Manual Durometro Digital
Manual Durometro Digital
Instruction Manual
Figure 1.1
z Unique structure. Adopting special indenter structure, See figure 1.1, it can be used not only
to conduct the general test which can be made by traditional hardness tester, but also to test
surfaces such as the inner surface of annular and pipe parts, which not possible to be tested
by common hardness tester.
z Wide measurement range. It can test following 15 hardness scales: HRA, HRB, HRC, HRD,
HRE, HRF, HRG, HRH, HRK, HRL, HRM, HRP, HRR, HRS, HRV,
z Powerful auxiliary functions. Besides general hardness testing, the tester has been provided
with following auxiliary functions: upper and lower limit setting and warning; data statistics of
mean value, standard deviation, maximum and minimum; scales conversion: the tested
results can be converted into values of HB, HV, HLD and σb (strength); Correction for the
effect of testing on cylindrical surface and spherical surface.
z LCD display with backlight.
2 Specifications
3.1 Configuration
Standard delivery:
Main body 1
Hardness test block, scale B 1
Hardness test block, scale C 1
Steel Ball indenter (1/16″) 1
Spare balls (1/16″) 5
Diamond cone indenter 1
Mounting screws for indenter 2
Screwdriver 1
Large Vee anvil 1
Power supply cable 1
Dust cover 1
Protection sheath of indenter 1
Instruction manual 1
TIME Certificate 1
Warranty card 1
Optional accessories:
Short diamond indenter (TH3FJ01)
Chisel diamond indenter (TH3FJ02)
Slender diamond indenter (TH3FJ03)
1/8″ ball indenter (TH3FJ04) and spare ball
1/4″ ball indenter (TH3FJ05) and spare ball
1/2″ ball indenter (TH3FJ06) and spare ball
Micro printer (TA220S)
Communication cable
Flat anvil φ225 (TH3FJ07)
Flat anvil φ150 (TH3FJ08)
Vee/point anvilφ70 (TH3FJ09)
Vee/Flat anvil (TH3FJ10)
Assistant joist (TH3FJ11)
Assistant support (TH3FJ12)
Hardness reference blocks for other scales
Key sound setting +/- — upper/ lower limitation setting Σ — data statistics
HR= — scale conversion Φ — curved surface correcting MENU — main menu
▲、▼ — rolling direction ENTER — confirm
Figure 3.2
1- interface RS232, 2- power supply switch 3-220V/110V switch 4- power socket
Figure 3.3
4.1 Refer to figure 4.1a, b. The tester should be mounted on a stationary base platform with no
vibration. The bearing capacity of the platform should be more than 150kgs. Three Φ12mm holes
should be prepared on the platform. To ensure the free elevating of the leadscrew, the distance
between the center of front holes and the margin of platform should be less than 152mm; otherwise, a
Φ70mm annular hole should be prepared at the place 187mm away from front holes. The limitation of
the leadscrew moving downward is 250mm under bench surface, and no obstacle should exist within
this range. If there are walls or other obstacles behind the tester, a clearance of 70mm at least should
be preserved to make the cables connecting or extracting easily.
4.2 To unpack the tester, run the hand wheel anticlockwise, remove the rubber cushion. 4.3 Place the
tester on the prepared platform. Turn the hand wheel in anti clockwise to lower the anvil and remove
the rubber pad. Then place a level with the precision of 0.2mm/m on the anvil. Adjust the foundation
screws of tester to make the levelness to be 1mm/m and lock the nuts. Tightly fix the tester to the base
platform using M10 screws. Refers to figure 4.1.
4.4 Dismantle the bolt fixing the top cover, lift up the front part of the cover a little, pull out the cable
and then remove the cover complately.
4.5 There are 4 weights with marks in all for this tester. The 4 marks can be seen in figure 4.1. open
the side door of the tester, remove all of the rubber protection belts. Then start installing the weights in
the order of 1, 4, 3, 2, in which the weight 1 is hung in the hanging groove of underside lever and the
weights 4,3,2 are all hung on the upside lever, please refer to the figure 4.2 for weights 1, 2, and figure
4.3 for weights 3, 4. Do make sure they are hung in the groove.
4.6 Connect the cable which is pull out in step 4.4, put on the cover and fit on the bolt.
4.7 select the power supply of 220V or 110V according to the local power, connect the cable properly.
The primary set of power is 220V before first using.
Figure 4.2a Figure 4.2.b
5 Operation
Figure 5.1.3
Pay attention that the indenter selected according to the prompt must be mounted physically on the
indenter base. When installing the indenter, make sure the installing surface is clean, without foreign
matters, and tightly fixed by screws. The installation method is shown as figure 5.1.3.
Examine to ensure the test force selected is proper. The selection is performed by the switch handle
on the right top of the tester, which include 60kgf(588.4N), 100kgf(980.7), 150kgf(1471N).
The test can be performed directly according to the procedure 5.4 if none of the parameters above
need to be modified. Following procedures should be observed if the modification is necessary:
Figure 5.2.
The manual of this tester can be switched between Chinese and English language by pressing “▲”
or “▼”keys
5.3 Setting of testing parameters
Press “MENU” key, figure 5.3.1 (Rockwell) or 5.3.2 (Rockwell superficial) will be shown. It is ready
to set the parameters.
Figure 5.3.1
Press “▲” or “▼”key to move the cursor to the scale which will be selected, and press “ENTER”
to finish the selection. Now the screen will show what is in figure 5.3.3 according to new selected
scales, which is showing which kind of indenter and testing force to use.
Press key “ENTER” to get back to figure 5.3.1
5.3.2 Set dwell loading
Dwell loading refers to the duration of the total loading test force (i.e. time of the pre-load and the
additional load actuating simultaneously). For samples without continuous deformation with time after
the additional load has been loaded, the additional load is normally set as 1~3s. For samples with
continuous slowly deformation with time after the additional load has been loaded, the dwell loading is
normally set as 6~8s. For samples with obvious deformation with time after the additional load has
been loaded, the dwell loading is normally set as 20~25s.
Press “▲” or “▼”key to move the cursor to “Dwell loading” in figure 5.3.1 or 5.3.2 and
press ”ENTER” key. Figure 5.3.4 will be shown on the screen.
Press “▲” or “▼”key to reset dwell loading (1s ~ 50s). Press the “ENTER” key to complete the
selection. The display will return to figure 5.3.1
Figure 5.3.6
The tester is provided with a series-digital-communication-port RS232 (baud rate 9600bps), which can
be connected with a printer or external computer. Results of each test will be printed or transmitted to
an external computer in real time.
The cable of printer or computer should be connected prior to using. Press “▲” or “▼”key to move
the cursor to “RS232” in figure 5.3.1. Press ”ENTER” key. Then, figure 5.3.6 will be showed. Press
“▲” or “▼”key to move the cursor to “printer”, “comms” or “off”, and then press “ENTER” key to
complete the selection. The display will return to figure 5.3.1
5.3.5 Return
After the setting was finished completely, press “MENU” to return to figure 5.1.1
5.4 Testing
5.4.1 Preload
Place the sample to be tested on the anvil. Smoothly rotate the hand wheel in clockwise to uplift the
anvil until the sample has been contacted with indenter. The schematic diagram of indenter moving
course will be showed on screen. Refer to 5.4.1, the last symbol is the end position of the pre-load.
Rotate the hand wheel slowly and smoothly until the cursor in the figure reaches the end position.
Buzzing warning will be given out at the same time and the rotating should be stopped at once.
Test results and precision would not be affected by the slight over-rotation of the hand wheel. In case
of excessive rotation occurs, The tester will automatically warn and give prompt. See figure 5.4.2.
Redo is necessary.
Figure 5.4.3
5.4.3 Unload
Rotate the hand wheel in anti-clockwise to low the anvil, and the test force will be unloaded completely.
The display will return to figure 5.1.1. And all test parameters will be stored for the next test.
Repress “+/-”key to cancel the function, and the indicating light will go off.
The value of serial number n and N will be showed simultaneously with the display of the test result
each time after this function is set. As shown in figure 5.5.5, 10 tests will be performed totally, current
test is the sixth, and the hardness value tested is 55.0HRC. In case of n=N, that is to say the last test
has been completed, the tester will automatically calculate the mean value X , standard deviation S,
maximum value Max, minimum value Min and range R after unloading by turning the hand wheel
anti-clockwise. The display is shown as figure 5.5.6.
Figure 5.5.6
Of which, the mean value, standard deviation and range are calculated according to the following
equations:
X= 1
N ∑X i ,
1
S=
N −1
∑ ( X − X i )2
R=Max-Min
Press “ENTER” key to return to figure 5.1.1. The statistics results can be output automatically if the
printer or computer is connected (Refer to 5.3.4)
Normally, the serial number will plus 1 after each test until the n is equal to N (i.e. all of the N tests
were completed). Then the statistical calculation begins. In case that the current test need to be
cancelled due to some reasons, move the cursor to “×” as figure 5.5.5. Then unload (i.e. lower the
anvil) and redo. In this case, 1 will not be added to n, and the current test will not take part in the
statistical calculation, and the result will not be printed out or output.
Repress “Σ” key, the function will be cancelled and the indicating light will go off.
For instance, if we intend to convert the HRC value of current scale into HLD, the work piece is
stainless steal, the procedures are as following: Press“HR=”key once. The indicating light in top-left
corner of the key is illuminating and figure 5.5.7 will be displayed. Press “▲” or “▼”key to move
the cursor to "HRC—HLD stainless ”, and then press “ENTER” to confirm. The display will return to
figure 5.1.1.
The value will be showed simultaneously with the displaying of hardness value each time after the
function is set into effect. As shown in figure 5.5.8, the hardness value is 28.0HRC, and the Leeb
hardness after converting is 556Mpa. Each conversion is only performed within the applicable scope
which conversion is possible. “E” will be showed in case of the converting value beyond the scope. As
shown in figure 5.5.9, the hardness measured is 65.5HRC. The converted value HB is beyond the
available scope, thus “E” is displayed.
Repress “HR=” key, the function is cancelled and the indicating light will go off.
Figure 5.5.9
Figure 5.5.10
The measuring results should be corrected if the surface of sample is the convex surface of cylindrical
part or spherical surface. press “Ф” key once, the indicating light in top-left corner of the key is
illuminating and figure 5.5.10 will be displayed on screen. Press “▲” or “▼”key to move the cursor
to select correcting mode, then press “ENTER” to confirm.
In case of cylindrical surface correction, figure 5.5.11 will be displayed on screen. Press “▲” or“▼”
to determine the curvature radius (applicable scope is 3~19mm) note 1. Then, press “ENTER” to return
to figure 5.1.1. In case of spherical surface correcting, figure 5.5.12 will be displayed on screen. Press
“▲” or“▼” to determine the diameter of sphere (applicable scope is 4~25mm) note 2. Then press
“ENTER” to return to figure 5.1.1.
Note 1, note 2:
In the ISO6508-2, Metallic Materials - Rockwell Hardness Test - Part 2: Verification and Calibration of
testing machines, only cylindrical surface correction for scale A, C, D, B, F, G, N and T; and spherical
surface correction for scale C, as well as the scope are regulated. However, there is no regulation for
the values and scales beyond the scope. Therefore, the tester will not correct the values beyond the
scope, either. Users should consider if it is necessary to correct them.
The measuring result as well as the correction value will be given in the course of each measurement
after the function setting has been taken into effect. Note that the screen will display respectively
testing value (the direct testing result without correcting) and correction value. As shown in figure
5.5.13, the measuring result is 38.5HRC and the correction value is “+1.5” HRC.
Figure 5.5.13
Repress the “Ф” key to cancel the function, and the indicating light will go off.
The four auxiliary functions can be all selected simultaneously. In this case, the measurement result is
as 5.5.14:
Figure 5.5.14
6.1 The tester will perform self-checking, display the type of tester, and determine if the motor is at
original position when powered on. If the anvil is not put down, the display will prompt to lower the
anvil after self-checking. Figure 5.1.1 will be displayed if everything is in good condition.
6.2 If there is not any display after powering on, Check whether the power supply switch is on,
whether the position of voltage selection switch on back panel is proper, whether the fuse of power
supply is in good condition and is installed properly, whether the transformer output plug at the inner
side of back panel is connected properly with the power supply socket board, and whether the three
fuses on power supply socket board had been fused.
6.3 If the display on LCD is abnormal, check whether the position of voltage selection switch on back
panel is proper, or uplift the canopy to check whether the connecting of the cables for LCD is perfect.
6.4 If a roar sound of locked motor is given out after powering on, Shut down the power immediately
and contact with the supplier.
6.5 If there is no diagram indication on screen when rotate the hand wheel to load, check whether
the test was begun with the status of main menu as figure 5.1.1.
6.6 The light on the key is not illuminating or the key operation is failure. Check to determine whether
the connection of cables is perfect. Or contact with the supplier.
6.7 In case of other malfunctions, no hesitate to contact with the supplier.
7.1 The weights and indenter must be removed when transporting the hardness tester. And a piece of
shockproof rubber pad must be placed between the indenter base and anvil. The original package
should be packed prior to a remote transport.
7.2 When perform any adjusting and checking such as loading/unloading the weights, removing the
canopy, plugging/ extracting cables or opening the side door for checking, the power supply should be
cut off for safety.
7.3 The changing of test force is forbidden under the status of indenter contacting with the sample
measured.
7.4 When installing/uninstalling the indenter, much attention should be paid to ensure the tip to avoid
damage and contaminating. And the mounting surface should be clean without any foreign matters.
The indenter should be removed to store carefully in case of idle for a long term, and rust protection
measures are necessary.
7.5 The surface of the anvil and Hardness reference block should be clean without contamination and
scratch, scoring and bruising. Coating of rust protection oil is necessary when store.
7.6 The sample measured must be placed steadily when testing. And the supporting should be
reliable to ensure no movement or distortion during the course of testing.
7.7 Dustproof and corrosive medium prevention should be considered in daily operating environment.
Regular rust prevention measures should be adopted in moist area.
7.8 Regular lubrication should be carried out for the leadscrew. Steps in detail: remove the anvil and
the protection ring. Loose the leadscrew protection sleeve. Drip some light oil to the leadscrew. Rotate
hand wheel to elevate the leadscrew again and again to distribute the lubrication oil uniformly. Finally
reassemble the parts such as the protection sleeve. Note no much lubrication oil should be dripped.
7.9 The regular verification and calibration of hardness tester should be performed according to the
relevant standard and rule of state.
7.10 Not to dismantle or adjust the parts and components that are used for fastening and assembling
by yourself when abnormal condition happens. Please contact with the supplier.
8. Non-warranty parts
Indentor
Appendix 1: Test principle of Rockwell hardness
For Rockwell hardness measurement, the displacement is measured actually. And the displacement
obtained under the specified testing condition will be converted into hardness value. The detail
procedures are as following. Refer to attached diagram 1.
F0 F1 F0
F0
h2-h0
h2
h1
h0
Attached diagram 1
(1)First, load test force F0 to press the indenter into the sample surface, and detect the original
displacement h0.
(2)Then, load the additional load F1 and hold for a period. The displacement of indenter is h1 at this
time. Thereafter, unload F1.
(3)Maintain the pre-load F0, and measure the indenter displacement at this moment h2.
(4)Calculate the Rockwell hardness value according to the following equations (displacement unit:
mm):
Rockwell hardness has 15 types of scales which applicable for materials with different hardness.
Scale depends on the combination of indenter and test force. The corresponding relations and
application scopes are as following list:
Hardness
Scale Indenter Test force Application examples
scope
Carbide alloy, , superficial case
HRA 588.4N/60kgf 20-88
hardened steel
Diamond Medium case hardened steel,
HRD 980.7N/100kgf 40-77
indenter pearlite forgeable iron and etc.
Quenched steel, hardened and
HRC 1471N/150kgf 20-70 tempered steel, harden cast steel
and etc.
Anneal copper alloy, mild sheet plate
HRF 588.4N/60kgf 60-100
Steel ball alloy
indenter
Copper alloy, mild steel, aluminum
HRB Diameter: 980.7N/100kgf 20-100
alloy
1.5875mm
(1/16”) Forgeable iron, copper-nickel-zinc
HRG 1471N/150kgf 30-94
alloy
The minimum thickness of sample depends on the material hardness and scale used. There should
be no visual distortion on the back of sample after measurement.
Sample minimum thickness, material hardness and scale used refer to attached diagram 2 and
attached diagram 3.
1 .6
1 .4
1 .2
1 .0
0 .8
0 .6
0 .4
H R C
H R D
H R A
2 .2
2 .0
1 .8
1 .6
1 .4
1 .2
1 .0
0 .8
H R H
H R E
H R F
H R K
H R G
H R B
Attached diagram 3: Test with steel ball indenter(HRB, HRE, HRF, HRG, HRH and HRK)
In general, TH300 hardness tester had been strictly calibrated prior to leaving factory, thus the
adjusting by user is not necessary and forbidden. However, it can be adjusted according to following
method if minute error occurs in operation due to some reasons. Following rules should be especially
observed when adjusting:
1.2 Verification should be performed according to the testing methods in GB/T 230 or GB/T 1818,
and five valid impression should be made on each standard block to measure its hardness. (six time
will be conducted for pressing, of which, the first impression will be omitted and the mean value of
later five times will be recorded)
1.3 The reading errorδ of hardness tester will be calculated according to equation (1):
δ = HR − HR …………………………(1)
of which: HR ⎯ the arithmetic mean value of five hardness measurements;
HR ⎯ the standard value of standard Rockwell hardness block.
1.4 Repetitiveness E will be calculated according to equation (2):
b = HRmax − HRmin …………………………(2)
of which: HRmax ⎯ the maximum in five hardness measurements;
HRmin ⎯ the minimum in five hardness measurements;
1.5 The reading error δ of hardness tester and repetitiveness E should conform to the regulations in
table 2 of 5.6.
Hardness tester had been checked and calibrated prior to leaving factory, and it can only delivery if
passed. However, some error may occur due to the remote transport, the difference of standard
hardness block itself, uniformity of standard block, class difference of standard block and other factors,
i.e. the reading inaccuracy in popular words. What frequently occurs is good repetitiveness, however,
minute deviation that can be adjusted is present between reading and standard block value. Specific
method is introduced as following (taking HRC as example):
* If problems occur, which scale occurs deviation should be determined firstly. For example: HRC
scale occurs deviation, the differential between testing results and standard block value is over the
tolerant scope.
* Determine the hardness scope section which deviation occurs; take each 10 unit as a hardness
scope section. For example: HRC scale occurs deviation; if the nominal value of standard block is
63.5, testing result is about 65, then the hardness scope section deviation occurs is 60~70.
* Determine correcting value. Measurements should be performed by six times with standard block in
this hardness scope section. The procedures should be in accordance with the testing method in GB/T
230; six valid impressions will be made on each standard block; then measure and record the
hardness values. The value of first impression will be omitted and the mean value of later five
impressions is required. For example: the nominal value of standard block is 63.5HRC, the mean
value of five measurements is 65.2HRC, thus the correcting value should be “-1.7”. Note: the standard
block should be moved on the surface of sample bench, no uplift; six points should be evenly
distributed on the standard block; any distance between impressions, the distance between
impression and standard block margin should be not less than 2mm.
* Set out to correct. Under the initial menu status, continuously press “ENTER”, “ENTER”, “▲” and
“ENTER” key in turn; picture of figure 1 displaying at this moment indicates that correcting status has
entered.
Figure 1
* Press “▲” and “▼” key to move cursor to the hardness scale needs to correct. For example: cursor
has been moved to “HRC”, then press “ENTER” to confirm, and picture of figure 2 will be showed.
Figure 2
* Press “▲” and “▼” key to choose the hardness scope section needs to correct. For example: press
“▲” until “50-60” has changed into “60-70”, then press “ENTER” to confirm. Cursor will be at “+” at this
moment.
* Press “▲” and “▼” key to choose “+” or “-“. For example: if the correcting value is “-1.7”, then
“-“ should be selected, press “ENTER” to confirm. Cursor will be at “0.5” at this moment.
* Press “▲” and “▼” key until “0.5” has been changed into “1.7”, then press “ENTER” to confirm.
* Press “MENU”, screen will back to figure 1; other selection can be made to correct other scales.
* If no correcting is necessary for other scales, press “MENU” to exit correcting status; return to main
menu.
* The correcting method for any scale and hardness scope section is as the same as this example.
* The correcting value of any scale and hardness scope section should be recorded for requisition
later.