Computing Scale (3.1 Version)

This document provides instructions for setting up and calibrating the MRP-15 scale. It outlines the following key steps: 1. To set the unit of measurement (e.g. kg, lb, oz), press the corresponding number key when the display shows 98-04, 3.1, -1-. 2. To set the decimal place for the total price display, press UNIT SET and TARE/ZERO followed by 0, 1 or 2. 3. To reset the data for IC93C46, press and hold CE for 3-5 seconds when the display shows 98-04, 3.1, -1- and then enter 110. 4. To calibrate

Uploaded by

Micky Boza
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
106 views3 pages

Computing Scale (3.1 Version)

This document provides instructions for setting up and calibrating the MRP-15 scale. It outlines the following key steps: 1. To set the unit of measurement (e.g. kg, lb, oz), press the corresponding number key when the display shows 98-04, 3.1, -1-. 2. To set the decimal place for the total price display, press UNIT SET and TARE/ZERO followed by 0, 1 or 2. 3. To reset the data for IC93C46, press and hold CE for 3-5 seconds when the display shows 98-04, 3.1, -1- and then enter 110. 4. To calibrate

Uploaded by

Micky Boza
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 3

MRP-15 < Version 3.

1>
 SETUP 
Instruction Display Note

1. Press [CHANGE] key and WEIGHT: 98-04


switch on the scale at the UNIT PRICE: 3.1
same time. Do not
TOTAL PRICE: -1-
release your hand from
[CHANGE] key till display
shows the version of
scale.

1. Setup the weight unit:


Press the following keys while LCD shows [98-04] [3.1] [-1-]

Key in: Function:

1 Power on the scale at Kg weight units

2 Power on the scale at Lb weight units

3 Power on the scale at Oz weight units

4 Power on the scale at 100g weight units

0 Show internal A/D Value (Press [CHANGE] key to re-boot the scale)

2. Setup the decimal point in Total Price Display.


While LCD shows [98-04] [3.1] [-1-]:

Key in: Decimal Point on Total Price Display:

[UNIT SET] → [TARE/ZERO] → [0] None

[UNIT SET] → [TARE/ZERO] → [1] 0.0

[UNIT SET] → [TARE/ZERO] → [2] 0.00

3. Setup the primitive data for IC93C46


While LCD shows [98-04] [3.1] [-1-], please press [CE] key for 3 ~ 5 seconds, and then follow the
instructions as below:

Instructions: Descriptions
1. Key in [1] → [1] → [0], and then the scale will enter 010 means MRP-15kg / 5g
Calibration Mode automatically.

2. Please calibrate the scale (refer to the instructions of


calibration on next page)

!! Notice: After re-setting the data for IC93C46, the


scale will work under Lb weight units. Please
remember to change the weight units into Kg.

 Weight Calibration 

Instructions Display Descriptions

1. Power on the scale, and wait till [0.000] [0] [0]


Display shows 0.

2. Press [CHANGE] key and hold SETUP Enter SETUP mode


till Display shows SETUP.

3. Press [] once to enter -CAL- Enter calibration mode


calibration mode.
V Figure = Capacity/ Standard Weight
4. Press [CE] key to select the V
Example:
figure (0 ~ 10). If you use a 5kg 3
GRP-15Kg / 5g
standard weight, please select 3.
Standard Weight: V figure:
5kg 3
15kg 1

5. Press [CHANGE] key to confirm CAL-1


the V figure. And then Display ↓
will show CAL-1 → Internal
XXXX
Offset Value.

6. Press [CHANGE] key again CAL-2


while the Offset Value is stable. ↓
And then Display will show CAL-
0
2→0

7. Put the standard weight on he


platter. And then Display will XXXX
show Span Value.

8. Press [CHANGE] key twice Finish setting the data for IC93C46.
when the Span Value is stable.
And then the scale will re-boot
automatically.

Notice!! Please remove the


standard weight from the platter
while the scale is re-booting.

 Data-Lose: Err-1 
1. After switching on the scale, if Display shows [0098] [0400] [Err-1], please key in [0] → [1] →
[0] directly, and then the scale will enter Calibration Mode automatically.

2. If the scale does not enter Calibration Mode, please change a new IC93C46.

3. If the scale cannot be calibrated normally, please make sure if the battery is already under-
voltage.

 Recharge the Battery 


1. TOTAL PRICE Display shows “oo”:

Please recharge the battery.

2. The scale sounds “BiBiBiBiBiBi” three times and three LCD Displays show “- - - - -“ continuously:

The battery has completely run down. Please recharge the battery immediately.

3. Charging Indicator – LED:

LED in red light: the battery is under recharging.

LED in green light: the battery is full.