Particle Size Analysis
Particle Size Analysis
LASER DIFFRACTION
Abstract:
Laser-based techniques for particle-size measurement have become increasingly important in
combustion research and many other disciplines. Instruments are continually being
developed and improved to meet the demanding geometric, accuracy and other
requirements associated with current research and industrial applications. The operating
associated with applying these methods, are discussed. Light diffraction is one of the
most widely used techniques for measuring the size of a wide range of particles from
very fine to very coarse. The method is popular because it is quick and easy to use,
flexible, and it can be adapted to measure samples presented in various physical forms.
The method depends on the analysis of the diffraction pattern produced when particles are
particle size, mathematical analysis can produce an accurate, repeatable picture of the
size distribution.
Abbreviations:
i
TABLE OF CONTENTS
1. Introduction vii
ii
LIST OF TABLES
LIST OF FIGURES
iii
1. Introduction:
the research and development and industrial communities working in a broad range of
disciplines. Table 1 describes some of the many fields in which particle-size analysis is
essential and lists some of the primary uses of particle-sizing methods in these areas
For example, particle size and number density are critical parameters in two-phase
combustion experiments and modelling, since the particles or droplets represent the energy
source in reacting two-phase flows. Thus, knowledge of the particle size distribution is a key
parameter in modelling radiative heat transfer, as well as in characterizing other properties,
iv
such as droplet evaporation rates and total droplet/particle burnout. The particles and the
associated heat release also have a large impact on the flow field in the combustion
region. Laboratory work with laser-based sizing instrumentation includes work on the
characterization of all types of two-phase flows, measurements of droplet size
distributions in clouds, quality control of injection fluids for medical use and studies of
respiratory illness caused by particle inhalation. Laser-based methods of particle-size
analysis are used extensively in spray characterization and have seen industrial
applications in such diverse fields as pulverized coal combustion, spray combustion,
powder metallurgy, pharmaceuticals manufacturing, agriculture, powder and liquid paint
production and pollution monitoring.
Fig.1 shows a diagram outlining several broad methods of particle-size analysis, including
those based on optical techniques and those mechanical in nature.
v
1.3 Important elements in particle-size measurement
Laser-based methods of particle-size analysis have several common elements that must be
addressed and understood before a method can accurately be used for particle-size
analysis. A knowledge of particle-sizing terminology and particle size distribution
functions is essential to understand the measurements from the various systems. Issues
to resolve when applying a technique include that of a Gaussian intensity profile
across the laser beam diameter, the scattering characteristics of the particle, and the
measurement environment. Problems related to the Gaussian beam profile must be
resolved for all types of laser-based systems.
A number of terms have specific meanings related to the description of the shape and
size of the particles. Table 2 lists some of the more common terms related to aspects of
particle shape, along with their associated meanings
Most descriptions of the particle diameter involve relating some aspect of the particle
to a particle having an equivalent spherical diameter. If the particle under consideration is
spherical, then all diameters become equal. The wide variation in particle shape leads to
terms such as the volume diameter (where the diameter is that of a sphere having the same
volume as the actual particle), the surface diameter (where the diameter is that of a
sphere having the same surface area as the actual particle) and others, as defined in
Table 3.
vi
Table 3. Definitions and descriptions of mean particle (Barth, 1993)
Each of these diameter definitions stems from application of one or more methods of
size analysis. Diameter definitions, such as the projected area diameter, Martin's diameter
and Feret's diameter, come from analyses using microscopy where these diameters are
easily measured. Others, such as the drag diameter, come from sedimentation methods of
particle-size analysis. All laser-based methods of particle sizing measure an equivalent
spherical diameter.
Laser-based particle analyzers generally determine the distribution of the equivalent spherical
particle sizes. The mean particle size of the distribution can then be defined in different
ways. Some of the most common are the arithmetic mean (d10), the surface mean (d20),
the volume mean (d30) and the Sauter mean (d32). Each mean diameter is defined to be
the most useful or convenient measure to use given the phenomena under investigation.
For example, d32 is used in combustion related fields, d43 is used in the field of
chemical kinetics and d30 is common, since diffraction- based instrumentation typically
report this parameter. f (di) is the value of the particle size distribution function at the
discrete values of the particle size.
vii
Table 4. Common size distributions and their associated parameters (Barth, 1993)
Particle size distribution functions are mathematical expressions that relate some aspect of
the particulates in the flow, such as the volume or concentration, with the particle diameter.
These distributions functions, given the symbol rid), are a simple mathematical method
of characterizing particulates without the need for detailed plotting of the experimental
distribution. They allow communication of detailed information about the particles in a
concise manner. The use of size distribution functions also provides a way to
extrapolate information about particle sizes outside the measured range of values,
makes calculation of relevant information simple (such as the average particle diameters
and parameters describing the distribution shape), and provides insight into the
processes involving the particles. Distribution functions of particles and droplets can be of
several types, depending on the nature of the particles and the environment in which
they are used.
Some of the most common distribution functions for powders and sprays are the
normal, log-normal and Rosin-Rammler distributions, The mathematical formulas of the
normal, log-normal and Rosin- Rammler distributions along with the equations for the
mean and standard deviation for each distribution are shown in Table 4.
The normal distribution is rare, usually applicable only when the range of particle sizes is
narrow. Most distributions of powders in manufacturing processes, pulverized coal, fly
ash and any powders produced by grinding processes fit a log-normal distribution. The
log-normal distribution is similar to the normal distribution and is given by letting z =
ln(d). Fine powders and sprays can be described in terms of the Rosin-Rammler
viii
distribution, written in terms of the percentage undersized R; and the particle diameter
d. The variables b and n are a measure of the spread in the particle size and a
characteristic of the particle material, respectively.
Light diffraction is one of the most widely used techniques for measuring the size of
a wide range of particles from very fine to very coarse. The method is popular
because it is quick and easy to use, flexible, and it can be adapted to measure samples
presented in various physical forms. The method depends on the analysis of the
diffraction pattern produced when particles are exposed to a collimated beam of light .
As the patterns are characteristic of the particle size, mathematical analysis can produce
an accurate, repeatable picture of the size distribution.
ix
However, the technique since has been broadened t o include light scattering in a
wider angular range by application of the Mie theory, in addition to the Fraunhofer
approximation and anomalous diffract ion. The measurement of particle size is an
import ant aspect in the characterization of raw materials and drug formulations. To
monitor production and particle stability, efficient and rapid methods for particle sizing are
needed. The electronic measurement of samples offers a convenient solution to this
problem. However, the electronic measurement will often have to be accompanied by
microscopical investigation to determine the type of particles being investigated.
Modern drug formulations may also be particulate formulations where the particle size
will be below 100 µm, and reliable and reproducible methods are needed for the
quality control of these drug products.
2.1 Principle
2.2 Apparatus
x
A representative sample, dispersed at an adequate concentration, is passed through the
light beam in a measuring zone by a transporting fluid ; This measuring zone should be
within the working distance of the lens used. In some cases the particle stream in a process is
illuminated directly by the laser beam for measurement, as in the case of sprays, aerosols, and
air bubbles in liquids. In other cases (such as emulsions, pastes, and powders), representative
samples can be dispersed in suitable liquids. Often dispersants (such as wetting agents or
stabilizers) or mechanical forces (such as agitation ) is applied to deagglomerate particles
and to stabilize the dispersion. For these liquid dispersions, a recirculating system is
most commonly used, consisting of an optical measurement cell, a dispersion bath
(usually equipped with stirrer and ultrasonic element s), a pump, and tubing.
Dry powders can also be converted into aerosols through the use of dry powder
dispersers, which apply mechanical forces for deagglomeration. In this case, a dosing device
feeds the disperser with a constant mass flow of sample. The disperser uses the energy of
compressed gas or the differential pressure to a vacuum to disperse the particles. It
outputs an aerosol that is blown through the measuring zone, usually into the inlet of a
vacuum pipe that collects the particles.
There are two posit ions in which the particles can enter the laser beam. In the conventional
case, the particles enter the parallel beam before and within the working distance of the
collecting lens. In the so-called reversed Fourier optics case, the particles enter behind
the collecting lens, and thus in a converging beam.
The advantage of the conventional set up is that a reasonable path length for the sample
is allowed within the working distance of the lens. The second set up allows only small
path lengths but enables measurement of scattered light at larger angles, which is useful
when submicron particles are present .
The interact ion of the incident light beam with the dispersed particles results in a
scattering pattern with different light intensities at various angles. The total angular
intensity distribution, consisting of both direct and scattered light , is then focused by a
positive lens or an assembly of lenses onto a multi element detector. The lens (es) provide(s)
for a scattering pattern which, within limits, is not dependent upon the location of the
particles in the light beam. Thus the continuous angular intensity distribution is
converted into a discrete spatial intensity distribution on a set of detect or elements.
It is assumed that the recorded scattering pattern of the particle ensemble is identical to
the sum of the patterns from all individual single scattering particles presented in
xi
random relative posit ions. Note that only a limited angular range of scattered light is
collected by the lens (es), and thus by the detector.
Sprays, aerosols, and gas bubbles in liquid should be measured directly, provided that
their concentration is at an appropriate level, since sampling or dilution is generally
impossible without altering the particle size distribution.
The dispersion procedure is adjusted t o the purpose of the measurement : for example,
whether agglomerates should be detected or broken down to primary particles.
The particle concentration in the dispersion should be above a minimum level, which
for many instruments will correspond to about 5% obscuration, in order to produce an
acceptable signal - to - noise ratio in the detector. Likewise, it should be below a
maximum level in order to avoid multiple scattering.
xii
A variety of liquids are available for the dispersion of powders and must have the
following characteristics:
be transparent at the laser wavelength and free from air bubbles or other
particles
be compatible with the materials used in the instrument
not dissolve or alter the size of the particulate materials
favour easy and stable dispersion of the particulate material
have suitable viscosity to enable calculation
have a refractive index that differs from that of the material (for the Mie
calculation), and
not be hazardous to health and meet safety requirements
Most often either the Fraunhofer approximation or the Mie theory is used, though
other approximations are sometimes applied for calculation of the scattering matrix.
Below approximately 25 µm, the differences between the optical models become more
significant. In this size range, the proper application of the Mie theory (assuming
accurate real and imaginary refractive index values) provides the greatest accuracy.
When using the Mie theory, the refractive indices of particulate and medium, or their ratio,
are entered into the instrument to allow calculation of the model matrix. Often, small values
of the imaginary part of the refractive index (about 0. 01–0. 1) are applied to cope with
the surface roughness of the particles. In order to obtain traceable results, it is essential that
the refractive index values used are reported.
xiii
The scattered intensity can be calculated by replacing the particle by a disc of same shape and
size as the projected area of the particle in the direction of propagation of the incident light.
Therefore, far field intensity distribution in the forward direction is independent of refractive
index and only a function of particle size, shape and the properties of the incident wave. For a
spherical cross section and small scattering angles (sinθ = θ), the Fraunhofer Diffraction
pattern is given by
Where
x = Mie parameter
This angular intensity structure contains side lobes, which decrease in maximum intensity
with increasing scattering angle. For larger particles, the overall intensity decreases with
scattering angle.
When an ensemble of particles in a finite volume is illuminated, all the diffraction pattern of
the individual particles overlap in the far field. For a time varying particle ensemble and a
finite illumination time, interference effects are averaged and the resulting intensity s the
summation over the intensities of all contributing particles.
xiv
Where p(x) is the probability density function of the Mie parameter, i.e., of the size
distribution. Therefore, this equation represents a relation between particle size distribution
and far field distribution of scattered light from these illuminated particles.
This deconvolution step is the inverse of the calculation of a scattering pattern for a
given particle size distribution. The fact that rapidly measured data always contain some
random and systematic errors may cause erroneous size distribution results. Several
mathematical procedures have been developed for use in the different instruments
available. They allow for some weighting of deviations between measured and calculated
scattering patterns (such as least squares), some constraints (such as non negativity for
amounts of particles), or (or in combination with) some smoothing of the size
distribution curve. A new procedure uses the observed fluctuations of the detect or
signals to introduce proper weighting of these data and to calculate confidence intervals
for the particle size distribution.
The algorithms used are specific to each make and model of equipment and are
proprietary. The differences in the algorithms between different instruments can give rise
to differences in the particle size statistics. For this reason, when reporting the particle
size distribution and statistics for a given material, also report the cell type, sample state
and preparation, together with the make and model of the equipment .
2.7 Replicates
The required precision of the method is dependent on the characteristics of the material
(milled versus not milled, robust versus fragile), and also on the requirements of the
application (formulation type and technique). Appropriate measurement conditions are
established experimentally, in relation to the desired precision. In general, at least three
different representative samples from the same batch are measured. The repeatability of the
particle size distribution parameter is as follows: for any central value of the
distribution, for example the median (x50), the coefficient of variation is less than 10%.
For values away from the centre of the distribution, for example x10 and x90, the coefficient of
variation cannot exceed 15%. Below 10µm, these maximum values are doubled.
xv
The system suitability test is used to verify that the precision and accuracy are adequate
for the analysis to be done. The test is based on the concept that the equipment ,
electronics, and analytical operations constitute an integral system that can be evaluated
as such. This can be done by measuring at regular time intervals a control material
of known size distribution. In general, unless otherwise specified in the individual
monograph, the mean values of three measurements must deviate from the established
value by less than 10% for x50, and by less than 15% for x10 and x90. Below 10µm,
these maximum values are doubled.
The distribution statistics are usually reported by the instrument data system. Most
common parameters are calculated from the cumulative distribution by interpolation.
Percentile sizes, xm, represent the particle size in relation to which m percent of the
distribution is smaller. (The notation dm is also used and is equivalent to xm.) Qy
represents the percent smaller than y microns. Mean sizes, such as D4,3, the arithmetic
volume mean diameter, can also be calculated by representing the distribution as a
collection of spherical particles with diameters of the size band midpoints. Unless
otherwise stated, parameters are calculated on the volume or mass basis.
Primarily, qualification validation can be made with any certified or standard reference
material, accept able for use in the particular industry. Here, the total measurement
procedure is being examined, including sampling, sample dispersion, sample transport
through the measuring zone, measurement , and the deconvolution procedure. It is
essential that the total operational procedure is adequately described in full detail.
xvi
than 3%. The mean values for x10 and x90 must not exceed the certified range of
values by more than 5%. For repeatability, the coefficient of variation must be less than
3% for x50 and less than 5% for x10 and x90. Below 10 µm, these maximum values are
doubled.
Although the use of spherical materials is preferable, non spherical ones may also be used.
Preferably, these have certified or typical values coming from laser diffraction analyses
according to an agreed, detailed operational procedure. If the reference values come
from methods other than laser diffraction, a significant bias may result . The reason for
this bias is that the different principles applied in the various methods may lead to
different responses to the particles, and thus to different equivalent -sphere diameters for
the same non spherical particle.
xvii
3. Comparison of Fraunhofer and Mie scattering
As diameter increases, intensity per particle increases and the location of first peak shifts left
i.e. to the smaller angle.
xviii
Fig.5 Effect of imaginary part of the refractive index (Hirleman, E., 1988)
In general, refractive index is expressed as m= n-ik, where the real part n is the refractive
index of the particle and the imaginary part k is the absorptivity of the particle.
Fig.6 Effect of real part of the refractive index (Hirleman, E., 1988)
xix
Practical application:
THE FUTURE
Laser diffraction systems are presently perceived ass productive, flexible workhorses – a
reliable part of company’s analytical armoury. However, there are challenges remaining as
industries evolve. One main issue for laser diffraction instrument manufacturers is to make
sure the performance gains over the last decade are fully exploited by the users.
Laser is complementary to, and complemented by, these newer techniques. It is highly
effective to exploit them in combination. The benefit of laser diffraction suggests that the
technique will retain its place as the preferred choice for particle size measurement.
xx
REFERENCES
Barth, H. G. and Sun, S (1993)., Particle size analysis, Anal. Chem. 65, 55R-66R.
Farmer (1972), W. M., Measurement of particle size, number density and velocity using a
laser interferometer, Appl. Opt. 11, 2603-2612.
Holve (1991), D. J., Using ensemble diffraction to measure particle size distribution, Powder
Bulk Engg 5, 15-19.
Stanley-Wood, N. G. and Lines, R. W., (1992) (Eds), Particle Size Analysis, Royal Society of
Chemistry, Cambridge.
Zhang, H. J. and Xu, G. D. (1992), The effect of particle refractive index on size
measurement, Powder TechnoL 70, 189-192.
xxi