100% found this document useful (1 vote)
2K views

A707V User Manual

This document is a manual for the ARGUS 707 V volumetric infusion pump. It provides introductions and instructions on operating the pump, including its control panel functions, set up procedures, special functions, safety information, cleaning and disinfection, warranty, and technical specifications. The pump is designed for precise delivery of fluids and features an electronic pressure sensor, programmable options, selection of IV sets, and other functions.

Uploaded by

eri
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
2K views

A707V User Manual

This document is a manual for the ARGUS 707 V volumetric infusion pump. It provides introductions and instructions on operating the pump, including its control panel functions, set up procedures, special functions, safety information, cleaning and disinfection, warranty, and technical specifications. The pump is designed for precise delivery of fluids and features an electronic pressure sensor, programmable options, selection of IV sets, and other functions.

Uploaded by

eri
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 38

Manual for Volumetric Infusion Pump

ARGUS 707 V
Made in Switzerland

CODAN ARGUS AG
CH-6340 Baar / Switzerland

(a member of the CODAN group)

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 1


TABLE OF CONTENTS
1 INTRODUCTION ..................................................................................................................... 5
1.1 GENERAL INFORMATION ................................................................................................................................... 5
1.2 USE ................................................................................................................................................................ 5
1.3 SCOPE OF SUPPLY........................................................................................................................................... 5
1.4 MAINTENANCE................................................................................................................................................. 5
1.5 SUPPORT ........................................................................................................................................................ 6
1.6 SYMBOLS ........................................................................................................................................................ 6
2 CONTROL PANEL .................................................................................................................. 7
2.1 OPERATIONS AND ALARM DISPLAYS .................................................................................................................. 7
2.2 SPECIAL KEY FUNCTIONS ............................................................................................................................... 10
3 SET-UP .................................................................................................................................. 11
3.1 GENERAL INFORMATION ................................................................................................................................. 11
3.2 INSTALLATION ............................................................................................................................................... 11
3.3 PUMP SET-UP ................................................................................................................................................ 12
3.4 DELIVERY OPERATION WITHOUT A PRESET VOLUME ......................................................................................... 13
3.5 DELIVERY OPERATION WITH A PRESET VOLUME (VTBI) .................................................................................... 14
3.6 INPUT OF VOLUME AND INFUSION TIME WITH AUTOMATIC RATE CALCULATION .................................................... 14
3.7 CHANGE THE INFUSION RATE WITHOUT INFUSION INTERRUPTION ...................................................................... 15
4 SPECIAL FUNCTIONS ......................................................................................................... 15
4.1 THE ELECTRONIC PRESSURE SENSOR ............................................................................................................. 15
4.2 PROGRAMMABLE OPTIONS ............................................................................................................................. 16
4.3 SELECTING OR CHECKING AN IV SET............................................................................................................... 17
4.4 FILL IV SET ................................................................................................................................................... 18
4.5 INPUT OF THE BOLUS RATE AND THE BOLUS VOLUME........................................................................................ 19
4.6 MANUAL BOLUS APPLICATION ........................................................................................................................ 20
4.7 AUTOMATIC BOLUS APPLICATION.................................................................................................................... 21
4.8 BATTERY CAPACITY ....................................................................................................................................... 21
4.9 SETTING OF OCCLUSION PRESSURE LIMIT ....................................................................................................... 22
4.10 ACTIVATION OF THE NEONATOLOGY MODE AND NEONATOLOGY OCCLUSION PRESSURE LIMIT SETTING ................ 23
4.11 PATIENT TRANSPORT ..................................................................................................................................... 25
4.12 CLEAR "ML INF." (VOLUME INFUSED) IN STOP AND RUN MODE ........................................................................... 26
4.13 DISPLAY OF ACCUMULATED "ML INF." SINCE LAST POWER UP (BALANCE) ........................................................... 26
4.14 DATA-LOCK (KEYBOARD LOCK) ....................................................................................................................... 27
4.15 SETTING THE STAND-BY ALARM TIME .............................................................................................................. 27
4.16 INPUT OF MEDICATION NAME .......................................................................................................................... 28
4.17 TIMER ALARM ................................................................................................................................................ 28
4.18 AIR BUBBLE DETECTION ................................................................................................................................ 29
5 SAFETY INFORMATION ...................................................................................................... 30
5.1 RISK AND DANGER ......................................................................................................................................... 30
5.2 SAFETY STANDARD CHECKS (SSC) ................................................................................................................ 31
6 CLEANING / DISINFECTION ................................................................................................ 31
6.1 GENERAL REFERENCES ................................................................................................................................. 31
7 WARRANTY .......................................................................................................................... 32
7.1 W ARRANTY DURATION ................................................................................................................................... 32
7.2 W ARRANTY LIMITATIONS ................................................................................................................................ 32
8 ACCESSORIES ..................................................................................................................... 33
9 SPECIFICATIONS ................................................................................................................. 34
APPENDIX: RECOMMENDED IV SETS ................................................................................... 38

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 2


Drop detector

Park area
Drop chamber drop detector

Display

Global alarm

automatic
„free-flow“-clamp Air bubble
detector

Pressure sensor Pressure sensor


„upstream“ „downstream“

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 3


Universal clamp Staff alert and
RS-232-connection

Drop detector
connection
AC power
connection
Identification plate
with pump serial
number

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 4


1 Introduction

1.1 General information


Congratulations on selecting the Swiss high-tech and top quality ARGUS 707 V infusion
pump. This medical device meets all the provisions of the directive MDD 93/42/EEC
which apply to it.
The ARGUS 707 V infusion pump is characterized particularly by the following ad-
vantages:
- Swiss made high-tech quality volumetric pump
- Intuitive, very easy operation
- Modern design, light-weight and compact
- Uses standard sets
- User-selectable occlusion alarm limit
- Upstream & downstream pressure monitoring with bar graph display of line pressure
- ARGUS multifunctional clamping system
- Table top operation possible
- ARGUS Docking Station A60/100 compatible
- Flash memory for fast software updates

1.2 Use
The ARGUS 707 infusion pump can be used in standard applications requiring the
highest degree of accuracy, as well as in special cancer therapies, blood transfusions
and parenteral nutrition.

Caution! The ARGUS 707 V infusion pump may only be used with calibrated
consumables and sets specified in the appendix and recommended by
CODAN ARGUS AG. The safety of the patient may be endangered.

1.3 Scope of supply


ARGUS 707 V infusion pump with power cord, external drop detector and user manual.
Options: Bottle holder, rail combi clamp.

1.4 Maintenance
No special maintenance is necessary for the ARGUS 707 V infusion pump, apart from the
technical safety check. There are no wear and tear parts that will require preventive re-
placement.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 5


1.5 Support
Servicing should only be carried out by CODAN ARGUS AG trained personnel or by an
approved local distributor. In case of repair, send the unit with the filled out “repair order
form” (see service manual) to the local distributor. Further information is available from:

CODAN ARGUS AG
CH-6340 Baar / Switzerland
EMAIL: [email protected]
www.codanargus.com

1.6 Symbols

Caution: consult accompanying documents

IPX2 Protected against dripping (±15° tilte d) when operated in horizontal position

Complies with MDD 93/42/EEC directive

Applied parts CF type device (leak currents protection)

Double insulation

Staff alerting system

Conform to WEEE 2002/96/EC directive


(Waste in Electrical and Electronic Equipment)

Data communication interface

Drop detector

Infusion flow direction

18–30°C Recommended temperature range for the solution and IV sets

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 6


2 Control panel

2.1 Operations and alarm displays

Drop Battery alarm

Battery operation KVO-Infusion * Occlusion / Set


installationalarm

External supply Air bubble


(charging battery) alarm

Door alarm
Global alarm

Drop alarm

* KVO-operation (Keep Vein Open):


- 3 ml/h for infusion rates ≥ 10 ml/h
- 1 ml/h or set infusion rate, whichever is less, for infusion rates <10 ml/h

Alarm situation: - An intermittent acoustic sound is released (can be momentarily muted


with „Mode“ key)
- The staff alert system is activated
- A pictogram is turned on and the global alarm is flashing

Caution! The ARGUS infusion pump cannot be started, if:

• the tube is not properly installed in the pump


• the pressure in the tube is too high or the infusion solution too cold
• the infusion tube contains air bubbles inside air detector
• the infusion rate is = “0.0” ml/h
• the door is open
• the battery is depleted (applies only when operated on battery)
• a non recommended IV set might be used

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 7


Occlusion alarm Downstream Occlusion
(The patient side pressure in the tube exceeds the electronic pressure
alarm level, the entire pressure bar graph turns on, an intermittent acoustic
alarm sounds, the alarm lights are blinking)
Needle obstructed? Kinked line patient side?
Roller clamp patient side closed? Bad set position inside door?
➔ Check above points and resolve problem.
Caution!
The automatic pressure reduction can withdraw blood from the patient.

Upstream-Occlusion
(exceeded under pressure bottle side, the right segment of the bar graph
turns on, an intermittent acoustic alarm sounds, the alarm lights are blink-
ing)
Filter in drop chamber obstructed? Kinked line bottle side?
Drop chamber ventilation cap closed? Bottle empty?
Bad set position inside door?
➔ Check above points and resolve problem. If no obvious problem is
found, close roller clamp, open pump door, shift set approximately
12cm towards the right, close pump door, reopen roller clamp and start
pump again.

Air bubble alarm Air was detected


Air bubbles in the IV set? Tube not properly positioned?
➔ Remove air bubbles, reinstall the IV set or check if recommended IV set
is used

Door alarm Door open

Drop alarm Deviation of the total number of drops with regard to the set infusion rate
Infusion bottle empty? The tube roller clamp is closed?
The level of the liquid in the drop chamber is too high? Free-flow?
Deviation of the total number of drops with regard to the set infusion rate?
➔ Check above points and resolve problem

Battery alarm Pre alarm battery low


(The battery pictogram turns on, an intermittent acoustic alarm sounds)
Battery almost empty, pump will stop in approx. 15 minutes

Battery empty

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 8


(The battery pictogram turns on, an intermittent acoustic alarm sounds, the
alarm lights are blinking)
• The pump goes in stop mode while the battery alarm continues for 6 min.
Thereafter the pump is switched off automatically in order to prevent a to-
tal discharge of the battery.
➔ Immediately connect the power cord to the mains outlet and continue to
infuse, battery will be recharged automatically.

End alarm Infusion end


The volume total has been reached (→ KVO operation)

Safety check Reminder alarm for safety standard check


If the display flashes "CtrL" several times after switching the pump on,
it is time for the safety standard check.
➔ Please refer to your technical staff

Stand-by alarm No manipulation has been made for 2 minutes


(pump in stop mode)
➔ Press key “MODE”

Technical alarm F-code (F-XX)


Technical alarm (failure) with continuous acoustic sound.
➔ If F-code is invisible, press the “MODE” key

NURSE CALL Staff alerting system


A connecting socket allows the unit to be hooked up to an external paging
system. The optical and acoustic alarms of the pump are not affected.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 9


2.2 Special key functions

ALARM MUTING “MODE” muting system


By pressing the “MODE” key the acoustic alarm can be interrupted for
2 minutes. After the muting time has elapsed the alarm is reactivated
automatically.

CLEARING Clearing of the alarm condition


After the cause of the failure has been corrected, press the “Start/Stop”
key to clear the alarm condition and to restart the infusion.

"ON/OFF"
This key is used to switch the pump on and off.
To switch the unit off, press the key 2 seconds.

"100", "10", "1", "0.1"


These keys are used for all numerical inputs.

"MODE"
The key "MODE" has the following 4 functions:
- Acoustic and flashing global alarm muting system (for 2 minutes)
- Input mode (select the display for an input)
- Interrogation mode for "ml total" and "h.min"
- Selection of the programmable features (see chapter 4.2)

"START/STOP"
This key is used to start or stop an infusion. In alarm state, this key starts
the pump and mutes the acoustic and global alarm. In stopped state "KVO"
operation is switched on (default, configurable). If the stopped state lasts for
more than 2 minutes, an acoustic reminder signal is activated.

Clear the display


The selected display will be set to zero
by pressing both keys simultaneous.

Auto repeat
Pressing a key for longer than 1 sec., the correspond-
ing number is set automatically forward.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 10


Software release display and display test
Keep the key "MODE" pressed and press the key "ON/OFF".
The left hand display shows "707", the right hand display "rx.xx" (soft-
ware release) during 3 sec.
Afterwards the visual display test starts as follows:
Display of "2", "4", "7", "F.", "ml total", "ml inf", h.min", pressure
display, operation symbols, alarm symbols and "ALARM" with
acoustical beeps.

Call back of the last infusion values


Keep the key "1" pressed and press the key "ON/OFF".
The following values are now available in the display:
Infusion rate, preset volume, infused volume, infusion time, pres-
sure limit, medicament No. and the last pre-selected bolus volume.

PC-configuration
Keep the key "10" pressed and press the key "ON/OFF". For further in-
formation about this function please refer to the technical staff.

3 Set-up

3.1 General information


The infusion pump must only be used under the supervision of qualified clinical or nursing
staff. The user has the responsibility to read and to observe the following instructions. Al-
so use only standard IV administration sets in accordance with the instructions of the
manufacturer, check for valid set selection and be sure that the inserted set is calibrated!
Under normal conditions, we recommend changing the IV administration set every
24 hours or after 2.5 liters infused.

Caution! Only use the recommended accessories, consumables and IV sets with
Luer- Look connections (see appendix). The functional safety of the pump
is not guaranteed if non approved IV sets are used. The safety of the pa-
tient may be endangered. This also applies in the case the patient line is
connected with other infusion systems.

3.2 Installation
The ARGUS 707 V infusion pump can be mounted on a table top, an IV-stand / ceiling
pole (up to diameter 38mm), a standard rail system (optional accessory) and on a pole on
the ARGUS Docking Station A60/100.
When fixed to an IV stand, the equipment should not be mounted higher than 1.2 m over
the floor so that the stability remains.
To fix the equipment to a rail system, use the optional available ARGUS multifunctional
clamping system.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 11


3.3 Pump set-up
The ARGUS 707 V infusion pump must only be used under the supervision of qualified
clinical or nursing staff. The user has the responsibility to read and to observe the
following instructions.
Only use standard IV administration sets in accordance with the instructions of the
manufacturer.

Caution! The connection of several infusion types (gravity, syringe pumps, peristal-
tic pumps, etc.) together into the same tube can be very dangerous.
The combination is allowed with the ARGUS 707 V if at least an IV set
equipped with a back check valve is used in every line.
This type of connection should only be used if expressively specified in the
operation manual of each device and/or approved by a notified body and
applied under a trained qualified clinical or nursing staff.

a) Connect the power cord to the AC line


b) Open the IV set packaging, slide the roller clamp down the line to be able to place
it between pump and patient, then close roller clamp and connect the IV set to the
fluid container
c) Fill the drop chamber 1/3 to max. 1/2, open the roller clamp and fill the entire set
(make sure to remove all air bubbles)
d) Close the roller clamp again
e) Slide the drop detector over the drop chamber as shown in picture B and C.
Observe the notch and do not pull on the spring (see picture D).

A B C D

Drop detector WRONG !

CODAN-set B.Braun-set

f) Open the pump door by pulling up the door handle


g) Insert the tube into the pump from the left, place the tube slightly stretched in the
tube guides (observe flow direction left to right) and push the tube properly into
the notches both sides of the pump
h) Close the pump door and open the tube roller clamp
i) Check that there is no "free-flow"

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 12


j) Connect the IV set to the patient IV catheter
k) Be sure that the pressure in the tube is zero (= 0 bar)
l) Switch the pump ON
m) Proceed to the flow rate input in accordance with the following chapters

Caution! Please mind the position of the decimal point:

For display values up to 999.9 For display values ≥ 1000

Display auto increment: If a numerical key is pressed for longer than 1 sec., the corre-
sponding number is automatically incremented. If key “100” is pressed for automatic in-
crement, check for correct decimal point position since 1000 values can be entered this
way.

3.4 Delivery operation without a preset volume

This function is only available if drop detector is enabled!

1. Turn pump ON! 3. Start the infusion!

2. Input the desired infusion


rate (e.g. 200 ml/h)!

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 13


3.5 Delivery operation with a preset volume (VTBI)

2. Input the desired infusion


1. Turn pump ON! rate (e.g. 200 ml/h)!

5. Start the infusion!

4. Input the desired infusion 3. Press this key once; indica-


volume (e.g. 300 ml)! tion VTBI "ml total" turns on!

3.6 Input of volume and infusion time with automatic rate calculation

3. Input the desired infusion 2. Press this key once; indica-


volume (e.g. 20 ml)! tion VTBI "ml total" turns on!

1. Turn pump
ON!

6. By pressing this key, the


Hours calculated flow rate will be
5. Input the de-
displayed!
sired infusion Minutes Please check if the dis-
time!
played infusion rate suits
your application before you
4. Press this key until
start the infusion!
indication “h.min” Pressing this key again will
turns on! start the infusion!
10 h 1h 10 min 1 min

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 14


3.7 Change the infusion rate without infusion interruption

1. The present running infusion


flow rate is e.g. 20.2 ml/h.

3. If the new infusion rate is not


confirmed with the START
key within 2 seconds the dis-
play ml/h starts to flash! The
pump still infuses with the
previous flow rate!
If the new flow rate is not con-
firmed within 15 seconds the
2. Input the new infusion
acoustic alarm will be re-
rate (e.g. 42.7 ml/h)! leased and the previous infu-
sion rate will be displayed
again. The pump still infuses
with the previous flow rate!
Confirm old flow rate and
4. Mute the alarm clear the alarm with this key!
with this key!

4 Special functions

4.1 The electronic pressure sensor

The electronic pressure sensors provide a fast alarm reaction time and a very low occlu-
sion bolus volume. Both patient side (downstream) and container side (upstream) occlu-
sions are recognized.

Caution! Place the IV set in the pump before the pump is switched on.

The electronic pressure sensor can be used in two different modes:

a) Fixed pressure level


If the pressure in the downstream system exceeds the programmed pressure alarm level,
the infusion will be stopped and an occlusion alarm released. The ARGUS 707 V will then
automatically reduce the pressure in the set; possibly patient blood could flow back in-
to the tube. Before starting the infusion again, search carefully for the cause and elimi-
nate the problem.
Caution! If the door has been opened, switch the pump off, close the door and
switch the pump on again while maintaining the “1” key pressed
(recall of previous infusion data).

b) Adjustable pressure level


At any time the staff has the possibility to adjust the pressure alarm level in the menu
"PrL" from 100 to 1000 mbar in 10 steps of 100 mbar (10 to 100 kPa, 75 – 750 mmHg).

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 15


If the pressure alarm level is changed while the infusion is running and no key is pressed
during 5 seconds (programmable), both displays change back to the basic position!

After switching the pump off and on again or opening the door, the default programmed
alarm level will be set and the sensor reinitialized.

The full scale of the bar graph pressure display equals the selected patient side pressure
alarm level.

The bottle side pressure alarm level is a fixed preprogrammed value. The alarm condition
is shown with a single bar on the right side of the bar graph.

4.2 Programmable options


If one of the following option is to be used, please get in touch with the local distributor or
with CODAN ARGUS AG service department.

a) Display of the infusion time


The elapsed infusion time is indicated in hours and minutes.
In the VTBI-mode the remaining time is indicated.
b) SBS (step by step)
If the preset volume is reached and increased afterwards only the difference between
the new and the old value is infused after the pump is restarted.
c) VTBI (volume to be infused)
The volume to be infused is indicated.
d) Set rate "ml/h" automatically to “0.0” when the pump is switched on again.
e) The last preset volume "ml” will automatically appear when the pump is switched on
again.
f) Automatic return on default infusion set 1 after power up if set 2, 3 or 4 was used be-
fore power off
g) Neonatology option with inline pressure indication and precision occlusion pressure
limit adjustment (see chapter 4.10)
h) Choice of the occlusion threshold display unit (mbar, mmHg, kPa, cmH2O, Psi)
i) No automatic pressure release after occlusion.
j) Air detector, air bubble size programmable (50...1000 µl)
The air bubble size is max.100 µl for rates <10 ml/h
k) Air detector, air volume accumulated over time (e.g. 1 ml over 0,5 h)
l) No acoustic acknowledgement when pump starts to infuse
m) Buzzer volume adjustable
n) Display brightness adjustable
o) KVO options (KVO only at infusion end)
p) Second rate input is required
q) no detection of the upstream occlusion
r) Additional functions:

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 16


"SEt" "-x-" IV-set selection (see chapter 4.3)
"SEt" "FILL" Fill IV set (see chapter 4.4)
"boLu" Bolus application (see chapters 4.5 - 4.7)
"CAP" Battery capacity (see chapter 4.8)
"PrL" Pressure limit (see chapter 4.9)
"trA" Transport mode (see chapter 4.11)
"CLr" Clear "ml inf." (volume infused) (see chapter 4.12)
"InF" Display of accumulated "ml inf." since last power up (balance) (4.13)
"dLo" Data-lock (see chapter 4.14)
"Stb" Stand-by (see chapter 4.15)
"MEd" Medication name (see chapter 4.16)
"tM" Timer (see chapter 4.17)
Air Bubble Detection (see chapter 4.18)

4.3 Selecting or checking an IV set

This function is only available if it was enabled by the technical service.It allows selecting or
checking one of the configured IV set number (brand/type).

The infusion line should be already inserted in accordance with chapter 3.3. The IV set
selection is only available after switching on the pump and if more than one IV set is enabled
(do not press the "START" key). As soon as the pump was started once, this function allows
only displaying the selected IV set (both in stop and infusion mode).
The last used IV set will be stored at switching off the pump.
Insert the chosen set in the pump, close the door and turn on the pump (“ON/OFF” key).

1. Press this key until the function


"SEt" "- X -" appears in the display

2. Press this key until the 3. Confirm your choice


desired set number ap- with this key.
pears in the right display!
If no key is pressed within
5 seconds (programma-
ble) the pump changes
back to the basic position.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 17


4.4 Fill IV set

With this function the user can fill an empty IV set. This function (only accessible after
the pump is switched on) is only available if it was enabled by the technical service.
While the function "SEt" "FILL" is active, important alarm functions are suppressed!

Caution! Do not connect the IV-set to the patient!

1. Press this key until the


function "SEt" "FILL"
appears in the displays!

2. As long as this key is pressed


the pump runs at the maximum
speed. The display "FILL" is
flashing. After 60 seconds the
delivery will be interrupted. In
order to pump more, you have
to press the key again!
If no key is pressed within 5
seconds (programmable) the
pump changes back to the
basic position.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 18


4.5 Input of the bolus rate and the bolus volume
This function (accessible in stop- and run-mode) is only available if it was enabled by the
technical service.
1. Press this key until the
function "bolu" "MAn" or
"bolu" "Auto" appears in
the displays!

2. Pressing this key to activate


pre-selection of the bolus
rate. The left hand display
shows the last active bolus
rate. "boLr"!

Input of bolus rate:


This function (accessible in stop- and run-mode) is only available if it was enabled by the
technical service.

3. Use these keys to enter a


new bolus rate!
If no key is pressed during 5
seconds (programmable),
both displays change back to
the basic position!

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 19


Input of bolus volume:
This function (accessible in stop- and run-mode) is only available if it was enabled by the
technical service. To activate the automatic bolus, it is mandatory to input a bolus volume.

4. Press this key to activate


the pre-selection of the bolus
5. Use these keys to enter a volume. The left display
new bolus volume (e.g.10.0)! shows “tot”, the right hand
If no key is pressed during 5 display shows the last active
seconds (programmable), bolus volume!
both displays change back to
the basic position!

4.6 Manual Bolus application


This function (accessible in stop- and run-mode) is only available if it was enabled by the
technical service.

1. Press this key until the


function "bolu" "MAn"
appears in the displays!

2. Start the bolus application with


this key. Hold it down until the
desired bolus volume is infused
or the maximum pre selected
volume is reached!

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 20


4.7 Automatic Bolus application
This function (accessible in stop- and run-mode) is only available if it was enabled by the
technical service and a bolus rate and a bolus volume has been entered (see chap. 4.5).
1. Press this key until the
function "bolu" "Auto"
appears in the displays!

2. Press this key to start the auto-


matic bolus application!
Press this key again to interrupt
the running automatic bolus;
pump will then continue to infuse
at the normal rate!

4.8 Battery capacity


This function is accessible in stop- and run-mode.

1. Press this key until the


function "CAP" appears
in the left display!

2. The right display shows the battery


running time in hours and minutes
(e.g. 2 h 30 min).
If no key is pressed within 5 sec-
onds (programmable) the display
changes back to the basic position!

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 21


4.9 Setting of occlusion pressure limit
This function is accessible in stop- and run-mode.

1. Press this key until the


function "PrL" appears in
the left display!

- Decrease + Increase

2.Press these keys to set


the occlusion pressure
limit!
In run-mode, if no key is
pressed within 5 seconds
(programmable) the dis-
play changes back to the
basic position!

Occlusion pressure limits


mbar mmHg kPa Psi cmH2O
100 75 10 1.4 101
200 150 20 2.9 203
300 225 30 4.3 305
400 300 40 5.8 407
500 375 50 7.2 509
600 450 60 8.7 611
700 525 70 10.1 713
800 600 80 11.6 815
900 675 90 13.0 917
1000 750 100 14.5 1019

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 22


4.10 Activation of the neonatology mode and neonatology occlusion pressure limit setting
This function (only accessible after a successful START) is only available if it was enabled by
the technical service and the pump was started once after start up.

This special function allows selecting a new occlusion pressure limit based on the present
line pressure and adding a configured step value according to the following formula

New occlusion pressure limit = Present line pressure + « Configured step »*

*
The step value is configured by the technical service.
1. Press this key until the
function "PrL" appears in
the left display!

2. Press this key and the new occlusion


pressure limit appears in the right
display! (Limit is set according to
above mentioned formula). The neo-
natology mode is now activated.
If no key is pressed within 5 seconds
(programmable) the display changes The maximum bar graph level
back to the basic position! corresponds now to the « Configured step »

4.10.1 Line pressure display in neonatology mode


In normal neonatology infusion mode, an approximate value of the line pressure is
displayed alternatively with the rate and volume display:

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 23


4.10.2 Deactivation of the neonatology mode

1. Press this key until the


function "PrL" appears in
the left display!

Decrease Increase

2. Press these keys to set a new


occlusion pressure limit!
In run-mode, if no key is pressed
within 5 seconds (programmable)
the display changes back to the
basic position!
The neonatology mode is now
deactivated.

CAUTION : Opening the door, turning OFF the pump or the occlusion
alarm will automatically deactivate the neonatology mode.

In all these cases, the neonatology mode must be explicitly


reactivated.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 24


4.11 Patient transport
This function (accessible in stop- and run-mode) is only available if it was enabled by
the technical service.

If this function is activated (On), no drop alarm is released if too many drops (splashes
by vibration) are detected. A drop alarm is released only if no drop is detected within a
certain time.

This special function must only be used in uncritical applications and for the following
special circumstances:
-Transport of patients in the internal hospital area.
-Rescue ambulance or in other rescue equipment.
-Infusion of protein or vitamin based solutions.

1. Press this key until the


function "trA" appears in
the left display!

2. Use this key to switch the transport


mode On or OFF (toggle function)!
If no key is pressed within 5 seconds
(programmable) the display changes
back to the basic position.

In run mode, if this function is turned on, the message "trAn" "SPOr" is blinking
alternatively with the infusion rate display.

If the pump is switched off, the transport mode is also automatically switched off.
If needed again, it must be reactivated (On).

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 25


4.12 Clear "ml inf." (volume infused) in stop and run mode
This function is only available if it was enabled by the technical service.

1. Press this key until the


function "CLr" appears
in the left display!

2. Press both keys simultane- 3. RUN MODE: If the clear


ously to clear "ml inf"! function is not confirmed with
Right display will be set to 0. this key within 15 seconds,
the acoustic alarm will be re-
leased and the previous val-
ue will be displayed again!
Confirm old value and clear
the alarm with this key!

4.13 Display of accumulated "ml inf." since last power up (balance)


This function (accessible in stop- and run-mode) is only available if it was enabled by
the technical service
1. Press this key until the
function "InF" appears in
the left display!

2. The total infused volume since last


power up of the pump is displayed!
If no key is pressed within 5 seconds
(programmable) the display changes
back to the basic position.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 26


4.14 Data-lock (keyboard lock)
All input keys can be locked with this function.
This function (accessible in stop-mode) is only available if it was enabled by the
technical service.
1. Press this key until the function
"dLo" appears in the left display!

2. Use this key to switch the data lock


ON or OFF (toggle function)!
If no key is pressed within 5 sec-
onds (programmable) the display
changes back to the basic position.

4.15 Setting the stand-by alarm time


With this function you can set an individual stand-by alarm time longer then 2 minutes
(default value). This function (accessible in stop-mode) is only available if it was enabled
by the technical service.
1. Press this key until the function
"Stb" appears in the left display!

Hours Minutes

2. Input the desired stand by time!


(>2 minutes, e.g. 2 hours 30 minutes)
The entered time is activated as soon
as the displays change back to the
basic position. During that time, no
stand-by alarm is released. 10 h 1h 10 min 1 min

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 27


4.16 Input of medication name
With this function you can select a medication name from a pre-programmed list which
will be displayed. This function (accessible in stop-mode) is only available if it was ena-
bled by the technical service. If no key is pressed within 5 seconds (programmable) the
display changes back to the basic position!
1. Press this key until the function
"MEd" appears in the left display!

2.Use this key to scroll the pre-programmed list and to


select the desired medication name!
If no key is pressed within 5 seconds (programmable)
the displays change back to the basic position!

4.17 Timer alarm


With this function you can set an individual timer alarm. After alarm the pump will stop!
This function (accessible in stop-mode) is only available if it was enabled by the
technical service.
1. Press this key until the
function "tM" appears in
the left display!

3. The message: "tM" "End" ap-


pears with an acoustic alarm if
2. Input the desired alarm time! the timer has run out!
The entered time is activated Press this key to acknowledge
as soon as the displays change the message. The displays
back to the basic position. changes back to the basic po-
sition.

The remaining time can be checked by selecting the function timer "tM".
The timer runs in stop and run mode.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 28


4.18 Air Bubble Detection

Pumps with a standard factory setting detect air bubbles featuring a volume of 250 µl (~
0.247 ml or 35 mm length) upwards and trigger off a specific air-bubble alarm. Your tech-
nical department can adjust this value by means of the configuration menu to within a
range from 50 to 1,000 µl.

Depending on the operating range of the device, it is also possible – again by means of
the configuration menu – to switch on an additional function for monitoring the accumula-
tion of small air bubbles (the volume of the bubbles can be set to anywhere between 100
and 2,000 µl) over a given period of time (settable between 8 and 64 minutes). Setting
this function is recommended in particular for solutions that tend to foam.

The size of the air bubbles to be detected by the pump shall be defined by the medical
personnel in charge.

The following precautions should be observed in order to minimise the risk of air
bubbles forming:

1. The drop chamber must always be at least one third full to a maximum of half full. When
high flow rates are used, it should be half full.

2. The tube must be filled completely and checked before connection to the patient; any air
bubbles must be removed.

3. Y fittings must be properly sealed and 3-way valves must be set correctly.

4. Solutions that have to be stored in a refrigerated condition must be warmed at least to


room temperature before use.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 29


5 Safety information

5.1 Risk and danger


Caution! The infusion pump must only be used under the supervision of qualified
clinical or nursing staff.
Caution! The ARGUS 707 V pump was designed and manufactured to be used only
as an IV infusion pump.
Caution! The ARGUS 707 V infusion pump may only be used with spare parts, ac-
cessories, consumables and sets with Luer-Lock connections (specified in
the appendix) recommended by CODAN ARGUS AG. The functional safety
of the pump is not guaranteed if non approved materials are used. The
safety of the patient may be endangered.
Caution! The connection of several infusion types (gravity, syringe pumps, peristaltic
pumps, etc.) together into the same tube can be very dangerous. This type
of connection should only be used if expressively specified in the operation
manual of each device and/or approved by a notified body and applied un-
der a trained qualified clinical or nursing staff.
Caution! Keep the ARGUS 707 V pump clean and dry. If inadvertently liquid is
poured over the unit, immediately remove the AC power cord of the unit or
the ARGUS Docking Station and contact the relevant hospital department
for cleaning and drying.
Caution! The ARGUS 707 V must not be operated in hazardous locations and explo-
sive gaze environment.
Caution! After each falling the pump has to be tested by the tech. service.
Caution! The ARGUS 707 V pump has to be connected to mains supplies in accord-
ance to the limits specified in chapter 9. The pump can only completely
be disconnected from the mains by removing the mains power cord.
Caution! No interferences by external high frequency electromagnetic fields (e.g. in
combination with surgical equipment) are known that could influence the
safe operation of the pump. In case of doubt we suggest that you contact
your local distributor.
Caution! Be aware that a free-flow and/or under-flow will not be detected when the
pump is operated without a drop detector!
Furthermore the following security steps have to be taken:
You must install the roller clamp on the patient side of the pump.
An infusion volume (ml total smaller than the content of the bottle) must
be entered.
Caution! The pump must not be operated with a defective battery pack. Pumps that
do not have a functioning battery pack may present a danger to patients
because the pump will turn off immediately and without prior warning in the
case of a power failure. For safety reasons the pump should only be fitted
with battery packs provided by CODAN ARGUS AG.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 30


Emboli: To avoid this risks flush the IV set and the extension line before use, make
sure no air bubbles remain in the entire system!
Pulmonary
Oedema: An excessive or to rapid infusion may endanger the patient or cause death!

5.2 Safety standard checks (SSC)


The maintenance safety standard check has to be performed at least every 24 months or
after maximum 10’000 hours of operation. The checks have to be performed according to
the description available in the service manual.

6 Cleaning / Disinfection

6.1 General references


Caution! The pump has to be switched off and must be unplugged from the power
line before cleaning! Remove all connections and cables.
Caution! It is strictly forbidden to autoclave the ARGUS infusion pump or to dip it in-
to liquid.
Caution! Take care that no liquid gets into the unit or the plugs.
Caution! Keep the ARGUS 707 V pump clean and dry. If inadvertently liquid is pour-
ed over the unit, immediately remove the AC power cord of the unit or the
ARGUS Docking Station and contact the relevant hospital department for
cleaning and drying.

The pump must only be cleaned by "swabbing". Only alcoholic disinfectants may be used.

Caution! Do not use scrubbing agents for cleaning!

In order to keep the pump full operational, we recommend regular cleaning. Use a cloth
moistened with lukewarm water for cleaning. Alcoholic cleaning agents must only be us-
ed diluted.

For more information regarding the supply of suitable cleaning agents and disinfectants
please contact the specialists in your house.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 31


7 Warranty

7.1 Warranty duration


The warranty period is determined by the distributor and is subject to its general condi-
tions of sale. The warranty covers the repair and replacement of defective parts in case
of manufacturing or material faults.

7.2 Warranty limitations


The warranty terminates in the event of modifications or repairs carried out by non-
authorized persons and in case of non-adherence to the inspection /maintenance inter-
vals.
The warranty does not include the batteries, failures that are due to wrong manipulation,
inexpert handling, liquid ingress or normal wear and tear.

The supplier assumes responsibility for the safety, reliability and performance of the unit
only if all following conditions are met:

a) Exclusively authorized persons have carried out the assembly, additions, readjust-
ments, modifications or repairs.
b) The electrical installations in the room where the unit is operated meets the require-
ments of the IEC regulations.
c) The unit has been operated in compliance with the instructions for use.

Caution! The ARGUS 707 V infusion pump may only be used with spare parts, ac-
cessories, consumables and sets with Luer-Lock connections (specified
in the appendix) recommended by CODAN ARGUS AG. The functional
safety of the pump is not guaranteed if non approved materials are used.
The safety of the patient may be endangered.

This manual contains the latest data available. It is subject to further modifications in ac-
cordance with technical improvements.

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 32


8 Accessories
Bottle holder Drop detector Combi clamp
(45cm / 60 cm) (basic)
REF 11.005 / 11.043 REF 10.089 REF 10.087

Power distributor Barcode Reader Combi clamp AMService Utility


with holder (upgrade kit) Software
REF 90.009 REF 90.151 REF10.108 -10.111

ARGUS IV-stands Docking Station Transport Unit


various REF 90.100 REF 90.052

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 33


9 Specifications
ARGUS 707 V
Designation Volumetric peristaltic infusion pump ARGUS 707 V
Order number 18.1110 (230V AC) / 18.1111 (115V AC)

CONFORMITY CE-marked (CE 0120), Directive 93/42/EEC Appendix ll


IP-protection against liquid ingress IPX2 (protected against dripping, 15° tilte d)
Applied part Type CF
Protection class II
Medical device classification: IIb
Regulations & Electrical safety EN 60601-1-1, EN 60601-1-4, EN 60601-2-24
Electromagnetic compatibility EN 61000-3-2, EN 61000-3-3, EN 60601-1-2
Certification ISO 13485, ISO 9001

INPUT
Volume / Rate Infusion rate 0.1 - 999.9 ml/h (in 0.1-step up to 999.9 ml/h)
Volume total (VTBI) 0.1 - 9999 ml
(in 0.1-step up to 999.9 ml and in 1-step from 1000 to 9999 ml)
Set Fill (prime) rate 1 - 999 ml/h (in 1-step up to 999 ml/h)
Rate calculation Volume total (VTBI) and infusion delivery time
Bolus Bolus rate 0.1 - 1200 ml/h programmable without flow interruption
(in 0.1-step up to 999.9 ml/h and in 1-step from 1000 to 1200 ml/h)
Bolus volume 0.1 - 999 ml programmable without flow interruption
(automatic & manual) (in 0.1-step up to 999 ml)
Time Infusion delivery time 1 min - 99 h 59 min (in 1 min-step up to 99:59 h)
KVO KVO rate (KOR) 0.1 - 3 ml/h (depending on entered infusion rate)

ACCURACY
Rate Flow rate deviation ≤ ± 5%
• For rates from 1 to 999.9 ml/h
• IV-set changed every 24 hours or after 2.5 liters infused
• Maximum back pressure +/- 100 mmHg
• Depends on the infusion set used
Flow discrepancy in the event
of a technical failure ≤ ± 10%
Tech. Technical deviation < 1%

OPERATING REQUIREMENT
Temperature range 5 °C -40 °C
Medication temperature 18 °C -30 °C
Storage temperature range 0 °C -40 °C
Relative humidity (permissible) 20 -90 %; no condensation

POWER SUPPLY
Battery type NiMH- 12 V / 1.5 Ah (maintenance-free)
Battery operation 5 h @ 25 ml/h
Battery charging time 16 h
External DC power supply 20 V / 0.3 A
Power consumption max. 12 VA
Line fuse 125 mAT
AC power supply 230 V +/- 10%, 50 - 60Hz
Optional AC power supply 115 V +/- 10%, 50 - 60Hz

INTERFACE
Data interface 2 x RS-232 (1 x galvanic isolated)

DISPLAY
LED 1 large 4-digit LED-display (left) Infusion rate, additional information
1 large 4-digit LED-display (right) Volume infused (0.1- 9999ml), volume total (VTBI),
infused time (1 min – 99 h 59 min), additional information
Pictograms and LED'S Operation and alarm conditions
LED bar graph Pressure monitoring display

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 34


ALARM
Acoustic alarm Volume adjustable in 6 steps (cannot be completely turned off)
Pre alarm Battery near empty ca. 15 minutes before infusion stop
Battery depleted ca. 6 minutes before power turn-off
Occlusion Patient side
Bottle side
Volume Total (VTBI) reached
Empty Infusion bottle empty
Drop Deviation too big
Air bubble
Door open
KVO Reminder alarm (KOR)
Service Suggestion
Failure Technical

SAFETY & INFORMATION


Air bubble Air bubble detector (single bubble) 50 - 1000 µl (programmable)
Air bubble accumulation 100 - 2000 µl (50 µl steps) within 8 -64 min (8 min-steps), config.
Drop detector Configurable drop window 10 - 65 drops/ml
Occlusion Pressure reduction YES
Pressure limit (adjustable) 100 - 1000 mbar (10 - 100 kPa, 75 - 750 mmHg)
in 10 steps programmable without flow interruption
Alarm reaction time
and related bolus vol. See separate table below
IV set Approved sets See appendix
Nurse call Staff-alerting system 24 V / 0.2 A (potential free change over contact; static/dynamic)
History 400 events
Installation Horizontal
Mounting possibilities - Table top
- IV-stand / ceiling pole (up to diameter 38mm)
- Standard rail system (optional accessory)
- ARGUS Docking Station A60/100
Storage time 3 months recharge the battery after each storage time or at latest all 3 months
to maintain the battery capacity specified
Transport In original packaging
Disposal Recyclable

MEASUREMENT / MATERIAL
Dimensions 190 x 160 x 130 mm (W x H x D) excluded combi clamp
Weight 2 kg including battery (without accessories)
Housing ASA (high performance plastic)

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 35


Typical alarmdelay at occlusion volumetric infusion pump A707

10000

1000

100
Alarmdelay [s]

Measured values
Measured values
PrL = 700 mbar Typical values
Typical values
10

1
PrL = 200 mbar

0.1
0 100 200 300 400 500 600 700
Rate [ml/h]

Bolus volume after occlusion (typical values) of the A707 volumetric pump

600

500

400
Bolus volume in [µl]

300

200

Important remark:
100 Bolus volume with automatic pressure release: typical 100 µl !

0
1 2 3 4 5 6 7 8 9 10
Pressure limit steps in 100 [mbar]

Test I in the first two hours of the test period of 24 hours at 1.0 ml/h (initial period)
Trumpet graph Start-up graph

15 2.0
1.8
Percentage error(%)

10 Ep(max) 1.6
1.4
Flow(ml/h)

5 1.2
1.0
0 0.8
-5 Ep(min) 0.6
Overall percentage erro r -2.09% 0.4 Set rate 1.0 ml/h
-10 0.2
0.0
-15 -0.2
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 0 10 20 30 40 50 60 70 80 90 100 110 120
Observation window(min) Time(min)

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 36


Test II in the last two hours of the test period of 24 hours at 1.0 ml/h (end period)

Trumpet graph

15

Percentage error(%)
10
Ep(max)
5
0
-5 Ep(min)
Overall percentage erro r -0.1%
-10
-15
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30
Observation window(min)

Test I in the first two hours of the test period of 24 hours at 25 ml/h (initial period)
Trumpet graph Start-up graph

15 50.0
45.0
Percentage error(%)

10 40.0
Ep(max) 35.0
Flow(ml/h)

5 30.0
25.0
0 20.0
-5 Ep(min)
15.0
10.0 Set rate 25 ml/h
Overall percentage erro r -1.19% 5.0
-10
0.0
-15 -5.0
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 0 10 20 30 40 50 60 70 80 90 100 110 120
Observation window(min) Time(min)

Test II in the last two hours of the test period of 24 hours at 25 ml/h (end period)
Trumpet graph

15
Percentage error(%)

10
Ep(max)
5
0
-5 Ep(min)
Overall percentage erro r -3.00%
-10
-15
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30
Observation window(min)

All measurements are done under laboratory conditions!

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 37


Appendix: Recommended IV sets
Caution! The ARGUS 707 V infusion pump may only be used with the recommended
sets listed below. The sets must have Luer-Lock connections. The function-
al safety of the pump is not guaranteed if non approved materials are used.
The safety of the patient may be endangered.

Caution! Each time you change set brand (manufacturer) or tube material you have
to perform a new set calibration (contact your service department).

Caution! Unless otherwise specified by the customer, the ARGUS 707 V has been
calibrated with the CODAN L86 Art. 43.3030 (NoDEHP) infusion set.

Manufacturer IV set Order Number


CODAN V86-P / S86-P / L86-P (NoDEHP) various
green line V86 (NO PVC) 43.4825
CYTO-Z various

B.Braun Intrafix Air P 406 2990


Intrafix Primeline Comfort 406 2981L
Intrafix Primeline Classic 406 2957
Intrafix Safeset 406 3000
Fresenius Infudrop-Air PD 288 63 51
Becton Dickinson R87 P 3963.50
Becton Dickinson R87 P with backflow-stop 3963.53

The accuracy of ± 5 % cannot be guaranteed for infusion rates > 600 ml/h for the
following IV set:

Clinico Perfudrop Air P 484 036 08


CODAN V86-I.V.STAR 10 (NoDEHP) 43.4401

14.204.J_A707 en V4.4X.docx CODAN ARGUS AG 05.08.2014/ VZ 38

You might also like