This document provides information on the GA-L reagent for measuring glycated albumin (GA) levels in serum and plasma on the Cobas 6000 analyzer. It describes the calibration method as a 2-point linear calibration using standards, with measurement wavelengths of 546nm and 700nm. Sample volume is 4.5 microliters and reaction time is 10 minutes.

Uploaded by

Elang Sudewa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
27 views2 pages

Jurnal Coomb Test

This document provides information on the GA-L reagent for measuring glycated albumin (GA) levels in serum and plasma on the Cobas 6000 analyzer. It describes the calibration method as a 2-point linear calibration using standards, with measurement wavelengths of 546nm and 700nm. Sample volume is 4.5 microliters and reaction time is 10 minutes.

Uploaded by

Elang Sudewa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

GA-L REAGENT GA

For measurement of % GA in serum and plasma

Cobas 6000

TEST NO. *
TEST NAME GA
TEST CODE * << CALIB >> << OTHERS >>
MODULE C CALIBRATION METHOD LINEAR STD (1) (CALIB. CODE) 901
SAMPLE TYPE Ser / Pl CALIBRATION POINT 2 (CONCENTRATION) 0
<< ANALYZE >> SPAN POINT 2 (RACK-POS.) *
WEIGHT 0 (VOLUME) 4.5
ASSAY 2 POINT END (CALIB. PRIORITY NUM.) 0.0
REACTION TIME 10 0
MEASUREMENT POINT (1) 34 UPDATE TYPE NONE
MEASUREMENT POINT (2) 70 (POINT)
MEASUREMENT POINT (3) 0 STD (2) (CALIB. CODE) *1)
MEASUREMENT POINT (4) 0 SD LIMIT (CONCENTRATION) *2)
999
(RACK-POS.) *
DUPLICATE LIMIT (%) 99
WAVELENGTH (PRIMARY) 546 (VOLUME) 4.5
DUPLICATE LIMIT (ABS) 32000
WAVELENGTH (SECONDARY) 700 SENSITIVITY LIMIT (LOW) -99999 0.0
SENSITIVITY LIMIT (HIGH) 99999 0
S. VOLUME (NORMAL) 4.5 S1 ABS (LOW) -32000
0.0 S1 ABS (HIGH) 32000 STD (3) (CALIB. CODE)
0 AUTO MASKING NO
S. VOLUME (DECREASE) 3.0 (CONCENTRATION)
0.0 (RACK-POS.)
CHANGE OVER (TO CASSETTE) CANCEL (VOLUME) 10.0
0 AUTO CALIB. TIMEOUT YES
S. VOLUME (INCREASE) 6.0 0.0
CASSETTE CALIB. (TO 0
0.0 CASSETTE) CANCEL
0 (TIME OUT) CANCEL
(DAYS) STD (4) (CALIB. CODE)
CASSETTE CODE * (CONCENTRATION)
CASSETTE EXP. DAYS 99 AUTO CALIB. QC VIOLATION NO (RACK-POS.)
(METHOD) BLANK (VOLUME) 10.0
R. VOLUME (R1) 180 (RULE) 1s 0.0
(DIL) 0 (CONTROL 1) 0
(CONTROL 2)
R. VOLUME (R2) 0 (CONTROL 3)
(DIL) 0 STD (5) (CALIB. CODE)
(CONCENTRATION)
R. VOLUME (R3) << RANGE >> (RACK-POS.)
45
(DIL) 0 UNIT g/dL (VOLUME) 10.0
REPORT NAME GLYCAT 0.0
LINEARITY LIMIT (%) 0 ED 0
0 ALBUMI
(ABS) 32 N STD (6) (CALIB. CODE)
32000 (CONCENTRATION)
PROZONE LIMIT -32000 DATA MODE ACTIVE (RACK-POS.)
32000 AUTOMATIC RERUN YES (VOLUME) 10.0
( POINT) TECHNICAL LIMIT (LOW) -99999 0.0
TECHNICAL LIMIT (HIGH) 999999 0
INSIDE REPEAT LIMIT (LOW) -99999
(Minimum rate of Total) 0 REPEAT LIMIT (HIGH) 999999
(Minimum margin rate) 0 CONTROL INTERVAL 0 *1) Fill with code calib of GA-
ABSORBANCE LIMIT 32000 L 0 L Calibrator
INCREASE H 0 *2) Input the GA value of the
CELL DETERGENT DETERGENT 1 I 0 GA-L Calibrator
STIRRING LEVEL 1
GA-L REAGENT ALB
For measurement of % GA in serum and plasma

Cobas 6000
TEST NO. *
TEST NAME ALB
TEST CODE * << CALIB >> << OTHERS >>
MODULE C CALIBRATION METHOD LINEAR STD (1) (CALIB. CODE) 901
SAMPLE TYPE Ser / Pl CALIBRATION POINT 2 (CONCENTRATION) 0
<< ANALYZE >> SPAN POINT 2 (RACK-POS.) *
WEIGHT 0 (VOLUME) 2.0
ASSAY 2 POINT END (CALIB. PRIORITY NUM.) 0.0
REACTION TIME 10 0
MEASUREMENT POINT (1) 34 UPDATE TYPE NONE
MEASUREMENT POINT (2) 70 (POINT)
MEASUREMENT POINT (3) 0 STD (2) (CALIB. CODE) *1)
MEASUREMENT POINT (4) 0 SD LIMIT (CONCENTRATION) *2)
999
(RACK-POS.) *
DUPLICATE LIMIT (%) 99
WAVELENGTH (PRIMARY) 600 (VOLUME) 2.0
DUPLICATE LIMIT (ABS) 32000
WAVELENGTH (SECONDARY) 660 SENSITIVITY LIMIT (LOW) -99999 0.0
SENSITIVITY LIMIT (HIGH) 99999 0
S. VOLUME (NORMAL) 2.0 S1 ABS (LOW) -32000
0.0 S1 ABS (HIGH) 32000 STD (3) (CALIB. CODE)
0 AUTO MASKING NO
S. VOLUME (DECREASE) 1.5 (CONCENTRATION)
0.0 (RACK-POS.)
CHANGE OVER (TO CASSETTE) CANCEL (VOLUME) 10.0
0 AUTO CALIB. TIMEOUT YES
S. VOLUME (INCREASE) 3.0 0.0
CASSETTE CALIB. (TO 0
0.0 CASSETTE) CANCEL
0 (TIME OUT) CANCEL
(DAYS) STD (4) (CALIB. CODE)
CASSETTE CODE * (CONCENTRATION)
CASSETTE EXP. DAYS 99 AUTO CALIB. QC VIOLATION NO (RACK-POS.)
(METHOD) BLANK (VOLUME) 10.0
R. VOLUME (R1) 150 (RULE) 1s 0.0
(DIL) 0 (CONTROL 1) 0
(CONTROL 2)
R. VOLUME (R2) 0 (CONTROL 3)
(DIL) 0 STD (5) (CALIB. CODE)
(CONCENTRATION)
<< RANGE >> (RACK-POS.)
R. VOLUME (R3) 75
(DIL) 0 UNIT g/dL (VOLUME) 10.0
REPORT NAME ALBUMI 0.0
LINEARITY LIMIT (%) 0 N 0
0
(ABS) 32 DATA MODE ACTIVE STD (6) (CALIB. CODE)
32000 AUTOMATIC RERUN YES (CONCENTRATION)
PROZONE LIMIT -32000 TECHNICAL LIMIT (LOW) -99999 (RACK-POS.)
32000 TECHNICAL LIMIT (HIGH) 999999 (VOLUME) 10.0
( POINT) REPEAT LIMIT (LOW) -99999 0.0
REPEAT LIMIT (HIGH) 999999 0
INSIDE CONTROL INTERVAL 0
(Minimum rate of Total) 0 L 0
(Minimum margin rate) 0 H 0 *1) Fill with code calib of GA-
ABSORBANCE LIMIT 32000 I 0 L Calibrator
INCREASE *2) Input the Albumin value of
CELL DETERGENT DETERGENT 1 the GA-L Calibrator
STIRRING LEVEL 1