Quick Start Guide STVI With SMRT
Quick Start Guide STVI With SMRT
* SMRT36 shown with “P” Plus option (Extra Binary Inputs / Outputs and battery simulator)
29 Sep 2012
5 31963
This guide will outline the information that you will need in order to set up and use the STVI hand held
controller with the SMRT Relay Test Systems (SMRT1, SMRT33, SMRT36, SMRT410, and MPRT General
Descriptions are located in the Addendums). We will discuss how the unit operates, some of the
different types of testing you can do and how to save and view your test results with the Smart Touch
View Interface (STVI). It will also cover running the RTMS software using the PowerDB software on your
PC.
On the Test Report you can have your company logo appear in the upper right hand corner. Create a
BMP or JPG file of your logo using a graphics software package such as Adobe Photoshop or Microsoft
Paint. The overall image dimension should be 400 pixels wide by 240 pixels high. The actual logo need
not cover the entire image. Select Tools>Options from the menu bar. Click the right Browse button and
navigate to where you have stored the logo and select the image.
Select an Instrument
Click on the appropriate instrument button (i.e. SMRT). In the Instrumentation Configuration window
select the appropriate model from the Model pull down window. If using the Ethernet port on your PC
This section describes basic operating procedures for using the RTMS software with the SMRT or MPRT
units for such applications as basic pickup or dropout, basic timing test, paralleling current outputs,
conducting harmonic restraint tests, series of potential sources to provide higher than rated potential,
and forming various three phase voltage outputs.
① Connection button
Press or click on the button and the PC, or STVI hand-held controller, will auto detect the SMRT unit
connected and automatically set the IP address through the Ethernet ports. If the button shows
two red Xs it indicates that there is no communication to the SMRT unit. If the background color is
house icon will appear in the upper left-hand corner. Pressing the house icon will return you
to the manual test screen.
The Battery Simulator button – Turns the Battery Simulator ON and OFF by pressing the button, the
background color changes red for ON and gray for OFF. The voltage to be applied is displayed in the
button and can be changed by pressing the configuration button.
This button will add the present test result to the report. It also displays the report and allows the
user to name the test, enter limits, comments or deficiencies. Reports can be saved to the STVI
internal memory and transferred to PowerDB via a USB memory stick. Previous tests results can be
loaded and the ‘Retest’ option can be used to repeat the test using the same parameters as the
previous test.
⑤Nameplate button
To access the nameplate data window press this button. Here the user can input import information
relative to the relay under test such as manufacturer, model number, serial number, CT and PT
information.
To access the File Management system, touch the File Folder top center of the test screen. This icon
only appears on the STVI hand-held controller and the SMRT-D series of units (not on PC version). It
provides the user the ability to save tests, or open saved tests.
Ramp button
Press either the I Ramp or V Ramp buttons to go to the Ramp Test screen. This button is used for
doing pick up and drop out tests on any type of relay. It can also be used to perform a general
purpose linear step ramp, pulse ramp, or pulse ramp binary search.
There are three Timing Test buttons, I (Overcurrent) V (Over/Under Voltage), and f (Under/Over
Frequency). Press the appropriate button to go to the desired Timing Test screen to easily time
overcurrent, under/over voltage and frequency relays. The RTMS software has ANSI, IEEE and IEC
standard time curve algorithms built-in. In addition, it includes time curves and time curve
algorithms for hundreds of different specific relays selectable by manufacturer, relay model number,
and curve shape (inverse, very inverse, definite time etc.).
Press this button to go to the Sequencing Test screen used for testing reclosing type relays, setting
up multiple vectors and general multi-state sequencing.
Impedance button
Press this button to go to the Click On Fault relay test screen for testing impedance relays.
Differential button
Press this button to go to the Transformer Differential test screen for testing three phase current
differential relays.
Transducer button
Pressing this button will turn the meter mode on (for displaying metered amplitudes). When the
meter mode is on the STVI will displayed the measured output amplitudes in real-time.
Pressing this button will reset the unit back to power up default settings. Use this button to reset
the VIGENS after they have alarmed off due to either a short circuit on the voltage channels, or an
open circuit on the current channels.
⑩ Help button
Pressing this button will provide Help for both software and hardware. Some test screens the Help
button is sensitive to the test. For example, in the Click On Fault Impedance test screen pressing the
Help button will bring up information relative to testing impedance relays.
This display shows the phases and angles of the test values. Pressing on the screen display provides
a full screen display of the test vectors with amplitudes and phase angles. Pressing it again reduces
it back to its original size. If used with the Symmetrical Components (see Configuration button) the
display will display the positive, negative and zero sequence component values.
The Binary Inputs selection bar and More button – The first 3 binary inputs are displayed showing
their present state. Pressing binary input windows #2 and above will display the dialog shown in the
following Figure A. For conducting a timing test, press binary input #1 will display the dialog shown
in Figure B.
The Debounce time is set in milliseconds. The trip contacts must stay closed for the bounce time for
the time test to be true. If the contacts open in less than the set bounce time the timer will continue
to run. Once the input condition is true then the time test will conclude. The trip time displayed will
be the total test time less the debounce time.
Pressing or clicking on the binary more button (next to binary input button) reveals more Binary
Input and Output options.
Pressing these buttons will toggle and set the Prefault and Fault amplitudes, phase angles and/or
frequencies. If the outputs are on, toggling between the two will apply the prefault and fault values
repeatedly with each toggle. This is an especially helpful tool for the user who has to adjust
mechanical contacts. It is also used in conjunction with the following Items 14, 15 and 18 for simple
timing tests.
⑮ Play button
Pressing or clicking the blue play button will apply the prefault vector, start the countdown of the
Prefault Time setting, then step to the Fault values and look for the relay under test to operate.
Pressing or clicking on this button either turns all of the selected outputs ON, or if one or more of
the outputs are ON, it will turn all of the outputs OFF. The center color of this button changes to
green when one or more outputs are ON and indicates that pressing the button will turn all outputs
off, with the exception of the Battery Simulator. To turn off the Battery Simulator press the Bat SIM
button. This button is also used when ramping the battery simulator output.
This button works in junction with the # 16 All ON/OFF button. Pressing the button the center of the
button changes color indicating the output is selected to turn ON. After pressing the All ON button
turns the selected outputs on. Pressing the channel button individually turns the selected channels
ON and OFF after the All ON/OFF button is ON. This allows the user to turn individual outputs ON
and OFF without affecting other channels.
If using the PC version, a right mouse click in the selected window will present the user with a
numeric keypad and dialog box window for setting either individual values or setting multiple values
swiftly and easily. For example, to change the default voltage values for all three phases, enter the
desired voltage value in the entry window and then press the Balance button. All of the values
will change to the desired output voltage. The same is true for both Phase Angle and Frequency.
If using the PC version, a right mouse click in the selected window will present the user with the
Fault Calculator Input Screen.
Mode Options
Overcurrent Mode button: The default selection window will present Overcurrent Mode. There are
two fields available to enter values, Normal (Prefault) and Fault Values. Enter in the provided
windows the desired values of amplitude, and phase angle. Depending on the Selected Fault Type,
upon pressing or clicking on the green check mark, the Fault values will appear in the appropriate
value windows in the manual test screen.
Voltage Mode button: Press or click on the Voltage Mode button. There are two fields available to
enter values, Normal (Prefault) and Fault Values. Enter in the provided windows the desired values
of amplitude, and phase angle (phase angle will be the A phase current relative to the A phase
voltage). Depending on the Selected Fault Type, upon pressing or clicking on the green check mark,
the Fault values will appear in the appropriate value windows in the manual test screen.
Frequency Mode button: Press or click on the Frequency Mode button. The Fault (Frequency)
Values entry window will be provided. Enter the desired fault frequency. The Prefault Frequency will
be the default value.
Impedance Mode button: Press or click on the Impedance Mode button. Pressing A Phase to ground
fault the user will see the following screen.
Test Model button: The input screen defaults to Constant Current test mode. In this screen the user
simply inputs the Ohmic Reach of the relay, at the desired test angle. If the relay requires Prefault
values prior to stepping to the fault values, the user will need to input the desired prefault current
and load angle (Normal Values). Click or press the green check button to return to the test screen.
Selecting Constant Voltage test mode will present the user with a similar input screen, where the
user inputs the desired fault voltage. Selecting Constant Source Impedance the user will be
presented with the following screen.
In this screen the user is required to input the source impedance and angle. The fault calculator will
calculate the resistive and reactive values based upon the user inputs. Upon returning to the test
screen the test values of fault voltage(s), current(s) and angles will be displayed.
3.1.20.1.4.3 Compensation button: This button only appears when you select the Phase to Ground
Fault type. Press or click on the Compensation Mode button to access the selection menu. There are
two types of compensation formulas available, KN and Z0/Z1. Residual compensation factor, KN, is a
complex number that is used to express the earth-return impedance, ZN, in terms of the positive-
sequence impedance reach setting, Z1. This factor is calculated as:
Symmetrical Mode button: Press or click on the Symmetrical button to access the Symmetrical input
setting screen, see the following figure.
To simulate unbalanced fault conditions, a set of 3-phase unbalanced currents or voltages may be
resolved into 3 sets of balanced components of Positive, Negative, and Zero Sequence values. Zero
Sequence currents and voltages occur as the result of a Phase to Ground Fault on the system. If
testing for Zero Sequence enter the Zero Sequence Current or Voltage value into the screen above.
Upon returning to the test screen the appropriate Current and/or Volte values will be displayed and
ready for testing. Negative Sequence is a result of a three phase unbalance condition. Enter the
desired negative Sequence values of Voltage and Current and upon returning to the test screen all
the three phase values will ,be calculated and displayed ready for testing.
Pressing or clicking on the Fault Type Selection button reveals a selection list for different types of
Fault test options.
User may select Phase to Earth (i.e. L1), or Phase to Phase (i.e. L1-L2), or Three Phase fault
simulation (i.e. L1-L2-L3). All values calculated for the type fault selected will be automatically
calculated and entered into the appropriate value windows.
Operation
Initial Start Up
Connection and Power Up - With the Ethernet cable supplied with the unit connect the STVI
Ethernet Port on the SMRT unit to the Ethernet port on the top of the Smart Touch View Interface
(STVI). If using the STVI PC version software, connect the PC/OUT Ethernet Port on the SMRT unit to
the PC Ethernet port. Turn the test set on. As the SMRT unit goes through its power up sequence,
in less than a minute the STVI power up screen will appear. The STVI will auto-detect the SMRT unit
(does not require the user to input an IP address). If using the PC version it too can auto-discover
the SMRT unit connected to the PC. On the PC version, the unit might not auto detect due to
firewall settings. In this case the firewall can be turned off or you can enter the IP address directly
using the PowerDB instrument configuration screen by clicking on the Instrument Setup icon on the
PowerDB tool bar . Click or Press the MPRT/SMRT button. From the Instrument
Configuration Screen, shown in the following Figure, click off the check mark in the Auto Discover
Unit box. Once it auto-discovers the unit, and determines the configuration of the SMRT unit
connected will appear.
Here the user can enter the IP address directly into the box highlighted in red. The IP address of the
unit can be determined by counting the number of times the Binary Output led flashes at the end of
the boot up cycle. The address is 169.254. <#flashes> .0. If the unit flashed four times, the address
would be 169.254.4.0. If the unit is on a network with a DHCP server, the user must use the Auto
Discovery mode.
1. Configuration – Pressing the Configuration button (#2) will allow the user to configure both the
SMRT hardware and software for items such as language, phase angle rotation, etc. Pressing this
button will display the following screen.
Operation Modes – The user can select the output configuration. The connection picture will
change with the selection indicating to the user how to connect the outputs. With the selection
the STVI display will also change in combination with the selected outputs. For example, if a
user needs more than 60 amps, the ‘3 Voltages – 1 Current @ 180 amps’ option can be selected
and will allow the user to enter the value directly. The display will change to a single current
channel and the value entered will automatically be distributed across all three current
generators.
Battery – For units with the ‘P’ System Board Option that includes the battery simulator, the
user can either select one of the common battery voltage levels, or enter the desired voltage
level in the window provided. Upon returning to the test screen the voltage value will be
displayed in the Battery Simulator button . Press this button to turn the output on/off. The
button changes color with the change in output. Note: SMRT1 units do not have a Battery
Simulator.
Last V As Battery – Selecting this button will convert the last voltage channel (normally #3 in an
SMRT36) to a battery simulator. This is especially useful if the unit does not have the system
board with a battery simulator (‘P’ option).
Phase Angles – Select the desired phase angle display for the Phase Vector Screen (#9). The
phase angle designations can be set to 0 to 360 degrees Lead/Lag, or ± 180 degrees (positive
angles are leading). The rotation can also be set to either counter clockwise or clockwise
rotation.
Fault (Phase Vector) Labels – This button allows the user to set labels for the test reports such
as ABC, RST, L1L2L3, RYB, UVW, etc.
Language – This button allows the user to select the desired display language.
Color Options- Allows the user to select the colors for the voltage and current vectors displayed
in the Phase Vector Screen, the background color, and other various colors associated with
moving vectors, grid colors, and characteristic colors.
Change State Immediately / Change on Zero Cross – Unit defaults to the Immediate Mode
where amplitudes, phase angles and frequency changes take place immediately upon command
(point on waveform). The Zero Cross Mode is used to force all amplitude, phase angle or
frequency changes to take place at the positive going zero crossing of the sine wave (normally
used when testing frequency relays).
Symmetrical Components - Pressing this button will change the vector display to show positive,
negative and zero sequence vectors instead of amplitude and phase.
High Burden- Associated with the 1 and 6 Amp ranges in the current amplifier. When enabled
the 1 Amp range is disabled, thus providing full compliance voltage of 50 Volts down to the
smallest current values. Where high accuracy is required at the smaller currents (1 amp and
below) when disabled the 1 amp range is enabled with a lower compliance voltage of 15 volts.
With lower compliance voltage the digital feedback loop (and metering capability) stay on down
to 40 milliamps, thus provides higher accuracy at very low test currents. Test currents below 40
mA are provided as setting values, and have good accuracy due to the lower compliance
voltage. Lower compliance voltage means lower signal to noise ratio.
Voltage Character -The character used to define and label the voltage output channels can be
either V or U. Press this button to change the character. Be sure to press the Save as Default
button to save the changes.
Default Settings Options- Use to change or reset factory default power up settings. Users can
change, save, or reset power up settings such as Phase Angle designations, Phase Rotation,
display settings, etc.
Display Versions – Pressing this button will display firmware and driver versions and build dates.
Auto Frequency – The SMRT will measure and determine the input frequency and automatically
set the default output frequency to the line frequency in the Automatic mode. Other choices
are 50 Hz, 60 Hz, and Custom so that the output frequency can be something other than the
input line. In addition, the user may select Line Sync so that the output phase angles are in a
direct relationship with the positive going zero crossing of the input line frequency.
Logging – Select this button to log commands sent to the SMRT unit.
Set Date and Time – Press button to set the internal date and time reference.
Adjust Screen Brightness – Press this button to adjust the brightness of the STVI screen. Use the
Control Knob to increase or decrease the brightness.
Number Format- Press this button to change the format that numbers are displayed. Typical
format for North America is where a comma delimiter is used to separate thousands from
hundreds, and a decimal point is used to separate tenths from integer numbers. The second
format is similar to Comma Separated Values where commas are used in place of decimal
points.
Screen Exit - To exit the screen and return to the previous screen press the green check
button (you will see this same button on other screens).
Pressing an amplitude, phase or frequency button (right clicking if running on a PC) will display the
following pop-up keypad to enter the value you want to change.
The back arrow < key will delete the right most digit displayed.
Use the numeric keys to type in the value you want and press either the OK button or the
Balance button. The check button will enter the value and take you back to the test screen.
Press the Balance button if you want all the values to be the same amount. If setting phase angles,
and you want all three phases to be shifted the same amount from the defaulted values, enter the
amount of phase shift desired and press Balanced. For example, the default is 0, 120 and 240
degrees. Press on A phase current phase angle and enter 30, press Balance, and the test screen will
now show 30, 150, 270 degrees.
When this button is pressed the window around the channel magnitude is highlighted indicating
that it is now set to be ramped manually using either the Control Knob on the STVI, or the up down
arrows/mouse wheel on the PC keyboard (PC version). If the channel is already selected for
ramping, this button will be labeled ‘Remove channel from ramping’. The magnitude may now be
ramped up or down using the default increment setting. If the user wants to ramp more than one
channel, change the increment, or change the value to be ramped (Amplitude, Phase or Frequency),
on the STVI display screen press the Manual Ramp Options button (Control Knob or Up Down
Arrows Icon #15) to display the following screen.
Increment – Select the desired increment. Color change will indicate the value selected.
Channel – Select the desired channel(s). The channel button will change color indicating the
selected channel(s) to be ramped.
Mode – Select Amplitude, Phase, Frequency, or Battery as the value to be ramped. Press or click on
the green check button to return to the test screen. One click on the STVI Control Knob or one press
of the Up Down arrow on the keyboard will change the output by the Increment setting. If the Auto
Increment button is selected the STVI will automatically select the increment depending on how fast
the control knob is being rotated, the faster the rotation the larger the increment.
1. On the STVI display screen press the Manual Ramp Options button. In Channel Increment
Selection Screen, see Figure 8 above, press or click on the Battery button. Select the desired
increment level of the Battery Simulator, 1 or 5 volt increments. Press on the green check
button.
2. Upon returning to the main test screen, note the Battery Simulator will be set for the value
setting in the configuration screen. If another starting value is desired, go to the configuration
screen and enter the starting value in the window provided, and press the green check
button. For manual ramping press the All ON/OFF button, note button turns green.
3. Press the battery button and note that it changes to red indicating that the battery output
is on, and it will have a yellow arrow through the box with the dc starting value to be
ramped. Use the control knob (pc version use the cursor up down arrows) to vary the dc voltage
1. Select the channel value(s) to be ramped by pressing the appropriate magnitude window(s).
Using the numeric keypad enter the starting value.
2. Connect the appropriate output terminal(s) for the selected channel(s) to be ramped.
3. Connect the desired Binary Input terminal to sense the relay contacts closing or opening. Select
the desired Binary Input and set the appropriate sensing Continuity or Voltage modes.
4. Press the Output Selector Adjustment icon, and enter the increment, channel number(s), and
whether Amplitude, Phase and Frequency. Note that the selected channel(s) should have a
highlighted border around the magnitude window(s).
5. Select the output(s) to be turned on by pressing the ON/OFF icon button for the selected
channel(s). Turn the selected output(s) on by pressing All ON/OFF button. Using either the
Control Knob on the STVI, or the PC up and down arrows, manually ramp the output(s) until the
relay contacts either picks up or drops out, depending on the start value entered in Step 1.
6. Turn the output(s) off by pressing the ON/OFF icon button(s), or press the ALL OFF icon button.
7. Press the Add To Report (#3) button if you would like to document this test in your report.
1. Press the Prefault button. Press the Test/Prefault/Post Time Setting button. Set the desired
Prefault duration in seconds in the window provided.
2. Set the Prefault value(s) by pressing the appropriate magnitude window(s), and using the
numeric keypad, enter the Prefault value(s).
3. Connect the appropriate output terminal(s) for the selected channel(s) to be used.
4. Connect the desired Binary Input terminal to sense the relay trip contacts.
5. Press the selected Binary Input and check the Use as Trip Contact window. Select for the
appropriate sensing Normally Open, Normally Closed, Voltage Applied, or Voltage Removed.
The results can be moved up or down to change the order of the results presentation. The result can be
deleted, or a retest performed. In addition, the user can add or hide Comments or Deficiencies. Press
either Close report to return to the test screen or press Cancel to return to the report. To exit the report,
press the Check button in the top left corner or select the Options button followed by the ‘Close Report’
button.
SMRT36 General Description For detailed descriptions and safety related instructions read the
STVI SMRT MPRT Users Manual prior to operation. The unit’s design is a "modular" concept. All inputs
and outputs are clearly marked and logically grouped. The unit’s Top Panel will appear different among
units, since each unit may have up to three optional Voltage/Current Generator (VIGEN) Modules
installed and one of two system boards. The ‘N’ version consists of a system board with only the power
and communication ports. The ‘P’ version adds 8 additional binary inputs, 4 additional binary outputs
and a battery simulator. For this guide it is assumed the unit is a complete three channel unit.
1. Binary Outputs 1 and 2 are rated for 300 V at 8 Amps and are used to simulate circuit breaker
contacts.
2. Binary Inputs 1 and 2 will sense dry normally open / normally closed contacts, or voltage
applied or removed with a range of 5 to 300 VAC or 5 to 250 VDC.
3. Three Phase Voltages and Currents are noted by the red and yellow surrounding each output
channel. Phases 1, 2 and 3 voltage channels are denoted by the red color. Phases 1, 2 and 3
current channels are denoted by the yellow color. When the voltage generators are converted
to current generators, they will change on the STVI display as current phases 4, 5 and 6.
Input Line Cord Using the power cord supplied with the unit, with the power ON/OFF
switch in the OFF position, connect the unit to a suitable power
source. The unit will operate from any input voltage from 100 to 240
VAC 50/60 Hz. Depending on the outputs in use and the load the test
set can draw up to 1,800 VA on the input. Therefore, make sure the
power outlet is properly sized for the test application.
SMRT-1 General Description For detailed descriptions and safety related instructions read the
STVI SMRT MPRT Users Manual prior to operation. The unit’s design is a "modular" concept. All inputs
and outputs are clearly marked and logically grouped.
SMRT-410 General Description For detailed descriptions and safety related instructions read the
STVI SMRT MPRT Users Manual prior to operation. The unit’s design is a "modular" concept. All inputs
and outputs are clearly marked and logically grouped so continual reference to the instruction manual
should not be necessary once the operator is acquainted with the test system. The unit’s Top Panel will
appear different among units, since each unit may have up to four optional Voltage/Current Generator
(VIGEN), and either a Double Current Generator (DIGEN) or Voltage Only Generator (VGEN) Modules
installed. The ‘N’ version consists of a system board with only the power and communication ports. The
‘P’ version adds 8 additional binary inputs, 4 additional binary outputs and a battery simulator.
Top Panel SMRT410 (Pictured with Floating Returns and DIGEN Option)
1. Binary Outputs – the first two VIGEN modules include Binary Inputs and Binary Outputs.
Therefore, there are 2 Binary Outputs located on the top panel (numbered 1 and 2). More
Binary Outputs are available with the P option see Front Panel section for more information.
Each Binary Output can be configured as Normally Open or Normally Closed contacts providing
logic to the device under test. The Top Panel Binary Outputs can switch up to 300 VAC or 250
2. Binary Inputs – with a minimum 2 channel unit there are 2 Binary Inputs located on the top
panel. More Binary Inputs are available with the P option see Front Panel section for more
information. The Binary Inputs will accept a voltage range of 5 to 300 VAC, or 5 to 250 VDC, or
dry Normally Open / Normally Closed contacts.
3. Voltage/Current Generator Module (or VIGEN) –There are four available slots for the VIGEN
Modules. The slots are numbered 1 to 4 from bottom to top, with the topmost VIGEN numbered 4.
The four voltages and currents are noted by the red and yellow surrounding each output channel.
Phases A, B, C and D Voltage Channels (V1, V2, V3 and V4) are denoted by the red color.
Phases A, B, C and D Current Channels (I1, I2, I3 and I4) are denoted by the yellow color. When
the voltage generators are converted to current generators, they will change on the STVI display
as V1 = I5, V2 = I6, V3 = I7, and V4 = I8.
4. Current/Current Generator Module (or DIGEN) –There is one available slot for the DIGEN
Module. The DIGEN current channels are numbered 5 and 6. When the voltage generators are
converted to current generators, they will change on the STVI display as V1 = I7, V2 = I8, V3 = I9,
and V4 = I10.
5. USB 2.0 Interface requires a Type B “downstream” connector and is primarily used as a
communication and control port when used with a PC and Megger AVTS or the PowerDB STVI
screens. Since the SMRT36 defaults to a baud rate of 115,200 the user will need to configure
their USB output com port to match.
6. PC/OUT Ethernet Port is a 10/100BaseTX port, and is the primary PC connection port. This port
supports MDI/MDI-X auto cross over configuration, which means both standard and “crossover”
Ethernet cables may be used. This port provides the optimal method for downloading EMTP files,
DFR streaming, and updating the unit’s firmware as required. The SMRT comes standard with a
crossover cable. This port may also be used for connecting to the IEC 61850 substation bus for
use in testing IEC 61850 devices. For multiple unit operation, the unit providing the OUT link is
providing the master phase reference to all units “downstream”.
7. IN - 61850 Ethernet Port is a 10/100BaseTX port, and is primarily used to interconnect multiple
SMRT units together for synchronous multi-unit operation. It may also be used to provide
access to the substation IEC 61850 network. Note that the IN and STVI ports share a common
physical port and cannot be used at the same time. With the PC connected to the PC Port, the
1. Incoming Power / Line Cord – The input voltage rating varies depending upon the model
number. If the model number has a letter F or G in the 5th digit, the input voltage may be from
100 to 240 VAC, ± 10%, 50/60 Hertz. If the letter is either C or E, the input voltage may be from
220 to 240 VAC, ± 10%, 50/60 Hertz. Total maximum output power will be limited for CE marked
units. For CE mark units only (C or E in the 5th digit), a derating of the simultaneously available
total output power of the current / voltage amplifiers and battery simulator will occur. The
maximum output power of a single amplifier is not affected. Input current required varies with
the number of output modules in use, load, and input voltage value. The maximum input
current is 15 Amps. The input is protected by an ON/OFF switch / circuit breaker.
2. Earth Ground Jack – use this terminal to connect chassis ground to earth ground.
A chassis ground (earth) point on the front panel is provided as an
additional safety ground.
3. POWER ON/OFF Switch – used to switch unit on and off. The switch illuminates when power is
on.
4. Battery Simulator – the SMRT410 with the P option provides a variable dc output voltage from 5
to 250 Volts, at 100 Watts (4 Amps Max) providing logic voltage for solid-state relays. When
5. Binary Outputs – the P option provides 4 additional Binary Outputs, numbered 3, 4, 5 and 6.
Each Binary Output can be configured as Normally Open or Normally Closed contacts providing
logic to the device under test. Binary Outputs 3 and 4 have an AC Rating of 400 V max., Imax: 8
amps, 2000 VA maximum breaking capacity, and a DC Rating of 300 V max., Imax: 8 amps, 80
Watts, with a Response Time: < 10ms. Binary Outputs 5 and 6 are high speed and have an AC /
DC Voltage Rating of 400 V peak, Imax: 1 amp, with a Response Time: < 1ms typical. The
programmable wait duration is from 1 millisecond to 10,000 milliseconds. An LED directly above
the terminals indicates the status of the contact. ON indicates closed, and OFF indicates open.
The SMRT410 with the N option does not include the additional Binary Outputs 3 through 6.
6. Binary Inputs – the P option provides 8 additional (numbered 3 through 10), independent,
galvanically isolated, Start/Stop or Monitor circuits to monitor operation of relay contacts or trip
SCR. A continuity light is provided for each input gate. Upon sensing continuity, or voltage
applied, the lamp will glow. In addition to serving as wet/dry contacts the Binary Inputs may be
programmed to trigger binary output sequence(s). Binary Inputs can also be programmed using
Boolean logic for more complex power system simulations. The Binary Inputs will accept a
voltage range of 5 to 300 VAC or 5 to 250 VDC, or dry Normally Open / Normally Closed
contacts. The SMRT410 with the N option does not include the additional Binary Inputs 3
through 10.
MPRT 8445 General Description For detailed descriptions and safety related instructions read the
STVI SMRT MPRT Users Manual prior to operation. The unit’s design is a "modular" concept. All controls
and outputs are clearly marked and logically grouped so continual reference to the instruction manual
should not be necessary once the test operator is acquainted with the operation of the test system.
NOTE: The unit can be powered from an input source with a rating of
120 VAC to 240 VAC, 1800 VA max. The unit automatically adjusts to
the available power if it is within the specified range.
2. Earth Ground Jack Use this terminal to connect chassis ground to earth ground.
The chassis ground (earth) point on the front panel is provided as an
additional safety ground.
V2 ⇒ I6
V3 ⇒ I7
V4 ⇒ I8
6. Binary Outputs – There are 6 Binary Outputs. Each Binary Output can be configured as normally
open or normally closed contacts providing logic to the device under test. Binary Outputs 1to 4
have a rating of 300 V AC/DC, 8 amperes and a maximum of 2000 VA breaking capacity (80
watts DC), with a response time of less than 10ms. Binary Outputs 5 and 6 are high speed and
have an AC/DC voltage rating of 400 volts peak, 1 ampere and a response time typically less
than 1ms.
7. Binary Inputs – There are 10 Binary Inputs. Each is independent, galvanically isolated, and serve
as Start/Stop or Monitor circuits to monitor the operation of relay contacts or trip SCR. A
MPRT 8415 and MPRT 8430 General Description For detailed descriptions and safety related
instructions read the STVI SMRT MPRT Users Manual prior to operation. The unit’s design is a "modular"
concept. All controls and outputs are clearly marked and logically grouped. The unit’s rear panel will
appear different among units since each unit may have up to four V/I Generator Modules and may have
either one or two Ethernet ports depending on the age of the unit. There must be at least one V/I
Generator Module installed for this unit to operate properly.
V2 ⇒ I6
V3 ⇒ I7
V4 ⇒ I8
NOTE: The unit can be powered from an input source with a rating of
100 VAC to 240 VAC, 2100 VA max. The unit automatically adjusts to
the available power if it is within the specified range.
2. Earth Ground Jack – use this terminal to connect chassis ground to earth ground.
A chassis ground (earth) point on the back panel is provided as an
additional safety ground.
3. RS-232C Serial Port - The serial port can operate at a maximum baud rate of 115,200 and will
send / receive data in a serial fashion.
Note: For use with the Model STVI-2 (STVI-1 with Power-Over-Ethernet (POE) interface
adapter) connect the MPRT LAN1 port to the Data In port on the POE adapter, and connect the
POE Data and Power Out port to the STVI-1. Note that MPRT units built before November 2007
have a 10BaseT interface. Units built after November 2007 has a 10/100BaseT port. Units built
starting in October 2009 have the dual 10/100BaseT Ethernet ports as shown in Figure 2. LAN2
is the port for connecting to the IEC 61850 substation bus for use in testing IEC 61850 devices.
6. USB Interface – Used to update unit firmware.
7. EPOCH High Current Interface Port –The High Current Interface port is provided on the rear
panel of the unit. It is designed to interface with the Multi-Amp Models EPOCH-II or EPOCH-20
High-Current Output Units. This port interfaces with the optional interface module box (Megger
Model MEIU) to provide control of up to three EPOCH-20 or EPOCH-II current amplitude, phase
angle and frequency outputs. Status of the EPOCH-20/II output can be seen on the TVI display.
8. External Clock –The external clock has separate input and output connections. The Clock In is
used in conjunction with a Clock Out from another unit or other clock source for multiple unit
operations or other special test applications. The Clock Out provides a 14.4 KHz clock signal to
another unit that phase locks these units together.
9. External Trigger –The External Trigger has separate input and output connections. They enable
the unit to sync with another unit, or TTL (+ 5Vdc) signal source such as a GPS satellite receiver.
The Trigger In is used in conjunction with another unit to establish a trigger for a special
operation or programmed event. Typically the Trigger In synchronizes to the Trigger Out unit
and acts as a slave to it. The Trigger Out provides a TTL digital signal to another unit in order to
synchronize a multiple unit operation when it needs to establish a trigger for a special operation
or programmed event. The TTL signal and pulse width for the GPS receiver is:
Voltage Input Signal: 3V peak minimum Pulse Width: 50µS nominal