Carestation 620 Spec Sheet Rev5
Carestation 620 Spec Sheet Rev5
Key Features
• Elegant modern design in a slim, compact frame well
suited for constrained environments
• Simple and easy to use 15" touchscreen ventilator display
• Intuitive CARESCAPE™ inspired user interface for the
unified Carestation user experience
• Integrated Respiratory Module
Ventilation
• Small, compact breathing system specifically designed
for low flow anesthesia
• Fast gas kinetics for rapid wash-in and wash-out
• Digitally controlled flow valve ventilator supports all
patient types from neonates to adults
• Continual fresh gas flow with fresh gas flow compensation
during mechanical ventilation
Design
• Durable wheels for mobility and stability
• Robust and easy to clean surfaces
• Top shelf display mounting
• Two vaporizer configuration
• Bi-level work surface illumination
• Absorbent canister designed for ease of use and long life
• Intelligent lighting that highlights active flow controls
and auxiliary ports when in use
Physical Specifications
Product Description
Carestation 620 A1
Dimensions
Height: 135 cm/53.1 in
Width: 82.5 cm/32.4 in
Depth: 69.1 cm/ 27.2 in
Weight: 145 kg/320 lb*
Top shelf
Weight limit: 25 kg/55 lb
Width: 41.3 cm/16.3 in
Depth: 27.0 cm/10.6 in
Height:
Top and middle: 8.6 cm/3.4 in
Bottom: 13.3 cm/5.2 in
Width: 34 cm/13 in
Depth: 37 cm/14.6 in
Casters
Diameter: 12.5 cm/4.9 in
Brakes: Caster Brakes
3
Ventilator Components Airway Modules
Flow transducer General
Type: Variable orifice flow sensor E-sCAiO, E-sCAiOV, N-CAiO
(autoclavable) Size (HxWxD),
Location: Inspiratory outlet and expiratory inlet excluding water trap: 113 x 38 x 205 mm/4.4 x 1.5 x 8.1 in
Weight: 0.7 kg/1.5 lb
Oxygen sensor
Sampling rate: 120 mL/min ±20 mL
Type: Optional galvanic fuel cell or
Automatic compensation for atmospheric pressure variation
paramagnetic with Airway
(495 to 795 mmHg) temperature and CO2/N2O and CO2/O2
Module option
collision broadening effect. Parameter display update
Ventilator screen interval typically breath-by-breath. Functional alarms for
blocked sample line, D-fend check and D-fend replacement.
Display size: 15 inch
Pixel format: 1024 x 768 Non-disturbing gases:
Ethanol, acetone, isopropanol, methane, nitrogen, nitric
Battery backup oxide, carbonmonoxide, water vapor, freon R134A (for CO2,
Backup power: Battery time is 90 minutes when O2 and N2O):
fully charged, which supports full Maximum effect
system functionality and ventilation. on readings: CO2 < 0.2 vol %; O2, N2O < 2 vol %,
Battery type: Internal rechargeable sealed AA < 0.15 vol%
lead acid
Carbon dioxide (CO2)
Communication ports
EtCO2: End-tidal CO2 concentration
RS-232C compatible serial interface
FiCO2: Inspired CO2 concentration
Ethernet
Datex-Ohmeda device interface solutions port CO2 waveform
USB port Measurement range: 0 to 15%
VGA Output (0 to 15 kPa, 0 to 113 mmHg)
Accuracy: ±0.2 vol % + 2 % of reading
Anesthetic Agent Delivery Datex-Ohmeda infrared sensor
Adjustable low and high alarm limits for EtCO2 and FiCO2
Delivery
Respiration rate (RR)
Vaporizers: Tec™ 6 Plus, Tec 7
Measurement range: 4 to 100 breaths/min
Number of positions: 2
Detection criteria: 1% variation in CO2
Mounting: Tool-free installation Selectatec™
Adjustable low and high alarm limits for respiration rate;
manifold interlocks and isolates
alarm for apnea
vaporizers
Patient Oxygen (O2)
FiO2: Inspired O2 concentration
EtO2: End-tidal O2 concentration
FiO2-EtO2: Inspired-expired difference
O2 Measurement
Measurement range: 0 to 100%
Accuracy: ±1 vol % +2 % of reading
Datex-Ohmeda differential paramagnetic sensor
Adjustable low and high alarm limits for FiO2 and EtO2;
alarm for FiO2 < 18%
4
Nitrous Oxide (N2O) Compliance
Measurement range: 4 to 100 1 to 100
Measurement range: 0 to 100%
mL/cmH2O mL/cmH2O
Accuracy: ±2 vol % +2 % of reading
Airway resistance
Anesthetic Agent (AA) Measurement range: 0 to 200 cmH2O/L/s
Halothane, Isoflurane, Enflurane
Sensor specifications
Measurement range: 0 to 6%
Accuracy: ±(0.15 vol% +5% of reading) D-lite/ Pedi-lite/
D-lite(+) Pedi-lite(+)
Sevoflurane
Measurement range: 0 to 8% Dead Space: 9.5 mL 2.5 mL
Accuracy: ±(0.15 vol% +5% of reading) Resistance
Desflurane at 30 L/min: 0.5 cmH2O
Measurement range: 0 to 20% at 10 L/min: 1.0 cmH2O
Accuracy: ±(0.15 vol% +5% of reading)
Waveform displayed
Electrical Specifications
MAC value displayed (Airway Gas Option modules)
Current leakage
MACage value displayed (CARESCAPE modules)
Identification threshold: 0.15 vol%** 100/120 V: < 300μA
Agent mixture detection 220/240 V: < 500μA
Adjustable high and low alarm limits for EtAA, FiAA Power
Patient Spirometry™ (optional) Power input: 100-120 Vac, 50/60 Hz
220-240 Vac, 50/60 Hz
Pressure-volume loop
120/220-240 Vac ± 10%, 50-60 Hz
Pressure-flow loop
Power cord:
Flow-volume loop
Length: 5 m/16.4 ft
Airway pressure and flow waveforms
Rating: 10A @ 220-240 Vac or
Adjustable low and high alarm limits for Ppeak, PEEPtot and MVexp 15A @ 100-120 Vac
Alarms for MVexp << MVinsp and for MVexp low. Detection 10A @ 120/220-240 Vac
through D-lite™ or Pedi-lite™ flow sensor and gas sampler
with following specifications:
Inlet modules
100/120 V:
CARESCAPE Airway Modules Without outlets: 2A
D-lite(+) Pedi-lite(+) With outlets: 10A
Respiration rate: 4 to 35 4 to 70 220/240 V:
breaths/min breaths/min Without outlets: 1A
Tidal volume With outlets: 5A
Measurement range: 150 to 2000 mL 5 to 300 mL
Outlet modules (optional)
Accuracy**: ±6% or 30 mL ±6% or 4 mL
100/120 V:
Minute volume
3 outlets on side, 1-3A, 2-2A, individual breakers,
Measurement range: 2 to 20 L/min 0.1 to 5 L/min
isolation transformer (optional)
Airway pressure
220/240 V:
Measurement range: -20 to +100 cmH2O
3 outlets on side, 1-2A, 2-1A, individual breakers,
Accuracy**: ±1 cmH2O isolation transformer (optional)
Display units: cmH2O, mmHg, kPa, mbar, hPa
120/220-240 V:
Flow No outlets
Measurement range: -100 to -25 to 25 L/min
100 L/min
I:E
Measurement range: 1:4.5 to 2:1
**Typical value 5
Pneumatic Specifications O2 concentration range: 21% to 100% when Air is available
O2 Cell accuracy: ±2.5% plus 2.5% of reading
Auxiliary O2 (optional) Compensation: Temperature and atmospheric
pressure compensated to standard
Connection: 7-10 mm hose barb port
conditions of 20°C and 101.3 kPa
O2 concentration range: 100% O2
Hypoxic guard: Mechanical Link-25:
Flow range: 0 to >10 L/min Provides a nominal minimum
Auxiliary common gas outlet (optional) 25% concentration of oxygen
in O2/N2O mixture.
Connector: ISO 22 mm OD and 15 mm ID
Materials
Gas supply
All materials in contact with patient breathing gases are not
Pipeline input range: 280 kPa to 600 kPa made from natural rubber latex.
(41 psig to 87 psig)
Pipeline connections: DISS-male, DISS-female, AS4059,
S90-116, or NIST
Environmental Specifications
All fittings available for O2, N2O, and
Air, and contain pipeline filter and System operation
check valve. Temperature: 10° to 40°C (50° to 104°F)
Secondary O2 pipeline inlet available. Humidity: 15 to 95% relative humidity
Cylinder input: Pin indexed in accordance with (non-condensing)
CGA-V-1 or DIN-477 (nut and gland); Altitude: -440 to 3565 m
contains input filter and check valve. (500 to 800 mmHg)
Large cylinder kit available for O2 -440 to 4000 m
and N2O (with DIN-477). (without Airway Module)
Note: Maximum 3 cylinders (475 to 800 mmHg)
6
Breathing Circuit Specifications Expiratory
resistance in bag
mode:
Carbon dioxide absorbent canister
Pexp Pexp
Absorbent capacity: Reusable canister 1370 mL/1150 g Absorber canister Absorber canister
Disposable canister 1437 mL/1200 g Flow rate Installed Removed
All materials in contact with exhaled patient gases are Low vacuum, Low vacuum 12.7 mm hose barb,
autoclavable, except O2 cell, and Airway Modules. low flow: 25 to 50 L/min 25 mm hose barb,
All materials in contact with patient gas are not made ISO 1L or 30 mm ISO taper
from natural rubber latex. Passive: Passive system 30 mm/1.2 in
with air break M ISO taper
Breathing circuit parameters
Compliance:
Bag mode: 1.81 mL/cmH2O
(filled disposable absorber canister)
1.74 mL/cmH2O
(filled reusable absorber canister)
Mechanical mode: Automatically compensates for
compression losses within the
absorber and bellows assembly
Volume: 2006 mL Ventilator side
500 mL Bag side
1004 mL Reusable canister
985 mL Disposable canister
7
© 2015 General Electric Company – All rights reserved.
General Electric Company reserves the right to make
changes in specifications and features shown herein, or
discontinue the product described at any time without
notice or obligation. Contact your GE representative for
the most current information.
GE and the GE Monogram are trademarks of the
General Electric Company.
TM
Carestation, CARESCAPE, PSVPro, Tec, Selectatec,
Patient Spirometry, D-lite, Pedi-lite are trademarks of
General Electric Company.
Not for sale in all markets.
Please check with your sales representative.
Always refer to the complete instructions manuals
before use.
Datex-Ohmeda, Inc. a General Electric Company.
www.gehealthcare.com
GE provides transformational medical technologies and services that
are shaping a new age of patient care. Our broad expertise in medical
imaging and information technologies, medical diagnostics, patient
monitoring systems, drug discovery, biopharmaceutical manufacturing
technologies, performance improvement and performance solutions
services help our customers to deliver better care to more people around
the world at a lower cost. In addition, we partner with healthcare leaders,
striving to leverage the global policy change necessary to implement a
successful shift to sustainable healthcare systems.
Datex-Ohmeda, Inc.
3030 Ohmeda Drive
PO Box 7550
Madison, WI 53707-7550
USA
Carestation 620 A1
DOC1649440 Rev5