July 2018 A1c

This document is a quality control report from the laboratory department of an unnamed organization. It summarizes the results of monthly quality control and coefficient of variation monitoring for a Cobas analyzer when testing HbA1c levels over July 2018. Key metrics reported include the mean, standard deviation, and CV for two levels of control material tested across multiple runs and days. The lab mean, SD, and CV are also provided to check that the instrument is performing within acceptable specifications.

Uploaded by

ARIF AHAMMED P
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
70 views

July 2018 A1c

This document is a quality control report from the laboratory department of an unnamed organization. It summarizes the results of monthly quality control and coefficient of variation monitoring for a Cobas analyzer when testing HbA1c levels over July 2018. Key metrics reported include the mean, standard deviation, and CV for two levels of control material tested across multiple runs and days. The lab mean, SD, and CV are also provided to check that the instrument is performing within acceptable specifications.

Uploaded by

ARIF AHAMMED P
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
You are on page 1/ 1

Division/Department/Section: Laboratory Reference Number: QR/LAB/BIO-IMM-3.

2
Subject : Monthly QC and CV Monitoring - Cobas Interga Issue Date : 01.04.2018 Version: 03

JULY 2018
DATE L1 L2 Analyst Arif Ahammed
1 5.55 10.81 Test HbA1c
2 5.54 10.87 Control Material PC A1c N PC A1c P
3 5.65 10.95 Lot No. 203132 203133
4 5.44 10.65 QC LOT EXPIRY: SEP 18 SEP 18
5 5.57 10.75 T. Mean 5.65 11.3
6 5.62 10.85 T. SD 0.33 0.7
7 5.55 10.97 M - 2 SD 4.99 9.9
8 5.43 10.76 M + 2 SD 6.31 12.7
9 5.45 10.88 Manufacturer's CV 5.84 6.19
10 5.5 10.98 LEVEL 1
11 5.5 10.7
12 5.54 10.73 6.85
13 5.62 10.86
14 5.59 10.87
6.25
15 5.55 11.04
16 5.68 10.89
5.65
17 5.69 10.85
18 5.62 11.25
19 5.62 11.1 5.05
20 5.6 11.11
21 5.63 11.14 4.45
22 5.71 10.79
23 5.69 10.91 LEVEL 2
24 5.64 11.24
12.9
25 5.71 10.89
26 5.59 10.73
11.9
27 5.61 11.07
28 5.65 11.33
29 5.55 10.88 10.9
30 5.55 10.91
31 5.72 10.98 9.9
MEAN 5.59 10.93
SD 0.08 0.17 8.9
CV 1.41 1.53
LAB MEAN 5.65 10.9
SD 0.3 0.5
CV 5.31 4.59
M-2SD 5.05 9.90 CV IS MONITORED
M+2SD 6.25 11.90
M-4SD 4.45 8.90
M+4SD 6.85 12.90 Technologist Pathologist