0% found this document useful (0 votes)
267 views

CS4 Engl 122016

This document provides information about the CS4 free chlorine sensor from Xylem Analytics Germany GmbH. The CS4 uses an amperometric, potentiostatic 3-electrode system to measure free chlorine concentration with reduced dependence on pH. It has various measuring ranges from 0.01-2 ppm up to 0.01-20 ppm, with analog output. The sensor requires a voltage supply between 12-30 VDC and has a response time of approximately 2 minutes. It is designed to measure free chlorine in applications like swimming pools, drinking water, and seawater.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
267 views

CS4 Engl 122016

This document provides information about the CS4 free chlorine sensor from Xylem Analytics Germany GmbH. The CS4 uses an amperometric, potentiostatic 3-electrode system to measure free chlorine concentration with reduced dependence on pH. It has various measuring ranges from 0.01-2 ppm up to 0.01-20 ppm, with analog output. The sensor requires a voltage supply between 12-30 VDC and has a response time of approximately 2 minutes. It is designed to measure free chlorine in applications like swimming pools, drinking water, and seawater.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Xylem Analytics Germany GmbH

Sensortechnik Meinsberg
Kurt-Schwabe-Str. 6, 04736 Waldheim, Germany
Tel +49 34327 6230, Fax +49 34327 623-79
[email protected] www.meinsberg.de 12/2016

CS4 – free chlorine sensor


indicator Free chlorine with reduced pH-dependence
e. g. swimming pool water, drinking water, sea water
Application Surfactants (tensides) are partially tolerated.
Membrane covered, amperometric potentiostatic 3-electrode
Measuring system
system with electronic inside
- current output analog
- not galvanically isolated electronics
Electronic - output signal: analog (analog-out/analog)
- only valid for sensors with 2-pole-terminal and
CE-mark: poti for (restricted) adjustment of measuring range
Slope of a sensor can vary production-related or application-
related between 65% and 150% of the nominal slope

-> Recommendation to determine the suitable measuring range


Information about the
or the suitable sensor:
measuring range of
Concentration to be measured x factor 1.5 = measuring range
sensors with 4-20 mA
of the sensor

Example: Concentration to be measured 1.6 ppm x 1.5 = 2.4


-> recommended sensor with a measuring range of 5 ppm
Accuracy - Measuring range 2 mg/l: at 0.4 mg/l <1%
after calibration at repeatability at 1.6 mg/l <1%
conditions (25°C, pH 7.2 in drinking - Measuring range 20 mg/l: at 4 mg/l <1%
water) of the upper full scale
at 16 mg/l <3%
Working temperature 0 – <45 °C (no ice crystals in the measuring water)
Temperature
Automatically, by an integrated temperature sensor
compensation
Max. allowed working
3 bar, no pressure impulses and/or vibrations
pressure
Flow rate approx. 15-30L/h, small flow rate dependence is given
pH-range pH 4 – pH 9, reduced dependence on pH-value
Run-in time First start-up approx. 2 h
Response time T90: approx. 2 min
Zero point adjustment Not necessary
Slope calibration At the device, by analytical determination, DPD-1-Method
Connection 2-pole terminal (2 x 1 mm²)
Microporous hydrophilic Membrane, PVC-U, stainless steel
material
1.4571
diameter: approx. 25 mm
Size
Length: approx. 220 mm (2-pol-terminal)
Probe: Frost-protected, dry and without electrolyte
no limit
at >5 - <40 °C
storage Membrane cap: Used membrane caps can not be stored!
Electrolyte: in original bottle protected from sunlight
min. 1 year
at >10 - <35 °C
Regularly control of the measuring signal, min. once a week
The following specifications depend on the water quality:
maintenance
Change of the membrane cap: once a year
Change of the electrolyte: once a year
Measuring range:

Measuring
resolution Nominal
range Output Voltage supply
in ppm slope
in ppm
CS4MA2 0.01…2.00 0,01 8.0 mA/ppm
CS4MA5 0.01…5.00 0,01 3.2 mA/ppm 12…30 VDC
CS4MA10 0.01…10.00 0,01 4…20 mA 1.6 mA/ppm RL 50Ω…900Ω
uncalibrated
CS4MA20 0.01…20.00 0,01 0.8 mA/ppm
CS4MA-200 0.03… 20.00 0,01 0.08 mA/ppm

www.wtw.com www.si-analytics.com www.ebro.com www.meinsberg.de


Xylem Analytics Germany GmbH Amtsgericht München HRB 145685 Bank: Deutsche Bank AG, BLZ 700 700 10, Kto.-Nr. 201585700
Dr.-Karl-Slevogt-Str. 1 USt-IdNr./ VAT No.: DE275429984 IBAN: DE16700700100201585700
82362 Weilheim, Germany Managing Directors: BIC: DEUTDEMMXXX
Dr. Robert Reining · Ulrich Schwab · Matthias Rohwedder Steuernummer: 119/116/90391

You might also like