MANUAL

TR-200
Surface roughness tester
Changes in products and/or product
specifications can emerge due to new
technologies and continuous development.
We reserve the right to change or modify
specifications of products without prior
notice.
We recommend you to contact our sales
office for up-to-date information.
© All rights reserved
CONTENTS

1 OVERVIEW 4
1.1 Measurement principles 4
1.2 Standard configuration 5
1.3 Names of each part of the instrument 5
1.4 Basic connection method 6
1.4.1 Installation and removing pickup 6
1.4.2 Power adapter and charging of battery 7

2 MEASURING OPERATION 8
2.1 Preparation for measurement 8
2.2 Basic measurement status 9
2.3 Modifying conditions of measurement 12
2.3.1 Sampling length 12
2.3.2 Evaluation length 12
2.3.3 Standard 13
2.3.4 Range 13
2.3.5 Filter 14
2.3.6 Parameter 14
2.4 System setup 14
2.4.1 Language 15
2.4.2 Unit 15
2.4.3 Display 15
2.5 Function selection 16
2.5.1 Printing 17
2.5.2 Primary profile 18
2.5.3 Stylus position 18
2.5.4 Calibration 18
2.6 Communication with pc 19

3 OPTIONS AND USAGE 20


3.1 Adjustable supporter and sheath of pickup 20
3.2 Measurement stand 21
3.3 Extension rod 21
3.4 Connection rod of magnetic gauge base 22
3.5 Curvd surface pickup 22

Manual TR-200 1
4 TECHNICAL PARAMETER AND FEATURES 23

5 GENERAL MAINTENANCE 25
5.1 Troubleshooting 25
5.2 Fault information 26

6 BATTERY SWITCH 27

7 REFERENCES 28
7.1 Profile and filter 28
7.1.1 Profile 28
7.1.2 Filter 28
7.2 Central line 28
7.3 Traversing length 29
7.3.1 RC filter 29
7.3.2 PC-RC filter 29
7.3.3 Gauss filter 29
7.3.4 D-P Direct profile 29
7.4 Definition of TR-200 roughness parameters 30
7.4.1 Ra arithmetical mean deviation of profile 30
7.4.2 Rq root-mean-square deviation of profile 30
7.4.3 Rz ten point height of irregularities 30
7.4.4 Ry (ISO) maximum height of profile 31
7.4.5 Ry (DIN) maximum height of profile 31
7.4.6 Rt total peak-to-valley height 31
7.4.7 Rp maximum depth of profile peak 31
7.4.8 Rm maximum depth of profile valley 31
7.4.9 Sm mean spacing of profile irregularities 31
7.4.10 S mean spacing of local peaks of profile 32
7.4.11 tp profile bearing length ratio 32
7.4.12 Sk skewness of the profile 32
7.4.13 R3z third maximum peak-to-valley height 32

2 Manual TR-200
Start key

Parameter key

Esc key

Menu key

Scroll key

“Enter” key

Power key

Manual TR-200 3
1. OVERVIEW

TR-200 hand-held roughness tester is a new product . This tester applies to


production site and can be used to measure surface roughness of various
machinery-processed parts, calculate parameters according to selected
measuring conditions and clearly display all measurement parameters and
profile graphs on LCD.

Feature:
- Multi-parameter measuring: Ra, Rz, Ry, Rq, Rp, Rm, Rt, R3z, Rmax,
Sk, S, Sm, tp;
- High accuracy inductance pickup;
- Four filtering methods of RC, PC-RC, GAUSS and D-P;
- Compatible with four standards of ISO, DIN, ANSI and JIS;
- 128×64 dot matrix LCD displays all parameters and graphs;
- DSP chip is used to control and process data with high speed and low
power consumption;
- Built-in lithium ion chargeable battery and control circuit have high
capacity, without memory effect. Consecutive work time is longer than
20 hours;
- Design of mechanical and electrical integration is adopted to achieve
small bulk, light weight and easy usage;
- Can be connected to TA-220s printer to print all parameters and
graphs;
- Built-in standard RS232 interface enables communication with PC;
- Automatic switch off, memory and various prompt instructions;
- Accessories of curved surface pickup, measurement stand, sheath of
pickup, adjustable supporter and extending rod are available.

1.1 MEASUREMENT PRINCIPLE


When measuring roughness of part surface, the pickup is placed on the surface
of the part and then tracing the surface at constant rate. The pickup acquires
the surface roughness by the sharp stylus in pickup. The roughness causes
displacement of pickup which results in change of inductive value of induction
coils thus generate analogue signal which is in proportion to surface
roughness at output end of phase-sensitive rectifier. This signal enters data
collection system after amplification and level conversion. After that, those
collected data are processed with digital filtering and parameter calculation
by DSP chip and the measuring result can be read on LCD, printed through
printer and communicated with PC.

4 Manual TR-200
1.2 STANDARD CONFIGURATION
Table 1-1 Standard Configuration List
- Standard sensor
- Main processor
- Standard sample plate
- Power adapter
- RS232 communication cable
- Sheath of pickup
- Adjustable supporter

1.3 NAMES OF EACH PART OF THE INSTRUMENT

Socket

Skid Stylus Protection sleeve Main body

Figure 1-1-1 Pick up

Start key Display key Exit key Menu key Scroll key

Display “Enter” key

Sensor Power switch

Standard sample
plate

Figure 1-1-2 Front view of instrument ew of instrument

Manual TR-200 5
RS232 interface

Installation Hole
of accessory

Power socket

Figure 1-1-3 Side view of instrument

1.4 BASIC CONNECTION METHOD


1.4.1 Installation and removing of pickup

For installation, hold the main part of pickup with hand, push it into connection
sheath at the bottom of the instrument as shown in Figure 1-2 and then slightly
pushed it to the end of the sheath. To remove, hold the main part of pickup or
the root of protective sheath with hand and slowly pull it out.

Connection sheath Sensor

Figure 1-2 Installation and removing of pickup


Tip:
1. The stylus of pickup is key part of this tester and great
attention should be paid to it.
2. During installation and unloading, the stylus should not
be touched in order to avoid damage and affecting
measurement.
3. Connection of pickup should be reliable during
installation.

6 Manual TR-200
1.4.2 Power adapter and charging of battery

When battery voltage is too low (that is, battery voltage symbol flashs on screen
to prompt low voltage), the instrument should be charged as soon as possible.
As shown in Figure 1-3, the plug of power adapter should be plugged into
power socket of the instrument. Power adapter could be connected to power of
220V 50Hz .then charging will begin. Input voltage for power adapter is AC
220V with DC 6V of output, about 500mA of maximum charge current,
charging time of 2.5 hours. This instrument adopts lithium ion chargeable
battery without memory effect and charging can be fulfilled at any time
without affecting normal operation of the instrument.

Power adapter

Power socket Power top guide

Figure 1-3 Connection of power adapter


Tip:
1. Layout of connection lines shall not affect measuring part
under charging state.
2. Meanings of battery voltage prompts are:
indicates normal voltage and measurement can be
carried out; the black part inside prompt shows
capacity of battery;
indicates too-low voltage and battery should be
charged as soon as possible;
indicates that battery is being charged;
indicates that battery is full and charging power
should be cut off as soon as possible;
3. Relative high noises of power source may affect
measurement of weak signal to some extent when
battery is being charged;
4. The instrument needs to monitor the process of charging
so it is not necessary to turn it off. The instrument will
turn on automatically even it's turned off.
5. Keep the battery switch on unless the TR-200 will not be used
for a long period of time (more than 2 to 3 weeks). If the
battery switch is off, the measurement results will be lost.
6. When the TR-200 is delivered the battery switch is off,
User should set the switch on firstly before use it.

Manual TR-200 7
2 MEASURING OPERATION

2.1 PREPARATION FOR MEASUREMENT


a. Switch-on to check if battery voltage is normal;
b. Clear the surface of part to be measured;
c. Refer to Figure 2-1 and Figure 2-2 to place the instrument correctly,
stably and reliably on the surface to be measured;

Figure 2-1 Front view

Figure 2-2 Side view


d. Refer to Figure 2-3,trace of the pickup must be vertical to the
direction of process line of the measured surface.

Figure 2-3 Measuring direction


Instruction:
Correct and standard operation is the premise for accurate
measurement result, please make sure to follow it.

8 Manual TR-200
2.2 BASIC MEASUREMENT STATUS
Press and release Power key to switch on. The instrument automatically displays
model, name and information of manufacturer, and then enters basic
measurement status, as shown in Figure 2-4.

On/off switch

Portable roughness
tester

Standard Battery indicator


measurement Parameter value indicator
conditions Measurement conditions
indicator

Figure 2-4 Switch-on process


Instruction:
Contents of basic measurement status entered in the first
switch-on are default settings of this instrument. Settings and
data of last switch-off will be displayed in the next switch-on.
Basic measurement status will be entered automatically for each
switch-on (as shown in Figure 2-4).

Please don’t keep pressing for a long time to switch on the instrument.

Manual TR-200 9
In basic measurement status, the follows can be perform:
- Measurement
Press Start key to start measurement,as shown in Figure 2-5.

Press to start measurement

Press to stop
Pickup trace on surface

Filtering

Sampling ends, filtering is in


process
Calculating
parameter
Filtering ends, calculating parameter

Measurement ends, return to basic


measurement status, display
measurement result

Figure 2-5 Measurement process


- Enter menu operation status
Press Menu key to enter menu operation status. For detailed
operation, see descriptions in corresponding chapters and sections
later.

- Display measurement parameters


Press Parameter key first to shows all parameter values of this
measurement.
Press Scroll key to scroll pages; secondly, press Parameter key to
displays profile graphs of this measurement. Press Scroll key to roll
profile graphs with other sampling lengths; thirdly, press Parameter
key to displays tp curve and tp value of this measurement; Press the
keys again will repeat above descriptions. Press Esc key in each
status to return basic measurement status (as shown in figure below).

10 Manual TR-200
Figure 2-6 Display of parameters
- Display position of stylus
Press Enter key to show the position of stylus in shortcut mode, which
is easy to use in practical measurement.

Figure 2-7 Position of stylus


Instruction:
1. Tester automatically stores results and conditions of last
measurement before it is turned off, and will
automatically enter this status while it is turned on again.
2 After entering basic measurement status, press Start key
to measure, if measuring conditions are not needed to
be changed.
3. If position of stylus is close to the limit of range or
beyond it, adjust the position of pickup slightly. Note to
keep in line with instructions in 2.1 Preparation for
Measurement (adjustment is not needed generally).

Manual TR-200 11
2.3 MODIFYING CONDITIONS OF MEASUREMENT
Under basic measurement status, press Menu key to enter menu operation
status.
Press Scroll key to select setting function of measured conditions, and then press
Enter key to enter Set Measurement Conditions status. In this status, all
measurement conditions can be modified (as shown in Figure 2-8).

Figure 2-8 Select set measurement conditions status

2.3.1 Sampling length

After entered setting status, press Scroll key to select Set Sampling Length.
Press Enter key to cycle with 0.8 mm . 2.5mm . auto . 0.25mm (as shown in
Figure 2-8). Stop at the value you want and press Scroll key to modify others.

2.3.2 Evaluation length press menu key

to enter menu operation status and press Scroll key to select Set Measurement
Conditions. Press Enter key to enter Set Measurement and press Scroll key to
select Set n*cutoff. Press Enter to cycle with 1.2.3.4.5 (as shown in
Figure 2-9) Stop at the value you want and press Scroll key to modify others.

Figure 2-9 Modifying Evaluation Length


Instruction:
When sampling length is set to be automatic, evaluation length
will automatically display value 5 to match. This value can not be
modified.

12 Manual TR-200
2.3.3 Standard press menu key

to enter menu operation status and press Scroll key to select Set Measurement
Condition. Press Enter key for measurement setting and press Scroll key to
select Set Standard. Press Enter key to cycle with ISO. DIN. JIS . ANSI.

Figure 2-10 Modifying standard

Table 2 standard code and name


Code Standard name
ISO 4287 International standard
DIN 4768 German standard
JIS B601 Japanese industrial standard
ANSI B46.1 American standard

2.3.4 Range press menu key

to enter menu operation status and press Scroll key to select Set Measurement
Conditions. Press Enter key for measurement setting and press Scroll key to
select Set Range. Press Enter key to cycle with ±20µm.±40µm .±80µm .auto.

Figure 2-11 Modifying range

Manual TR-200 13
2.3.5 Filter

Press menu key to enter menu operation status and press Scroll key to select Set
Measurement Conditions. Press Enter key for measurement setting and press
Scroll key to select Set Filter. Press Enter key to cycle with RC.PC-RC.Gauss.D-P.

Figure 2-12 Modifying filter

2.3.6 Parameter

Press menu key to enter menu operation status and press Scroll key to select Set
Measurement Conditions. Press Enter key for measurement setting and press
Scroll key to select Parameter. Press Enter key to cycle with Ra..Rz.Ry.Rq (of
which: five parameters of Ra.Rz.Ry.Rmax .Rq are available for ANSI[American
Standard] and DIN[German Standard]). After confirmation, parameters
selected will be displayed in basic measurement status.

Figure 2-13 Modifying parameters to display

2.4 SYSTEM SETUP


Press Menu key to enter menu operation status and press Scroll key to select
System Setup. Press Enter key for system setting. In System Setup (as shown in
Figure 2-14), modify contents of system settings.

Figure 2-14 Select system setup function

14 Manual TR-200
2.4.1 Language

Press menu key to enter menu operation status and press Scroll key to select
System Setup. Press Enter key for system setting and press Scroll key to select
Language. Press Enter key for language selection and press to select the
language you want. Press Enter key to confirm.

Figure 2-15 Language selection function

2.4.2 Unit

Press menu key to enter menu operation status and press Scroll key to select
System Setup. Press Enter key for system setting and press Scroll key to choose
Unit. Press Enter key to switch between metric system and British system.

Figure 2-16 Metric/British system switching function

2.4.3 Display

a. LCD Back-light Press Menu key to enter menu operation status and
press Scroll key to select System Setup. Press Enter key for system
setting and press Scroll key to select LCD Back-light On/Off.
Press Enter key to switch between On and Off.

Figure 2-17 Back-light On/Off Function

Manual TR-200 15
Tip:
press after turn on the TR-200 for back-light.

b. LCD Brightness Press Menu key to enter menu operation status and
press Scroll key to select System Setup. Press Enter key for system
setting and press Scroll key for LCD Brightness Adjustment.
Press Enter key to enter LCD Brightness Adjustment and press Scroll
key to adjust the brightness to satisfactory degree.

Figure 2-18 LCD Brightness Adjustment Function

2.5 FUNCTION SELECTION


Press Menu key to enter menu operation status and press Scroll key to select
Function Selection. Press Enter key to enter Function Selection status.
After that, select appropriate functions you want.

Figure 2-19 Choose function selection

16 Manual TR-200
2.5.1 Printing

Before printing, connect the tester to the printer with the communication cable
as shown in Figure 2-20, and set the printer to On-line status.

Figure 2-20 Connect printer instruction:


This tester is used with TA-series printers only. TA-210 only prints values of
parameters. TA-220s prints values of parameters, profile graph and tp curve.
a. Print Parameters Press Menu key to enter menu operation status and
press Scroll key to choose Function Selection. Press Enter key to enter
Function Selection status and press Scroll key to select Print
Parameter. Press Enter key to print all measurement parameters
(as shown in Figure 2-19).

Instruction:
Under basic measurement status, press key to print all parameter
values.

b. Print Parameter and Profile Press Menu key to enter menu operation
status and press Scroll key to choose Function Selection.
Press Enter key to enter Function Selection status and press Scroll key
to select Print Parameter and Profile. Press Enter key to start printing.
The contents to print include all measurement parameters, profile
figures after filtering and tp figure.

Figure 2-21 Choose print parameter and profile function

Manual TR-200 17
2.5.2 Primary profile

Press Menu key to enter menu operation status and press Scroll key to choose
Function Selection. Press Enter key to enter Function Selection status and press
Scroll key to select Primary Profile. Press Enter key to display Primary Profile
(i.e. Direct Profile or Original Profile) in this measurement on LCD.

Figure 2-22 Select primary profile function

2.5.3 Stylus position press menu key

to enter menu operation status and press Scroll key to choose Function Selection.
Press Enter key to enter Function Selection status and press Scroll key to select
Stylus Position. Press Enter key to display the Stylus position.

Figure 2-23 Select display stylus position function

2.5.4 Calibration

Press menu key to enter menu operation status and press Scroll key to choose
Function Selection. Press Enter key to enter Function Selection status and press
Scroll key to select Display Value Calibration function. Press Enter key to enter
Calibration status and press Scroll key to change calibration coefficients.
Press Enter key to move cursor.

Figure 2-24 Select display value calibration function

18 Manual TR-200
Instruction:
1. While using correct measuring method to test random
sampling plate, if measured value exceeds ±10% of the
value of sampling plate, use Display Value Calibration
function to calibrate according to the percentage of real
deviation with calibration range within ±20%.
2. Generally, instrument has been strictly tested before
delivery so as to ensure display value error to be much
less than ±10%. In this case, user is suggested not to
frequently use Display Value Calibration function.

2.6 COMMUNICATION WITH PC


Before communicating with PC, connect the instrument to serial interface of PC
with communication cable attached to the instrument as shown in Figure 2-25,
and enter private operation software Data View on PC.

Figure 2-25 Connect PC instruction: to establish the communication between this


instrument and PC, use Data View private software. Please refer to the
Software Instruction for operation.

Manual TR-200 19
3. OPTIONS AND USAGE

3.1 ADJUSTABLE SUPPORTER AND SHEATH OF PICKUP


When measured surface of part is smaller than the bottom surface of the
instrument, sheath of pickup and adjustable supporter of TR-200 options can be
used for auxiliary support to complete the measurement (as shown in
Figure 3-1 and 3-2).

TR-200

Protection sleeve Adjustable leg

Figure 3-1 Connections of adjustable supporter and sheath of pickup


Sensor protection TR-200 Adjustable leg
sleeve

Item to be measured

Working table

Figure 3-2 Usage of adjustable supporter and sheath of pickup


Tip:
1. L above shall not be shorter than driving stroke of this
measurement to prevent pickupfrom dropping out of
part during measurement.
2. Locking of adjustable supporter shall be reliable.

20 Manual TR-200
3.2 MEASUREMENT STAND
TA series measurement Stand can adjust the positions between tester and
measured part conveniently with flexible and stable operation and wider
application range. Roughness of complex shapes can also be measured.
TA series measurement stand enable the adjustment of the position of stylus to
be more precise and measurement to be more stable. If Ra value of measured
surface is relatively low, Using measurement platform is recommended.

Measurement plat-
form

Figure 3-3 Measurement platform

3.3 EXTENDING ROD


Extending rod increases the depth for pickup to enter the part. Length of
extending rod is 50mm.

Sensor Extension rod TR-200

Figure 3-4 Extending Rod

Manual TR-200 21
3.4 CONNECTION ROD OF MAGNETIC GAUGE BASE
Connection rod connects the tester with magnetic gauge base so as to measure
various surfaces of parts flexibly and easily as shown in figure 3-5. It's
particularly suitable for production sites.
Magnetic gauge Connection rod
base

Platform connect-
ing
plate part

Figure 3-5 Connect the tester with magnetic gauge base


3.5 CURVED SURFACE PICKUP
Curved surface pickup can measure convex or concave surfaces of parts, as shown
in the figure below.

Side transformation
rod TR-200

Sensor

Curved surface sensor

Topguide
Item to be measured Stylus
Figure 3-7 Curved Surface Pickup

22 Manual TR-200
4 TECHNICAL PARAMETER AND FEATURES

- Pickup
Test principle Inductance type
Measurement range 160µm
Stylus tip radius 5µm
Stylus tip material Diamond
Measuring force 4mN (0.4gf)
- Stylus tip angle 90°
- Radius of skid
curvature 45mm
- Maximum drive range: 17.5mm
- Traversing speed: measuring sampling length = 0.25mm
Vt=0.135mm/s
sampling length = 0.8mm Vt=0.5mm/s
sampling length = 2.5mm Vt=1mm/s
returning V=1mm/s
- Accuracy Less than or equal to ±10%
- Repeatability of
displayed value Less than or equal to 6%
- Display content
- Menu modify measurement conditions, calibration
display value and select communication
with PC or printing.
- Parameter parameters of roughness compatible with
four standards of ISO, DIN, ANSI and JIS.
- Graph Primary profile, filtered profile and tp curve.
- Prompt information measurement, menu prompt, errors, battery
capacity and switch-off prompt information.
- Profile and filter
Filtered Profile RC
Filtered Profile PC-RC
Filtered Profile Gauss
Non-Filtered Profile D-P
- Sampling length /cut-off
length Automatic, 0.25mm,0.8mm,2.5mm
- Evaluation length (1~ 5) *sampling length

Manual TR-200 23
- Roughness parameter and display range
Parameter Display range
Ra 0.005µm ~ 16µm
Rq 0.005µm ~ 16µm
Rz
R3z
Ry
Rt
Rp
Rm 0.02µm ~ 160µm
Sk 0 ~ 100%
S
Sm 1mm
tp 0 ~ 100%
Measuring rang and resolution
Automatic 0.01µm ~0.04µm
±20µm 0.01µm
±40µm 0.02µm
±80µm 0.04µm
- Power supply AC adapter Rating 6V, 500mA
Supply voltage: 220V
Built-in battery(Lithium ion battery ):
Charging time: 2.5 hours
Consecutive work time per charge: more
than 20 hours
Auto-sleep function: display turns off if not
operated for 5 minutes
- Temperature Working environment 0°C~ 40°C
- Humidity: < 90% RH
- Store and transportation: Temperature: - 40°C ~ 60°C
Humidity: < 90% RH
- Dimensions L×W×H: 141×56×48mm
- Weight Mass: approximately 430g
- Connect with PC Standard RS-232 serial interface
- Connect to printer Connect with TA series printers only.
TA-210 printer only prints parameters.
TA-220s can print parameters as well as
profile figures.

24 Manual TR-200
5. GENERAL MAINTENANCE

- Avoid crash, intensive vibration, heavy dust, humidity, grease stain


and strong magnetic field;
- Pickup is a precise part and should be protected carefully.
Put it back in the package box after operation of each time;
- Protect standard sample plate provided with tester carefully so as to
avoid calibration fault caused by scratching.

5.1 TROUBLESHOOTING
When the tester breaks down, handle the troubles according to measures
described in next section Fault Information. If troubles still exist, please return
the instrument to factory for repair. Users should not dismantle and repair the
device by themselves. Returned instrument should be accompanied with
warranty card and sample plate attached. Phenomenon of problem should be
explained.

Manual TR-200 25
5.2 FAULT INFORMATION
Table 6
Display content Cause Solutions
Out of Range Maximum value of 1. Press Esc key to return;
measured signal 2. Enter menu setting status,
exceeds measuring increase measuring range,
range; press Esc key to return;
3. Measure again.
No data Wrong operation results 1. Press Esc key to return;
in failed measuring; 2. Check if preparation for
measuring is correct;
3. Switch on and measure
again.
A/D failure Hardware circuit fault; Scheme 1: Switch off and
switch on again;
Scheme 2: Press Reset key;
Scheme 3: Return to factory
to repair.
Motor failure Mechanical fault; Scheme 1: Switch off and
switch on again;
Scheme 2: Press Reset key;
Scheme 3: Return to factory
to repair.
Pickup failure Pickup is in automatic 1. Press Esc key to return and
return process wait till pickup returns to
the start position;
2.Measure again.
Working abnormally 1. Turnoff and then restart
2. switch the battery off, then
switch the battery on 10
seconds later.

26 Manual TR-200
6. BATTERY SWITCH

The battery switch is at the bottom of the instrument.

Battery switch

Figure 6-1 Battery Switch


When the TR-200 is delivered the battery switch is off, User should set the switch
on before use it. If TR-200 don’t start up after the battery switch is turned on,
press ON/OFF button. Keep the battery switch on unless the TR-200 will not be
used for a long period of time. If the battery switch is off, the measurement
results will be lost.

Manual TR-200 27
7. REFERENCES

7.1 PROFILE AND FILTER


7.1.1 Profile

a. Primary profile:
non-filtered profile signal obtained by pickup from measured surface.
b. Filtered profile:
profile signal after primary profile is filtered to remove waviness.

7.1.2 Filter

a. RC filter:
analogue 2RC filter with phase difference;
b. PC-RC filter:
RC filter with phase-correction;
c. Gauss filter:
DIN4777
d. D-P (direct-profile):
adopt central line of Least Square Algorithm.

7.2 CENTRAL LINE


TR-200 adopts central line of Least Square Algorithm.

7.3 TRAVERSING LENGTH


7.3.1 RC Filter

1 1

Traversing length

28 Manual TR-200
7.3.2 PC-RC filter

1 ½ ½

Traversing length

7.3.3 Gauss filter

1 ½ ½

Traversing length

7.3.4 D-P Direct profile

Traversing length

Manual TR-200 29
7.4 DEFINITIONS OF TR-200 ROUGHNESS PARAMETERS
This section gives definitions of TR-200 measurement parameters.

7.4.1 Arithmetical mean deviation of profile

Ra is arithmetic mean of the absolute values of profile deviation (Yi) from mean
within sampling length.

7.4.2 Rq root-mean-square deviation of profile

Rq is the square root of the arithmetic mean of the squares of profile deviation
(Yi) from mean within sampling length.

7.4.3 Rz ten point height of irregularities

The sum of the mean height of the five highest profile peaks and the mean depth
of the five deepest profile valley from mean within the sampling length.

30 Manual TR-200
7.4.4 Ry (ISO) maximum height of profile

Ry is The sum of height Rp of the highest profile peak from the mean line and
depth Rv of the deepest profile valley from the mean line within sampling length.

7.4.5 Ry (DIN) maximum height of profile

To get Ry (DIN) value: firstly calculate Ryi in each sampling length, the maximum
of which is the Ry (DIN) for the evaluation length.

7.4.6 Rt Total peak-to-valley height

Rt is the sum of the height of the highest peak and the depth of the deepest
valley over the evaluation length.

7.4.7 Rp maximum depth of profile peak

Rp is the height from the highest profile peak line to mean line within sampling
length.

7.4.8 Rm maximum depth of profile valley

Rm is the depth from the deepest profile valley line to mean line within sampling
length.

7.4.9 Sm mean spacing of profile elements

Sm is the mean spacing between profile peaks at the mean line within sampling
length.

Manual TR-200 31
7.4.10 S Mean spacing of local peaks of profile

S is the mean spacing of adjacent local peaks of the profile within sampling length.

7.4.11 tp Profile bearing length ratio

The bearing ratio tp is the length of bearing profile ,at a depth c below the
highest peak. Tp(%) is the ratio at the depth c.

7.4.12 Sk Skewness of the profile

Sk is the quotient of the mean cube value of the profile deviation (Yi) and the
cube of Rq within sampling length.

7.4.13 R3z third maximum peak-to-valley height


R3z is the mean of the sum of the third profile peak height and the third profile
valley depth of each sampling length over evaluation length.

32 Manual TR-200
EC-DECLARATION OF CONFORMITY

This certifies that the following designated product


TR-200 (Surface roughness tester)
complies with the essential protection requirements of
Council Directive 89/336/EWG approximation of the laws of the
Member States relating to electromagnetic companies.

This declaration applies to all specimens manufactured


in accordance with the manufacturing drawings
which form part of this declaration.

Assessment of compliance of the product with the requirements relating


to the compatibility was based on the following standards:
EN55022, EN60555-2, EN60555-3, EN50082-1

This declaration is the responsibility of the manufacturer/importer:

Company name: INNOVATEST Europe BV

Address: Borgharenweg 140

Telephone / Telefax: +31-43-3520060 / +31-43-3631168

Name of the subscriber: Nicole Paulissen-Schiffer

Position: General Manager

Maastricht
______________________ _____________________________________
City Authorised representative signature

Manual TR-200 33