Icpms2 PDF
Icpms2 PDF
GLOSSARY
Table of Contents
4.7 INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETRIC DETERMINATION OF
ARSENIC, CADMIUM, CHROMIUM, LEAD, MERCURY, AND OTHER ELEMENTS IN FOOD USING
MICROWAVE ASSISTED DIGESTION .................................................................................................................1
4.7.1 SCOPE AND APPLICATION .............................................................................................................................2
4.7.2 SUMMARY OF METHOD .................................................................................................................................2
4.7.3 EQUIPMENT AND SUPPLIES ...........................................................................................................................3
4.7.4 REAGENTS AND STANDARDS ........................................................................................................................5
4.7.5 DIGESTION PROCEDURE ................................................................................................................................8
4.7.6 METHOD QUALITY CONTROL ...................................................................................................................... 12
4.7.7 DETERMINATION PROCEDURE..................................................................................................................... 15
4.7.8 CALCULATIONS ........................................................................................................................................... 22
4.7.9 REPORT ....................................................................................................................................................... 23
4.7.10 METHOD VALIDATION ............................................................................................................................ 23
4.7.11 METHOD REVISION HISTORY ................................................................................................................... 24
4.7.12 REFERENCES ...................................................................................................................................... 24
FDA Elemental Analysis Manual (Section 4.7 ICP-MS Method)
Typical analytical limits were calculated per §3.2 and are listed in
4.7 Table 1 but will vary depending on the specific instrumentation, dilution factor and blank
quality. Significantly lower LODs and LOQs have been achieved for a number of target analytes
for different matrices and larger sample masses. Achieving the lowest limits will require
meticulous attention to operating conditions and the highest level of quality control for each set
of analyses.
Disclaimer: The use of trade names in this method constitutes neither endorsement nor
recommendation by the U. S. Food and Drug Administration. Equivalent performance
may be achievable using apparatus and materials other than those cited here.
Use high purity or trace metals grade reagents at all times. Blank levels will be < ASQL if using
best laboratory practices and high purity reagents.
Safety Notes: Reagents should be regarded as potential health hazards and exposure
to these materials should be minimized. Follow universal precautions. Wear gloves, a
lab coat, and safety glasses while handling reagents.
Exercise caution when handling and dispensing concentrated acids. Always add acid
to water. Acids are caustic chemicals that are capable of causing severe eye and skin
damage. If acids or bases come in contact with any part of the body, quickly wash the
affected area with copious quantities of water for at least 15 minutes.
Reagents
(1) Reagent water—Water meeting specifications for ASTM Type-I water3.
(2) Argon supply—High purity (99.99%) argon.
(3) Helium for collision cell—Ultra high purity (99.999%)
(4) High purity nitric acid—Concentrated (67-70%, sp. Gr. 1.42), double distilled. The
trade name for double distilled grade will vary by manufacturer.
(5) High purity hydrochloric acid—Concentrated (30-35%, sp. Gr. 1.18), double distilled.
Solutions
(1) Hydrochloric acid 10% (v/v)—Dilute 200 mL (236 g) high purity HCl to 2,000 mL with
reagent water.
Recommendation: Prepare solution in an empty bottle originally used for
concentrated hydrochloric acid. Dilute gravimetrically on a top loading balance
with a capacity of a least 2500 g. Tare bottle. Fill with approximately 1000 mL
reagent water. Note mass. Add approximately 200 g acid while pouring slowly from
the stock bottle. Add the remaining acid from a Teflon squeeze bottle to enable fine
control of acid addition. The total mass of concentrated hydrochloric acid added
should be 236 g (200 mL * 1.18 g/mL = 236 g). Add reagent water until a total
solution mass of ~2036 g is reached (1800 g water + 236 g HCl). Cap bottle and
mix.
(2) Diluent and rinse solution 5% HNO3 & 0.5% HCl (v/v)—Dilute 100 mL HNO3 (142 g)
and 10 mL (11.8 g) HCl to 2,000 mL with reagent water.
Recommendation: Use an empty bottle originally used for concentrated
hydrochloric or nitric acid. Dilute gravimetrically on a top loading balance with a
capacity of a least 2500 g if making 2L of solution. Tare bottle. Fill with
approximately 1000 g reagent water. Note mass. Add 11.8 g (10 mL) high purity
HCl (double distilled, 30-35%). Swirl to mix. Add 142 g (100 mL) high purity
HNO3 (double distilled, 67-70%). Dilute with reagent water to 2L or ~2044 g. It is
recommended to add concentrated acids either with a high purity bottle top acid
dispenser or Teflon PFA squeeze bottle.
(3) Internal standard solution (ISTD)—Multi-element solution prepared by diluting an
appropriate volume of stock standard. ISTD matrix is 1% HNO3, 0.5% HCl and 4%
isopropanol. The presence of isopropanol will help equalize arsenic and selenium
sensitivities due to residual carbon post digestion.4 The ISTD dilution factor is 1:1 if the
autosampler and internal standard peristaltic pump tubes are equal inner diameter. The
analytical solution pumped into the nebulizer will be approximately 2% isopropanol.
a. ISTD solution may be prepared volumetrically. The exact concentration is not as
important as maintaining the same concentration over an analytical sequence.
b. ISTD elements and suggested concentrations: 20 ng Ge/g, 2.5 ng Rh/g, 5 ng Ir/g,
and 2.5 ng Bi/g. These concentrations are only suggestions. Labs may alter at
their discretion as long as QC passes.
c. If using default Agilent tubing (1.02 mm i.d. sample white/white and 0.25 mm
i.d. orange/blue) then increase IPA and ISTD elemental concentrations
approximately 8x. The exact concentration is not important.
(4) Recommended Tuning Solution—2 µg/L Li, Co, Y, Ce, and Tl solution in 5% HNO3 –
0.5% HCl used to tune ICP-MS.
The method suggests sample tubing and ISTD tubing to be equal diameter,
diluting tune solution by 2×. Therefore, tune solution should be 2 µg/L so that 1
µg/L is aspirated into the ICP (see 4.7 Figure 2).
reaction subsides. If a clean hood is unavailable, place caps on vessels without pressing
down fully or, if so equipped, cap vessels but loosen the pressure relief nut (with the
safety membrane) to allow pressure to escape. If, however, it appears that excessive
foaming would result in the sample-acid mixture expanding out of the vessel then cap
the vessel and tighten to appropriate torque to prevent loss of sample or acid.
(5) Add 1 mL high purity 30% H2O2 to each vessel. It may be necessary to pre-digest for
more than 20 minutes before adding H2O2 if samples foam excessively.
(6) Seal vessels, apply correct torque to cap (tighten pressure relief nuts if equipped) and
run the digestion program in 4.7 Table 3.
(7) After vessels have cooled to less than 50 °C move to an exhausting clean hood and vent
excess pressure slowly. Quantitatively transfer digests to a clean container and dilute
digestion solution to approximately 50 g with reagent water followed by 0.5 mL (1.2 g)
high purity HCl (or 5 mL of 10% HCl solution). Add more reagent water for a final
volume of 100 mL and record final analytical solution mass. The mass of a 100 mL 5%
HNO3 – 0.5% HCl analytical solution is approximately 102 g.
Note: Quartz or TFM vessels are recommended. Ensure that samples are
completely wetted by the acid.
(10) After the digestion finishes, allow chamber temperature (T1) to cool to 60 °C and
release pressure no faster than 8 bar/min. Effervescence may occur at higher pressure
release rates resulting in sample loss.
(11) Move cooled and depressurized vessels to an exhausting clean hood. Quantitatively
transfer each digest to a clean container and dilute digestion solution to approximately
25 g with reagent water followed by 0.25 mL (0.6 g) high purity HCl (or 2.5 mL of 10%
HCl solution). Dilute with reagent water to approximately 50 g. Record weight of final
analytical solution to nearest 0.01 g.
(12) Assuming 50% oxidative acid consumption, the final matrix composition is 5% HNO3
and 0.5% HCl.
Failure of any of the QC elements described below to meet performance criteria requires
reanalysis of samples analyzed prior to the loss of method control measures. A single element’s
QC failure does not automatically fail other elements. For example, if Zn QC fails but As QC
passes, it is acceptable to report As results.
The following items are common causes of erroneous results and the associated QC items
follow:
Vessel contamination may be identified when analyzing duplicate analytical portions.
When RPD of two duplicate analytical portions exceeds 20% and concentrations > LOQ,
it is possible that one of the analytical portions has been contaminated. Analyzing
duplicate portions will not indicate if the bulk composite is contaminated.
Inadequate spectral interference mitigation is identified through measurement of
multiple isotopes. Where possible, EAM 4.7 calls for measurement of two isotopes.
Elemental concentrations calculated from each isotope should agree to within 20% RPD
when concentrations > LOQ. If they do not, this suggests a spectral interference at one of
the isotope m/z and should be further investigated and remedied.
Matrix effects may cause unwanted enhancement or depression of sample signal.
Internal standards and have been chosen to compensate for matrix effects, but are not
100% assured. Fortified analytical portion and fortified analytical solution QC failure can
indicate a matrix effect. FAPs and FASs do not correct for or indicate spectral
interferences (or indicate a lack thereof).
The following is the minimum number of quality control samples analyzed with each analytical
sequence:
2 method blanks (MBKs)
o Minimum of 2 MBKs and concentration of both MBKs are ≤ MBKC. If 3 or more
MBKs are analyzed then at least two-thirds of MBKs are ≤ MBKC (§3.6). MBKs
exceeding MBKC should be uncommon.
o If MBKC has not been established, subtract average MBK if MBK > ASDL as in
§4.7.8.
o If a failure occurs due to contamination, the source of contamination should be
investigated and remedied. One of the most common contamination issues is spot
(microwave vessel) contamination. When spot contamination is suspected, wash
vessels and run a digestion with vessels as MBKs. Place vessels back in the same
rotor position (or label vessels to keep track). Analyze MBKs and find possible
contaminated vessels. Repeat and/or remove contaminated vessels from use.
o If a failure occurs due to polyatomic interference, increase helium flowrate and/or
energy discrimination and reanalyze the entire sequence. If a failure is still
present, vessels should be thoroughly cleaned and new analytical portions must be
digested.
Stability Check
o Demonstrate instrument stability by analyzing a midrange multi-element standard
containing the analytes (e.g. CCV). Relative standard deviation (RSD) of ion
signals must be ≤10%. If RSD > 10%, determine and correct problem before
standardization. Stability problems are usually related to sample introduction.
1 certified reference material (CRM)
Match reference material matrix as closely as possible to the food matrix. In-house RMs
are acceptable if no CRM is available and/or the in-house RM is well characterized.
o RM % true value recovery: 80 – 120% when concentrations > LOQ or within
concentration uncertainty (converted to percent relative uncertainty) supplied on
certificate, whichever is greater.
o If acceptable values are not obtained, the analytical solution may be reanalyzed
once. If acceptability is still not met, recalibrate and reanalyze the entire analytical
sequence and/or prepare and digest new analytical portions.
Analyze duplicate analytical portions at a frequency of 10%. Analyze at least one
duplicate analytical portion of each food sample type and for any foods where non-
homogeneity is a concern. It is highly recommended that duplicate analytical portions are
analyzed for each food sample.
o RPD < 20% for replicate analytical portions when concentration > LOQ (§3.4.5)
o If RPD < 20% is not achieved, reanalyze replicate analytical solutions once. If
acceptable RPD is still not achieved, the source of imprecision should be
investigated and remedied. The entire analytical sequence may need to be
reanalyzed and/or new analytical portions be digested.
1 fortified analytical portion (FAP) per sample type. It is recommended that the food
sample analyzed in duplicate also be spiked for FAP.
o FAP % recovery can fail due to inappropriate fortification levels. We recommend
analyzing the food once as a ‘test’ sample and then fortifying at the proper levels
in a subsequent digestion (report only the second analysis result) or b) fortify
duplicate portions at ‘low’ and ‘high’ levels e.g. fortify one portion at ~100 ng/g
and a separate analytical portion at ~500 ng/g.
o FAP preparation: Spike 50-300% of the native elemental concentration into the
digestion vessel with the analytical portion. If the native concentration range is
unknown, spike at a low level standard analytical solution concentration – (i.e.
using 4.7 Table 2 as an example, spike at level 3 calibrant concentration).
o FAP % marginal recovery: 80 – 120%
o If acceptable recovery is not obtained, ensure spike level is appropriate and
reanalyze analytical solution once. If FAP fails again, reanalyze samples that
followed the last acceptable FAP. If FAP fails again, prepare and digest new
analytical portions.
1 fortified analytical solution (FAS) per sample type. It is recommended that the food
sample analyzed in duplicate also be spiked for FAS.
o FAS preparation: Spike 50-300% of the analytical solution concentration. If the
native concentration range is unknown, spike at a low level standard analytical
solution concentration – (i.e. using 4.7 Table 2 as an example, spike at level 3
calibrant concentration).
If FAS fails due to inappropriate fortification levels, then a new FAS
solution must be made.
o FAS % marginal recovery: 90 – 110%
o If acceptable recovery is not obtained, ensure spike level is appropriate and
reanalyze analytical solution once. If FAS fails again, reanalyze samples that
followed the last acceptable FAS. If FAS fails again, prepare a new FAS solution.
Optional fortified method blank
o FMB preparation: Spike approximately 2x ASQL – 8x ASQL
o FMB % marginal recovery: 90 – 110%
Method parameters are listed in 4.7 Table 5. Internal standards help compensate for matrix
effects and instrumental drift.
Instrument Setup
(1) See §3.6.1.4 for additional details on ICP-MS.
(2) Perform manufacturer recommended or laboratory start-up procedures.
(3) Program data acquisition method as shown in 4.7 Table 5.
a. Elements that will not be reported may be removed to save time. Ensure that
proper internal standard isotopes are still measured
b. Use spectrum helium mode and kinetic energy discrimination (Agilent specific
nomenclature – other manufacturers will have different names)
c. Reaction gases are not allowed
d. Program the autosampler probe to go to the rinse station for at least 10 seconds
after analyzing an analytical solution and then to a rinse bottle. Multiple (3) rinse
bottles are recommended. The rinse time must be great enough so that a standard
blank solution analyzed after the highest standard results in all analytes <ASQL
e. An “intelligent rinse” or “smart rinse” feature may be used if so equipped
Analyte levels must return to within 10% RPD of the average CCB before
moving to the next analytical solution
f. Use 3 points per peak and at least 3 replicates for integration. Use the mean of
the integrations for reporting
(4) Optimize instrument
a. Tune instrument according to the guidelines in the manufacturer’s tuning guide.
The instrument must exceed minimum manufacturer specifications.
Note: During tuning, the internal standard tubing is placed in reagent water.
b. HCl is added to the tuning solution to create chloride based interferences that
would be found in food samples high in salt.
c. Use at least 3 (typically 3 – 4) volts energy discrimination (difference between
octapole and quadrupole biases) and at least 3 (typically 3 – 5) mL/min He flow
rate. Higher He flow rate is allowable and recommended when analyzing ultra-
trace concentrations (< 10 ng/g) of isotopes that suffer from severe polyatomic
interferences such as 52Cr.
e. Keep a record of instrument parameters such as sample gas flow rate, sensitivity,
oxide formation, doubly charged ratio, and stability (count rate %RSD).
(5) Set up method interference correction equations. These equations can also be edited
after data has been acquired.
a. 3 Pb isotopes are summed to account for isotopic variations between standards
and samples. Use the method edit function to sum Pb isotopes.
i. 208: (206)*1 + (207)*1 + (208)*1
74
b. Ge has an isobaric interference with 74Se, which is especially problematic in a
selenium fortified dietary supplement or other selenium fortified food. The Ge
concentration in the ISTD should be high enough that an correction for 74Se on
74
Ge is unnecessary, but an analyst may choose to use 72Ge or 103Rh for high
selenium products.
78
c. Se has an isobaric interference with 78Kr. The Kr isotope is minor abundance but
is occasionally found as an argon impurity. Analysts may use a mathematical
correction for the Kr overlap of Se. Correction factors are calculated from the
natural abundance of Kr and do not incorporate mass bias factors.
i. 78: (78)*1 - (83)*0.031
ii. 82: (82)*1 - (83)*1.008
Recommended Optional
ISTD ISTD
103 105 74
Rh Pd, Ge
74 72 105
Ge Ge, Pd
209 205 175
Bi Tl, Lu
193 195 197
Ir Pt, Au
c. RSD of replicate integrations must be ≤ 10% for all solutions when instrument
response > 5 times ASQL or greater than 5,000 cps, whichever is greater. If RSD
exceeds 10%, determine source of noise and remedy before proceeding.
d. Continuing calibration blank (CCB) analyzed at a frequency of 10% and at the
end of the analytical sequence and must be < ASQL to proceed. If CCB fails,
reanalyze one time. If CCB fails again, reanalyze samples analyzed after the last
acceptable CCB. If CCB fails a third time, restart analytical sequence with a
longer washout time and/or prepare new digests, standards and QC solutions.
e. Analytical solution concentrations must be less than the highest standard
concentration. Gravimetrically dilute analytical solution if necessary.
(7) Suppression or enhancement of ISTD response may indicate a matrix effect is present.
Monitor internal standard signals and dilute any analytical solution where the internal
standard signal differs by more than 40% from the standard blank.
a. It is helpful to monitor 13C as an ISTD element. Carbon enhanced ionization can
cause false positives of high ionization potential isotopes. Use 13C as an easy
proxy to determine if samples may need further dilution due to high carbon
content (e.g. greater than 200% 13C vs standard blank).
b. In the same way as 13C, monitor 37Cl as an ISTD element. Foods high in salt
(NaCl) may form Cl-containing polyatomic interferences (see table 6).
Determination of Analyte Concentration Using Standard Additions
Quantification by the method of standard additions can also be used.
(1) Analyze 3 additional portions of analytical solution with added varying amounts of
analyte.
(2) Additions should be no less than 0.5 and no greater than 3 times native amount.
(3) Correlation coefficient (r) of linear regression must be ≥ 0.9975. If correlation
coefficient is < 0.9975, repeat analysis. If repeat analysis still fails to meet control limits
then dilute sample by a factor of 2 and re-analyze using additions based on the level in
analytical solution and the dilution factor.
4.7.8 CALCULATIONS
Calculate the concentration (mass fraction) of the analyte in the analytical portion according to:
𝜇𝑔 𝑀
𝐶𝑜𝑛𝑐𝑒𝑛𝑡𝑟𝑎𝑡𝑖𝑜𝑛 ( ⁄𝑘𝑔) = [(𝑆 × 𝐷𝐹) − 𝑀𝐵𝐾𝐿 ] ×
𝑚 × 𝑀𝐶𝐹
where:
𝑆 = concentration of analyte in analytical solution (or diluted analytical
solution) (ng/g)
𝑀𝐵𝐾𝐿 = laboratory MBK (ng/g) (subtract if average of the three MBK is greater than
ASDL)†
𝑀 = Mass (g) of analytical solution (usually 50 – 100 g)
𝑚 = mass of analytical portion (g)
𝐷𝐹 = dilution factor (1 if analytical solution not diluted)
𝑀𝐶𝐹 = mass correction factor (1 if no water or other solvent was added to aid
homogenization)
Report concentration to no more than 3 significant figures. Concentration may be converted to
other convenient units (e.g., mg/kg, ng/kg for solids or ng/L for liquids providing that the
specific gravity is known).
† MBKL subtraction may not be appropriate for all analyses or labs (i.e. when MBKL is not well
established or when multiple analysts work on a single analytical portion). In this case replace
MBKL in the above equation with the average MBK concentrations from the digestion batch
when MBK > ASDL.
𝐶𝑥+𝑠 − 𝐶𝑥
% 𝑅𝑒𝑐𝑜𝑣𝑒𝑟𝑦 = [ ] × 100
𝐶𝑠 𝑀𝑠
( 𝑀 )
𝑥
where:
𝐶𝑥+𝑠 = concentration determined in spiked sample (g/kg)
𝐶𝑥 = concentration determined in unspiked sample (g/kg)
𝐶𝑠 = concentration of spiking solution (g/kg)
𝑀𝑠 = mass of spiking solution added to analytical portion (g)
𝑀𝑥 = mass of analytical portion (g)
4.7.9 REPORT
Report results after all quality control criteria for an analytical sequence have been met. Report
average concentration when replicate analytical portions are analyzed.
Report results that are ≥LOQ as concentration followed by the units of measurement.
Report results that are ≥LOD and <LOQ as concentration followed by the units of
measurement and the “Trace” data qualifier that indicates analyte is present at a trace
level that is below the limit of reliable quantification. Trace values are documented by a
“TR” after the result.
Report results that are <LOD as 0 followed by the units of measurement and the qualifier
that indicates analyte is below the level of reliable detection or is not detected (ND).
Example: LOQ = 10 µg/kg; LOD =3 µg/kg. Levels found for three different
samples were 10 µg/kg, 5 µg/kg and 2 µg/kg.
In-house validation.
EAM 4.7 has undergone a level 2 single lab validation (SLV) as described in FDA’s Guidelines
for the Validation of Chemical Methods for the FDA Foods Program (Office of Foods and
Veterinary Medicine).6 The cumulative LODs and LOQs for method blanks measured over the
course of 1 year during the SLV (n = 143) are listed in table 1 and are adequate for the intended
purpose of this method. All quality control criteria were followed and met during the SLV.
Analyses were performed on 25 different foods that were similar to those collected in FDA’s
Total Diet Study but purchased from local grocers. Foods were analyzed several (N ≥ 5) times.
Fortified analytical portions (3 spike levels each) were also prepared and analyzed. Repeatability
and reproducibility of the method was < 20% relative standard deviation for elements with
concentrations > LOQ. Fortification and reference material recoveries were in the range of 80-
120%.
Uncertainty.
A result above LOQ has an estimated combined uncertainty of 10%. Use of a coverage factor of
2 to give an expanded uncertainty at about 95% confidence corresponds with the RM recovery
control limit of ± 20%. A result above LOD but below LOQ is considered qualitative and is not
reported with an uncertainty.
A more detailed discussion of method uncertainty is presented in §3.3. This method conforms to
Interlaboratory trial.
A mutli-laboratory validation is currently underway. An updated method will be published with
the multi-lab validation results when the exercise is complete.
4.7.12 REFERENCES
(1) Link, D., Walter, P., and Kingston, H. (1998) Development and Validation of the New EPA Microwave
Assisted Leach Method 3051A, Environ. Sci. Technol. 32, 3628-3632.
(2) U.S. Environmental Protection Agency (2007) SW-846 EPA Method 6020A rev. 1: Inductively Coupled
Plasma-Mass Spectrometry. Available from EPA
(3) ASTM International (2006) ASTM D 1193-06, Standard Specification for Reagent Water. Available from
ASTM
(4) Larsen, E., and Sturup, S. (1994) Carbon-enhanced Inductively Coupled Plasma Mass Spectrometric
Detection of Arsenic and Its Application to Arsenic Speciation, J. Anal. At. Spectrom. 9, 1101-1105.
(5) Official Methods of Analysis of AOAC INTERNATIONAL (2005) 18 th Ed., AOAC International,
Gaithersburg MD, USA, Appendix D: Guidelines for Collaborative Procedures To Validate Characteristics
of a Method of Analysis.
(6) U.S. Food and Drug Administration (2012) Guidelines for the Validation of Chemical Methods for the
FDA Foods Program, Version 1. Available from FDA
.