CETA-Compounding-Isolator-Testing Guide PDF
CETA-Compounding-Isolator-Testing Guide PDF
CAG-002-2006
Revised December 8, 2008
1.0 Overview:
The purpose of this document is to establish an industry-based minimum set of testing criteria appropriate
for all Compounding Isolators used pursuant to USP Chapter 797 1. Compounding Isolators consist of
Compounding Aseptic Isolators used for compounding sterile preparations and Compounding Aseptic
Containment Isolators used for compounding sterile hazardous drug preparations in pharmacy applications.
While this document gives general guidance or referenced guidance through relevant industry documents, it
is not the intention to set the specific acceptance criteria. It is the manufacturer’s responsibility to
determine exact testing procedures consistent with these guidelines and assign appropriate values
pertaining to acceptance criteria that is consistent with user requirements. This guide has also been
established to create a uniform approach for field certifiers to allow consistent and repeatable testing at all
facilities.
An isolator certified by the manufacturer as meeting the criteria established in CETA CAG-002-2006 shall
post the following on the front of the cabinet in a location readily visible to the user:
1. Notice of pass or failure of the Preparation Ingress and Egress Test (2.09).
2. Whether or not designed for use with volatile hazardous drug compounds, toxic, flammable or
explosive materials.
3. Indication of the results of the Recovery Time Determination Test (2.07).
NOTE: Instructions for material transfer should be developed by individual user/ facility, based
on their SOPs taking into account purge and recovery times listed by the isolator manufacturer.
a. Purge time for the pass through
b. Purge time after materials are entered into the isolator
An isolator field certified to the statement “in accordance to CETA CAG-002-2006” shall have the
following information posted on the front of the cabinet in a location readily visible to the user:
A. Company name
B. Address
C. Phone number
X = Mandatory
O = Optional
D = Design criteria qualification
Ambient
Environment external to the compounding isolator
Critical Site:
A location that includes any component or fluid pathway surfaces (e.g. vial septa, injection ports,
beakers) or openings (e.g. opened ampule, needle hubs) exposed and at risk of direct contact with
air (e.g. ambient room or HEPA filtered), moisture (e.g. oral and mucosal secretions), or touch
contamination. Risk of microbial particulate contamination of the critical site increases with the
size of the openings and exposure time.
Main Chamber:
The primary chamber containing the work area, critical work zone, and means for manipulation of
the preparation. This includes plenums and areas covered with diffusers.
Pass-through Chamber:
The chamber designed to separate the surrounding ambient space from the main chamber while
facilitating the movement of materials between the two areas.
Particle Elevation:
Airborne room particles elevation to enable certain tests. This elevation may be accomplished
through the use of aerosol generator, theatrical smoke generator, or other smoke sources that
provide suitable particle size and quantity
Surrogate Manipulation:
A manipulation of a preparation that involves the use of substituted materials that closely resemble
actual materials used in sterile compounding to provide a similar impact upon the environment
tested and to resemble actual operational conditions during testing
Unidirectional Airflow:
An airflow moving in a single direction in a robust and uniform manner and at sufficient speed to
reproducibly sweep particles away from the critical processing or testing area.
Work Area
The horizontal and vertical space within the isolator from wall to wall and from the base or floor
to the top within the main chamber. This excludes plenums and areas covered with diffusers.
The manufacturer shall be responsible for identifying special accommodations, or ancillary equipment that
may be needed for accurately field testing the Compounding Isolator design. These may include, but are
not limited to the methods, sources, or devices to provide accurate and repeatable testing such as:
INFORMATIONAL NOTE:
Metric equivalents (SI) are provided in parentheses ( ) behind English
measurement values throughout this document as an aid or cross-reference.
Apparatus:
A calibrated thermal anemometer with an accuracy of ± 3.0 fpm (0.015 m/s) or ± 3% of the
indicated velocity, whichever is greater.
Procedure:
Main Chamber:
Measure the air velocity in the Compounding Isolator using equally spaced points in the horizontal plane
that produces the most repeatable readings.
2. For non-unidirectional purged pass-through, the manufacturer will determine the optimum
location and method that produces the most repeatable readings or determine an alternate
repeatable method to determine that the specified purge objective is met
Acceptance Criteria:
The manufacturer will determine the appropriate acceptance criteria for the particular design.
Unidirectional Airflow Compounding Isolator velocity criteria shall be expressed in terms of a
range of velocities in feet per minute (fpm) (meters per second (m/s)) along with a velocity
uniformity designator. For example, a statement would be presented as the average velocity shall
fall between __fpm (m/s) and __fpm (m/s) and __% uniformity from the average
Apparatus:
A calibrated mechanical or electronic differential pressure indicating device.
Procedure:
1. Utilize the factory installed test port or sealed access location to main and pass-through chambers.
2. Attach test instrument tubing into chamber and allow pressure levels to stabilize.
Test 1:
Record pressure level of main chamber with outer and inner pass-through doors closed and
isolator gloves / gauntlets extended into Compounding Isolator.
Fully open outer pass-through door. Document any change to the pressure.
Confirm that the pressure does not change from positive to negative or from negative to positive.
Close outer pass-through door. Document any change to the pressure.
Test 2:
Record pressure level of main chamber with outer and inner pass-through doors closed and
isolator gloves / gauntlets extended into isolator.
Using Compounding Isolator glove, fully open inner door. Document the change to the pressure.
Confirm that the main chamber pressure does not change from positive to negative or from
negative to positive.
Close inner door. Determine if pressure level returns to original levels.
Test 3a:
Record pressure level of Compounding Aseptic Isolator main chamber with outer and inner pass-
through doors closed and isolator gloves / gauntlets extended into isolator.
Insert hands into gloves and pull back from isolator to simulate operator hand removal. An
example of a reasonable pull back time is no greater than 3 seconds from fully extended to fully
extracted (gloves at a point even with the view panel)
Determine the change in pressure and verify that the pressure does not change from positive to
negative.
Test 3b:
Record pressure level of Compounding Aseptic Containment Isolator main chamber with outer
and inner pass-through doors closed and isolator gloves / gauntlets extended outside isolator.
Insert hands into gloves and push into main chamber to simulate operator hand insertion.
Determine the change in pressure and verify that the pressure does not change from negative to
positive.
Acceptance:
Compounding Isolator operating pressure range shall be determined by the manufacturer.
The pressure shall not change from positive to negative or from negative to positive during any
manipulations.
Purpose:
These tests are performed to verify:
The airflow or pressure setpoint(s) where an audible and/or visual alarm will activate to
signify unfavorable operating conditions within the compounding isolator and/or the remote
exhaust blower.
Functional pass-through door interlock operation.
Proper canopy or exhaust connection performance.
Procedure:
Acceptance:
Airflow or Pressure Alarm Test
The setpoint(s) shall be determined by the manufacturer as appropriate for operator safety, product
sterility and/or containment.
Apparatus:
A calibrated thermal anemometer with an accuracy of ± 3.0 fpm (0.015 m/s) or 3% of the
indicated velocity, whichever is greater.
Procedure:
1. Determine that all other operating parameters have been met.
2. Remove one gauntlet from chamber view screen.
3. Measure velocity at the geometric center of the opening. A minimum of 3 samples at a minimum
of 1 location shall be taken.
4. Record results.
Apparatus:
A calibrated Aerosol Photometer capable of indicating 100% upstream concentration with an
aerosol of between 10 and 90 µg per liter of polydispersed dioctylphthalate (DOP or DEHS) CAS#
117-81-7, polyalphaolefin (PAO) CAS# 68649-12-7 or an equivalent fluid. Unit must have a
threshold sensitivity of at least 10-3 micrograms per liter and be capable of measuring
concentrations over a range of 105 times the threshold sensitivity. The sampling rate shall be 1
CFM (28.3 lpm) (±10%) with an inlet probe having sufficient area (1.7 square inches) (11.0 cm 2)
as to maintain a probe inlet velocity of 90 fpm (0.46 m/s) or slightly higher.
Laskin Nozzle aerosol generator or equivalent.
Procedure:
Main and Pass-through Chamber HEPA filters:
Acceptance Criteria:
For filters that can be scanned:
Sustained aerosol penetration shall not exceed 0.01% of the upstream concentration.
Filter patch size shall not exceed 3% of the effective filter area of the side being patched. The maximum
width of any one patch shall not exceed 1.5 inches (3.8 cm).
Acceptance:
Any detected leaks do not exceed ISO class 5 conditions of 3,520 particles per cubic meter (ppcm)
at 0.5µm size particles and larger when the probe is held stationary within 1” (2.5 cm) of the leak.
INFORMATIONAL NOTE:
Elevating operating pressures during this test may aid in locating leak sources.
Procedure:
1. Set the particle counter sample time to 6 second sample periods with a 14 second hold time in
“Concentration” mode to report in particles per cubic meter or foot.
2. With the Compounding Isolator running, measure particle levels at the center of the work-surface
of the main chamber.
3. Collect multiple samples to determine average baseline particle levels using 6 second samples.
4. Three consecutive baseline particle level samples should be established and noted.
5. Turn off particle counter sample pump and cover the particle counter sampling probe.
6. Turn off Compounding Isolator and fill the chamber with particulate using a Laskin nozzle
generator set at 1 Laskin nozzle at 20 psi (138 kPa) for five seconds per cubic foot (per 0.028
cubic meter) of chamber interior space.
7. Turn on the Compounding Isolator and start timer.
8. To prevent sampling above the particle counter’s coincidence loss rate or damaging the device,
wait until the smoke is visibly cleared from the chamber and remove particle counter probe cover
and begin sampling.
9. Maintained particle levels are achieved when three consecutive counts are at or below the
originally determined baseline particle level.
10. Total recovery time is considered from Compounding Isolator blower turn-on time to the first
particle count where maintained particle levels were achieved.
Acceptance criteria:
The manufacturer will determine the appropriate acceptance criteria for their particular design.
Recovery time will be expressed in minutes and seconds to recover to baseline from excursion.
Apparatus:
A source of visible smoke that is generally neutrally buoyant
Procedure:
1. Pass the smoke along the work area beginning 1 inch (2.5 cm) from all main chamber walls and
view screens at a height determined to authenticate the purpose of test.
2. Pass smoke over gloves, gauntlets, IV bars, interior lights, or other extruding features that may be
installed within the main chamber.
Acceptance criteria:
The smoke shall show smooth downflow with a minimum of dead spots or reflux (upward flow) at
obstructions and across the critical work zone and it is removed to the returns without reentry.
Apparatus:
A calibrated discrete particle counter with a particle size discrimination capability of 0.5 µm.
Aerosol generator or smoke tubes.
Empty perforated transfer tray.
Time keeping device capable of reading in seconds.
Procedure:
1. Verify the background count in the testing room is at least 3,532,000 particles per cubic meter
(ppcm) (100,000 particles per cubic foot (ppcf)).
2. If the count is too low, elevate the background levels using an aerosol generator or smoke
generator.
3. Place the particle counter probe in the Main Chamber approximately 6 to 8 inches (15.2 to 20.3
cm) off the isolator floor surface, approximately 2 inches (5.1 cm) outside the normally used path
of the inner pass-through chamber door swing and within the area that would be affected by the
airflow caused by the movement of the door. Probe placement should be so that the operator’s
arms will not pass directly over the probe when removing material from the pass-through.
4. Verify the particle counts meet ISO Class 5 levels before beginning the test cycle.
5. Set the particle counter for a one minute count with no more than a one second hold time.
6. Open the outside pass-through door.
7. Place a perforated empty transfer tray into the pass-through and close the outer door.
8. Wait for the manufacturer’s recommended purge time or start the purge cycle.
9. After completion of the purge cycle, open the inside pass-through door and move the transfer tray
from the pass-through to the work area.
10. Close the inside pass-through door.
11. Document the particle counts during the transfer process and for a period of one minute after the
transfer.
Acceptance Criteria:
The particle counts shall not exceed the ISO Class 5 @ 0.5 µm and larger class limit at any time
during this test.
Apparatus:
A calibrated discrete particle counter with a particle size discrimination capability of 0.5 µm.
Clean surrogate materials and transfer tray
Acceptance Criteria:
ISO 14644-1:1999 Class 5 at 0.5 µm and larger At Rest and Operational.
INFORMATIONAL NOTE:
Generally, conditions are met when none of the individual counts
exceed the class limit or equivalent of 3,520 particles per cubic meter (ppcm) at
0.5µm size particles and larger.
Where between 2 and 9 locations are sampled, a statistical analysis
of the upper 95% confidence level (UCL) must confirm that sample levels fall
within the acceptance criteria per ISO 14644-1:1999
Purpose:
This test 10 is to verify that Compounding Aseptic Containment Isolator properly connected to building
exhaust provides worker protection from the escape of volatilized hazardous drugs during all aspects of
compounding operations.
Apparatus:
A calibrated Infrared Portable Ambient Air Analyzer, Electron Capture Device, or equivalent.
Instrument shall be a continuous reading instrument for the tracer gas of choice. Range of
detection shall be at least from 0.01 ppm (parts per million) to 100 ppm. The accuracy of the
instrument shall be ± 10% of the reading for concentrations above 0.1 ppm and ± 25% for
concentrations between 0.01 ppm and 0.1 ppm. The repeatability of the instrument shall be ± 1%
of the reading at 50 ppm tracer gas concentration. The response time shall not exceed 10 seconds
for 90% indication of actual concentration. The instrument shall not exhaust more than 50 lpm
(liters per minute). The detectors shall be calibrated with a known concentration of tracer gas
within 24 hours preceding a test. The method for calibration shall be those furnished or specified
by the detector manufacturer and shall use the tracer gas that was selected as a standard. The
analyzer shall have a strip chart recorder or data logger.
Tracer Gas – Sulfur Hexafluoride (SF6), (or a gas of similar molecular weight and stability,
supplied from a cylinder capable of maintaining 30 psig (207 kPa) at the test release rate for at
least one hour). The tracer gas release rate shall be 4.0 lpm. The tracer gas shall be commercial
grade or reagent grade. Since the detection instrument is calibrated by the actual tracer gas, a
100% pure gas is not required. Low-grade mixtures are inappropriate since they significantly
reduce the limit of detection for the test.
Tracer Gas Ejector System - The tracer gas is piped to the ejector. The trace gas passes through a
critical orifice, entrains air through the holes in the side of the ejector tube, and is distributed
through a wire mesh outlet diffuser.
Critical Orifice - the flow rate of the tracer gas is determined by the upstream pressure and size of
the orifice. The size for the orifice using sulfur hexafluoride as tracer gas, at a flow rate of 4.0
lpm and a nominal upstream pressure of 30 psig (207 kPa) is 0.025 inches (0.635 mm). The
orifice size and pressure will give the approximate flow rate of tracer gas; however, the actual
flow rate must be measured. Convert the measured value to standard conditions: 70°F (21.1°C)
at one atmosphere. The ejector system release rate shall be calibrated within 24 hours preceding
a test and each time the orifice plate is changed.
Flow meter - capability of reading 4 lpm, accuracy of ± 3% of full scale or better calibrated per
manufacturer's specifications.
Pressure Gauge - capability of reading 30 psig (207 kPa), accuracy of ± 2% of full scale calibrated
per manufacturer's specifications.
Surrogate materials and transfer tray.
Procedure:
Acceptance Criteria:
Tracer gas concentration level shall not exceed 0.01 ppm at any time during these tests.
Purpose:
This test 10 is to verify that the partial-recirculating Compounding Aseptic Containment Isolator that is
properly connected to building exhaust system provides personnel protection from the escape of volatilized
hazardous drugs during all aspects of compounding operations. Due to its design, any volatiles released in
the workzone will partially recirculate and partially exhaust from the building.
Apparatus:
A calibrated Infrared Portable Ambient Air Analyzer, Electron Capture Device, or equivalent.
Instrument shall be a continuous reading instrument for the tracer gas of choice. Range of
detection shall be at least from 0.01 ppm (parts per million) to 100 ppm. The accuracy of the
instrument shall be ± 10% of the reading for concentrations above 0.1 ppm and ± 25% for
concentrations between 0.01 ppm and 0.1 ppm. The repeatability of the instrument shall be ± 1%
of the reading at 50 ppm tracer gas concentration. The response time shall not exceed 10 seconds
for 90% indication of actual concentration. The instrument shall not exhaust more than 50 lpm
(liters per minute). The detectors shall be calibrated with a known concentration of tracer gas
within 24 hours preceding a test. The method for calibration shall be those furnished or specified
by the detector manufacturer and shall use the tracer gas that was selected as a standard. The
analyzer shall have a strip chart recorder or data logger.
Tracer Gas – Sulfur Hexafluoride (SF6), (or a gas of similar molecular weight and stability,
supplied from a cylinder capable of maintaining 30 psig (207 kPa) at the test release rate for at
least one hour). The tracer gas release rate shall be 4.0 lpm. The tracer gas shall be commercial
grade or reagent grade. Since the detection instrument is calibrated by the actual tracer gas, a
100% pure gas is not required. Low-grade mixtures are inappropriate since they significantly
reduce the limit of detection for the test.
Procedures:
1. Place sealed plastic bag containing SF6 gas-filled 60 mL syringe in main chamber.
2. Record room background levels of SF6.
3. Perform surrogate manipulation by opening bag and releasing 60 mL of tracer gas into workzone.
Start timer upon release of tracer gas. Place syringe back into bag and seal. Place sealed bag into
pass-through chamber, then open outer pass-through door and remove bag.
4. Operator places analyzer detector in least favorable position within 1” (2.5 cm) of exterior pass-
through chamber door based on Isolator design. (i.e. measure just below door on static pass-
through or measure just above door on integral vertical flow pass-through).
NOTE: Either a main chamber or pass-through chamber purge or delay time may be
required to meet acceptance criteria. Repeat Phase 2 with adequate purge or delay time
until acceptance criteria is met.
Acceptance Criteria:
SF6 concentration level shall not exceed 0.01 ppm at any time during these tests.
INFORMATIONAL NOTE:
With the test results, a risk analysis can be performed for the hazardous drug
preparation process equating amount of volatiles generated by the hazardous drug
compounding process to purge time required for the safe removal of compounded
materials.
To assure personnel protection, the test uses a known quantity of volatile tracer gas
(SF6), released into the main chamber airstream, while monitoring the material
transfer process for tracer gas just outside the external interchange door. The test is
designed to determine the volatile purge time based on the known release volume.
Purpose: This test is to verify that Compounding Aseptic Containment Isolator provides worker
protection from the escape of hazardous drug particles during all aspects of compounding operations.
NOTE: This test is not required if Section 2.11A or 11B is successfully completed.
Apparatus:
A calibrated discrete particle counter with a particle size discrimination capability of 0.5 µm.
A calibrated aerosol photometer capable of indicating 100% upstream concentration with an
aerosol of between 10 and 90 µg per liter of polydispersed dioctylphthalate (DOP),
polyalphaolefin (PAO) or an equivalent fluid. Unit must have a threshold sensitivity of at least 10 -
3
micrograms per liter and be capable of measuring concentrations over a range of 10 5 times the
threshold sensitivity. The sampling rate shall be 1 CFM (28.3 lpm) (±10%) with an inlet probe
having sufficient area (1.7 square inches) (11.0 cm2) as to maintain a probe inlet velocity of 90
fpm (0.46 m/s) or slightly higher.
Laskin Nozzle aerosol generator or equivalent.
Aerosol diffuser (shown in Figure 1 below)
Particle diluter.
Time keeping device capable of reading in minutes and seconds.
Cleaned preparation transfer tray and safe surrogate compounding manipulation materials.
Figure 1
INFORMATIONAL NOTE:
The aerosol diffuser is used to reduce turbulence caused by the aerosol generator
within the isolator chamber.
The diffuser should be centered between the glove ports and centered between the back
wall and the view screen.
Procedure:
Acceptance Criteria:
Purpose:
This test is to determine the appropriate purge cycle time to be used after placing materials in the pass-
through prior to transferring them into or out of the main chamber.
Apparatus:
A calibrated discrete particle counter with a particle size discrimination capability of 0.5 µm.
A source of visible smoke that is generally neutrally buoyant or aerosol particulate detectible by
the particle counter. Source is to be determined by the manufacturer.
NOTE: If the ambient count at 0.5 µm and larger is not above the
equivalent of 3,520,000 ppcm, elevate the counts using smoke or
particulate source to achieve at least this level.
Acceptance Criteria:
The maximum acceptable purge time shall be determined by the manufacturer and confirmed
using this procedure.
Supplies:
2 x 10 mL syringe with needle
1 x 20 mL Sterile Water for Injection (SWFI) vial
1 x 100 mL bag of Dextrose 5% Water (D5W/NS) or 0.9% Sodium Chloride
4 Sterile Alcohol wipes
Isopropyl Alcohol (IPA) or equivalent disinfectant
1 blank label (to be affixed to the bag prior to placing into pass-through)
1) All supplies introduced into the isolator environment for immediate compounding, are to be
picked from the warehouse or inventory area. All products will be removed from their master
cartons (non-laminated or corrugated cardboard), or non-plastic over wrap prior to leaving the
warehouse or inventory area.
2) After inspection, the products and/or supplies are transferred into the isolator pass-through in
plastic bin or tote.
3) The operator will insert hands into glove ports of the isolator, seat hands in gloves; disinfect the
gloves with an appropriate disinfectant (IPA).
4) The operator will disinfect the interior surfaces of the primary compounding chamber with an
appropriate disinfectant and allow the surfaces to dry after establishing that it is safe to do so
where flammable materials are used.
5) The operator will open the interior pass-through door, retrieve the bin/tote, close the pass-through
door and place the bin/tote in corner of isolator, gather supplies out of bin/tote, and place
immediately in front of the operator.
a. Open one syringe by separating the two halves of the packaging, by slowly peeling the
paper half, from the plastic outer wrap, laying the syringe onto the work surface (place
packaging in the bin/tote).
b. Remove the cap of the SWFI vial and disinfect with IPA pad (place pad and packaging in
bin/tote after use).
c. Pick up the syringe with the free dominant hand.
d. Utilizing just the pinkie finger of the non-dominant hand, dislodge the needle cap, and
pierce the septum of the vial and withdraw 3 mL of SWFI.
e. Remove needle for vial and transfer the SWFI from the syringe into the 100 mL bag of
D5W/NS.
f. Repeat the transfer from the vial into the bag for a total of 5 injections.
6) Upon completion of transfers, place 100 mL bag onto the work surface.
7) Place the syringe and vial of SWFI in the bin/tote.
8) Peel back label and label 100 mL bag with blank label.
9) Place the labeled bag and label backing into the bin/tote.
10) Pick up the bin/tote and open inner pass-through door and place bin/tote into pass-through.
11) Disinfect the interior surfaces of the isolator with the appropriate disinfecting agent.
12) Place cleaning wipes and waste in the pass-through or trash disposal port.
13) Operator will remove hands from glove ports and retrieve bin/tote from pass-through via the outer
pass-through door.