Cardiocare 2000 Operation Manual
Cardiocare 2000 Operation Manual
OPERATION MANUAL
Ver. 2.41
www.ebionet.com
CardioCare2000 Operation Manual
Terms of Warranty
- This product is manufactured and passed through strict quality control and
thorough inspection. Compensation standard concerning repair,
replacement, refund of the product complies with “Consumer’s protection
law” noticed by Ministry of Finace and Economy.
2 / 61
Ver. 2.41
CardioCare2000 Operation Manual
How to reach us …
The following are telephone numbers and addresses for contacting various service,
product supplies and sales personnel
※ In the event of a malfunction or failure, contact Service Dept. of Bionet Co., Ltd along with
the model name, serial number, date of purchase and explanation of failure.
※ If you need the supply circuit diagram, component list, description and calibration instruction
etc. you can contact us we will provide you with it.
3 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Warning
To inform that it may cause serious injury or death to the patient, property
damage, material losses against the “Warning” sign.
Caution
To inform that it may cause no harm in life but lead to injury against the
“Caution” sign.
Note
4 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Service Requirements
Refer servicing for equipment to Bionet Co., Ltd authorized service personnel. Any
attempt to repair equipment under warranty will void that warranty.
It is the responsibility of users requiring service to report the need for service to Bionet
Co., Ltd., or to one of their authorized agents.
Failure on the part of the responsible individual, hospital, or institution using this
equipment to implement a satisfactory maintenance schedule may cause undue
equipment failure and possible health hazards.
If there are any problems with the equipment, please follow the steps below:
Contact the Bionet Oversea Service Department immediately. After gathering the
model name, Serial Number, date of purchase, and description of the problem
contact Bionet with the information shown below.
Try to solve the problem over the phone with the service department personnel. If
the problem cannot be solved over the phone, the service personnel can came and
fix the problem directly.
5 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Avoid placing in an
area exposed to
Avoid exposure to
moist. Do not touch the
direct sunlight
equipment with wet
hand.
Avoid placing in an
area where there is a
high variation of
temperature.
Operating Avoid in the vicinity
temperature of Electric heater
ranges from 10C
to 40C. Operating
humidity ranges
from 30% to 85%.
Avoid placing in an
Avoid placing in an
area where there is
area where there is
an excessive
an excessive shock or
humidity rise or
vibration.
ventilation problem.
Avoid placing in an
area where Avoid being inserted
chemicals are dust and especially
stored or where metal material into
there is danger of the equipment
gas leakage.
6 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Note
The equipment should not be placed in the vicinity of electric generator, X-ray, broadcasting
apparatus to eliminate the electric noise during operation. Otherwise, it may cause incorrect
result.
Self-power line and stabile grounding is important for EKG 2000.
To use same power source with other electric instruments may cause incorrect result.
Note
Note
Note
Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC standards ( e.g. IEC 950 for data processing equipment
and IEC 601-1 for medical equipment ). Furthermore all configurations shall comply with the
system standard EN 60601-1-1:1993.
Everybody who connects additional equipment to the signal input part or signal output part
configures a medical system, and is therefore responsible that the system complies with the
requirements of the system standard IEC 601-1-1:1993. If in doubt, consult the technical
service department or your local representative. Diagnosis has to be confirmed by the
doctor.
7 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Safety Symbols
Symbols contents
Identifies the point where the system safety ground is fastened to the
chassis. Protective earth connected to conductive parts of Class I
equipment for safety purposes.
8 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Table of Contents
9 / 61
Ver. 2.41
CardioCare2000 Operation Manual
10 / 61
Ver. 2.41
CardioCare2000 Operation Manual
1) Introduction
3) Equipment description
Standard Accessories
Options
Main Unit
LCD Panel
Control Panel
AC Power
4) System Installation
5) Menu Structure
6) System Settings
Initial Setting
Date and Time
User Identification
Language
11 / 61
Ver. 2.41
CardioCare2000 Operation Manual
1) Introduction
EKG2000 is 12ch ECGs equipment which measure and record ECGs of the patient.
EKG2000 provides users with ECGs record of the patient, measurement parameters for
diagnosis and auto analysis.
Patient information and user identification printed along with ECGs on the output report is
effective for hospital chart control.
For convenient operation, ECGs can be measured and recorded by pressing short key. Then,
ECGs is printed in A4 sized report form after applying filter, calculating the measurement
parameter and auto analysis.
In addition to an AC power, it can be continuously operated with a built-in rechargeable battery.
This enables the EKG2000 to be used in an ambulance, in a visit to patients or during patient
transportation.
12 channel ECGs is printed in a various channel form of 3ch+1rhy, 6ch+1rhy, 12ch rhy, 60s 1rhy
in A4 sized report form.
1. 1 channel rhythm is recorded for 60 seconds and printed in A4 sized report form.
2. 12 channel rhythms are printed simultaneously and continuously in real time..
3. Measurement parameters such as heart rate, PR interval, QRS duration, QT/QTc, P-R-T
axes required for diagnosis is printed along with ECGs on the output report after calculating
automatically.
4. Auto analysis enables 75 diagnosis.
5. For convenient diagnosis, ECGs can be printed after changing system setting such as filter,
signal level, printing speed, channel form, rhythm cannel of the recorded ECG data.
6. Easy to carry with built-in rechargeable battery.
7. Patient information and user identification can be entered and printed along with ECGs for
effective hospital chart control.
12 / 61
Ver. 2.41
CardioCare2000 Operation Manual
3) Equipment description
Standard accessories
④ Chest electrode(1SET)
Options
① Cart
② Hanger
③ Built-in Battery
13 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Main Unit
▣ Front View
①
⑤
② ③
① Knob
② Printer door
③ Printer door release button
④ LCD
⑤ Control panel
14 / 61
Ver. 2.41
CardioCare2000 Operation Manual
▣ Rear view
① ②
③ ④ ⑤
③ AC power socket
⑤ LAN port
① Handle
15 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Warning
Do NOT disassemble the case of the equipment to avoid the electric shock. Refer servicing
to Bionet Co., Ltd qualified personnel or your local authorized dealer.
16 / 61
Ver. 2.41
CardioCare2000 Operation Manual
LCD Panel
LCD panel displays system setting status after indicating the version of the system and
manufacturer name for 2 seconds when power is turned on.
① ② ③ ④
1 0 1 2 . 5 O F F O F F
1 0 0 6 C H + 1 R H Y a V R
⑤ ⑥ ⑦
① Display ECG signal level out of 5, 10, 20, aut (Auto Gain)
② Display printing speed out of 12.5, 25, 50
③ Display base line filter setting on or off
④ Display EMG filter setting on or off
⑤ Display heart rate
⑥ Display channel form of the output report out of 3ch+1rhy, 6ch+1rhy, 12ch rhy,
60s 1rhy.
⑦ Display rhythm channel setting out of I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5,
V6..
17 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Control panel
▣ Indicator lamp
In this case, check off which one of lead connection is in lead-off status
through monitor mode output.
Indicate the battery capacity status: Green and Red colors are used
to display 2 different statuses.
The battery indicator is red when the device is operational without having
an A/C power supply. When an A/C power is supplied, the battery
indicator will be either red or green depending on the charge level. No
color will be shown, if there is no battery installed in the device.
▣ Short keys
Select on or off whether or not to activate the filter that eliminates base
line drift
Select on or off whether or not to activate the filter that eliminates EMG.
ECG signal can be distorted by applying this filter.
18 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Call up the menu to enter patient ID, name, age, sex, height, weight
Print the same report as previously printed one in record mode or prints
after changing system setting such as filter, signal level, output speed,
channel form, rhythm cannel of previously recorded ECGs data in record
mode.
19 / 61
Ver. 2.41
CardioCare2000 Operation Manual
AC Power
When AC power is supplied to the equipment, AC power indicator lamp at the front side is
illuminated green and battery is being charged automatically.
If no battery is installed, battery indicator lamp will be off. Once a battery is plugged in and
the device is powered on, it will change its power mode to supply power to the battery. If
the device is running on the power from the battery, battery charge indicator lamp will light
up red color. If the battery is running low on power, three continuous buzzer alarms will
sound. If this occurs, turn off the device and connect the A/C power before turning the
device back on.
Caution
For the protection of environment, do NOT throw wasted battery away. Please inquire at
the biomedical engineering lab of the hospital and dispose according to the appropriate
procedure to the authorized place(according to national regulations).
20 / 61
Ver. 2.41
CardioCare2000 Operation Manual
4) System Installation
Installation precautions
- Check the power cord connection and probe should be handled with care.
- The entire system settings are stored on the internal memory even when system
power is turned off and on again
The equipment operates when the power cable end connect into the EKG2000 power cord.
Replace the fuse of the equipment if power is not turned on despite that power cord is connected
and battery power is in normal condition. Contact service dept. of Bionet Co.,Ltd if the equipment is
not operated after replacing fuse.
Caution
For the continuous protection against the fire hazard, the fuse must be changed to the
same type and rate of fuse by technician (T3A250V).
For the protection of environment, do NOT throw wasted fuse away in a dangerous place
for the example fire hazard
21 / 61
Ver. 2.41
CardioCare2000 Operation Manual
- Connect the patient cable into the connection port at the rear of the equipment.
- Connect the limb electrodes to the N, F, R, L lead and chest electrodes to C1, C2, C3, C4,
C5, C6 of the patient cable.
22 / 61
Ver. 2.41
CardioCare2000 Operation Manual
5) Menu structure
EKG2000 system provides 3 basic modes with preparation mode, output mode, and menu
mode.
Preparation mode is an initial condition when system starts operating. LCD displays system
settings and heart rate.
In preparation mode, system setting can be changed by pressing each short keys SENS,
SPD, BASE, MUSC, FORM, LEAD, PAT, MON, REC, COPY of the control panel. key
calls up the menu mode.
Output mode is to activate print operation in the preparation mode with pressing short keys
MON, REC, COPY. Only with MON, REC, COPY, key of the control panel stops
printing in output mode.
Menu mode is activated by pressing key in preparation mode and following window
displays on the LCD.
First line indicates menu selection and second line indicates an explanation of the menu items
or selected value.
< PATient Info > Enter the patient information including patient ID, name,
PAT
age, sex, height, weight
< FiLTer setup > Select and adjust filter setting of base line drift filter, EMG
FLT
filter, ac filter, low pass filter.
< PriNter setup > Select and adjust signal level, printing speed, grid,
PRN
numbers of channel, rhythm channel select, test printout
< SYStem setup > Enter system setting including date, time, hospital name,
SYS
user name.
23 / 61
Ver. 2.41
CardioCare2000 Operation Manual
6) System Setting
Initial setting
Press 1key of the control panel for 3 seconds to initialize the system at the factory preset. After
3 seconds, following window message appears on the LCD for 1sec and system sets to initialize.
* FACTORY MODE *
initialize
24 / 61
Ver. 2.41
CardioCare2000 Operation Manual
To set the present date of EKG2000, move the menu to * DATE SETUP * as following steps;
SYS select --
2000-03-01
DATE Select --
* DATE SETUP *
2000-03-01
, ESC
Number is entered by pressing 0,…, 9 numeric key in the current cursor position and cursor
automatically moves to the right . To exit the date setting press key or ESC key that
25 / 61
Ver. 2.41
CardioCare2000 Operation Manual
User identification
User identification is to enter the hospital name and doctor name that operates the
system. To enter the hospital name, move the menu to * ENTER HOSP. * as
following steps.
SYS select --
ESC
HOSP select -- , ▷
* ENTER HOSP *
_
, ESC
Hospital name can be entered in up to 16 characters. To enter the name, select the required
character in the current cursor position. As for [1ABC] key, the key toggles in the order of A, B, C,
1, A, …whenever pressing that key.
So, if you want to enter B character, press ▷ key to input B after pressing the [1ABC] key 2
times in the current cursor position. Then moves the cursor to the next. To exit the hospital
name setting, press key or ESC key that moves to the menu above.
To enter the user name who operates the system, move the menu to SYS DOC * ENTER
DOCTOR * as shown above. Then enter the user name in the same way.
26 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Language
To set a language for CARDIOCARE-2000, press and hold the ◁ key and turn the power on.
You will then be redirected to the “Language” menu. In the ‘Language’ menu, press a
number key (1 through 7) to select a desired language (1: English; 2: Chinese; 3: German; 4:
French; 5: Russian; 6: Spanish; and 7: Italian).
After selecting a language, press the key and the selected language will be applied to
the system.
Language Set
ENGLISH
27 / 61
Ver. 2.41
CardioCare2000 Operation Manual
1) Electrode positions
2) Electrode Connection
Check before connection
Connecting the Electrodes
Connecting the patient cable
Lead-off troubleshooting
28 / 61
Ver. 2.41
CardioCare2000 Operation Manual
1) Electrode positions
To record the 12 standard leads ECGs [ I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 ] ,
connect the electrodes to the patient as follows ;
29 / 61
Ver. 2.41
CardioCare2000 Operation Manual
2) Electrode connection
Check before connection
Check the system connection before measuring the patient’s ECG.
- Check that there’s no mechanical hazard.
- Check the connection of external cable and accessories.
- Check all the measurement device to measure the patient’s
Put the patient to bed in relaxed state and apply some ECG GEL. Connect the electrodes
described in “Electrode Positions”
Warning
Use only electrodes and patient cables provided from Bionet Co.,Ltd. Otherwise Bionet
Co.,Ltd takes no responsibility for accidents.
Lead-Off troubleshooting
If lead connection status is O.K, READY indicator lamp lights up when power switch is turned
on. If indicator lamp still does not light up after 10 seconds, it indicates lead-off status.
You can detect which one of leads is in lead-off status from operating MONITOR mode output.
Lead-off status happens in 2 cases.
First, lead comes off from the patient body. In case of this, connect the lead to the patient body
again.
Second, Conductivity between patient body and electrodes is poor. Incase of this, apply some
ECG gel to the electrodes and reapply it to the patient..
In exceptional circumstances if you are still troubled with poor EKG signal after troubleshooting
as described above, it can be a problem with the patient cable. Contact Service Center of Bionet
Co.,Ltd..
30 / 61
Ver. 2.41
CardioCare2000 Operation Manual
1) Introduction
2) Basic Setting
General Information
Signal Level Setting
Signal Display Setting
Filter Setting
Channel Form Setting
Rhythm Lead Setting
Grid Setting
Monitoring Form Setting
Monitoring Channel Setting
Diagnosis on/off setting
Enter patient data
31 / 61
Ver. 2.41
CardioCare2000 Operation Manual
1) Introduction
1. Turn the power switch on after connecting leads to the patient according to preparation
for ECGs recording of Chap2.
2. Set the filter, signal level, display speed, channel form, rhythm lead according to “Basic
Setting” of Chapter3..
3. Enter the patient information according to “Basic Setting –Enter Patient Data” of
Chapter 3.
4. Press the REC key to record patient ECGs after READY indicator lamp lights up
5. In case READY indicator lamp does not light up, press MONI key to check out
the ECG signal. If you are still troubled with poor signal, take necessary steps
according to “Lead-off troubleshooting” of Chapter 2 .
6. Press COPY key to print the same report as previously printed one. Press
COPY: Prints the same report as previously printed one in record mode
or prints after changing system setting such as filter, signal level, output
speed, channel form, rhythm cannel of previously recorded ECGs data in
record mode
32 / 61
Ver. 2.41
CardioCare2000 Operation Manual
2) Basic setting
General information
LCD displays current system setting in the order of signal level, printing speed, base line
filter ,EMG filter, heart rate, channel form, rhythm channel when system is turned on . These
settings can be changed by two kinds of methods. First method is to use short keys. System
setting can be easily changed by pressing short keys of control panel that accord with the icons
displayed on the LCD screen. Second method is to use menu. Once, move to menu mode while
LCD displays current system setting by presage key and then select “PAT”, “FLT”, “PRN”.
“PAT” can be used to enter the patient information. , “FLT” can be used to select the filter.
“PRN” can be used to adjust printing.
The signal level should be adjusted in case that signal overlaps with neighboring traces or the
amplitude of pulse is too small to interpret.
Signal Level setting is adjustable in four ways. The entire 12 channel is set to each out of
5mm/mV, 10mm/mV, 20mm/mV and aut (Auto gain control) which sets I, II, III, aVR, aVL, aVF to
10mm/mV , V1, V2, V3, V4, V5, V6 to 5mm/mV .
As for 5mm/mV, it indicates that the amplitude of 1mV pulse is amplified into 5mm.
To change the signal level, press SENS key to select the required one.
SENS key toggles between 5, 10, 20, aut that are displayed on the LCD.
The selected setting is shown on the bottom line of the output report indicated by ‘all channels:
5mm/mV, all channels: 10mm/mV, all channels: 20mm/mV’ as for 5, 10, 20 and ‘ I-aVF’:
10mm/mV, V1-V6: 5mm/mV ‘ as for aut.
33 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Filter Setting
AC power noise, base line drift caused by breathing , EMG signal can be recorded along with
the output ECG signal. To set the AC noise filter, move to “FLT” menu to call up “AC” menu
and press key. key toggles between off, 50hz, 60hz. Off indicates that filter does
not suppress ac interference. As for 50Hz or 60Hz, filter suppresses 50Hz or 60Hz AC power
noise. In case of battery power, ECG signal is clear only with Off setting because AC noise
interference hardly exist. The selection is 50Hz for most of Europe and 60Hz for USA and other
countries. The selected setting is shown on the bottom line of the output form indicated by Cut
off AC.
Base line drift is caused by patient breath. Base line drift filter can be set by pressing key.
BASE key switches on or off. On or off indicates base lien filter is applied or not. The
selected setting is shown on the bottom line of the output form indicated by ‘0.1hz-‘ as for filter
on and ‘0hz-‘ as for filter off.
EMG signal is induced from muscle and internal organs of the patient.
For clear ECG signal output, it is necessary to eliminate EMG signal interference.
EMG filter can be set by pressing MUSC key. MUSC key switches on or off. On
or off indicates EMG filter is applied or not. The selected setting is shown on the bottom line of
the output form indicated by Cut off EMG.
Low pass filter can be selected to optimize ECG signal after applying above three filters and if
ECG signal is still interfered. To set the low pass filter, move the menu to “FLT” to call up “LPF”
menu and press key. key toggles between off, 40hz, 100hz, 150hz. As for 40Hz,
it indicates filter suppresses higher frequency than 40Hz. The selected setting is shown on the
bottom line of the output form indicated by ‘-250hz-‘ as for off, by’-40hz’ as for 40hz, by’-100hz’
as for 100hz, by ’-150hz’ as for 150hz.
To apply filter could result in ECG distortion. So, filter should be appropriately applied to reduce
the fluctuation without affecting the ECG signal when output signal is badly interfered only.
For your reference, base line file and ac filter should be always active and EMG filter can be
applied appropriately. It is recommended for LPF to set 150Hz during diagnosis.
34 / 61
Ver. 2.41
CardioCare2000 Operation Manual
As for 6ch+1rhy, 10 seconds of ECG data are recorded such as I, II, III, aVR, aVL, aVF for first 5
seconds time segment and V1, V2, V3, V4, V5, V6 for next 5 seconds time segment. And 1
channel rhythm is recorded for 10 seconds in lower part of the output form.
As for 12ch rhy, 12 channel rhythms are simultaneously and continuously recorded in the order
of I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6.
As for 60s 1rhy, 1 channel rhythm is recorded with 6 lines, 10 seconds to every line, for 60
seconds.
To set the channel form, press FORM key. FORM key toggles between 3ch+1rhy, 6ch+1rhy,
12ch rhy, 60s 1rhy that are displayed on the LCD.
The selected setting is shown on the top line of the output form indicated by ‘3 Channels +
1Rhythm Report’ as for 3ch+1rhy, ‘6Channels + 1Rhythm Report’ as for 6ch+1rhy, ‘12Channels
Rhythm Report’ as for 12ch rhy, ‘1Rhythm Report (60sec) as for 60s 1rhy.
Rhythm channel setting is to adjust the lead of output form. To set rhythm lead, press LEAD key.
LEAD key toggles between I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 that are displayed on
the LCD. The selected rhythm lead recorded in lower part of the output form for 10 seconds is
the standard lead compared with other channel.
35 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Grid setting
On
key switches on or off . If set to on, grid is recorded on the recording paper.
If set to off, grid is not recorded on the recording paper.
36 / 61
Ver. 2.41
CardioCare2000 Operation Manual
To set the monitoring form, moves the menu to * MONI * as following steps.
CONTINUE
key switches continue or A4 report. If set to continue, the lead will be printed out
continuously until you press the key on the control panel. If set to A4 report, print
37 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Monitoring channel setting is to adjust the numbers of channel of monitoring form. This system
provides four kinds of channel form of 3ch, 6ch, 12ch.
To set the monitoring channel, moves the menu to * M-CH * as following steps.
12
key switches 3, 6 and 12 . If set to 3, print 3 channels at the same time (I ~ III / aVR ~
aVF / V1 ~ V3 / V4 ~ V6) If you want to change the channel, press the on the
control panel.
If set to 6, print 6 channels at the same time (I ~ aVF / V1 ~ V6) If you want to change
38 / 61
Ver. 2.41
CardioCare2000 Operation Manual
On
key switches on or off . If you set up ON, the result of diagnosis is printed out, and if you
set up OFF, the result of diagnosis is not printed out on the report.
39 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Patient information including patient ID, name, age, sex, height, weight can be entered. There
are two methods to enter patient data. First, move the menu to patient data menu by pressing
PAT key. Second, move the menu to “ID” as following steps.
When there are previous patient data, following window appears on the LCD whether to change
data of the previously recorded one or to select new patient.
New Patient?
yes no Select No
Select ID
* Enter ID *
40 / 61
Ver. 2.41
CardioCare2000 Operation Manual
All the other patient data are deleted with selecting yes and then move the menu to “ENTER
ID” for new patient data entry.
Previous patient data are displayed on the LCD as follows with selecting no;
New Patient?
yes no
Select No
12345678
Select ID
* Enter ID *
12345678_
To enter patient ID, move the menu to “ENTER ID” as following steps;
Select ID
* Enter ID *
12345678_
41 / 61
Ver. 2.41
CardioCare2000 Operation Manual
To enter patient name, move the menu to “ENTER NAME ” as following steps;
Select PAT
Select Name
* Enter ID *
Patient name can be entered in up to 16 characters. To enter the name, select the required
character in the current cursor position. As for [1ABC] key, the key toggles in the order of A, B, C,
1, A, …whenever pressing that key.
So, if you want to enter B character, press ▷ key to input B after pressing the [1ABC] key two
times in the current cursor position. Then moves the cursor to the next. To exit the patient name
press key or ESC key that moves to the menu above.
42 / 61
Ver. 2.41
CardioCare2000 Operation Manual
PAT 선택
ID NAME AGE >>
Select Age
* Enter Age*
years
Age can be entered in up to 3 figures. Age is entered by the unit of year. Age is entered with
pressing 0,…, 9 numeric key in the current cursor position and cursor automatically moves to
the right. To exit Age press key or ESC key that moves to the menu above.
43 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Select PAT
Select SEX
* Enter Sex *
male female
Sex can be selected between male and female with ◁▷ key. To exit sex press key or
ESC key that moves to the menu above.
Select PAT
Select HEI
* Enter Height *
_ cm
Height can be entered in up to 3 figures. Height is entered by the unit of cm. Height is entered
with pressing 0,…, 9 numeric key in the current cursor position and cursor automatically moves
to the right. To exit height press key or ESC key that moves to the menu above.
44 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Select WEI
* Enter Weight *
_ kg
Weight can be entered in up to 3 figures. Weight is entered by the unit of kg. Weight is entered
with pressing 0,…, 9 numeric key in the current cursor position and cursor automatically moves
to the right. To exit Weight press key or ESC key that moves to the menu above.
45 / 61
Ver. 2.41
CardioCare2000 Operation Manual
BASE LINE On
Muscle filter Off
AC Rejection Filter 60hz
Low Pass Filter 40hz
Printing speed can be set to 12.5 mm/s, 25mm/s, 50mm/s and signal level can be set to
5mm/mV, 10mm/mV, 20mm/mV, aut.
Output method
Filter setting value is fixed. Set the required printing speed and signal level and press MON
key. Then, the system starts printing according to the selected speed with the following
message on the LCD.
* MONITOR *
Printing
Press key or key to stop printing. Then system stops printing and displays system
setting after indicating following message for 1 second on the LCD.
* MONITOR *
aborted!
Output form
In monitor mode output, 12 channel rhythms are printed simultaneously and continuously as
shown below.
Indicated heart rate is the average of previous 8 heart rates.
If the MONI M-CH is set at 3 or 6, please press key to start printing.
When you press key again the lead is changed to next group and the printing starts.
If you want to stop in the middle of printing, press the key.
46 / 61
Ver. 2.41
CardioCare2000 Operation Manual
In record mode output, ECGs is recorded on the memory first and displayed according to the
selected signal level, output display speed, channel form after applying the selected filter and
calculating the measurement parameter including heart rate, PR interval, QRS duration, QT/QTc,
P-R-T axes of the recorded ECGs.
Press key when channel form is set to 3ch+1rhy, 6ch+1rhy, 12ch rhy
Then, the system starts recording data for 10 seconds on the memory with the following
message indicates recording time on the LCD.
* RECORD(10s) *
10sec
After recording data for 10 seconds, the system starts applying the selected filter and calculating
the measurement parameter of the recorded ECGs with the following message on the LCD.
* RECORD(10s) *
Processing
After data processing , the system starts printing with the following message on the LCD.
* RECORD(10s) *
Printing
Press key to exit during recording or printing data. Then, system stops recording or
printing and displays system setting after following message for 1 second on the LCD.
* RECORD(10s) *
Aborted!
47 / 61
Ver. 2.41
CardioCare2000 Operation Manual
* RECORD(10s) *
60 sec
After recording data for 60 seconds, the system starts applying the selected filter and calculates
the heart rate of the recorded ECGs with the following message on the LCD.
* RECORD(60s) *
Processing
After data processing , the system starts printing with the following message on the LCD.
* RECORD(60s) *
Printing
Press key to exit during recording or printing data. Then, system stops recording or
printing and displays system setting after indicating following message for 1 second on the LCD.
* RECORD(60s) *
Aborted!
48 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Output form
The printing forms are described and the example of printing form by each channel form is
provided below.
49 / 61
Ver. 2.41
CardioCare2000 Operation Manual
50 / 61
Ver. 2.41
CardioCare2000 Operation Manual
51 / 61
Ver. 2.41
CardioCare2000 Operation Manual
52 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Output form
These are the explanation of the output form and examples according to each channel form.
53 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Copy mode output function is to print the same report as previously printed one or print after
changing setting such as filter, signal level, output display speed, channel form, rhythm cannel
of previously recorded ECG data .
System starts printing the same report as previously printed one with the following message on
the LCD by pressing key after record mode printing.
* COPY *
Printing
If copy mode printing is tried without operating record mode printing, system displays system
setting after indicating following message for 1 second on the LCD.
* COPY *
If filter setting is changed to the previously recorded ECG data, the system starts applying
changed filter with the following message on the LCD.
* COPY *
processing
After filter processing, the system starts printing with the following message on the LCD.
* COPY *
printing
After printing is finished, the system displays system setting after indicating following message
for 1 second on the LCD.
* COPY *
Finished!
Press key to exit during printing data. Then, system stops printing and displays system
setting after indicating following message for 1 second on the LCD.
* COPY *
Aborted!
54 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Warning
Warning
When you are connecting to the electrodes or the patient cable, make sure that the
connectors never come into contact with other conductive parts, or with earth. In
particular, make sure that all of the ECG electrodes are attached to the patient, to
prevent them from contacting conductive parts or earth.
Warning
Use the provided ECG patient cable for measuring ECG in the operating room. Patient
cable cannot be used for measuring respiration.
Caution
Caution
Users must use same type of electrode or any other biocompatibility certified ones that
have been proved by International Standard.
Caution
This device must be operated under health care professional when applying to
pacemaker treated patient.
55 / 61
Ver. 2.41
CardioCare2000 Operation Manual
2) Regular Inspection
3) Troubleshooting
56 / 61
Ver. 2.41
CardioCare2000 Operation Manual
The warranty does not cover the damage resulting from using materials (unauthorized material)
which may cause damage of the product even during the warranty period.
Caution!
Check the main unit and electrodes thoroughly after cleaning. Do not use the old and
damaged equipment. Please consult with your hospital biomedical Engineer to
determine the device condition.
Main unit and electrodes should be wiped with soft cloth dampened with warm water to keep
clean. Wipe with absorbent cotton soaked with alcohol once a month. Do not use lacquer,
ethylene, acidic liquids.
Cable should be kept clean not to stain dust or filth and wiped with soft cloth dampened with
warm water(40˚C / 104 F). Wipe with absorbent cotton soaked with clinical alcohol once a
month
Do not, under any circumstance, immerse the equipment and ECG cable into cleaning liquids or
detergents. Do not insert any kind of liquids into the equipment and ECG cable.
WARNING
57 / 61
Ver. 2.41
CardioCare2000 Operation Manual
2) Regular Inspection
In common with all medical equipment, it is recommended that a safety check of EKG2000
should be carried out once a year. Refer to Service manual offered by Bionet Co., Ltd. on the
extent of the check.
3) Troubleshooting
1. In case that buzzer alarms for 1 second during printing and following message appears on
the LCD ;
* RECORD *
Paper empty!
This indicates that printer run out of printing paper. Operate again after replacing printing
paper.
2. In case that nothing can be printed out or printout fades ; Top cover of the printer is not
closed completely. Operate again after closing firmly
3. In case that buzzer alarms three times in series when system is on battery power, following
message appears on the LCD.:
* Low Battery *
Power off
This indicates that the battery is almost discharged. Turn the system power off and connect
the ac power source. Then turn the system power on again.
.
4. In case that ECGs is interfered with power noise;
First, move the menu to FLT AC and check if AC is set to 60Hz..
If ECGs is still interfered even if ac filter is set to 60Hz, connect the potential equalization
connector with the common ground conductor. Do not connect the potential equalization
conductor with the ac power system ground. The metallic frame of the patient’s bed or another
metallic object connected with building can be used for grounding.
58 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Chapter 5. Specification
TECHNICAL DATA
- Input circuit : Isolated and defibrillation protected
- ECG leads : Standard 12 leads
- Sensitivity selection : 5,10,20 mm/mV 5 %
- Calibration voltage : 1mV 2 %
- Electrode offset tolerance : 250mV
- Resolution : 2 V, 500 Sample/Sec
- Frequency response: 0.05 ~ 150Hz
- Common mode rejection : > 100 dB
- Input impedance : 10 M
- Patient leakage : < 50 A
- Signal quality control : Disconnected lead detection
- Communication : PC connection with RS-232 interface and LAN
- Display : 2x16 char LCD Display
- Registration Resolution :
Vertical : 8 points/mm; Horizontal: 25 m at 25 mm/S
Paper type : Thermal, roll paper
Paper width : A4: 210 mm or 8.5” ; Length : A4 : 300 mm or 11”
- Keyboard : Membrane keyboard
- Electrical :
Power supply : AC or built-in Battery(option)
Voltage rating : 100 - 240Vac (50/60 Hz)
- Environment :
· Operating Temperature:15°C to 30°C (59°F to 86°F)
· Storage Temperature:- 10°C to 60°C (14°F to 140°F)
· Operating/Storage Humidity:20% to 95% RH, non-condensing
· Operating Altitude:70(700) to 106Kpa(1060mbar)
- RS232C Interface
Protocol: asynchronous
Baud rate: 19200
Byte format: 8 data bit, 1 stop bit, no parity bit
Connection socket: 9 pole female, wired as DTE(Data Terminal Equipment)
Pin connection: 3=TXD(out), 2=RXD(in), 6=DSR, 4=DTR, 5=GND
59 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Warranty
Approval No.
Approval Date
Serial No.
Date of purchase
Customer Hospital :
Address :
Name :
Tel :
Sales Agency
Manufacturer
60 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Bionet Co.,Ltd.
Product Name: CardioCare-2000
61 / 61
Ver. 2.41