Operation Manual: Transmission
Operation Manual: Transmission
OPERATION MANUAL
TABLE OF CONTENTS
TOPIC PAGE
1 GENERAL
Introduction I-I
Failure Messages I-I
Front Panel I-I
Front Panel Reference Guide I-I
3 SETUP
DC Power Option 3-1
Power On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 3-1
Setting The Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Test Setup 3·1
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
8 NOISE·WITH·TONE MEASUREMENT
Notch Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
,
Worst IS Minutes 9-1
Impulse Noise 9-1
10 P/AR MEASUREMENTS
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-1
'-...../
Measurements " 10-1
P/AR Systems 10-1
Transmit Procedure 10-1
Measurement Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-1
11 DIAL PROCEDURE
12 FAILURE MESSAGES
13 SPECIFICATIONS
INDEX
"-../
ISS 2, SECTION 1125 - 301
TABLE OF CONTENTS (CONn
LIST OF ILLUSTRATIONS
FIGURE PAGE
TABLE/CHART PAGE
1·A Front Panel Reference Guide 1-3
1.01 Introt.Wctioll: The TTl 1125 Transmission 1.03 Frollt JHlMl.: All operator commands to the
Test Set is a multifunction instrument design- 1125 are made through an entry keypad.
ed for measuring transmission impairment with VF, Functiol18 on the front panel are numbered, and the
program, and wideband telecommunications circuits. status of all selectable functions is indicated by light·
emitting diodes (LEOs).
(a) An internal oscillator provides calibrated test
signals. The unit features full duplex circuit 1.U Throughout this section, references to front
interface, ac operation, and a real-time clock with panel functiol18 are designated by brackets
recall of the worst IS-minute period of impulse around the numeric code corresponding to the func·
noise. tion, i.e., [13] refers to the TRANSMIT function,
while [30] is FREQUENCY SET.
(b) A dc power option is available.
1.05 &f. .race guide: Table l·A lists each section
(c) A P/AR (Peak-to-Average Ratio) option is of the front panel, and the selections available
available. in each. Item numbers in the left-hand column refer
to Fig. I-I. Additional explanations of front panel
(d) Appendix A lists the operating specifications. selections and procedures are referenced in the table
and start in Topic 2.
1.02 Failure "..Nage,: The 1125 uses an onboard
microprocessor to control all internal cir-
cuitry and to initiate a self·diagnostic routine when
power is applied. Malfunctions are indicated by
failure messages in the display.
~~
...-
•• _S£I • • " ,... c:au.TWnl •• OIS'"." iliff crs •• ,_
... .0_ •• _ lilt"' Cl'S
.a I.... .11 • II lMllIT . . . . - T 11_ • • 045'lAY .11 11_
••••
•• c.- •• ,... •• nn_ ... "!IT """.... en ...._
.l.... •
•••
IIITal . . . _ . "00 I01lI , _ "... • II ea-T
• I . Nt lOSS . . , . . , _ II Q,J 10
r ,_IT ) ~r;;~;;;;;JR~@V£~!k;~;;;;;;~) ~
• ~
••••
lilt ......
...
"" ......
•
"--.--/
TAlLE l·A
FRONT PANEL REFERINCE GUIDE
,
ITIM
SICTlON SILICTlON' IIPI.INCI
20 LEVEL dBm
21 N·G dBm
22 CMSG
23 3 kHz
1 LeveUNoise 2.11
24 15 kHz
25 PROGR
26 50 KB
27 NOTCH
10 TMS
2 Measure 11 IMP N 2.10
12 P/AR
13 TRAN
3 Display 2.07
14 REC
COUNTING
4 IMP N Status 1004 Hz PRES 2.14
1004 Hz LOSS
30 FREQ SET
31 LEVEL SET
32 SF SKIP
5 Oscillator 33 404 Hz 2.12
34 1004 Hz
35 2713 Hz
36 2804 Hz
6 Left Display 2.08
CLOCK
7 Left Display (Function) dBm 2.08
dBm
40 TOTAL COUNTSITH
41 WORST 15 MIN
42 REF SET
8 Impulse Functions 43 IMP N W/O TONE 2.13
44 1004 Hz LOSS RESET
45 DISPLAY REF CTS
46 DISPLAY MID CTS
'--./
,
ITIM
SICTlON S.LICTION' R•••R.NCI
47 DISPLAY HI CTS
48 125 ms BLANK
49 10 ms BLANK
Impulse Functions 50 5 MIN
8 (cont) 2.13
51 15 MIN
52 60 MIN
53 CONT
54 CLOCK SET
135
150
600
16 Receive 2.04
900
1200
DIAL
17 Binding Posts 2.03
135
150
600
19 Transmit 2.04
900
1200
DIAL
20 Speaker 2.15
I ---..J
Page 1-4 ISS 2, SECTION 1125 . 301
FRONT PANEL FEATURES TOPIC 2
EDmON 1
(b) Unit automatically switches to battery power (1) Values change by + 100 Hz in the FRE-
in the event of ac power failure. QUENCY SET mode. Hold the key in for
continuous incrementing/decrementing.
2.02 TRANSFER SWITCH (Item 14): Controls
the direction of test operations. When the (2) Values change by +0.1 dBm in LEVEL
transfer switch is in, the A binding posts and 310 jack SET mode.
are connected to the transmit side of the circuit, and
the B binding posts and 310 jack are connected to the (3) Values change by + 1 dBrn in REFER·
receive side of the circuit under test. With the ENCE SET mode.
transfer switch in the out position, the A connections
are connected to the receive side of the circuit and the (b) The + j- key is used when entering positive
B connections are connected to the transmit side of values. A default negative value is assigned to
the circuit. all level inputs unless this key is depressed.
2.03 BINDING POSTS (Item 11): Labeled T (Tip) 2.07 DISPLAY Section (Item 3): LEOs indicate
and R (Ring). Used to connect a butt-in whether the measurements on the TRANS-
handset to the telephone circuit connected to the 310 MIT [13] or the RECEIVE [14] side of the circuit
jack. Refer to Topic 11 for the dial-up procedure. under test are being displayed.
2.04 RECEIVE SWITCH (Item 16) and 2.01 LEFT DISPLAY (Item 6): Will display a
TRANSMIT SWITCH (Item 19): Matches the value as determined by the mode selected;
1125's terminating impedance to the characteristic LED to the right of the display will indicate the unit
impedance of the line. When set to the DIAL position, of measurement.
the internal measuring circuitry (including line
holding) is disconnected and a butt-in handset may be (a) TMS Mode:
used to dial and hold through the 310 jack to the line,
via the binding posts. (1) Transmit signal level in dBm
(b) IMPULSE NOISE Mode: 2.11 LEVELINOISE S«tioll (Item 1): Operator
may select the desired filter or function by
(1) 24-hour real time while test is in progress, pressing the ENTER key, followed by the
or when CLKSET 54 is selected numeric code of the desired filter or function.
2.09 RIGHT DISPLA Y (Item 9): Will display a 24 15 kHz (15 kHz flat filter) (Note 2)
value as determined by the mode selected; the
LED to the right of the display will indicate the' unit 25 PROGR (Program filter) (Note 2)
of measurement.
26 50 KB (50 kBit wideband filter) used for
(a) TMS Mode: Receive or transmit signal fre- testing digital transmission local loops, in-
quency in kilohertz. cluding DSO level.
(b) IMPULSE NOISE Mode: Displays counts 27 NOTCH (1010 Hz notch filter)
(hits) above threshold. Either total counts per
threshold, worst 15·minute counts, or REF, MID Note 1: If both the LEVEL LED and the N-G LED
or HI counts (depending on selection) may be are off, then the 1125 is in the NOISE mode (for "-.J'
displayed. measuring metallic noise).
(c) PIAR Mode: Receive or transmit P/AR Note 2: Characterization of noise filter requires a
measurement reading in P/AR units. continuous signal at a preferred reference level of
-10.0 dBm.
2.10 MEASUREMENT S«tioll (Item 2): Opera-
tor selects the measurement to be made by 2.12 OSCILLATOR Section (Item 6): Test opera·
pressing the ENTER key on the keypad, followed by tor selects the level and frequency of
the numeric code of the desired measurement type. oscillator functions by entering the actual frequency
and level desired, or by entering the numeric codes of
10 TMS (Transmission Measurement Set) the fixed frequencies provided.
---../
Page 2·2 ISS 2. SECTION 1125·301
FRONT PANEL FEATURES TOPIC 2 (Cont)
EDmON 1
2.13 IMPULSE FUNCTIONS Section (Item 8): 49 10 ms BLANK: Disables counting for 10 ms
With the 1125 in the IMP N mode and set after a hit.
to RECEIVE, the operator may set the threshold for
measuring hits and select the total time for making 50 5 MIN: Length of measurement period is 5
measurements by pressing the ENTER key, followed minutes.
by the desired function. An audible signal will sound
on each occurrence of a noise spike. Tone volume can 51 15 MIN: Length of measurement period is 15
be adjusted with the volume control (Fig. 1·1, Item minutes.
12).
52 60 MIN: Length of measurement period is 60
40 TOTAL COUNTS/TH: The 1125 will count all minutes.
hits at or above the preset threshold value.
53 CONT: Continuous measurement.
41 WORST 15 MIN: The 1125 will recall from its
memory a record of the worst 15-minute 54 CLOCK SET: Used to set and/or display the
measurement period, along with the real time of real time.
occurrence. The 1125 allows a maximum of 255
counts per threshold during anyone minute; if ex- 2.14 IMP N STATUS (Item 4): Impulse noise
ceeded, display will read "OR15". measurement status is indicated by three
LEDs.
42 REF SET: Enables the operator to set the
reference value (threshold) for impulse noise (a) Counting LED: Lit during the counting inter-
measurements. vals of 5 minutes, 15 minutes, 60 minutes, and
continuous. The LED will light when the start but·
43 IMP N W/O TONE: Impulse noise without ton is pressed, and go out after the selected time
tone. interval has elapsed.
44 1004 Hz LOSS RESET: Resets and arms the (b) 1004 Hz PRES: Lit when a 1004 Hz tone is
1004 Hz loss indicator LED (see 2.14). present, and go out when the tone goes below
- 40 dBm for more than 125 mao The LED will
45 DISPLAY REF CTS: The total number of relight when tone reappears for more than 125 mao
hits above the reference threshold is
displayed. (c) 1004 Hz LOSS: Lit if a 1004 Hz tone is not
present, or if the 1004 Hz baa dropped below
46 DISPLAY MID CTS: The total number of hits - 40 dBm for more than 126 ms. To tum off the
4 dB above the MID threshold is displayed. 1004 Hz LED:
3. SETUP 3.04 Setting the clock: Set the real time clock by
entering the following sequence:
3.01 This topic describes setup procedures for the
use of the 1125 Transmission Test Set. ENTER 5 4
Specific procedures for a variety of tests may be
found in Topics 4 through 11. The time should appear on the left display, and the
CLOCK SET LED should blink. Time is displayed on
3.02 DC /IOUWI' option: Units equipped with the dc a 24-hour basis (3 p.m. is displayed as 15.00). If the
power option operate on 110 V ac, with a time displayed is incorrect, enter the correct time at
battery back-up system. If the unit is connected to an the keypad, then enter ENTER 5 4 again to exit
ac outlet, the battery is charging. The 1125 may be from the CLOCK SET mode. The CLOCK SET LED
used on battery power for up to 3 1/z hours. A low bat- will stop blinking. Note that the clock circuitry is
teryindicator on the front panel (Item 11, Fig. 1·1) in· powered by battery even when the 1125 is turned off
dicates operational time is limited to 15 to 30 minutes; and unplugged.
the unit should be connected to ac power to fully
charge the battery. 3.01 Tut anup: Set up the unit for testing a cir-
cuit:
(a) If the unit must be left-on while charging, it
will receive a trickle charge. (a) Connect a 310 cord from the circuit under test
to the front panel jacks on the 1125. Press the
(b) With the unit turned off, battery will be fully transfer switch to the appropriate position:
charged in approximately 16 hours.
(1)If the receive side of the circuit under test
3.OS POlDer on: Pull the ON/OFFNOL switch out is to be connected to the B 310 jack, the
to turn the unit on. TRANSFER switch should be in.
(a) All the front panel LEDs and all segments of (2) If the receive side of the circuit under test
both displays should light. No error codes is to be connected to the A 310 jack, the
should be displayed. TRANSFER switch should be out.
(1) The transmit signal is automatically looped (b) Set the TRANSMIT switch (Item 19, Fig. I-I)
through the filters to the receiver section. to the position which matches the impedance
of the transmit side of the circuit under test.
(2) The output of each filter is checked to
verify that the signal is within specified (c) Set the RECEIVE switch (Item 16, Fig. I-I)
parameters. "Out-of-spec" signals generate er- to the position which matches the impedance
ror codes. Refer to Topic 12. of the receive side of the circuit under test.
(b) The 1125 will default to TMS[10], TRAN[13], (d) If the RECEIVE circuit is to be measured in
and LEVEL{20], at 0.0 dBm and 1.004 kHz. the BRIDGED mode, push the BRIDGEI
TERM switch (Item 16, Fig. I-I) so that it is in the
(c) The operator may exit from any function out position.
mode by entering any other function.
3.06 Instructions for performing tests begin in
Topic 4.
4. TlANSMlml LEVEL AND FIEQUENCY 4.02 Cluuaging level 4ntl freq,"MII: Chart 4·B
shows the steps used to set or change the level
4.01 ll1ititJl Ntup: Chart 4·A shows the keypad en· and frequency of a signal being transmitted by the
tries used to set up the 1125 to display the 1125.
level and frequency of the signal being generated by
the test set.
CHAIT 4·A
TlANSMmEI LEVEL AND FIEQUENCY
O,DATOI ACTION
fUNCTION INDlCATOa
(KIYSTIOKU)
CHART A·I
CHANGING LEVEL/FREQUENCY
Nota:
1. Refer to 2.06 to increment or decrement the value entered.
2. The most frequently changed parameter should be entered last to reduce the number of
operator keystrokes.
3. Values entered are negative unless the + I - key is depressed. Depress the + I - key only when positive
values are to be entered.
5. LEVEL AND 'REQUENCY MEASUREMENTS 5.03 SiglltJl level rtuage.. If the signal level is
greater than + 10.3 dEm, an over range (Or)
5.01 Monitoring level and !nqruM1/: The se- symbol will appear in the left display. If the signal is
quence given in Chart 5-A allows an operator less than -50.0 dEm, an under range (Ur) symbol will
to monitor the level and frequency of a signal on a cir- appear.
cuit being tested by the 1125.
5.OA FrequeM1/ rollfle: If the frequency is less
5.02 The level of the signal will be displayed on the than approximately 20 Hz, LO appears in
left display; the frequency of the signal will the right display. If the frequency is higher than
appear on the right display. about 110 kHz, HI appears in the right display.
CHART 5·A
LEVEL AND FREQUENCY MEASUREMENTS
1 Set transfer
switch as required -
5.05 Entering Quiet Termination Mode: The 5.06 Exiting Quiet Termination Mode: The quiet
quiet termination mode can be entered from termination mode is released from either a
either a transmit level or noise level function by transmit level or transmit noise mode by following
following the steps in Chart 5-B. the steps in Chart 5-C.
CHART 5·8
ENTERING QUIET TERMINATION MODE
1 ENTER Transmitter is still on until re- Transmit level will display OFF. Frequency is
ceive mode is entered, or the blanked.
frequency is changed.
2 ENTER 1 4 Receive mode. Transmitter is REC LED lit.
turned of(and stays off until a
level set operation is performed.
CHART 5·C
EXITING QUIET TERMINATION MODE I
STEP ACTION FUNCTION INDICATORS "---'
NOTE: Quiet termination u dropped after any valid level u entered during level aet operation.
.~
6. MITALLIe NOISI MEASUREMENTS 6.02 Metallic noise level (10 to 100 dBm) will aIr
pear in left display. Ur indicates the noise
6.01 Procedure: Metallic noise level in dBm may level is less than 10 dBm.
be measured by entering the sequence given
in Chart 6-A (Note 1). Note 1: If both the LEVEL LED and the N·G LED
are off, the 1125 is in the (metallic) noise mode. The
noise filter in use is indicated by the illuminated LED;
2.11 lists noise filters available.
CHART 6·A
METALLIC NOISE MEASUREMENTS
1 Set TRANSFER
SWITCH as required
2 ENTER 1 0 TMS mode [10] [10] TMS LED lit
3 ENTER 1 4 Receive [14] side [14] REC LED lit
CHAIT 7·A
NOISE·TO·GROUND MEASURIMENn
I. NOISE·WITH·TONE MEASUREMENT 1.02 Procetlun: Chart S·A shows the keypad se-
quence used for initiating noise-with·tone
1.01 Notela filter: The 1125 is equipped with a 1010 measurements.
Hz filter for making noise with tone (notch
noise) measurements. 1.03 Read the noise level (10 to 100 dBm) on the
left display.
CHART I·A
NOISE·WITH·TONE MEASUREMENTS
CHART '·A
IMPULSE NOISI MIASUREMENTS
,
REF SET [42] LED
5 ENTER 4 2 Set reference level should blink
"--'
Pageg·2 ISS 2, SECTION 1125·301
IMPULSE NOISE MEASUREMENTS TOPIC 9 (Cont)
EDmONt
...
III IINGING IlANKID
>
...
~
III
C
10-
D h
7J
}~U NTASHm
TIM.
10. PIAl MIASUIEMENn (b) The receiver measures the peak and average
values of the received signal, and displays
10.01 Purpo.e: The optional P/AR system is design- their ratio, which is the basis for the PI AR rating.
ed to measure the simultaneous effects of
bandwidth reduction, envelope delay distortion, and (c)A P/AR rating of 100 signifies that the pulse
return 1088 due to gain and phase ripples on voiceband train measured has suffered no degradation,
data signals. compared to the pulse train transmitted.
10.02 Meauremenu: P/AR measurements are not (d) P/AR specifications are listed in Topic 13. Ad·
affected by transient phenomena, frequency ditional information on PI AR systems
shifts, or noise. is in Bell System Publication 41009.
10.03 PIAR 'I/.tem: The P/AR system consists of a 10.lW '1'rtJIqmit proeedure: To transmit a PI AR
transmitter and a receiver connected to op- signal, use the sequence given in Chart 10·A.
posite ends of a voiceband transmission system.
10.01 Meaurement procedure: To measure a PI AR
(a) The transmitter generates a complex line signal, use the sequence given in Chart 10-B.
signal with no even harmonics, producing a
pulse train with half·wave symmetry. The peak.to-
average ratio of the signal is known.
CHART 10·A
TIANSMmlNG PIAl SIGNALS
CHART 10·.
MIASUIING '1 AI SIGNALS
.-
11. DIAL PaOCIDUII (d) Use the butt-in handset to dial the far end of
the circuit.
11.01 The 1125 has the ability to hold wet dial-up
lines. Use the following sequence: (e) After the circuit is answered, switch to the
desired impedance to continue testing.
. (a) Connect patch cord from circuit under test to ,
the A or B 310 jack. Refer to Fig. 11-1 fori (f) Disconnect the butt·in handset (preferable) or
butt-in and circuit connection to the 1125. I return its mode switch to the "monitor" posi·
tion (Not. 1).
(b) Connect a butt-in handset to the binding posts
of the selected 310 jack. (g) To disengage, put the switch back in the
DIAL position.
(c) Set either the TRANSMIT or the RECEIVE
switch, 88 appropriate, to the DIAL position. Not. 1: The bridging 1088 caused by leaving the
Seize the circuit by switching the handset from handset connected may degrade the measurement
"monitor" to "talk." of a tone on the circuit.
------'
- •• .~ 0 ..··
'I
.O~_ -=tf-.~
_.
-1l:J:1Ill~==11.'
_1._.
.. ..••••
II
II·· ....
'I .-.-
'I
I
:
e
_.. - •• -..••
•.
,.
-......... cu.,
..
• • • .,
. . . _ _.~ ....
!!fCIM
0I .0--, r-,_ _
••
_ ,
I
. .•.•. ._ . ......
.1•m• •
1m.....
•
m
,.
•••
.'..
~
'~
!
I
( DAti EitIt
••••
! _ •
1 !1\MMf II •
I-I! 11 . .
·_·'
. _00','00<,
"c•
.'.."" -,
'
I '00
'".
ill
""'"
" : OI ..UAL-'O......
'ILICO."".IC
O
310
ItATCH CORD
I
'-...-J
'---"
Page 11·2 ISS 2, SECTION 1125· 301
FAILURE MESSAGES TOPIC 12
EDmON 1
TAILI12·A
FAILURI MISSAGU
24 15 kHz filter
25 PROG filter
26 50 kB filter
27 1004 Hz notch filter
RECEIVEI
F.....uency Mealuremenhz
Lev" Mealuremenhz
Nol.. Mealuremenhz
"-./
Transmltt.r:
L f
E + 10
V +3
-~ -w -~ -40
+6 '-J
E -~ -~ -w -~ -~
L -30
d -~ -~ -~ -~ -~
-40
B 60Hz 100Hz 200Hz 4kHz 85kHz 110 kHz
m
FIEQUENCY
-Measured per Pub 41009, Table L
PIAl (optIonal):
General:
Acceuorf...