Management Plan contained the Technical Report.
Date:
Customer:
Information regarding possible accidents that may have occurred while using the collimator.
Directive 2007/47/CE states that accidents (such as death or grievous injury to a patient) that involve the
collimator described herein, must be reported to the Ministry of Health and to the Manufacturer.
The present form is provided to report to Ralco srl post-free.
Date:
Customer:
info_cl.fm
info_cl.fm