EB3100 Lecture Note 1-Intro - 8
EB3100 Lecture Note 1-Intro - 8
Use data
Select further Final
to narrow the
tests diagnosis
diagnosis
More than Only one
one likely likely
Feedback in measurement systems…
• Figure shows that the measurand is measured by a
sensor converting the variable to an electrical signal,
which can undergo signal processing. Sometimes the
measurement system provides feedback through an
effector to the subject.
Outputs
Feedback
Data Data
Effector
storage communication
…Feedback in measurement systems
• Figure (a) shows that a patient reading an instrument usually lacks
sufficient knowledge to achieve the correct diagnosis.
• Figure (b) shows that by adding the clinician to form an effective
feedback system, the correct diagnosis and treatment result.
Patient
Instrument
Patient Clinician
Instrument
(a) (b)
…Feedback in measurement systems
• In certain circumstances, proper training of the patient and a
well-designed instrument can lead to self-monitoring and self-
control (one of the goals of bioinstrumentation).
– An example of such a situation is the day-to-day monitoring of glucose
by people suffering from diabetes. Such an individual will contact a
clinician if there is an alert from the monitoring instrument.
Abnormal
Clinician readings Patient Instrument
Some characteristics of BMI
• Deals with biological tissues, organs and organ systems and their
properties and functions
• Always consider the underlying physiological mechanism in
measurement interpretation
• Bio-phenomena:
– bioelectricity, biochemistry, biomechanics
Actuator
Computation Unit
Variable
Conversion
Element
Electrical
Interface
Components of BM Instrumentation System…
• A sensor
Converts a physical measurand (biochemical, bioelectrical, or
biophysical parameter) to an electrical output
– (Ideally) should respond only to the targeted form of energy
present in the measurand
– Should be minimally invasive (ideally noninvasive)
– Provides a safe interface with biological materials
• An actuator
– Delivers external agents via direct or indirect contact
– Controls biochemical, bioelectrical, or biophysical parameters
– Provides a safe interface with biologic materials
Sensor/Actuator Technical Specification:
example (1)
• Bioinstrumentation should be designed with a specific signal in mind.
– Table shows a few specifications for an electrocardiograph
– The values of the specifications, which have been agreed upon by a committee,
are drawn from research, hospitals, industry, and government.
Specification Value
Input signal dynamic range ±5 mV
Dc offset voltage ±300 mV
Slew rate 320 mV/s
Frequency response 0.05 to 150 Hz
Input impedance at 10 Hz 2.5 M
Dc lead current 0.1 A
Return time after lead switch 1s
Overload voltage without damage 5000 V
Risk current at 120 V 10 A
Sensor/Actuator Technical Specification:
example (2)
• The specifications for a typical blood pressure sensor.
– Sensor specifications for blood pressure sensors are
determined by a committee composed of individuals from
academia, industry, hospitals, and government
Specification Value
Pressure range –30 to +300 mmHg
Overpressure without damage –400 to +4000 mmHg
Maximum unbalance ±75 mmHg
Linearity and hysteresis ± 2% of reading or ± 1 mmHg
Risk current at 120 V 10 A
Defibrillator withstand 360 J into 50
Sensor/Actuator Technical Specification:
compare with the biosignal range!
Measurement Range Frequency, Hz Method
Blood flow 1 to 300 mL/s 0 to 20 Electromagnetic or ultrasonic
Blood pressure 0 to 400 mmHg 0 to 50 Cuff or strain gage
Cardiac output 4 to 25 L/min 0 to 20 Fick, dye dilution
Electrocardiography 0.5 to 4 mV 0.05 to 150 Skin electrodes
Electroencephalography 5 to 300 V 0.5 to 150 Scalp electrodes
Electromyography 0.1 to 5 mV 0 to 10000 Needle electrodes
Electroretinography 0 to 900 V 0 to 50 Contact lens electrodes
pH 3 to 13 pH units 0 to 1 pH electrode
pCO2 40 to 100 mmHg 0 to 2 pCO2 electrode
pO2 30 to 100 mmHg 0 to 2 pO2 electrode
Pneumotachography 0 to 600 L/min 0 to 40 Pneumotachometer
2 to 50
Respiratory rate 0.1 to 10 Impedance
breaths/min
Temperature 32 to 40 °C 0 to 0.1 Thermistor
…Components of BM Instrumentation System…
Sensor Sensor
signal signal
Measurand Measurand
(a) (b)
Biomedical Sensing
• Most sensors are analog and provide a continuous range of
amplitude values for output (a).
• Other sensors yield the digital output (b).
– Digital output has poorer resolution, but does not require conversion
before being input to digital computers and is more immune to
interference
Amplitude
Amplitude
Time Time
(a) (b)
Biomedical Sensing
5 mV
(a) Dynamic
Time Range
-5 mV
Amplitude
1V
(b)
Time
-1 V
Biomedical Sensing
• DC offset voltage is the amount a signal may be moved from
its baseline and still be amplified properly by the system.
Figure shows an input signal without (a) and with (b) offset.
Amplitude
(a) Time
Amplitude
(b) Dc offset
Time
Biomedical Sensing
• The frequency response of a device is the range of
frequencies of a measurand that it can handle.
• Frequency response is usually plotted as gain versus
frequency.
• Figure shows Frequency response of the electrocardiograph.
1.0
Amplitude
0.1
0.05 Hz 150 Hz
Frequency
Biomedical Sensing
Output Output
Input Input
(a) (b)
Biomedical Sensing
Amplitude
Time Time
(a) (b)
Biomedical Sensing
• Signal conditioning
– Amplify, filter, match the impedance of the sensor
to the display
– Convert analog signal to digital
– Process the signal
Clinical Biosignal Measurement