FlowSys TopFlow Multiphase Meter
FlowSys TopFlow Multiphase Meter
FlowSys TopFlow
Multiphase Meter
Operations Manual
Issue/Rev. 0.1 (5/05) Bulletin MNMP003
Contents
Section 1 — Introduction .......................................................................................................................... Page 3
Section 2 — TopFlow MultiPhase Meter ................................................................................................... Page 4
Sensor Unit ............................................................................................................................................... Page 5
Field Electronics ........................................................................................................................................ Page 6
Transmitters .............................................................................................................................................. Page 6
Computer Interface Enclosure ................................................................................................................... Page 6
Industrial Computer ................................................................................................................................... Page 7
User Interface ........................................................................................................................................... Page 7
Section 3 — Operating Instructions ......................................................................................................... Page 8
Operating Instructions ............................................................................................................................... Page 8
Manual Valves and Manifolds .................................................................................................................... Page 8
Switching On the Equipment ...................................................................................................................... Page 8
Switching Off the Equipment ...................................................................................................................... Page 8
User Interface Software (UIS) .................................................................................................................... Page 8
Well Tests ................................................................................................................................................. Page 9
Well Monitoring ......................................................................................................................................... Page 9
Section 4 — Related Publications ............................................................................................................. Page 10
Copyright
All technical information contained in this manual, including any drawings and technical specifications, shall remain the property of FMC
Measurement Solutions and may not be used (other than for the operation of this product), copied, multiplied, passed on or communicated to
a third party without prior written permission of FMC Measurement Solutions.
© 2004 FMC Measurement Solutions
Trademarks
Products listed are trademarks of their respective manufacturers. Company names listed are trade names of their respective companies.
Introduction
The FMC Measurement Solutions FlowSys TopFlow Multiphase Meter is your best choice for virtually any multiphase
measurement application. Consider these advantages:
• Measures oil, water, and gas flow rates without any mixing or fluid separation
• Accurate across a wide range of water cuts, viscosities, flow conditions, and salinities –all at a minimum pressure drop
• Totally free from radioactive sources, so it’s safe for both environment and operator
• Compact, lightweight design that’s easy to install and operate
• Uses standard, off-the-shelf parts for low-cost maintenance
• In-line units are flanged in as part of the piping connections to the customers specifications
• No moving parts
Hazardous Safe
Area Area
T Customer
B Data
a Acquisition
P r
Metering r
Processing
Section i
P e
Unit
r
s
Sensor FlowSys
Electronics User/Service
Software
(Ex-ia)
Sensor Unit
The sensor unit is designed and manufactured from materials and procedures in compliance with NACE, B31.3 B16.5.
The sensor unit is sized according to data provided by the customer at the time of the order.
The TopFlow Sensor Unit, Field electronics and transmitters are shown below in Figure 2.
Sensor
Unit
Transmitter
Field
Electronics
The TopFlow meter is installed in a vertical upwards position. There are no special requirements for upstream or
downstream piping applications.
Field Electronics
The Field Electronics is intrinsically safe and complies with the following standards:
Applicable CSA Standards C22.2
UL STD. 508
UL STD 913
UL/CSA E60079-0
UL/CSA E60079-11
EN50014
EN50020
Field Electronics:
CSA-US Listed 05CSA1606592
Class I, Div 1, Group C&D
Class I, Zone 0, Group IIB T5
AEx ia IIB T5
Ex ia IIB T5
Transmitters
The three types of process transmitters used on the TopFlow meter are:
1. Differential Pressure
2. Pressure
3. Temperature
FMC Measurement Solutions will select a transmitter from a qualified transmitter supplier. If a transmitter is specified by
the customer, FMC Measurement Solutions will supply the transmitters to the customers specifications as long as the
specified transmitter has no negative impact on the performance of the TopFlow meter.
The selected transmitter will comply with the following:
Type: Smart transmitters
Certification: Explosion proof, weather proof, intrinsically safe (CENELEC and FM/UL Standards)
Output: Two wire, 4-20 mA
Ambient temperature: -40 °C – 85 °C
The differential pressure transmitter has a five-way manifold with integral vents for each line, isolation for each line and a
equalizer valve normally equipped with the meter. During operation the vents and the equalizer valves are closed while
the isolation valves are open.
Separate documentation for each transmitter will be included in the meters documentation package.
Safety Barriers
Appropriate safety barriers are used for all transmitters and field electronics.
There are three fuses mounted on the RS422 Transceiver Board.
Power Supply
The TopFlow Multiphase Meter can be supplied with 100 - 240 VAC, 50/60 Hz..
An internal power supply is used to distribute all necessary voltages to the internal parts of the Multiphase Flow Meter.
Optionally the meter can be supplied as a 24 Vdc unit and power is distributed from the customer supplied 24 Vdc supply.
(The option of AC or DC power is specified at time of order.)
Industrial Computer
The computer processes all signals coming from the meter and calculates the flow rates of oil, water and gas. The
computer prepares these measurements for the TopFlow User Interface as well as the customer’s data acquisition
system.
User Interface
The user interface software (UIS) is developed on a LabView platform. The software can be installed using Windows 9X,
Windows 2000 or Windows XP.
The user interface is used for configuration of the TopFlow Multiphase Meter.
If the customers data acquisition system is not connected to the TopFlow meter, the TopFlow meter will be a stand-alone
system and the user interface can then be used to monitor and display the measured data or data can be collected on
discs and analysed on any other PC.
Operating Instructions
Operation
The TopFlow Multiphase Meter can be operated as a stand-alone unit and/or interfaced to customer’s data acquisition
system.
The operation of the TopFlow multiphase meter can all be accomplished via the TopFlow User Interface Software
located on the PC. The TopFlow User Interface Software is described in a separate User Interface Software Manual.
Differential Pressure
There is normally a five-way manifold connected to the Differential Pressure (DP) transmitter. During normal operation
the valves should be in the following position:
• 1 x Equalize : Closed
• 2 x Isolate : Open
• 2 x Bleed : Closed
In addition the meters could be equipped with double block valves to isolate the piping from the instrument tubing. During
normal operation these valves are open.
Pressure
There is normally a two-way manifold connected to the pressure transmitter. During normal operation the valves should
be in the following position:
• 1 x Isolate : Open
• 1 x Bleed : Closed
In addition the meters could be equipped with double block valves to isolate the piping from the instrument tubing. During
normal operation these valves are open.
Well Tests
There is flexibility in using the TopFlow meter for well testing.
The well testing can be controlled from the TopFlow User Interface software or from the customer’s control system.
Two examples of steps required for well testing using TopFlow User Interface Software are shown below:
Selecting a Well Test Start and Stop Times While Viewing the Historical Display
A well test can also be defined while viewing historical data. The advantage is that a time period where the well has been
flowing stable can be selected as the time period for the well test.
1.Enter the historical display in the UIS.
2.Select a time window for the well test.
3.View or save the results to a file.
If data is transferred to the customer’s data acquisition system, well testing can be defined and reported by the
customer’s control system using the available options in that system. In such a case there is no need for the TopFlow
meter to know the start and stop timings of the well test.
Well Monitoring
There is no routine operation required for using the TopFlow meter for well monitoring.
If the fluid properties such as base densities or water salinity change, the parameters must be updated through the User
Interface Software.
For well monitoring it can be useful to set the long term averaging time to 24 hours. The daily production will then be
displayed continuously in the User Interface.
For well monitoring the TopFlow meter can be set to print a report for instance once a day (if a printer is connected to the
TopFlow computer). This report can also be written to a file.
Additional literature can be obtained from FMC Measurement Solutions Literature Fulfillment at [email protected] or
online at www.fmcmeasurementsolutions.com. When requesting literature from Literature Fulfillment, please reference
the appropriate bulletin number and title.
The specifications contained herein are subject to change without notice and any user of said specifications should verify from the manufacturer that the specifications are
currently in effect. Otherwise, the manufacturer assumes no responsibility for the use of specifications which may have been changed and are no longer in effect.