A707V User Manual
A707V User Manual
ARGUS 707 V
Made in Switzerland
CODAN ARGUS AG
CH-6340 Baar / Switzerland
Park area
Drop chamber drop detector
Display
Global alarm
automatic
„free-flow“-clamp Air bubble
detector
Drop detector
connection
AC power
connection
Identification plate
with pump serial
number
1.2 Use
The ARGUS 707 infusion pump can be used in standard applications requiring the
highest degree of accuracy, as well as in special cancer therapies, blood transfusions
and parenteral nutrition.
Caution! The ARGUS 707 V infusion pump may only be used with calibrated
consumables and sets specified in the appendix and recommended by
CODAN ARGUS AG. The safety of the patient may be endangered.
1.4 Maintenance
No special maintenance is necessary for the ARGUS 707 V infusion pump, apart from the
technical safety check. There are no wear and tear parts that will require preventive re-
placement.
CODAN ARGUS AG
CH-6340 Baar / Switzerland
EMAIL: [email protected]
www.codanargus.com
1.6 Symbols
IPX2 Protected against dripping (±15° tilte d) when operated in horizontal position
Double insulation
Drop detector
Door alarm
Global alarm
Drop alarm
Upstream-Occlusion
(exceeded under pressure bottle side, the right segment of the bar graph
turns on, an intermittent acoustic alarm sounds, the alarm lights are blink-
ing)
Filter in drop chamber obstructed? Kinked line bottle side?
Drop chamber ventilation cap closed? Bottle empty?
Bad set position inside door?
➔ Check above points and resolve problem. If no obvious problem is
found, close roller clamp, open pump door, shift set approximately
12cm towards the right, close pump door, reopen roller clamp and start
pump again.
Drop alarm Deviation of the total number of drops with regard to the set infusion rate
Infusion bottle empty? The tube roller clamp is closed?
The level of the liquid in the drop chamber is too high? Free-flow?
Deviation of the total number of drops with regard to the set infusion rate?
➔ Check above points and resolve problem
Battery empty
"ON/OFF"
This key is used to switch the pump on and off.
To switch the unit off, press the key 2 seconds.
"MODE"
The key "MODE" has the following 4 functions:
- Acoustic and flashing global alarm muting system (for 2 minutes)
- Input mode (select the display for an input)
- Interrogation mode for "ml total" and "h.min"
- Selection of the programmable features (see chapter 4.2)
"START/STOP"
This key is used to start or stop an infusion. In alarm state, this key starts
the pump and mutes the acoustic and global alarm. In stopped state "KVO"
operation is switched on (default, configurable). If the stopped state lasts for
more than 2 minutes, an acoustic reminder signal is activated.
Auto repeat
Pressing a key for longer than 1 sec., the correspond-
ing number is set automatically forward.
PC-configuration
Keep the key "10" pressed and press the key "ON/OFF". For further in-
formation about this function please refer to the technical staff.
3 Set-up
Caution! Only use the recommended accessories, consumables and IV sets with
Luer- Look connections (see appendix). The functional safety of the pump
is not guaranteed if non approved IV sets are used. The safety of the pa-
tient may be endangered. This also applies in the case the patient line is
connected with other infusion systems.
3.2 Installation
The ARGUS 707 V infusion pump can be mounted on a table top, an IV-stand / ceiling
pole (up to diameter 38mm), a standard rail system (optional accessory) and on a pole on
the ARGUS Docking Station A60/100.
When fixed to an IV stand, the equipment should not be mounted higher than 1.2 m over
the floor so that the stability remains.
To fix the equipment to a rail system, use the optional available ARGUS multifunctional
clamping system.
Caution! The connection of several infusion types (gravity, syringe pumps, peristal-
tic pumps, etc.) together into the same tube can be very dangerous.
The combination is allowed with the ARGUS 707 V if at least an IV set
equipped with a back check valve is used in every line.
This type of connection should only be used if expressively specified in the
operation manual of each device and/or approved by a notified body and
applied under a trained qualified clinical or nursing staff.
A B C D
CODAN-set B.Braun-set
Display auto increment: If a numerical key is pressed for longer than 1 sec., the corre-
sponding number is automatically incremented. If key “100” is pressed for automatic in-
crement, check for correct decimal point position since 1000 values can be entered this
way.
3.6 Input of volume and infusion time with automatic rate calculation
1. Turn pump
ON!
4 Special functions
The electronic pressure sensors provide a fast alarm reaction time and a very low occlu-
sion bolus volume. Both patient side (downstream) and container side (upstream) occlu-
sions are recognized.
Caution! Place the IV set in the pump before the pump is switched on.
After switching the pump off and on again or opening the door, the default programmed
alarm level will be set and the sensor reinitialized.
The full scale of the bar graph pressure display equals the selected patient side pressure
alarm level.
The bottle side pressure alarm level is a fixed preprogrammed value. The alarm condition
is shown with a single bar on the right side of the bar graph.
This function is only available if it was enabled by the technical service.It allows selecting or
checking one of the configured IV set number (brand/type).
The infusion line should be already inserted in accordance with chapter 3.3. The IV set
selection is only available after switching on the pump and if more than one IV set is enabled
(do not press the "START" key). As soon as the pump was started once, this function allows
only displaying the selected IV set (both in stop and infusion mode).
The last used IV set will be stored at switching off the pump.
Insert the chosen set in the pump, close the door and turn on the pump (“ON/OFF” key).
With this function the user can fill an empty IV set. This function (only accessible after
the pump is switched on) is only available if it was enabled by the technical service.
While the function "SEt" "FILL" is active, important alarm functions are suppressed!
- Decrease + Increase
This special function allows selecting a new occlusion pressure limit based on the present
line pressure and adding a configured step value according to the following formula
*
The step value is configured by the technical service.
1. Press this key until the
function "PrL" appears in
the left display!
Decrease Increase
CAUTION : Opening the door, turning OFF the pump or the occlusion
alarm will automatically deactivate the neonatology mode.
If this function is activated (On), no drop alarm is released if too many drops (splashes
by vibration) are detected. A drop alarm is released only if no drop is detected within a
certain time.
This special function must only be used in uncritical applications and for the following
special circumstances:
-Transport of patients in the internal hospital area.
-Rescue ambulance or in other rescue equipment.
-Infusion of protein or vitamin based solutions.
In run mode, if this function is turned on, the message "trAn" "SPOr" is blinking
alternatively with the infusion rate display.
If the pump is switched off, the transport mode is also automatically switched off.
If needed again, it must be reactivated (On).
Hours Minutes
The remaining time can be checked by selecting the function timer "tM".
The timer runs in stop and run mode.
Pumps with a standard factory setting detect air bubbles featuring a volume of 250 µl (~
0.247 ml or 35 mm length) upwards and trigger off a specific air-bubble alarm. Your tech-
nical department can adjust this value by means of the configuration menu to within a
range from 50 to 1,000 µl.
Depending on the operating range of the device, it is also possible – again by means of
the configuration menu – to switch on an additional function for monitoring the accumula-
tion of small air bubbles (the volume of the bubbles can be set to anywhere between 100
and 2,000 µl) over a given period of time (settable between 8 and 64 minutes). Setting
this function is recommended in particular for solutions that tend to foam.
The size of the air bubbles to be detected by the pump shall be defined by the medical
personnel in charge.
The following precautions should be observed in order to minimise the risk of air
bubbles forming:
1. The drop chamber must always be at least one third full to a maximum of half full. When
high flow rates are used, it should be half full.
2. The tube must be filled completely and checked before connection to the patient; any air
bubbles must be removed.
3. Y fittings must be properly sealed and 3-way valves must be set correctly.
6 Cleaning / Disinfection
The pump must only be cleaned by "swabbing". Only alcoholic disinfectants may be used.
In order to keep the pump full operational, we recommend regular cleaning. Use a cloth
moistened with lukewarm water for cleaning. Alcoholic cleaning agents must only be us-
ed diluted.
For more information regarding the supply of suitable cleaning agents and disinfectants
please contact the specialists in your house.
The supplier assumes responsibility for the safety, reliability and performance of the unit
only if all following conditions are met:
a) Exclusively authorized persons have carried out the assembly, additions, readjust-
ments, modifications or repairs.
b) The electrical installations in the room where the unit is operated meets the require-
ments of the IEC regulations.
c) The unit has been operated in compliance with the instructions for use.
Caution! The ARGUS 707 V infusion pump may only be used with spare parts, ac-
cessories, consumables and sets with Luer-Lock connections (specified
in the appendix) recommended by CODAN ARGUS AG. The functional
safety of the pump is not guaranteed if non approved materials are used.
The safety of the patient may be endangered.
This manual contains the latest data available. It is subject to further modifications in ac-
cordance with technical improvements.
INPUT
Volume / Rate Infusion rate 0.1 - 999.9 ml/h (in 0.1-step up to 999.9 ml/h)
Volume total (VTBI) 0.1 - 9999 ml
(in 0.1-step up to 999.9 ml and in 1-step from 1000 to 9999 ml)
Set Fill (prime) rate 1 - 999 ml/h (in 1-step up to 999 ml/h)
Rate calculation Volume total (VTBI) and infusion delivery time
Bolus Bolus rate 0.1 - 1200 ml/h programmable without flow interruption
(in 0.1-step up to 999.9 ml/h and in 1-step from 1000 to 1200 ml/h)
Bolus volume 0.1 - 999 ml programmable without flow interruption
(automatic & manual) (in 0.1-step up to 999 ml)
Time Infusion delivery time 1 min - 99 h 59 min (in 1 min-step up to 99:59 h)
KVO KVO rate (KOR) 0.1 - 3 ml/h (depending on entered infusion rate)
ACCURACY
Rate Flow rate deviation ≤ ± 5%
• For rates from 1 to 999.9 ml/h
• IV-set changed every 24 hours or after 2.5 liters infused
• Maximum back pressure +/- 100 mmHg
• Depends on the infusion set used
Flow discrepancy in the event
of a technical failure ≤ ± 10%
Tech. Technical deviation < 1%
OPERATING REQUIREMENT
Temperature range 5 °C -40 °C
Medication temperature 18 °C -30 °C
Storage temperature range 0 °C -40 °C
Relative humidity (permissible) 20 -90 %; no condensation
POWER SUPPLY
Battery type NiMH- 12 V / 1.5 Ah (maintenance-free)
Battery operation 5 h @ 25 ml/h
Battery charging time 16 h
External DC power supply 20 V / 0.3 A
Power consumption max. 12 VA
Line fuse 125 mAT
AC power supply 230 V +/- 10%, 50 - 60Hz
Optional AC power supply 115 V +/- 10%, 50 - 60Hz
INTERFACE
Data interface 2 x RS-232 (1 x galvanic isolated)
DISPLAY
LED 1 large 4-digit LED-display (left) Infusion rate, additional information
1 large 4-digit LED-display (right) Volume infused (0.1- 9999ml), volume total (VTBI),
infused time (1 min – 99 h 59 min), additional information
Pictograms and LED'S Operation and alarm conditions
LED bar graph Pressure monitoring display
MEASUREMENT / MATERIAL
Dimensions 190 x 160 x 130 mm (W x H x D) excluded combi clamp
Weight 2 kg including battery (without accessories)
Housing ASA (high performance plastic)
10000
1000
100
Alarmdelay [s]
Measured values
Measured values
PrL = 700 mbar Typical values
Typical values
10
1
PrL = 200 mbar
0.1
0 100 200 300 400 500 600 700
Rate [ml/h]
Bolus volume after occlusion (typical values) of the A707 volumetric pump
600
500
400
Bolus volume in [µl]
300
200
Important remark:
100 Bolus volume with automatic pressure release: typical 100 µl !
0
1 2 3 4 5 6 7 8 9 10
Pressure limit steps in 100 [mbar]
Test I in the first two hours of the test period of 24 hours at 1.0 ml/h (initial period)
Trumpet graph Start-up graph
15 2.0
1.8
Percentage error(%)
10 Ep(max) 1.6
1.4
Flow(ml/h)
5 1.2
1.0
0 0.8
-5 Ep(min) 0.6
Overall percentage erro r -2.09% 0.4 Set rate 1.0 ml/h
-10 0.2
0.0
-15 -0.2
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 0 10 20 30 40 50 60 70 80 90 100 110 120
Observation window(min) Time(min)
Trumpet graph
15
Percentage error(%)
10
Ep(max)
5
0
-5 Ep(min)
Overall percentage erro r -0.1%
-10
-15
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30
Observation window(min)
Test I in the first two hours of the test period of 24 hours at 25 ml/h (initial period)
Trumpet graph Start-up graph
15 50.0
45.0
Percentage error(%)
10 40.0
Ep(max) 35.0
Flow(ml/h)
5 30.0
25.0
0 20.0
-5 Ep(min)
15.0
10.0 Set rate 25 ml/h
Overall percentage erro r -1.19% 5.0
-10
0.0
-15 -5.0
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 0 10 20 30 40 50 60 70 80 90 100 110 120
Observation window(min) Time(min)
Test II in the last two hours of the test period of 24 hours at 25 ml/h (end period)
Trumpet graph
15
Percentage error(%)
10
Ep(max)
5
0
-5 Ep(min)
Overall percentage erro r -3.00%
-10
-15
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30
Observation window(min)
Caution! Each time you change set brand (manufacturer) or tube material you have
to perform a new set calibration (contact your service department).
Caution! Unless otherwise specified by the customer, the ARGUS 707 V has been
calibrated with the CODAN L86 Art. 43.3030 (NoDEHP) infusion set.
The accuracy of ± 5 % cannot be guaranteed for infusion rates > 600 ml/h for the
following IV set: