08 Image Quality
08 Image Quality
Improved by:
1 - Masks and
diaphragms
2 - Filters
3 - Lead screens
4 - Potter-Bucky
diaphragm
Film contrast refers to the slope (steepness) of the characteristic curve of the film.
It depends on the type of film, the processing it receives, and the density. It also
depends on whether the film is exposed with lead screens (or direct) or with
fluorescent screens. Film contrast is independent, for most practical purposes, of
the wavelengths and distribution of the radiation reaching the film, and hence is
independent of subject contrast.
Definition refers to the sharpness of outline in the image. It depends on the types
of screens and film used, the radiation energy (wavelengths, etc), and the
geometry of the radiographic setup.
SUBJECT CONTRAST
Subject contrast decreases as kilovoltage is increased. The decreasing slope
(steepness) of the lines of the earlier exposure chart as kilovoltage increases
illustrates the reduction of subject contrast as the radiation becomes more
penetrating. For example, consider a steel part containing two thicknesses, 3/4 inch
and 1 inch, which is radiographed first at 160 kV and then at 200 kV.
Column 5 shows that, at 160 kV, the intensity of the x-rays passing through the 3/4-
inch section is 3.8 times greater than that passing through the 1-inch section. At
200 kV, the radiation through the thinner portion is only 2.5 times that through the
thicker. Thus, as the kilovoltage increases, the ratio of x-ray transmission of the
two thicknesses decreases, indicating a lower subject contrast.
FILM CONTRAST
The dependence of film contrast on density must be kept in mind when considering
problems of radiographic sensitivity. In general the contrast of radiographic films,
except those designed for use with fluorescent screens, increases continuously with
density in the usable density range. Therefore, for films that exhibit this continuous
increase in contrast, the best density, or upper limit of density range, to use is the
highest that can conveniently be viewed with the illuminators available. Adjustable
high-intensity illuminators that greatly increase the maximum density that can be
viewed are commercially available.
The use of high densities has the further advantage of increasing the range of
radiation intensities that can be usefully recorded on a single film. This in turn
permits, in x-ray radiography, the use of lower kilovoltage, with resulting increase
in subject contrast and radiographic sensitivity.
The image on an x-ray film is formed by countless minute silver grains, the
individual particles being so small that they are visible only under a microscope.
However, these small particles are grouped together in relatively large masses,
which are visible to the naked eye or with a magnification of only a few diameters.
These masses result in the visual impression called graininess.
All films exhibit graininess to a greater or lesser degree. In general, the slower films
have lower graininess than the faster. Thus, Film Y would have a lower graininess
than Film X.
The graininess of all films increases as the penetration of the radiation increases,
although the rate of increase may be different for different films. The graininess of
the images produced at high kilovoltages makes the slow, inherently fine-grain
films especially useful in the million- and multimillion-volt range. When sufficient
exposure can be given, they are also useful with gamma rays.
The use of lead screens has no significant effect on film graininess. However,
graininess is affected by processing conditions, being directly related to the degree
of development. For instance, if development time is increased for the purpose of
increasing film speed, the graininess of the resulting image is likewise increased.
Conversely, a developer or developing technique that results in an appreciable
decrease in graininess will also cause an appreciable loss in film speed. However,
adjustments made in development technique to compensate for changes in
temperature or activity of a developer will have little effect on graininess. Such
adjustments are made to achieve the same degree of development as would be
obtained in the fresh developer at a standard processing temperature, and
therefore the graininess of the film will be essentially unaffected.
Another source of the irregular density in uniformly exposed areas is the screen
mottle encountered in radiography with the fluorescent screens. The screen mottle
increases markedly as hardness of the radiation increases. This is one of the factors
that limits the use of fluorescent screens at high voltage and with gamma rays.
PENETRAMETERS
A standard test piece is usually included in every radiograph as a check on the
adequacy of the radiographic technique. The test piece is commonly referred to as
a penetrameter in North America and an Image Quality Indicator (IQl) in Europe.
The penetrameter (or lQI) is made of the same material, or a similar material, as
the specimen being radiographed, and is of a simple geometric form. It contains
some small structures (holes, wires, etc), the dimensions of which bear some
numerical relation to the thickness of the part being tested. The image of the
penetrameter on the radiograph is permanent evidence that the radiographic
examination was conducted under proper conditions.
Codes or agreements between customer and vendor may specify the type of
penetrameter, its dimensions, and how it is to be employed. Even if penetrameters
are not specified, their use is advisable, because they provide an effective check of
the overall quality of the radiographic inspection.
Hole Type Penetrameters
The common penetrameter consists of a small rectangular piece of metal,
containing several (usually three) holes, the diameters of which are related to the
thickness of the penetrameter (see the figure below).
The ASTM (American Society for Testing and Materials) penetrameter contains
three holes of diameters T, 2T, and 4T, where T is the thickness of the
penetrameter. Because of the practical difficulties in drilling minute holes in thin
materials, the minimum diameters of these three holes are 0.010, 0.020, and
0.040 inches, respectively. These penetrameters may also have a slit similar to the
ASME penetrameter described below. Thick penetrameters of the hole type would
be very large, because of the diameter of the 4T hole. Therefore, penetrameters
more than 0.180 inch thick are in the form of discs, the diameters of which are 4
times the thickness (4T) and which contain two holes of diameters T and 2T. Each
penetrameter is identified by a lead number showing the thickness in thousandths
of an inch.
Some sections of the ASME (American Society of Mechanical Engineers) Boiler and
Pressure Vessel Code require a penetrameter similar in general to the ASTM
penetrameter. It contains three holes, one of which is 2T in diameter, where T is
the penetrameter thickness. Customarily, the other two holes are 3T and 4T in
diameter, but other sizes may be used. Minimum hole size is 1/6 inch.
Penetrameters 0.010 inch, and less, in thickness also contain a slit 0.010-inch wide
and 1/4 inch long. Each is identified by a lead number designating the thickness in
thousandths of an inch.
In practically all cases, the penetrameter is placed on the source side of the
specimen--that is, in the least advantageous geometric position. In some instances,
however, this location for the penetrameter is not feasible. An example would be
the radiography of a circumferential weld in a long tubular structure, using a source
positioned within the tube and film on the outer surface. In such a case a "film-
side" penetrameter must be used. Some codes specify the film-side penetrameter
that is equivalent to the source-side penetrameter normally required. When such a
specification is not made, the required film-side penetrameter may be found
experimentally. In the example above, a short section of tube of the same
dimensions and materials as the item under test would be used to demonstrate the
technique. The required penetrameter would be used on the source side, and a
range of penetrameters on the film side. If the penetrameter on the source side
indicated that the required radiographic sensitivity was being achieved, the image
of the smallest visible penetrameter hole in the film-side penetrameters would be
used to determine the penetrameter and the hole size to be used on the production
radiograph.
Sometimes the shape of the part being examined precludes placing the
penetrameter on the part. When this occurs, the penetrameter may be placed on a
block of radiographically similar material of the same thickness as the specimen.
The block and the penetrameter should be placed as close as possible to the
specimen.
Wire Penetrameters
A number of other penetrameter designs are also in use. The German DIN
(Deutsche Industrie-Norm) penetrameter (See the figure below) is one that is
widely used. It consists of a number of wires, of various diameters, sealed in a
plastic envelope that carries the necessary identification symbols. The thinnest wire
visible on the radiograph indicates the image quality. The system is such that only
three penetrameters, each containing seven wires, can cover a very wide range of
specimen thicknesses. Sets of DIN penetrameters are available in aluminum,
copper, and steel. Thus a total of nine penetrameters is sufficient for the
radiography of a wide range of materials and thicknesses.
On the other hand, the hole penetrameter can be made of any desired material but
the wire penetrameter is made from only a few materials. Therefore, using the hole
penetrameter, a quality level of 2-2T may be specified for the radiography of, for
example, commercially pure aluminum and 2024 aluminum alloy, even though
these have appreciably different compositions and radiation absorptions. The
penetrameter would, in each case, be made of the appropriate material. The wire
penetrameters, however, are available in aluminum but not in 2024 alloy. To
achieve the same quality of radiographic inspection of equal thicknesses of these
two materials, it would be necessary to specify different wire diameters--that for
2024 alloy would probably have to be determined by experiment.
Special Penetrameters
Special penetrameters have been designed for certain classes of radiographic
inspection. An example is the radiography of small electronic components wherein
some of the significant factors are the continuity of fine wires or the presence of
tiny balls of solder. Special image quality indicators have been designed consisting
of fine wires and small metallic spheres within a plastic block, the whole covered on
top and the bottom with steel approximately as thick as the case of the electronic
component.
In the case of a wire image quality indicator of the DIN type, the visibility of a wire
of a certain diameter does not assure that a discontinuity of the same cross section
will be visible. The human eye perceives much more readily a long boundary than it
does a short one, even if the density difference and the sharpness of the image are
the same.
For routine viewing of high densities, one of the commercially available high-
intensity illuminators should be used. These provide an adjustable light source, the
maximum intensity of which allows viewing of densities of 4.0 or even higher.