0% found this document useful (0 votes)
381 views

CETA-Compounding-Isolator-Testing Guide PDF

This document establishes testing criteria for compounding isolators used for sterile compounding. It outlines general responsibilities and minimum criteria for airflow, pressure, HEPA filter integrity and other tests. Responsibility for manufacturer testing lies with the manufacturer, while field certification testing lies with third-party certifiers. A set of information must be displayed on certified isolators including pass/fail of ingress/egress tests and recovery time.

Uploaded by

santanu paul
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
381 views

CETA-Compounding-Isolator-Testing Guide PDF

This document establishes testing criteria for compounding isolators used for sterile compounding. It outlines general responsibilities and minimum criteria for airflow, pressure, HEPA filter integrity and other tests. Responsibility for manufacturer testing lies with the manufacturer, while field certification testing lies with third-party certifiers. A set of information must be displayed on certified isolators including pass/fail of ingress/egress tests and recovery time.

Uploaded by

santanu paul
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 20

CETA Compounding Isolator Testing Guide

CAG-002-2006
Revised December 8, 2008
1.0 Overview:
The purpose of this document is to establish an industry-based minimum set of testing criteria appropriate
for all Compounding Isolators used pursuant to USP Chapter 797 1. Compounding Isolators consist of
Compounding Aseptic Isolators used for compounding sterile preparations and Compounding Aseptic
Containment Isolators used for compounding sterile hazardous drug preparations in pharmacy applications.
While this document gives general guidance or referenced guidance through relevant industry documents, it
is not the intention to set the specific acceptance criteria. It is the manufacturer’s responsibility to
determine exact testing procedures consistent with these guidelines and assign appropriate values
pertaining to acceptance criteria that is consistent with user requirements. This guide has also been
established to create a uniform approach for field certifiers to allow consistent and repeatable testing at all
facilities.

1.01 General Responsibilities:


The criteria outlined are intended to describe currently accepted practice. Responsibility for test
performance lies with the manufacturer during design and production, and field certifiers at the installation
site. INFORMATIONAL NOTES are inserted to aid users of this guide by providing a format or general
reference for the information desired. The engineering and design concepts employed are up to the
individual manufacturer’s discretion. The general assumptions in determining the minimum criteria for
Compounding Aseptic and Compounding Aseptic Containment Isolators covered in this document are:

used for compounding sterile preparations


used in either a classified or unclassified space
use of “open” design, using pass-through chamber systems to transfer material in and out
may be either positive or negative pressure depending on the hazard of the drugs
no expectation of end user employing gaseous decontamination of the Compounding Isolator

An isolator certified by the manufacturer as meeting the criteria established in CETA CAG-002-2006 shall
post the following on the front of the cabinet in a location readily visible to the user:

1. Notice of pass or failure of the Preparation Ingress and Egress Test (2.09).
2. Whether or not designed for use with volatile hazardous drug compounds, toxic, flammable or
explosive materials.
3. Indication of the results of the Recovery Time Determination Test (2.07).

NOTE: Instructions for material transfer should be developed by individual user/ facility, based
on their SOPs taking into account purge and recovery times listed by the isolator manufacturer.
a. Purge time for the pass through
b. Purge time after materials are entered into the isolator

An isolator field certified to the statement “in accordance to CETA CAG-002-2006” shall have the
following information posted on the front of the cabinet in a location readily visible to the user:

1. Information to be completed by testing technician:


A. Date tested
B. Reference report number
C. Due for retest date (month and year)
D. Tested by:
1) Name printed
2). Signature

CETA CAG-002-2006 1 12/08/2008


E. If the cabinet does not have manufacturer’s listed information noted above.
1) Pass/ fail status of the material ingress/ egress test
2) Recovery time of the main chamber.
3) Indication that the operator has to wait at least the recovery time before
initiating operations after material transfer.
4) Ventilation appropriate for use with volatile / non-volatile hazardous
drugs.

2. Information to be pre-printed on label:

A. Company name
B. Address
C. Phone number

1.02 Test Responsibility

Procedure Manufacturer Field Test


2.01 Airflow Test X X
2.02 Chamber Pressure Test X X
2.03 Site Installation Assessment Tests D X
2.04 Gauntlet Breach Air Velocity Test D O
2.05 HEPA Filter Integrity Test X X
2.06 Particle Containment Integrity and Enclosure Leak Test X X
2.07 Recovery Time Determination Test D O
2.08 Airflow Smoke Pattern Test X X
2.09 Preparation Ingress and Egress Test D X
2.10 Particle Count Tests X X
2.11 Volatile Hazardous Drug Containment Tests D O
2.12 Hazardous Particle Containment Test D O
2.13 Pass-through Particle Purge Time Determination Test D O

X = Mandatory
O = Optional
D = Design criteria qualification

1.03 Glossary of Terms

Ambient
Environment external to the compounding isolator

Compounding Aseptic Isolator:


A form of isolator specifically designed for compounding pharmaceutical ingredients or
preparations. It is designed to maintain an aseptic compounding environment within the isolator
throughout the compounding and material transfer processes. Air exchange into the isolator from
the surrounding environment should not occur unless the air has first passed through a microbially
retentative filter (HEPA minimum).

Compounding Aseptic Containment Isolator:


A compounding aseptic containment isolator is designed to provide worker protection from
exposure to undesirable levels of airborne drug throughout the compounding and material transfer
processes, and to provide an aseptic environment for compounding sterile preparations. Air
exchange with the surrounding environment should not occur unless it is first passed through a
microbially retentive filter (HEPA 3 minimum) system capable of containing airborne
concentrations of the physical size and state of the drug being compounded. Where volatile
hazardous drugs are prepared, the exhaust air from the isolator shall be appropriately removed by
properly designed building ventilation. 2

CETA CAG-002-2006 2 12/08/2008


Critical Area
An ISO Class 5 environment.

Critical Site:
A location that includes any component or fluid pathway surfaces (e.g. vial septa, injection ports,
beakers) or openings (e.g. opened ampule, needle hubs) exposed and at risk of direct contact with
air (e.g. ambient room or HEPA filtered), moisture (e.g. oral and mucosal secretions), or touch
contamination. Risk of microbial particulate contamination of the critical site increases with the
size of the openings and exposure time.

Critical Work Zone:


The defined specific localized area starting 6” (15.2 cm) below the IV bar including horizontal and
vertical space within the work area where aseptic manipulations are performed.

Design Acceptance Test:


A manufacturer test performed to verify that the design meets intended criteria. Must be
successfully completed on at least 2 units of each model and size upon introduction and
modification of that design.

Direct Compounding Area (DCA)


A critical area within the ISO Class 5 primary engineering control (PEC) where critical sites are
exposed to unidirectional HEPA-filtered air, also known as first air.

HEPA (high-efficiency particulate air) Filter 3


An extended-medium, dry-type filter in a rigid frame when tested at rated airflow having a
minimum particle collection efficiency of 99.97% for 0.3-µm mass median diameter particles of
DOP when tested in accordance with MIL-STD-282 4.

Main Chamber:
The primary chamber containing the work area, critical work zone, and means for manipulation of
the preparation. This includes plenums and areas covered with diffusers.

Pass-through Chamber:
The chamber designed to separate the surrounding ambient space from the main chamber while
facilitating the movement of materials between the two areas.

Particle Elevation:
Airborne room particles elevation to enable certain tests. This elevation may be accomplished
through the use of aerosol generator, theatrical smoke generator, or other smoke sources that
provide suitable particle size and quantity

Primary Engineering Control (PEC)


A device or room that provides an ISO Class 5 environment for the exposure of critical sites when
compounding CSPs. Such devices include, but may not be limited to, laminar airflow work-
benches (LAFWs), biological safety cabinets (BSCs), compounding aseptic isolators (CAIs) and
compounding aseptic containment isolators (CACIs).

Surrogate Manipulation:
A manipulation of a preparation that involves the use of substituted materials that closely resemble
actual materials used in sterile compounding to provide a similar impact upon the environment
tested and to resemble actual operational conditions during testing

Unidirectional Airflow:
An airflow moving in a single direction in a robust and uniform manner and at sufficient speed to
reproducibly sweep particles away from the critical processing or testing area.

CETA CAG-002-2006 3 12/08/2008


Volatile
A gas, vapor, or micro-aerosol potentially generated during the pharmacy compounding process
that can pass through a High Efficiency Particulate Air (HEPA) filter.

Work Area
The horizontal and vertical space within the isolator from wall to wall and from the base or floor
to the top within the main chamber. This excludes plenums and areas covered with diffusers.

1.04 Safety Concerns


Testing in accordance with this document may involve hazardous materials, equipment and operations. It
is the responsibility of the user to consult and establish appropriate safety and health practices and to
determine the applicability of regulatory limitations prior to use of this document.

2.00 Compounding Isolator Tests:


The following tests are offered to establish that the Compounding Isolator design is appropriate for the task,
that test methods are repeatable and accurately reproducible in the field, to provide detailed guidance for
the field certifier through a specific field test documentation and / or through a designated section of the
product owner’s manual, and to set appropriate acceptance values.

The manufacturer shall be responsible for identifying special accommodations, or ancillary equipment that
may be needed for accurately field testing the Compounding Isolator design. These may include, but are
not limited to the methods, sources, or devices to provide accurate and repeatable testing such as:

correcting for impact of operating pressures where applicable


provision of ports to support the use of diagnostic equipment
supplementary equipment, materials, products to facilitate tests
providing and updating performance information as necessary
For turbulent flow isolators, provide chamber volume needed to calculate air changes.

INFORMATIONAL NOTE:
Metric equivalents (SI) are provided in parentheses ( ) behind English
measurement values throughout this document as an aid or cross-reference.

2.01 Airflow Test:


Purpose:
This test verifies the appropriate airflow parameters are met for the manufacturer’s design intent.

Apparatus:
A calibrated thermal anemometer with an accuracy of ± 3.0 fpm (0.015 m/s) or ± 3% of the
indicated velocity, whichever is greater.

Procedure:
Main Chamber:
Measure the air velocity in the Compounding Isolator using equally spaced points in the horizontal plane
that produces the most repeatable readings.

CETA CAG-002-2006 4 12/08/2008


INFORMATIONAL NOTE:
As an example, the horizontal plane would be set at X inches (cm) below
the filter diffuser using a X inch (cm) grid.

An example of uniformity ranges may be a) +/- Y % of the average or +/-


X fpm (m/s), whichever is greater, or b) +/- Y % of the average velocity.

Procedure – All Compounding Isolators:


Pass-through Chamber
In a purged pass-through chamber, the manufacturer shall develop a repeatable method of determining
airflow volume or velocity to achieve the stated purge objective.
1. For unidirectional purged pass-through, measure the air velocity at multiple points across the pass-
through using equal points in a plane that produces the most repeatable readings or determine an
alternate repeatable method to determine that the specified purge objective is met.

2. For non-unidirectional purged pass-through, the manufacturer will determine the optimum
location and method that produces the most repeatable readings or determine an alternate
repeatable method to determine that the specified purge objective is met

INFORMATIONAL NOTE A: Airflow velocity and / or volume and chamber


pressurization will interact with each other. Both tests should be verified before
recording final airflow and pressure values.

INFORMATIONAL NOTE B: Ideally the manufacturer’s recommended pass-


through purge time should relate to and specify an acceptable level of airborne
particulate at conclusion of the purge time. For example, a purge time of X
seconds or minutes should result in airborne particle levels consistent with ISO
Class Y at 0._ µm and larger size particulate classification 5 within the pass-
through chamber.

Acceptance Criteria:
The manufacturer will determine the appropriate acceptance criteria for the particular design.
Unidirectional Airflow Compounding Isolator velocity criteria shall be expressed in terms of a
range of velocities in feet per minute (fpm) (meters per second (m/s)) along with a velocity
uniformity designator. For example, a statement would be presented as the average velocity shall
fall between __fpm (m/s) and __fpm (m/s) and __% uniformity from the average

INFORMATIONAL NOTE: Airflow velocity and / or volume and chamber


pressurization will interact with each other. Both tests should be verified before
recording final airflow and pressure values.

2.02 Chamber Pressure Test:


Purpose:
This test is performed to determine that pass-through and main chamber pressures are adequate to aid in
providing Compounding Isolator separation between the main chamber and ambient spaces.

Apparatus:
A calibrated mechanical or electronic differential pressure indicating device.

CETA CAG-002-2006 5 12/08/2008


INFORMATIONAL NOTE: Differential pressure indicating devices should
have a fast reaction or update capability (time constant) of less than 1 second
in order to detect momentary or large changes in differential pressure.

Procedure:
1. Utilize the factory installed test port or sealed access location to main and pass-through chambers.
2. Attach test instrument tubing into chamber and allow pressure levels to stabilize.

Test 1:
Record pressure level of main chamber with outer and inner pass-through doors closed and
isolator gloves / gauntlets extended into Compounding Isolator.
Fully open outer pass-through door. Document any change to the pressure.
Confirm that the pressure does not change from positive to negative or from negative to positive.
Close outer pass-through door. Document any change to the pressure.

Test 2:
Record pressure level of main chamber with outer and inner pass-through doors closed and
isolator gloves / gauntlets extended into isolator.
Using Compounding Isolator glove, fully open inner door. Document the change to the pressure.
Confirm that the main chamber pressure does not change from positive to negative or from
negative to positive.
Close inner door. Determine if pressure level returns to original levels.

Test 3a:
Record pressure level of Compounding Aseptic Isolator main chamber with outer and inner pass-
through doors closed and isolator gloves / gauntlets extended into isolator.
Insert hands into gloves and pull back from isolator to simulate operator hand removal. An
example of a reasonable pull back time is no greater than 3 seconds from fully extended to fully
extracted (gloves at a point even with the view panel)
Determine the change in pressure and verify that the pressure does not change from positive to
negative.

Test 3b:
Record pressure level of Compounding Aseptic Containment Isolator main chamber with outer
and inner pass-through doors closed and isolator gloves / gauntlets extended outside isolator.
Insert hands into gloves and push into main chamber to simulate operator hand insertion.
Determine the change in pressure and verify that the pressure does not change from negative to
positive.

Acceptance:
Compounding Isolator operating pressure range shall be determined by the manufacturer.
The pressure shall not change from positive to negative or from negative to positive during any
manipulations.

2.03 Site Installation Assessment Tests:

Purpose:
These tests are performed to verify:
The airflow or pressure setpoint(s) where an audible and/or visual alarm will activate to
signify unfavorable operating conditions within the compounding isolator and/or the remote
exhaust blower.
Functional pass-through door interlock operation.
Proper canopy or exhaust connection performance.

CETA CAG-002-2006 6 12/08/2008


Apparatus:
A calibrated thermal anemometer with an accuracy of ± 3.0 fpm (0.015 m/s) or ± 3% of the
indicated velocity, whichever is greater.
A calibrated mechanical or electronic differential pressure indicating device.
Alarm adjustment/alignment tool.
A source of visible smoke - such as a chemical smoke tube.

Procedure:

Test 1: Airflow or Pressure Alarm Test:


Verify the alarm setpoint(s) using the compounding isolator or alarm manufacturer’s
performance verification procedures.
Adjust as necessary.
Return isolator to certification setpoint.

Test 2: Door Interlock Verification Test:


Verify the inner and outer door interlock system operates as per manufacturer’s design
criteria.

Test 3a: Exhaust System Performance – Canopy Connections:


Using a visible smoke source, verify negative pressure at the gap.
Measure the duct static pressure between the air gap and any duct-mounted balancing
dampers.

Test 3b: Exhaust System Performance – Hard ducted connections:


Measure the duct static pressure between the isolator discharge point and any duct-
mounted balancing dampers.

Acceptance:
Airflow or Pressure Alarm Test
The setpoint(s) shall be determined by the manufacturer as appropriate for operator safety, product
sterility and/or containment.

Door Interlock Verification Test


Determined by the manufacturer.

Exhaust System Performance


No smoke shall escape into the room once it enters the exhaust system

2.04 Gauntlet Breach Air Velocity Test:


Purpose:
This test is performed on containment Compounding Isolators to assure some level of operator protection in
the event of a significant glove or gauntlet integrity failure.

Apparatus:
A calibrated thermal anemometer with an accuracy of ± 3.0 fpm (0.015 m/s) or 3% of the
indicated velocity, whichever is greater.

Procedure:
1. Determine that all other operating parameters have been met.
2. Remove one gauntlet from chamber view screen.
3. Measure velocity at the geometric center of the opening. A minimum of 3 samples at a minimum
of 1 location shall be taken.
4. Record results.

CETA CAG-002-2006 7 12/08/2008


Acceptance:
The velocity shall be determined by the manufacturer as appropriate for operator safety.

INFORMATIONAL NOTE: Published sources show various ranges of


airflow velocities that pertain to gloveboxes and other containment devices.
These sources include:
NSF/ANSI Standard 49 (75 to 100 lfpm) (0.38 to 0.51 lm/s)6
ANSI/AIHA Z9.5-2003 (> 100 lfpm) (0.51 lm/s) 7

2.05 HEPA Filter Integrity Test:


Purpose:
This test determines the integrity of the all HEPA filters, filter housings, and filter mounting frames. The
cabinet shall be operated at operational airflow for this test.

Apparatus:
A calibrated Aerosol Photometer capable of indicating 100% upstream concentration with an
aerosol of between 10 and 90 µg per liter of polydispersed dioctylphthalate (DOP or DEHS) CAS#
117-81-7, polyalphaolefin (PAO) CAS# 68649-12-7 or an equivalent fluid. Unit must have a
threshold sensitivity of at least 10-3 micrograms per liter and be capable of measuring
concentrations over a range of 105 times the threshold sensitivity. The sampling rate shall be 1
CFM (28.3 lpm) (±10%) with an inlet probe having sufficient area (1.7 square inches) (11.0 cm 2)
as to maintain a probe inlet velocity of 90 fpm (0.46 m/s) or slightly higher.
Laskin Nozzle aerosol generator or equivalent.

Procedure:
Main and Pass-through Chamber HEPA filters:

1. Turn on the Compounding Isolator blowers and lights.


2. Remove the HEPA filter diffuser screen(s) if present and where possible.
3. Introduce the aerosol in a manner that will create a well mixed aerosol upstream of the filter.
4. If a specific introduction location is needed to achieve a well mixed aerosol, that location should
be documented by the device manufacturer.
5. Sample or calculate the upstream challenge and verify that a 10-90 µg per liter has been
introduced.
6. Adjust the photometer to 100% based on the challenge levels used.
7. With the nozzle of the probe held not more than 1.0 inch (2.5 cm) from the area being tested, scan
the entire downstream side of the HEPA filter(s), the perimeter of each filter pack and mounting
frame by passing the photometer probe in slightly overlapping strokes at a traverse rate of not
more than 2 inches (5.1 cm) per second when using a rectangular probe.
8. For filters that cannot be scanned, the manufacturer shall develop a test procedure that allows for
probe testing at a location where adequate mixing and safety to the certifier can be achieved.

INFORMATIONAL NOTE: The manufacturer shall verify both temporal and


spatial uniformity of challenge as outlined in IEST-RP-CC034 8.

Acceptance Criteria:
For filters that can be scanned:
Sustained aerosol penetration shall not exceed 0.01% of the upstream concentration.

CETA CAG-002-2006 8 12/08/2008


For filters that cannot be scanned:
Sustained aerosol penetration shall not exceed 0.005% of the upstream concentration.

Filter patch size shall not exceed 3% of the effective filter area of the side being patched. The maximum
width of any one patch shall not exceed 1.5 inches (3.8 cm).

2.06 Particle Containment Integrity and Enclosure Leak Test:


Purpose:
This test is performed on Compounding Aseptic Containment Isolators only. This test can aid in
determining the particle integrity of the cabinet construction joints, seams, access panels and seals, glove
ports and entry/exit points into the main chamber or where elevated particle levels have been detected at or
near the work surface and where other potential leak sources have been eliminated (e.g. damaged HEPA
filters, damaged glove/gauntlet, etc.).
NOTE: A suitable alternative test method to 2.06 is the AGS-G001–2007 Guideline for
Gloveboxes 9 section # 9.6.2 Leak Detection. Acceptance criteria is less than or equal to 0.5% of
the device volume per hour.
Apparatus:
A calibrated discrete particle counter with a particle size discrimination capability of 0.5 µm.

Procedure –Compounding Aseptic Containment Isolators:


1. The isolator shall be operated at operational airflow.
2. Verify the background count in the testing room is at least 3,520,000 particles per cubic meter
(ppcm) or 100,000 particles per cubic foot (ppcf). If the count is too low, elevate the background
levels using an aerosol generator or smoke generator.
3. Scan interior of main chamber within 1 inch (2.5 cm) of surfaces using a scan rate of no more than
2 inches (5.1 cm) per second.
4. Areas of interest include:
a) all welded or gasketed seams
b) side glass gasket/seams
c) pass-through chamber door seals
d) glove ports perimeters and gloves
e) front view screen gasket seals
f) trash or sharps exit ports/chutes
g) lights installed inside the main chamber
h) any penetrations into the main chamber

Acceptance:
Any detected leaks do not exceed ISO class 5 conditions of 3,520 particles per cubic meter (ppcm)
at 0.5µm size particles and larger when the probe is held stationary within 1” (2.5 cm) of the leak.

INFORMATIONAL NOTE:
Elevating operating pressures during this test may aid in locating leak sources.

This test is most effective when performed in uncontrolled environments


(>3,520,000ppcm or >100,000ppcf ).

2.07 Recovery Time Determination Test:


Purpose:
This test is performed to determine the amount of time it takes the main chamber to recover to ISO Class 5
particle levels after an event such as an open door or a surge of process generated contamination.

CETA CAG-002-2006 9 12/08/2008


Apparatus:
Laskin Nozzle aerosol generator or equivalent.
A calibrated discrete particle counter with a particle size discrimination capability of 0.5 µm and
variable sample times.
Time keeping device capable of reading in minutes and seconds.
Particle sample probe cover.

Procedure:
1. Set the particle counter sample time to 6 second sample periods with a 14 second hold time in
“Concentration” mode to report in particles per cubic meter or foot.
2. With the Compounding Isolator running, measure particle levels at the center of the work-surface
of the main chamber.
3. Collect multiple samples to determine average baseline particle levels using 6 second samples.
4. Three consecutive baseline particle level samples should be established and noted.
5. Turn off particle counter sample pump and cover the particle counter sampling probe.
6. Turn off Compounding Isolator and fill the chamber with particulate using a Laskin nozzle
generator set at 1 Laskin nozzle at 20 psi (138 kPa) for five seconds per cubic foot (per 0.028
cubic meter) of chamber interior space.
7. Turn on the Compounding Isolator and start timer.
8. To prevent sampling above the particle counter’s coincidence loss rate or damaging the device,
wait until the smoke is visibly cleared from the chamber and remove particle counter probe cover
and begin sampling.
9. Maintained particle levels are achieved when three consecutive counts are at or below the
originally determined baseline particle level.
10. Total recovery time is considered from Compounding Isolator blower turn-on time to the first
particle count where maintained particle levels were achieved.

Acceptance criteria:
The manufacturer will determine the appropriate acceptance criteria for their particular design.
Recovery time will be expressed in minutes and seconds to recover to baseline from excursion.

2.08 Airflow Smoke Pattern Test:


Purpose:
This test determines that the airflow within the compounding chamber conforms to the manufacturer’s
design criteria.
Airflow within the Direct Compounding Area is downward with no dead spots or refluxing in the
Critical Site, that ambient air does not enter the chamber other than through the supply HEPA
filter, and that once the air enters the Direct Compounding Area, it is removed to the returns
without reentry.

Apparatus:
A source of visible smoke that is generally neutrally buoyant

INFORMATIONAL NOTE: Chemical smoke tubes or glycol-based smoke


generators are examples of acceptable smoke sources.

Procedure:
1. Pass the smoke along the work area beginning 1 inch (2.5 cm) from all main chamber walls and
view screens at a height determined to authenticate the purpose of test.

2. Pass smoke over gloves, gauntlets, IV bars, interior lights, or other extruding features that may be
installed within the main chamber.

CETA CAG-002-2006 10 12/08/2008


3. For Compounding Aseptic Containment Isolators, pay particular attention to all penetrations, pass-
through door, trash disposal tubes and gloves where turbulence may occur or may indicate
significant penetration of the Compounding Isolator.

Acceptance criteria:
The smoke shall show smooth downflow with a minimum of dead spots or reflux (upward flow) at
obstructions and across the critical work zone and it is removed to the returns without reentry.

2.09 Preparation Ingress and Egress Test:


Purpose:
This test is performed to determine if the Compounding Isolator pass-through system is capable of
supporting material transfer while maintaining the designated cleanliness classification during the transfer.

Apparatus:
A calibrated discrete particle counter with a particle size discrimination capability of 0.5 µm.
Aerosol generator or smoke tubes.
Empty perforated transfer tray.
Time keeping device capable of reading in seconds.

Procedure:
1. Verify the background count in the testing room is at least 3,532,000 particles per cubic meter
(ppcm) (100,000 particles per cubic foot (ppcf)).
2. If the count is too low, elevate the background levels using an aerosol generator or smoke
generator.
3. Place the particle counter probe in the Main Chamber approximately 6 to 8 inches (15.2 to 20.3
cm) off the isolator floor surface, approximately 2 inches (5.1 cm) outside the normally used path
of the inner pass-through chamber door swing and within the area that would be affected by the
airflow caused by the movement of the door. Probe placement should be so that the operator’s
arms will not pass directly over the probe when removing material from the pass-through.
4. Verify the particle counts meet ISO Class 5 levels before beginning the test cycle.
5. Set the particle counter for a one minute count with no more than a one second hold time.
6. Open the outside pass-through door.
7. Place a perforated empty transfer tray into the pass-through and close the outer door.
8. Wait for the manufacturer’s recommended purge time or start the purge cycle.
9. After completion of the purge cycle, open the inside pass-through door and move the transfer tray
from the pass-through to the work area.
10. Close the inside pass-through door.
11. Document the particle counts during the transfer process and for a period of one minute after the
transfer.

Acceptance Criteria:
The particle counts shall not exceed the ISO Class 5 @ 0.5 µm and larger class limit at any time
during this test.

2.10 Particle Count Tests:


Purpose:
This is a two part test and is performed to verify that the Compounding Isolator main chamber operates
within ISO class 5 conditions at 0.5 µm and larger during Static (At-Rest) and Dynamic Operating
(Operational) conditions to determine that air within the isolator and the immediate proximity of exposed
sterilized preparation operations would be of appropriate particle quality during compounding operations.
Air shall be sampled in a manner that characterizes the true level of extrinsic particle contamination to
which the product is exposed. Classified pass-though chambers shall be tested in the Static (At Rest) state.

Apparatus:
A calibrated discrete particle counter with a particle size discrimination capability of 0.5 µm.
Clean surrogate materials and transfer tray

CETA CAG-002-2006 11 12/08/2008


2.10A Procedure- Static Particle Level:
1. Sample ports compatible with particle counting equipment should be used to facilitate accurate
and repeatable testing.
2. Determine particle locations per ISO 14644-1:1999. Note that a reasonable sample plan is one
that addresses the entire work surface and may result in more than minimum sample locations
allowed by ISO 14644-1:1999. An example of a reasonable plan for the Main Chamber is one
where the locations are in each of the four corners (6” (15.2 cm) from interior walls) and one at the
geometric center of the work-area. A reasonable sample plan for the Pass-through would be one
location at the geometric center with a minimum of 3 samples.
3. Determine that particle sampling tube has been appropriately purged of particles.
4. Perform the particle counter background noise count rate (zero count) test prior to sampling.
5. The Compounding Isolator should be empty, clean, and safe for access when tested.
6. Position the particle counter isokinetic probe at the 1st location to sample the air as it reaches the
clean zone, the probe height being typically no more than 12” (30.5 cm) above the work surface.
7. Operate the Compounding Isolator for a period of time appropriate for the design to clean up the
work area.
8. Sample the assigned locations within the space.
9. Note the particle levels at the specified locations.

2.10B Procedure- Dynamic Operating Test (Monitoring):


1. Sample ports compatible with particle counting equipment should be used to facilitate accurate
and repeatable testing.
2. Determine that particle sampling tube has been appropriately purged of particles.
3. Perform the particle counter background noise count rate (zero count) test prior to sampling.
4. Compounding Isolator should contain all of the cleaned surrogate compounding components that
would support surrogate manipulation operation during the test.
5. Particle counter isokinetic probe location(s) should be placed near the direct compounding area
where there is most potential risk to the exposed sterile preparation, containers, and closures. The
particle counting probe should be placed in an orientation demonstrated to obtain a meaningful
sample.
a. Position the particle counter isokinetic probe within the airflow not more than 12 inches
(30.5 cm) away from the location of the Critical Site within the Direct Compounding
Area.
6. Purge the compounding isolator as specified in the operating procedures prior to conducting the
particle count survey. If not specified in the operating procedures, allow the isolator to purge for 5
minutes.
7. To simulate actual preparation, perform surrogate manipulation using both gloves during particle
testing to determine whether activity affects particle levels at the sample point.
8. Document the particle levels.

Acceptance Criteria:
ISO 14644-1:1999 Class 5 at 0.5 µm and larger At Rest and Operational.

INFORMATIONAL NOTE:
Generally, conditions are met when none of the individual counts
exceed the class limit or equivalent of 3,520 particles per cubic meter (ppcm) at
0.5µm size particles and larger.
Where between 2 and 9 locations are sampled, a statistical analysis
of the upper 95% confidence level (UCL) must confirm that sample levels fall
within the acceptance criteria per ISO 14644-1:1999

CETA CAG-002-2006 12 12/08/2008


INFORMATIONAL NOTE:
When selecting a Discrete Particle Counter (DPC), the instrument should have the
ability of displaying and recording the count and size of discrete particles in air with
a size discrimination capability to detect the total particle concentration in the
appropriate particle size ranges for the class under consideration. 5

2.11A Volatile Hazardous Drug Containment Test -Total Exhaust Compounding


Isolator:

Purpose:
This test 10 is to verify that Compounding Aseptic Containment Isolator properly connected to building
exhaust provides worker protection from the escape of volatilized hazardous drugs during all aspects of
compounding operations.

Apparatus:
A calibrated Infrared Portable Ambient Air Analyzer, Electron Capture Device, or equivalent.
Instrument shall be a continuous reading instrument for the tracer gas of choice. Range of
detection shall be at least from 0.01 ppm (parts per million) to 100 ppm. The accuracy of the
instrument shall be ± 10% of the reading for concentrations above 0.1 ppm and ± 25% for
concentrations between 0.01 ppm and 0.1 ppm. The repeatability of the instrument shall be ± 1%
of the reading at 50 ppm tracer gas concentration. The response time shall not exceed 10 seconds
for 90% indication of actual concentration. The instrument shall not exhaust more than 50 lpm
(liters per minute). The detectors shall be calibrated with a known concentration of tracer gas
within 24 hours preceding a test. The method for calibration shall be those furnished or specified
by the detector manufacturer and shall use the tracer gas that was selected as a standard. The
analyzer shall have a strip chart recorder or data logger.
Tracer Gas – Sulfur Hexafluoride (SF6), (or a gas of similar molecular weight and stability,
supplied from a cylinder capable of maintaining 30 psig (207 kPa) at the test release rate for at
least one hour). The tracer gas release rate shall be 4.0 lpm. The tracer gas shall be commercial
grade or reagent grade. Since the detection instrument is calibrated by the actual tracer gas, a
100% pure gas is not required. Low-grade mixtures are inappropriate since they significantly
reduce the limit of detection for the test.
Tracer Gas Ejector System - The tracer gas is piped to the ejector. The trace gas passes through a
critical orifice, entrains air through the holes in the side of the ejector tube, and is distributed
through a wire mesh outlet diffuser.
Critical Orifice - the flow rate of the tracer gas is determined by the upstream pressure and size of
the orifice. The size for the orifice using sulfur hexafluoride as tracer gas, at a flow rate of 4.0
lpm and a nominal upstream pressure of 30 psig (207 kPa) is 0.025 inches (0.635 mm). The
orifice size and pressure will give the approximate flow rate of tracer gas; however, the actual
flow rate must be measured. Convert the measured value to standard conditions: 70°F (21.1°C)
at one atmosphere. The ejector system release rate shall be calibrated within 24 hours preceding
a test and each time the orifice plate is changed.
Flow meter - capability of reading 4 lpm, accuracy of ± 3% of full scale or better calibrated per
manufacturer's specifications.
Pressure Gauge - capability of reading 30 psig (207 kPa), accuracy of ± 2% of full scale calibrated
per manufacturer's specifications.
Surrogate materials and transfer tray.

CETA CAG-002-2006 13 12/08/2008


INFORMATIONAL NOTE: For compounding of volatile drugs, Compounding
Aseptic Containment Isolators which use negative pressure plenums and ducts are
preferred. For those compounding isolators with positive pressure plenums or
ducts, the seam leak test of step #6 should also be performed along all duct
connections and plenum seams.

Procedure:

1. Place surrogate materials into main chamber via pass-through chamber.


2. Record room background levels of SF6 (or tracer gas).
3. Activate SF6 gas flow and verify 4 LPM flow rate through ejector within isolator.
4. Perform surrogate manipulations.
5. Upon conclusion of surrogate manipulation, stop SF6 gas flow and place surrogate materials
in pass-through chamber.
6. Seal inner door and follow established purge time.
7. Upon completion of established purge time, open exterior pass-through door and retrieve
surrogate preparation.
8. Tester takes SF6 samples at the least favorable position within 1” (2.5 cm) of exterior pass-
through chamber door based on Isolator design (i.e. measure just below door on static pass-
through or measure just above door on integral vertical flow pass-through).
9. Remove materials from pass-through chamber.
10. Document results.

Acceptance Criteria:
Tracer gas concentration level shall not exceed 0.01 ppm at any time during these tests.

2.11B Volatile Hazardous Drug Containment Test-Partial-Recirculating


Compounding Isolator:

Purpose:
This test 10 is to verify that the partial-recirculating Compounding Aseptic Containment Isolator that is
properly connected to building exhaust system provides personnel protection from the escape of volatilized
hazardous drugs during all aspects of compounding operations. Due to its design, any volatiles released in
the workzone will partially recirculate and partially exhaust from the building.
Apparatus:
A calibrated Infrared Portable Ambient Air Analyzer, Electron Capture Device, or equivalent.
Instrument shall be a continuous reading instrument for the tracer gas of choice. Range of
detection shall be at least from 0.01 ppm (parts per million) to 100 ppm. The accuracy of the
instrument shall be ± 10% of the reading for concentrations above 0.1 ppm and ± 25% for
concentrations between 0.01 ppm and 0.1 ppm. The repeatability of the instrument shall be ± 1%
of the reading at 50 ppm tracer gas concentration. The response time shall not exceed 10 seconds
for 90% indication of actual concentration. The instrument shall not exhaust more than 50 lpm
(liters per minute). The detectors shall be calibrated with a known concentration of tracer gas
within 24 hours preceding a test. The method for calibration shall be those furnished or specified
by the detector manufacturer and shall use the tracer gas that was selected as a standard. The
analyzer shall have a strip chart recorder or data logger.
Tracer Gas – Sulfur Hexafluoride (SF6), (or a gas of similar molecular weight and stability,
supplied from a cylinder capable of maintaining 30 psig (207 kPa) at the test release rate for at
least one hour). The tracer gas release rate shall be 4.0 lpm. The tracer gas shall be commercial
grade or reagent grade. Since the detection instrument is calibrated by the actual tracer gas, a
100% pure gas is not required. Low-grade mixtures are inappropriate since they significantly
reduce the limit of detection for the test.

CETA CAG-002-2006 14 12/08/2008


Tracer Gas Ejector System - The tracer gas is piped to the ejector. The trace gas passes through a
critical orifice, entrains air through the holes in the side of the ejector tube, and is distributed
through a wire mesh outlet diffuser.
Critical Orifice - the flow rate of the tracer gas is determined by the upstream pressure and size of
the orifice. The size for the orifice using sulfur hexafluoride as tracer gas, at a flow rate of 4.0 lpm
and a nominal upstream pressure of 30 psig (207 kPa) is 0.025 inches (0.635 mm). The orifice
size and pressure will give the approximate flow rate of tracer gas; however, the actual flow rate
must be measured. Convert the measured value to standard conditions: 70°F (21.1°C) at one
atmosphere. The ejector system release rate shall be calibrated within 24 hours preceding a test
and each time the orifice plate is changed.
Flow meter - capability of reading 4 lpm, accuracy of ± 3% of full scale or better calibrated per
manufacturer's specifications.
Pressure Gauge - capability of reading 30 psig (207 kPa), accuracy of ± 2% of full scale calibrated
per manufacturer's specifications.
60 mL syringe filled with tracer gas Sulfur Hexafluoride (SF6) (or a gas of similar molecular
weight and stability) and placed within a sealed plastic bag.

Procedures:

1. Place sealed plastic bag containing SF6 gas-filled 60 mL syringe in main chamber.
2. Record room background levels of SF6.
3. Perform surrogate manipulation by opening bag and releasing 60 mL of tracer gas into workzone.
Start timer upon release of tracer gas. Place syringe back into bag and seal. Place sealed bag into
pass-through chamber, then open outer pass-through door and remove bag.
4. Operator places analyzer detector in least favorable position within 1” (2.5 cm) of exterior pass-
through chamber door based on Isolator design. (i.e. measure just below door on static pass-
through or measure just above door on integral vertical flow pass-through).

NOTE: Either a main chamber or pass-through chamber purge or delay time may be
required to meet acceptance criteria. Repeat Phase 2 with adequate purge or delay time
until acceptance criteria is met.

Acceptance Criteria:
SF6 concentration level shall not exceed 0.01 ppm at any time during these tests.

INFORMATIONAL NOTE:
With the test results, a risk analysis can be performed for the hazardous drug
preparation process equating amount of volatiles generated by the hazardous drug
compounding process to purge time required for the safe removal of compounded
materials.

To assure personnel protection, the test uses a known quantity of volatile tracer gas
(SF6), released into the main chamber airstream, while monitoring the material
transfer process for tracer gas just outside the external interchange door. The test is
designed to determine the volatile purge time based on the known release volume.

The NIOSH Alert: Preventing Occupational Exposures to Antineoplastic and


Other Hazardous Drugs in Health Care Settings (2004) 11 recommends against
the use of recirculating cabinets for the manipulation of volatile hazardous
drugs. This test procedure should not be misconstrued as an endorsement of
recirculating isolators with volatile drugs.

CETA CAG-002-2006 15 12/08/2008


2.12 Hazardous Particle Containment Test:

Purpose: This test is to verify that Compounding Aseptic Containment Isolator provides worker
protection from the escape of hazardous drug particles during all aspects of compounding operations.

NOTE: This test is not required if Section 2.11A or 11B is successfully completed.

Apparatus:
A calibrated discrete particle counter with a particle size discrimination capability of 0.5 µm.
A calibrated aerosol photometer capable of indicating 100% upstream concentration with an
aerosol of between 10 and 90 µg per liter of polydispersed dioctylphthalate (DOP),
polyalphaolefin (PAO) or an equivalent fluid. Unit must have a threshold sensitivity of at least 10 -
3
micrograms per liter and be capable of measuring concentrations over a range of 10 5 times the
threshold sensitivity. The sampling rate shall be 1 CFM (28.3 lpm) (±10%) with an inlet probe
having sufficient area (1.7 square inches) (11.0 cm2) as to maintain a probe inlet velocity of 90
fpm (0.46 m/s) or slightly higher.
Laskin Nozzle aerosol generator or equivalent.
Aerosol diffuser (shown in Figure 1 below)
Particle diluter.
Time keeping device capable of reading in minutes and seconds.
Cleaned preparation transfer tray and safe surrogate compounding manipulation materials.

Figure 1

INFORMATIONAL NOTE:
The aerosol diffuser is used to reduce turbulence caused by the aerosol generator
within the isolator chamber.

The diffuser should be centered between the glove ports and centered between the back
wall and the view screen.

Procedure:

Operational test – material transfer test:


1. This test should be performed in a controlled atmosphere such as a cleanroom. The interior of
isolator main and pass-through chambers should be clean.
2. Place the aerosol diffuser in the isolator main chamber and connect to the aerosol generator.
3. Operate the isolator normally.
4. Place cleaned preparations transfer tray and safe surrogate materials via the pass-through into the
main chamber.
5. Activate aerosol supply.
6. Perform surrogate manipulations.

CETA CAG-002-2006 16 12/08/2008


7. Upon conclusion of surrogate manipulations, stop aerosol generator.
8. Place surrogate materials and tray into pass-through chamber and follow established purge time
using time keeping device to verify.
9. With the particle counter (with installed particle diluter), and photometer test probes, 2.5” (6.4 cm)
directly under the center of the door swing of the exterior pass-through door, monitor particle
levels, remove materials and transfer tray from pass-through chamber via exterior door.
10. Record the particle counts at 0.5 µm and larger (calculate actual particle levels based upon diluter
rating) and photometer readings as the materials are removed from the pass-through chamber and
for a period of 30 seconds after the door is closed. Verify that arm movement in opening the door
does not create a particle stream.

INFORMATIONAL NOTE: To avoid the impact of arm movement creating a particle


stream when opening the door, a non-shedding remote handle or door handle actuator
should be considered.

Acceptance Criteria:

No elevation in particle level during completion of transfer process.

2.13 Pass-through Particle Purge Time Determination Test:

Purpose:

This test is to determine the appropriate purge cycle time to be used after placing materials in the pass-
through prior to transferring them into or out of the main chamber.

Apparatus:

A calibrated discrete particle counter with a particle size discrimination capability of 0.5 µm.
A source of visible smoke that is generally neutrally buoyant or aerosol particulate detectible by
the particle counter. Source is to be determined by the manufacturer.

Procedure for non-unidirectional “Purge” and static pass-through designs:


1. Place the particle counter probe at the geometric center of the pass-through work surface at a height
of 6” (15.2 cm).
2. Place a transfer basket on the pass-through work surface.
3. Measure the ambient particle count concentration in the pass-through.

NOTE: If the ambient count at 0.5 µm and larger is not above the
equivalent of 3,520,000 ppcm, elevate the counts using smoke or
particulate source to achieve at least this level.

4. Initiate purge and particle count cycles simultaneously.


5. Three second particle count samples in concentration mode (normalized to particles per cubic
meter) with a one second delay to be taken until counts are at manufacturer’s recommended levels
for material transfer into the main chamber. This will usually be less than the equivalent of 3,520
ppcm at 0.5 µm and larger particle sizes.
6. Document the time required to get to the manufacturer’s recommended levels. This time period
will be communicated to the end user in the owner’s manual.

CETA CAG-002-2006 17 12/08/2008


Procedure for “constant operation” unidirectional and non-unidirectional pass-through designs:
1. Establish a background room ambient particle concentration at least equivalent to 3,520,000 ppcm
at 0.5 µm and larger. Document the background count.
2. Place the particle counter probe at the geometric center of the pass-through work surface at a
height of 6” (15.2 cm).
3. Open the outer pass-through door and place a transfer basket on the pass-through work surface.
4. Initiate particle count cycle immediately after closing the outer pass-through door.
5. Three second particle count samples in concentration mode (normalized to particles per cubic
meter) with a one second delay to be taken until counts are at manufacturer’s recommended levels
for material transfer into the main chamber. This will usually be less than the equivalent of 3,520
ppcm at 0.5 µm and larger particle sizes.
6. Document the time required to get to the manufacturer’s recommended levels. This time period
will be communicated to the end user in the owner’s manual.

Acceptance Criteria:

The maximum acceptable purge time shall be determined by the manufacturer and confirmed
using this procedure.

CETA CAG-002-2006 18 12/08/2008


3.0 Reference Material:
1
USP 32-NF27: United States Pharmacopeial Convention, Inc., 12601 Twinbrook Parkway, Rockville,
MD 20852, www.usp.org.
2
CAG-001-2005 (revised 12/08/2008): Applications Guide for the use of Compounding Isolators in
Compounding Sterile Preparations in Healthcare Facilities, Controlled Environment Testing Association,
1500 Sunday Drive, Suite 102, Raleigh, NC 27607, www.cetainternational.org.
3
IEST-RP-CC001.4: HEPA and ULPA Filters, Institute of Environmental Sciences and Technology, 5005
Newport Drive, Suite 506, Rolling Meadows, IL 60008, USA, www.iest.org
4
MIL-STD-282: MIL/QQ (Military Standard) Standardization Document Order Desk, 700 Robbins Ave,
Bldg 4, Section D, Philadelphia, PA 19111-5094, http//.dodssp.daps.dla.mil.
5
ISO 14644-1:1999: Cleanrooms and associated controlled environments-Classification of air cleanliness,
International Organization for Standardization, Case Postale 56, CH-1211 Geneve 20, Switzerland,
www.iest.org.
6
NSF/ANSI 49-2008: Class II (laminar flow) Biosafety Cabinetry, NSF International, P.O. Box 130140,
Ann Arbor, MI 48113-0140, www.nsf.org.
7
ANSI/AIHA Z9.5-2003: American Industrial Hygiene Association, 2700 Prosperity Ave., Suite 250,
Fairfax, VA 22031, www.aiha.org
8
IEST-RP-CC034.2: HEPA and ULPA Filter Leak Tests, Institute of Environmental Sciences and
Technology, 5005 Newport Drive, Suite 506, Rolling Meadows, IL 60008, USA, www.iest.org.
9
AGS-G001 – 2007, 2nd Edition: Guideline for Gloveboxes, American Glovebox Society, P.O. Box 9099,
Santa Rosa, CA 95405, www.gloveboxsociety.org.
10
ANSI/ASHRAE 110-1995: Method of Testing Performance of Laboratory Fume Hoods, American
Society of Heating, Refrigerating and Air-Conditioning Engineers, Inc. 1791 Tullie Circle, NE, Atlanta,
GA 30329, www.ashrae.org.
11
NIOSH Alert for Preventing Occupational Exposure to Antineoplastic and Other Hazardous Drugs in
Health Care Settings, Department of Health and Human Services, CDC, NIOSH, NIOSH – Publications
dissemination, 4676 Columbia Parkway, Cincinnati, OH 45226-1998, www.cdc.gov/niosh.

CETA CAG-002-2006 19 12/08/2008


APPENDIX:
The following is an example of surrogate compounding procedures used during Dynamic Particle Counting
procedures (2.10B).

Supplies:
2 x 10 mL syringe with needle
1 x 20 mL Sterile Water for Injection (SWFI) vial
1 x 100 mL bag of Dextrose 5% Water (D5W/NS) or 0.9% Sodium Chloride
4 Sterile Alcohol wipes
Isopropyl Alcohol (IPA) or equivalent disinfectant
1 blank label (to be affixed to the bag prior to placing into pass-through)

1) All supplies introduced into the isolator environment for immediate compounding, are to be
picked from the warehouse or inventory area. All products will be removed from their master
cartons (non-laminated or corrugated cardboard), or non-plastic over wrap prior to leaving the
warehouse or inventory area.
2) After inspection, the products and/or supplies are transferred into the isolator pass-through in
plastic bin or tote.
3) The operator will insert hands into glove ports of the isolator, seat hands in gloves; disinfect the
gloves with an appropriate disinfectant (IPA).
4) The operator will disinfect the interior surfaces of the primary compounding chamber with an
appropriate disinfectant and allow the surfaces to dry after establishing that it is safe to do so
where flammable materials are used.
5) The operator will open the interior pass-through door, retrieve the bin/tote, close the pass-through
door and place the bin/tote in corner of isolator, gather supplies out of bin/tote, and place
immediately in front of the operator.
a. Open one syringe by separating the two halves of the packaging, by slowly peeling the
paper half, from the plastic outer wrap, laying the syringe onto the work surface (place
packaging in the bin/tote).
b. Remove the cap of the SWFI vial and disinfect with IPA pad (place pad and packaging in
bin/tote after use).
c. Pick up the syringe with the free dominant hand.
d. Utilizing just the pinkie finger of the non-dominant hand, dislodge the needle cap, and
pierce the septum of the vial and withdraw 3 mL of SWFI.
e. Remove needle for vial and transfer the SWFI from the syringe into the 100 mL bag of
D5W/NS.
f. Repeat the transfer from the vial into the bag for a total of 5 injections.
6) Upon completion of transfers, place 100 mL bag onto the work surface.
7) Place the syringe and vial of SWFI in the bin/tote.
8) Peel back label and label 100 mL bag with blank label.
9) Place the labeled bag and label backing into the bin/tote.
10) Pick up the bin/tote and open inner pass-through door and place bin/tote into pass-through.
11) Disinfect the interior surfaces of the isolator with the appropriate disinfecting agent.
12) Place cleaning wipes and waste in the pass-through or trash disposal port.
13) Operator will remove hands from glove ports and retrieve bin/tote from pass-through via the outer
pass-through door.

CETA CAG-002-2006 20 12/08/2008

You might also like