Prisma Guide
Prisma Guide
Conventional UT
Phased Array
TOFD
Table of contents
1 Introduction to your new device ........................................................................................................................4
2 Inserting and charging the battery for the first time .........................................................................................5
3 Booting the unit ..................................................................................................................................................6
4 Connecting probes and USB memory stick ........................................................................................................7
5 Load an existing configuration ...........................................................................................................................9
6 Create a configuration ......................................................................................................................................11
7 Starting data acquisition ...................................................................................................................................13
8 Understanding the conventional UT screen layout ..........................................................................................15
9 Navigating through menus and views ..............................................................................................................18
10 Keypad ..............................................................................................................................................................19
10.1 Keypad definition table ..........................................................................................................................20
11 Adjusting gain ...................................................................................................................................................21
12 Automatic 80% gain adjustment ......................................................................................................................22
13 Calibration wizards (for conventional UT) ........................................................................................................23
14 TOFD option......................................................................................................................................................25
14.1 Creating a TOFD setup ...........................................................................................................................25
14.2 Calibrating a TOFD setup .......................................................................................................................27
14.3 Understanding what you look at ...........................................................................................................29
15 Phased array Option .........................................................................................................................................30
15.1 Understanding at the phased array screen layout ................................................................................30
15.2 Moving the A-scan Extractor cursor ......................................................................................................33
15.3 Moving the Top/End view box ...............................................................................................................33
15.4 Adjusting the focus (focal laws) .............................................................................................................33
15.5 Calibration wizards (for phased array)...................................................................................................34
16 Manipulating cursors (Cartesian, Angular, Gate, Extractor Box, and Extractor) ..............................................35
17 Adding cursors ..................................................................................................................................................36
18 Making measurements with gates ...................................................................................................................37
19 Customizing the Measurements bar ................................................................................................................38
20 Taking screenshots and building reports ..........................................................................................................40
21 Defining scan plan geometry (cartesian/axis reference system) .....................................................................41
21.1 Wedge references ..................................................................................................................................42
21.2 Group reference.....................................................................................................................................42
21.3 Part datum .............................................................................................................................................43
22 Setting up position-encoded or time-encoded scans.......................................................................................44
23 Setting up an encoded scan..............................................................................................................................45
24 Selecting a layout with appropriate views .......................................................................................................46
25 Menu tree (menu items vary according to the inspection setup)....................................................................47
26 Ruler and axis color map ..................................................................................................................................49
27 Power LED Behavior .........................................................................................................................................50
28 Color palettes ...................................................................................................................................................51
29 Features not covered in this document ...........................................................................................................52
30 Encoder Connector Pin out...............................................................................................................................53
31 I/O Connector Pin out .......................................................................................................................................54
32 Definitions ........................................................................................................................................................55
32.1 Scan types ..............................................................................................................................................55
32.1.1 Sectorial scan ........................................................................................................................................... 55
32.1.2 Linear scan ............................................................................................................................................... 55
32.1.3 Mono scan (conventional UT) .................................................................................................................. 55
32.1.4 TOFD scan................................................................................................................................................. 55
33 Copyright, disclaimers, and certifications ........................................................................................................56
33.1 Copyright ................................................................................................................................................56
33.2 Disclaimer of liability ..............................................................................................................................56
33.3 Certification (electromagnetic compatibility) ........................................................................................57
V1.1 3
1 Introduction to your new device
Before starting, make sure that your Prisma kit includes the following:
Make sure you have the appropriate probes, wedges, and encoder for the type of inspection you intend
to perform. You may also use the Prisma demo kit, which includes a basic set of accessories necessary to
complete the scenarios described in this document.
V1.1 4
2 Inserting and charging the battery for the first time
Unscrew the back door and insert the battery. Remove the waterproof rubber cover over the AC input
socket and connect the AC power adaptor. Before using it for the first time, the battery must be charged
for at least 6 hours; however, the unit can be used while charging.
An optional external battery charger and additional battery packs can be added in order to reduce down-
time and increase productivity.
V1.1 5
3 Booting the unit
V1.1 6
4 Connecting probes and USB memory stick
4. Drawing: UT connectors
In order to connect phased array probes, unscrew and remove the grey protective cap that covers the
I-PEX connector located on the right-hand side of the Prisma. The cap can be stored on the unit by
screwing it on the back panel of the unit, over the CE mark.
IMPORTANT
When a phased array probe is not connected to the unit, the connector protective cap must be
reinstalled in order to prevent water and dust ingress.
If the phased array probe does not have an I-PEX connector, Sonatest can supply an appropriate
adaptor.
V1.1 7
Your Prisma will be equipped with either BNC or Lemo 1 connectors for conventional UT and TOFD
probes. Sonatest can similarly supply appropriate cables and adaptors to suit probes with other types of
connectors.
You may use either, or both, channels A and B for conventional UT or TOFD. For pulse-echo scans,
connect the probe to the Tx/Rx connector, labelled “T” on the case. For pitch-catch scans, connect the
transmitter probe to the Tx/Rx “T” connector and the receiver probe to the Rx “R” connector.
2. Insert a USB memory stick into USB port 1. A miniature USB memory stick is supplied, which allows the
cover to close while in use.
Three USB connectors and an Ethernet connector are located on the side of the unit; protected by the
waterproof cover. The USB ports can be used to connect external memory devices, a keyboard, a mouse,
or a USB hub should you need to connect additional devices. The Ethernet port can be used for remote
session support and to drive an external display via a VNC connection. If a USB memory device is
connected to the Prisma then the user may opt to save configuration (.utcfg) and data (.utdata) files to it.
Otherwise, by default and in the absence of an external memory device, these will be saved to the 5-GB
internal memory.
V1.1 8
5 Load an existing configuration
1. Press the key to open the startup screen where you can select a user-defined or an example
configuration file.
V1.1 9
3. Press the button to toggle through the available memory location options.
If you have connected a USB memory device, the list will include the following:
a. The Internal partition is where you can save your own files.
7. Select the appropriate file in the list using the Click Wheel .
V1.1 10
6 Create a configuration
The following is the procedure to create a new configuration with the help of the New Inspection Wisard.
1. To create a new configuration, press the key to open the startup screen.
2. Use the Click Wheel to select Start New Inspection Wizard and press .
3. Follow the instructions appearing on screen to set the parameters one by one.
4. Select the required inspection type (conventional, TOFD, or phased array) and press on the
keypad.
V1.1 11
The number of steps required to complete a configuration depends on the option selected. A typical
sequence of steps is as follows:
V1.1 12
7 Starting data acquisition
The Prisma software simplifies your workflow by using three operational modes:
• Configuration
• Acquisition/Recording
• Analysis
These modes are controlled using the STOP , PLAY , and REC keys. The current operation
mode is always displayed at the top-right corner of the screen by the and icons.
1. After opening a configuration file (.utcfg), the Prisma always starts in Configuration mode, also referred
to as the STOP mode. Before starting the data acquisition it is good practice to review your setup.
In Configuration mode the Prisma displays the following screen:
1. Menu
2. Configuration summary view
3. 3D view
4. Contextual help view
V1.1 13
2. Press the key to select the Menu bar then use the Click Wheel to navigate.
V1.1 14
8 Understanding the conventional UT screen layout
It is important to understand the components of the screen layout. This section guides you through the
main elements of the Acquisition or PLAY mode.
1. Menu
2. A-scan view
3. C-scan View
4. B-scan view
A number of alternate screen layouts can be selected from the View menu
V1.1 15
16. Drawing: UT analysis mode (component)
1. Measurements bar
2. Calibration indicators
3. Instrument state (Acquisition, Configuration, Recording, Analysis) and battery status
4. Scroll bar
5. View layout (current in green)
6. Encoder position
7. View measures
The division of the screen layout into areas is called a “layout”. Different layout options are available
depending on the inspection type. To change the layout press to select the Menu bar on the left
of the screen and use the key to select the View tab, use the Click Wheel to select Select
Layout, press and use the Click Wheel again to select the appropriate layout.
In the image above, the layout comprises one A-scan and one B-scan. This is the result of an encoded scan
of a 0° dual crystal probe over a step block. At the bottom, the colored spots represent the amplitude of
the underlying A-scan sample. The X-axis is the encoder scan axis. The Y-axis is the depth into the test
item.
Each view has its own cross-hair cursors, which can be disabled from the View tab. In drawing 16 above
the A-scan view has one gate and one Cartesian cursor (cross-hair). The B-scan view has two Cartesian
cursors. The Cartesian cursor location will report the depth (D) and true surface distance offset (SD). If
you move the cursor over a red region in the B-scan it will provide you the exact location of the flaw in
the part, regardless of the angle at which the peak signal occurs. All measurements are referenced from
V1.1 16
the wedge reference (Wedge Reference) point. The wedge reference location depends on the type of
wedge you are you using. The wedge reference is represented by the red dot in the 3D and scan plan
views displayed in Configuration mode. For more details see the “Defining scan plan geometry
(cartesian/axis reference system)” section.
V1.1 17
9 Navigating through menus and views
1. Press to select the Menu bar, and use the Click Wheel to navigate.
2. Press to edit a parameter and press again to confirm the new value.
OR
Press the key to cancel making a change.
3. Press to toggle between the different views.
4. Press to toggle between the different cursors of the selected view, then use the Click Wheel to
move the selected cursor (displayed in blue).
V1.1 18
10 Keypad
The keypad comprises the following areas. Details of the functionality of each button are provided in Table 2
1. Click wheel
2. Main controls to access menus, views and cursors
3. Alphanumerical keypad and shortcuts
4. Mode selection: PLAY/PAUSE, STOP, REC
5. Files save and load, use these buttons to:
a. LOAD
i. Create a new configuration (setup)
ii. Open a configuration (setup, .utcfg file)
iii. Open a report (PDF)
iv. Open a data file (.utdata file)
b. SAVE
i. Save configuration (setup, .utcfg file)
ii. Save screen capture image
iii. Create and save Inspection report
V1.1 19
10.1 Keypad definition table
CANCEL Escape Cancels an ongoing value change and resets to the previous value. Exit from a popup window.
Rotate your finger clockwise and counter clockwise to scroll up and down through a menu, to
Click Wheel --- increase/decrease a parameter value, to move cursor in view area or zoom in/out in 3D view.
Also acts as LEFT/RIGHT key to move during text entry.
UP / DOWN Moves from one menu item to another. Changes a parameter value. Moves up and down the
Arrows
ARROWS selected cursor or gate.
LEFT / RIGHT
Arrows Moves from one menu tab to another. Moves the selected cursor or gate to left and right.
ARROWS
VIEW V Toggles between the various views. Display the Measurements bar (in Live mode)
Toggles between the various cursors of the currently selected view. Also acts as Backspace
CURSOR C
during text entry.
Changes the gain value of the selected scan. If pressed and held for 3 seconds, it will
dB D
automatically adjust the gain to 80% of the selected gate (if RefAmplitude is set to 80%).
WELD W Shows or hides the weld overlay for the selected view.
F1 / HELP F1 Access the Contextual Help, Quick Start Guide and User Guide Documentation.
F2 / INFO F2 Displays the view information menu or for wizards returns to previous step or page.
Resets the encoders position to the origin, or for wizards advances to the next step or page. .
F3 / SET REF F3
Starts the recording mode permitting data to be saved on completion or termination of the
RECORD R
scan.
ALARM1 &2 --- Illuminated when any configured gates are triggered
ON/OFF
Powers the instrument on (Hold down) or shuts it down. Lightis green when the device is ON.
& ---
Refer to the section 27 “Power LED Behaviour”
Battery indicator
V1.1 20
11 Adjusting gain
To adjust the gain, press the key to open the gain items and set the gain of the current scan.
Note: If you have multiple scans (only in UT), you should select the appropriate scan by pressing
before pressing the key.
V1.1 21
12 Automatic 80% gain adjustment
1. Press the gate key , move the gate over the peak to be maximized using the Click Wheel
arrows , , , and
2. Press and hold the key for 3 seconds and it will automatically adjust the gain to the value of the
Ref Amplitude parameter of the selected gate, which is 80% by default.
V1.1 22
13 Calibration wizards (for conventional UT)
The Calibration tab displays the available calibration wizards. The menu items on this tab are displayed in
the order in which the calibrations should normally be performed. If you are using a multi-scan setup
each scan must be calibrated independently. The wizards will facilitate the calibration of the unit taking
into account the reflector type, type of calibration block, user supplied parameters, and the type of
probe/wedge used.
Type Description
Velocity & Zero Guides you through a calibration accounting for material
velocity and probe zero.
Wizard...
TCG/DAC/DGS Guides you through the setup of TCG, DAC, or DGS curves.
Wizard...
Encoder Guides you through the calibration of the encoder resolution.
Wizard...
Clear Calibrations Resets one or more calibrations.
V1.1 23
Figures 19 and 20 show the examples of the Velocity & Zero and TCG Wizards.
V1.1 24
14 TOFD option
It is possible to either load an existing configuration or use the New Inspection Wizard to create and
calibrate a new TOFD setup.
This section shows how the New Inspection Wizard is used to create a new configuration.
1. Press the key to open the startup screen and, using the Click Wheel , select Start New
V1.1 25
22. Drawing: New Inspection Wizard (step2)
the parameters have been entered press to move to the next step.
5. Configure the probe by adjusting the parameters and press once complete.
6. Configure the wedge the same way.
V1.1 26
7. In the Geometry step, the wizard will propose a PCS derived from the values previously entered. Set the
probe to this exact distance (exit point to exit point).
8. In the Path step the wizard shows a list of parameters computed from information entered previously.
Modify the Gain, the Start Path, and the Range.
9. Make sure the lateral wave, the back wall, and the converted mode are visible. Make adjustments
accordingly and then press to move to the next step.
10. Configure the encoder by appropriately adjusting the parameters and press once done.
11. Modify the measurement type if desired (measurement using hyperbolic cursors is automatically
Having set the PCS, it remains to calibrate the TOFD setup by adjusting the speed of sound and the ruler
position.
1. Move to the “Scan” tab with the or arrow keys.
2. Select the “TOFD” category by pressing the or , or rotate your finger around the Click Wheel
5. Make sure that the Ruler BW Pos item is set so that the blue cursor in the A-scan is aligned to the first
negative peak of the backwall echo.
If it is not aligned, try adjusting the material velocity on the Part menu, the zero on the Scan menu, or
the probe spacing on the scanner.
V1.1 28
14.3 Understanding what you look at
The following instructions will guide you through the essential elements of the Acquisition or PLAY mode
.
1. A-scan view
2. TOFD view
The division of the screen layout into areas is called a “layout”. Different layout options are available
appropriate to each inspection type.
To change the layout:
1. Press to activate the Menu bar on the left of the screen and use the key to select the View
tab.
2. Use the Click Wheel to select Select Layout, press , and use the Click Wheel again to select
the appropriate layout.
In the image above, the layout comprises one A-scan view and one TOFD view (the B-scan is called the
TOFD view when making TOFD inspections). This is the result of an encoded scan. The grey is 0% FSH, the
white is +100% FSH and the black is -100% FSH.
The orange cursors in the TOFD view are called “Hyperbolic” cursors. The double cross-hair cursor
indicates which A-scan is showed in the A-scan view. Changing the extractor angle in the TOFD scan view
will therefore change the A-scan signal displayed in the right-side view.
There are several other functions which perform lateral wave straightening and lateral wave removal
These are explained in detail in the “User Guide”.
V1.1 29
15 Phased array Option
It is important to understand the components of the screen layout. This section guides you through the
main elements of the phased array Acquisition mode.
V1.1 30
27. Drawing: phased array analysis and acquisition (component)
The division of the screen layout into areas is called a “layout”. Different layout options are available
appropriate to each inspection type.
To change the layout:
1. Press to activate the Menu bar on the left of the screen and use the key to select the View
tab.
2. Use the Click Wheel to select Select Layout, press and use the Click Wheelagain to select the
appropriate layout.
3. Use the Click Wheel to select Select Layout, use the Click Wheelto select the appropriate layout.
In the drawing 27 above, the “pie-shaped” image is referred to as a sectorial scan or S-scan. It is an image
created from the samples of 71 beams from 35° to 70°, with a 0.5° step. The colors code the amplitude of
the underlying A-scans.
The dashed angled line on the S-scan is referred to as the Angular Extractor cursor. This cursor indicates
the origin of the A-scan which is displayed in the A-scan view. Changing the extractor angle in the S-scan
V1.1 31
will change the A-scan angle displayed in the A-scan view. This is one advantageous feature of phased
array, i.e. it permits the interrogation of multiple beam angles in real-time.
The cross-hair cursor on the S-scan is called a Cartesian cursor. The cross-hair location will report the
depth (D) and true surface distance (SD) offset. If the Cartesian cursor is moved over a coloured region on
screen it will return the location of that position regardless of the angle at which the peak signal occurs.
All measurements are referenced from the wedge reference point. The wedge reference location
depends on the type of wedge you are you using, if any. The wedge reference is represented by the red
dot in the 3D and scan plan views used in the Configuration mode.
For more details see the “Defining scan plan geometry (cartesian/axis reference system)” section.
Each view has its own rulers which can be used to quickly evaluate the location of flaws.
28. Drawing: phased array analysis and acquisition layout, ruler identification
Each A-scan has independent signal gates. The gates can be positioned along the time/distance axis. Each
gate has a threshold level which if exceeded by the signal triggers a measurement, the result of which is
displayed next to the relevant gate; whenever the signal crosses the gate, a measurement is calculated
and displayed on the side of the gate. Measurements are also presented at the top of the screen.
The green triangle at the top right of the screen indicates that real-time imaging mode is active and that
ultrasonic pulses are continually being generated. Real-time data can be recorded by pressing or the
image frozen by pressing
V1.1 32
15.2 Moving the A-scan Extractor cursor
1. In acquisition mode press the key and select the desired S-scan view.
2. Press the key (once or multiple times) to select the extractor in the current view.
3. Use the Click Wheel to move the selected Extractor cursor to the desire position.
4. Press to modify the box size then use the Click Wheel to change dimensions and press again to
accept them.
When using phased array it is important to be able to adjust the focusing of the ultrasonic sound beam.
This is achieved using the Focus Dist item. The focusing capability is restricted to the near-field length. In
other words, beam focusing is only achievable to distances shorter than the near-field length. For weld
testing, a non focused beam is often preferred and this can be achieved by setting the focal distance to a
value much greater than the near-field length. Refer to the 3D view to know the near-field length of the
probe.
2. Use the RIGHT ARROW key on the Click Wheel to reach the Scan tab.
3. Use the DOWN ARROW key to reach the Focus Dist item.
V1.1 33
15.5 Calibration wizards (for phased array)
The calibration wizards are listed in the sequence that they are normally used. For a multi-scan inspection
each scan must be calibrated separately.
Type Description
Velocity & Zero Wizard Guides you through calibration of the material velocity and the
measurement zero.
Wedge Delay Wizard Guides you through calibration of the Wedge Delay. Only
applicable to phased array.
Element Check Wizard Guides you through the process for determining if an element
is defective.
V1.1 34
16 Manipulating cursors (Cartesian, Angular, Gate, Extractor Box, and Extractor)
The following is the procedure to move the different cursors available in all the views.
1. In Acquisition mode use the key (once or multiple times) to select the appropriate view.
2. Use the key (once or multiple times) to select the appropriate cursor in the current view.
4. To change the gate size press then change size by using the Click Wheel.
V1.1 35
17 Adding cursors
Cartesian Cartesian cursors are used to measure surface distance (SD) and depths
in S-scans, L-scans and A-scans.
Extractor The Extractor Box is used to extract Top and End views from S-scans and
Box L-scans.
Hyperbolic Hyperbolic cursors are used to evaluate surface distance (SD) and depth
measurements in a TOFD (Time of Flight Diffraction) scan. The TOFD
view is a B-scan view from a pitch-catch probe geometry. The
hyperbolic cursor is used because the depth measurement is non-linear.
1. In Acquisition mode, press the key (once or multiple times) to select the appropriate view.
2. Press and select the View tab.
V1.1 36
18 Making measurements with gates
Gates are the most common and precise way to make measurements. A gate measurement is made as
follows:
1. In Acquisition mode, use the gate key (once or multiple times) to select the appropriate gate.
2. Use the Click Wheel to move the selected gate to the desired position.
3. To change the gate size, press and change size using the Click Wheel.
V1.1 37
19 Customizing the Measurements bar
The Measurements bar is shown at the top of the screen in PLAY mode, when a menu is not selected and
a view is selected (in PLAY mode press the button). It can also be shown permanently in PLAY
mode by enabling the Keep Measures preference.
To select desired measures from cursors:
3. Select one of the measurements items (1 through 5) and press to modify it.
Each measurement item is composed of three parts; the cursor identifier, the reference point, and the
type of measurement. For example, “G1 ^ 2” means “the true depth in the second half skip of the
peak detected in the gate number 1”.
V1.1 38
The following three tables present the details of each measurement type.
V1.1 39
20 Taking screenshots and building reports
The following is the procedure to make screen capture and produce reports.
1. To make a screenshot or create a report press the key.
2. Select Report or Screen Capture (refer to sections about loading and saving files).
V1.1 40
21 Defining scan plan geometry (cartesian/axis reference system)
The scan plan is defined on the Geometry tab where the position of the probes and wedges are set.
Table 7 lists the icons used to visually identify the principal data on the scan graphic.
V1.1 41
21.1 Wedge references
Table 8 defines the wedge reference ( ) for various different probe-wedge combinations for phased array
and conventional UT configurations. Note that the wedge reference is different in each case.
When an inspection involves more than one probe, the probes are positioned relative to a group
reference point, or Grp Ref, identified by . This can be located anywhere on the probe retention
assembly, but it is a good practice to choose a position coincident with the centre line of the weld or scan
axis and the probe/wedge assembly. See drawing 33.
V1.1 42
33. Drawing: group reference (“Grp Ref”)
The part datum, identified by , is a point chosen on the inspected part which acts as the spatial origin
for all of the acquired data. Where possible it is recommended that the part datum is coincident with the
Grp Ref point. Alternatively, an offset between the Scan Start Position on the Encoding tab and the Grp
Ref can be defined.
V1.1 43
22 Setting up position-encoded or time-encoded scans
2. Use the RIGHT ARROW key on the Click Wheel to access the Encoding tab.
Note: In order to connect two encoders a Y-splitter cable is used for the connector marked “E”.
V1.1 44
23 Setting up an encoded scan
OK button to move up or down the menu. Press to select from the following parameters to edit:
• Geometry: W1 Index Offset (W1 stands for "Wedge 1", or "Probe 1" if there’s no wedge)
• Geometry: W1 Scan Offset (W1 stands for "Wedge 1", or "Probe 1" if there’s no wedge)
• Geometry: W1 Rotation (W1 stands for "Wedge 1", or "Probe 1" if there’s no wedge)
4. Review the parameters on the Encoding tab use an encoder is intended:
• Encoding: Encoding Setup
• Encoding: Scan Enc. Type (the type of encoder used)
• Encoding: Scan Start Pos (the start position )
• Encoding: Scan Distance
• Encoding: Scan Step
5. Press PLAY to change to the Acquisition mode.
V1.1 45
24 Selecting a layout with appropriate views
The type of inspection determines the views that are available. Table 9 lists the different
view types.
View Description
A A-scan view
B B-scan view
C C-scan view
D D-scan view
S S-scan view
TOP Top view
END End view
TOFD TOFD view
MTOP Merged Top view
MC Merged C-scan
B-LOG Thickness B-Scan view
Table 9: List of available view types
1. Select the View tab and press to open the list of available views.
V1.1 46
25 Menu tree (menu items vary according to the inspection setup)
V1.1 47
Scan Scan (continued) View
• Add… RX • Select Layout…
• Delete… • Signal Rectification • View Orientation
• Digital Filter • Add Cursor…
• Scan Type
• Analog filter Palette Properties
Gain
• Sub-Sampling • View Palette
• Gain
• Rejection • Palette Position
• Reference Gain
• Reject Threshold • Palette Ampl. Low
• Set Reference Gain
• Smoothing • Palette Ampl. High
• Software Gain
• Contouring • Palette Depth Low
• Auto FSH%
• Cont. Decay Rate • Palette Depth High
• Reference Amplitude
• Averaging Data Extraction
Acquisition Area
• IFT Active • Depth/Amp Mode
• Resolution
- IFT Start, Range, Threshold
• Start Angle • Depth/Path Mode
• Acq. Freq
• Stop Angle • Data Gate
Probe Connect • Reference Gate
• Angle
• Probe TX/RX • IFT Culling
• Start Path
Elements Rulers & Grids
• Range Path
• First Element TX/RX • Rulers
• Stop Path
• Last Element TX/RX • Depth/Path Mode
• Zero
Statistics • Grids
• Wave Mode
• Beams Quantity Scroll & Zoom
• Travel Mode
• Samples Quantity • View Scrolling Mode
Focusing
• Path Resolution • Frame Start
• Number of Active Element
• Double Resolution Cursor • Zoom
Amplitude
A-scan
Sound path
Scan
B-scan
Sound path
Scan
C-scan
Surface distance
Index
D-scan
Sound path
Surface distance
S-scan
Depth
Scan
Top
Surface distance
Scan
End
Depth
Scan
TOFD
Depth (non-linear)
Table 10: List of ruler color map
V1.1 49
27 Power LED Behavior
V1.1 50
28 Color palettes
3 level threshold
ABF Amplitude
Aeronautic
Cubic Law
Fire Ice
Rainbow
Spectrum
Thermal
Greyscale
TOFD
V1.1 51
29 Features not covered in this document
The following list of features are not covered in this quick start guide, but are addressed in the full “User
Guide”.
• Calibration on each type of reflectors
• DAC
• Split DAC
• DGS
• Split DGS
• API
• AWS measurements
• TCG
• BEA
• TOFD lateral wave straightening
• TOFD lateral wave removal
• Data analysis
• Detailed documentation of all type of measurements
• Example of typical inspections
• Enabling software options
V1.1 52
30 Encoder Connector Pin out
V1.1 53
31 I/O Connector Pin out
Pin # Description
1 (red dot) Power 5V, 300mA
max.
2 5V TTL, Trig-Sync
3 Output, Gate 1
4 Output, Not yet
defined
5 Output, Not yet
defined
6 Input, Not yet
defined
7 Input, Not yet
defined
8(center pin) Power return –
Ground
V1.1 54
32 Definitions
The Type parameter on the Scan tab allows the user to select between Sectorial, Linear, Mono, and TOFD
scans. Scans can be configured to be either pulse-echo (PE) or pitch-catch (PC), with the exception of
TOFD which is always pitch-catch. The scan types are described in the following sections.
The sectorial scan, also called azimuthal scan or more commonly S-scan, is a cross-sectional image of the
inspected volume. It represents an angular coverage of the volume under the probe.
The linear scan, or L-scan, is represented on screen as a parallelogram view, directly under the probe or at
a specific angle. The L-scan is created from multiple A-scans from different exit points, but with the same
beam angle. A small group of elements are activated in order to generate each beam (focal law) and the
group of activated elements is cycled along the elements array for each subsequent beam.
Mono scan refers to the conventional A-scan representation acquired using a single element or pair of
elements in the case of a dual probe. The A-scan is a waveform representing the amplitude of the
ultrasound signal as a function of time or propagation distance. Events in the A-scan , or echoes, occurs
when the ultrasound wave packet encounters a discontinuity, which may be a defect, such as a crack, or
may be due to the intrinsic geometry of the item inspected.
The time-of-flight-diffraction (TOFD) scan involves a pitch- catch arrangement of conventional UT probes
from which the data is displayed using a greyscale B-scan to encode multiple A-scans.
V1.1 55
33 Copyright, disclaimers, and certifications
33.1 Copyright
All statements, technical information and recommendations contained in this manual or any other
information supplied by Sonatest Limited in connection with the use, features and qualifications of the
Prisma is based on tests believed to be reliable, but the accuracy or completeness thereof is not
guaranteed. Before using the product you should determine its suitability for your intended use based on
your knowledge of ultrasonic testing and the characteristics of materials. You bear all risk in connection
with the use of the product.
You are reminded that all warranties as to merchantability and fitness for purpose are excluded from the
contract under which the product and this manual have been supplied to you. The Seller’s only obligation
in this respect is to replace such quantity of the product proved to be defective.
Neither the seller nor the manufacturer shall be liable either in contract or in tort for any direct or
indirect loss or damage (whether for loss of profit or otherwise), costs, expenses or other claims for
consequential or indirect compensation whatsoever (and whether caused by the negligence of the
company, its employees or agents or otherwise).
V1.1 56
33.3 Certification (electromagnetic compatibility)
Warning: This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy, and, if not installed and used in
accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his or her own expense.
The authority to operate this equipment is conditioned by the requirement that no modifications will be
made to the equipment unless the changes or modifications are expressly approved by Ascend
Communications, Inc.
The Prisma also complies with EN 12668-1:2010, Non-destructive testing and verification of ultrasonic
examination equipment – Part 1: Instruments.
V1.1 57