Introduction To Flow Cytometry: Principles Data Analysis Protocols Troubleshooting
Introduction To Flow Cytometry: Principles Data Analysis Protocols Troubleshooting
Flow Cytometry
Principles
Data analysis
Protocols
Troubleshooting
Data analysis
Protocols
Troubleshooting
Misha Rahman
Marketing Communications Manager
2
Contents
Chapter 1 Principles of the flow cytometer
Fluidics system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Optics and detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Signal processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Electrostatic cell sorting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Chapter 5 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Recommended reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3
1
Chapter
Principles of the
flow cytometer
Fluidics system
One of the fundamentals of flow cytometry is the ability to measure the
properties of individual particles. When a sample in solution is injected into
a flow cytometer, the particles are randomly distributed in three-dimensional
space. The sample must therefore be ordered into a stream of single particles
that can be interrogated by the machine’s detection system. This process is
managed by the fluidics system.
Essentially, the fluidics system consists of a central channel/core through
which the sample is injected, enclosed by an outer sheath that contains faster
flowing fluid. As the sheath fluid moves, it creates a massive drag effect on
the narrowing central chamber. This alters the velocity of the central fluid
whose flow front becomes parabolic with greatest velocity at its center
and zero velocity at the wall (see Figure 1). The effect creates a single file
of particles and is called hydrodynamic focusing. Under optimal conditions
(laminar flow) the fluid in the
Central core central chamber will not mix
with the sheath fluid.
The flow characteristics
of the central fluid can be
estimated using Reynolds
Sheath Number (Re):
fluid
Re = pVD
µ
CH A PTER 1 5
Long pass 500 nm long pass
transmits > 500 nm
Dichroic
Light passed
mirror
Blocked light
deflected
Signal processing
When light hits a photodetector a small current (a few microamperes) is
generated. Its associated voltage has an amplitude proportional to the
total number of light photons received by the detector. This voltage is then
amplified by a series of linear or logarithmic amplifiers, and by analog to
digital convertors (ADCs), into electrical signals large enough (5–10 volts) to
be plotted graphically.
Log amplification is normally used for fluorescence studies because it expands
weak signals and compresses strong signals, resulting in a distribution that is
easy to display on a histogram. Linear scaling is preferable where there is not
such a broad range of signals e.g. in DNA analysis.
The measurement from each detector is referred to as a ‘parameter’ e.g.
forward scatter, side scatter or fluorescence. The data acquired in each
6 CHAPTER 1
Fluidics system
Filters Detectors
635
nm
FSC
48 8 nm
Lens
Lasers PMT
(FL-1)
PMT
(FL-2)
Filters
PMT
(FL-3)
PMT
(FL-4)
Screen
SSC
Detector
parameter are known as the ‘events’ and refer to the number of cells
displaying the physical feature or marker of interest.
CH A PTER 1 7
Laser interrogation
+
Break-off point
(charging)
–
– +
Voltage
+ plates
+ –
The speed of flow sorting depends on several factors including particle size
and the rate of droplet formation. A typical nozzle is between 50–70 µM in
diameter and, depending on the jet velocity from it, can produce
30,000–100,000 droplets per second, which is ideal for accurate sorting.
Higher jet velocities risk the nozzle becoming blocked and will also decrease
the purity of the preparation.
8 CHAPTER 1
Principles of fluorescence
2
Chapter
Fluorochromes and light
Fluorochromes are essentially dyes, which accept light energy (e.g. from
a laser) at a given wavelength and re-emit it at a longer wavelength.
These two processes are called excitation and emission. The process of
emission follows extremely rapidly, commonly in the order of nanoseconds,
and is known as fluorescence. Before considering the different types of
fluorochrome available for flow cytometry, it is necessary to understand the
principles of light absorbance and emission.
Light is a form of electromagnetic energy that travels in waves. These waves
have both frequency and length, the latter of which determines the color of
light. The light that can be visualized by the human eye represents a narrow
wavelength band (380–700 nm) between ultraviolet (UV) and infrared (IR)
radiation (Figure 5). Sunlight, for example, contains UV and IR light that,
although invisible to the eye, can still be felt as warmth on the skin and
measured scientifically using photodetectors. The visible spectrum can further
be subdivided according to color, often remembered by the mnemonic
‘ROY G BV’ standing for red, orange, yellow, green, blue and violet. Red light
is at the longer wavelength end (lower energy) and violet light at the shorter
wavelength end (higher energy).
Higher Lower
energy energy
400 nm 500 nm 600 nm 700 nm
Ultraviolet Infrared
Visible spectrum
Stokes Shift
When light is absorbed by a fluorochrome, its electrons become excited and
move from a resting state (1) to a maximal energy level called the ‘excited
electronic singlet state’ (2). The amount of energy required will differ for each
CH A PTER 2 9
fluorochrome and is depicted in Figure 6 as Eexcitation. This state only lasts for
1–10 nanoseconds because the fluorochrome undergoes internal
conformational change and, in doing so, releases some of the absorbed
energy as heat. The electrons subsequently fall to a lower, more stable,
energy level called the ‘relaxed electronic singlet state’ (3). As electrons
steadily move back from here to their ground state they release the
remaining energy (Eemission) as fluorescence (4).
As Eemission contains less energy than was originally put into the fluorochrome
it appears as a different color of light to Eexcitation. Therefore, the emission
wavelength of any fluorochrome will always be longer than its excitation
wavelength. The difference between Eexcitation and Eemission is called Stokes
Shift and this wavelength value essentially determines how good a
fluorochrome is for fluorescence studies. After all, it is imperative that the
light produced by emission can be distinguished from the light used for
excitation. This difference is easier to detect when fluorescent molecules have
a large Stokes Shift.
Heat
OUT
2
Excited singlet state
Eexcitation
Eemission
Light
OUT
Light
IN
Ground state
1 4
10 CHAPTER 2
Excitation
laser line A B
Peak or Peak or
maximum maximum
Wavelength (nm) 400 500 600 700 400 500 600 700
Spectrum
Excellent Weak
(blue light) signal signal
(green) (orange)
CH A PTER 2 11
Why use a fluorescent probe?
The purpose of a fluorescent probe, such as a fluorochrome-conjugated
antibody, is to directly target an epitope of interest and to allow its biological
and biochemical properties to be measured more easily by the flow cytometer.
Fluorescent probes are useful in a wide range of applications including:
identifying and quantifying distinct populations of cells, cell surface receptors
or intracellular organelles; cell sorting; immunophenotyping; calcium influx
experiments; determining nucleic acid content; measuring enzyme activity, and
for apoptosis studies. By changing the excitation light and using more than one
fluorochrome, it is possible to analyze several parameters of the sample at any
one time. This forms the basis of multicolor fluorescence studies.
Single dyes:
Some of these single dyes e.g. FITC have been in use for the past 30 years but
are now facing competition from alternatives like Alexa Fluor® dyes, which
offer the user greater photostability and increased fluorescence.
Tandem dyes:
In a tandem dye, a small fluorochrome takes a ‘piggy-back’ ride on another
larger fluorochrome. When the first dye is excited and reaches its maximal
singlet state, all its energy transfers to the second dye (an acceptor molecule),
located in close proximity. This activates the second fluorochrome, which then
produces the fluorescence emission. The process is called FRET (fluorescence
resonance energy transfer). It is a clever way to achieve higher Stokes Shifts
and, therefore, increase the number of colors that can be analyzed from a
single laser wavelength.
The majority of tandem dyes have been manufactured for the standard 488 nm
laser, which is found in most flow cytometers. Tandem dyes are very useful for
multicolor fluorescence studies especially in combination with single dyes.
For example, Alexa Fluor® 488, Phycoerythrin, PerCP-Cy5.5 and PE-Cy7 can
all be excited at 488 nm, but will produce green, yellow, purple and infrared
emissions respectively, which can be measured using separate detectors.
12 CHAPTER 2
Single Dyes for Flow Cytometry/Microscopy
Fluorochrome Fluorescence Maximal Maximal Spectrally similar dyes
color absorbance (nm) emission (nm)
Alexa Fluor® 405 401 421 Cascade Blue , DyLight® 405
Alexa Fluor® 488 495 519 Cy2, DyLight® 488, FITC
Alexa Fluor® 647 650 665 APC, Cy5, DyLight® 649
Alexa Fluor® 700 Infrared 702 723 Alexa Fluor® 647
APC 650 661 Alexa Fluor® 647, Cy5
Cy5 649 670
DyLight® 405 400 420 Alexa Fluor® 405, Cascade Blue
DyLight® 488 493 518 Alexa Fluor® 488, FITC
DyLight® 549 562 576 Alexa Fluor® 546, Alexa Fluor® 555,
Cy3, TRITC
DyLight® 649 654 673 Alexa Fluor® 647, Cy5
DyLight® 680 692 712 Alexa Fluor® 650, Cy5.5
DyLight® 750 Near Infrared 752 778 Alexa Fluor® 750
DyLight® 800 Infrared 777 794 IR Dye® 800
FITC 490 525 Alexa Fluor® 488, Cy2, DyLight® 488
Pacific Blue™ 410 455
PerCP 490 675
PE 490; 565 578
Texas Red® 596 615
TRITC 596 570 Alexa Fluor® 488, Cy3
Abbreviations
APC Allophycocyanin
FITC Fluorescein isothiocyanate
PE Phycoerythrin
PerCP Peridinin-chlorophyll-protein complex
CH A PTER 2 13
Fluorescence compensation
One consideration to be aware of when performing multicolor fluorescence
studies is the possibility of spectral overlap. When two or more fluorochromes
are used during a single experiment there is a chance that their emission
profiles will coincide, making measurement of the true fluorescence emitted
by each difficult. This can be avoided by using fluorochromes at very different
ends of the spectrum e.g. Alexa Fluor® 405 and Phycoerythrin; however, this
is not always practical.
Instead, a process called fluorescence compensation is applied during data
analysis, which calculates how much interference (as a %) a fluorochrome
will have in a channel that was not assigned specifically to measure it.
Figure 8 helps to explain the concept.
The graphs show the emission profiles of two imaginary fluorochromes ‘A’
and ‘B’ which are being detected in FL-1 and FL-2 channels respectively.
Because the emission profiles are so close together, a portion of fluorochrome
A spills over into FL-2 (red shade) and conversely, some of fluorochrome B
reaches FL-1 (dark blue shade).
To calculate how much compensation needs to be applied to the dataset if
both dyes are used simultaneously, some control readings must first be taken.
Fluorochrome A should be run through the flow cytometer on its own and the
% of its total emission that is detectable in FL-2 (spillover) determined. The
procedure should be repeated with fluorochrome B, except that this time FL-1
is spillover.
Suppose the results are:
Spillover Fluorescence
FL-1 FL-2
Fluorochrome A N/A 17%
Fluorochrome B 5% N/A
This means that when the two fluorochromes are used for a dual-color
experiment, the true reading for fluorochrome A in FL-1
= (total fluorescence measured in FL-1) minus (5% of fluorochrome B’s total fluorescence)
Similarly, the true reading for fluorochrome B in FL-2
= (total fluorescence measured in FL-2) minus (17% of fluorochrome A’s total fluorescence)
Fortunately, modern flow cytometry analytical software applies fluorescence
compensation mathematics automatically, which simplifies matters
considerably.
14 CHAPTER 2
FL-1 FL-2
Spectral properties
of two imaginary
fluorochromes,
Fluorescence
Wavelength
FL-1 FL-2
Spectral overlap.
Dark blue shade represents
the proportion of B
that overlaps into the
Fluorescence
FL-1 channel.
Red shade represents
B the proportion of A
that interferes with
A
FL-2 channel
measurements.
Wavelength
CH A PTER 2 15
Data analysis
3
Chapter
Gates and regions
An important principle of flow cytometry data analysis is to selectively
visualize the cells of interest while eliminating results from unwanted
particles e.g. dead cells and debris. This procedure is called gating.
16 CHAPTER 3
On the density plot, each dot or point represents an individual cell that has
passed through the instrument. Yellow/green hotspots indicate large numbers
of events resulting from discreet populations of cells. The colors give the
graph a three-dimensional feel. After a little experience, discerning the various
subtypes of blood cells is relatively straightforward.
Contour diagrams are an alternative way to demonstrate the same data.
Joined lines represent similar numbers of cells. The graph takes on the
appearance of a geographical survey map, which, in principle, closely
resembles the density plot. It is a matter of preference but sometimes discreet
populations of cells are easier to visualize on contour diagrams e.g. compare
monocytes in Figure 9.
Newer gating strategies utilize fluorescence parameters along with scatter
parameters. Once again, blood can be used to demonstrate this principle.
Above on the left is a FSC/SSC plot for human lysed whole blood using
smaller numbers of cells than in Figure 9. The lymphocytes, monocytes and
granulocytes have been gated as region 1 (R1), region 2 (R2) and region
3 (R3), respectively. ‘Region’ simply refers to an area drawn on a plot
displaying flow cytometry data.
On the right the same cells are now plotted as SSC on the y-axis versus CD45
fluorescence on the x-axis. CD45 is a marker expressed on all white blood
cells at varying intensities but is absent on red blood cells. In relative terms,
lymphocytes have a low SSC and high CD45 count (R4), granulocytes have
a high SSC and low CD45 count (R6), while monocytes are somewhere in
between the other two (R5). The major difference between the lymphocytes
CH A PTER 3 17
gated in R1 and those gated in R4 is the absence of red blood cells in the
latter, making it a much purer preparation. This highlights the usefulness
of gating strategies that combine a scatter parameter with a fluorescence
parameter.
Single-parameter histograms
These are graphs that display a single measurement parameter (relative
fluorescence or light scatter intensity) on the x-axis and the number of events
(cell count) on the y-axis.
The histogram in Figure 11 looks very basic but is useful for evaluating the
total number of cells in a sample that possess the physical properties selected
for or which express the marker of interest. Cells with the desired
characteristics are known as the positive dataset.
Ideally, flow cytometry will produce a single distinct peak that can be
interpreted as the positive dataset. However, in many situations, flow analysis
is performed on a mixed population of cells resulting in several peaks on
the histogram. In order to identify the positive dataset, flow cytometry should
be repeated in the presence of an appropriate negative isotype control
(see Figure 12).
18 CHAPTER 3
Figure 12 Which is the positive dataset? LEFT, Using rat anti-mouse F4/80 conjugated to
FITC to stain mouse peritoneal macrophages produces two peaks. RIGHT, By running
an appropriate isotype control (rat IgG2b negative control conjugated to FITC) and
overlaying its image on the histogram (blue outline) the positive dataset is identified
as the taller red peak on the right.
CH A PTER 3 19
In Figure 13, 99.83% of the negative control (blue outline) is in R2.
28.14% of cells (red shade) ‘stain negative’ for F4/80 (R2) compared to
71.86% in the positive dataset (R3). Additional statistics about the peaks
(median and standard deviation) is also provided automatically here but this
will vary with the software. A similar type of analysis will be generated for
two-parameter histograms.
Two-parameter histograms
These are graphs that display two measurement parameters, one on the
x-axis and one on the y-axis, and the cell count as a density (dot) plot or
contour map. The parameters could be SSC, FSC or fluorescence. Some
examples of two-parameter histograms were illustrated in Figures 9 and 10.
Another example is the dual-color fluorescence histogram presented below.
Lymphocytes were stained with anti-CD3 in the FITC channel (x-axis) and
anti-HLA-DR in the PE channel (y-axis). CD3 and HLA-DR are markers for
T cells and B cells, respectively.
20 CHAPTER 3
Currently, flow cytometry can be performed on samples labeled with up to 17
fluorescence markers simultaneously1. Therefore a single experiment can yield
a large set of data for analysis using various two-parameter histograms.
Intracellular antigens
Staining intracellular antigens like cytokines can be difficult because
antibody-based probes cannot pass sufficiently through the plasma
membrane into the interior of the cell. To improve the situation, cells
should first be fixed in suspension and then permeabilized before adding
the fluorochrome. This allows probes to access intracellular structures while
leaving the morphological scatter characteristics of the cells intact. Many
commercial kits are available today that provide the reagents to carry out
these crucial steps e.g. Leucoperm™ (see Figure 15).
(a) BEFORE
(b) AFTER
1Perfetto,
S. et al (2004) Seventeen colour flow cytometry: unravelling the immune system.
Nature Reviews Immunology 4:648–655
CH A PTER 3 21
Immunophenotyping
All normal cells express a variety of cell surface markers, dependent on the
specific cell type and degree of maturation. However, abnormal growth may
interfere with the natural expression of markers resulting in overexpression
of some and under-representation of others. Flow cytometry can be used
to immunophenotype cells and thereby distinguish between healthy and
diseased cells. It is unsurprising that today immunophenotyping is one of the
major clinical applications of flow cytometry, and is used to aid the diagnosis
of myelomas, lymphomas and leukemias. It can also be used to monitor the
effectiveness of clinical treatments.
The differences between the blood profiles of a healthy individual and one
suffering from leukemia, for instance, are very dramatic. This can be seen
from the FSC v SSC plots in Figure 16. In the healthy person the cell types are
clearly defined, whereas blood from a leukemia patient is abnormal and does
not follow the classic profile.
Figure 16 Immunophenotyping
22 CHAPTER 3
Testing the patient’s lymphocytes for specific cell surface markers also reveals
more about the condition.
Normal person
CD3
A normal person has a significant proportion of CD3-positive lymphocytes.
In the patient with leukemia, staining for CD3 is absent.
CD20
In the leukemia patient there are a large number of cells staining positive for
CD20. In the healthy person only a few stain positive.
HLA-DR
The leukemia patient is HLA-DR-positive. In the normal person only a small
number of cells stain positive.
Being CD3-negative, CD20-positive and HLA-DR-positive, a clinician could
diagnose with certainty that this patient is suffering from a B cell lineage
leukemia or lymphoma. The precise classification of disease may be
determined using further antibodies.
CH A PTER 3 23
Common protocols
4
Chapter
Sample preparation
Single cells must be suspended at a density of 105–107 cells/ml to prevent
the narrow bores of the flow cytometer and its tubing from clogging up.
The concentration also influences the rate of flow sorting, which typically
progresses at 2000–20,000 cells/second. However, higher sort speeds may
decrease the purity of the preparation.
Phosphate buffered saline (PBS) is a common suspension buffer and the most
straightforward samples for flow cytometry include non-adherent cells from
culture, water-borne micro-organisms, bacteria and yeast. Even whole blood
is easy to use – red cells are usually removed by a simple lysis step; it is then
possible to quickly identify lymphocytes, granulocytes and monocytes by their
FSC/SSC characteristics (see page 16).
However, researchers may also wish to analyze cells from solid tissues
e.g. liver or tumors. In order to produce single cells, the solid material must
be disaggregated. This can be done either mechanically or enzymatically.
Mechanical disaggregation is suitable for loosely bound structures
e.g. adherent cells from culture, bone marrow and lymphoid tissue. It involves
passing a suspension of chopped tissue through a fine-gauge needle several
times, followed by grinding and sonication as necessary.
Enzymes are used to disrupt protein-protein interactions and the extracellular
matrix that hold cells together. Their action is dependent on factors including
pH, temperature and co-factors, so care must be taken when choosing
an enzyme. For example, pepsin works optimally between pH 1.5–2.5
but the acidic conditions would damage cells if left unneutralized for too
long, and cell surface antigens of interest may be lost. Chelators like EDTA
and EGTA can remove divalent cations responsible for maintaining cell
function and integrity but their presence may inhibit certain enzymes, for
instance, collagenase requires Ca2+ for activity. Enzymatic and mechanical
disaggregation is often a trial and error process to optimize the isolation of
the epitope under investigation.
To study intracellular components e.g. cytokines by flow cytometry, the
plasma membrane of the cell must be permeabilized to allow dyes or
1 Preparation of cells
(a) Cells stored in liquid nitrogen
1 Prepare PBS/BSA buffer (phosphate buffered saline pH 7.4 and 1% BSA).
2 Carefully remove cells from liquid nitrogen storage.
3 Thaw rapidly using PBS/BSA buffer and place into a 15 ml conical
centrifuge tube.
4 Centrifuge at 400 g for 5 minutes.
5 Discard supernatant and resuspend pellet in an appropriate amount of
PBS/BSA buffer.
(b) Tissue culture cell lines in suspension
1 Prepare PBS/BSA buffer (phosphate buffered saline pH 7.4 and 1% BSA).
2 Decant cells from tissue culture flask into 15 ml conical centrifuge tube(s).
3 Centrifuge at 400 g for 5 minutes.
4 Discard supernatant and resuspend pellet in 10 ml of PBS/BSA.
5 Centrifuge at 400 g for 5 minutes.
6 Discard supernatant and resuspend pellet in an appropriate amount of
PBS/BSA.
(c) Adherent tissue culture cell lines
1 Prepare PBS/BSA buffer (phosphate buffered saline pH 7.4 and 1% BSA).
2 Harvest cells by gentle scraping using 2 ml of PBS/BSA buffer.
3 Transfer cells to a 15 ml conical tube and add buffer up to 10 ml.
4 Centrifuge at 400 g for 5 minutes.
5 Discard supernatant and resuspend pellet in fresh PBS/BSA (10 ml).
6 Centrifuge at 400 g for 5 minutes.
7 Discard supernatant and resuspend pellet in an appropriate amount of
PBS/BSA buffer.
CH A PTER 4 25
(d) Preparing cells from solid/lymphoid tissues
1 Place tissue on a sterile Petri dish. Remove cells by gently perfusing
the tissue using a syringe and needle containing approximately 15 ml
of PBS/BSA (phosphate buffered saline pH 7.4 and 1% BSA).
2 Transfer the cell suspension from the Petri dish into a 15 ml conical
centrifuge tube.
3 Centrifuge at 400 g for 5 minutes.
4 Discard the supernatant and resuspend the pellet in PBS/BSA.
5 Add 10 ml of ammonium chloride lysis buffer.
6 Mix and incubate for 2 minutes. DO NOT EXCEED THIS TIME.
7 Centrifuge at 400 g for 5 minutes.
8 Add 10 ml of PBS/BSA and mix.
9 Centrifuge again at 400 g for 5 minutes.
10 Discard the supernatant and resuspend the pellet to a final volume of
10 ml with PBS/BSA.
11 Count cells using a hemocytometer.
12 Adjust the cell suspension, if necessary, to give a final count of
0.7–1.2 × 107 cells/ml.
26 CHAPTER 4
[To the blood suspension add freshly prepared red cell lysis buffer e.g. 2 ml
of AbD Serotec’s Erythrolyse and mix well. Incubate for 10 minutes at room
temperature. Centrifuge at 400 g for 5 minutes and discard the supernatant].
Resuspend cells in 0.2 ml of PBS/BSA or with 0.2 ml of 0.5%
paraformaldehyde in PBS/BSA if required.
Acquire data by flow cytometry. Appropriate standards should always be
included e.g. an isotype-matched control sample.
The stimulation conditions described are suitable for IFN gamma, IL-2 and
TNF alpha. Different conditions may be needed for other cytokines.
The procedure requires a reagent kit to fix and permeabilize cells. There are
several available but we recommend Leucoperm™.
28 CHAPTER 4
Note. All blood samples must be collected into heparin anti-coagulant. EDTA
interferes with the cell stimulation process and, therefore, must be avoided.
1 Aliquot 0.5 ml of blood separately into 2 tubes, then add 0.5 ml of cell
culture medium (without any additives) to each sample.
2 To one tube (the resting population), add monensin to a final
concentration of 3 mM.
3 To the other tube (activated cells), add PMA, ionomycin and monensin to
a final concentration of 10 ng/ml, 2 mM and 3 mM, respectively.
4 Incubate for 2–4 hours at 37°C in a 5% CO2 atmosphere.
5 At the end of the incubation period aliquot 100 µl samples into the
appropriate number of tubes.
6 Add cell surface antibodies at this stage (if needed for your experiment)
and incubate for 15 minutes.
7 Add 100 µl of Leucoperm™ Reagent A per tube and incubate for
15 minutes. This reagent fixes cells in suspension.
8 Wash twice with PBS containing 0.1% sodium azide and 1% BSA.
9 Add 100 µl of Leucoperm™ Reagent B (permeabilizes cells) and the
required anti-cytokine antibodies.
10 Incubate for 20 minutes.
11 Wash twice using the PBS buffer, and analyze by flow cytometry.
CH A PTER 4 29
5 Wash once in wash buffer.
6 Resuspend cells in Leucoperm™ Reagent B (cell permeabilization agent)
using 50 µl per 1 × 106 cells.
7 Aliquot 50 µl of cell suspension into the required number of tubes
containing directly-conjugated antibodies. Incubate for 30 minutes at
room temperature.
8 Wash once in wash buffer, and then resuspend in 0.25 ml of 0.5%
paraformaldehyde in PBS.
9 Store at 4°C until acquisition on the flow cytometer, preferably within
24 hours.
30 CHAPTER 4
Troubleshooting
5
Chapter
If something doesn’t work check through the following list to resolve the
problem. If there are still difficulties and you have purchased an AbD Serotec
antibody our Technical Services Team will be happy to offer further advice.
PE antibody does not stain 1 PE conjugate may have been frozen. If so, purchase another vial
but same FITC antibody gives of antibody.
good results 2 Paraformaldehyde (PFA) may be a problem. Breakdown of PFA
may release methanol, which will affect staining. Make up fresh
paraformaldehyde. Cells can be analyzed immediately without
fixing.
CH A PTER 5 31
Unusual scatter profiles 1 Make sure that cells are used as fresh as possible. Profile may be
showing dead cells and debris.
2 Activation methods may affect scatter characteristics of cells.
3 If you are using lysing solution, confirm that this is fresh and has
been made up correctly.
Unexpected staining 1 Some reagents may affect certain antigens and, therefore, may
need reviewing e.g. EDTA will affect some platelet markers.
2 Lysing solutions may affect certain antigens. Select a method
that does not interfere with antigen detection.
3 Some antigens are expressed intracellularly and, therefore, cell
permeabilization methods may be required. Check manufacturer’s
datasheet for correct permeabilization reagent.
32 CHAPTER 5
Recommended reading
Flow Cytometry: A Practical Approach, 3rd Edition. (Practical Approach Series).
Edited by M.G. Ormerod. Oxford University Press (2000)
Alexa Fluor® and Pacific Blue™ are trademarks of Molecular Probes Inc., OR, USA
Cy containing products or portions thereof are manufactured under license from Carnegie Mellon
University under U.S. Patent Number 5,268,486 and related patents
Cy® and CyDye® are registered trademarks of GE Healthcare Limited
DyLight® is a registered trademark of Thermo Fisher Scientific and its subsidiaries
FACS™ (Fluorescence Activated Cell Sorter) is a trademark of Becton, Dickinson and Company, CA, USA
IRDye® Infrared Dyes are a registered trademark of LI-COR Biosciences
Leucoperm™ is made for AbD Serotec by AN DER GRUB Bio Research GmbH
33
Primary and secondary antibodies
Wide range of species, formats, and labels
www.abdserotec.com/flowcytometry
North & South America Tel: +1 800 265 7376 Europe Tel: +49 (0) 89 80 90 95 21
Fax: +1 919 878 3751 Email: [email protected]
Email: [email protected]
LIT.FLOW.MAY.2014.1 © Bio-Rad Laboratories Inc 2014. All rights reserved. Published by AbD Serotec, A Bio-Rad Company. Endeavour House,
Langford Business Park, Langford Lane, Kidlington, Oxford OX5 1GE, UK.