This document provides instrument settings and parameters for performing a protein quantification assay using the biuret method and spectrophotometry. Key details include using a wavelength of 540 nm, a sample volume of 4 μL, and reagent volumes of 200 μL. The dynamic range is 0 to 150 g/L protein concentration. A two-point linear calibration is specified using assigned standard values and factors to calculate concentration from absorbance readings.

Uploaded by

Enrique Duarte
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
53 views

Prot

This document provides instrument settings and parameters for performing a protein quantification assay using the biuret method and spectrophotometry. Key details include using a wavelength of 540 nm, a sample volume of 4 μL, and reagent volumes of 200 μL. The dynamic range is 0 to 150 g/L protein concentration. A two-point linear calibration is specified using assigned standard values and factors to calculate concentration from absorbance readings.

Uploaded by

Enrique Duarte
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

PROTEIN

Spectrophotometric
BIURET

The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the
instrument. It is advisable to validate the use when there is any change in software or reagent versions.

Instrument: AU 400 / 640


Reagent preparation
Reagent is ready to be used.

Instrument settings

Specific Test Parameters


Sample: Volume 4 µL Dilution 0 µL
Reagents: R1 Volume 200 µL Dilution 0 µL Reagent OD limit:
Wavelength: Pri. 540 Sec. First L -0.5 First H 0.1
Method: END Last L -0.5 Last H 0.1
Reaction slope: + Dynamic Range:
Measuring Point 1: First 0 Last 27 L 0 H 150
Measuring Point 2: First 0 Last 0 Correlation Factor:
Linearity: % A 1 B 0
No Lag Time: On-board stability period:

L H
7. None selected
8. Out of range 65 80
L H
Panic Value Unit: g/L Decimal places: 0

Calibration Specific

Calibration Type: AB Formula: Y=AX+B Counts: 2 Process: CONC.

Cal. Nº OD CONC Factor/OD-L Factor/OD-H


Point 1: # * -9999 9999

* Enter the assigned standard value


# User defined

Vers. 0206