0% found this document useful (0 votes)
59 views

Short Notes On PEC

The document discusses the principles and parameters of Pulsed Eddy Current (PEC) testing. PEC uses electromagnetic induction to measure the thickness of ferromagnetic materials like carbon steel. It works by inducing eddy currents in the material when a magnetic field is pulsed. Factors that affect the PEC signal include wall thickness, lift-off distance, material properties, and interferences from insulation, corrosion, or reinforcement materials between the probe and steel surface. The signal decay rate and transition point are used to determine wall thickness.

Uploaded by

Gaurav Chopra
Copyright
© © All Rights Reserved
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
59 views

Short Notes On PEC

The document discusses the principles and parameters of Pulsed Eddy Current (PEC) testing. PEC uses electromagnetic induction to measure the thickness of ferromagnetic materials like carbon steel. It works by inducing eddy currents in the material when a magnetic field is pulsed. Factors that affect the PEC signal include wall thickness, lift-off distance, material properties, and interferences from insulation, corrosion, or reinforcement materials between the probe and steel surface. The signal decay rate and transition point are used to determine wall thickness.

Uploaded by

Gaurav Chopra
Copyright
© © All Rights Reserved
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 6

PEC

 PEC scope limited to ferromagnetic materials. E.g. - carbon steel and low alloyed steel.
 Ferromagnetic material is material with magnetic permeability of µr»1
 PEC method based on Lenz’s law of electromagnetic induction
 Lenz law- an electrically conductive object when subjected to change in magnetic field will produce
eddy current that oppose change in magnetic field.
 ‘Lift-off’- Distance between probe and steel surface.
 After primary field has been switched off, eddy current decay and at same time diffuse into the steel.
 Change in sec. field inside receiver coil induces voltage over its terminal, this voltage as function of
time is called ‘PEC signal’.
 Point between straight section and curved section in PEC signal is called ‘transition point’. Occurs
when PEC signal reaches back wall of steel.
 ‘Amplitude’(A)- signal strength at the beginning of PEC signal
 ‘Slope’ (n) - rate of decay at the straight section. Slope parameter n≈1.5; n>1.5
 ‘Transition Time (τ) – onset of fast decay.
 PEC signal is displayed on log-log graph.
 𝑊𝑇 = √𝜏 ÷ 𝜎𝜇, 𝜎- electrical conductivity, 𝜇- permeability of steel is electromagnetic properties of
steel.
 The velocity of diffusion is higher for a thin wall than for a thick wall.
 If the wall thickness at reference spot is unknown the readings can be presented as percentage relative
to reference measurement and are ‘relative wall thickness reading.’
 If the wall thickness at reference spot is known then it can be expressed in mm (inches) and wall
readings are ‘absolute.’

 Above equation valid only if σ1µ1=σ ref µref.

 Electromagnet properties may vary across test specimen if one part of test sample has received heat
treatment different than other (e.g. furnace tube).
 Variation of electromagnetic properties limits accuracy of PEC, to typically ±10% of wall thickness.
 PEC does not show internal or external wall loss, it is only sensitive only to wall thickness.
 Time scale is expressed in milliseconds (ms).
 Limiter line method- for the determination of transition time 𝜏 ref
 Starting point at normalisation time is 30% below the reference signal, with amplitude of 0.7V ref.
 Normalisation means multiplying the PEC signal with a factor.
 The strength of the primary field is largest when the PEC sensor is closest to the surface.
 Lower lift-off- higher amplitude and vice versa.
 Real PEC signals are lift-off dependence.
 Degree of lift-off dependence varies with wall thickness and geometry.
 Typical dependence varies from 0.1% decrease to 0.25% increase in PEC reading per mm lift-off.
 PEC readings vary strongly with distance.
Insulation covers
 Material between lift-off can affect PEC signals if they are electrically conductive or magnetic.
Aluminium covers
 Electrical conductivity of aluminium 5 times higher than steel. Therefore strong contribution of
aluminium covers.
 Aluminium covers results in early tail in time in the PEC signal.
 Strength of early tail depends on the thickness of insulation cover.
 If less than 0.2mm it will hardly influence
 If 1mm thick then will influence signal.
 The effect is stronger where there is an insulation overlap.
Steel reinforcement bars
 Inside concrete, have same effect as aluminium covers.
 Strength of early tail depends on the density of the reinforcement mesh.
 For both aluminium covers and steel bars, the normalisation time should be chosen after the early tail.
 For Al covers, the norm time should be chosen when effect of Al has become weaker and signal is
straight again. It is after 6-8ms.
 Prolonged and strong tail- for thick Al sheet over lap and dense reinforcement.
S.S covers
 Electrical conductivity of Stainless steel 5 times less than steel. Are weakly magnetic.
 No noticeable effect on PEC signal.
Galvanised insulation sheeting
 Made of carbon steel.
 Magnetic and good electrical conductor.
 Distorts PEC signal
Chicken wire mesh
 No influence to signal if made of S.S.
 No influence if made of carbon steel but if it is properly fixed that it does not move.
 Influence signal if magnetic and free to move i.e. loose.
Corrosion product
 Ferrous oxide (Fe2O3) - highly magnetic but do not conduct electric current.
 Magnetised by primary field but no eddy currents are generated.
 Instead field decays very slowly
 Corrosion product results tail very late in time.
 Fe3+ (Fe2O3) are magnetic, Fe2+ (FeS) are non magnetic.
 Corrosion product thinner than 5mm can be ignored.
 Stand off – distance between bottom of PEC probe and corrosion product.
 Used for suppressing the influence of corrosion product by placing plastic shims in between.
 Contribution of corrosion product strongly reduced by higher stand offs (at least 20mm).
 Insulation components ensures this stand offs.
 Such case observed when inspecting annular ring of tank floor from outside.
 Corrosion product on far side of the steel has no effect on the signal
 PEC signal are sensitive to corrosion thickness layer above 5mm, ensure a standoff of at least 20mm or
if the insulation thickness exceeds 40mm.
 ADC sample time at least 600ms if corrosion product more than 5mm and lift-off less than 40mm.
Clipping or Saturation
 ADC has max input amplitude.
 ADC output is max for a signal with max signal input amplitude.
 ADC output will not increase if signal input increased above the max output. It will remain max
 This is referred as clipping.
 PEC signal cannot be used before clipping time.
 Gain of amplified signal inside probe adjusts such clipping occurs around 1.5ms
High frequency noise
 Always superimposed on PEC signal.
 This noise comes from amplifiers, environment and steel itself.
 High frequency noise can be reduced by HF filters.
50/60 pick up
 PEC probes picks up mains-related signals.
 This pick comes from electromagnetic fields from power cables, electric motors of pumps or other
electrical equipment.
 Frequency 50Hz in Europe and 60Hz in USA.
 Pick up is often accompanied in harmonics of 100, 150, 200Hz etc, for 50Hz and 120Hz, 180Hz,etc for
60Hz
 Filter used in software that uses frequency analysis (Fourier filter) to remove 50/60 and higher
harmonics.
 Fourier filter suppress mains related noise provided it is symmetrical.

Probe movement
 Movement causes erratic signal that is superimposed on the PEC signal.
 PEC signal verified for signal distortion if PEC measurement differ by more than ±10% from the
average over the other PEC readings.
Criteria for an acceptable PEC signal
 A straight and a curved section after smoothing and filtering.
 Amplitude of distortion is less than the signal amplitude at intersection point.
 Amplitude of PEC signal at norm point is within a factor 4 of the amplitude of that of the reference
signal.
 Repeatability of signal should be preferably within 1%.
Footprint
 Twice of the insulation thickness and steel thickness.
 PEC readings are averages over footprint.
 Time of onset of fast decay, i.e. the transition time, is measure of thickness of test specimen.

Quantitative model for a PEC signal


 Electric resistance experienced by eddy current decreases as eddy current diffuse downwards.
 Gradual decrease of eddy current is due to gradual increase of vertical cross section of eddy current
path.

 Signal Amplitude A depends on:


a) Properties of transmitter coil: diameter, no. of turns and configuration.
b) Excitation of current in transmitter coil
c) Properties of receiver coil: diameter, no. of turns and configuration.
d) Electronics (amplification factor, coil resistance)
e) Lift-off
 (a), (c) & (d) are constant, unless probe breaks down.
 Excitation current is monitored.
 Variation in amplitude (A) is due to lift-off

 PEC wall thickness readings slightly affected by lift-off. Requires lift-off correction.
 PEC footprint size increases with lift-off.

 Slope parameter n depends on:


a) Electromagnetic properties of steel, i.e. σ & µ.
b) Geometry of test specimen (e.g. curvature of pipe).
c) Sensor lift-off.
d) Probe configuration.

 Range of n varies small, 1.4≤n≤1.75, nominal value of n=1.5

 Two PEC signal, difference is slope parameter:


a) Not recorded on same type of steel.
b) Not at the same lift-off.
c) Not at same geometry, e.g. different piece of steel.

 The electromagnetic properties also depend on steel grade and can vary from one batch of steel to next.
 PEC measurement also depends on temperature of steel, because electromagnetic properties σ & µ also
depends on temperature.
 Wall thickness reading decrease by 0.1% per ˚C. This can be ignored for wall thickness mapping but
need to be corrected for corrosion mapping.

 Deviant readings at insulation overlaps, separately calibrate insulation overlaps with normalisation late
in time.
.
Corrosion product
 Corrosion i.e. iron oxide, poor electrical conductor.
 Does not conduct eddy current.
 Contributes to signal if between probe and steel, due to magnetic permeability of rust.
 Effect is a ‘tail’ on the PEC signal.
 Apparent transition point increased due to it.
 This results in increase in PEC wall thickness reading.
 Magnitude of increase depends on:
a) Thickness of corrosion product.
b) Distance of probe to corrosion product.
 Sampling Time- PEC signal sampled during a pre set time
 Sampling frequency- ADC records PEC signal at a pre set frequency, typically 20Hz, i.e. one ADC
sample every 20µs(micro second)
 Measurement Repeats- referred to as current pulses. Each repeat results in one PEC signal, which can
be added to produce one PEC signal, which is sent to computer.
 Cycle Delay- time between end of one cycle and beginning of next cycle.

 PEC signal from trailing edge (when the current is off) is recorded. Signal from trailing
discarded as its not stable though

INSTRUMENT PROBES
 Four standard PEC probes:
1. CP (contact probe)
2. AP-50 (probe with coils without ferrite, ‘Air probe’)
3. FP-50 (probe with ferrite inside the coils)
4. TB (for tank bottom inspection)

 Selection of probe determined by ‘Lift-off’.


 Preferred to keep lift-off as constant as practical,
1. Within about ±10mm if no lift-off correction can be applied; and
2. Within about ±30mm if lift-off correction can be applied.

 Substantial piece of steel, such as support or nozzle, affects PEC readings if PEC probe is closer than
twice lift-off to the inspection plate.

Notes:
1. The criteria for an acceptable PEC signal are:
a) A straight and a curved section of the PEC signal can be observed after filtering and
Smoothing;
b) No signal clipping after 3ms;
c) The amplitude of any distortions of the PEC signal is less than the signal amplitude at the
Intersection point;
d) The amplitude of the PEC signal at the normalization point is within a factor 4 of the amplitude
Of that of the reference signal, this is automatically verified by the PEC software.

You might also like