This document provides instrument settings and parameters for analyzing albumin levels using a spectrophotometric bromocresol green method on a BS-300 instrument. It specifies reagent preparation, instrument settings like wavelengths, volumes, and incubation times. It also outlines calibration details including using a linear method and entering standard values. Maintaining validated parameters when software or reagents change is advised.

Uploaded by

Enrique Duarte
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
55 views

Albumin: Instrument

This document provides instrument settings and parameters for analyzing albumin levels using a spectrophotometric bromocresol green method on a BS-300 instrument. It specifies reagent preparation, instrument settings like wavelengths, volumes, and incubation times. It also outlines calibration details including using a linear method and entering standard values. Maintaining validated parameters when software or reagents change is advised.

Uploaded by

Enrique Duarte
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

ALBUMIN

Spectrophotometric
BROMOCRESOL GREEN

The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the
instrument. It is advisable to validate the use when there is any change in software or reagent versions.

Instrument: BS-300
Reagent preparation
Reagent is ready to use.

Instrument settings

PARAMETERS

Nº … Test ALB Method End-Point Direction Ascend


Unit g/L Decimal 0 Prim.Wavelength 630 Sec. Wavelength 700
Sample Volume 3 R1 Volume 300 R2 Volume 0 Line Limit N/A

Response
Incubation -- Check Antigen No Lower -- Sample Volume …
Reaction 0 -- 10 Substrate N/A Upper -- Dilution …

R1 Blank Mix R Blank Linearity Full Name …


Lower … Lower … Lower 1.4 Print Number …
Upper … Upper … Upper 70

Calibration
Method Linear Std. 0 0
Std. 1 (*)
Std. 2 --
Std. 3 --
Std. 4 -- (*) Enter the standard value
Std. 5 -- … Value entered by the user

Version 0707