66 User Manual 2
66 User Manual 2
This USER MANUAL, contains all the information you need in order
to safely operate your GETINGE STERILIZATION AB equipment.
Please read and follow the entire manual carefully, and please pay
special attention to the framed warnings and caution remarks. See
“Attention symbols” under SAFETY
In the event of an accident or incident you are obliged to report this. See
“Reporting an accident or incident” under SAFETY
GETINGE STERILIZATION AB shall not be liable for any damages,
directly or indirectly, due to actions, which are not in accordance with
this manual.
By following this manual you are ensured quality equipment for many
years ahead. Should any unforeseen problems occur, you are of course
welcome to contact our local representative.
Manufacturer address:
Getinge Sterilization AB
Ekebergsvägen 26
305 75 Getinge
Sweden
Fax +46(0) 35 54952
1
2
SAFETY
Attention symbols
Some of the warnings, instructions and advice in this manual are so
important that we use the following special symbols to draw attention
to them. The designs and symbols used are:
Warnings
Instructions
Advice
SAFETY - 3
Symbols on the unit
Hot surface
Electrical danger
Other dangers
4 - SAFETY
Reporting an accident or incident
SAFETY - 5
6 - SAFETY
INTRODUCTION
HS 66
HS 66 is the designation of a series of GETINGE high-pressure auto-
claves with vertically sliding automatic doors, all with the door opening
660 x 660 mm
There are both stand-alone and built-in versions.
INTRODUCTION - 7
European standard EN 285 (Australia/New Zeeland, Standard AS 1410)
specifies the use of fixed programs only. However, the control equip-
ment of HS 66-autoclaves allows the programs to be made variable dur-
ing the time testing is in progress.
As not only the processes but door operation too are controlled automat-
ically, the step to full automation of the handling of sterile items is not
a long one. The control unit is therefore prepared for controlling an
autoclave having a GETINGE ASF automatic loading and unloading
unit.
As with other GETINGE double-jacketed autoclaves, the external sur-
face of the chamber is more than half covered by all-welded U-sec-
tions.. These stiffen the flat walls of the pressure vessel, and at the same
time accommodate the steam which keeps the chamber walls warm in
order to minimize condensation in the chamber. The design principle
makes all welds accessible for visual inspection after the thermal insu-
lation has been removed.
A vacuum pump of the water-ring type removes air, steam and conden-
sation from the chamber and door-gasket groove. The sealing water of
the pump is taken from an open tank which in some cases also supplies
the feedwater pump, on autoclaves with a steam generator, with water.
The condenser between the chamber and vacuum pump converts waste
steam to water. This reduction of the pump handling volume contributes
greatly to improving pump efficiency and protecting the pump and
waste pipes against high temperatures.
A well-damped discharge system after the vacuum pump and well-
dimensioned thermal insulation around the chamber mean that the auto-
clave affects the surroundings to an insignificant extent and is therefore
easy to position.
HS 66-autoclaves have to be supplied with electricity, cold water, com-
pressed air and steam. When steam is supplied from a central steam-pro-
ducing plant, its distribution system is required to be correctly pressure-
reduced, protected by a safety valve and produce a quality of steam
which is suitable in the context of sterilization. (For some versions the
pressure reducer is already fitted)
If the steam is not suitable in the context of sterilization, a steam gener-
ator can be installed inside the trim plates on all models. The task of the
8 - INTRODUCTION
steam generator is to generate pure steam using the centrally produced
steam.
If there is no steam network or there are reasons for not using the steam
network, an electric steam generator may be installed inside the trim
plates on all models.
Safety components
Every unit is equipped with a number of components with the specific
purpose of ensuring the safety of personnel. These items are marked
with a warning triangle below in the following documents:
• electrical diagrams
• pipework diagrams
• spare parts lists
INTRODUCTION - 9
Doors
The doors are closed and locked while the process is running and remain
locked in the event of media loss. The doors cannot be opened until the
pressure in the chamber has been equalized with the ambient pressure,
not even if fault management gives an opening command. The sliding
doors are operated by a compressed air cylinder which has a severely
throttled inlet. A drain valve with a considerably greater flow capacity
than the inlet is located in the bottom of the cylinder, which is pointing
upwards. Objects that obstruct the upward motion of the door press
down a metal plate which opens the drain valve, causing the door
motion to stop. This prevents operator injuries and material damage. A
support latch holds the door so that it does not open if the pneumatic
operating cylinder loses pressure. When the pressure of the compressed
air is so low that the door cylinder cannot support the weight of the door,
the air is not capable of pulling away the support latch.
Emergency stop
In the front panel, at the side of the door, there is a pushbutton with these
functions:
A On door operation.
to stop the door motion immediately during door operation. Press-
ing the button triggers an alarm. This alarm stops the door immedi-
ately:
B When the door is closed
The current process is aborted and all valves for media to the cham-
ber are closed. This also triggers an alarm.
Trim plates
Panels are possible to dismount by means of tools or keys. The panels
form a barrier for operators, but not for specially trained technicians.
Stainless steel surfaces within reach of the operating personnel are insu-
lated and cooled to a harmless temperature.
Valves
If the equipment has electrically and/or pneumatically actuated valves
they are normally made to close by means of spring force. This prevents
10 - INTRODUCTION
undesired flow through the valves in case of power failure. A fail safe
function is achieved.
Pressure vessels
The sterilizer chamber jacket and steam generator are pressure vessels,
designed and built in accordance with standards laid down by official
bodies charged with monitoring this area of safety.
The internal steam supply to the sterilizer is protected against excessive
supply pressure by a safety on the steam generator, the sterilizer or both.
The connected supply lines from external steam, water or compressed
air networks must be protected against excessive pressure by safety
valves. The sterilizer supplier is not responsible for these valves. Per-
mitted supply media pressures are stated in “Utility data”.
Safety valves
Safety valves limit the system pressure, providing the last line of pro-
tection to prevent the design pressure of the vessel from being exceeded.
Safety valves are required to be inspected at prescribed intervals.
Operation
The heating power is divided among a number of groups of elements.
The elements are controlled by a pressure switch with double changeo-
ver contacts, one for each heater element circuit. One group of contacts
limits the operating pressure of the steam generator. The other controls
reconnection of the power when the minimum pressure is reached.
Pressures higher than those stated on the electrical diagram should not
be set, since the steam will be superheated when admitted to the steri-
lizer chamber. On the other hand, if the operating pressure is allowed to
fall too far, the steam tends to become wet, which is undesirable for ster-
ilisation purposes.
INTRODUCTION - 11
A feedwater pump supplies the steam generator with water from an inte-
gral water tank with air trap. The water level of the steam generator is
sensed by a device which controls the operation of the pump so that the
level is kept almost constant. Boil-dry protection cuts off the supply of
electric power to the elements when the water level is low.
See also “Steam generator settings and controls” under Advice and
instructions in the TECHNICAL MANUAL.
Control switch
The steam generator and sterilizer are controlled by the same switch. On
sterilizers with med double steam systems there is an extra shutoff but-
ton marked Steam generator.
Spark-reduction connection
As well as shortening the life of electrical contacts, sparking at the con-
tacts may also cause interference with nearby electronic equipment.
The anti-spark protection is based on fast-acting auxiliary relays taking
over the load from the slow-breaking contacts of the pressure switch for
operating pressure, so largely eliminated sparking.
12 - INTRODUCTION
The task of the level device is to sense and control the water level of the
steam generator. The main components are a sensitive relay and two
electrodes, the ends of which touch the water surface at the upper and
lower fill levels. Contact of the upper electrode with the water stops the
feedwater pump, whilst absence of contact at the lower electrode starts
the pump. A third, lowest, electrode cuts off the electric power supply
to the heating elements when the electrode loses contact with water; see
boil-dry protection below.
The type and sensitivity of the level relays can be adapted to varying
water conductivity. If the feedwater is de-ionised, a high-sensitivity
level relay is required and the sensitivity setting must be increased as the
conductivity of the water decreases.
The sensitivity of the level relay is automatically adjusted at power up
to match varying conductivity.
The distance between the electrodes is another factor that affects the
overall sensitivity of the level control system.
Version with float
The GETINGE float-based level control system is intended for water
with a conductivity of 0 to 2 μS/cm, but it can also be used for higher
conductivity.
The task of the level device is to sense and control the water level in the
steam generator. The main components are a float with two level con-
tacts and two relays. The level contacts change over at the upper and
lower fill levels. The upper contact stops and starts the feedwater pump,
whilst the lower contact cuts off the electric power supply to the heating
elements; see boil-dry protection below. The purpose of the relays is to
protect the float contacts from harmful currents.
Water-level monitoring
A lamp in the “reset level” button on the steam generator control panel
is lit when the water level is correct. On some models, the water level
can also be read off on a sight glass. The sight glass has two valves
which can be closed to prevent loss of water and steam in the event of
breakage of the sight glass.
Boil-dry protection
If the water filling fails, the boil-dry protection cuts off the electric
power supply after a certain time, when the water level has fallen below
a minimum limit.
Overheat protection
This protection prevents the steam generator from overheating if the
boil-dry protection should fail to operate.
INTRODUCTION - 13
Overfilling protection
The timer relay prevents the steam generator over-filling with feedwater
if the level control system stops working.
RCD
Some types of steam generator must be connected to the electric power
supply via a residual current device (RCD). See the Installation instruc-
tions chapter in the INSTALLATION MANUAL.
Safety switch
It must be possible to isolate all electric power supplies to the steam
generator with a safety switch. The safety switch must be lockable in the
OFF position. See also chapter Installation in the INSTALLATION
MANUAL.
14 - INTRODUCTION
THE PROCESS
A Pre-treatment
B Sterilizing
C Post-treatment
A B C
Pre-treatment
It has been found that moisture is an essential element in achieving ste-
rility with steam. It is therefore important that the steam comes into
close contact with the micro-organisms to be killed.
A pre-treatment phase consisting of a number of pressure variations fol-
lowing a certain pattern effectively removes air from various types of
goods and produces the moisture required in the subsequent sterilizing
phase. Depending on the equipment of the sterilizer, steam may pass
through before the evacuations.
Air removal when autoclaving liquids in open containers is by applying
a vacuum or by flowing steam, to ensure that the liquid does not boil
during the decompression periods in a pulsating procedure.
Sterilizing
For sterilizers equipped with a computerized control unit, the sterilizing
time starts at the instant in the pre-treatment phase of the process when
the chamber temperature sensor signals a temperature equal to or higher
than the sterilizing temperature specified for the current program. An
alarm sounds if the temperature sensor registers a value outside the tem-
perature band, or if the temperature differs by more than the permitted
amount from the chosen sterilizing temperature.
THE PROCESS - 15
When sterilizing liquids, the temperature is measured with a tempera-
ture sensor in the load. The condition for countdown of sterilizing time
is that both temperature sensors in the chamber and the sensor in the
load indicate the programmed sterilizing temperature.
Post-treatment
The purpose of the post-treatment is to normalize the temperature and
moisture content of the goods. All goods except liquids are therefore
exposed for a certain time to a vacuum below 70 mbar (a). After such
post-treatment. textiles may appear to increase in weight by about 1%.
This represents the normal addition of water that occurs during steam
autoclaving process.
Post-treatment of liquids that have been sterilized in open or half-closed
containers consists of a self-cooling period. During this period the pres-
sure and temperature are lowered very slowly until the temperature is
well below the boiling point of the liquid. This process is speeded up by
applying a slight vacuum to the chamber when its pressure approaches
atmospheric.
Process adaptation
A large number of factors such as type of goods, goods carrier, packag-
ing material, etc. affect the result of the process. The control equipment
offers a number of features that can be used to optimize the processes to
suit the requirements of different customers. For further information,
see "User-programmable functions" in the TECHNICAL MANUAL.
16 - THE PROCESS
Maintenance program
Leak test
Fully automatic process for steam autoclaves
With autoclaves equipped with an analogue pressure transducer the leak
test can easily be automatically performed. The leak test is to be per-
formed with an empty chamber. The chamber additionally has to have
been warm, for example as a result of a heating program having been
run beforehand.
The checking process has its own program button (see program list)
which starts with the vacuum system evacuating the chamber. Depend-
ing on the equipment of the sterilizer, steam may pass through before
the evacuation. The pump in the vacuum system stops when a deep vac-
uum has been created. For a short period of time after this evacuation a
slight rise in the chamber pressure takes place which is not due to leak-
age but results from evaporation of condensate and temperature-volume
changes in the recently rarefied residual steam.
It is therefore not possible to establish that an increase in pressure in the
chamber is caused by leakage until conditions in the chamber have sta-
bilized. It is not until about 10 minutes after the pump has stopped that
the pressure and time start to be measured.
A satisfactory leak test allows for a maximum permissible pressure rise
of 13 mbar / 10 minutes.
With double-ended autoclaves, the loading door opens after a com-
pleted leak test either automatically or manually, depending on how it
has been programmed.
THE PROCESS - 17
See Components in the TECHNICAL MANUAL for printer settings
and maintenance.
The system stores the last process so that an “emergency printout” can
be produced afterwards if there is a problem with transmission of the
printout or with the printer itself.
If the system is fitted with a SUPERVISOR, the printer is connected to
that. Otherwise it is normally connected to the control system. SUPER-
VISOR is an independent measuring system that records the process
data of the sterilizer. See also Control unit in the TECHNICAL MAN-
UAL for further information.
13:47 29 06 04
SIGNATURE ……… 6
..
PROCESS OK 5
00:24:58 71.2 132.5 85.2 0.995
PROCESS COMPLETE
00:23:39 75.5 132.5 85.2 0.039
EQUALIZATION
00:18:35 51.1 132.2 88.6 0.096
00:16:14 134.9 135.4 134.8 3.076
DRYING
HIGHEST TEMP 135.2C
LOWEST TEMP 134.9C
134.9C
00:16:08 135.0 135.5 134.9 3.094 4
POST-TREATMENT
00:12:07 134.1 134.3 134.0 3.039
STERILIZING
00:09:10 106.5 132.6 103.6 1.088
18 - THE PROCESS
HEATING
00:08:50 115.8 132.5 115.6 1.880
00:08:33 107.8 132.6 104.3 1.107
POS PULSING
00:08:14 115.9 132.5 115.7 1.876
00:07:57 107.2 132.6 103.1 1.103
POS PULSING
00:07:38 116.0 132.6 115.7 1.880
00:07:21 106.0 132.6 100.1 1.119
POS PULSING
00:07:02 116.1 132.2 114.9 1.877
00:06:44 104.5 132.4 77.2 1.090
POS PULSING
00:06:24 116.6 132.3 87.4 1.875
00:05:52 92.0 132.5 34.0 0.835
POS PULSING
00:05:32 78.4 132.7 33.7 0.065
00:04:22 91.3 132.5 33.9 0.836
NEG PULSING
00:04:01 75.1 132.5 33.4 0.066
00:02:54 81.4 132.5 33.7 0.849 4
NEG PULSING
00:02:37 81.9 132.6 33.4 0.099
00:00:15 46.6 132.6 33.4 0.997
NEG PULSING
00:00:00 46.5 132.5 33.4 0.997
START
PROG. TIME AI00 AI01 AI02 AI03
PROGRAM: P1 POROUS LOAD 134
AI03 CHAMBER PRESSURE
AI02 AIR DETECTOR
THE PROCESS - 19
AI01 JACKET TEMP 3
A100 DRAIN TEMP
SIGNALS
POS. PULSE 5
NEG. PULSE 3
PARAMETER SET-
TINGS
CYCLE COUNTER :16
MACHINE NAME :HS55 1
PROCESS START 13:20:45
DATE 29/06/2004
1. The lower part shows the date and time when the process was
started and the type designation, number and cycle counter of the
sterilizer.
2. Above the heading “Parameter settings” there is a list of the param-
eters of the current program that can be changed with a parameter
code. In this example, sterilizing temperature in °C and times for
various sub-processes in hours, minutes and seconds.
3. Above the heading “Signals” there is a list of the parameters chosen
for printing.
4. Information about which program has been started, followed by
process logging.
5. Printed out if an error occurs during the process.
6. Signature line
20 - THE PROCESS
See Components in the TECHNICAL MANUAL for printer settings
and maintenance.
Printout is possible in four modes. The mode shown below is preset
before delivery and is recommended by GETINGE.
The system stores the last process so that an “emergency printout” can
be produced afterwards if there is a problem with transmission of the
printout or with the printer itself.
If the system is fitted with a PACS SUPERVISOR, the printer is con-
nected to that. Otherwise it is normally connected to the control system.
PACS SUPERVISOR is an independent measuring system that records
the process data of the sterilizer. See also Control unit in the TECHNI-
CAL MANUAL for further information.
POS. PULSE 5
STERILIZING TEMP134.0 C
STERLIZING TIME 00:07:00
DRYING TIME 00:05:00
DRYING STEAMPULSE 00:00:00
DRYING AIRPULSE00:00:00
SIGNATURE:__________________________________ 6
Page: 3
DATE : 16/05/2001 PARAMETERS
THE PROCESS - 21
PROCESS START : 08:01:36 NEG. PULSE 3
AUTOCLAVE NAME : HS6613-2 POS. PULSE 5
AUTOCLAVE NUMBER :1 STERILIZING TEMP134.0 C
CYCLE COUNTER :5 STERLIZING TIME 00:07:00
DRYING TIME 00:05:00
DRYING STEAMPULSE 00:00:00
DRYING AIRPULSE00:00:00
1 In the top part, the date and time when the process was
started and the type designation, number and cycle coun-
ter of the sterilizer are shown.
2 Parameters of the current program that can be changed
with a parameter code are listed under PARAMETERS.
In this example, sterilizing temperature in °C and times
for various sub-processes in hours, minutes and seconds.
3 The parameters chosen for printing are listed under
SIGNALS.
4 Information about which program has been started, fol-
lowed by process logging.
5 Printed out if an error occurs during the process.
6 These lines are printed out after a faulty process. Identi-
fication code of the person who entered the password and
a line for a signature.
7 Graphical representation of the process
22 - THE PROCESS
INSTRUMENTS
Loading side
3 4
1 6
Door-blocking key
See chapter Introduction, “Safety devices, an overview”.
Emergency stop
See chapter Introduction, “Safety devices, an overview”.
INSTRUMENTS - 23
Unloading side
2 1
Recorder
See recorder operating instructions in the TECHNICAL MANUAL
chapter Components.
Emergency stop
See chapter Introduction, “Safety devices, an overview”.
24 - INSTRUMENTS
OPERATION
Program start
Where applicable, the operator must state process parameters and other
information requested by the control system. On equipment where an
authorisation code for programs not intended for routine sterilization is
activated, the operator must also enter an authorisation code immedi-
ately after startup.
Via operator panel Avanti
To start a cycle:
OPERATION - 25
Authorization protected programs only:
Message
If the operator chooses to attempt to start the sterilizer, even though
starting is blocked, the operator panel display shows a message refer-
ring to an essential condition that is not met.
Alarms - Avanti
If a fault occurs, in standby or during a cycle, the system goes to an
alarm phase. This is an exceptional situation, but it is normally not
dangerous. The current cycle is stopped and output signals from the
control system go to settings that maintain the safety of personnel, the
unit, and the load.
26 - OPERATION
• An alarm signal goes off and the relevant fault code is
displayed on the cycle screen, while the cycle screen’s
background turns red.
• After the alarm signal goes silent, the
Alarm indicator in the status field blinks.
In case of alarm:
Read the error message and correct the fault.
OPERATION - 27
Alarms in standby mode
An alarm from an idle sterilizer prevents it being started.
• Call a technician.
To quick stop:
• Depending on how the cycle is set, the cycle stops
automatically or the cycle stops until the user presses
Start.
28 - OPERATION
• If the unit is equipped with a printer, the alarm text is
also printed out as part of the cycle log.
Call a technician.
In case of a serious fault, always call a technician.
After the technician activates the key switch to the Stepping/Authorized
user setting, the technician has the following alternatives:
• Restart the cycle where it stopped.
• Move the cycle to another phase and start the cycle from there or
move it to the end.
To restart:
• Press Start.
OPERATION - 29
• Press More.
To reset certain types of faults, a technician must first activate the key
switch for Stepping / Authorized users.
Repeated alarms
If the process repeatedly hangs at the same alarm point, call a technician
to step the process to the end. If several faults occur, only the last one is
displayed.
Doors
When a fault has occurred, the unloading door of a double-ended steri-
lizer will not open. Exceptions are errors that occur so late in the process
that unlocking of the door has begun.
The door can only be opened when the correct password has been
entered. For further information, see the menu description in the Con-
trol unit chapter.
30 - OPERATION
NOTE!
Important where a steam generator safety valve has been
blowing off for a long time and the chamber is under pres-
sure at the same time.
Only switch off the sterilizer control switch as a last resort,
because doing so disables certain safety functions.
Instead, call a service technician who can trip the steam
generator circuit-breakers.
Error codes
The error messages listed below are defined in the control system. If
more than one error occurs consecutively, only the error code for the
last error is displayed.
If the sterilizer is equipped with a printer, the error message is printed
out. For secondary errors, all error messages are printed out.
Information about the last twenty faults is saved in the control system
and is accessible in the service menu. If the sterilizer has a printer, this
information can be printed out.
Service codes
When the word “SERVICE” appears on the display, service is required
and must be carried out by a service technician. See below for explana-
tions of the codes.
NOTE!
No error has occurred.
The sterilizer can still be used.
OPERATION - 31
Error code/ Explanation
Message
High pressure The chamber pressure has been too high dur-
ing the sterile phase. This may indicate the
presence of air or non-condensable gases. The
actual pressure in the chamber is compared to
a theoretical saturation pressure calculated
from the sterile temperature + 3°C. A message
is printed in the alarm log when the error
occurs.
Low pressure The chamber pressure has been too low during
the sterile phase. This may indicate a faulty or
improperly calibrated sensor. The actual pres-
sure in the chamber is compared with a theo-
retical saturation pressure calculated from the
sterile temperature. A message is printed in
the alarm log when the error occurs.
HIGH JACKET The jacket temperature has been too high dur-
TEMP ing part of the process.
GASKET Door gasket pressure too low.
FAILURE
DOOR FAILURE Door(s) not closed and interlocked.
PUMP FAILURE Vacuum pump has stopped.
HIGH TEMPER- The chamber temperature has been above the
ATURE permitted limit during the sterilization phase.
32 - OPERATION
Error code Explanation
OPERATION - 33
Error code Explanation
Heating of load
Heating of items to be processed may jeopardie sterilization result due
to poor humidification. Make sure appropriate process has been chosen
before loading the sterilizer.
34 - OPERATION
Fire hazard
There have been reports of textile loads catching fire in the sterilizer
chamber. In all cases this has been due to the load becoming excessively
dry and hot. This can happen in two ways:
• The load has been placed in a heated chamber and left for a long time
without the process being started. Ignition is believed to take place
when the load is moistened again on the admission of steam to the
chamber.
• The load is left in the chamber for a long time without the process
being completed. This probably happens when the process has been
interrupted because of a fault and the load has not been taken out of
the chamber.
Ignition takes place when the process is completed and the load is
exposed to air.
External cleaning.
Clean stainless steel surfaces on the outside of the sterilizer with a
standard household cleaner that does not contain abrasives. Take care
when cleaning painted surfaces, texts, and plastic parts.
OPERATION - 35
The sterilizer should have cooled down before cleaning.
When the key switch cannot be constantly observed, the
person doing the cleaning must always take the key with
them. This is to ensure that no-one mistakenly resets the
key switch.
36 - OPERATION
tion time may be adapted to the performance of the sterilizer but must
not exceed 3.5 minutes.
When the indicator paper has been sufficiently exposed to steam at a
particular temperature, further exposure tends to hide the indication of
any deficiencies. GETINGE’s Bowie Dick programme is therefore
tested with the shortest possible sterilization time so that any deficien-
cies are not concealed by overexposure.
In day-to-day work, the above, standardised test packs for the sake of
convenience are sometimes replaced by various types of disposable
packs which are manufactured by a large number of manufacturers. As
these packs are generally even more sensitive to overexposure than the
standardised test pack, it is extremely important to choose the correct
make. Test packs in accordance with BS 7720:1995 must be used for
GETINGE’s Bowie Dick programme.
Approved consumables
Consumables are used for different purposes as packaging for goods to
be sterilized, for testing, for replacement of sterilizer parts with limited
life or a sterilization media for low-temperature sterilizing. The list of
consumables below is intended as a guide to be used with GETINGE’s
CE-marked sterilizers.
OPERATION - 37
Packaging material for formaldehyde
For formaldehyde sterilizing, we recommend paper/plastic bags
intended for steam sterilizing.
38 - OPERATION
Where relevant, details of concentration, volume, packaging, handling
and properties are described under Operating instructions and under
“General advice on formaldehyde sterilizers”.
Accessories
The sterilizer's CE label makes the manufacturer responsible for
ensuring that its product safety and function satisfies the strict
requirements laid out in the relevant EC directive. Accessories that are
mechanically, electrically or otherwise connected to the sterilizer must
be compatible with this. As a result, only equipment that has been
evaluated by Getinge AB and registered as approved equipment may be
used together with the sterilizer.
Approved accessories
The accessories listed below are compatible with the sterilizer once the
sterilizer has been configured to function with the accessory in question.
Such configuration shall be performed by a professional authorized by
Getinge Sterilization AB.
OPERATION - 39
9 Rails for AGS X X
10 Support bar (Rails for basket X X
11 Loading Platform X X
12 Loading Module X X
13 L-Rails (Extendable Shelves) X X
14 Extendable Shelves (perfo) X X
15 Extendable Shelves (Wire) X X
16 Sterilizer Basket (DIN) X X
(600*300*150/290)
17 600*400*115/150/290) X X
18 Sterilizer Basket (ISO) X X
(100) / (200)
19 Sterilizer Basket (SPRI) X X
(100) / (200)
20 Sterilizer Basket (SPRI- X X
Swedish) (100) / (200)
21 Automatic Loader (Basket) X X
22 Automatic Unloader X X
(Basket)
23 Automatic Loader (Racks) X X
24 Automatic Unloader (Racks) X X
25 Basket Feeder (Manual) X X
26 Loading Trolley Fix Height X X X
27 Loading Trolley Height Adj. X
(Racks)
28 LT FIXED 6606/6613HL/ X
633-07
29 LT FIXED 6610/6620HL/ X
633-10
30 LT FIXED FOR 6613/633- X
13
31 LT FIXED FOR 6617 X
32 LT FIXED SAL FOR 6610/ X
633-10
33 LT FIXED SAL FOR 6613/ X
633-13
40 - OPERATION
34 LT FIXED SAL PD 6610/ X
633-10
35 LT FIXED SAL PD 6613/ X
633-13
36 LT H/A 6610/6620HL/6910/ X X
633-10
37 LT H/A FOR 6613/6913/ X X
633-13
38 LT H/A FOR 6617 X
39 LT H/A SAL 6x10/633-10 X
40 LT H/A SAL 6613/6913/ X X
633-13
41 LT H/A SAL PD 6x10/633- X
10
42 LT H/A SAL PD 6x13/633- X
13
Other accessories.
43 Air Compressor 175 EO-24 X X
44 Air Compressor 55E-24 X X
45 Water Treatment System X X
"RO 51"
46 Water Softener Princess X X
MIDI
47 Water Softener Princess X X
TURBO
48 Water Chiller MINI X X X
49 Water Chiller MIDI X X X
50 Water Chiller MAXI X X X
51 AGS X X
52 NetCOM X X X X X
53 Free standing steam X X X
generator
OPERATION - 41
42 - OPERATION
OPERATION INSTRUCTIONS
Use
OPERATION INSTRUCTIONS - 43
Daily preparations
• Learn the functions of the control buttons and signal lamps by stud-
ying the Instruments and Control unit chapters. Find out how the
user-programmable parameters are set on this unit.
• Open valves for water, air and, where applicable, steam.
• Switch on the mains switch of the unit.
• Run the first steam sterilization process of the day with the chamber
empty. This heats up and drains the system pipework.
Running a process
• Select the desired program. The selected program is indicated on the
control panel.
Parameter password
44 - OPERATION INSTRUCTIONS
If the lamp does not light up, start is prevented for some
reason (also see chapter Operation and Control unit).
Process check
Check that the completed process has been carried out correctly by
comparing the process printout with a corresponding type load printout
from the sterilizer validation.
OPERATION INSTRUCTIONS - 45
Goods handling
Remember that the goods may contain a considerable amount of heat
when unloaded. Position them with regard to a good working environ-
ment.
• Turn off the control power supply by the switch on the front.
• Inspect the strainer in the bottom of the chamber and clean it if nec-
essary.
• Close valves for air, water and, if used, steam and gas.
If necessary:
• Shut down the sterilizer for long enough to let it cool down, eg over-
night.
• Clean the inside of the chamber. Use a chlorine-free cleaner if nec-
essary.
46 - OPERATION INSTRUCTIONS
To interpret the “Finished cycle” indicator
OPERATION INSTRUCTIONS - 47
On power failure or if the control switch or working switch
of the sterilizer had to be turned off during a process, there
is a risk of leakage at the door. Safety measure when there
is positive pressure in the chamber:
KEEP WELL AWAY FROM THE DOOR AND CALL A TECH-
NICIAN!
48 - OPERATION INSTRUCTIONS
CONTROL UNIT PACS 3500
CONTROL UNIT - 49
A number of special terms
STERILIZATION refers to the entire series of treatments that make up
a process aimed at achieving the total killing of all living organisms.
This applies to sterilizers and usually includes air removal, heat treat-
ment and a drying phase.
STERILIZING refers to the actual killing part of the process, the heat
treatment.
On the same basis as the two terms above, STERILIZATION TIME
refers to the duration of the entire process from the start until the objects
can be taken out of the sterilizer. The PROCESS TIME is the same as
the sterilization time.
The STERILIZING TIME is only that part of the process for which the
programmed STERILIZING TEMPERATURE exists in the chamber.
In this context, PARAMETERS means FACTORS THAT INFLU-
ENCE THE sterilization process. Examples of parameters in the sterili-
zation process are temperature, pressure, time, humidity, gas
concentration, etc.
PARAMETER VALUES may be permanently set in the program, be
adjusted by the operator, be included in selectable recipes or down-
loaded from a higher-level system.
50 - CONTROL UNIT
Control panel type Avanti
General description
Screen
The screen can be divided into three areas: status field, middle field and
button field.
Status field
The status field is displayed in all cycle images. The field contains
information about the machine's current status, such as the current cycle
and phase and the time these have lasted. The message is displayed in a
blue box to the left of the indicators.
In the event of an alarm, the status field color changes to red and an error
message displays in a red message box to the left of the indicators.
CONTROL UNIT - 51
Middle field
The middle field is specific for the screen and displays unique functions
and information for the respective menus; see the section Screens in the
Main Menu chapter.
Button field
The button field shows the application buttons that are used to give
commands.
52 - CONTROL UNIT
Buttons and symbols
CONTROL UNIT - 53
Buttons Help No function. For information, see
"User Manual" or "Service
manual."
Cycles Open menu to select cycle.
Note: Not displayed on the
unloading side.
Home Display latest start screen.
Menu
Calibrate
54 - CONTROL UNIT
Close Close machine door.
CONTROL UNIT - 55
Scroll list If there are more rows than space
available on the screen, a scroll list
is displayed on the right. To reach
the row above or below, just drag
the button.
Boot screen
When the panel starts, the screen below displays. The blue bar moves
from left to right. When the panel is ready and the control system is
ready, the panel automatically switches to the pie chart screen.
56 - CONTROL UNIT
Pop-up windows and dialogue windows
Alarm pop-ups
Shown below is an example of an alarm pop-up.
CONTROL UNIT - 57
The window remains on the screen until a selection is made by pressing
the desired button.
Green information window
The window and an acknowledgement button lights up to signal that the
process is ready.
An audible signal sounds.
58 - CONTROL UNIT
Inputting window
Here are two examples of inputting windows.
.
To edit the time, mark the number you wish to edit using the arrow keys
at the bottom. Press the desired number. To save the entry, press Save
at the bottom right.
Different inputting windows exist depending on the input data required.
CONTROL UNIT - 59
Operator menu Avanti
Start screen
The start screen allows access to the main functions for starting a cycle.
Open/Close door.
The door is opened and closed with the Open and Close buttons,
respectively.
Cycle selection
Cycles are selected with the button in the bottom left corner of
the button field.
A list of optional cycles is shown.
60 - CONTROL UNIT
To select a cycle, press the desired cycle. The previous screen is
displayed again.
The selected cycle is displayed in the top left corner of the status field.
Note! The function is not available from the unloading side.
Start cycle
Selected cycles are started with the Start button. If the sterilizer is not
ready to start, a message is shown.
Note! The function is not available from the unloading side.
Various information about the current cycle is displayed depending on
the current start screen. For further information, see the Screens section
in Main menu.
The length of time that the cycle or current phase has been running is
displayed in the status field. Also displayed here is the current cycle
indicator.
Login/logout
Main menu
The Menu button returns you to the main menu.
System menu
About
A list of machine specific data is displayed.
CONTROL UNIT - 61
The following information is displayed:
• Machine name
• Application version
• Panel version
• PACS version
• PROM date
• Serial number
• Total amount of cycles
The pie chart graphically displays the time remaining for the cycle. The
circle diagram also displays the time remaining in digit form.
The following cycle parameters are displayed:
Bar graph
The start screen displays two cycle parameters, as well as a bar graph
for the respective parameters and cycle time remaining.
62 - CONTROL UNIT
The following cycle parameters are displayed:
Plot graph
The start screen displays a plot graph with various cycle parameters.
The status field also shows other cycle parameters. The remaining cycle
time is displayed to the right in the status field.
CONTROL UNIT - 63
In order to enlarge the graph, press Zoom in. When the graph is enlarged
it is possible to navigate by pressing somewhere on the screen and then
dragging with a finger.
Details Graph
The start screen displays a list of parameters that shows information
about the system and the current cycle. The remaining cycle time is
displayed on the top right-hand side, above the list.
Documentation
An pull-down menu with optional printout modes is displayed. The
possibility of choosing printouts varies depending on the user.
The following printout modes exist:
• Print last cycle, latest cycle run.
• Print alarm list, list of the most recent alarms. Password required.
• Print cycle
A list of optional print cycles is displayed. To print, select the print
cycle of your choice and press Printout. Password required.
• Print test page
• Print all cycles. Password required.
• Print machine setup. Password required.
For more detailed information about an alarm, press the alarm you wish
to select.
64 - CONTROL UNIT
The following alarm information is displayed:
CONTROL UNIT - 65
66 - CONTROL UNIT
Program combination 2530290-090
Liquids must not be processed with any of these programs.
Hazardous waste and explosive materials must not be processed in this sterilizer.
RAX2000
RAX2001
RAX2002
Password required!
Parameter Range Delivered Actual
Negative pulses --- 1 1 ................
Positive pulses --- 2 2 ................
1 Sterilizing temperature °C 134 134 ................
2 Sterilizing time min 3 - 90 4 ................
3 Postvacuum time min 0 - 90 3 ................
B3111_Rev 2
P 0 4 Bowie & Dick test
RAX2003
Password required!
Parameter Range Delivered Actual
Negative pulses --- 3 3 ................
Positive pulses --- 5 5 ................
1 Sterilizing temperature °C 121 - 135 134 ................
2 Sterilizing time min 0 - 15 1 ................
3 Postvacuum time min 0 - 90 3 ................
RAX2004
Password required!
Parameter Range Delivered Actual
1 Stabilizing time min 10 - 90 10 ................
2 Test time min 10 10 ................
B3111_Rev 2
Print date 2014-06-30