Department of Chemical Engineering School of Engineering and Architecture Saint Louis University Laboratory Report Evaluation Sheet
Department of Chemical Engineering School of Engineering and Architecture Saint Louis University Laboratory Report Evaluation Sheet
By
Feria, Gabriel V.
July 2017
3
ACKNOWLEDGEMENT
We are grateful to our dear instructor, Engineer Ray Limuel C. Laureano, who has guided us
We are also grateful to our laboratory technicians, who have provided us all the apparatus that
We are also grateful to our classmates for the support and assistance.
4
TABLE OF CONTENTS
TITLE PAGE 2
ACKNOWLEDGEMENT 3
TABLE OF CONTENTS 4
LIST OF TABLES 5
LIST OF FIGURES 6
ABSTRACT 7
Chapter 1: INTRODUCTION 9
REFERENCES 16
APPENDICES 17
5
LIST OF TABLES
6
LIST OF FIGURES
7
ABSTRACT
In order to control the flow of a pipe, flow must be measured. This experiment introduces
measurement of flow by using Orifice meter. Orifice meter is a device basically used for
measuring the rate of flow of a fluid through a pipe. It has two distinct uses. The first is to restrict
flow where high accuracy is generally not important and the second is flow measurement where
calculation accuracy is critical. The Orifice consists of flat circular plate which has circular sharp
edged hole called orifice, which is concentric with pipe. It is one half smaller in diameter of the
pipe. Based from the experiment, the velocity increases while flow rate increases. The head loss
Keywords: Orifice Meter, Coefficient of Discharge , Flow Rate, and Head Loss.
8
CHAPTER 1
INTRODUCTION
constriction in accord with the principle of mass continuity, while its static pressure must
decrease in accord with the principle of conservation of mechanical energy. Thus any gain in
kinetic energy a fluid may accrue due to its increased velocity through a constriction is balanced
by a drop in pressure.By measuring the change in pressure, the flow rate can be determined, as in
various flow measurement devices such as venturi meters, venturi nozzles and orifice plates.
An orifice meter is a conduit and a restriction to create a pressure drop. An hour glass is
a form of orifice. A nozzle, venturi or thin sharp edged orifice can be used as the flow restriction.
In order to use any of these devices for measurement it is necessary to empirically calibrate them
. That is, pass a known volume through the meter and note the reading in order to provide a stand
ard for measuring other quantities. Due to the ease of duplicating and the simple construction, th
e thin sharp edged orifice has been adopted as a standard and extensive calibration work has been
done so that it is widely accepted as a standard means of measuring fluids. Provided the standard
mechanics of construction are followed no further calibration is required. The minimum cross se
Orifice meter or orifice plate is a device used for measuring the rate of flow of a fluid
through a pipe. It works on the same principle as a venturimeter. It consists of a flat circular plate
which has a circular sharp edged hole called orifice, which is concentric with the pipe. The
orifice diameter is 0.5 times the diameter of the pipe. A differential manometer is connected at
section 1 which is at a distance of about 1.5 to 2 times the pipe diameter upstream from the
9
orifice plate, and at section 2, which is at a distance of about half the diameter of the orifice on
As the fluid approaches the orifice the pressure increases slightly and then drops
suddenly as the orifice is passed. It continues to drop until the “vena contracta” is reached and
pressure point is reached that will be lower than the pressure upstream of the orifice. The
decrease in pressure as the fluid passes thru the orifice is a result of the
The purpose of this experiment is to calibrate the Orifice Meter. The purpose of this
obtain data on pressure drop versus water flow rate and to be able to plot the calibration curves
for the meter on a rectangular coordinate paper in terms of pressure drop vs. flow rate.
10
CHAPTER 2
Experimental procedure:
In this experimental study, the group needed the assistance of the Laboratory Technician
as well as the Supervising Instructor. First, the group with the assistance of the technician
selected the pipe run containing the orifice meter by closing the isolating valves for all other
horizontal pipe runs. Then, the group studied the design and construction of the apparatus to
become thoroughly familiar with the method of operation. Then the group recorded the initial
weight of the tank 1 that is placed on the platform balance. Afterwards the group opened the
water valve but the group ensured that the discharge is located on the tank 2. And again, the
group recorded the monometer reading when the flow is already established. The group then,
transferred the discharge to the tank 1 and recorded the weight of the water collected for some
time. After this step, computation was made possible for the volumetric flow rate by using the
density of water. The group also computed the velocity at the pipe, the pressure drop and the
coefficient of discharge using the suitable equations. This procedural step was repeated for 5
11
Experimental Set-up
12
CHAPTER 3
sharp-edged orifice. Also, to obtain data on pressure drop versus water flow rate and to be able to
plot the calibration curves for the meter on a rectangular coordinate paper in terms of pressure
13
Head Loss Versus Flow Rate
3500
3000
2500
2000
1500
1000
500
0
0 100 200 300 400 500 600
Based on the data gathered, it was observed that increasing the velocity will cause an
increase in the manometer reading thereby, also increasing the head loss. This means that there is
In the case of coefficient of discharge however errors were encountered and hence, the
experimental data was not able to precisely and accurately prove the theoretical value of 0.61.
the closest data gathered was that of 160 gph with the discharge coefficient equivalent to 0.7402.
14
CHAPTER 4
The students conclude that orifice meter is somewhat similar to that venturi meter.
In an orifice, flow rates are set in which the trend is increasing. The head loss increases
with an increase in flow rate. The flow rates affect the velocity which has an increasing order.
The coefficients of discharge vary from one another. Its constants are found from 0.5463 up to
0.5691.
In a venturi meter, the head loss is nearly same value of that head loss in orifice meter. It
has also the same trend for head loss. The calculated velocity in this experiment has also near
values of that velocity of orifice. In terms of constant, the venturi is lesser by 0.1 with the same
The students then recommend that the presence of bubbles seen inside pipes should be
removed or at least wait for it to be collided by liquids. It is believed that this causes big
difference on the data with respect to flow rate. A keen eye from the group is also needed for
15
REFERENCES
2010 Daniel Measurement and Control, Inc. , " Fundamentals of Orifice Meter
http://www2.emersonprocess.com/siteadmincenter/PM%20Daniel%20Documents/Funda
16
APPENDICES
APPENDIX A
Definition of Apparatus
Fluid flow apparatus- is an apparatus used to determine the relationship between Fanning
Friction Fa tor and the Reynolds number and was used to understand the behavior of fluid in a
straight pipe.
Mercury manometer - consist of a column of liquid in a tube whose ends are exposed to
different pressures. The column will rise or fall until its weight (a force applied due to gravity) is
in equilibrium with the pressure differential between the two ends of the tube (a force applied
due to fluid pressure). A very simple version is a U-shaped tube half-full of liquid, one side of
which is connected to the region of interest while the reference pressure (which might be the
atmospheric pressure or a vacuum) is applied to the other. The difference in liquid level
has two important elements: (1) a temperature sensor (e.g. the bulb of a mercury-in-glass
thermometer) in which some physical change occurs with temperature, and (2) some means of
converting this physical change into a numerical value (e.g. the visible scale that is marked on a
17
regulate processes, in the study of weather, in medicine, and in scientific research.
Orifice meter - is a device used for measuring the rate of flow of a fluid through a pipe. It works
APPENDIX B
Definition of Terms
Head Loss - In any real moving fluid, energy is dissipated due to friction; turbulence dissipates
even more energy for high Reynolds number flows. This dissipation, called head loss, is divided
into two main categories, "major losses" associated with energy loss per length of pipe, and
Flow Rates - is the volume of fluid which passes per unit time; usually represented by the
symbol Q (sometimes V̇). The SI unit is m3/s (cubic metres per second). Another unit used is
18
APPENDIX C
Computations
Computation
𝑘𝑔
Density of mercury: 𝜌 = 13533.1248
𝑚3
𝑘𝑔
From handbook (Table 2-305) Water at 𝑇 = 24℃: 𝜌 = 997. 2503 𝑚3
𝜇 = 919.4356𝑃𝑎. 𝑠
𝑄 𝜋𝐷 2
VELOCITY: 𝑣 = 𝐴 ; 𝐴= ; 𝐷 = 16.92 𝑚
4
𝑔𝑎𝑙 3.7853 𝐿 1 𝑚3 1 ℎ𝑟
160 ( )(1000 𝐿)(3600 𝑠)
ℎ𝑟 1 𝑔𝑎𝑙
𝑣1 = = 0.7482 𝑚/𝑠
16.92
𝜋(10000)2
4
𝑔𝑎𝑙 3.7853 𝐿 1 𝑚3 1 ℎ𝑟
260 ( )(1000 𝐿 )(3600 𝑠)
ℎ𝑟 1 𝑔𝑎𝑙
𝑣2 = = 1.2159 𝑚/𝑠
16.92
𝜋(10000)2
4
𝑔𝑎𝑙 3.7853 𝐿 1 𝑚3 1 ℎ𝑟
360 ( )(1000 𝐿 )(3600 𝑠)
ℎ𝑟 1 𝑔𝑎𝑙
𝑣3 = = 1.6835 𝑚/𝑠
16.92 2
𝜋(10000)
4
𝑔𝑎𝑙 3.7853 𝐿 1 𝑚3 1 ℎ𝑟
460 ( )(1000 𝐿)(3600 𝑠)
ℎ𝑟 1 𝑔𝑎𝑙
𝑣4 = = 2.1511 𝑚/𝑠
16.92 2
𝜋(10000)
4
19
𝑔𝑎𝑙 3.7853 𝐿 1 𝑚3 1 ℎ𝑟
560 ( )(1000 𝐿 )(3600 𝑠)
ℎ𝑟 1 𝑔𝑎𝑙
𝑣5 = = 2.6189 𝑚/𝑠
16.92
𝜋(10000)2
4
𝑔
HEAD LOSS: ∆𝑃 = 𝑔 (𝜌𝐻𝑔 − 𝜌𝐻2 𝑂 )
𝑐
At 160 gph:
9.81
∆𝑃 = (13600 − 997. 2503)(4/1000) = 494.5319 Pa
1
At 260 gph:
9.81
∆𝑃 = (13600 − 997. 2503)(7.5/1000) = 927.2293Pa
1
At 360 gph:
9.81
∆𝑃 = (13600 − 997. 2503)(12/1000) = 2348.9809Pa
1
At 460 gph:
9.81
∆𝑃 = (13600 − 997. 2503)(19/1000) = 4203.5211 Pa
1
At 560 gph:
9.81
𝑃 = (13600 − 997. 2503)(26/1000) = 3214.3948Pa
1
𝐶𝑣 2𝑔𝑐(𝑃1−𝑃2)
VENTURI CONSTANT:𝑣 = √
𝐷2 𝜌
√1.( )^4
𝐷1
At 160 gph:
20
𝐶𝑣 2(1)(494.5390)
0.7482 = √ 𝐶𝑣 = 0.7401
√1 + (0.4148)^4 997. 2503
At 260 gph:
𝐶𝑣 2(1)(927.2293Pa)
1.2159 = √ 𝐶𝑣 = 0.8785
√1 + (0.4148)^4 997. 2503
At 360 gph:
𝐶𝑣 2(1)(2348.9809Pa)
1.6835 = √ 𝐶𝑣 = 0.9626
√1 + (0.4148)4 997. 2503
At 460 gph:
𝐶𝑣 2(1)(4203.5211 Pa)
2.1511 = √ 𝐶𝑣 = 0.9764
√1 + (0.4148)4 997. 2503
At 560 gph:
𝐶𝑣 2(1)(3214.3948Pa)
2.6189 = √ 𝐶𝑣 = 1.0162
√1 + (0.4148)^4 997. 2503
21