Electronics PDF
Electronics PDF
Helmuth Spieler
Physics Division, Lawrence Berkeley National Laboratory∗
Berkeley, California 94720, U.S.A.
1 Introduction
Figure 1: Basic detector functions: Radiation is absorbed in the sensor and converted into
an electrical signal. This low-level signal is integrated in a preamplifier, fed to a pulse shaper,
and then digitized for subsequent storage and analysis.
2 Example Systems
PHOTOMULTIPLIER VTH
Figure 2: In a scintillation detector absorbed energy is converted into visible light. The
scintillation photons are commonly detected by a photomultiplier, which can provide sufficient
gain to directly drive a threshold discriminator.
TOKEN IN
ANALOG SIGNAL PROCESSING
SPARSIFICATION
ANALOG SIGNAL PROCESSING CONTROL
DIGITAL
ANALOG SIGNAL PROCESSING CONTROL DATA OUT
OUTPUT
ANALOG SIGNAL PROCESSING DRIVERS TOKEN OUT
ANALOG SIGNAL PROCESSING
Figure 3: Circuit blocks in a representative readout IC. The analog processing chain is
shown at the top. Control is passed from chip to chip by token passing.
very simple, as the PMT output is plugged directly into a charge-sensing ADC.
A detector array combines the sensor and the analog signal processing
circuitry together with a readout system. The electronic circuitry is often
monolithically integrated. Figure 3 shows the circuit blocks in a representa-
tive readout integrated circuit (IC). Individual sensor electrodes connect to
parallel channels of analog signal processing circuitry. Data are stored in an
analog pipeline pending a readout command. Variable write and read pointers
are used to allow simultaneous read and write. The signal in the time slot of
interest is digitized, compared with a digital threshold, and read out. Circuitry
is included to generate test pulses that are injected into the input to simulate
a detector signal. This is a very useful feature in setting up the system and
is also a key function in chip testing prior to assembly. Analog control levels
are set by digital-to-analog converters (DACs). Multiple ICs are connected to
a common control and data output bus, as shown in Figure 4. Each IC is as-
signed a unique address, which is used in issuing control commands for setup
and in situ testing. Sequential readout is controlled by token passing. IC1 is
the master, whose readout is initiated by a command (trigger) on the control
bus. When it has finished writing data it passes the token to IC2, which in
turn passes the token to IC3. When the last chip has completed its readout
Analog and Digital Electronics for Detectors 113
STRIP DETECTOR
TOKEN
PASSING
CONTROL BUS
DATA BUS
Figure 4: Multiple ICs are ganged to read out a silicon strip detector. The right-most chip
IC1 is the master. A command on the control bus initiates the readout. When IC1 has written
all of its data it passes the token to IC2. When IC2 has finished it passes the token to IC3,
which in turn returns the token to the master IC1.
the token is returned to the master IC, which is then ready for the next cycle.
The readout bit stream begins with a header, which uniquely identifies a new
frame. Data from individual ICs are labeled with a chip identifier and channel
identifiers. Many variations on this scheme are possible. As shown, the readout
is event oriented, i.e. all hits occurring within an externally set exposure time
(e.g. time slice in the analog buffer in Figure 3) are read out together. For a
concise discussion of data acquisition systems see ref. [3].
In colliding beam experiments only a small fraction of beam crossings
yields interesting events. The time required to assess whether an event is po-
tentially interesting is typically of order microseconds, so hits from multiple
beam crossings must be stored on-chip, identified by beam crossing or time-
stamp. Upon receipt of a trigger the interesting data are digitized and read
out. This allows use of a digitizer that is slower than the collision rate. It is
also possible to read out analog signals and digitize them externally. Then the
output stream is a sequence of digital headers and analog pulses. An alternative
scheme only records the presence of hit. The output of a threshold comparator
signifies the presence of a signal and is recorded in a digital pipeline that retains
the crossing number.
Figure 5 shows a closeup of ICs mounted on a hybrid using a flexible poly-
imide substrate[4]. The wire bonds connecting the IC to the hybrid are clearly
visible. Channels on the IC are laid out on a ∼ 50 µm pitch and pitch adapters
fan out to match the 80 µm pitch of the silicon strip detector. The space be-
114 Helmuth Spieler
tween chips accommodates bypass capacitors and connections for control busses
carrying signals from chip to chip.
The minimum detectable signal and the precision of the amplitude mea-
surement are limited by fluctuations. The signal formed in the sensor fluctu-
ates, even for a fixed energy absorption. In addition, electronic noise introduces
baseline fluctuations, which are superimposed on the signal and alter the peak
amplitude. Figure 6 (left) shows a typical noise waveform. Both the amplitude
and time distributions are random. When superimposed on a signal, the noise
alters both the amplitude and time dependence, as shown in Figure 6 (right).
As can be seen, the noise level determines the minimum signal whose presence
can be discerned.
In an optimized system, the time scale of the fluctuations is comparable
Figure 5: Closeup of ICs mounted on a hybrid utilizing a flexible polyimide substrate. The
high-density wire bonds at the upper edges connect via pitch adapters to the 80 µm pitch of
the silicon strip detector. The ground plane is patterned as a diamond grid to reduce material.
(Photograph courtesy of A. Ciocio.)
Analog and Digital Electronics for Detectors 115
TIME TIME
Figure 6: Waveforms of random noise (left) and signal + noise (right), where the peak signal
is equal to the rms noise level (S/N = 1). The noiseless signal is shown for comparison.
to that of the signal, so the peak amplitude fluctuates randomly above and
below the average value. This is illustrated in Figure 7, which shows the same
signal viewed at four different times. The fluctuations in peak amplitude are
obvious, but the effect of noise on timing measurements can also be seen. If
the timing signal is derived from a threshold discriminator, where the output
fires when the signal crosses a fixed threshold, amplitude fluctuations in the
TIME TIME
TIME TIME
Figure 7: Signal plus noise at four different times, shown for a signal-to-noise ratio of about
20. The noiseless signal is superimposed for comparison.
116 Helmuth Spieler
leading edge translate into time shifts. If one derives the time of arrival from
a centroid analysis, the timing signal also shifts (compare the top and bottom
right figures). From this one sees that signal-to-noise ratio is important for all
measurements – sensing the presence of a signal or the measurement of energy,
timing, or position.
The sensor signal is usually a short current pulse is (t). Typical durations
vary widely, from 100 ps for thin Si sensors to tens of µs for inorganic scintilla-
tors. However, the physical quantity of interest is the deposited energy, so one
has to integrate over the current pulse
�
E ∝ Qs = is (t)dt . (1)
This integration can be performed at any stage of a linear system, so one can
3. amplify the current pulse and use an integrating ADC (“charge sensing”
ADC),
4. rapidly sample and digitize the current pulse and integrate numerically.
DETECTOR AMPLIFIER
tc
VELOCITY OF v
CHARGE CARRIERS
t
Cd Vi Ci Ri RATE OF INDUCED dqs
CHARGE ON SENSOR dt
ELECTRODES
t
Qs
SIGNAL CHARGE qs
Cf
Qi
DETECTOR !A
Cd vi vo
TEST
INPUT !V
CT
Q-AMP
Cd Ci
DYNAMIC INPUT
CAPACITANCE
Figure 10: Adding a test input to a charge-sensitive amplifier provides a simple means of
absolute charge calibration.
DETECTOR AMPLIFIER
ii
is Cd vi Ri
The preceding discussion assumed that the amplifiers are infinitely fast,
so they respond instantaneously to the applied signal. In reality this is not the
case; charge-sensitive amplifiers often respond much more slowly than the time
duration of the current pulse from the sensor. However, as shown in Figure 11,
this does not obviate the basic principle. Initially, signal charge is integrated on
the sensor capacitance, as indicated by the left hand current loop. Subsequently,
as the amplifier responds the signal charge is transferred to the amplifier.
Nevertheless, the time response of the amplifier does affect the measured
pulse shape. First, consider a simple amplifier as shown in Figure 12.
The gain element shown is a bipolar transistor, but it could also be a field
effect transistor (JFET or MOSFET) or even a vacuum tube. The transistor’s
V+
RL
io Co
vo
vi
Figure 12: A simple amplifier demonstrating the general features of any single-stage gain
stage, whether it uses a bipolar transistor (shown) or an FET.
120 Helmuth Spieler
output current changes as the input voltage is varied. Thus, the voltage gain
dvo dio
Av = = · ZL ≡ gm ZL . (6)
dvi dvi
The parameter gm is the transconductance, a key parameter that determines
gain, bandwidth, and noise of transistors. The load impedance ZL is the parallel
combination of the load resistance RL and the output capacitance Co . This
capacitance is unavoidable; every gain device has an output capacitance, the
following stage has an input capacitance, and in addition the connections and
additional components introduce stray capacitance. The load impedance is
given by
1 1
= + iωCo , (7)
ZL RL
where the imaginary i indicates the phase shift associated with the capacitance.
The voltage gain
� �−1
1
Av = gm + iωCo . (8)
RL
At low frequencies where the second term is negligible, the gain is constant
Av = gm RL . However, at high frequencies the second term dominates and
the gain falls off linearly with frequency with a 90◦ phase shift, as illustrated
in Figure 13. The cutoff (corner) frequency, where the asymptotic low and
high frequency responses intersect, is determined by the output time constant
τ = RL Co , so the cutoff frequency
1 1
fu = = . (9)
2πτ 2πRL Co
In the regime where the gain drops linearly with frequency the product of gain
and frequency is constant, so the amplifier can be characterized by its gain–
bandwidth product, which is equal to the frequency where the gain is one, the
unity gain frequency ω0 = gm /Co .
The frequency response translates into a time response. If a voltage step
is applied to the input of the amplifier, the output does not respond instanta-
neously, as the output capacitance must first charge up. This is shown in the
second panel of Figure 13.
In practice, amplifiers utilize multiple stages, all of which contribute to
the frequency response. However, for use as a feedback amplifier, only one
time constant should dominate, so the other stages must have much higher
cutoff frequencies. Then the overall amplifier response is as shown in Figure 13,
except that at high frequencies additional corner frequencies appear.
Analog and Digital Electronics for Detectors 121
V0
Av = 1
1 !0 log ! V = V0 (1 # exp(#t / $) )
RLCo
$ = RLCo
Figure 13: The time constants of an amplifier affect both the frequency and the time
response. Both are fully equivalent representations.
We can now use the frequency response to calculate the input impedance
and time response of a charge-sensitive amplifier. Applying the same reasoning
as above, the input impedance of an inverting amplifier as shown in Figure 9,
but with a generalized feedback impedance Zf , is
Zf Zf
Zi = ≈ (A � 1) . (10)
A+1 A
At low frequencies the gain is constant and has a constant 180◦ phase shift,
so the input impedance is of the same nature as the feedback impedance, but
reduced by 1/A. At high frequencies well beyond the amplifier’s cutoff frequency
fu , the gain drops linearly with frequency with an additional 90◦ phase shift,
so the gain
ω0
A = −i . (11)
ω
In a charge-sensitive amplifier the feedback impedance
1
Zf = −i , (12)
ωCf
i 1 1
Zi = − · ω = ≡ Ri . (13)
ωCf −i 0 ω0 Cf
ω
122 Helmuth Spieler
From this we see that the rise time of the charge-sensitive amplifier increases
with sensor capacitance. As noted above, the amplifier response can be slower
than the duration of the current pulse from the sensor, but it should be much
faster than the peaking time of the subsequent pulse shaper. The feedback ca-
pacitance should be much smaller than the sensor capacitance. If Cf = Cd /100,
the amplifier’s gain–bandwidth product must be 100/τi , so for a rise time con-
stant of 10 ns the gain–bandwidth product must be 1010 radians = 1.6 GHz. The
same result can be obtained using conventional operational amplifier feedback
theory.
The mechanism of reducing the input impedance through shunt feedback
leads to the concept of the “virtual ground”. If the gain is infinite, the input
impedance is zero. Although very high gains (of order 105 to 106 ) are achievable
in the kHz range, at the frequencies relevant for detector signals the gain is
much smaller. The input impedance of typical charge-sensitive amplifiers in
strip detector systems is of order kΩ. Fast amplifiers designed to optimize
power dissipation achieve input impedances of 100 to 500 Ω [5]. None of these
qualify as a “virtual ground”, so this concept should be applied with caution.
Apart from determining the signal rise time, the input impedance is critical
in position-sensitive detectors. Figure 14 illustrates a silicon-strip sensor read
out by a bank of amplifiers. Each strip electrode has a capacitance Cb to
the backplane and a fringing capacitance Css to the neighboring strips. If
the amplifier has an infinite input impedance, charge induced on one strip will
capacitively couple to the neighbors and the signal will be distributed over many
strips (determined by Css /Cb ). If, on the other hand, the input impedance of
the amplifier is low compared to the interstrip impedance, practically all of the
charge will flow into the amplifier, as current seeks the path of least impedance,
and the neighbors will show only a small signal.
Analog and Digital Electronics for Detectors 123
Figure 14: To preserve the position resolution of strip detectors the readout amplifiers
must have a low input impedance to prevent spreading of signal charge to the neighboring
electrodes.
5 Signal Processing
Figure 15: Signal and baseline fluctuations add in quadrature. For large signal variance
(top) as in scintillation detectors or proportional chambers, the baseline noise is usually neg-
ligible, whereas for small signal variance as in semiconductor detectors or liquid Ar ionization
chambers, baseline noise is critical.
case, low noise is critical. Baseline fluctuations can have many origins, external
interference, artifacts due to imperfect electronics, etc., but the fundamental
limit is electronic noise.
6 Electronic Noise
1
NORMALIZED COUNT RATE
Qn
0.78
FWHM= 2.35 Q n
0.5
Q s /Q n
Figure 16: Repetitive measurements of the signal charge yield a Gaussian distribution whose
standard deviation equals the rms noise level Qn . Often the width is expressed as the full
width at half maximum (FWHM), which is 2.35 times the standard deviation.
V2
P = = I 2R , (18)
R
the spectral voltage and current noise densities
Since the spectral noise components are non-correlated (each black body ex-
citation mode is independent), one must integrate over the noise power, i.e.
the voltage squared. The total noise increases with bandwidth. Since small
bandwidth corresponds to large rise-times, increasing the speed of a pulse mea-
surement system will increase the noise.
This is a very important result – the signal-to-noise ratio for a given signal
charge is inversely proportional to the total capacitance at the input node.
Note that zero input capacitance does not yield an infinite signal-to-noise ratio.
As shown in ref. [1], this relationship only holds when the input time constant
is greater than about ten times the sensor current pulse width. The dependence
of signal-to-noise ratio on capacitance is a general feature that is independent of
amplifier type. Since feedback cannot improve signal-to-noise ratio, eqn 22 holds
for charge-sensitive amplifiers, although in that configuration the charge signal
is constant, but the noise increases with total input capacitance (see[1]). In the
noise analysis the feedback capacitance adds to the total input capacitance (the
passive capacitance, not the dynamic input capacitance), so Cf should be kept
small.
8 Pulse Shaping
Pulse shaping has two conflicting objectives. The first is to limit the
bandwidth to match the measurement time. Too large a bandwidth will increase
the noise without increasing the signal. Typically, the pulse shaper transforms
a narrow sensor pulse into a broader pulse with a gradually rounded maximum
at the peaking time. This is illustrated in Figure 17. The signal amplitude is
measured at the peaking time TP .
The second objective is to constrain the pulse width so that successive
signal pulses can be measured without overlap (pileup), as illustrated in Figure
18. Reducing the pulse duration increases the allowable signal rate, but at the
expense of electronic noise.
In designing the shaper it is necessary to balance these conflicting goals.
Usually, many different considerations lead to a “non-textbook” compromise;
128 Helmuth Spieler
TP
Figure 17: In energy measurements a pulse processor typically transforms a short sensor
current pulse to a broader pulse with a peaking time TP .
AMPLITUDE
TIME TIME
Figure 18: Amplitude pileup occurs when two pulses overlap (left). Reducing the shaping
time allows the first pulse to return to the baseline before the second pulse arrives.
Analog and Digital Electronics for Detectors 129
e-t /!d
Figure 19: Components of a pulse shaping system. The signal current from the sensor is
integrated to form a step impulse with a long decay. A subsequent high-pass filter (“differ-
entiator”) limits the pulse width and the low-pass filter (“integrator”) increases the rise-time
to form a pulse with a smooth cusp.
1.0
SHAPER OUTPUT
n= 1
0.5 2
n= 8
0.0
0 1 2 3 4 5
TIME
Figure 20: Pulse shape vs. number of integrators in a CR-nRC shaper. The time constants
are scaled with the number of integrators to maintain the peaking time.
illustrated in Figure 20. The integration and differentiation time constants are
scaled to maintain the peaking time. Note that the peaking time is a key design
parameter, as it dominates the noise bandwidth and must also accommodate
the sensor response time.
Another type of shaper is the correlated double sampler, illustrated in
Figure 21. This type of shaper is widely used in monolithically integrated
circuits, as many CMOS processes (see Section 11.1) provide only capacitors
130 Helmuth Spieler
To determine how the pulse shaper affects the signal-to-noise ratio consider
the detector front-end in Figure 22. The detector is represented by the capac-
itance Cd , a relevant model for many radiation sensors. Sensor bias voltage is
applied through the resistor Rb . The bypass capacitor Cb shunts any external
interference coming through the bias supply line to ground. For high-frequency
signals this capacitor appears as a low impedance, so for sensor signals the “far
end” of the bias resistor is connected to ground. The coupling capacitor Cc
blocks the sensor bias voltage from the amplifier input, which is why a capaci-
SIGNAL
S1
S1
V1
V1 vn
Vo
S2
SIGNALS V2
S2
NOISE vs + vn
V2
!v= vs + vn vs
Vo
vn
Figure 22: A detector front-end circuit and its equivalent circuit for noise calculations.
tor serving this role is also called a “blocking capacitor”. The series resistor Rs
represents any resistance present in the connection from the sensor to the am-
plifier input. This includes the resistance of the sensor electrodes, the resistance
of the connecting wires or traces, any resistance used to protect the amplifier
against large voltage transients (“input protection”), and parasitic resistances
in the input transistor.
The following implicitly includes a constraint on the bias resistance, whose
role is often misunderstood. It is often thought that the signal current generated
in the sensor flows through Rb and the resulting voltage drop is measured. If
the time constant Rb Cd is small compared to the peaking time of the shaper TP ,
the sensor will have discharged through Rb and much of the signal will be lost.
Thus, we have the condition Rb Cd � TP , or Rb � TP /Cd . The bias resistor
must be sufficiently large to block the flow of signal charge, so that all of the
signal is available for the amplifier.
To analyze this circuit we’ll assume a voltage amplifier, so all noise con-
tributions will be calculated as a noise voltage appearing at the amplifier input.
Steps in the analysis are 1. determine the frequency distribution of all noise
voltages presented to the amplifier input from all individual noise sources, 2.
integrate over the frequency response of the shaper (for simplicity a CR-RC
shaper) and determine the total noise voltage at the shaper output, and 3. de-
termine the output signal for a known input signal charge. The equivalent noise
charge (ENC) is the signal charge for which S/N = 1.
The equivalent circuit for the noise analysis (second panel of Figure 22)
includes both current and voltage noise sources. The “shot noise” ind of the
sensor leakage current is represented by a current noise generator in parallel
with the sensor capacitance. As noted above, resistors can be modeled either
as a voltage or current generator. Generally, resistors shunting the input act as
noise current sources and resistors in series with the input act as noise voltage
132 Helmuth Spieler
sources (which is why some in the detector community refer to current and
voltage noise as “parallel” and “series” noise). Since the bias resistor effectively
shunts the input, as the capacitor Cb passes current fluctuations to ground, it
acts as a current generator inb and its noise current has the same effect as the
shot noise current from the detector. The shunt resistor can also be modeled
as a noise voltage source, yielding the result that it acts as a current source.
Choosing the appropriate model merely simplifies the calculation. Any other
shunt resistances can be incorporated in the same way. Conversely, the series
resistor Rs acts as a voltage generator. The electronic noise of the amplifier
is described fully by a combination of voltage and current sources at its input,
shown as ena and ina .
Thus, the noise sources are
where e is the electronic charge, Id the sensor bias current, k the Boltzmann
constant and T the √ temperature.√Typical amplifier √noise parameters ena and
ina are of order nV/ Hz and fA/ Hz (FETs) – pA/ Hz (bipolar transistors).
Amplifiers tend to exhibit a “white” noise spectrum at high frequencies (greater
than order kHz), but at low frequencies show excess noise components with the
spectral density
Af
e2nf = , (23)
f
where the noise coefficient Af is device specific and of order 10−10 – 10−12 V2 .
The noise voltage generators are in series and simply add in quadrature.
White noise distributions remain white. However, a portion of the noise currents
flows through the detector capacitance, resulting in a frequency-dependent noise
voltage in /(ωCd ), so the originally white spectrum of the sensor shot noise and
the bias resistor now acquires a 1/f dependence. The frequency distribution of
all noise sources is further altered by the combined frequency response of the
amplifier chain A(f ). Integrating over the cumulative noise spectrum at the
amplifier output and comparing to the output voltage for a known input signal
yields the signal-to-noise ratio. In this example the shaper is a simple CR-RC
shaper, where for a given differentiation time constant the noise is minimized
when the differentiation and integration time constants are equal τi = τd ≡ τ .
Then the output pulse assumes its maximum amplitude at the time TP = τ .
Analog and Digital Electronics for Detectors 133
Although the basic noise sources are currents or voltages, since radiation
detectors are typically used to measure charge, the system’s noise level is con-
veniently expressed as an equivalent noise charge Qn . As noted previously, this
is equal to the detector signal that yields a signal-to-noise ratio of one. The
equivalent noise charge is commonly expressed in Coulombs, the corresponding
number of electrons, or the equivalent deposited energy (eV). For the above
circuit the equivalent noise charge
� 2 � �� � �
e 4kT � � C2
Q2n = 2eId + + i2na · τ + 4kT Rs + e2na · d + 4Af Cd2 .
8 Rb τ
(24)
The prefactor e2 /8 = exp(2)/8 = 0.924 normalizes the noise to the signal gain.
The first term combines all noise current sources and increases with shaping
time. The second term combines all noise voltage sources and decreases with
shaping time, but increases with sensor capacitance. The third term is the
contribution of amplifier 1/f noise and, as a voltage source, also increases with
sensor capacitance. The 1/f term is independent of shaping time, since for
a 1/f spectrum the total noise depends on the ratio of upper to lower cutoff
frequency, which depends only on shaper topology, but not on the shaping time.
Just as filter response can be described either in the frequency or time
domain, so can the noise performance. Detailed explanations are given in papers
by Goulding and Radeka[7] [8] [9] [10]. The key is Parseval’s theorem, which
relates the amplitude response A(f ) to the time response F (t).
�∞ �∞
2 2
|A(f )| df = [F (t)] dt . (25)
0 −∞
The left hand side is essentially integration over the noise bandwidth. The
output noise power scales linearly with the duration of the pulse, so the noise
contribution of the shaper can be split into a factor that is determined by the
shape of the response and a time factor that sets the shaping time. This leads
to a general formulation of the equivalent noise charge
C2
Q2n = i2n Fi TS + e2n Fv + Fvf Af C 2 , (26)
TS
where Fi , Fv , and Fvf depend on the shape of the pulse determined by the
shaper and TS is a characteristic time, for example the peaking time of a
CR-nRC shaped pulse or the prefilter time constant in a correlated double
sampler[1]. As before, C is the total parallel capacitance at the input. The
134 Helmuth Spieler
�∞ �∞ � �2
1 2 TS dW (t)
Fi = [W (t)] dt , Fv = dt . (27)
2TS 2 dt
−∞ −∞
For time-invariant pulse shaping W (t) is simply the system’s impulse response
(the output signal seen on an oscilloscope) with the peak output signal normal-
ized to unity. For a time-variant shaper the same equations apply, but W (t) is
determined differently. See refs. [7], [8], [9], and [10] for more details.
A shaper formed by a single CR differentiator and RC integrator with
equal time constants has Fi = Fv = 0.9 and Fvf = 4, independent of the
shaping time constant, so for the circuit in Figure 19 eqn 24 becomes
� �
4kT � � C2
Q2n = 2qe Id + + i2na Fi TS + 4kT Rs + e2na Fv + Fvf Af C 2 . (28)
Rb TS
Pulse shapers can be designed to reduce the effect of current noise, e.g. mitigate
radiation damage. Increasing pulse symmetry tends to decrease Fi and increase
Fv , e.g. to Fi = 0.45 and Fv = 1.0 for a shaper with one CR differentiator and
four cascaded RC integrators.
Figure 23 shows how equivalent noise charge is affected by shaping time.
At short shaping times the voltage noise dominates, whereas at long shaping
times the current noise takes over. Minimum noise obtains where the current
and voltage contributions are equal. The noise minimum is flattened by the
presence of 1/f noise. Also shown is that increasing the detector capacitance
will increase the voltage noise contribution and shift the noise minimum to
longer shaping times, albeit with an increase in minimum noise.
For quick estimates one can use the following equation, which assumes an
FET amplifier (negligible ina ) and a simple CR-RC shaper with peaking time
τ . The noise is expressed in units of the electronic charge e and C is the total
parallel capacitance at the input, including Cd , all stray capacitances, and the
amplifier’s input capacitance.
� � � 2 � � �
e2 5 e kΩ τ e2 ns C2
Q2n = 12 Id τ + 6 · 10 + 3.6 · 10 4
e2n
nA · ns ns Rb 2 2
(pF) (nV) /Hz τ
(29)
The noise charge is improved by reducing the detector capacitance and
leakage current, judiciously selecting all resistances in the input circuit, and
Analog and Digital Electronics for Detectors 135
4
10
EQUIVALENT NOISE CHARGE (e)
3
TOTAL
10
TOTAL
1/f NOISE
CURRENT VOLTAGE
NOISE NOISE
102
0.01 0.1 1 10 100
SHAPING TIME (µs)
Figure 23: Equivalent noise charge vs. shaping time. At small shaping times (large band-
width) the equivalent noise charge is dominated by voltage noise, whereas at long shaping
times (large integration times) the current noise contributions dominate. The total noise as-
sumes a minimum where the current and voltage contributions are equal. The “1/f ” noise
contribution is independent of shaping time and flattens the noise minimum. Changing the
voltage or current noise contribution shifts the noise minimum. Increased voltage noise is
shown as an example.
choosing the optimum shaping time constant. The noise parameters of a well-
designed amplifier depend primarily on the input device. Fast, high-gain tran-
sistors are generally best.
In field effect transistors, both junction field effect transistors (JFETs) or
metal oxide semiconductor field effect transistors (MOSFETs), the noise cur-
rent contribution is very small, so reducing the detector leakage current and
increasing the bias resistance will allow long shaping times with correspond-
ingly lower noise. The equivalent input noise voltage e2n ≈ 4kT /gm , where gm
is the transconductance, which increases with operating current. For a given
current, the transconductance increases when the channel length is reduced,
so reductions in feature size with new process technologies are beneficial. At
a given channel length minimum noise obtains when a device is operated at
maximum transconductance. If lower noise is required, the width of the device
can be increased (equivalent to connecting multiple devices in parallel). This
increases the transconductance (and required current) with a corresponding de-
136 Helmuth Spieler
crease in noise voltage, but also increases the input capacitance. At some point
the reduction in noise voltage is outweighed by the increase in total input ca-
pacitance. The optimum obtains when the FET’s input capacitance equals the
external capacitance (sensor + stray capacitance). Note that this capacitive
matching criterion only applies when the input current noise contribution of
the amplifying device is negligible.
Capacitive matching comes at the expense of power dissipation. Since
the minimum is shallow, one can operate at significantly lower currents with
just a minor increase in noise. In large detector arrays power dissipation is
critical, so FETs are hardly ever operated at their minimum noise. Instead,
one seeks an acceptable compromise between noise and power dissipation (see
[1] for a detailed discussion). Similarly, the choice of input devices is frequently
driven by available fabrication processes. High-density integrated circuits tend
to include only MOSFETs, so this determines the input device, even where a
bipolar transistor would provide better performance.
In bipolar transistors the shot noise associated with the base current IB is
significant, i2nB = 2eIB . Since IB = IC /βDC , where IC is the collector current
and βDC the direct current gain, this contribution increases with device current.
On the other hand, the equivalent input noise voltage
2(kT )2
e2n = (30)
eIC
decreases with collector current, so the noise assumes a minimum at a specific
collector current
�
C � kT � Fv 1
Qn,min = 4kT √
2
Fi Fv at IC = C βDC . (31)
βDC e Fi TS
AMPLITUDE
AMPLITUDE
VT 2!n
VT dV = max
dt
TIME TIME
2!t
Figure 24: Fluctuations in signal amplitude crossing a threshold translate into timing fluc-
tuations (left). With realistic pulses the slope changes with amplitude, so minimum timing
jitter occurs with the trigger level at the maximum slope.
noise current contribution is significant. Due to the base current noise bipolar
transistors are best at short shaping times, where they also require lower power
than FETs for a given noise level.
When the input noise current is negligible, the noise increases linearly with
sensor capacitance. The noise slope
�
dQn Fv
≈ 2en · (33)
dCd T
depends both on the preamplifier (en ) and the shaper (Fv , T ). The zero inter-
cept can be used to determine the amplifier input capacitance plus any addi-
tional capacitance at the input node.
Practical noise levels range from < 1 e for CCDs at long shaping times
to ∼ 104 e in high-capacitance liquid Ar calorimeters. Silicon strip detectors
typically operate at ∼ 103 electrons, whereas pixel detectors with fast readout
provide noise of 100 – 200 electrons. Transistor noise is discussed in more detail
in[1].
10 Timing Measurements
σn tr
σt = ≈ , (34)
(dS/dt)ST S/N
where σn is the rms noise and the derivative of the signal dS/dt is evaluated
at the trigger level ST . To increase dS/dt without incurring excessive noise the
amplifier bandwidth should match the rise-time of the detector signal.The 10 –
90% rise time of an amplifier with bandwidth fu (see Figure 13) is
2.2 0.35
tr = 2.2τ = = . (35)
2πfu fu
For example, an oscilloscope with 350 MHz bandwidth has a 1 ns rise time.
When amplifiers are cascaded, which is invariably necessary, the individual rise
times add in quadrature
�
tr ≈ t2r1 + t2r2 + ... + t2rn . (36)
11 Digital Electronics
AMPLITUDE
VT
TIME
!T = “WALK”
Figure 25: The time at which a signal crosses a fixed threshold depends on the signal
amplitude, leading to “time walk”.
A A
AND EXCLUSIVE
B OR B
A A
B B
D FLIP-FLOP D
A (LATCH)
OR
B CLK
D Q
A
Q
B CLK
?
Figure 26: Basic logic functions include gates (AND, OR, Exclusive OR) and flip flops. The
outputs of the AND and D flip flop show how small shifts in relative timing between inputs
can determine the output state.
140 Helmuth Spieler
S Q
R Q
D Q
CLK Q
Figure 27: Some common logic symbols. Inverted outputs are denoted by small circles or
by a superimposed bar, as for the latch output Q. Additional inputs can be added to gates
as needed. An R-S flip-flop sets the Q output high in response to an S input. An R input
resets the Q output to low.
VDD
VDD
0 0 0
Figure 28: In an NMOS inverter the transistor conducts when the input is high (left),
whereas in a PMOS inverter the transistor conducts when the input is low (right). In both
circuits the input pulse is inverted, whether the input swings high or low.
load capacitance increases the rise time of the logic pulse and as a result delays
the time when the transition crosses the logic threshold. The energy dissipated
in the wiring resistance R is
�
E = i2 (t)R dt . (37)
When pulses occur at a frequency f , the power dissipated in both the positive
and negative transitions
P = f CV 2 . (40)
Thus, the power dissipation increases with clock frequency and the square of
the logic swing.
Fast logic is time-critical. It relies on logic operations from multiple paths
coming together at the right time. Valid results depend on maintaining mini-
mum allowable overlaps and set-up times as illustrated in Figure 26. Each logic
VDD
VDD
VDD
VDD VDD
VDD 0
0 0
0
VDD
V R i
0
VTH C VTH
T WIRING T+!t
RESISTANCE
SUM OF INPUT
CAPACITANCES
Figure 30: The wiring resistance together with the distributed load capacitance delays the
signal.
circuit has a finite propagation delay, which depends on circuit loading, i.e.
how many loads the circuit has to drive. In addition, as illustrated in Figure 30
the wiring resistance and capacitive loads introduce delay. This depends on the
number of circuits connected to a wire or trace, the length of the trace and the
dielectric constant of the substrate material. Relying on control of circuit and
wiring delays to maintain timing requires great care, as it depends on circuit
variations and temperature. In principle all of this can be simulated, but in
complex systems there are too many combinations to test every one. A more
robust solution is to use synchronous systems, where the timing of all tran-
sitions is determined by a master clock. Generally, this does not provide the
utmost speed and requires some additional circuitry, but increases reliability.
Nevertheless, clever designers frequently utilize asynchronous logic. Sometimes
it succeeds . . . and sometimes it doesn’t.
tions. This is illustrated in Figure 31. Field Programmable Gate or logic Arrays
(FPGAs) are a common example. A representative FPGA has 512 pads usable
for inputs and outputs, ∼ 106 gates, and ∼ 100K of memory. Modern design
tools also account for propagation delays, wiring lengths, loads, and tempera-
ture dependence. The design software also generates “test vectors” that can be
used to test finished parts. Properly implemented, complex digital designs can
succeed on the first pass, whether as ASICs or as logic or gate arrays.
For data storage and subsequent analysis the analog signal at the shaper
output must be digitized. Important parameters for analog-to-digital converters
(ADCs or A/Ds) used in detector systems are
Figure 31: Complex logic circuits are commonly implemented using logic arrays that as an
integrated block provide the desired outputs in response to specific input combinations.
Analog and Digital Electronics for Detectors 145
Vref COMPARATORS
INPUT ENCODER
R
R
DIGITIZED
OUTPUT
R
conversion bin. For example, an 8-bit converter requires 256 comparators. The
conversion is always monotonic and differential non-linearity is determined by
the matching of the resistors in the threshold divider. Only relative matching
is required, so this topology is a good match for monolithic integrated circuits.
Flash ADCs are available with conversion rates > 500 MS/s (megasamples per
second) at 8-bit resolution and a power dissipation of about 5 W.
The most commonly used technique is the successive approximation ADC,
shown in Figure 33. The input pulse is sent to a pulse stretcher, which follows
the signal until it reaches its cusp and then holds the peak value. The stretcher
output feeds a comparator, whose reference is provided by a digital-to-analog
converter (DAC). The DAC is cycled beginning with the most significant bits.
The corresponding bit is set when the comparator fires, i.e. the DAC output
becomes less than the pulse height. Then the DAC cycles through the less
significant bits, always setting the corresponding bit when the comparator fires.
Thus, n-bit resolution requires n steps and yields 2n bins. This technique makes
efficient use of circuitry and is fairly fast. High-resolution devices (16 – 20 bits)
with conversion times of order µs are readily available. Currently a 16-bit ADC
with a conversion time of 1 µs (1 MS/s) requires about 100 mW.
A common limitation is differential non-linearity, since the resistors that
set the DAC levels must be extremely accurate. For DNL < 1% the resistor
determining the 212 -level in a 13-bit ADC must be accurate to < 2.4 · 10−6 . As
a consequence, differential non-linearity in high-resolution successive approxi-
mation converters is typically 10 – 20% and often exceeds the 0.5 LSB (least
significant bit) required to ensure monotonic response.
PULSE CONTROL
COMPARATOR
STRETCHER LOGIC
ANALOG
INPUT DIGITIZED
OUTPUT
DAC
ADDRESS
DAC
Figure 33: Principle of a successive approximation ADC. The DAC is controlled to se-
quentially add levels proportional to 2n , 2n−1 , . . . 20 . The corresponding bit is set if the
comparator output is high (DAC output < pulse height).
Analog and Digital Electronics for Detectors 147
START STOP
VBL
PULSE
STRETCHER COMPARATOR
ANALOG
INPUT
IR
VBL
PEAK COUNTER
DETECTOR
OUTPUT
CLOCK
DIGITIZED
OUTPUT
Figure 34: Principle of a Wilkinson ADC. After the peak amplitude has been acquired,
the output of the peak detector initiates the conversion process. The memory capacitor is
discharged by a constant current while counting the clock pulses. When the capacitor is
discharged to the baseline level VBL the comparator output goes low and the conversion is
complete.
CLOCK
COUNTER
START S Q
DIGITIZED
STOP R OUTPUT
START STOP
Figure 35: The simplest form of time digitizer counts the number of clock pulses between
the start and stop signals.
Analog and Digital Electronics for Detectors 149
V+
START
STOP
IT COMPARATOR
IR C
VBL
COUNTER
CLOCK
DIGITIZED
OUTPUT
Figure 36: Combining a time-to-amplitude converter with an ADC forms a time digitizer
capable of ps resolution. The memory capacitor C is charged by the current IT for the
duration Tstart − Tstop and subsequently discharged by a Wilkinson ADC.
as illustrated in Figure 36. This technique provides high resolution, but at the
expense of dead time and multi-hit capability.
14 Signal Transmission
PRIMARY PULSE
REFLECTED
PULSE
2td
PULSE SHARPE
AT ORIGIN
Figure 37: Voltage pulse reflections on a transmission line terminated either with a short
(left) or open circuit (right). Measured at the sending end, the reflection from a short at
the receiving end appears as a pulse of opposite sign delayed by the round trip delay of the
cable. If the total delay is less than the pulse width, the signal appears as a bipolar pulse.
Conversely, an open circuit at the receiving end causes a reflection of like polarity.
must be smaller and the reflected voltage wave has the opposite sign. If the
termination is greater than the line impedance, the voltage wave is reflected
with the same polarity. Conversely, the current in the reflected wave is of like
sign when the termination is less than the line impedance and of opposite sign
when the termination is greater. Voltage reflections are illustrated in Figure 37.
At the sending end the reflected pulse appears after twice the propagation delay
of the √cable. Since in the presence of a dielectric the velocity of propagation
v = c/ ε, in typical coaxial and ribbon cables the delay is 5 ns/m.
Cable drivers often have a low output impedance, so the reflected pulse is
reflected again towards the receiver, to be reflected again, etc. This is shown in
Figure 38, which shows the observed signal when the output of a low-impedance
pulse driver is connected to a high-impedance amplifier input through a 4 m
long 50 Ω coaxial cable. If feeding a counter, a single pulse will be registered
multiple times, depending on the threshold level. When the amplifier input is
terminated with 50 Ω, the reflections disappear and only the original 10 ns wide
pulse is seen.
There are two methods of terminating cables, which can be applied either
Analog and Digital Electronics for Detectors 151
1
2
VOLTAGE (V)
VOLTAGE (V)
1
0.5
0
-1
0
-2
0 200 400 600 800 1000 0 200 400 600 800 1000
TIME (ns) TIME (ns)
Figure 38: Left: Signal observed in an amplifier when a low-impedance driver is connected
to the amplifier through a 4 m long coaxial cable. The cable impedance is 50 Ω and the
amplifier input appears as 1 kΩ in parallel with 30 pF. When the receiving end is properly
terminated with 50 Ω, the reflections disappear (right).
152 Helmuth Spieler
Z0
RT = Z0
RT = Z0 Z0
Figure 39: Cables may be terminated at the receiving end (top, shunt termination) or
sending end (bottom, series termination).
Qi
Cd
R
VBIAS
!V C
Figure 40: Noise on the detector bias line is coupled through the detector capacitance to
the amplifier input.
V1
COMMON
GROUND BUS
!V
V2
Figure 41: Shared current paths introduce common voltage drops to different circuits.
couple through the capacitor C. Naively, one might assume the ground plane
to be “clean”, but it can carry significant interference for the following reason.
One of the most common mechanisms for cross-coupling is shared current
paths, often referred to as “ground loops”. However, this phenomenon is not
limited to grounding. Consider two systems. The first is transmitting large
currents from a source to a receiver. The second is similar, but is attempting
a low-level measurement. Following the prevailing lore, both systems are con-
nected to a massive ground bus, as shown in Figure 41. Current seeks the path
of least resistance, so the large current from source V1 will also flow through the
ground bus. Although the ground bus is massive, it does not have zero resis-
tance, so the large current flowing through the ground system causes a voltage
drop ∆V .
154 Helmuth Spieler
In system 2 (source V2 ) both signal source and receiver are also connected
to the ground system. Now the voltage drop ∆V from system 1 is in series
with the signal path, so the receiver measures V2 + ∆V . The cross-coupling has
nothing to do with grounding per se, but is due to the common return path.
However, the common ground caused the problem by establishing the shared
path. This mechanism is not limited to large systems with external ground
busses, but also occurs on the scale of printed circuit boards and micron-scale
integrated circuits. At high frequencies the impedance is increased due to skin
effect and inductance. Note that for high-frequency signals the connections can
be made capacitively, so even if there is no DC path, the parasitic capacitance
due to mounting structures or adjacent conductor planes can be sufficient to
close the loop.
The traditional way of dealing with this problem is to reduce the impedance
of the shared path, which leads to the “copper braid syndrome”. However,
changes in the system will often change the current paths, so this “fix” is not
very reliable. Furthermore, in many detector systems – tracking detectors, for
example – the additional material would be prohibitive. Instead, it is best to
avoid the root cause.
+VDET
+V
Q2
Q1 Q3
OUTPUT
DETECTOR
–V
–VDET
Figure 42: The signal is transferred from the sensor to the input stage and from stage to
stage via local current loops.
Analog and Digital Electronics for Detectors 155
critical parts of this chain are the input, which is the most sensitive node, and
the output driver, which tends to circulate the largest current. Circuit diagrams
usually are not drawn like Figure 42; the bottom common line is typically shown
as ground. For example, in Figure 40 the sensor signal current flows through
capacitor C and reaches the return node of the amplifier through “ground”.
Clearly, it is critical to control this path and keep deleterious currents from this
area.
SUPPORT /
COOLING STAVE
DETECTOR
SIGNAL OUTPUT
DETECTOR BIAS
ISOLATION
RESISTORS
DETECTOR
SIGNAL OUTPUT
DETECTOR BIAS
ISOLATION
RESISTORS
Figure 43: Capacitive coupling between detectors or detector modules and their environment
introduces interference when relative potentials and stray capacitance are not controlled.
However superfluous grounding may be, one cannot let circuit elements
simply float with respect to their environment. Capacitive coupling is always
present and any capacitive coupling between two points of different potential
will induce a signal. This is illustrated in Figure 43, which represents individual
detector modules mounted on a support/cooling structure. Interference can
couple through the parasitic capacitance of the mount, so it is crucial to reduce
156 Helmuth Spieler
this capacitance and control the potential of the support structure relative to the
detector module. Attaining this goal in reality is a challenge, which is not always
met successfully. Nevertheless, paying attention to signal paths and potential
references early on is much easier than attempting to correct a poor design after
it’s done. Troubleshooting is exacerbated by the fact that current paths interact,
so doing the “wrong” thing sometimes brings improvement. Furthermore, only
one mistake can ruin system performance, so if this has been designed into the
system from the outset, one is left with compromises. Nevertheless, although
this area is rife with myths, basic physics still applies.
16 Conclusion
References
[2] http://www-physics.lbl.gov/∼spieler
[4] T. Kondo et al., Construction and performance of the ATLAS silicon mi-
crostrip barrel modules. Nucl. Instr. and Meth. A485 (2002) 27–42
[6] S. Ramo, Currents Induced by Electron Motion. Proc. IRE 27 (1939) 584–
585
[8] F. S. Goulding and D.A. Landis, Signal Processing for Semiconductor De-
tectors, IEEE Trans. Nucl. Sci. NS-29/3 (1982) 1125–1141
ABSTRACT
We present a concise review of the status of the Standard Model and of the
models of new physics.
The results of the electroweak precision tests as well as of the searches for the
Higgs boson and for new particles performed at LEP and SLC are now available
in nearly final form. Taken together with the measurements of mt , mW and
the searches for new physics at the Tevatron, and with some other data from
low energy experiments, they form a very stringent set of precise constraints
[1] to compare with the Standard Model (SM) or with any of its conceivable
extensions. When confronted with these results, on the whole the SM performs
rather well, so that it is fair to say that no clear indication for new physics
emerges from the data [2].
All electroweak Z pole measurements, combining the results of the 5 ex-
periments, are summarised in Table 1. Information on the Z partial widths are
contained in the quantities:
Here �� is the partial decay width for a pair of massless charged leptons. The
partial decay width for a given fermion species are related to the effective vector
and axial-vector coupling constants of the neutral weak current:
GF m3 � 2 �
Γf f̄ = NCf √ Z gAf CAf + gVf
2
CVf + ∆ew/QCD , (2)
6 2π
where NCf is the QCD colour factor, C{A,V}f are final-state QCD/QED cor-
rection factors also absorbing imaginary contributions to the effective coupling
162 Guido Altarelli
Table 1: Summary of electroweak precision measurements at high Q2 [3]. The first block
shows the Z-pole measurements. The second block shows additional results from other ex-
periments: the mass and the width of the W boson measured at the Tevatron and at LEP-2,
the mass of the top quark measured at the Tevatron, and the contribution to α(m2Z ) of the
hadronic vacuum polarisation.
The Electroweak Interactions in the Standard Model and beyond 163
constants, gAf and gVf are the real parts of the effective couplings, and ∆ con-
tains non-factorisable mixed corrections.
Besides total cross sections, various types of asymmetries have been mea-
sured. The results of all asymmetry measurements are quoted in terms of the
asymmetry parameter Af , defined in terms of the real parts of the effective
coupling constants, gVf and gAf , as:
where T3f is the third component of the weak iso-spin and qf the electric charge
of the fermion. The effective electroweak mixing angle is thus given indepen-
dently of the ρ parameter by the ratio gVf /gAf and hence in a one-to-one relation
by each asymmetry result.
The various asymmetries determine the effective electroweak mixing angle
lept
for leptons with highest sensitivity. The results on sin2 θeff are compared in
Figure 1. The weighted average of these six results, including small correlations,
is:
lept
sin2 θeff = 0.23150 ± 0.00016 . (5)
Note, however, that this average has a χ2 of 10.5 for 5 degrees of freedom,
corresponding to a probability of 6.2%. The χ2 is pushed up by the two most
lept
precise measurements of sin2 θeff , namely those derived from the measurements
of A� by SLD, dominated by the left-right asymmetry A0LR , and of the forward-
backward asymmetry measured in bb production at LEP, A0, b
FB , which differ by
about 2.9 standard deviations. No experimental effect in either measurement
has been identified to explain this, thus the difference is presumably either the
164 Guido Altarelli
Final
0,l
A fb 0.23099 0.00053
A l(SLD) 0.23098 0.00026
A l(Pt) 0.23159 0.00041
had
Q fb 0.2324 0.0012
Preliminary
0,b
A fb 0.23212 0.00029
0,c
A fb 0.23223 0.00081
(5)
2 Dahad = 0.02761 0.00036
10
mt = 178.0 4.3 GeV
on top of the small window between ∼ 2 and ∼ 3 which is allowed, on the one
side, by the direct search limit (mH > ∼ 114 GeV from LEP-2 [8]), and, on the
other side, by the theoretical upper limit on the Higgs mass in the minimal SM,
mH <∼ 600 − 800 GeV [9].
Fit 1 2 3
Measurements mW , ΓW mt mt , mW , ΓW
mt (GeV) 178.5+11.0
−8.5 177.2 ± 4.1 178.1 ± 3.9
mH (GeV) 117+162
−62 129+76
−50 113+62
−42
log [mH (GeV)] 2.07+0.38
−0.33 2.11 ± 0.21 2.05 ± 0.20
αs (mZ ) 0.1187 ± 0.0027 0.1190 ± 0.0027 0.1186 ± 0.0027
2
χ /dof 16.3/12 15.0/11 16.3/13
mW (MeV) 80386 ± 23
Table 2: Standard Model fits of electroweak data. All fits use the Z pole results and
(5)
∆αhad (m2Z ) as listed in Table 1, also including constants such as the Fermi constant GF . In
addition, the measurements listed in each column are included as well. For fit 2, the expected
W mass is also shown. For details on the fit procedure see [3].
10
epsilon . 103
1000
3
5 1
3
1
1000
0
m H = 114 —1000 GeV CDF - D0
-5 b
b
2
114
-10 2
Figure 2: The epsilon variables: comparison of the data with the SM predictions. The data
should be horizontal bands but they are shown here near the central value of mt .
When confronted with these results, on the whole the SM performs rather
well, so that it is fair to say that no clear indication for new physics emerges
from the data. However, as already mentioned, one problem is that the two
lept
most precise measurements of sin2 θeff from ALR and A0, b
FB differ nearly three
standard deviations. In general, there appears to be a discrepancy between
lept
sin2 θeff measured from leptonic asymmetries ((sin2 θeff )l ) and from hadronic
asymmetries ((sin2 θeff )h ), see also Figure 1. In fact, the result from ALR is
in good agreement with the leptonic asymmetries measured at LEP, while all
hadronic asymmetries, though their errors are large, are better compatible with
the result of A0, b
FB .
The situation is shown in Figure 3 [15]. The values of (sin2 θeff )l , (sin2 θeff )h
and their formal combination are shown each at the mH value that would corre-
spond to it given the central value of mt . Of course, the value for mH indicated
lept
by each sin2 θeff has an horizontal ambiguity determined by the measurement
error and the width of the ±1σ band for mt . Even taking this spread into ac-
count it is clear that the implications on mH are sizably different. One might
imagine that some new physics effect could be hidden in the Zbb̄ vertex. Like
for the top quark mass there could be other non decoupling effects from new
170 Guido Altarelli
178.0 GeV
lept
eff
sin2 q
182.3 GeV
lept.asymm
0.231
MH [GeV]
Figure 3: lept
The data for sin2 θeff are plotted vs mH . For presentation purposes the measured
points are shown each at the mH value that would ideally correspond to it given the central
value of mt (updated from [15]).
heavy states or a mixing of the b quark with some other heavy quark. However,
it is well known that this discrepancy is not easily explained in terms of some
new physics effect in the Zbb̄ vertex. In fact, A0, b
FB is the product of lepton-
and b-asymmetry factors: A0, b 0, b
FB = (3/4)Ae Ab . The sensitivity of AFB to Ab
is limited, because the Ae factor is small, so that a rather large change of the
b-quark couplings with respect to the SM is needed in order to reproduce the
measured discrepancy (precisely a ∼ 30% change in the right-handed coupling,
an effect too large to be a loop effect but which could be produced at the tree
level, e.g., by mixing of the b quark with a new heavy vectorlike quark [16]).
But then this effect should normally also appear in the direct measurement of
Ab performed at SLD using the left-right polarized b asymmetry, even within
the moderate precision of this result, and it should also be manifest in the accu-
rate measurement of Rb ∝ gRb 2
+ gLb
2
. The measurements of neither Ab nor Rb
confirm the need of a new effect. Even introducing an ad hoc mixing the overall
The Electroweak Interactions in the Standard Model and beyond 171
fit is not terribly good, but we cannot exclude this possibility completely. Alter-
natively, the observed discrepancy could be due to a large statistical fluctuation
or an unknown experimental problem. The ambiguity in the measured value of
lept
sin2 θeff could thus be larger than the nominal error, reported in Equation 5,
obtained from averaging all the existing determinations.
We have already observed that the experimental value of mW (with good
agreement between LEP and the Tevatron) is a bit high compared to the SM
prediction (see Figure 4). The value of mH indicated by mW is on the low side,
lept
just in the same interval as for sin2 θeff measured from leptonic asymmetries. It
is interesting that the new value of mt considerably relaxes the previous tension
lept
between the experimental values of mW and sin2 θeff measured from leptonic
asymmetries on one side and the lower limit on mH from direct searches on the
other side [17, 18]. This is also apparent from Figure 4.
80.5
M W [GeV]
80.45
MW word average
80.4
Mt=182.3 GeV
178 GeV
80.35
173.7 GeV
80.3
Figure 4:The world average for mW is compared with the SM prediction as a function of
mH (updated from [15]).
The main lesson of precision tests of the standard electroweak theory can
be summarised as follows. The couplings of quark and leptons to the weak gauge
bosons W± and Z are indeed precisely those prescribed by the gauge symmetry.
The accuracy of a few per-mille for these tests implies that, not only the tree
level, but also the structure of quantum corrections has been verified. To a lesser
172 Guido Altarelli
accuracy the triple gauge vertices γW+ W− and ZW+ W−�have also been found
in agreement with the specific prediction of the SU (2) U (1) gauge theory.
This means that it has been verified that the gauge symmetry is unbroken in
the vertices of the theory: the currents are indeed conserved. Yet there is
obvious evidence that the symmetry is otherwise badly broken in the masses.
Thus the currents are conserved but the spectrum of particle states is not at
all symmetric. This is a clear signal of spontaneous symmetry breaking. The
practical implementation of spontaneous symmetry breaking in a gauge theory
is via the Higgs mechanism. The Higgs sector of the SM is still very much
untested. What has been tested is the relation m2W = m2Z cos2 θW , modified by
computable radiative corrections. This relation means that the effective Higgs
(be it fundamental or composite) is indeed a weak isospin doublet. The Higgs
particle has not been found but in the SM its mass can well be larger than the
present direct lower limit mH >
∼ 114 GeV obtained from direct searches at LEP-
2. The radiative corrections computed in the SM when compared to the data
on precision electroweak tests lead to a clear indication for a light Higgs, not
too far from the present lower bound. No signal of new physics has been found.
However, to make a light Higgs natural in presence of quantum fluctuations
new physics should not be too far. This is encouraging for the LHC that should
experimentally clarify the problem of the electroweak symmetry breaking sector
and search for physics beyond the SM.
Given the success of the SM why are we not satisfied with that theory?
Why not just find the Higgs particle, for completeness, and declare that particle
physics is closed? The reason is that there are both conceptual problems and
phenomenological indications for physics beyond the SM. On the conceptual
side the most obvious problems are that quantum gravity is not included in
the SM and the related hierarchy problem. Among the main phenomenological
hints for new physics we can list coupling unification, dark matter, neutrino
masses, baryogenesis and the cosmological vacuum energy.
The computed evolution with energy of the effective SM gauge couplings
clearly points towards the unification of the electro-weak and strong forces
(Grand Unified Theories: GUT’s) at scales of energy MGU T ∼ 1015 − 1016 GeV
which are close to the scale of quantum gravity, MP l ∼ 1019 GeV . One is led
to imagine a unified theory of all interactions also including gravity (at present
superstrings provide the best attempt at such a theory). Thus GUT’s and the
realm of quantum gravity set a very distant energy horizon that modern parti-
cle theory cannot ignore. Can the SM without new physics be valid up to such
The Electroweak Interactions in the Standard Model and beyond 173
large energies? This appears unlikely because the structure of the SM could
not naturally explain the relative√smallness of the weak scale of mass, set by
the Higgs mechanism at µ ∼ 1/ GF ∼ 250 GeV with GF being the Fermi
coupling constant. This so-called hierarchy problem is related to the presence
of fundamental scalar fields in the theory with quadratic mass divergences and
no protective extra symmetry at µ = 0. For fermion masses, first, the�diver-
gences are logarithmic and, second, they are forbidden by the SU (2) U (1)
gauge symmetry plus the fact that at m = 0 an additional symmetry, i.e. chiral
symmetry, is restored. Here, when talking of divergences, we are not worried of
actual infinities. The theory is renormalisable and finite once the dependence
on the cut off is absorbed in a redefinition of masses and couplings. Rather
the hierarchy problem is one of naturalness. We should see the cut off as a pa-
rameterization of our ignorance on the new physics that will modify the theory
at large energy scales. Then it is relevant to look at the dependence of phys-
ical quantities on the cut off and to demand that no unexplained enormously
accurate cancellations arise.
The hierarchy problem can be put in very practical terms: loop corrections
to the higgs mass squared are quadratic in Λ. The most pressing problem is
from the top loop. With m2h = m2bare + δm2h the top loop gives
3GF 2 2
δm2h|top ∼ √ mt Λ ∼ (0.3Λ)2 (10)
2π 2
If we demand that the correction does not exceed the light Higgs mass
indicated by the precision tests, Λ must be close, Λ ∼ o(1 T eV ). Similar
constraints arise from the quadratic Λ dependence of loops with gauge bosons
and scalars, which, however, lead to less pressing bounds. So the hierarchy
problem demands new physics to be very close (in particular the mechanism
that quenches the top loop). Actually, this new physics must be rather special,
because it must be very close, yet its effects are not clearly visible (the ”LEP
Paradox” [19]). Examples of proposed classes of solutions for the hierarchy
problem are:
Supersymmetry. In the limit of exact boson-fermion symmetry the
quadratic divergences of bosons cancel so that only log divergences remain.
However, exact SUSY is clearly unrealistic. For approximate SUSY (with soft
breaking terms), which is the basis for all practical models, Λ is replaced by the
splitting of SUSY multiplets, Λ ∼ mSU SY − mord . In particular, the top loop
is quenched by partial cancellation with s-top exchange.
Technicolor. The Higgs system is a condensate of new fermions. There
are no fundamental scalar Higgs sector, hence no quadratic devergences asso-
ciated to the µ2 mass in the scalar potential. This mechanism needs a very
174 Guido Altarelli
top (the loop corrections increase with log m̃t ). Stringent naturality constraints
2
also follow from imposing that the electroweak symmetry breaking occurs at the
right place: in SUSY models the breaking is induced by the running of the Hu
mass starting from a common scalar mass m0 at MGU T . The squared Z mass
176 Guido Altarelli
m2Z can be expressed as a linear combination of the SUSY parameters m20 , m21/2 ,
A2t , µ2 ,... with known coefficients. Barring cancellations that need fine tuning,
the SUSY parameters, hence the SUSY s-partners cannot be too heavy. The
LEP limits, in particular the chargino lower bound mχ+ > ∼ 100 GeV , are suffi-
cient to eliminate an important region of the parameter space, depending on the
amount of allowed fine tuning. For example, models based on gaugino univer-
sality at the GUT scale are discarded unless a fine tuning by at least a factor of
20 is not allowed. Without gaugino universality [27] the strongest limit remains
on the gluino mass: m2Z ∼ 0.7 m2gluino + . . . which is still compatible with the
present limit mgluino > ∼ 200 GeV .
The non discovery of SUSY at LEP has given further impulse to the quest
for new ideas on physics beyond the SM. Large extra dimensions [28] and ”little
Higgs” [29] models are the most interesting new directions in model building.
Large extra dimension models propose to solve the hierarchy problem by bring-
ing gravity down from MP l to m ∼ o(1 T eV ) where m is the string scale.
Inspired by string theory one assumes that some compactified extra dimensions
are sufficiently large and that the SM fields are confined to a 4-dimensional
brane immersed in a d-dimensional bulk while gravity, which feels the whole
geometry, propagates in the bulk. We know that the Planck mass is large be-
cause gravity is weak: in fact GN ∼ 1/MP2 l , where GN is Newton constant.
The idea is that gravity appears so weak because a lot of lines of force escape
in extra dimensions. Assume you have n = d − 4 extra dimensions with com-
pactification radius R. For large distances, r >> R, the ordinary Newton law
applies for gravity: in natural units F ∼ GN /r2 ∼ 1/(MP2 l r2 ). At short dis-
tances, r < ∼ R, the flow of lines of force in extra dimensions modifies Gauss
law and F −1 ∼ m2 (mr)d−4 r2 . By matching the two formulas at r = R one
obtains (MP l /m)2 = (Rm)d−4 . For m ∼ 1 T eV and n = d − 4 one finds that
n = 1 is excluded (R ∼ 1015 cm), for n = 2 R is at the edge of present bounds
R ∼ 1 mm, while for n = 4, 6, R ∼ 10−9 , 10−12 cm. In all these models a
generic feature is the occurrence of Kaluza-Klein (KK) modes. Compactified di-
mensions with periodic boundary conditions, as for quantization in a box, imply
a discrete spectrum with momentum p = n/R and mass squared m2 = n2 /R2 .
There are many versions of these models. The SM brane can itself have a thick-
ness r with r <∼ 10−17 cm or 1/r >∼ 1 T eV , because we know that quarks
and leptons are pointlike down to these distances, while for gravity there is no
experimental counter-evidence down to R <∼ 0.1 mm or 1/R >∼ 10−3 eV .
In case of a thickness for the SM brane there would be KK recurrences for SM
fields, like Wn , Zn and so on in the T eV region and above. There are models
with factorized metric (ds2 = ηµν dxµ dxν + hij (y)dy i dy j , where y (i,j) denotes
the extra dimension coordinates (and indices), or models with warped metric
The Electroweak Interactions in the Standard Model and beyond 177
(ds2 = e − 2kR|φ|ηµν dxµ dxν − R2 φ2 [30]. In any case there are the towers of
KK recurrences of the graviton. They are gravitationally coupled but there are
a lot of them that sizably couple, so that the net result is a modification of
cross-sections and the presence of missing energy.
Large extra dimensions provide a very exciting scenario [31]. Already it is
remarkable that this possibility is compatible with experiment. However, there
are a number of criticisms that can be brought up. First, the hierarchy problem
is more translated in new terms rather than solved. In fact, the basic relation
Rm = (MP l /m)2/n shows that Rm, which one would apriori expect to be 0(1),
is instead ad hoc related to the large ratio MP l /m. In this respect the Randall-
Sundrum variety is more appealing because the hierarchy suppression mW /MP l
could arise from the warping factor e−2kR|φ| , with not too large values of kR.
Also it is not clear how extra dimensions can by themselves solve the LEP
paradox (the large top loop corrections should be controlled by the opening of
the new dimensions and the onset of gravity): since mH is light Λ ∼ 1/R must
be relatively close. But precision tests put very strong limits on Λ. In fact, in
typical models of this class, there is no mechanism to sufficiently quench the
corrections. No simple, realistic model has yet emerged as a benchmark. But
it is attractive to imagine that large extra dimensions could be a part of the
truth, perhaps coupled with some additional symmetry or even SUSY.
In the extra dimension general context an interesting direction of devel-
opment is the study of symmetry breaking by orbifolding and/or boundary
conditions. These are models where a larger gauge symmetry (with or without
SUSY) holds in the bulk. The symmetry is reduced in the 4 dimensional brane,
where the physics that we observe is located, as an effect of symmetry breaking
induced geometrically by suitable boundary conditions. There are models where
SUSY, valid in n > 4 dimensions is broken by boundary conditions [32], in par-
ticular the model of Ref.[33], where the mass of the Higgs is computable and can
be extimated with good accuracy. Then there are ”Higgsless models” where it
is the SM electroweak gauge symmetry which is broken at the boundaries [34].
Or models where the Higgs is the 5th component of a gauge boson of an ex-
tended symmetry valid in n > 4 [35]. In general all these alternative models
for the Higgs mechanism face severe problems and constraints from electroweak
precision tests [36]. At the GUT scale, symmetry breaking by orbifolding can
be applied to obtain a reformulation of SUSY GUT’s where many problematic
features of ordinary GUT’s (e.g. a baroque Higgs sector, the doublet-triplet
splitting problem, fast proton decay etc) are improved [31], [37].
In ”little Higgs” models, the symmetry of the SM is extended to a � suitable
global group G that also � contains some gauge � enlargement of SU (2) U (1),
for example G ⊃ [SU (2) U (1)]2 ⊃ SU (2) U (1). The Higgs particle is a
178 Guido Altarelli
pseudo-Goldstone boson of G that only takes mass at 2-loop level, because two
distinct symmetries must be simultaneously broken for it to take mass, which
requires the action of two different couplings in the same diagram. Then in the
relation between δm2h and Λ2 there is an additional coupling and an additional
loop factor that allow for a bigger separation between the Higgs mass and the
cut-off. Typically, in these models one has one or more Higgs doublets at
mh ∼ 0.2 T eV , and a cut-off at Λ ∼ 10 T eV . The top loop quadratic cut-off
dependence is partially canceled, in a natural way guaranteed by the symmetries
of the model, by a new coloured, charge-2/3, vectorial quark χ of mass around
1 T eV (a fermion not a scalar like the s-top of SUSY models). Certainly these
models involve a remarkable level of group theoretic virtuosity. However, in the
simplest versions one is faced with problems with precision tests of the SM [38].
Even with vectorlike new fermions large corrections to the epsilon parameters
arise from exchanges of the new gauge bosons W � and Z � (due to lack of custodial
SU (2) symmetry). In order to comply with these constraints the cut-off must
be pushed towards large energy and the amount of fine tuning needed to keep
the Higgs light is still quite large. Probably these bad features can be fixed by
some suitable complication of the model (see for example, [39]). But, in my
opinion, the real limit of this approach is that it only offers a postponement
of the main problem by a few TeV, paid by a complete loss of predictivity at
higher energies. In particular all connections to GUT’s are lost.
Finally, we stress the importance of the cosmological constant or vacuum
energy problem [40]. The exciting recent results on cosmological parameters,
culminating with the precise WMAP measurements [41], have shown that vac-
uum energy accounts for about 2/3 of the critical density: ΩΛ ∼ 0.65, Trans-
lated into familiar units this means for the energy density ρΛ ∼ (2 10−3 eV )4
1/4
or (0.1 mm)−4 . It is really interesting (and not at all understood) that ρΛ ∼
Λ2EW /MP l (close to the range of neutrino masses). It is well known that in field
theory we expect ρΛ ∼ Λ4cutof f . If the cut off is set at MP l or even at 0(1 T eV )
there would an enormous mismatch. In exact SUSY ρΛ = 0, but SUSY is bro-
1/4
ken and in presence of breaking ρΛ is in general not smaller than the typical
SUSY multiplet splitting. Another closely related problem is ”why now?”: the
time evolution of the matter or radiation density is quite rapid, while the den-
sity for a cosmological constant term would be flat. If so, them how comes that
precisely now the two density sources are comparable? This suggests that the
vacuum energy is not a cosmological constant term, buth rather the vacuum ex-
pectation value of some field (quintessence) and that the ”why now?” problem
is solved by some dynamical mechanism.
Clearly the cosmological constant problem poses a big question mark on
the relevance of naturalness as a relevant criterion also for the hierarchy prob-
The Electroweak Interactions in the Standard Model and beyond 179
lem: how we can trust that we need new physics close to the weak scale out
of naturalness if we have no idea on the solution of the cosmological constant
huge naturalness problem? The common answer is that the hierarchy problem
is formulated within a well defined field theory context while the cosmological
constant problem makes only sense within a theory of quantum gravity, that
there could be modification of gravity at the sub-eV scale, that the vacuum en-
ergy could flow in extra dimensions or in different Universes and so on. At the
other extreme is the possibility that naturalness is misleading. Weinberg [42]
has pointed out that the observed order of magnitude of Λ can be successfully
reproduced as the one necessary to allow galaxy formation in the Universe. In a
scenario where new Universes are continuously produced we might be living in a
very special one (largely fine-tuned) but the only one to allow the development
of an observer. One might then argue that the same could in principle be true
also for the Higgs sector. Recently it was suggested [43] to abandon the no-
fine-tuning assumption for the electro-weak theory, but require correct coupling
unification, presence of dark matter with weak couplings and a single scale of
evolution from the EW to the GUT scale. A ”split SUSY” model arises as a
solution with a fine-tuned light Higgs and all SUSY particles heavy except for
gauginos, higgsinos and neutralinos, protected by chiral symmetry. Or we can
have a two-scale non-SUSY GUT with axions as dark matter. In conclusion, it
is clear that naturalness can be a good heuristic principle but you cannot prove
its necessity.
Supersymmetry remains the standard way beyond the SM. What is unique to
SUSY, beyond leading to a set of consistent and completely formulated models,
as, for example, the MSSM, is that this theory can potentially work up to the
GUT energy scale. In this respect it is the most ambitious model because it
describes a computable framework that could be valid all the way up to the
vicinity of the Planck mass. The SUSY models are perfectly compatible with
GUT’s and are actually quantitatively supported by coupling unification and
also by what we have recently learned on neutrino masses. All other main ideas
for going beyond the SM do not share this synthesis with GUT’s. The SUSY
way is testable, for example at the LHC, and the issue of its validity will be
decided by experiment. It is true that we could have expected the first signals
of SUSY already at LEP, based on naturality arguments applied to the most
minimal models (for example, those with gaugino universality at asymptotic
scales). The absence of signals has stimulated the development of new ideas
like those of large extra dimensions and ”little Higgs” models. These ideas are
180 Guido Altarelli
very interesting and provide an important reference for the preparation of LHC
experiments. Models along these new ideas are not so completely formulated
and studied as for SUSY and no well defined and realistic baseline has so far
emerged. But it is well possible that they might represent at least a part of the
truth and it is very important to continue the exploration of new ways beyond
the SM.
I would like to express my gratitude to the Organisers of the ICFA ’03
School for their invitation and their magnificent hospitality in Itacuruca. In
particular I would like to thank Bernard Marechal.
References
[3] The ALEPH, DELPHI, L3, OPAL, SLD Collaborations and the LEP Elec-
troweak Working Group, A Combination of Preliminary Electroweak Mea-
surements and Constraints on the Standard Model, hep-ex/0312023, and
references therein.
[5] The CDF Collaboration, the DØ Collaboration, and the Tevatron Elec-
troweak Working Group, Combination of CDF and DØ Results on the
Top-Quark Mass, hep-ex/0404010.
[8] The ALEPH, DELPHI, L3 and OPAL Collaborations, and the LEP Work-
ing Group for Higgs Boson Searches, Search for the Standard Model Higgs
Boson at LEP, hep-ex/0306033, Phys. Lett. B565 (2003) 61–75.
[9] Th. Hambye, K. Riesselmann, Matching Conditions and Higgs Mass Upper
Bounds Revisited, hep-ph/9610272, Phys. Rev. D55 (1997) 7255–7262.
[11] The NuTeV Collaboration, G.P. Zeller et al., Phys. Rev. Lett. 88 (2002)
091802.
[12] M. Yu. Kuchiev and V. V. Flambaum, Radiative Corrections to Parity
Non-Conservation in Atoms, hep-ph/0305053.
[13] The SLAC E158 Collaboration, P.L. Anthony et al., Observation of Parity
Non-Conservation in Moller scattering, hep-ex/0312035, A New Measure-
ment of the Weak Mixing Angle, hep-ex/0403010. We have added 0.0003 to
the value of sin2 θ(mZ ) quoted by E158 in order to convert from the MSbar
scheme to the effective electroweak mixing angle [4].
[14] S. Davidson et al, hep-ph/0112302.
[15] P. Gambino, The Top Priority: Precision Electroweak Physics from Low-
Energy to High-Energy, hep-ph/0311257.
[16] D. Choudhury, T.M.P. Tait, C.E.M. Wagner, Beautiful Mirrors and Pre-
cision Electroweak Data, hep-ph/0109097, Phys. Rev. D65 (2002) 053002.
[17] M. S. Chanowitz, Electroweak Data and the Higgs Boson Mass: A Case
for New Physics, hep-ph/0207123, Phys. Rev. D66 (2002) 073002.
[18] G. Altarelli, F. Caravaglios, G.F. Giudice, P. Gambino, G. Ridolfi, Indica-
tion for Light Sneutrinos and Gauginos from Precision Electroweak Data,
hep-ph 0106029, JHEP 0106:018, 2001.
[19] R. Barbieri and A. Strumia, hep-ph/0007265.
[20] K. Lane, hep-ph/0202255,
R. S. Chivukula, hep-ph/0011264.
[21] For a recent introduction see, for example, S. P. Martin, hep-ph/9709356.
[22] M. Dine and A. E. Nelson, Phys. Rev. D48 (1993) 1277;
M. Dine, A. E. Nelson and Y. Shirman, Phys. Rev. D51 (1995) 1362;
G. F. Giudice and R. Rattazzi, Phys. Rept. 322 (1999) 419.
[23] L. Randall and R. Sundrum, Nucl. Phys. B557 (1999) 79;
G.F. Giudice et al, JHEP 9812 (1998) 027.
[24] P. Langacker and N. Polonsky, Phys. Rev. D52 (1995) 3081.
[25] G. Altarelli, F. Feruglio and I. Masina, JHEP 0011:040,2000.
[26] For a review see, for example, G. Altarelli and F. Feruglio, hep-ph/0405048.
182 Guido Altarelli
ABSTRACT
There are instances of high energy particle interactions where one, or both, of
the interacting particles survives the proccess experiencing only a slight scatter-
ing that deviates it from the original beam. The identification of the particles
scattered at small angles is of uppermost importance in the study of such events.
Typical 4π detectors usually miss these particles. We describe here a series of
detectors specially built to operate very close to the beam enlarging the capa-
bility of the experiment to study the physics of this kind of event.
1 Introduction
The detection of the particles scattered at small angles requires the posi-
tioning of special detectors very close to the beam, usually inside the accelerator
beampipe. That must be done with consideration to the very critical conditions
existing in such environment, specially the ultra high vacuum (UHV) that may
be present in the area. The detectors may have to be moved close to beam or
removed to clear the area during accelerator insertion periods. Its location must
be known within some precision and accuracy and a high degree of repeatability
must be achieved.
Accelerator vacuum may reach up to 10−10 Torr, or even more, and any
apparatus built to operate in such environment must not disturd this vacuum.
We can use a simple description to help understanding the meaning of such
vacuum.
Consider a cube whose sides are 10.0 cm long (a 1 L volume). Imagine the
inside of this cube completely void of particles. That would mean a 0.0 Torr
vacuum. If we cover the cube surface with a single layer of N2 molecules we will
need 6.0×1017 N2 molecules. Throwing all these molecules into the cube volume
the vacuum will correspond to 1.7 × 10−2 Torr. This is 8 orders of magnitude
bigger than 10−10 Torr. To achieve UHV, we must remove N2 molecules from
inside the cube, leaving only 1 out of every 100, 000, 000 molecules in there !
This is not simple to do. The tinyiest impurity on the material surface
would make our efforts vain. Even when completely clean of impurities, the
material would degass (liberating particles that are embeded in the material,
like water, nitrogen). This degassing will also ruin the attempts to achieve
UHV.
A UHV chamber is built following a series of steps, from which we shall
mention:
• TIG (Tungsten Inert Gas) welding (a proccess where a cloud of inert gas
is set around the welding point). Welding is made in the interior of the
chamber, so there are no air pockets left inside.
3 Roman Pots
Roman pots are stainless containers that allow the position detector to
function outside of the machine ultra high vacuum, but close to the beam. The
scattered particle traverses a thin steel window at the entrance of the pot, goes
through the detector that rests in the pot and exit the pot through another thin
window. The pots are remotely controlled and can be moved close to the beam
during stable conditions.
The first ever built Roman pot[3] was used in 1970-1972 at the ISR by
the CERN-Rome group. It did not have a thin window and housed a small
hodoscope of scintillating counters. Figure1 shows the first Roman pot and a
later Roman pot that was used by UA4 at SPS. The UA4 Roman pot section
facing the beam is concave in shape, allowing a closer aproach to the beam. A
thin window is used to reduce the material between the beam and the detector
inside the pot.
More recently, CDF installed a set of three Roman pots with a scintillating
fiber detector, as viewed in Figure2. When hit by the scattered particle, the fiber
scintillates and the light is guided to a multianode photomultiplier that collects
the light. Figure2 also shows details of the CDF scintillating fiber detector.
186 Hélio da Motta
Figure 1: First Roman pot (left). UA4 Roman pot (right). The concave bottom of the UA4
Roman pot allows a closer aproach to the beam and the thin window reduces the material
the particle has to traverses before hiting the detector that rests inside the pot.
7.62
20
30
BO
Bellows
Recoil Track To MCPMT Fiber
Tracker
Detail
To MCPMT
CDF
P X
Y
Figure 2: CDF Roman pot system. Schematics of the detector is shown. It consists of an
array of scintillating fibers oriented in the X and Y direction. The scattered particle leaves a
signal in a pair of XY channel that is used to determine its position.
to move perpendicularly to the beam in a very precise, safe and accurate way.
The system is operated by a step motor and a set of reduction gears allows pot
motion with a precision of aproximately 5 µm. A set of cylindrical and conical
bearings allows adjustment of the pot alignment and a linear variable differential
transducer (LVDT) monitors the pot position. A steel bellows guaranties the
movement of the pot whithout affecting the vacuum.
The FPD driving assembly is basically a cylindrical tube threaded on
its outer surface that is actuated by a worm-gear system. The system has
a reduction of 120 times which accounts both for a precision of about 5µm
movement of the detector and for compensation for the inward force of about
2, 000 N resulting from the vacuum inside the castle. This arrangement makes
possible the use of small low torque motors and makes unecessary the use of any
vacuum compensation system. An exploded view of the driving system is shown
in Figure5. The compacteness of the whole piece makes it easy to employ in as
many units as needed. Details of the driving assembly structure is presented in
Figure6. The moving parts get a thin layer of a molibidenium lubricant.
188 Hélio da Motta
DO
Interaction
P
A2U A1U Region P1U P2U
P A2O A10 P10 P20 P
Figure 3: FPD layout at DØ experimental area. Quadrupole castles are named P or A when
placed on the p side or the p̄, respectively.
279.5
145.6
1234
MAPMT
Sci Fibers
4 mm
4.5 mm
Clear Fibers
30 mm
1 2
3 4
4 mm
2 mm
Fibers
II
II
PM
SCINTILLANTING
COUNTER
6 Conclusions
Near beam detectors have been used for over 30 years in different acceler-
ators with different beams. They allow the measurement of several proccesses
(like elastic scattering and diffraction dissociation at high energy) that can not
be as well studied without these detectors.
They are as more important as higher the energy. Consequently, they
shall be of great interest in the new LHC era that is to start in a few years. The
technology to design and built such detectors is already of common knowledge
and has been tested successfully. They are safe to operate and can become an
integrated part of the accelerator structure.
7 Acknowledgments
The author thanks the Organizers of the ICFA School for the invitation
to present this talk in Itacuruçá, and thanks Marcelo Juni (LNLS) and Regis
Neuenschwander (LNLS) for their very useful information on vacuum and me-
chanics. Special thanks goes to Alberto Santoro (UERJ) whose support and
determination has made the FPD possible.
References
ABSTRACT
Computing is one of the technologies associated to High Energy Physics (HEP).
The needs of the next generation CERN HEP experiments push the development
of new computing architectures. The scale of data will be orders of magnitude
larger than the one we currently have. In this talk we will present the current
status of GRID computing development.
1 Introduction
The Grid will permeate all science, allowing groups around the world to
collaborate objectively and giving different regions the oportunity to take part
in the science frontier. This technology will be present in our professional life
for a long time. New chips and computer memmories, new hard disks, the Raid
technology, are developments that help to advance the proposed computing
architecture for HEP even faster.
1.1 Origin
HEP physicists have always been involved with technology and, since
the begining, computing has been present as one of the main tools. Computing
is present in the accelerator, in the proposal, during the development, in each
detector project, in the data acquisition, in the analysis. Putting it simple,
computing is in all parts of a high energy physics experiment. Each experiment
builds its own instrumentation and new tools according to its needs. Grid
is a natural step forward. Frontiers of science imply frontier of technologies.
The World Wide Web was developed by Tim Bernes Lee team at CERN as
a need of HEP because collaborations were becoming more participative and
comunication was a priority.
The arrival of New Technologies (CPU, storage, networks, new lan-
guages) will be very useful in the very high energy collisions (14 TeV) environ-
ment with a much larger amount of events (tens of PetaBytes).
196 Alberto Santoro
1.2 Definitions
Before going ahead, I would like to give a few useful definitions acronyms
frequently used in the field.
OSG[2]: Open Science Grid. This is the simmilar organization for United
States Sciences.
HEP has one of the most advanced and suitable structures for GRID de-
velopment. A GRID has to take into acount each aspect of one HEP experiment
as, (i) Project for Particle Accelerators and Detectors; (ii) Data Acquisition Sys-
tems and Data Storage; (iii) Development of software and languages; (iv) Data
reconstruction using parallel processing; (v)Monitoring, control, simmulation,
security, networking; (vi) Data Analysis (one of the most important activities
in GRID computing for LHC era); and (vii) Video Conferencing (already used
at global level). All this defines a new world or a new way to work. The devel-
opment of GRID and applications will certainly represent a new revolution in
the internet. Several branches of society are being re-organized in view of these
new possibilities.
2 Projects
HEP GRID has a number of projects based on the planning of new collidder
experiments. These experiments and related Grid projects are summarized here.
should go to the Atlas[9] web page. Many technologies have been developed
by ATLAS collaboration to build one of the most interesting general purpose
detector. The goals of the experiment are: Detect the Standard Model Higgs
Boson, detect Supersymmetric states, study Standard Model QCD (Chromod-
inamics), EW (Electroweek), HQ (Heavy Quark) Physics, and new physics (to
be defined by each collaboration but, in general, we can say that new physics
is everything beyond the Standard Model). The Collaboration has been in
intensive collaborative development of GRID software. Some of the tools de-
veloped in collaboration with LHCb are: GridView ( Simple tool to monitor
status of testbed), Gripe (unified user accounts), Magda (Manager for Grid
Data), Pacman (package management and distribution tool), Grappa (web por-
tal using active notebook technology), GRAT (Grid Application Toolkit), Grd-
searcher (browser), GridExpert (Knowledge Database), VOToolkit (Site Au-
thentication,Authorization).
LHCb[10], is a detector dedicated mainly to b physics. The LHCb (Large
Hadron Collider Beauty Experiment) experiment has 563 physicists from 50 In-
stitutes from 12 Countries. The experiment expects to get 1012 bb pairs per year,
a much higher statistics than the current B factories. Another set of numbers ex-
pected for LHCb experiments is: (i) 200,000 reconstructed B 0 → J/ψKs events
per year; (ii) 26,000 reconstructed B 0 → π + π − ; (iii) all B Mesons and Barions.
The LHCb collaboration has produced many useful software for analysis in the
near future in cooperation with ATLAS. Two examples, in addition to those
pointed out above on ATLAS subsection are GANGA ( Gaudi ANd Grid), an
user interface for Grid, and DIRAC ( Distributed Infrastructure with Remote
Agent Control) for Monte Carlo event production. Details of this experiment
can be found on the LHCb[10] web page.
CMS[11] collaboration has approximately 2000 physicists from 160 Insti-
tutions from about 40 countries. It weights 12,500 Tons, is 15 m in diameter
and 22 m in length. It will use a Magnet field of 4 Tesla.
The Detector is composed by Silicon Microstrips as a central Tracker;
Electromagnetic and Hadronic Calorimeters; a Superconducting Coil; Iron Yoke;
Muon Barrel and Muon Endcaps. CMS will explore a big number of physics
topics. In the case of Higgs, the collaboration intends to explore the full range
of 100 - 1000 GeV as the allowed region for Higgs mass. Topics as QCD,
Heavy Flavor Physics, SUSY, New Phenomena are in the list of future analysis.
Diffractive Physics will be explored by the Totem Group (for elastic and total
cross section only) and by a large number of physicists interested in exploring
all innelastic diffraction and respective topologies. Hard diffraction will be the
central part of these studies.
200 Alberto Santoro
tion that consisted in transmitting a record of 101 Gbps. This was one of the
activities of the Bandwidth Challenge of the Supercomputing 2005 conference.
We expect that the T2-HEPGRID BRAZIL will soon be connected to the full
circuit of CMS Grid. Basic information about Grid systems can be find on the
web[16, 17, 18, 19]
5 Acknowledgments
The author thanks the Organizers of the ICFA School for the invitation to
present this talk in Itacuruçá, where we found a very good ambient created by
the participants. I would like to thank FAPERJ and CNPq by partial financial
support. Finally, I thank my colleagues P. Avery, H. Newman, D. Barberis, J.
Bunn, R. Gardness, R. Mount, S. Bgnaco, P. Cerello, R. Barbera, P. Buncic,
F. Caminati, P. Satz, G. Pulard, N. Brook, C.Eck, J. Marco, F. Gagliardi,
T. Wenaus and F. Harri for all the information about their projects. Finally
I would like to thank my colleague Helio da Mota for interesting discussions
during the preparation of this talk.
References
[1] http://public.eu-egee.org/
[2] http://www.opensciencegrid.org/
[3] http://lcg.web.cern.ch/LCG/
[4] http://www.ivdgl.org/
[5] http://www.griphyn.org/
[6] http://www.ppdg.net/
[7] Future Experiments - GRID and LHC Proceedings of the Simpósio Latino
Americano de Fı́sica de Altas Energias -Lima, Peru, July 12-17,2004.
[8] http://www.alice.cern.ch/
[9] http://www.atlas.cern.ch/
[10] http://www.lhcb.cern.ch/
[11] http://www.cms.cern.ch/
202 Alberto Santoro
[12] See talk ”Mamogrid: applying Grid Technology to Health Care (in
breast cancer diagnose)” by Salvator Roberto Amendolia (CERN) in
http://www.lishep.uerj.br/
[13] http://icfa-scic.web.cern.ch/ICFA-SCIC/;
http://www.weforum.org/site/homepublic.nsf/Content/Global+Digital+
Divide+Initiative; and the book ”The Digital Divide”, Edited by Benjamin
M. Campaine, 2003.
[14] http://www.hepgridbrasil.uerj.br/
[15] http://ultralight.caltech.edu/gaeweb/
[16] http://www.cs.wisc.edu/condor/
[17] http://www.globus.org
[18] (i) The Grid Blueprint for a New computing Infrastructure, Edited by Ian
Foster and Carl Kesselman, and the second volume too; (ii) Grid Comput-
ing Making the Global Infrastructure a Reality - Fram Berman , Anthony
J.G. Hey and Geoffrey C. Fox
[19] The Digital Divide - Facing a Crisis or Creating a Myth? - Edited by
Benjamin M. Compaine
Laboratory Courses
Laboratory Course on Silicon Sensors
Elisabetta Crescio, Marek Idzik
INFN, Torino, Italy
ABSTRACT
This laboratory course consists of five different mini sessions, in order to give
the student some hands-on experience on various aspects of silicon sensors and
related integrated electronics. The five experiments are:
• Characterization of silicon detectors (VCI measurements).
• Double light spot (measurement of the collection time of electrons and
holes separately versus voltage in a silicon diode).
• Understanding of the VA read-out chip operation, and measurement of its
noise versus detector capacitance characteristics.
• Measurement of the position resolution of a microstrip detector.
• Observation of spectra in a silicon diode.
1.1 Introduction
Near intrinsic n-type silicon with a metallised p-doped region is the most
frequently used semiconductor structure for detecting charged tracks in high-
energy physics experiments. A polarisation voltage is applied across the diode
structure, which depletes the silicon from charge carriers. Charged particles
or photons interacting with the silicon will create electron-hole pairs that drift
along the electric field lines to the contacts located on the silicon surface. A
schematic picture of a silicon sensor diode is shown in Figure 1.
A first step in constructing a particle detector based upon a silicon sensor
is to characterise the sensor without readout electronics attached. The static
characteristics of a sensor are usually adequate to determine if the sensor can
be used for particle detection. The leakage current behaviour as a function of
206 Elisabetta Crescio et al.
metal contact
charged particle −V
p−type implant
guard ring
n−type bulk
n+ type contact
electron GND
hole
voltage and the voltage needed to fully deplete the sensor are two important
parameters. The voltage needed to fully deplete the sensor can be determined
by measuring the capacitance between the diode implant and the backplane
of the sensor. In the final particle detector system, both the capacitance and
the leakage current will influence the performance of the readout electronics.
The capacitance and leakage current depend on the geometry of the sensor and
the quality of the material and manufactoring process. In a well controlled
and uniform process sensors with the same geometrical layout, processed on
the same substrate, should have the same behaviour. In reality, there may be
variation both in the process and in material and therefore there may be sensors
which differ largely from what we naively would expect. When constructing
an experiment consisting of many sensors we have to measure them in the
laboratory to find the good sensors that can be assembled into the experiment.
This session requires some knowledge of the basic principles of diodes and
will give experience handling unprotected diodes, operating the microscope and
probe manipulators. Because of the short time available we restrict ourselves
to simple DC-coupled diodes.
• vacuum pump;
• microscope;
In Figure 2 a schematic picture of the setup is shown.
CV-meter
probe
station
microscope
conductive
rubber
Vacuum
diode pump
Contribution (1) is generally well controlled and small giving a few nA/cm2 .
The contribution from (2) depends largely on the purity of the material since
recombination centres and trapping centres increase the creation of electron-
hole pairs. The magnitude is higher than that from (1), giving a few µA/cm2 .
The leakage current originating from thermal generation is of course tempera-
ture dependent. By lowering the temperature of the sensor we may reduce the
contribution. By decreasing the temperature by 10o C the leakage current will
typically be reduced to a third. In some cases the contribution from (3) may be
the dominant source of leakage current. The surface current may be caused by
effects on the non-depleted edge region or by a bad processing environment. The
leakage current originating from the surface may vary extensively from sensor
to sensor. To reduce the effects from surface current a guard ring structure is
processed on the silicon. The guard ring can be anything, from a single implant
around the diode to a complex structure of alternating implants and floating
metal rings around the silicon diode. We will now study the IV-characteristic
of a silicon diode sensor with 300 µ m thickness. Execute the steps below:
1. Place the silicon diode on the vacuum chuck with conductive rubber under
it.
2. Connect the diode with the probe needle to the negative pole of the battery
pack.
3. Connect the chuck (and thus, the silicon backplane) to the positive pole
of the battery.
4. Cover the probe station to prevent light from generating current in the
diodes, note down the voltage and current at 0 V.
5. Ramp up the voltage and note down the voltage and the corresponding
current.
inter pixel/strip capacitance do not change in the same way as the backplane
capacitance when applying reverse bias to the sensor. In this experiment, the
capacitance of the silicon diode is measured between the backplane and the
p-implant. Execute the steps below:
1. Find the capacitance C0 of the setup by leaving the circuit open.
2. Follow the procedure outlined for IV-measurements. Measure the full
capacitance (capacitance of the diode and capacitance of the setup), cal-
culate the capacitance of the diode by subtracting the value C0 from the
measured one, and note down the measurements.
102
I (pA)
C (pF)
2
10
10
10
0 10 20 30 40 50 60 70 80 0 10 20 30 40 50 60 70 80
Voltage (V) Voltage (V)
Figure 3: The current of the diode as a function of the applied voltage (left) and the
logarithm of the corrected capacitance vs. depletion voltage in V (right).
test_on
ckb shift_in_b
shift_out_b
holdb
Analogue bias analogue out
& adjustments
• possible resistance between the active element of the sensor and ground,
or the bias supply.
212 Elisabetta Crescio et al.
Holdb
T=1us
Shift_in_b
Ckb
Shift_out−b
Signal?
Analogue out
#1 #2 #3 #4 #5 #6 #127
First (=top) Last (=bottom)
channel channel
Time of
physics event
16-Jun-99 HARDCOPY
16-Jun-99 CHANNEL 1
10:22:45 output to
14:49:53 Input 2 Int. Printer
2 Normal 1 s Disk
GPIB
1 s H1BN
5.0 V R$ 232
Centronicos
2.00 V Coupling
2 D50 1 page feed
Grounded
1 DC1M 1 s
1 s Grounded
AC1M 100 mV
Off On
2.00 V protocol
V/div OFFset
Deskjet b/w
NORMAL
HP 7470
ECL TTL HP 7550
TIFF
1 Global BWL
TIFF compr.
Off On
( 30 MHz )
(a) (b)
Figure 6: (a)Screen shots showing hold and shift signals. (b)Screen shots showing
clocking and analogue output pulse.
The main sources of noise for silicon detector assemblies are: The Equiv-
alent Noise Charge (ENC) of the input FET of the pre-amplifiers proportional
to the load capacitance:
�
Ce 4kT
EN CF ET = (1)
q 3gm τ
Laboratory Course on Silicon Sensors 213
16-Jun-99 CHANNEL 1
TRIGGER SETUP 15:10:20
16-Jun-99 Input
Normal
15:02:08 Edge SMART
HIBU
Coupling
DC50
Grounded
1 1 Ocino
Grounded
2 s
trigger or
1 2 Ext
2 s AC1
clope 1
Pos Neg Global DNL
Window Off On
(38 MHz)
Probe Atten
holdoff x1
2 s x2
SHL OFF Time Evts x5
x10
x20
1.1 V DC 2 s BNL
2.2 V DC 1 C 0.60 V STOPPED
108 Ms/s 1.5 V AC
2.2 V OC 1 C 0.33 V STOPPED
100 Ms/s
(a) (b)
Figure 7: (a)Screen shots showing output waveform showing hits in two adjacent channels.
(b)Output from a single channel.
where C is the total load capacitance, Il is the leakage current, τ is the peaking
time of the amplifier, k the Boltzmann constant, T the absolute temperature in
Kelvin and R is parallel of the bias resistor and feedback resistor. The parameter
gm is the trasconductance of the MOS device. The total noise contribution is
the squared sum of the components listed above:
�
EN Ctotal = EN CF2 ET + EN Cl2 + EN CR
2 (4)
For the Viking chip the noise is typically 70 e− + 12 e− Cload [pF ]. It is obvious
that we want to keep the leakage current small by choosing a sensor that has
low leakage current at a voltage that fully depletes the sensor. The contribution
to the noise from the detector capacitance and series resistance will be studied
in this experiment.
214 Elisabetta Crescio et al.
4. NIM crate;
2.4 Measurement
In order to measure the noise, the setup has to be calibrated. This is done
by applying a known charge pulse Qcal on the input of the Viking chip and
measuring the output from the chip Voutl . The input charge is generated by
applying a known voltage pulse Vcal over a known calibration capacitor Ccal :
Ccal V cal
Qcal = (5)
e
where e is here the magnitude of the electron charge (e = 1.6 × 10−19 C. The
calibration capacitance Ccal is on our case 1.8 pF. The RMS of the noise Vnoise
from the Viking chip without test pulse in mV can be calculated using a modern
digital oscilloscope. Since the calibration is known, the measured value in mV
can be converted to ENC by the relation:
Vnoise Qcal
EN C = RM S e (6)
Vout
Six channels have been pre-wired with different load capacitances to the input
pads of the Viking chip. Determine the calibration and measure the correspond-
ing ENC noise for the channels, and note down the values.
1000
Noise RMS (e-)
χ2 / ndf 3017 / 4
900
p0 15.4 ± 0.7146
800
p1 63.89 ± 20
700
600
500
400
300
200
100
0 10 20 30 40 50
Capacitance (pF)
3.1 Introduction
Silicon microstrip sensors are the most commonly used device for high reso-
lution tracking in particle physics. The strip design allows a large sensitive area
with relatively few readout channels. The basic strip detector is read out on
one side giving information of the track position only in one dimension. Vari-
ous solutions to measure track position in two dimensions exist. The simplest
solution is to glue two single-sided sensors back-to-back, but a more demanding
design is to process strips on both sides of the sensor. In this experiment we
216 Elisabetta Crescio et al.
will study a single sided sensor which is illuminated by a pulsed laser source on
the strip side (because the back of the sensor is fully aluminised).
4. NIM crate;
5. Oscilloscope;
The silicon microstrip sensor is wire bonded to the Viking readout circuit, which
has been placed on a readout PCB (Printed Circuit Board). The assembly has
been mounted on a slide which can be precisely moved such that the translation
direction is orthogonal to the strips. An optical fibre has been mounted about
100 µ m away from the silicon surface. Figure 9 shows a photograph of the PCB
card and the detector mounted on the slide.
Figure 9: Picture of the PCB board and the detector mounted on the slide.
Laboratory Course on Silicon Sensors 217
3.3 Measurement
We will now try to determine the position resolution of the silicon microstrip
sensor. The peak of the light source is a few strips wide. We require the
information from a number of strips in order to accurately determine the peak
position. By moving the fibre closer to the silicon sensor the laser spot size
will reduce, but on the other hand we risk mechanically damaging the sensor.
Proceed as folows:
1. move the slide by turning the micrometer screw on the right hand side
of the box (seen from the repeater electronics board). You will now see
the signal from the light moving from one strip to another. You may also
notice that this translation is not very smooth. The reason for this is the
aluminium on top of the implanted strips which reflect a fraction of the
light and reduces the signal locally. This effect would be larger if we had
a better focused light spot.
3. Determine and write down the amplitudes of the channels in the peak by
moving the cursor on the oscilloscope.
5. Repeat 4. and 3.
Table 3.4 shows the measured amplitudes of the signal for different positions
of the sensor. Figure 10 shows the oscilloscope output without a laser spot (a),
and with the laser spot in two different positions (b and c).
One horizontal division corresponds to about two readout channels. Now
perform a gaussian fit for each position and calculate the peak positions. Table
3.4 shows the results from a previous measurement.
Since we are averaging the signals from the Viking in order to minimize
the electronics noise, and taking 10 points to determine the peak in the gaussian
fit, it can be seen that position accuracy down to a few µ m can be achieved.
What are the main error sources?
218 Elisabetta Crescio et al.
Strip n 0 µm 10 µ m 20 µ m 30 µ m 40 µ m 50 µ m
1 304 292 268 240 232 212
2 424 400 368 340 320 288
3 612 584 544 504 472 436
4 888 844 808 760 716 680
5 1030 1000 984 948 912 884
6 1150 1140 1130 1100 1080 1050
7 1140 1160 1170 1150 1130 1120
8 1010 1070 1100 1130 1160 1200
9 736 772 820 864 916 968
x1 x1 x1
x2 x2 x2
x5 x5 x5
1 x10 1 x10 1 x10
x20 x20 x20
2 !s 2.6 !s 2 !s 2.6 !s 2 !s 2.6 !s
1.2 V 50" 1.2 V 50" 1.2 V 50"
100 MS/s 100 MS/s 100 MS/s
2.5 V AC 2.5 V AC 2.5 V AC
2 DC -0.11 V 2 DC -0.11 V 2 DC -0.11 V
AUTO AUTO AUTO
Figure 10: Oscilloscope output without a laser spot (a), and with the laser spot in two
different positions (b and c).
0 µm 10 µ m 20 µ m 30 µ m 40 µ m 50 µ m
Calculated peak 264 274 282 292 302 312
Difference 10 8 10 10 10
Laboratory Course on Silicon Sensors 219
4.1 Introduction
In gaseous detectors the mobility for electrons is several orders of magni-
tude higher than for positive ions. In semiconductors the mobility for holes
is only slightly lower than for electrons. In general the signal propagation in
semiconductors is a few nanoseconds while the signal propagation in gaseous
detectors typically varies from microseconds to milliseconds. In order to study
the drifting of electrons and holes in silicon, very fast electronics is required.
The drift velocity for electrons and holes in silicon at low electric field strength
is given by:
ve = µe E (7)
vh = µh E (8)
where µe and µh are the mobilities for electrons and holes respectively, and E is
the electric field. The mobility in silicon at room temperature is 1350 cm2 /V s
for electrons, and 480 cm2 /V s for holes. At high field, the velocity saturates
with velocities of the order of 107 cm/s.
When the sensor is fully depleted, the signals from holes and electrons will
arrive almost at the same time. We can try to study the slower transit of holes
by shining a short laser pulse on the back side of a non-depleted n-type silicon
diode. By choosing a wavelength which does not penetrate far in the silicon
the charge can be generated close to the surface of the silicon. The holes have
now to drift to the other side of the sensor, while the electrons are formed at
the interface. If the electric field is low and the detector has a reasonably large
depleted region we will now see a difference in the total time between the signal
arising from electrons and holes. It is important to realise that the signal itself
arises immediately as the charge carriers start moving, and lasts until the last
charge carriers are collected.
By starting with a high depletion voltage the signal from the setup will have
the same shape as the light pulse from the laser which is shown in Figure 11.
The current signal which is induced on the diode is due to the movement
of both charge carriers in the electric field. The transit time for the carriers
can be calculated by integrating Equations 7 and , remembering that the field
varies as a function of position:
qND
E(x) = x + Emin (9)
�
where � is the dielectric constant, Emin is a function of the bias voltage, and is
220 Elisabetta Crescio et al.
4-Jun-99 MEASURE
10:30:31
A : Average (1) OFF Cursors
20 ns Parameters
176 mV mode
-573.4 mV Time
208 swps Amplitude
type
Relative
Absolute
show
Diff - Ref
1 Diff & Ref
20 ns
200 mV
-570 mV
Reference
cursor
Track OFF On
Difference
20 ns cursor
1 .2 V 50 t 7.3 ns 1 137 MHz
2 .2 V AC t 500 MS/s
3 .5 V 50 1 DC -0.344 V
4 .10 mV 50 NORMAL
Figure 11: The amplifier output from the laser pulse with a fully depleted silicon diode.
a minimum of zero for a barely depleted detector, then the transit times are:
� � w + (�/qN )E �
D min
th = ln (10)
µh ND x0 + (�/qND )Emin
� � x + (�/qN )E �
0 D min
te = ln (11)
µh ND (�/qND )Emin
where x0 is the depth of production, measured from the ohmic contact side.
The induced current is given by Ramos theorem, which states that the current
on the electrode of interest is equal to the charge value multiplied by the dot
product of the ”weighting field” and the charge velocity. The weighting field
is an hypothetical field calculated by putting unit potential on the electrode of
interest, zero on all other electrodes, and ignoring any static charges (ND in
our case). For a simple diode, this reduces to a constant 1/w, and the velocities
Laboratory Course on Silicon Sensors 221
are aligned with the electric field, so calculating the scalar velocities gives the
induced current:
1 � x0 qND ) �� � ND ��
ih = q µh Emin + exp µh q t (12)
w � �
1 � x0 qND ) �� � ND ��
ie = q µe Emin + exp − µe q t (13)
w � �
In fact these pulses become smeared by the amplifier bandwidth, and the re-
sulting pulse is what is seen on the oscilloscope.
Figure 12: Picture of the sensor setup.Top:box containing the detector with the two optic
fibers and cables for the detector polarization and read-out. Bottom: zoom on the detector
and electronics.
1. Oscilloscope;
6. NIM crate.
2-Jun-99 MEASURE
22:23:54
B : Average (A) Off Cursors
20 ns Parameters
49.5 mV mode
-107.26 mV
Time
139 swps
Amplitude
B : 1-4 type
20 ns Relative
44.0 mV Absolute
-105.5 mV
show
Diff - Ref
Diff & Ref
Reference
cursos
Track Off On
Dofference
cursor
20 ns
1. 20 mV 50 t -12.2 ns 1 -82.0 Mhz
2. 2 V AC t 500 MS/s
3. 5 V 50
Ext AC -240 mV 1M
4. 20 mV 50 NORMAL
Figure 13: The signal from an unbiased sensor (read out from the p-side) when laser is
shined on the p-side.
Turn on the sensor bias and ramp slowly up the voltage. The amplitude
of the signal gets larger but the shape stays approximately unchanged. The
sensor is depleted from the p-side and the n-side is conducting transporting the
electrons to the amplifier.
structure. In the double peak structure are visible the contribution coming from
fast arriving electrons and the contribution from holes with a lower mobility
which have to travel through the depleted region of the sensor.
4-Jun-99 MEASURE
10:36:27
A : Average (1) Off Cursors
20 ns Parameters
9.6 mV
-9.963 mV mode
451 swps 1 Time
Amplitude
type
Relative
Absolute
show
1 Diff - Ref
Diff & Ref
20 ns
10.0 mV
-8.55 mV
Reference
cursor
Track Off On
Difference
cursor
20 ns
1 10 mV 50 t 11.3 ns 1/!t 88.5 Mhz
2 .2 V AC 500 MS/s
3 5 V 50
4 10 mV 50 1 DC -30.6 mV
NORMAL
Figure 14: The signal amplitude at low bias voltage when laser is shined on n-side of the
silicon sensor (The signal is read out from the n-side).
The double peak structure disappears when the bias voltage gets higher.
At full depletion the amplitude of the signal is comparable to that obtained
when laser is shined on p and n sides (see Figure 15).
4-Jun-99 MEASURE
10:28:50
A : Average(1) Off Cursors
20 ns Parameters
44.0 mV mode
-161.95 mV Time
150 swps 1 Amplitude
type
Relative
Absolute
show
Diff – Ref
1 Absolute
20 ns
44.0 mV
-155.8 mV
Reference
cursor
Track Off On
Difference
cursor
20 ns
"t 13.4 ns 1/"t 74.6 Mhz
1 50 mV 50! 500 MS/s
2 .2 V AC 1 DC – 86 mV
3 5 V 50! NORMAL
4 10 mV 50!
Figure 15: The signal amplitude at full depletion when laser is shined on n-side (read out
from n-side).
The optical fibre cables are not labeled; can you discover, by
224 Elisabetta Crescio et al.
using one or the other cable, which cable points on the n-side, and
which on the p-side?
5.1 Introduction
During this experiment we will use a silicon sensor to see the spectrum form
a γ source and study its energy resolution.
1. Oscilloscope;
2. 241
Am source;
6. NIM crate.
Figure 16: Photograph of the setup. The sensor and the readout chip are mounted on a
printed circuit board which is placed into a aluminium box.
counts
90 241 Am
80
70
60
50
40
30
20
10
0
10 20 30 40 50 60 70 80
Channel
Figure 17: Spectrum from a 241 Am source. The source is placed under the printed circuit
board hosting the detector.
and Kβ lines for copper and silver are: Cu (Kα ) 8.03 keV, Cu (Kβ ) 8.90 keV,
Ag (Kα ) 21.99 keV and Ag (Kβ ) 24.94 keV.
Successively, other measurement are taken with the source placed into the
aluminium box on a “bridge” above the sensor. In this case only the peak cor-
responding to the 59.95 keV line is visible. In the following, the peaks obtained
reading out channels 12,36,58,78,102 are shown.
From Figures 18,19 and 20 it can be seen that the energy resolution de-
teriorates increasing the number of the readout pads, due to the increase of
the sensor capacitance. If it assumed that the energy resolution is limited by
226 Elisabetta Crescio et al.
counts
30
counts
graph12 graph12
Entries 9 25 Entries 10
Mean 54.15 Mean 54.77
25
RMS 1.323 RMS 1.998
20
20
15
channel 12 - 1 pad
15 channel 36 - 2 pads
10
10
5 5
0 0
40 45 50 55 60 65 70 75 80 40 45 50 55 60 65 70 75 80
Channel Channel
(a) (b)
Figure 18: Spectrum from a 241 Am source obtained from channel 12, with only one pad
connected (a), and from channel 36 with two pads connected (b).
counts
graph12
counts
graph12
25 Entries 10 11
25 Entries
Mean 54.74
RMS 1.556 Mean 54.78
RMS 2.463
20
20
15 15
channel 58 - 4 pads
channel 78 - 8 pads
10 10
5 5
0 0
40 45 50 55 60 65 70 75 80 40 45 50 55 60 65 70 75 80
Channel Channel
(a) (b)
Figure 19: Spectrum from a 241 Am source obtained from channel 58, with 4 pads connected
(a), and from channel 78 with 8 pads connected (b).
counts
graph12
25 Entries 16
Mean 56
RMS 2.741
20
10
0
40 45 50 55 60 65 70 75 80
Channel
Figure 20: Spectrum from a 241 Am source obtained from channel 102, with 21 pads con-
nected.
the electronic noise, then the noise can be calculated knowing the RMS of the
energy distribution:
RM S(eV )
N oise (RM S e−) = (14)
3.6eV
Laboratory Course on Silicon Sensors 227
where 3.6 eV represent the energy required to create an electron-hole pair from
and incident particle.