100% found this document useful (1 vote)
226 views

2 - Optima CT520

The document describes the Lilyum mammography system. It provides technical details on the system's features including its high frequency X-ray generator capable of 20-35 kV, 1-640 mAs, a perfectly counterbalanced C-arm, and automatic exposure control device. The system aims to provide excellent imaging, ergonomics, and patient comfort through innovations like its compression system and software. Standard accessories include an 18x24 cm Potter Bucky and optional items include a 24x30 cm Potter Bucky or automatic filter interchange.

Uploaded by

Vishnu0049
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
226 views

2 - Optima CT520

The document describes the Lilyum mammography system. It provides technical details on the system's features including its high frequency X-ray generator capable of 20-35 kV, 1-640 mAs, a perfectly counterbalanced C-arm, and automatic exposure control device. The system aims to provide excellent imaging, ergonomics, and patient comfort through innovations like its compression system and software. Standard accessories include an 18x24 cm Potter Bucky and optional items include a 24x30 cm Potter Bucky or automatic filter interchange.

Uploaded by

Vishnu0049
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

Lilyum

www.metaltronica.com
Lilyum
A FLOWER TO DEFEAT BREAST CANCER!

Excellent imaging technology combined with a


modern, ergonomic and winning design for highest patient
comfort.
New features to improve operation efficiency such as:

- C-Arm integrated patient and operator hand rail

- Control panel positioning possible on either side of the unit

- Fully balanced C-Arm with every table size

- User programmable projections views

H. F. GENERATOR
X-Ray Generator data:

- High Frequency: 100 kHz ripple

- Power: 5 kW

- kV Range: 20 - 35 kV (steps of 0.5 kV)

- mAs range: 1 - 640 mAs

The LILYUM in standard version is supplied with a X-Ray tube with


Molybdenum anode and Molybdenum filter.
TUBE AND FILTERS

Different X-Ray tubes can be chosen:

- with bimetallic anode (Mo and W)

- with biangular anode (10° and 16°)


- high speed tube (9.000 rpm)

Besides, optionally is available an automatic interchangeable filter


(Molybdenum 30 μm/Rhodium 30 μm).

Operators will appreciate the brightness and efficient design of


the LILYUM’s LCD graphic display on which are indicated the exposure
parameters and the following data:
- “C” Arm angle rotation
- Effectively applied compression force
DISPLAYS

- Compressed breast thickness


- Average Glandular Dose
- HU level of tube assembly

An auxiliary display placed frontally on the bottom of


the unit is available on request. It shows the following data:
- C-Arm angle rotation
- Effectively applied compression force
- Compressed breast thickness

www.metaltronica.com
Lilyum
Patients will appreciate the new compression system “μPress” allowing

“μPress” SYSTEM
a correct and at the same time more comfortable compression as the descending
compression paddle speed is in inverse relation to the compression force applied.

Higher patient comfort is also achieved by newly designed compression


paddles with softly curved and smooth lines available in different formats.

Really innovative is
the LILYUM’s intelligent and
ULTRA SOFTWARE

microprocessor controlled automatic


exposure device enhanced by the
new calibration software ULTRA.

This standard feature makes the


mammography system’s installation
incredibly rapid and simple.
O.D. linearity exceeds QA protocols. Film and CR operation also guaranteed by
means of special AEC characteristics. ULTRA includes state of the art diagnostic
functions and data logger for easy service.

LATERAL VIEW FRONT VIEW


1231

788

DIMENSIONS AND WEIGHT


max 2103 (C-arm)
1998,5
max 1651,5
min 946,5

max 1310
min 605

677 696

Weight: 250 kg

www.metaltronica.com
Lilyum
• High frequency X-Ray generator:
- 5 kW power, 100 kHz ripple
- 20 ÷ 35 kV (0,5 kV step)
- 1 ÷ 640 mAs
• “C” Arm perfectly counterbalanced with each Potter-Bucky format:
- motorized up/down movement
TECHNICAL FEATURES

- manual rotation with electromagnetic brake (+/- 180°)


• X-Ray tube, molybdenum anode, dual focal spot 0.1 and 0.3 mm
• Fixed focus film distance 65 cm
• Motor-driven (with friction) and manual compression device with safety release
• Adjustable compression force
• Interactive digital control panel with graphic display showing exposure parameters, for
messages to the operator in different languages to be selected during installation
• Additional display on unit bottom with digital indication of angle rotation, effectively applied
compression force and compressed breast thickness
• AGD (Average Glandular Dose) measurement
• Automatic Exposure Control (AEC) device, fitted with triple field solid state detector
selectable from control panel, with the following possibilities:
- fully automatic “ZERO POINT” technique (automatic kV, automatic mAs)
- semi-automatic “ONE POINT” technique (manual kV, automatic mAs)
- manual “TWO POINTS” technique (manual kV, manual mAs)
- 11 levels of radiographic density
• Three serial ports for printing device, calibration, last 1300 exposure parameters memory

METALTRONICA reserves all rights to improve any characteristics at any product without notice
transfer

STANDARD ACCESSORIES
• 18x24 cm Potter Bucky with carbon fiber grid, inclusive of marking device of projections on
film (interchangeable with 24x30 cm Potter Bucky or with other image receptors)
• 18x24 cm collimation plate
• Shifted compression paddle, for 18x24 cm format
• ULTRA, software for advanced automatic exposure control calibration with a maximum of
16 selectable film-screen combinations fully programmable (3 dedicated to CR cassettes)
• One pair of compression foot pedal switch
• X-ray control push-button with extensible cable
• Phantom for complete daily check of the mammography system
• PC cable

• 24x30 cm Potter Bucky with carbon fiber grid, inclusive of marking device of projections on
film (interchangeable with 18x24 cm Potter Bucky or with other image receptors)
OPTIONAL ACCESSORIES

• FILTROMAMM, automatic filter interchangeable Rh/Mo (motorized movement)


• Bimetallic Anode X-Ray tube (double material anode)
• Biangular Anode X-Ray tube
• High speed starter
• 18x24/24x30 cm automatic collimator
• Collimation plate for magnification
• Compression paddle for magnification
• Shifted compression paddle for 24x30 cm format
• Shifted compression paddle for axillary examen
• Device for geometric magnification complete of cassette holder without grid and with
magnification ratio continuously variable
• Lexan screen for protection of the X-Ray field from parts not under examination
• Second pair of compression foot/pedal
• Anti-X glass shield for operator protection

Via della Pisana, 431 - 00163 Rome - Italy


LILYUM DPT/LILYUM/003/U

Phone +39-06-66160206 - Fax +39-06-66160357

www.metaltronica.com