User Manual 4644
User Manual 4644
Kommunikationsnetze
Communication Networks
Rohrleitungsnetze
Water Networks
Leitungsortung
Line Locating
Issue: 4 (07/2009)
ENG
3
Consultation with SebaKMT
The present system manual has been designed as an operating guide and for reference. It is
meant to answer your questions and solve your problems in as fast and easy a way as possible.
Please start with referring to this manual should any trouble occur.
In doing so, make use of the table of contents and read the relevant paragraph with great
attention. Furthermore, check all terminals and connections of the instruments involved.
Should any question remain unanswered, please contact:
Seba Dynatronic Hagenuk KMT
Mess- und Ortungstechnik GmbH Kabelmesstechnik GmbH
Dr.-Herbert-Iann-Str. 6 Röderaue 41
D - 96148 Baunach D - 01471 Radeburg / Dresden
Phone: +49 / 9544 / 68 – 0 Phone: +49 / 35208 / 84 – 0
Fax: +49 / 9544 / 22 73 Fax: +49 / 35208 / 84 249
E-Mail: [email protected]
http://www.sebakmt.com
SebaKMT
All rights reserved. No part of this handbook may be copied by photographic or other means unless SebaKMT have
before-hand declared their consent in writing. The content of this handbook is subject to change without notice.
SebaKMT cannot be made liable for technical or printing errors or shortcomings of this handbook. SebaKMT also
disclaim all responsibility for damage resulting directly or indirectly from the delivery, supply, or use of this matter.
4
Terms of Warranty
SebaKMT accept responsibility for a claim under warranty brought forward by a customer for a
product sold by SebaKMT under the terms stated below.
SebaKMT warrant that at the time of delivery SebaKMT products are free from manufacturing
or material defects which might considerably reduce their value or usability. This warranty does
not apply to faults in the software supplied. During the period of warranty, SebaKMT agree to
repair faulty parts or replace them with new parts or parts as new (with the same usability and
life as new parts) according to their choice.
SebaKMT reject all further claims under warranty, in particular those from consequential
damage. Each component and product replaced in accordance with this warranty becomes the
property of SebaKMT.
All warranty claims versus SebaKMT are hereby limited to a period of 12 months from the date
of delivery. Each component supplied by SebaKMT within the context of warranty will also be
covered by this warranty for the remaining period of time but for 90 days at least.
Each measure to remedy a claim under warranty shall exclusively be carried out by SebaKMT
or an authorized service station.
To register a claim under the provisions of this warranty, the customer has to complain about
the defect, in case of an immediately detectable fault within 10 days from the date of delivery.
This warranty does not apply to any fault or damage caused by exposing a product to
conditions not in accordance with this specification, by storing, transporting, or using it
improperly, or having it serviced or installed by a workshop not authorized by SebaKMT. All
responsibility is disclaimed for damage due to wear, will of God, or connection to foreign
components.
For damage resulting from a violation of their duty to repair or re-supply items, SebaKMT can
be made liable only in case of severe negligence or intention. Any liability for slight negligence
is disclaimed.
5
Table of Contents
4 Design ................................................................................................................................ 11
6 Start-up............................................................................................................................... 12
6.1 Protective Earthing................................................................................................................. 12
6.2 Preparation the Cable to be Identified .................................................................................... 13
6.3 Power Supply ......................................................................................................................... 16
6.4 Switching on .......................................................................................................................... 16
7 Operation ........................................................................................................................... 17
7.1 Check Measurement .............................................................................................................. 17
7.2 Performing a Cable Identification ........................................................................................... 18
7.3 Erroneous Measurements ...................................................................................................... 19
7.4 Phase Determination During Cable Installation ...................................................................... 20
6
1 Safety Advice
Safety This manual contains basic advice for the installation and operation of the
precautions CI TX. It is essential to make this manual accessible to the authorised and
skilled operator. He needs to read this manual closely. The manufacturer is
not liable for damage to material or humans due to non-observance of the
instructions and safety advices provided by this manual.
Locally applying regulations have to be observed.
Working with All electrical regulations of the country where the system is operated have
equipment of to be observed as well as national regulations for prevention of accidents
SebaKMT and existing regulations for the safety and operation of equipment of the
involved companies.
Original accessories ensure safe operation of the equipment. It is not
allowed and the warranty is lost if other accessories than the original ones
are used with the equipment.
Intended The cable identifying generator CI TX may only be operated at
application de-energised low voltage cables and medium voltage cables,
measurement category CAT I (EN 61010-1), according to its intended
application.
Safe operation is only realised when using the equipment for its intended
purpose.
The limits described under technical data may not be exceeded.
Five safety rules The following five safety rules must always be followed when connecting
the CI TX to a cable to be identified:
1. De-energise
2. Protect against re-energising
3. Confirm absence of voltage
4. Ground and short-circuit
5. Cover up or bar-off neighbouring energised parts
7
2 Technical Description
Necessity The positive identification of a single cable from a bunch poses a technical
problem which is often faced by cable technicians.
The purpose of a cable identification is to avoid damage to cables in
service with utmost reliability and to protect the operating personnel.
It must however be mentioned right from the start, that a cable
identification, no matter how reliable it seems to be, should never prompt
the operator to neglect the prescribed VDE safety regulations.
Prior to cutting a cable, it must be made sure that the cable core is
earthed. This can only be achieved by means of an adequate safety
cutting system.
Function The cable identifying generator CI TX can be used for the selectively
identification of low voltage cables and medium voltage cables.
For such a cable identification procedure, the receiver CI RX is required
additionally. The handling of the receiver is described in a separate
operation manual.
The impulse generator sends out pulses up to a peak current value of 100
A into the cable to be identified. This test current generates an
electromagnetic field around the cable which is picked up by a flexible
identification clamp attached to the cable.
The test current of these identification permits a determination of the
current value and of the direction of the measuring pulse, thus leading to a
safe and reliable identification of the cable.
8
Technical Data Parameter Value
Indicators o Signal quality LED (red/yellow/green)
o Power / battery status LED (red/yellow/green)
Pulse voltage 55 VDC
Pulse current Max. 100 A
Pulse sequence 30 per minute
Pulse width 72 ms
Power supply
external: 100 V … 240 VAC, 50/60 Hz
internal: 12 VDC (optional accumulator)
Operation time 4h
(accumulator)
Charging time 6h
(accumulator)
Weight
without accumulator 0.8 kg
with accumulator 1.6 kg
Dimensions 201 mm x 120 mm x 80 mm
Protection class IP 54
Operating -10°C … 60°C
temperature
Measurement Power Supply: 300V / CAT II
Category Measurement connection: CAT I
(EN 61010-1) (de-energized cable)
9
3 Scope of Delivery
Standard scope The following items are included with the standard shipment of the
of delivery generator:
• Mains supply lead 2.0 m (NKG1)
• Test lead 2.0 m, black (MK31-B)
• Test lead 2.0 m, red (MK32-B)
• Earthing lead 1.5 m (EK4)
• Clamp, black (AK41-B)
• Clamp, red (AK42-B)
• Earthing clamp, green/yellow (AK49-B)
Optional In addition to the standard shipment, the following optional items are
accessories available:
• a case which is suitable for a complete cable identifier system
(including the generator LCI TX and the receiver CI RX)
• internal accumulator for operation with 12 VDC
10
4 Design
4 3
5 2
6
7
8 1
Item Description
1 Signal quality LED (red/yellow/green)
2 Fuse
3 Socket for screen connection
4 Socket for core connection
5 Socket for earthing connection
6 Power / battery status LED (red/yellow/green)
7 Mains connection (100 V … 230 V, 50/60 Hz)
8 On/off button
11
5 Safety Mechanisms
Overvoltage The generator has a built-in overvoltage protection. In case, the generator is
protection connected to a live cable - contrary to the safety advises - the fuse [ 2 ]
blows in order to prevent the unit from being damage (but no warranty can
be given).
After the overvoltage protection has been responded, it is necessary to
change the fuse (F 5/250 E) using a screwdriver in order to put the
generator back to operating state.
Over-temperature At too high temperatures, the pulse transmission is automatically stopped
protection until the temperature has dropped below a certain threshold. If the over-
temperature protection is active, the signal quality LED [ 1 ] is lit red
permanently and the audible indicator does not sound.
6 Start-up
Protective Earthing Inspite of the fact that the identification generator does not deliver
dangerous voltages, the safety practices must not be neglected.
On mains operation, protective earthing of the generator is provided from
the earth contact mains connection lead. Caution! In many substations, the
earth conductors in the sockets are not connected in order to avoid hum
pick up during measurements. Moreover, most of these sockets are not
properly marked.
If the instrument is operated from the optional accumulator, then the
earthing socket [ 5 ] has to be connected to system earth via the supplied
earthing lead (EK4).
12
6.2 Preparation the Cable to be Identified
Preparation the Both ends of the cable have to be disconnected and the far end has to be
cable to be connected to system earth.
identified
A direct connection between the end of the core and the screen has to be
avoided, since in this method the identification clamp can only evaluate the
differential current from the outgoing and returning currents. The two fields
generated by the outgoing and returning currents would cancel out and
could not be measured. This has to be taken into account especially when
dealing with new cable installations which are not yet in service. The return
current has to flow through the soil or on neighbouring cable sheaths or
neutrals.
Coupling with The generator has to be connected to a screened cable as shown in the
screened cables figure below:
13
Coupling with If the cable does not have a screen (e.g. NYY) or the system earth is not
unscreened accessible at the cable end, then one should proceed as shown in the figure
cables below. The far end of the core is connected directly to an existing earthing
system, e.g. a lightning protective system. At the near end of the cable, the
identification generator is connected to the core to be identified and to a
different earthing system.
14
Connection The generator can also be connected between two phases of a multi-
between two conductor cable. With this type of connection, the identification of the cable
phases is performed according to the “Twisted-Field” method by means of the
optional TFS CI sensor (moved along or around the cable).
The black and the red test lead must be connected to any two phases of the
cable. These two phases must be bridged at the far end.
15
6.3 Power Supply
Mains connection For mains supply, the mains socket [ 7 ] is connected to a earthing
contact socket (100 V … 240 VAC, 50/60 Hz) by means of the mains
connection lead NKG 1.
Operation with No special measures are required, except the separate protective
accumulator conductor connection. It should however be noted that the maximum
(optional) operating time of the identification generator is about 4 hrs, depending on
the output current. If necessary, charge the battery prior to the
measurement.
6.4 Switching on
After the test leads are connected to the cable to be identified, the
generator can be switched on using the On/off button [ 8 ].
Afterwards, the power / battery status LED [ 6 ] indicates the power supply
status:
Mains operation: green
Battery operation green fully charged
yellow partially discharged
red almost completely discharged
The signal quality LED [ 1 ] indicates the current output and, in this way,
the quality of the signal coupling into the cable to be identified:
Flashing green Good coupling conditions (>30A)
Flashing yellow Increased impedance (30 A < I < 10 A)
Flashing red Unsure coupling conditions (<10 A)
Permanently red No signal transmission
16
7 Operation
17
7.2 Performing a Cable Identification
At the location where the cable is to be identified from a bunch, the clamp
is attached to each individual cable with the arrow pointing towards the far
end, until a distinct deflection of the green LED bar is obtained on the
receiver. If required, readjust the sensitivity. Neighbouring cables should
cause a deflection of the red LED bar.
The following figure illustrates the procedure:
18
7.3 Erroneous Measurements
A cable identification in power cables does in no way replace the need for
determining whether the cable is dead as called for in the five safety
rules. However, a cable identification will most probably prevent a live
cable from being cut. This avoids danger to human life and in some
cases enormous material damage – including power failure.
Even reliably identified cables must not be directly cut or opened, since
despite clear deflections, the cable could be transposed. The figure below
shows the cause of such an erroneous measurement.
At the test points A and B different directions are indicated due to a loop
in the cable. This can lead to erroneous measurements, especially when
measuring in the vicinity of substations where sometimes cable loops are
installed.
19
7.4 Phase Determination During Cable Installation
Measurement 6 Measurement 3
Measurement 5 Measurement 2
Measurement 4 Measurement 1
20
When measuring at the near end of the cable, the core carrying the test
current can then be easily identified by a deflection of the green LED bar
(provided the clamp is aligned in the direction as shown in the figure).
The technician there will then notify his colleague via radio of the
designation of the phase conductor showing the deflection. Subsequently,
this measurement is also to be carried out on the two other phase
conductors (with the clamp aligned in the same direction).
One of the conductors should not carry a signal while the other
measurement should result in a deflection of the red LED bar. In this way,
the two remaining phase conductors are identified one after the other. A
transposition is not possible.
The phase position of the individual conductors to the far end is determined
using the same method.
It is important to note that in this test method, the short-circuiting and
earthing devices at both cable ends remain connected. They are
indispensable for this test method.
The accumulator is charged via the built-in charger. During the charging
process, the instrument should be switched off. The charging time is
automatically limited, i.e. when the accumulator is fully charged, the unit
switches over to trickle charge thus avoiding an overcharge.
The charging time is max. 6 hours and depends on the state of the
accumulator.
The power / battery status LED [ 6 ] indicates the status of the charging
process:
Yellow Accumulator is charging
Green Accumulator is fully charged
21
22
Operation manual
Cable Identifying Generator
LCI TX / LCI TX-440
Kommunikationsnetze
Communication Networks
Rohrleitungsnetze
Water Networks
Leitungsortung
Line Locating
Issue: 4 (07/2010)
3
Consultation with SebaKMT
The present system manual has been designed as an operating guide and for reference. It is
meant to answer your questions and solve your problems in as fast and easy a way as possible.
Please start with referring to this manual should any trouble occur.
In doing so, make use of the table of contents and read the relevant paragraph with great
attention. Furthermore, check all terminals and connections of the instruments involved.
Should any question remain unanswered, please contact:
Seba Dynatronic Hagenuk KMT
Mess- und Ortungstechnik GmbH Kabelmesstechnik GmbH
Dr.-Herbert-Iann-Str. 6 Röderaue 41
D - 96148 Baunach D - 01471 Radeburg / Dresden
Phone: +49 / 9544 / 68 – 0 Phone: +49 / 35208 / 84 – 0
Fax: +49 / 9544 / 22 73 Fax: +49 / 35208 / 84 249
E-Mail: [email protected]
http://www.sebakmt.com
SebaKMT
All rights reserved. No part of this handbook may be copied by photographic or other means unless SebaKMT have
before-hand declared their consent in writing. The content of this handbook is subject to change without notice.
SebaKMT cannot be made liable for technical or printing errors or shortcomings of this handbook. SebaKMT also
disclaim all responsibility for damage resulting directly or indirectly from the delivery, supply, or use of this matter.
4
Terms of Warranty
SebaKMT accept responsibility for a claim under warranty brought forward by a customer for a
product sold by SebaKMT under the terms stated below.
SebaKMT warrant that at the time of delivery SebaKMT products are free from manufacturing
or material defects which might considerably reduce their value or usability. This warranty does
not apply to faults in the software supplied. During the period of warranty, SebaKMT agree to
repair faulty parts or replace them with new parts or parts as new (with the same usability and
life as new parts) according to their choice.
SebaKMT reject all further claims under warranty, in particular those from consequential
damage. Each component and product replaced in accordance with this warranty becomes the
property of SebaKMT.
All warranty claims versus SebaKMT are hereby limited to a period of 12 months from the date
of delivery. Each component supplied by SebaKMT within the context of warranty will also be
covered by this warranty for the remaining period of time but for 90 days at least.
Each measure to remedy a claim under warranty shall exclusively be carried out by SebaKMT
or an authorized service station.
To register a claim under the provisions of this warranty, the customer has to complain about
the defect, in case of an immediately detectable fault within 10 days from the date of delivery.
This warranty does not apply to any fault or damage caused by exposing a product to
conditions not in accordance with this specification, by storing, transporting, or using it
improperly, or having it serviced or installed by a workshop not authorized by SebaKMT. All
responsibility is disclaimed for damage due to wear, will of God, or connection to foreign
components.
For damage resulting from a violation of their duty to repair or re-supply items, SebaKMT can
be made liable only in case of severe negligence or intention. Any liability for slight negligence
is disclaimed.
5
Table of Contents
4 Design ................................................................................................................................. 12
7 Operating............................................................................................................................. 20
6
7
1 Safety Advice
Safety This manual contains basic advice for the installation and operation of the
precautions LCI TX and LCI TX-440 cable identifying generators. It is essential to make
this manual accessible to the authorised and skilled operator. He needs to
read this manual closely. The manufacturer is not liable for damage to
material or humans due to non-observance of the instructions and safety
advices provided by this manual.
Locally applying regulations have to be observed.
Symbols used in Important instructions concerning the protection of staff and equipment as
this manual well as technical safety within this document are labelled with one of the
following symbols:
Symbol Description
Notes have important information and useful tips on the
operation of your equipment. Non-observance may result
in useless measurement results.
Indicates a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury or
material damage.
CAUTION
Working with All electrical regulations of the country where the system is operated have
equipment of to be observed as well as national regulations for prevention of accidents
SebaKMT and existing regulations for the safety and operation of equipment of the
involved companies.
Original accessories ensure safe operation of the equipment. It is not
allowed and the warranty is lost if other accessories than the original ones
are used with the equipment.
Intended The cable identifying generators LCI TX and LCI TX-440 may only be
application operated at live low-voltage cables, measurement category 600 V / CAT IV
(EN 61010-1), according to their intended application.
Safe operation is only realised when using the equipment for its intended
purpose.
The limits described under technical data may not be exceeded.
8
2 Technical Description
Function The cable identifying generators are used for selective cable identification
on 100 V … 240 V (LCI TX) or 240 V … 440 V (LCI TX-440) live low-
voltage cables.
For such a cable identification procedure, the receiver CI RX is required
additionally. The handling of the receiver is described in a separate
operation manual.
The generator sends out pulses up to a peak current value of 90 A into the
cable to be identified. This test current generates an electromagnetic field
around the cable which is picked up by a flexible identification clamp
attached to the cable.
The test current of these identification permits a determination of the
current value and of the direction of the measuring pulse, thus leading to a
safe and reliable identification of the cable.
9
Technical data Parameter Value
Indicators o Power status LED (green)
o LED for pulse, polarity and error
indication (red)
Operating voltage
o LCI TX 100 V … 240 VAC 50/60 Hz
o LCI TX-440 240 V … 440 VAC 50/60 Hz
Pulse current 80 A ±10 A
Pulse sequence 30 per minute
Pulse width 1.7 ms
Weight 0.5 kg
Dimensions 151 mm x 101 mm x 60 mm
Protection class IP 54
Operating temperature -10°C … 60°C
Measurement category
(EN 61010-1)
o LCI TX Connected via NKG 1: 300 V / CAT II
Connected via NK 9 and fused clip:
1000 V / CAT III, 600 V / CAT IV
10
3 Scope of Delivery
Standard scope The following items are included with the standard shipment of the LCI TX
of delivery generator:
• Mains supply lead 2.0 m (NKG1)
• Mains measuring lead 2.5 m (NK9-C)
The following items are included with the standard shipment of the
LCI TX-440 generator:
• Black measuring lead with fused alligator clip, 2.0 m (MK34)
• Blue measuring lead with fused alligator clip, 2.0 m (MK35)
• Yellow/green measuring lead with alligator clip, 2.0 m (MK36)
11
4 Design
The LCI TX(-440) identifying generator is built into a sturdy ABS plastic
housing. The unit is in protection class IP 54.
The following figure shows the generators and their indicators and sockets:
1 1
2 2
3 4
Item Description
1 Power status LED
2 LED for pulse, polarity and error indication
3 Mains socket for direct connection to power outlets or low
voltage lines up to 240 V (LCI TX only)
4 Laboratory sockets for direct connection to low voltage lines
up to 440 V (LCI TX-440 only)
12
5 Safety Mechanisms
Overvoltage The generator has a built-in overvoltage protection. If a voltage > 270 V
protection (LCI TX) or >460 V (LCI TX-440) is detected, the fuses blow and, thus,
protect the unit from being destroyed.
After the overvoltage protection has been responded, it is necessary to
change both internal 5A F fuses (high breaking capacity) in order to put the
generator back to operating state.
Over-temperature At too high temperatures, the pulse transmission is automatically stopped
protection until the temperature has dropped below a certain threshold. If the over-
temperature protection is active, both LEDs are lit permanently and the
audible indicator does not sound.
13
6 Electrical Connection to Cable Under Voltage
Introduction The generator has to be connected to the open distal (load) end of the
cable. Proper cable identification using the CI RX can only be performed
between the transformer and the generator.
Connection sequence
Greatest care must be taken when connecting the generator to ensure
that the protective and neutral conductor are connected securely first.
CAUTION Only then may the live phase conductor be connected. Disconnect in
reverse order: first disconnect the phase conductor, then the protective
and neutral conductor.
14
6.1 Connecting the LCI TX
PEN PEN
Transformer
15
6.2 Connecting the LCI TX-440 to open distribution systems
Measuring leads The LCI TX-440 is connected to open distributors by means of the
measuring leads MK34 (black), MK35 (blue) and MK36 (yellow/green). It is
essential that the measuring leads are connected to the generator in
accordance with the colour coding!
To prevent arcing in the event of a short circuit in the system measuring
lead, the blue and the black test terminals are fitted with 10 A fuses. The
maximum switch load of these fuses is 50 kA.
Connection When connecting the LCI TX-440 for normal cable identification according
between phase to the current impulse method, proceed as illustrated in the picture
and earth below:
PEN PEN
Transformer
16
Connection The LCI TX-440 generator can also be connected between two phases of
between phases a multi-conductor cable. With this type of connection, the identification of
the cable is performed according to the “Twisted-Field” method by
means of the optional TFS CI sensor (moved along or around the cable).
The black and the red test lead must be connected to any two phases of
the cable. The use of a protective conductor is not necessary from a
metrological point of view, is recommended however for safety reasons.
PEN PEN
Transformer
17
6.3 Direct Connection to LV HRC fuses (Optional)
18
Step Description
4 Insert the plug-in adapter in the LV HRC fuse replacement
handle.
5 Push the plug-in adapter onto the upper contact blade so that it
attaches securely to the fuse attachment.
19
7 Operating
Polarity check The identifying generator switches itself on automatically after being
connected to the cable.
Subsequently, the generator automatically checks the polarity. This is
necessary as the CEE 7/7 (Schuko) plug of the mains supply lead NKG1
may have been connected the wrong way round.
Depending on the polarity, the system responds as follows:
Correct polarity Wrong polarity
In the case of wrong polarity, the polarity of the connection lead has to be
changed (the NKG 1 CEE 7/7 (Schuko) plug has to be turned round). The
connection lead must not be reconnected before the device has totally
turned off after appr. 3 seconds (both LEDs go out).
If the polarity change does not affect the LED status, it must be assumed
that the protective earth conductor is not connected. In this case, testing
cannot be carried out.
Practical use Once the polarity check has been successfully completed, the identifying
generator should be in operating status. The LEDs should now indicate
that the unit is working perfectly. The pulse indicator and an audible signal
should indicate a measuring pulse every 2 seconds. Cable identification
can now be started using the CI RX identifying receiver with the flexible
clamp.
A detailed description of the procedure is provided in the operating CI RX
operating manual.
The identification of the test signal can be impaired by asymmetrical
operating currents in the cable as well as by pulse shaped noise.
20
21
22
Operation manual
Cable Identifying Receiver
CI RX
Kommunikationsnetze
Communication Networks
Rohrleitungsnetze
Water Networks
Leitungsortung
Line Locating
Issue: 6 (11/2011)
ENG
Consultation with SebaKMT
The present system manual has been designed as an operating guide and for reference. It is
meant to answer your questions and solve your problems in as fast and easy a way as possible.
Please start with referring to this manual should any trouble occur.
In doing so, make use of the table of contents and read the relevant paragraph with great
attention. Furthermore, check all terminals and connections of the instruments involved.
Should any question remain unanswered, please contact:
Seba Dynatronic Hagenuk KMT
Mess- und Ortungstechnik GmbH Kabelmesstechnik GmbH
Dr.-Herbert-Iann-Str. 6 Röderaue 41
D - 96148 Baunach D - 01471 Radeburg / Dresden
Phone: +49 / 9544 / 68 – 0 Phone: +49 / 35208 / 84 – 0
Fax: +49 / 9544 / 22 73 Fax: +49 / 35208 / 84 249
E-Mail: [email protected]
http://www.sebakmt.com
SebaKMT
All rights reserved. No part of this handbook may be copied by photographic or other means unless SebaKMT have
before-hand declared their consent in writing. The content of this handbook is subject to change without notice.
SebaKMT cannot be made liable for technical or printing errors or shortcomings of this handbook. SebaKMT also
disclaim all responsibility for damage resulting directly or indirectly from the delivery, supply, or use of this matter.
3
Terms of Warranty
SebaKMT accept responsibility for a claim under warranty brought forward by a customer for a
product sold by SebaKMT under the terms stated below.
SebaKMT warrant that at the time of delivery SebaKMT products are free from manufacturing
or material defects which might considerably reduce their value or usability. This warranty does
not apply to faults in the software supplied. During the period of warranty, SebaKMT agree to
repair faulty parts or replace them with new parts or parts as new (with the same usability and
life as new parts) according to their choice.
SebaKMT reject all further claims under warranty, in particular those from consequential
damage. Each component and product replaced in accordance with this warranty becomes the
property of SebaKMT.
All warranty claims versus SebaKMT are hereby limited to a period of 12 months from the date
of delivery. Each component supplied by SebaKMT within the context of warranty will also be
covered by this warranty for the remaining period of time but for 90 days at least.
Each measure to remedy a claim under warranty shall exclusively be carried out by SebaKMT
or an authorized service station.
To register a claim under the provisions of this warranty, the customer has to complain about
the defect, in case of an immediately detectable fault within 10 days from the date of delivery.
This warranty does not apply to any fault or damage caused by exposing a product to
conditions not in accordance with this specification, by storing, transporting, or using it
improperly, or having it serviced or installed by a workshop not authorized by SebaKMT. All
responsibility is disclaimed for damage due to wear, will of God, or connection to foreign
components.
For damage resulting from a violation of their duty to repair or re-supply items, SebaKMT can
be made liable only in case of severe negligence or intention. Any liability for slight negligence
is disclaimed.
4
Table of Contents
4 Design .................................................................................................................................. 9
5 Start-up .............................................................................................................................. 10
6 Operation ........................................................................................................................... 13
6.1 General Handling of the Device ............................................................................................. 13
6.2 Identifying a Cable Using the DC Pulse Method (Current Flow Direction Determination) ......... 14
6.3 Phase identification using the PAS CI (optional) .................................................................... 15
6.4 Using the Optional TFS CI Sensor (“Twisted-Field” Method and Load Current Detection) ..... 16
6.4.1 Cable Identification with the “Twisted-Field” Method .................................................... 16
6.4.2 Load Current Detection ................................................................................................ 19
7 Troubleshooting ................................................................................................................ 22
5
1 Safety Advice
Safety This manual contains basic advice for the installation and operation of the
precautions CI RX. It is essential to make this manual accessible to the authorised and
skilled operator. He needs to read this manual closely. The manufacturer is
not liable for damage to material or humans due to non-observance of the
instructions and safety advice provided by this manual.
Locally applicable regulations have to be observed.
Working with All electrical regulations of the country where the system is operated have
equipment of to be observed as well as national regulations for the prevention of
SebaKMT accidents and existing regulations for the safety and operation of
equipment of any involved companies.
The original accessories provided ensure the safe operation of the
equipment. It is not allowed and the warranty is voided if any accessories
other than the original ones are used with the equipment.
Products of SebaKMT are continuously being enhanced according to the
state of the technology but such enhancements shall not constitute any
ground for claims of any kind, particularly indemnity claims, for older
versions of the product.
Intended The cable identifying receiver CI RX may only be operated with low and
application medium voltage cables, measurement category 600 V / CAT IV
(EN 61010-1), according to the intended application as described in this
manual.
Safe operation is only realised when using the equipment for its intended
purpose.
The limits described under technical data may not be exceeded.
6
2 Technical Description
Function The cable identifying receiver CI RX can be used to identify low voltage
cables and medium voltage cables out of a bunch of cables.
The receiver has to be operated in combination with one of the generators
(LCI TX, LCI TX-440 or CI TX) which transmit specific pulses into the
cable to be identified.
These current pulses generate an electromagnetic field around the cable
which is picked up by the flexible identification clamp clamped around the
cable.
Thus, the operator is able to determine the cable to be identified securely.
Technical data Parameter Value
Display o Power status LED (green)
o LEDs for indication of signal strength and
gain stage (red/green)
Sensors o Flexible identification clamp, Ø 140 mm
o Flexible identification clamp, Ø 250 mm
(optional)
o PAS CI phase identification sensor
(optional)
o “Twisted-Field” sensor TFS CI (optional)
Gain stages 10-stage (-3 dB … 24 dB dynamic range)
Power supply 2 x 1.5 V AA batteries
Operating time >50 h
Weight 0,4 kg (with batteries and sensor)
Dimensions (w x h x d) 150 mm x 65 mm x 35 mm
Protection class IP 54
Operating temperature -10°C … 60°C
Measurement category 600 V / CAT IV
7
3 Scope of Delivery
Standard scope The following items are included with the standard shipment of the
of delivery receiver:
• Cable identifying receiver CI RX
• Flexible identification clamp, Ø 140 mm
• 2 x 1.5 V AA batteries
Optional The following accessories can be ordered by your sebaKMT
accessories representative, if required:
Order
Accessories Description
number
Case Suitable for a complete cable 820011292
identifier system
Flexible identification Identification clamp with a 820025178
clamp AZF 250–CI diameter of 250 mm; used to
embrace larger cables or bundles
Phase identification Used for the identification of a 820014535
sensor PAS CI single phase in low-voltage
distribution networks
“Twisted-Field” Used for cable identification with 820024979
sensor TFS CI "Twisted-Field-Method" and for
load current detection
8
4 Design
9
5 Start-up
Connecting the When using the DC pulse method to identify a cable, special attention has
flexible identification to be paid to the orientation of the identification clamp.
clamp When working with the CI TX generator, encircle the cable to be identified in
such a way that the direction arrow on the identification clamp points
towards the grounded end of the cable as shown in the figure below:
PEN PEN
Transformer
10
In order to minimize the risk of stray field induction, the flexible
identification clamp (including the connection cable) must be connected at
a distance of at least 10 cm from nearby cables. If there is not enough
room for this, then at least the clamp closure and the connection cable
should be positioned as far away as possible from the nearby cables (see
diagram).
11
Switching-on The unit can be switched-on by pressing the on/off button [ 5 ].
If the receiver is ready for operation, the power LED on the on/off button
lights up green. Subsequent to the switch-on process, the green [ 1 ] and
red [ 2 ] signal LEDs indicate for three seconds the preselected gain stage
(1 up to 10). The following figure shows an example for gain stage 7:
After the three seconds, the receiver starts to evaluate the signal level
picked up by the identification clamp. It may take a few seconds until the
receiver is able to clearly identify the incoming pulses and to indicate
them appropriately.
In the case of low batteries, all red and green signal LEDs are flashing for
a short period of time immediately after switch-on. Afterwards, the unit
switches off automatically. The batteries have to be replaced (see
section 8).
12
6 Operation
13
6.2 Identifying a Cable Using the DC Pulse Method (Current Flow
Direction Determination)
If more than one cable cause a deflection of the green LEDs, the
identification clamp may be placed the wrong way round (see previous
section).
14
6.3 Phase identification using the PAS CI (optional)
Generator
Transformer or
earthed cable end
The identification of the phase is basically performed the same way as the
identification of a cable (see section 6.2).
If the sensor orientation is correct, the green LEDs [1] should only light up
on the phase which is connected to the generator, while the red LEDs [2]
should light up on all other conductors.
Any return currents flowing through the conductors of adjacent cables
should always result in the red LEDs lighting up.
15
6.4 Using the Optional TFS CI Sensor (“Twisted-Field” Method and Load
Current Detection)
Purpose The cable identification with the ‘Twisted-Field’ method, which is required by
various standards, is used specifically in the following scenarios:
• Inaccessible cables (cannot be gripped by identification tongs)
• Cable systems without alternative earth return path (e.g. IT
networks)
• Undefined return current conditions (e.g. paper-insulated lead
covered (PILC) cables with metal sheath in contact with earth)
LCI TX-440
or CI TX
Process Due to the twist of the conductors along the cable, the fields formed around
the two conductors are counter rotating.
As the TFS CI sensor is moved along or around the cable to be identified,
the green [ 1 ] and red [ 2 ] LEDs on the receiver light up in rotation.
Adjacent cables do not produce any activation of the LEDs.
16
When examining the sensor travel along a twisted bifilar cable, the following
signal behavior becomes apparent:
Axial The gauge reacts to
Longitudinal motion
motion the connected cable
Current flow: →
17
The “Twisted-Field” method can also be applied to three- or four-core
cables without any problems.
For four-core cables, it is recommended to connect the identification
generator to two opposite strands to be able to identify the described
polarity reversal as clearly as possible (see images).
The following diagram illustrates the signal response when the sensor is
moved around a four-core cable in a radial direction:
LCI TX-440
or CI TX
The following diagram illustrates the signal response when the sensor is
moved along a four-core cable:
LCI TX-440
or CI TX
18
6.4.2 Load Current Detection
Purpose The determination of the current flow direction and the “Twisted-Field”
identification represent straightforward and reliable cable identification
methods. An additional recording of load currents with a frequency of
50 Hz or 60 Hz can further increase the reliability of the cable identification
in certain cases, e.g. in the case of two cables of which only one is
definitely under voltage.
Switching to load When the receiver is switched on, it is always in identification mode. To
current detection switch to load current detection mode, press the + [ 3 ] and − [ 4 ] buttons
simultaneously. After the mode has been switched successfully, the
following continuous display shows that load current detection is activated:
19
Gain adjustment Setting the gain is performed in a similar way to the identification mode
(see section 6.1). Unlike the identification mode, in load current detection
mode, LEDs 1 and 10 flash to display overload. If this happens even when
gain stage 1 is selected, then you should increase the distance between
the sensor and the cable (e.g. by placing your hand in between them).
Procedure To perform load current identification, the TFS CI sensor must be placed
onto the cable that is to be checked. The sensor should be kept in this
position for several seconds. If the display on the receiver does not
respond, this means that no load current could be detected. This result
should be verified at other points along the cable for good measure.
Even if no load current has been detected, the cable must not
necessarily be de-energized! The CI RX cannot be used to test the
absence of voltage!
WARNING If load current has been detected, on the other hand, the cable must be
considered as live.
20
If a detectable current is flowing through the cable, then this is signaled by
the red and green LEDs flashing every second. The following differentiation
is made:
Depending on the gain setting of the CI RX, the following conclusions can
be drawn:
Gain
1…5 6 7 8 9 10
Full-scale deflection >50 A
Half-scale >10 A >7 A >5 A >3,5 A >2,5 A >1.75 A
deflection
21
7 Troubleshooting
The battery housing is located on the back side of the unit and can be
opened using the slide closure. Two identical 1.5 V AA batteries have to be
used as replacement.
22