The demand for confirming the measurement quality 3D laser scanners that are exposed to such
of 3D laser scanners is increasing. More and more conditions should be checked periodically to ensure
companies or contractors are ISO certified and need that the measurement results meet the equipment
to confirm the accuracy of their measurement specifications.
equipment periodically to conform with regulations.
In a world’s first, Leica Geosystems’ Check & Adjust
Leica Geosystems 3D laser scanners are is an easy-to-use field procedure for checking and
manufactured, assembled and calibrated to the best updating 3D laser scanner system parameters. This
possible quality. Nevertheless, extreme temperature improves system reliability and accuracy, helps fulfil
changes and hard shocks can sometimes cause the ISO standards and complete the work with
deviations that may influence the system accuracy. greater confidence.
Angular Tilt Range
Parameters Compensator Offset
The Angular Parameter Check The Tilt Compensator Check Set Range Offset allows updating of
determines vertical offset, deviations determines tilt indices by turning the the range offset according to results
of axes and laser beam direction. instrument around its standing axis. of reliable baseline measurements.
For the angular parameter check, at The tilt compensator check requires a To check range offset, a standard
least five different targets in a levelled scanner set up on stable procedure is used which can be done
configuration pre-defined by Leica ground. This quick and efficient check on any suitable reference baseline of
Geosystems need to be scanned in can be performed as often as needed. known distance.
both faces from two different stations.
It is recommended that users set up a Once the check is completed, tilt Once the range offset is checked the
fixed installation of targets, or contact indices are displayed and user can apply a recommended
Leica Geosystems to organise a check recommended adjustments can then adjustment for all future range
at a nearby facility. be applied. measurements.