LabSpec v5 User Manual PDF
LabSpec v5 User Manual PDF
LabSpec5 manual 1
Table of contents
LabSpec5 User's Manual................................................................................................................................ 1
1 - LabSpec5 general overview................................................................................................................. 5
2 - LabSpec5 installation........................................................................................................................... 6
3 - LabSpec5 main window....................................................................................................................... 7
4 - The main menu bar...............................................................................................................................8
4.1 - File menu......................................................................................................................................8
4.1.1 - Open... function....................................................................................................................9
4.1.2 - Close function...................................................................................................................... 9
4.1.3 - Save function........................................................................................................................9
4.1.4 - Save All function............................................................................................................... 10
4.1.5 - Print function..................................................................................................................... 10
4.1.6 - Print Preview function....................................................................................................... 11
4.1.7 - Print Setup function........................................................................................................... 12
4.1.8 - Page function......................................................................................................................13
4.1.9 - Exit..................................................................................................................................... 16
4.2 - Edit menu .................................................................................................................................. 17
4.2.1 - Format function..................................................................................................................18
4.3 - Data menu ................................................................................................................................. 19
4.4 - Acquisition menu ...................................................................................................................... 20
4.4.1 - Custom function.................................................................................................................21
4.4.2 - Trigger function................................................................................................................. 21
4.4.3 - Real Time Display function............................................................................................... 22
4.4.4 - Options function.................................................................................................................23
4.4.5 - Autosave function.............................................................................................................. 26
4.4.6 - Multi window function...................................................................................................... 27
4.4.7 - Detector function............................................................................................................... 29
4.4.8 - Autofocus function.............................................................................................................30
4.4.9 - Video functions..................................................................................................................31
4.5 - Option menu ..............................................................................................................................33
4.6 - Setup menu.................................................................................................................................34
4.6.1 - Instrument.......................................................................................................................... 34
4.7 - Window menu............................................................................................................................ 36
5 - The control panel................................................................................................................................37
5.1 - Laser........................................................................................................................................... 37
5.2 - Filter........................................................................................................................................... 37
5.3 - Confocal hole............................................................................................................................. 38
5.4 - Slit.............................................................................................................................................. 38
5.5 - Spectro........................................................................................................................................38
5.6 - Options....................................................................................................................................... 39
5.7 - Exposure times........................................................................................................................... 39
5.8 - Motorized XY/Z Table ..............................................................................................................39
6 - The main icons bar............................................................................................................................. 40
6.1 - Data management tools.............................................................................................................. 40
6.1.1 - CUT....................................................................................................................................40
6.1.2 - OPEN................................................................................................................................. 40
6.1.3 - SAVE................................................................................................................................. 40
6.1.4 - PRINT................................................................................................................................ 40
6.1.5 - HELP..................................................................................................................................40
LabSpec5 manual 2
6.2 - Objects display and information................................................................................................ 41
6.2.1 - Scale normalization............................................................................................................41
6.2.2 - Intensity normalization...................................................................................................... 41
6.2.3 - Pointers centering...............................................................................................................41
6.2.4 - Data sizes........................................................................................................................... 41
6.2.5 - Data parameters..................................................................................................................41
6.3 - Data acquisition..........................................................................................................................42
6.3.1 - Spectrum Real Time Display ............................................................................................ 42
6.3.2 - Detector image Real Time Display....................................................................................42
6.3.3 - Spectra accumulation......................................................................................................... 42
6.3.4 - Spectral Images accumulation.......................................................................................... 42
6.3.5 - Spectral Image properties.................................................................................................. 43
6.3.6 - Multi window.....................................................................................................................43
6.3.7 - Video .................................................................................................................................43
6.4 - Basic data treatment................................................................................................................... 43
6.4.1 - Baseline correction ........................................................................................................... 43
6.4.2 - Data correction...................................................................................................................45
6.4.3 - Filtration.............................................................................................................................46
6.4.4 - Fourier transformation....................................................................................................... 47
6.4.5 - Arithmetical operation....................................................................................................... 48
6.4.6 - Peaks and Bands operations...............................................................................................49
6.4.7 - Profile.................................................................................................................................55
6.5 - High level data treatment........................................................................................................... 56
6.5.1 - Mapping ............................................................................................................................ 56
6.5.2 - New spectral profile...........................................................................................................57
6.5.3 - Modeling ........................................................................................................................... 58
7 - The graphics tools panel.....................................................................................................................59
7.1 - Activate the pointer ................................................................................................................... 59
7.2 - Integral....................................................................................................................................... 59
7.3 - Multiply data on the constant. ...................................................................................................60
7.4 - Add constant to data. .................................................................................................................60
7.5 - Remove ‘spike’ noise. ............................................................................................................... 60
7.6 - Correct data shape. .................................................................................................................... 60
7.7 - Linear shift of the scale limits. ..................................................................................................60
7.8 - Zoom scale. ............................................................................................................................... 60
7.9 - Change data intensity limits. .....................................................................................................60
7.10 - Change axes rectangle. ............................................................................................................60
7.11 - Change 3D axes. ......................................................................................................................60
8 - Window context menu....................................................................................................................... 61
8.1 - Format........................................................................................................................................ 61
8.2 - Norma.........................................................................................................................................62
8.3 - Colors......................................................................................................................................... 62
8.4 - Axes............................................................................................................................................63
8.5 - Graph..........................................................................................................................................64
8.5.1 - Spectrum............................................................................................................................ 64
8.5.2 - Spectra................................................................................................................................65
8.6 - Pointers.......................................................................................................................................65
8.7 - Image.......................................................................................................................................... 66
8.8 - Scale bar..................................................................................................................................... 66
8.9 - Image3D..................................................................................................................................... 67
8.10 - 3DAxes.....................................................................................................................................68
LabSpec5 manual 3
9 - Service................................................................................................................................................ 69
LabSpec5 manual 4
1 - LabSpec5 general overview
LabSpec 5 is a general data acquisition and data analysis software, based on open
NexGen module architecture. LabSpec 5 controls all the HORIBA Jobin-Yvon Raman
instruments, enables different data acquisition modes(eg. single spectrum, multidimensional data
set, video image etc.)
The data analysis routines include many commonly used Raman and FTIR tools, such as baseline
correction, linear and non-linear filters, peaks fitting Kramers-Kronig transform etc. The mapping
technique enables to generate and visualize map by using different spectral features : band
intensity, peak position and so on.
The module architecture allows to customize software to support specific operation, for example
read foreign data file format, implement data acquisition trigger etc.
The Visual Basic script language of Microsoft is supported, that yield possibility to create macro
commands or include specific data analysis function. The LabSpec 5 can be used also as Active X
control in third part application.
LabSpec5 manual 5
2 - LabSpec5 installation
LabSpec5 manual 6
3 - LabSpec5 main window
Window of the
active objects
LabSpec5 manual 7
4 - The main menu bar
The main menu bar is located on top of the LabSpec5 main window.
To access the File menu click on the File button of the main menu bar or type Alt+f or Alt+F keys.
LabSpec5 manual 8
4.1.1 - Open... function
The Open… command permits to select a previously saved file to load it.
To access this command select Open… from the File commands menu in the Main menu bar or use the
keyboard shortcut Ctrl + O or click directly on the following open icon in the main icons bar.
To access this command select Close from the File commands menu in the Main menu bar.
LabSpec5 manual 9
4.1.4 - Save All function
The Save All command permits to record all the objects of the active window in a file for future
retrieving.
To access this command select Save All from the File commands menu in the Main menu bar.
Print icon
File menu
Print…
LabSpec5 manual 10
4.1.6 - Print Preview function
The Print Preview command allows to display the to-print page before printing.
To access this command select Print preview from the File commands menu in the Main menu bar.
Ø The Print Preview window
The to-print page is displayed in the print preview window as it will appeared at print out so you can
check the page layout before starting the printing operation.
To modify the page layout , use the Page… function of the File menu.
On top of the print preview window some specific buttons commands are available.
Print Preview
specific commands
Click on this button or press P-key or p-key to start the print out.
Click on this button or press N-key or n-key to display the second page in case of a to-print
document on 2 pages. For the moment the layout of a to-print document is only on one page.
Click on this button or press T-key or t-key to display the 2 pages of the document. Useful
for to-print document on 2 pages (future expansion)
Click on this button or press I-key or i-key to zoom the previewed page.
LabSpec5 manual 11
Click on this button or press O-key or o-key to zoom-back the previewed page. This button is
available only if the Zoom function has been activated .
Click on this button or press C-key or c-key to quit the Print preview function and to return to
the LabSpec5 application.
LabSpec5 manual 12
4.1.8 - Page function
The Page… command permits to define the layout of the to-print document.
To access this command select Page… from the File commands menu in the Main menu bar.
Header page
Associated toole:
graphic object
Associated tool:
Image object
Associated tool:
To close the Page window without validating the current Page settings.
LabSpec5 manual 13
➢ Using the Page tools icons
To insert a component in the to-print page click on the corresponding tool. Then move the mouse on the
page so a cross appears. Then move the cross at the wished location, press on the left mouse button and
move the mouse to define the size of the corresponding component.
Select the component. Then when the following mouse cursor appears, click on left mouse button and
drag the mouse.
➢ Re-sizing a component of the to-print page
First, click on the object to select it. Selection handles appear around the selected object.
Move the mouse pointer on one of the selection handles until to have the size pointer:
or .Keep the mouse button pressed and move the selection handle and modify the size of the object.
➢ Modifying or creating the contents of a to-print page component
Except the data object, the contents of all the other objects could be modified by double-clicking on the
wished component.
Then a dedicated window appears with the tools to modify the corresponding component:
- The options window for the header page, the bottom of page and the Text object.
-The parameters window for the array of experiment information parameters.
-The Open window for the Image object.
➢ Modifying the general page presentation or display
Click anywhere on the to-print page on the right button of the mouse to display the following menu
Cut To remove the selected component
Format To define the parameters of the sheet of paper. (See Page Setup)
Turn To define the page orientation Portrait or Landscape
Zoom To define the display of the to-print page inside the Page window.
LabSpec5 manual 14
Ø The Parameters function
On the printed document the selected parameters are laid out in an array. The Parameters function
permits to select the data parameters you wish to be printed and to define the style of the array.
To access this command double click with the left mouse button on the information parameters object
when defining the layout of the to-print document with the Page… command
List of the non printed NSParameters tools to manage Print list: List of the
data parameters. the list of the data parameters to data parameters selected
print. to be printed.
Exposure is the
printed label for
the exposition
parameter.
Remark:
A column contains 2 fields: the data parameter label and the data parameter value.
LabSpec5 manual 15
In the print list move the selected parameter to the previous line.
In the print list move the selected parameter to the next line.
4.1.9 - Exit
The Exit command permits to close all the opened objects and quit the application.
To access this command select Exit from the File commands menu in the Main menu bar or click on the
Close icon in the upper right corner of the LabSpec5 main window.
Close
File button
menu
Exit
LabSpec5 manual 16
4.2 - Edit menu
The Edit commands menu enables to exchange data between different applications and facilitates the data
analysis process.
To access the Edit menu click on the Edit button of the main menu bar.
Copy (CTRL+X)....... To copy the active data in clipboard. You can select the data format with the
Format… command.
Paste (CTRL+V)....... To insert the clipboard contents. You can select the clipboard format with the
Format… command.
LabSpec5 manual 17
4.2.1 - Format function
This command permits to define the format of the data you export to the clipboard or and the one of the
data you import from the clipboard.
To access this command select Format… from the Edit commands menu in the Main menu bar.
• DataNative NextGen format. Can be used to create the exact copy of the data object.
LabSpec5 manual 18
4.3 - Data menu
The Data menu commands permit to activate a data object and to see the list of object of the active
window. This list is shown as a set of color radio buttons on the right pan of the screen.
To access the Data menu click on the Data button of the main menu bar.
You can activate one of the data object of the active window by clicking directly in
the list of the Data menu .
Datas… Select this command to show the full objects list of the active window.
LabSpec5 manual 19
4.4 - Acquisition menu
The Acquisition menu commands permit to setup the data acquisition parameters.
To access the Acquisition menu click on the Acquisition button of the main menu bar.
RTD........................... To set the properties of the Real Time Display acquisition mode.
Options......................To set the parameters of the data acquisition, such as exposure time, binning,
acquisition mode and so on….
Auto save...................To set the ‘Save’ parameters to save automatically the data at end of acquisition.
Multi window........... To set the parameters to define the acquisition spectral regions.
Detector.....................To set the detector properties such as readout zone, gain etc.
LabSpec5 manual 20
4.4.1 - Custom function
The Custom function allows to modify the data properties table during the data acquisition.
To access this command select Custom in the Acquisition menu.
To insert or remove properties click on the right mouse button and select Insert Row or Remove Row
menu item.
Check Open Dialog box to display this dialog automatically when the data acquisition process starts.
Start .........................Select the corresponding Use box to see the associated message each time the
acquisition starts. You can customize the message text.
Sample ......................Select the Use box to see the associated message each time when a new sample
point is selected for measurement. You can customize the message text.
LabSpec5 manual 21
4.4.3 - Real Time Display function
The Real Time Display Properties Dialog window allows to modify the data acquisition properties for
the Real time CCD-image readout ( Detector image RTD function) or the spectra readout in
adjustment mode (. RTD function Spectrum )
To access this function select RTD item in the Acquisition menu.
Image binning X ......Set the binning factor in the X direction for the CCD-image readout . This allows to
increase the acquisition speed and the signal intensity level.
Image binning Y ......Set the binning factor in the Y direction for the CCD-image readout. This allows to
increase the acquisition speed and the signal intensity level.
Image zone................ Define here the chip area you wish to load for the RTD CCD-image. You can
choice to read the full chip matrix (select Full) or only the chip area corresponding
to the spectrum readout area (select Spectrum).
Exposition time.........Set here the exposure time value in seconds.. This value appears also in the bottom
control panel.
................... Click on the OK button to validate the settings and to close the window
................... Click on the Help button to retrieve the informations concerning the parameters of
this window.
.................... Click on the Cancel button to close the window without validating the
changes.setting
LabSpec5 manual 22
4.4.4 - Options function
The Acquisition Options window allows to define the data acquisition common properties.
To access to this window select the Options item in the Acquisition menu.
Exposition time...................Set here the exposure time value in seconds. This value appears also in the
bottom control panel.
Accumulation number ...... Set here the number of times you wish to repeat the detector exposure.This is
useful in case of signal overflow. This value appears also in the bottom
control panel.
Data name........................... Set here the data label and the default file name
Refresh time........................This parameter is usefull in case of fast acquisition process or big data sizes.
The data display is not refreshed on the screen until this time is elapsed.
Binning factor.....................Define the number of data points per spectrum.The signal intensity is
proportional to this value, but the number of data point (and spectral
resolution) will be reduced in the same way.
Data mode........................... This parameter is used in case of data accumulation ( Accumulation number
>1.)
LabSpec5 manual 23
* Average... Final spectrum = ( Spectrum1+Spectrum2+....+Spectrumn) / n
Spike filter...........................Allows to remove or not the spike cosmic noise of the CCD detector.
Click on the down arrow icon to display the Spike
filter option list
Then select if you wish to remove the eventual spikes
or not.
* Off........... The autofocus procedure is not used in the data acquisition process.
* Before acquisition. The autofocus procedure is applied before each data readout
* Before each point.. The autofocus procedure is applied at each new measurement point
positionning
* Shift value.............. The shift value allows to adjust the difference between the position
where the laser reflection is at maximum level and the one where the spectroscopic signal is
at maximum level .
Scanning.............................. Allows you to define which scanning device has to be used to record a
spectral image the XY stage or the Scanner.
LabSpec5 manual 24
Use detector ............. On systems with 2 detectors this option allows you to acquire data either from only
the current active detector or from the 2 installed detectors.
............. Click on the OK button to validate the settings and to close the window
................... Click on the Help button to retrieve the informations concerning the parameters of
this window.
.................... Click on the Cancel button to close the window without validating the
changes.setting
LabSpec5 manual 25
4.4.5 - Autosave function
The Auto Save window allows to modify the properties of the acquisition auto save option.
To access this window select Auto Save item in the Acquisition menu.
Save acquired data............. Select this box to save automatically the data when the acquisition process is
finished.
Format.................................Select here the file format. The list of the different format supported is the
same one as for Open/Save functions.
Folder ................................. Edit here the path where to save the files. Click on the button to select
the folder in the tree structure.
Year..................................... Select this box to add the year number to the folder name.
Month.................................. Select this box to add the month number to the folder name.
Day.......................................Select this box to add the day number to the folder name.
Hour ....................................Select this box to add the hour to the file name.
Min ......................................Select this box to add the minute to the file name.
Count .................................. Select this box to add the counter value to the file name. The counter is
modified each time the auto save function is called. But you can edit the start
value of the counter sequence.
LabSpec5 manual 26
4.4.6 - Multi window function
The Multi Window function allows to set the properties of the acquisition in multi window mode. This
mode enables to record data automatically over an extended range with a defined integration time and
averaging. The extended range is covered by taking a number of individual single shot windows and
'gluing' these together..
To acces this Dialog select Multi Window item in the Acquisition menu.
From.......... Edit here the value of the first limit of the spectral window in the current spectral unit. Select
the following to activate the associated spectral window.
To............... Edit here the value of the end limit of the spectral window in the current spectral unit. Select
the following to activate the associated spectral window.
Time........... Edit the relative exposure time(in %) of this window. This value is used to compensate the
differences of signal level.
Min overlap (pix)..... Define here the overlap between the individual windows in pixels. A value of 50
gives generally good results.
SubPixel.....Selecting a value >1 allows to restore the fine shape features. The method is based on the
shifting of the spectrograph position on a distance less than the detector pixel size.
Use multi window.....Select this box to activate the multi window acquisition mode
LabSpec5 manual 27
Auto overlap............. Select this box to automatically define the overlapping size.
Merge data................Select this box to create one data object for each active spectral window. In the
other case a data object is generated for each spectral range corresponding to the spectrograph position.
Adjust intensity........ Select this box to adapt the intensity to the same values on the overlapping parts.
This adaptation is based on the linear shift of the data.
................... Click on the OK button to validate the settings and to close the window
................... Click on the Help button to retrieve the informations concerning the parameters of
this window.
.................... Click on the Cancel button to close the window without validating the
changes.setting
LabSpec5 manual 28
4.4.7 - Detector function
The Detector function allows to set the detector parameters. These parameters are common for any
acquisition process.
To access to this window select Detector item in the Acquisition menu.
➢ Type
Sensor........................ Select the active detector.
Mode..........................Select the readout mode. Single spectrum readout or CCD image mode.
➢ Sizes
Sizes........................... Edit the readout limits of the detector chip. To check the readout zone run the
detector image Real Time Display, by clicking on the icon.
➢ Options
This section groups together parameters specific to the detector model. So this section appears differently
following the model of the installed detector. The most commonly used parameters will be described
hereafter. It is possible that some parameters does not appear if the correponding option is not available
on the on your current detector.
High speed................ Check this box to enable high speed data transfer.
High gain...................Check this box to enable high gain level of the detector amplifier.
LabSpec5 manual 29
Offset ........................Check this box to subtract the dark current offset from the intensity values. Edit the
offset value to subtract in the associated field.
Temperature
From.......... Set the start position (in μm) for the autofocus movement.
To............... Set the stop position (in μm) for the autofocus movement.
Step ........... Set the step size (in μm) for the autofocus movement.
Engine........ Allows you to select the device used for the autofocusing
Auto: The software select automatically between the Z-motor and the piezo following
the size of the movement defined by the From and To parameters.
Z-motor: Force the software to use the Z-motor for autofocussing
Piezo: Force the software to use the piezo for autofocussing.
LabSpec5 manual 30
Z-motor & piezo. The software uses the both devices for autofocusing. A first approach
is done with the Z-motor. The research of the final position is then refined by using the
piezo.
Intensity.....Display the autofocus diode intensity.
➢ Image Scale
The Image scale window allows you to adjust the scale of the video image. These parameters are used
when using an XY scanning device (Stage or Scanner) during the acquisition process. To access this
window run the video preview and then select Scale item in the Acquisition/Video menu.
LabSpec5 manual 31
Corner points Move the XY stage to left-up rectangle corner and click on the point #1. Then move stage
to right-bottom corner and click on the point #2. Program will read the stage position and use this values
as rectangle sizes.
SizeX X size of the rectangle.
SizeY Y size of the rectangle
Stage Click on the button to modify video scale, used by XY stage. This operation
should be done for each microscope objective.
Scanner Click on the button to modify scanner scale.
Clear Clears SizeX and SizeY fields for stage joystick operation.
LabSpec5 manual 32
4.5 - Option menu
The Option menu commands allows to select the default working unit and to convert the active data
object to the selected units when necessary.
To access the Option menu click on the Option button of the main menu bar.
Unit........ Select this command to modify the default spectral and intensity units.
Multi...... Select this option to apply any data treatment operation to all the data of the active window.
LabSpec5 manual 33
4.6 - Setup menu
4.6.1 - Instrument
This function allows you to setup the instrument following the needs of the measurement and the
configuration of the instrument.
Actually this function is available only for the Aramis system.
To access this function select Instrument item in the Setup menu.
➢ Measure Localization
Retro ......... Select this oprion to set the appropriate mirror to capture the Retro-Raman signal
Macro 90....Select this option to set the appropiate mirror to capture the raman signal from the macro
plate.
Micro..........Select this option to set the appropiate mirror to capture the raman signal from the
microscope.
External..... Select this option to set the appropiate mirror to capture the raman signal from the fiber
input.
➢ Polarization
* Laser
Vertical...... Setup the λ/2 slide to obtain a vertical polarization of the laser.
Horizontal. Set in place the optical to retrieve the natural polarization of the laser.
Circular..... Set in place the λ/4 slide to obtain a circular polarization of the laser.
LabSpec5 manual 34
* Raman
Vertical...... Set in place the analyser on the raman path to polarize it vertically.
Horizontal. Set in place the analyser on the raman path to polarize it horizontally.
None........... Remove the analyser to have no polarization of the raman signal.
Scrambler option.
This option is available only if Retro or Macro90 or Micro option is selected.
Select the scrambler option to cancel any polarization effect on the raman signal.
➢ Visualization
Trino off.... Set in place the trino for the selected option.
Trino on
Trino FTIR
➢ Measure
Point mode Set in place the pinhole and the lenses assembly following the selected measure mode.
Line mode
LabSpec5 manual 35
4.7 - Window menu
The Window menu commands permit to arrange automatically the data windows inside the main
LabSpec5 window.
To access the Window menu click on the Window button of the main menu bar.
Functions of the
Window menu
LabSpec5 manual 36
5 - The control panel
The Control Panel located on the bottom of the screen contains divisions which are directly related to the
System configuration. This panel will show only the devices which are installed on the System. The
description below details all of the available sub-units.
5.1 - Laser
If your instrument is full automated ( Aramis instrument) the set of the necessary optical pieces will be
automatically set in place so the chosen excitation light reaches the sample.
In case this part is not automated on your instrument a warning message similar to the following on will
be displayed.
5.2 - Filter
Click here to reset the filter wheel and to set in
place the filter displayed in the filter field.
Note: The reset function of the filter wheel is usefull when for any reason the selected filter is not well in
place or does not match the selected one.
LabSpec5 manual 37
There are 6 neutral filters installed with the optical densities 0.3, 0.6, 1, 2,3 or 4.
[...] = no attenuation (P0)
[D0.3] = P0 /2
[D0.6] = P0 /4
[D1] = P0 /10
[D2] = P0 /100
[D3] = P0 /1000
[D4] = P0 /10000
5.5 - Spectro
Initialize the spectrograph grating
position. Move to the zero order position
Move the spectrograph and then move to the position value
grating to the zero order previously set.
position. (0nm)
LabSpec5 manual 38
5.6 - Options
Click here to select one of the available
spectrograph gratings
Enter here the exposure time value to be used for the spectrum
and spectral image accumulation mode.
Position of the different axis Select the small blank box attached to
of the motorized XY/Z table. each table moving axis to have the
Enter here also the position corresponding position value refreshed in
you want the table to be real time even when moving with the
moved. joystick.
LabSpec5 manual 39
6 - The main icons bar
In this bar you will find all the icons correponding to the LabSpec5 main functions for the spectral data
acquisition, managing and treatment.These icons are grouped together by interest center.
{
{
{
{
{
Data management Display and Data acquisition Basic data High level data
tools information treatment treatment
6.1.1 - CUT
6.1.2 - OPEN
6.1.3 - SAVE
6.1.4 - PRINT
Print to the current printer the active to-print page. For more details please report to the description of the
print , print preview, print setup and page function in the File menu chapter.
6.1.5 - HELP
LabSpec5 manual 40
6.2 - Objects display and information
6.2.1 - Scale normalization
This function reports the limits for each dimension of the active object in the following window.and
allows you to re-define it if necessary.
The Size field reports the number of the spectral points for the correponding axis.
..............Click on the Get button to update the limits of the current active displayed object.
..............Use the Scale button to apply the limits to the active displayed object.
..............Extract the data included in the limits defined in the corresponding fields to build a
new object..
Clicking on this icon displays the array of the acquisition parameters attached to the active object.
LabSpec5 manual 41
6.3 - Data acquisition
6.3.1 - Spectrum Real Time Display
The main purpose of this function is to offer a tool to allow you to adjust quickly the focus and the other
experimental conditions to maximize the Raman signal. Clicking on this icon starts in continuous the
following cycle:
- Acquisition in single shot window of a spectrum.
- Display of the acquired spectrum.
Each spectrum displayed replaces the previously displayed spectrum. There is no averaging or spectra
accumulation and no extended coverage possible. For this use the dedicated accumulation functions
described below.
The exposure time used is the one set in the RTD exposure time field in the control panel (Report to the
exposure time section in the control panel chapter).
The CCD pixels read are the one set in the Acquisition / RTD function.
To stop the continuous readout click on the STOP icon located at right end of the main icons bar.
LabSpec5 manual 42
Variable parameter Results Optional device used
Time Time profile No additional device
Temperature Temperature profile Heating/cooling stage
X and Y position Map Motorized XY table
Motorized XY/Z stage or
X,Y and Z positions Volume
Motorized XY stage and piezo
Clicking on this icon allows you to define the properties of the spectra acquisition over extended spectral
ranges. You can also acces to this function from the main menu bar . Acquisition / Multiwindow. So
please report to this section for detailed explanations.
6.3.7 - Video
Clicking on this icon starts the continuous video image grabbing. To freeze the image press on the STOP
icon located at the right end of the main icons bar.
LabSpec5 manual 43
➢ Manual Baseline Correction
The Manual Baseline Correction method is based on polynomial or line-segmented correction using
referenced points selected by the user.
1. In the Baseline window options select the baseline Type : Line Segmented or Polynomial.
2. In the Baseline window options select the Degree for the Polynomial baseline.
3. Then in the left object associated graphics tools panel select the Baseline tool . Moving the
mouse inside the window of the active objet moves a cross pointer.
This enables you to add or move the points by clicking at the desired place.
By setting the Attach parameter in the options of the Baseline window to Yes the selected baseline
points are attached to the spectrum data.Setting this Attach parameter to No allows you to freely place
the points anywhere in the window.
4. In the Baseline window options use the Style list to select the style to display the baseline calculated
from the selected points.
5. Click on Sub button to subtract the baseline from the data.
This operation could be applied to multidimensional data set, visualized in one-dimensional curve mode
(1D)
LabSpec5 manual 44
The baseline is displayed
in the selected style.
➢ Automatic Baseline Correction.
➢ Baseline Conversion.
The Baseline Correction operation replaces the data by the baseline trace. This
operation allows the baseline trace to be saved as a data file for future retrieving.
Click on the Convert button to start this function.
➢ Baseline Clear.
LabSpec5 manual 45
............. The Norma button normalizes all the traces to same area value (100).
............. The Zero button moves all the traces to the minimum intensity level.
............. The Get button takes the activated spectrum and put it in Corrector frame. This
data object will be used as parameter.
.............The Sub button subtracts the Corrector spectrum from all the traces.
..............The Mul button multiplies all the traces by the Corrector spectrum.
..............The Corr button subtracts the Corrector spectrum from all the traces. The
intensity of the Corrector is multiplied to fit the intensity of the trace
Limits Select the small blank box if you want that only a part of the trace is used for operation. This
parameter is used for Norma and Corr operation.
From and To Edit these fields to modify the limits of the selected region.
6.4.3 - Filtration
This function allows linear and non-linear smoothing and calculates the derivatives of first and second
degrees. The Linear Savitsky-Golay smoothing and derivative computing are based on the convolution
approach which performs a least squares fit of polynomial.
The larger is the Size value and the lower is the Degree value result in a higher smoothing effect.
➢ Degree
Set in this field the degree of the polynomial.
The lower is the degree the more intense is the
smoothing effect .
➢ Size
Set in this field the number of the adjacent data
points to be used for the calculation.
The larger is the Size value the more intense is the
smoothing effect
LabSpec5 manual 46
.............. Click on the Der 2 button to calculate the second degree derivate.
.............. Click on the Median button to apply the median smoothing. The Median
smoothing is a non-linear data filter method.
The Fourier Smoothing is based on the direct Fourier data transformation, applying filter and
apodization function and inverse the Fourier transformation. The dialog window shows the real and
imaginary Fourier functions and allows you then to select the smoothing property.
➢ Limit – Enter in this field the position of the cutoff point in %. This parameter defines the smoothing
factor 0 – full smoothed, 100 – no smoothing. Click on the GO button to apply the value or drag the
pointer to select the Limit visually.
➢ Apod – Select here the type of the apodization function with apodization value reaching zero at a
Limit point.
LabSpec5 manual 47
Traffic .......................Traffic function
For example the “sin(a)” formula replaces the data intensities by sinus function.
The “1000*log(abs(x)+1)” converts intensity to logarithm scale.
➢ Merge function
LabSpec5 manual 48
6.4.6 - Peaks and Bands operations
➢ General presentation
The Peak Fitting operation performs a peak search and curve fitting for a single data trace or a
multidimensional data set. The following list of capabilities is offered by the program:
Click on button on the right side of the Peaks window to call one of the following functions:
LabSpec5 manual 49
➢ Peaks selection
The Peak Edit Properties control the automatic and manual peak correction.
To select and define peaks manually use the following icons on the left graphics tool panel.
LabSpec5 manual 50
• Adjusting peak height and width
Click on this icon to define or modify the height and width of the selected peak.
• Removing a peak
Click on this
icon to remove
peaks.
➢ SEARCH function
The Peak Search operation performs a search for peaks. It iterates all the points of the data trace to
localize the local maxima.
➢ APPROX function
The Peak Approximation procedure can be used to estimate the initial values of the peak parameters.
The height and width of the peaks are modified by this method; the other parameters are not changed. The
height of the peak is replaced by the intensity value at the peak position point. The width is replaced by
the difference in position of the peak and the intensity value of the nearest point, less then half of height.
LabSpec5 manual 51
➢ FIT function
The Peak Fit algorithm uses the Levenberg-Marquart method of non-linear peak fitting. It is based on the
iteratively adjustment of every peak parameter to attempt the minimal value of χ2. The basic problem of
such an approach is that it can be the multitude of possible solutions and depending on the starting values
of peak parameters. To avoid such a situation you should select the position and shape of the peak as
much to presumed solution as possible.
The Approx procedure allows the initial values to be adapted to the intensity of data points around the
peak. The other way is to fix the parameters which are known. You can select the maximum number of
iteration to limit the treatment time. Increasing the number of skip data point allows to check quickly the
convergence of the algorithm and the quality of the initial value selection.
➢ PEAKS... function
The Peak Table dialog shows the peaks parameters. You can edit each of these parameters, add new
peaks and remove existing ones.
p, a, w… columns show the values of the peak parameters. The number of the parameter depends from
the peak shape formula.
LabSpec5 manual 52
p................................. peak position (cm-1)
a..................................amplitude (max. Intensity)
w.................................full width half max (cm-1)
g..................................gaussian contribution (1= max)
s ................................. integrated area of band
Fix ............................. Check this box to fix the parameters during the fitting operation
Map .......................... Check this box to display the map of the parameters. This option is available for
multidimensional data set.
Formula ................... Shows the peak shape formula.
Click on the right mouse button and select Insert row or Remove row menu items to add or remove
peaks.
➢ CLEAR function
Remove all peaks and baselines of the data.
➢ CONVERT function
Convert the sum shape of peaks and baselines to data trace.
➢ VAR... function
The Peak parameters Initialization dialog enables you to select the initial value for the peak
parameters. The parameters are displayed in a spreadsheet. To move inside the spreadsheet use the arrow-
keys.
Edit Name column to select the peak parameter variable name. This name must be the same as in the peak
shape formula. To edit the name double click on the variable name.
Edit Init column to select the initial value of the parameters. You can use number or mathematical
formulae with the the following predefined variables:
x .................................x coordinate of the peak insertion point
y..................................y coordinate of the peak insertion point
minx........................... low limit of x data axis.
maxx.......................... high limit of x data axis.
miny........................... low limit of y data axis.
maxy.......................... high limit of y data axis.
dx................................maxx-minx
dy............................... maxy-miny
LabSpec5 manual 53
Edit Min and Max columns to select the parameter limits. These limits will be used during the fit. You
can use number or mathematical formulae in the same way as for Init.
➢ SHOW... function
The Peak Fitting Show Properties allows you to customize the display of the fitting results in the data
object current window.
➢ SHAPES... function
The Peak Shapes allows to customize the shape of the peak.
Then use the peak parameters initialization function (VAR... button) dialog to set or modify the initial
value of these parameters.
LabSpec5 manual 54
6.4.7 - Profile
The Profile operation performs a visualization of the vertical or horizontal profile of an image.
To start a Profile operation :
Choose Hor or Ver control to select between horizontal and vertical profiles.
Click on the Corr button to replace the image profile. This operation can be useful when you apply some
operations to the profile (for example smoothing) and then you want to set data back to the image.
LabSpec5 manual 55
6.5 - High level data treatment.
6.5.1 - Mapping
The Mapping procedure generates a map by using the signal intensity in the selected spectral regions. Up
to 4 maps can be displayed when you load or acquire the spectral image: the first three are the average
signal intensity and the last is the signal ratio.
The selection of the spectral region is done with the Red, Green or Blue pointer in the spectral image
window.
Check Use box to generate You can edit From and To
the corresponding map. fields to adjust the limits of
the spectral regions.
Check Base box to subtract
baseline.
.................Click on the OK button to generate the map following the settings and close the
mapping window.
................ Click on the Convert button to convert the current map into spectral image and
close the mapping window.
.................Click on the Help button to retrieve the informations concerning the parameters of
this window.
................. Click on the Cancel button to close the Mapping window without without
validating the changes.setting
LabSpec5 manual 56
6.5.2 - New spectral profile
The Spectral Profile operation allows you to create and modify the all traces of spectral profile data
matrix.
................................... Click on the NEW button to take the activated spectrum and create a new spectral
profile . Edit the Name box to define the spectral profile.
............ Click on the Add button to take the activated spectrum and add it to
the end of spectral profile.
................................... Click on the Insert button to take the activated spectrum and insert it to the spectral
profile, selected by cursor pointer.
........... To removes the trace selected by cursor pointer from the spectral profile.
................................... Click on the Help button to retrieve the informations concerning the parameters of
this window.
................................... Click on the Cancel button to close the window without validating the changes
setting.
LabSpec5 manual 57
6.5.3 - Modeling
Click on Modeling icon to open DCLS Dialog.
The Direct Classical Least Squares finds the linear combination of data from pure components making up
the combination that most closely matches the original data. This operation, applied to all traces of the
original data sets, yields the maps of each component. The multiplicative constants used on the pure
spectra are proportional to the concentrations of respective components. The difference between the
sample and the linear combination is the residual spectrum, which could be useful for diagnostic
purposes. DCLS requires the spectra of all components presented in the sample, and assumes that this
spectra are not the changed when the component are mixed together. To apply DCLS mapping :
LabSpec5 manual 58
7 - The graphics tools panel
The graphics Tool Panel on the left part of the screen allows you to modify the data appearance and run
visual edit operations. Following the type of the current data object the tool panel is updated.
You will find hereafter a description of all the tools available.
If the current active data object is a 2D spectral image click on the SpImRed, or SpImGreen or
SpImBlue icon to activate the corresponding pointer.
In any case to initialize the pointer position (move to the window center) click on the pointers centering
icon in the main icons bar.
LabSpec5 offers also the possibility to customize the pointers display by using the pointer function of the
Window context menu.
7.2 - Integral
The Integral Operation can be used to calculate the area of a spectrum peak. Click on the Integral icon
in the graphics tool panel to open the Integral Operation dialog.
LabSpec5 manual 59
From..........................Edit here the the low integral limit. You can also drag the pointer to select the limit
interactively
To...............................Edit here the high integral limit. You can also drag the pointer to select the limit
interactively
Top.............................Shows the integral result without baseline.
Base........................... Shows the baseline value
Full............................ Shows the Top + Base value.
Format...................... Select here the number of digits to display the result numbers
Line ...........................Click on the control to modify the color and line style.
Fill..............................Click on the control to modify the color and style of the background.
Click on the Apply button to apply the From, To, Line and Fill settings.
Click on the Copy button to copy the values to the clipboard. This is usefull to retrieve these
values by using Copy/Paste functions from other softwares.
7.3 - Baseline
Use this tool to define the points to build the baseline to perform a manual baseline correction.
Moving the mouse inside the window of the active objet moves a cross pointer. This enables you
to add or move the points by clicking at the desired place.
Use this tool to remove some points of the baseline. Move the mouse inside the window of the
active object on the point you wish to be removed. When the square cursor appears click on the
left mouse button to remove the wished point.
LabSpec5 manual 60
7.8 - Peak tools
This tool allows you to display the wavelength or the
wavenumber of the selected peaks. The label is displayed
following the style parameters set in the show function of the
Peaks and Bands operations function. This tool is also used to
define manually peaks for the fitting operation.
Use this tool to remove displayed peaks label or to remove bands selected for bands fitting. For
details concerning this tool report to the Peaks and Bands operations chapter.
7.12 - Axes
Use this tool to position and dimension the axes rectangle inside the active window.
LabSpec5 manual 61
7.13 - Change 3D axes.
LabSpec5 manual 62
8 - Window context menu
8.1 - Format
You can open Vizualization Format dialog by selecting Format menu item
➢ Internal Click on this control to modify the color and style of the data field background
➢ External Click on this control to modify the color and style of the window border. Usually it is where
the axes are painted.
➢ Dimension
1D / 2D 3D
➢ Use scroll scale check this box to use the window bottom and right scroll bar
➢ Always show scroll bar check this box to always show the scroll bars, when the visual zone is large
then the data limits.
➢ One data per view if you check this box, the next time data is displayed a new window will be
created.
LabSpec5 manual 63
➢ Y scale = X scale uses the same scale for both axes; useful for 2D image presentation.
Scale .......... shows the scale properties for each of the window axes.
From ..........value of low axis limit
To ...............value of high axis limit
Freeze ........fixes axis limits.
The scale will be skipped when normalization function is applied.
Auto ...........enables auto scaling when data object axis is changed
Prop ...........enables proper scale limits for each data object.
Log .............enables logarithm scale for the axis
8.2 - Norma
8.3 - Colors
The Colors Dialog enables the image color properties to be modified. Such properties are used when
displaying images.
To access this function click on the right mouse button and select Colors menu items.
In the Colors frame you can turn on/off Red, Green or Blue color.
To make image negative choose Neg check box.
LabSpec5 manual 64
8.4 - Axes
The Axes dialog allows you to customize the appearance of the graph axes.
As shown on the picture above there are 4 axis. The Axes Dialog allows you to define independently each
component of each axis. You access to the Axes Dialog by selecting the Axes item of the window
context menu.
➢ Options
LabSpec5 manual 65
➢ Style
Frame ........click on the control to modify the color and style of axis edge.
This parameter is used for a non 3D look.
Main Grid .click on the control to modify the color and style of the main grid
Sub Grid ... click on the control to modify the color and style of the sub grid
Values ........click on the control to modify the values font.
Labels ........click on the control to modify the labels font
Ticks ..........click on the control to modify the color and style of the ticks.
3D Look .... Check this box to enable a 3D outlook for the edge
8.5 - Graph
This function allows you to customize the appearance of 1D dimension displayed data objects.( Spectrum
or Spectral Image)
8.5.1 - Spectrum
If the active object is a spectrum selecting Graph item of the window context menu opens the spectrum
Properties dialog which allows you to modify the display properties of the spectrum.
Lines Points
Bar
LabSpec5 manual 66
8.5.2 - Spectra
If the active object is a spectral image selecting Graph item of the window context menu opens the
spectra Properties dialog which allows you to modify the display properties of the SpIm object.
➢ Click Line control to modify the color, thickness and line type.
➢ Click Fill control to modify the color and fill style of the filled zone.
➢ Click Lines control to select the representation mode of all the spectra. You can select to display All
the spectra , leave the software to display automatically the necessary spectra so it is readable, or as
shown on the picture above not to display all the spectra.
8.6 - Pointers
The Pointers dialog allows you to customize the appearance of the pointers.You access it by selecting
the Pointers item of the window context menu.
➢ Click on the Style control to choose one of the following pointer types:
➢ Click on the Hidden control to choose the line style, when the pointer is deactivated.
➢ Check Transparent box to draw a pointer in transparent style. In such cases the image of the pointer
is superposed with the window background.
LabSpec5 manual 67
8.7 - Image
The Image Properties Dialog enables to modify the appearance of the images data objects. To access
this dialog, select Image item from the window context menu.
Unsmoothed
Smoothed along
line
➢ Click Color to modify the image color. This color will be used, when you select overlay mode for the
data display.
Scale bar
LabSpec5 manual 68
8.9 - Image3D
The Image3D Properties Dialog enables the modification of the appearance of a 3D representation of
the data object. To open this dialog, select the Image3D item from the window context menu.
LabSpec5 manual 69
8.10 - 3DAxes
The 3D Axes Dialog allows you to modify the appearance of the axes for 3D representation of data
objects. You access to this Axes dialog by selecting the Axes item from the window context menu.
➢ Options
➢ Style
Frame....Click on the control to modify the color and style of axis edge.
Values... Click on thes control to modify the font values
Ticks......Click on the control to modify the color and style of the ticks
Labels....Click on the control to modify the labels font
LabSpec5 manual 70
9 - Service
LabSpec5 manual 71