AEC
AEC
Rob McLeod
2
4.6 CR response as a function of kVp ................................................................ 40
4.7 The CR response as a function of dose......................................................... 40
4.8 Image processing with the Kodak CR reader ............................................... 42
4.8.1 The DICOM standard data interface .................................................... 44
4.9 Working with the processed image............................................................... 45
4.9.1 CR PSP erasure process ....................................................................... 45
4.9.2 Image storage and retrieval: PACS ...................................................... 46
4.10 Noise characteristics of CR imaging .......................................................... 46
4.12 Processing time delay effects on exposed IP .............................................. 48
4.13 Comparison of screen-film and CR ............................................................ 50
Chapter 5 AEC in a CR Environment ....................................................................... 53
5.1 Introduction................................................................................................... 53
5.2 AEC optimisation with CR ........................................................................... 53
5.2.1 AEC dose set-point with CR ................................................................ 54
5.2.2 Optimisation procedure for AEC with CR........................................... 55
5.3 Determination of AEC EI test limits with CR .............................................. 56
5.4 X-ray beam standardisation for test functions .............................................. 57
Chapter 6 Developing a test procedure to optimise AEC performance with CR .. 58
6.1 Introduction................................................................................................... 58
6.2 The selection of Ionisation Chamber ............................................................ 58
6.3 Profile of the Inovision ion chamber ............................................................ 59
6.4 Ion chamber test setup .................................................................................. 60
6.4.1 Minimising back scatter in the test setup................................................... 62
6.5 X-ray Bucky arrangement in test set-up ....................................................... 64
6.5.1 The anti-scatter grid ................................................................................... 64
6.6 The test procedure and measurement accuracy ............................................ 66
6.7 Summary....................................................................................................... 66
Chapter 7 Test results : the rationale for an optimised AEC .................................. 68
7.1 Introduction................................................................................................... 68
7.2 Specific test considerations........................................................................... 68
7.3 Test results for an AEC with generic kVp compensation applied for CR.... 69
7.4 Test results for an AEC with optimised kVp compensation applied for a CR
imaging system ................................................................................................... 71
7.5 Evaluation of Dose and EI responses ........................................................... 73
3
Chapter 8 Image Quality comparison of AEC setups............................................... 74
8.1 Image Quality testing.................................................................................... 74
8.2 Contrast-detail test ........................................................................................ 74
8.2.1 Contrast-detail test description............................................................. 75
8.2.2 Contrast-detail test results .................................................................... 76
8.3 The spatial resolution limit test..................................................................... 77
8.3.1 Spatial resolution limit test description................................................ 79
8.3.2 The spatial resolution limit test result .................................................. 80
8.4 A comparison of image noise between AECs with generic kVp
compensation and AECs with optimised kVp compensation............................. 81
8.5 Image quality assessment ............................................................................. 83
8.5.1 Clinical confirmation of image quality ................................................ 83
Chapter 9 A protocol to test AEC functionality by the measurement of dose
alone.............................................................................................................. 85
9.1 The advantages of AEC optimisation ........................................................... 85
9.1.1 Applicability of the AEC optimisation procedure ............................... 85
9.2 The impact of a new AEC testing protocol on the working environment of
the Medical Physicist .......................................................................................... 86
9.2.1 Test time reduction for AEC assessment using a dose alone protocol 87
9.3 General conclusions of thesis ....................................................................... 88
9.4 Final Conclusions and future work............................................................... 90
9.5 Formal Statement.......................................................................................... 90
References ..................................................................................................................... 92
Index of Figures............................................................................................................ 92
Index of Tables ............................................................................................................. 97
Acronyms ...................................................................................................................... 98
4
Abstract
An Experimental Approach to AEC Testing
A New Zealand Qualified Health Physicist (QHP) is required, under the Radiation
Protection Act of 1965 and the Radiation Protection Regulations 1982, to perform
auditing compliance tests on x-ray equipment at regular intervals to ensure that this
equipment conforms to the Code of Safe Practice with the use of x-rays. The protocol
for these tests must be approved by the National Radiation Laboratory (NRL).
One of these test protocols sets out the requirements for the functionality testing of the
x-ray machine Automatic Exposure Control (AEC). The current NRL protocol for
AEC testing is based on the radiographic film environment (NRL C5 1994). This
protocol was tested to determine its applicability to the digital computed radiography
(CR) imaging systems which are replacing screen-film systems. To begin this process
a comparison of the different exposure indexes for each image medium was required.
This proved to be achievable using a system of exposure dose comparison. The AEC
test process for both image modalities follows identical requirements but differ
slightly in the test methods used to achieve these. The most significant finding
throughout this stage was not the differences between protocols but was the
requirement to achieve consistent exposure index values over the clinical kVp range
for each image medium. This requirement, applicable to any x-ray image medium,
became the focus of this thesis.
The thesis has explored through experimentation, the effect of optimisation of AEC
kVp compensation for the variable kVp response of an image medium, on image
consistency. At Christchurch Hospital where this investigation took place the work has
shown that the performance of AEC devices can be optimised to improve image
consistency, indicated by a more consistent exposure index over the clinical kVp
range. The optimisation process also achieves a more consistent dose response to the
image plate. A dose variation of 8.3% from the average was achieved compared to
26% in the unoptimised version. No clinically significant changes to image quality
were apparent in test images. Under these conditions it was found that AEC
functionality could be assessed solely by the measurement of AEC dose to the image
plate (IP). Use of this test method provides quantifiable time management benefits for
the Medical Physicist and for the radiology departments in which they work.
5
Disclaimer
To the best of my knowledge and belief this thesis contains no material which has
been accepted for the award of any other degree or diploma in any university and
contains no material previously published or written by another person except where
due reference is made in the text of the thesis.
I give consent to this thesis being made available for photocopying and loan, if
accepted for the award of the degree.
Signed:
Acknowledgments
I give thanks to the Canterbury District Health Board for the opportunity and their
support to enable me to take part in the MSc program.
This work acknowledges the help of all the people in my work group in the Medical
Physics and Bioengineering Department at Christchurch Hospital who have given
freely of their time and ideas to aid me in this project.
I also acknowledge the assistance given to me from the radiology staff at Christchurch
Hospital who have accommodated me in their work domain during the times when I
needed to spend many hours in their x-ray rooms collecting the experimental data
required for this thesis.
I thank John Le Heron from NRL in his capacity as my external supervisor for part of
my candidature and to Associate Prof Tim Van Doorn and Dr Judith Pollard as
principal supervisors and administrative advisers from the University of Adelaide.
I thank the engineering staff of Philips Medical who assisted me in setting up an AEC
to conform to the requirements proposed by this thesis.
7
Chapter 1
Introduction
A New Zealand qualified health physicist (QHP) is required under the Radiation
Protection Act of 1965 and the Radiation Protection Regulations 1982, to perform
auditing compliance tests on x-ray equipment at regular intervals to ensure that this
equipment conforms to the Code of Safe Practice with the use of x-rays. The protocol
for these tests must be approved by The National Radiation Laboratory (NRL), a
specialist business unit within the Ministry of Health based in Christchurch, New
Zealand. NRL provides a resource of expert advice, service provision and research
capability on matters concerning public, occupational and medical exposure to
radiation, the performance of radiation equipment, and the measurement of radiation
and radioactivity. NRL also maintains links to international bodies concerned with
radiation protection and associated health issues.
One of the test protocols required by NRL sets out the requirements for testing x-ray
machine Automatic Exposure Control (AEC) functionality. The current NRL protocol,
NRL C5 section 5.28 to 5.31 in New Zealand, is based on AEC testing in a film
environment. With the phasing out of radiographic film and film processors, in the
wake of the move into the modern age of digital x-ray detectors and digital image
processing, it is an appropriate time to consider the relevance of existing AEC testing
protocols (Huda 1997).
AEC devices are designed to automatically terminate an x-ray exposure when their
detectors have recorded that sufficient radiation has been received by an image
medium to ensure a high quality x-ray image of the subject (Hale 1989). A correctly
operating AEC system ensures that consistent imaging quality will result given the
same exposure conditions. “As a tool to help ensure correct exposure levels, properly
functioning and calibrated automatic exposure control systems may be even more
important for Computed Radiography (CR), an example of Digital Radiography (DR),
than for film based receptors because there may not be any other technique or
stabilising influence as effective as ‘acceptable film optical density’ is for film”
(Goldman and Yester 2004).
The work for this thesis evolves from a historical view of the development of
diagnostic x-ray technology to the development of a procedure to assess AEC
functionality using the measurement of dose alone, as described in the following
chapter reviews.
Chapter 2 begins with a developmental timeline of x-ray devices and x-ray exposure
control. It includes the development of x-ray production and the early understandings
of image quality relating to x-ray tube design. Radiation safety issues soon became
apparent to the early pioneers of x-ray, with the effects of radiation induced cancers.
Radiation safety is an important part of the use of any x-ray devices and underpins
much of their developmental work including the use of AECs. This section is followed
by the development of x-ray measurements and measurement units used to quantify x-
ray dose administered to patients.
In an attempt to simplify the process of AEC functionality testing and to reduce the
time required to perform and assess AEC system performance, a number of test
methods and devices have been developed. Some of these, relating to radiographic
screen-film imaging, are discussed and evaluated in this chapter.
The final section in this chapter describes how the results of the AEC test procedure
are processed and analysed to form the basis of the AEC assessment.
9
Chapter 4 describes the increasing use of digital technology and lists some of the types
of digital devices available for use with digital imaging. Following this is a description
of CR technology as an example of a digital imaging system. This leads to a series of
sections which details the response of CR with x-rays along with the mechanisms and
attributes which are characteristic of CR.
The chapter is designed to lay out the particular elements of AEC operation within the
CR environment. Similar to the procedure for screen-film detailed in chapter 3, the
operational dose set-point in the control circuit of an AEC is determined from a
requirement to minimise patient dose and to provide maximum imaging information.
The parameters used to determine what this dose should be with CR are discussed
with respect to the wide exposure latitude characteristic of CR.
The procedure to set the AEC dose set-point with CR at Christchurch Hospital is
explained followed by the AEC calibration over the clinical kVp range. The Kodak
CR system reports a measure of image contrast assessment with processed images in
the form of an ‘exposure index’ (EI). Due to the different parameters used to assess
image contrast a method of correlating the exposure requirements for screen-film
system OD limits and the Kodak CR EI values was developed. The outcome was an EI
variation limit based on existing screen-film exposure parameters. A discussion
follows on whether this limit can be improved further with CR.
The chapter continues with a discussion on the use of a standardised x-ray beam for
the AEC testing process. Without a beam standard, AEC test results can vary between
10
different groups performing the same test, which causes confusion when comparing
results.
In Chapter 7, the results of applying the test procedures developed in chapter 6 are
presented here for an AEC with generic kVp compensation applied to match the kVp
response of the CR image medium and for an AEC which has been optimally
calibrated to match the kVp response of the CR image medium. The dose and EI
responses for each AEC setup scenario are evaluated and presented in a graphical
format. These are followed by the results of image quality and CR noise testing
procedures.
In Chapter 8 the test results for the two AEC test scenarios presented in chapter 7 are
comparatively assessed and analysed. It also includes a clinical assessment of images
produced under each scenario which will determine any impact on current clinical
practice. The introduction of an AEC dose alone measurement as an alternative AEC
testing protocol is discussed. This includes the impact of introducing this protocol into
a testing program relative to conventional methods and consequently a comparison of
test duration for conventional test procedures and the proposed new protocol.
11
Future work beyond this thesis is briefly discussed with a view to the application of
the protocol to other current CR or DR imaging systems and to new digital imaging
systems of the future.
The chapter concludes with the presentation of a proposal to adopt a new protocol for
performing the assessment of AEC functionality associated with any x-ray imaging
medium using the measurement of AEC dose alone.
12
Chapter 2
Development of x-ray production and exposure
control
2.1 Introduction
Within months of Roentgen’s discovery of x-rays in 1895, the phenomenon was
utilised for medical purposes to image broken bones and locate foreign bodies
embedded in patients. Since then the equipment and techniques employed to produce
x-rays have developed to become more efficient, to produce faster imaging times and
to improve image contrast and resolution. The modern x-ray machine is designed to be
safer to use for both patient and operator.
In 1913, William Coolidge invented an x-ray tube that featured a high vacuum and
used thermionic emission from a heated filament as the source of electrons. The small
amount of gas left in this type of tube did not contribute to the production of electrons
(Hendee 1979). Other inventions followed including the focussing cup that
concentrated the electron production from the filament to a small part of the anode
called the focal spot and the line focus design using an angled anode target to create a
small focal spot which produced sharper images.
The efficiency of x-ray production was increased with these inventions but as x-ray
production increased, so too did the heat production at the anode target of the tube.
13
Various materials were used for the anode target in order to find a substance that could
withstand the heat generated by the electron beam striking the target. As x-ray
production increased, the mass of these target materials had to increase to avoid the
destruction of the target. Target materials with higher atomic numbers performed best
and tungsten with an atomic number of 74 proved to be one of the more preferred
materials and remains so today.
In the 1930s the rotating target (anode) x-ray tube was developed. Figure 2.1 is a
schematic diagram of a rotating target (tungsten) x-ray tube showing the anode
connected to the end of a rotor shaft. This development allowed the target position to
change continuously, effectively increasing the area of the target material, enabling
greater heat conduction away from the actual target area (Mould 1995). All of these
inventions and developments have contributed to the design of modern x-ray tubes.
Figure 2.1 The rotating anode x-ray tube with line focus
14
therefore very important to protect both patient and operator from unnecessary x-ray
radiation.
The lack of radiation protection to the x-ray operator and to the un-imaged parts of the
patient is notable in the early photograph shown in Figure 2.2 of a portable military set
up being used by Ernest Harnack. Harnack was one of the first casualties of radiation-
induced cancer after the discovery of x-rays.
Figure 2.2 Ernest Harnack (around 1896) using a portable military set-up, including
a Ruhmkorff coil and early vacuum tube. (Photograph courtesy of the Radiology
Museum Brussels, Belgium)
The early x-ray tubes were not encased in radiation protective material, nor did the
operators stand behind radiation protective barriers or wear radiation protective
apparel, as is the case today. The continued use of these unprotected machines and the
lack of external protective barriers caused many cases of radiation-induced medical
problems to the early operators. As developments in x-ray machine technology
increased, so too did the awareness of the potential for harm from exposure to x-rays.
This led to the realisation that operators and patients required greater protection from
unnecessary x-ray exposure and that this protection should also include some form of
control over the exposures themselves.
15
The modern x-ray tube is encased in radiation protective material to minimize the
emission of unwanted (scattered) x-ray photons. While x-ray photons are emitted from
the anode target material in all directions, the modern tube design and construction
only allows photons to exit the tube easily through a thin window in the glass
envelope. This window is usually made of some material that is highly transmissive to
x-rays such as beryllium. The x-rays emitted through this window can then be shaped
by collimation to become the directional beam of x-ray photons that we are familiar
with today in our x-ray facilities.
16
Figure 2.3 The Holzknecht chromoradiometer scale (Photograph courtesy of the
Radiology Museum Brussels, Belgium)
17
1 R = 2.58 × 10 - 4 C/kg
A second unit of radiation quantity, the gray (Gy) was defined in 1979 by the
International Commission on Radiation Units and Measurements (ICRU report 33).
One gray is equal to the absorbed dose when the energy per unit mass imparted to
matter by ionising radiation is one joule per kilogram (J/kg). X-ray exposure is
measured using standard air filled ionisation chambers. Exposure is almost
proportional to absorbed dose in soft tissue over the diagnostic energy range, typically
10 keV – 150 keV, because the atomic numbers of air and soft tissue are similar
(approximately 7.5). The roentgen and the gray are the standard units of measurement
used for the test procedures involved in this work. The relationship between these two
units in air is (Bushberg et al 2002, pp 54-56):
1 R = 0.00876 Gy.
The settings made will depend on patient size and pathology. Control of these
exposures is achieved by adjustment of the beam energy in kilo voltage peak (kVp),
18
the x-ray tube current in milliamps (mA) and exposure time in seconds (s). The kVp
value gives the maximum photon energy within the spectrum produced by the x-ray
tube. The x-ray energy spectrum from a tungsten x-ray target is shown in Figure 2.4
for a 100 kVp exposure. A useful radiographic image is achieved only if all the
exposure factors are set optimally to match the exposure conditions. The manual
timers used for these exposures were initially controlled mechanically and later
electronically. An example of a type of electronic timer uses pulse-counting
techniques to accurately control the time of short exposures. The pulses are generated
by quartz crystals capable of high frequency modes of operation. A higher degree of
timer accuracy is obtained as the crystal’s operating frequency is increased
(Christensen et al 1990).
The invention of the ionisation chamber during the 1920s was the forerunner to the
development of ionisation timer devices for use with AECs to measure and control x-
ray exposure time. When used for this purpose the ionisation chamber is located in the
x-ray beam and produces a very small electric current (in the order of A). This
current is then used to charge a capacitor in the AEC control circuitry. When this
capacitor, used as an electronic storage device, reaches a predetermined charge, the
exposure-terminating switch is activated to terminate the exposure.
When an x-ray exposure is controlled using an AEC device, the technologist sets the
kVp and tube current as for the manual approach, but the exposure time is controlled
by the AEC exposure termination control circuit. This device terminates the exposure
only after a preset charge on its capacitive trigger circuit has been reached, indicating
that sufficient radiation has reached the imaging medium to produce a quality
radiographic image. Figure 2.5 is a simplified schematic diagram of an AEC control
circuit. The AEC detectors in this diagram respond to the x-rays emitted by the x-ray
tube and produce a small electric current. This current is directed through a buffer
amplifier and causes a charge to accumulate on the capacitor at a rate proportional to
the amount of radiation present in the x-ray beam. The exposure time is determined by
the state of the accumulated charge on this capacitive trigger circuit. When this charge
20
X-ray tube X-ray tube HV control
Exposure termination
control circuit
Voltage
AEC comparator
detectors Buffer amplifier Capacitor
reaches the preset reference voltage, the voltage comparator output activates the
exposure termination control which in turn switches off the high voltage (HV) to the
x-ray tube, terminating the exposure. The density control is able to alter the state of the
reference voltage up or down to accommodate for extremes of patient size or at the
discretion of the radiographer to achieve optimal image quality, whilst maintaining a
consistent response over the clinical kVp range.
A Philips x-ray machine is shown in Figure 2.6. The erect Bucky associated with this
machine (photo left centre) shows the position of the three AEC chambers. The x-ray
tube and collimator assembly is shown on the right of the photo.
AEC detectors are of two types: entrance detectors are located in front of the film
cassette tray; exit detectors are located behind the film cassette tray.
21
Each AEC chamber is independently selectable by the MRT. A single chamber can be
selected or combinations to match the exposure requirements of the anatomical area to
be imaged. Some manufacturers apply a differential to the operational sensitivity of
the individual AEC chambers. The differentials are designed to match the attenuation
coefficients of the main anatomical areas of the lungs, heart and abdomen.
Figure 2.6 X-ray machine erect Bucky (centre left of photo) showing the physical
arrangement of the three ionisation chambers used as x-ray detectors for the AEC.
22
area of a patient can be erroneous. The effects of this type of exposure inaccuracy may
be a total or partial exclusion of the selected AEC chamber from the primary x-ray
beam especially where close collimation is required. In this type of situation the AEC
will prolong the exposure until it records that sufficient dose has been collected or it
reaches its exposure safety limit. The result will be an over exposed image and a
higher exposure dose to the patient. The exposure may have to be repeated after
correcting the exposure setup, resulting in further dose to the patient.
Paediatric patients
The standard physical AEC chamber layout presents problems for paediatric
radiography. The physical size of the paediatric patient relative to the size and position
of the standard AEC chambers can lead to positional errors or exposure timing
problems similar to those discussed above.
KVp selection
The correct selection of kVp is a critical factor for image quality and minimising
patient dose. For example, if an inappropriately low kVp technique is set for a large
patient, the AEC will attempt to maintain the exposure until its programmed trigger
dose has been reached. However it may not achieve this point before the maximum
safe exposure time has been reached at which point the AEC terminates the exposure.
The result again will be poor image quality and unnecessary dose to the patient. The
exposure would need to be repeated at a more appropriate kVp setting, resulting in
further dose to the patient.
23
The use of Computed Radiology (CR) as the imaging medium with AEC
When CR technology is used as the imaging medium, the radiographic techniques
must be closely monitored to ensure that the wide dynamic range of CR and post
processing methods do not hide inefficient or inappropriate radiographic technique.
There are many benefits to be gained from the use of AEC devices. However, as
indicated by the problems highlighted above, the use of an AEC device does not
relieve the operator of the responsibility of using good quality radiographic
techniques.
24
Chapter 3
AEC in a screen-film environment
3.1 Introduction
AEC functionality in any imaging environment is assessed by testing the following
radiographic parameters: reproducibility, detector chamber response, kVp response,
patient thickness response determined using a patient equivalent phantom (PEP),
density control function and the maximum and minimum x-ray exposure times. The
focus of this thesis is on the AEC response to varying kVp and PEP thickness for the
particular image medium for which the AEC is calibrated. For the screen-film imaging
medium, this is achieved by recording and analysing the optical density (OD) of the
radiographic film images resulting from a range of test exposure conditions.
The AEC is expected to control the exposure time of x-ray procedures in order to
produce consistent image quality over the range of clinical kVp and patient thickness.
By recording the OD of homogenous images produced under different conditions of
kVp and patient thickness, a profile of the uniformity of the AEC function is obtained.
Any variations in this performance over the test exposure conditions must fall within
specified tolerance limits to conform to quality assurance requirements. Should the
AEC performance vary beyond these requirements then the problem causing the
erroneous AEC behaviour must be corrected by the service engineers.
25
permanent part of the internal structure of a typical x-ray film cassette into which the
radiographic film is placed. Because of this, I will refer to radiographic film as screen-
film in future references to this image medium, unless the film itself is the subject of
the reference.
Linear
region
The combination of a luminescent screen with radiographic film reduces patient dose,
x-ray generator loading, exposure duration and hence patient motion artefacts
(Heggie et al 2001). This is due to greater x-ray detection efficiency of the
luminescent screen compared to the radiographic film by itself. Screen-film
combinations are considered to be contrast limited in their response because their
range of lineal contrast changes with exposure is relatively small
(Christodoulou et al 2000).
26
The sensitivity of screen-film combinations varies with kVp as shown in the
experimental results plotted in Figure 3.2. The graph shows the response of a Kodak
T - Mat-G-RA screen-film system exposed to a fixed dose of 0.8 mR over the kVp
range 50 kVp to 125 kVp. For each plotted kVp value on this graph, exposures were
performed by adjusting the x-ray source to image distance (SID), tube current and
exposure time to achieve 0.8 mR at the plane of the image medium placed in a Bucky
tray. This graph is presented not to show absolute OD values at each kVp but rather to
obtain an understanding of the shape of the response for this medium, especially at the
lower end of this range (50 to 70 kVp). The results show that the response is poor at
the lower end of this kVp range and then steadily increases to exhibit a relatively even
OD response over the 70 to 125 kVp range. It is this type of image medium response
characteristic that the AEC must compensate for to maintain consistent image quality
over the clinical kVp range.
2.4
2.3
2.2
2.1
Optical Density (OD)
2.0
1.9
1.8
1.7
1.6
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
50 60 70 80 90 100 110 120 130
kVp
Figure 3.2 Typical kVp response of a Kodak T-Mat-G-RA type screen-film system to a
fixed dose of 0.8 mR
27
Radiographic film, like any emulsion film based on some substrate, must be placed in
chemical reagent baths which act on the chemicals of the film emulsion to develop the
latent image trapped within its chemical structure. Gelatine and silver halide are the
two most important components of the film chemical structure. The gelatine provides
a suspension medium for the silver halide and forms an emulsion layer on the
substrate of the film. When the silver halide is exposed to light from the fluorescing
screen its chemical structure changes to form latent image sites. These sites are clumps
of silver atoms that collectively represent the image. The film development stage locks
this image onto the film substrate and removes any other emulsion material that does
not contribute to the image.
Radiographic film and processing chemicals must be carefully managed since images
of consistent quality cannot be assured with expired film and chemicals. Bulk film
should be stored in a dry, temperature-controlled area isolated from direct x-ray
sources in light-tight containers to avoid fogging.
28
image. The second requirement is to minimise patient exposure dose and hence to
prevent the possible onset of medical complications. Higher exposure levels are
potentially harmful to patients and violate the legislated requirement to keep doses as
low as reasonably achievable (the ALARA principle).
The exposure used to determine the AEC dose set-point is specified by kVp, PEP
thickness, SID and AEC chamber selection. The specified values for this exposure at
Christchurch Hospital are 81 kVp, 150 mm Perspex PEP, an SID of 1000 mm and the
selection of the central AEC chamber. The dose delivered to the plane of the screen
film, in the Bucky tray, under this test exposure is measured and compared to the
required dose set-up value. The AEC dose control is adjusted and the procedure
repeated until the correct value is attained. The dose measured at the plane of the
image medium under these circumstances is approximately 3.0 µGy. Having achieved
the dose required a second exposure is performed and the OD value of the
homogenous film image is measured from the image centre and recorded. This OD
value becomes the target exposure index for the process of setting the AEC dose
control to compensate for the kVp response of the image medium.
To set up the AEC the AEC dose control is adjusted for each kVp setting to achieve
the target OD on each exposed film image. The AEC control voltage references,
described in section 2.6, which control these doses, are then locked into the AEC
control circuit memory. These doses are all based on the value at the AEC dose set-
point.
29
Provision can be made in the design of an AEC for manual adjustment of the AEC
dose-point when required to compensate for extremes of patient sizes. This is achieved
on the Philips Optimus x-ray machine used for this thesis by varying the density
control setting. When this is done the AEC control algorithm maintains its normal
control function over the kVp response of the image medium but the dose set-point it
is normally based on is increased or decreased as required. The density control
function is typically enabled by a manually operated switch function.
The standard range of physics tests for AECs listed in section 3.1 is performed by the
Medical Physicist. All tests are performed at the SID for which the Bucky and its
associated grid have been designed. The SID is typically 100 cm for table Buckys and
140 cm for erect Buckys. The AEC test procedure applied by the Medical Physics
group at Christchurch Hospital utilises 75 mm thick Polymethyl Methacrylate
(Perspex) slabs for its patient equivalent phantom (PEP) simulation material. For the
full range of tests, up to four of these Perspex slabs are used to simulate a reasonable
range of patient size. The PEPs are placed in front of the Bucky assembly and
positioned to cover all three AEC chambers. The x-ray beam is collimated to form a
300 mm × 300 mm field size at the Bucky. The film cassettes, loaded with film, are
placed in the cassette tray of the Bucky assembly. A film image is required for each
part of each test function, resulting in the production of twenty eight or more
individual films.
The ODs from the test images are measured and recorded for all test exposure
conditions. These recorded values should lie within the range of acceptable OD values
1.2 - 1.6 (NRL C5 1994).
30
water by composition. Perspex is also radiographically similar to water as shown in
Table 3.1 and therefore to the human body.
In table 3.1 Z is the average atomic number of the material, A is the average mass
number and I is the mean excitation energy per ion pair. The perspex PEP is simpler
and safer to use around mechanical and electrical equipment compared to the more
traditional water phantoms. However, there are weight and size considerations with its
handling and transportation. Each PEP weighs approximately 8.5 kg and has
dimensions of 370 mm × 320 mm × 75 mm. The longer dimension has been machined
to form a 50 mm carrying handle. The Perspex slabs are physically stable,
radiographically uniform and offer good positional accuracy relative to the AEC
chambers associated with x-ray machines. These attributes of perspex were important
in the decision to use the perspex slabs as the PEP material for the experimental work
required for this thesis.
NOTE:
This table is included on page 31 of the print copy of
the thesis held in the University of Adelaide Library.
Table 3.1 Radiographic comparison: PEP to human tissue. (Data reproduced from
National Institute of Standards and Technology (NIST) x-ray mass attenuation tables)
The lead window test method is described in a paper by Hunt and Plain (1993). In the
version of this method used at Christchurch Hospital, a single 35 mm 35 mm film
cassette has been modified to mount a 2 mm thick lead sheet over its whole surface
with a 200 mm diameter hole cut out of the centre. This hole is filled with a 2 mm
thick lead disc which has a diameter slightly less than the hole in the lead plate so it is
free to rotate within the lead plate. The lead disc has a single cut-out sector as shown
in Figure 3.3. This diagram shows the sector window ready for exposure 1. The lead
disc is rotated one step for each separate test exposure, to open a window to a new
sector of the film contained within the cassette. The 2 mm thickness of lead protects
31
the other parts of the film from radiation. The size of the lead disc cut-out is designed
to provide sufficient sectors to enable as many images as required to complete a
section of the AEC test sequence. The only physical constraint on the number of
sectors in the disks circumference is the need to ensure a reasonable sample area to
measure the OD of the film image and also to minimise any radiation or scatter from
adjacent exposures. The rotating lead disc used at Christchurch Hospital provides for
18 separate exposures. In order to perform the full range of AEC tests, 28 exposures
are performed which can be accommodated on two films. If individual films are used
for each test then 28 films would be required as stated in section 3.4.3.
Figure 3.3 Physical layout of the AEC test cassette showing film window at position 1
The use of the lead window method is limited only by the physical size of the test
cassette. The test cassette with its lead disc assembly must be physically thin enough
to fit inside the limited space available within the Bucky assembly used for normal
screen-film cassettes. The depth of this space is approximately 20 mm for a standard
Philips Optimus x-ray machine Bucky but this may vary for machines of different
manufacture. The depth of the test cassette used at the hospital is 18 mm. This
compares to a depth of 15 mm for a standard x-ray screen-film cassette
32
Average OD values lower than the standard OD range result in images which are
lighter and lack sufficient contrast for good image diagnosis. Higher OD values
indicate a darker image where fine or lower contrast image detail may be lost and also
indicates higher clinical x-ray doses. The maximum OD variation of the individual
results from the average within each separate assessment area is also calculated. For
repeatability tests using the same exposure conditions this variation limit is ≤ 10%
while the limit for variable exposure conditions is ≤ 20%. If the result exceeds the test
limit, the AEC is recorded as having failed the particular function.
The AEC dose set-point is checked with a measurement of the dose from the centre
chamber test as described in section 3.4.1.
The complete set of AEC results and their assessment is documented and becomes part
of a report produced by the Medical Physicist which details the complete operational
status of the x-ray machine. It is the responsibility of the equipment maintenance
engineers to correct any non-conformance of the machine’s performance, including
AEC functionality.
33
Chapter 4
The Digital Environment with Computed
Radiography
4.1 Introduction to digital imaging
The development of digital imaging systems which includes computed radiography
(CR) has brought about a steady decline in the use of radiographic film systems as the
primary mode of x-ray imaging. Consequently, a review of all physics test procedures
including AEC functionality test procedures is required to assure their compatibility
with the imaging characteristics of digital imaging systems (Huda 1997).
There are many digital imaging media available on the market including
complementary metal oxide silicon (CMOS) devices, charge coupled devices (CCDs)
and thin film transistor (TFT) detector arrays. The calibration procedures described in
this thesis in relation to CR can be adapted to any of these systems. At the time of
writing this thesis, NRL had no published documentation or guidelines on the
requirements for AEC functionality testing in any digital imaging environment.
4.2 CR technology
CR technology has developed since its conception to become one of the most common
digital imaging modalities in radiology departments today. It became commercially
available in 1983 with the introduction of the Fuji FCR101 system (Bradford et
al 1999). CR technology is adaptable to digital radiography that enables the
application of advanced image enhancement techniques (Crawford and Brixner 1991).
“The broad acceptance of CR has been due to its large dynamic range, digital nature,
easy portability and uniqueness rather than its intrinsic image quality”
(Rowlands 2002).
34
Kodak CR PSPs are formed onto a rigid substrate called an image plate (IP) which is
contained in a protective cassette similar to a conventional screen-film cassette. They
are compatible with conventional x-ray machines, enabling hospital x-ray departments
to convert to a CR system without the need to invest in dedicated digital x-ray
machines. Most installations closely emulate radiographic film radiography with the
use of cassette based image systems. Although CR image plates are initially
expensive, they can be reused for thousands of exposures.
The CR cassette is a light tight enclosure for the IP. The purpose of the light-tightness
of the cassette is entirely different to that for screen-film systems. An x-ray film is
sensitive to normal room lighting and will be completely fogged (darkened) by
exposure to this light. Conversely, exposure of a CR PSP to the same light will cause
any latent image on the PSP, produced by exposure to x-rays, to be erased.
35
The sites containing trapped charges have strong optical absorption bands especially in
the 633 nm (red) band, which is the wavelength of He-Ne gas lasers commonly used
to scan the image in a CR reader as described in section 4.8. The particular CR PSP
referred to in this work is the barium fluoro halide, BaFBr 0.85 I 0.15 : Eu2+ using bromine
as the halide. This phosphor is now the most commonly used by many CR PSP
manufacturers. In comparison to the simpler form BaFBr I: Eu2+ used in earlier
versions, this new phosphor has a marginal increase in x-ray absorption and has a
better spectral match to the wavelength of maximum phosphor stimulation with solid-
state diode lasers. Solid state lasers have a spectral wavelength of 680 nm.
The mechanism of the photostimulable luminescence has been reported as being not
completely understood with at least two models being proposed, one being
monomolecular and the other bimolecular (Bogucki et al 1995). A band model using
the bimolecular approach offers a possible explanation for the process.
Figure 4.1 Energy level diagram of PSP (Reproduced courtesy of Kenji Takahashi
with permission from Elsevier Limited)
This model shown in Figure 4.1 should be viewed with the following explanation
quoted from Takahashi (2002): “When exposed to ionizing radiation, a number of
electron–hole pairs are generated, proportional to the radiation energy absorbed by the
phosphor particles. They immediately recombine to either emit light from the excited
states of the Eu2+ ions or create quasi-stable F-centres. An F-centre consists of an
electron trapped by the F+-centre. An F+-centre is a halogen ion vacancy created by
controlling the formation process of the phosphor. These F-centres and corresponding
36
hole traps store the energy of the absorbed ionizing radiation. When exposed to red
laser light, used for reading the IP, energy is absorbed by the F-centres and the trapped
electrons are raised to excited states where they are thermally excited to the
conduction band. They recombine with holes trapped by Eu2+ ions, which forces the
excited Eu2+ ions to emit light corresponding to the 5d–4f transitions.”
The absorption K-edge of a material is the applied energy required to eject an electron
from the first or ‘K’ shell of an atom. When this energy is applied, electrons in this
shell are released; de-excitation of the excited atom results in the emission of
characteristic radiation. In section 2.5, Figure 2.4 shows an example of how this
characteristic radiation appears on the radiographic spectrum for tungsten which has
two main characteristic radiation peaks. The most energetic of these occurs at about
69.0 keV from interactions involving electrons in the higher energy ‘K’ shell of
tungsten’s atomic structure.
Figure 4.2 shows the x-ray absorption response for gadolinium oxysulphide
(Gd 2 O 2 S), a luminescent screen used for screen-film systems and a PSP
(BaFBr 0.85 I 015 : Eu2+). The relative intensity of the primary x-ray beam is also
displayed in this graph to provide a perspective on the relevance of the absorption
response at each energy measured. The data for this graph were collated from two
sources. Firstly, the x-ray spectrum data were obtained from NRL’s ‘Specgen’
computer program developed by Glen Stirling and based on the Birch and Marshall
model. Secondly the absorption spectrum for the luminescent screen and PSP
37
phosphor was obtained from the International Atomic Energy Agency (IAEA)
computer program ‘xmudat’ version 1.0.1 (Nowotny 1998).
50 0.8
Gd2 O2 S
45 0.7
Ba f Br0.85 I0.015
40 Primary spectrum
0.6
Absorption coefficient
25 0.4
20
0.3
15
0.2
10
5 0.1
0 0.0
10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100
keV
The ‘K’ edge for the luminescent screen appears around 50 keV, characteristic of the
main absorbing medium gadolinium. The main absorbing material in the PSP is
barium which has a ‘K’ edge around 37.4 keV. The absorption characteristic between
35 keV and 50 keV is greater for the PSP, which supports one of the theories for the
increased sensitivity of CR to scatter radiation. However analysis of the absorption
characteristic over the whole range of the primary x-ray beam spectrum (10 –
100 keV) by calculating the area under the absorption characteristic for each material
produces similar results. This does not support the general theory that a PSP has a
poorer overall absorption characteristic although comparisons at individual keV points
not included in the 35 keV to 50 keV range do exhibit slightly poorer results.
In conclusion, we can see that the response characteristics for these two phosphor
materials are almost identical except for the 35 keV to 50 keV range where the PSP
exhibits a higher sensitivity. Any scatter radiation within this keV range will be
preferentially absorbed by the PSP compared to the luminescent screen. This scatter
increases the noise (signal content not directly related to the imaged object) content of
38
a CR image and reduces the SNR. The noise sources contributing to noise in CR
images are discussed in section 4.8.
Scattered radiation emitted from a patient or PEP placed in the x-ray beam will be of
lower energy than the primary x-ray beam. A phosphor material that discriminates
between this lower energy radiation and that of the primary beam can, partially at
least, filter out this unwanted and confounding contributor to the resulting x-ray image
(Bogucki et al 1995). The gadolinium based luminescent screen has this ability for
scatter energies within the range 35 keV to 50 keV.
The exposure index indicator for the image processing system should match the dose
value applied to the image medium according to its designed algorithm. A regular
39
program of CR equipment maintenance is required to ensure a consistent exposure
index indication for the same applied x-ray dose to the imaging medium. The medical
physicist confirms these values for a range of applied doses. At Christchurch Hospital
these tests are carried out in air and are not as susceptible to the level of scattered
radiation present when measurements are performed within the confines of an x-ray
Bucky tray as used in the experimentation for this thesis.
2200
2100
2000
PEP material
1900
EI 0.5mmCu+1mmAL
1800
150mm Perspex
1700
1600
1500
40 50 60 70 80 90 100 110 120 130
kVp
Figure 4.3 shows the results of tests performed experimentally on a Philips x-ray
machine using a fixed dose of 1.0 mR. A Kodak CR reader was used to process the
test images. The error bars indicate a possible range of ± 10 exposure index values.
This error has been determined from experimental data and is based on the readout
resolution of the Kodak reader as explained in section 4.12. The experiment compared
the results for two PEP materials, 0.1 mm Cu with 1 mm Al and 150 mm Perspex. The
response obtained is typical of those found in many reports and papers for Kodak CR
phosphors and shows significant variation in EI with kVp, especially for the Cu and Al
PEP (Lu et al, 2002). The kVp response of an image medium is required to provide the
data which enables the AEC of an x-ray machine to be calibrated optimally to provide
consistent exposures over the clinical kVp range.
40
4.7 The CR response as a function of dose
Current CR readers use photomultiplier tubes (PMTs) to detect the stimulated light
output from CR PSPs and these achieve reasonable linearity over four orders of
exposure magnitude. An experiment was conducted to derive the exposure range
response for a Kodak CR system at Christchurch Hospital. In order to obtain the
required data for this, two x-ray exposures were required at each dose setting. The first
was used to measure the exposure dose that would be received by a cassette in the
Bucky tray. This was achieved by placing a 15 cc ion chamber in the centre of the
Bucky tray in the same plane as a CR cassette. The second exposure was required to
expose a CR IP and obtain the EI data from the processed image. The x-ray machine
was set to perform these exposures at 81 kVp which enabled a comparison with
published responses. The dose values required to show the complete range of the
response of a CR PSP were achieved using a range of SID, tube current and exposure
time combinations.
4000
3800
3600
3400
3200
3000
2800
Exposure Index
2600
2400
2200
2000
1800
1600
1400
1200
1000
800
600
400
200
0
0.001 0.01 0.1 1 10 100
Dose (m R) (Log scale)
The range of possible EI values for a Kodak CR system is 0 to 4096. The response
obtained from the acquired data is shown in Figure 4.4. The results compare well with
typical exposure range responses of CR PSPs given in many reports and papers such
as the AAPM Task Group 10 report and the paper by Wilkinson and Heggie (1996).
The dashed blue line on the graph represents the Kodak dose / EI relationship given in
41
section 4.5 for GP-25 PSPs. Within experimental error the test results for doses above
0.25 mR conform well to the Kodak formula. The gradual deviation away from the
ideal Kodak formula line of the experimental results below 0.25 mR is a result of the
decreasing signal to noise relationship of the PSP. This effect is also exacerbated by
the offset factors for variation to laser power output which is added to image data.
This leads to the situation where an erased PSP is unlikely to produce a zero EI value
and can be up to 100 EI values (Bogucki et al 1995).
Each CR manufacturer employs their own unique method of processing their PSPs.
The IP contained within the Kodak CR cassette has a unique identifying bar code. The
MRT moves this bar code in front of the bar code reader attached to a remote
operators panel (ROP) to register the cassette for a particular examination. The ROP is
a computer workstation and is generally located near the x-ray room it serves, with a
communications link to the CR reader. The bar-code data is automatically inserted into
the appropriate area of the patient demographic data form on the ROP screen. The
ROP serves as an image viewing station where the MRT can verify image quality and
perform digital image enhancements. It is also the interface by which the patient’s
demographic data and hospital patient recording data is input and linked to the
particular CR cassette used for a patient’s examination.
42
Once the patient’s demographic data has been input to the ROP panel the MRT is able
to select pre-processing options which provide exam specific optimisation. It achieves
this optimisation by adjusting the processing sensitivity at certain points on the image
pixel histogram. The cassette is then placed into the CR reader. The machine opens the
cassette in a light sealed cavity and the IP is extracted. The PSP is scanned by a He-Ne
(633 nm) or diode laser beam which begins the process of photostimulated
luminescence (PSL) (Rowlands 2002). The sampling aperture is the instantaneous area
of the phosphor screen stimulated by the laser beam. The ratio of the sampling rate
and the scan rate across the PSP determines the size of the sampled area known as a
pixel. Also the speed of the PSP movement in the sub-scan (vertical) direction is
coordinated with the horizontal movement across the PSP such that the pixel length is
equal to its width. The pixels are square. The pixel dimension in the scan direction is
typically between 100 µm and 200 µm dependant on PSP size. As the laser beam
passes over an area of the phosphor it releases charges trapped by the reaction to
incident x-rays and this stimulates the emission of light from the phosphor. The light
emitted is guided to a photomultiplier tube (PMT) within the reader. The PMT detects
this light and converts it to an electrical signal proportional to the amount of light
produced. This analogue signal is then digitised to form the raw data that represents
the imaged object. This is achieved by quantising the average signal amplitude at the
individual pixel location. The quantising of the analogue signal pixels into discrete
integer values is determined by the amplitude of the signal present in each sample and
the total number of possible discrete values. Kodak has 4096 possible discrete values.
A pixel clock coordinates the time at which a particular signal is digitally encoded to a
physical location on the scan line (AAPM Task Group 10 2002). This timing is
determined by the pixel dimension in the horizontal laser scan direction and the total
number of pixels in that direction relative to the CR cassette size. The discrete signal
amplitude at each pixel location and the timing of their measurement ultimately define
the system resolution (Bogucki et al 1995).
NOTE:
This table is included on page 43 of the print copy of
the thesis held in the University of Adelaide Library.
Kodak PSPs are manufactured to several different physical sizes. Each has a unique
image size measured by the number of pixels in each orientation as shown in
Table 4.1.
43
Kodak initially quantises the PMT signal range to 216 (65536) bits or gray levels in
order to implement a digital logarithmic transformation to the final 212 (4096) bit
image data. Pixel sampling data is derived from the 16 bit raw image data output from
an analogue to digital converter. The raw 16 bit linear image data is mapped to 12 bit
logarithmic data to enable correct weighting of the digitisation across the four decades
of radiographic exposure to which the PSP is able to respond (Bogucki et al 1995).
This dynamic range compression preserves digitisation accuracy over a limited
number of discrete gray levels. Logarithmic conversion also provides a linear
relationship of incident exposure to output signal amplitude (Seibert 2004). The
transformation to 212 bits provides weighting of the digital signal for better
visualisation on a monitor (Rowlands 2002).
The 212 bit digital stream representing the exposed image is the data that is stored in
the PACS image storage system and which is later retrieved and viewed on
workstations at the request of radiologists or clinicians for image analysis and
diagnostic reporting.
A 28 bit digital stream is derived from the 212 bit data and used to generate the image
produced on the ROP that the MRT uses for image verification following an
examination. The image data from the CR reader used for image presentation and
storage is formatted to the Digital Imaging and Communications in Medicine
(DICOM) standard described in section 4.8.1.
Once an IP has been scanned by the laser it is exposed to white light to erase any
remaining image latency. Modern CR readers can process a PSP, including erasure
time, in less than one minute. The IP is returned to the cassette and then expelled from
the machine.
DICOM 3.0 is the current version of this standard and Part 14 of this describes the
grayscale Standard Display Function (GSDF) used in the display of the digitised CR
images. It specifies a function that relates pixel values to displayed luminance levels
on the viewing monitors. Other parts of this DICOM Standard specify how digital
image data can be moved from system to system and how these images should be
displayed.
The ROP image presentation available to the MRT is designed for local assessment of
image quality and positioning only. It is not designed to perform to a diagnostic image
standard.
Once a PSP is read by the CR processor the remaining latent image is erased. A bank
of high intensity fluorescent tubes within the CR processor performs this task and
ensures that the process of erasure is thorough and completed within a short time
interval. The total CR IP processing time for a CR 900 reader is approximately 1
minute. The IP erasure process occupies approximately 20 s of this total time. The
erasure process is required to remove all image information. In practice natural
45
radioactivity and cosmic radiation can accumulate on IP’s such that there will always
be some residual activation within the PSP.
Quantum noise (quantum mottle) is the dominant noise factor on under exposed
images. Quantum mottle occurs in both screen-film and CR systems, and is caused by
the non-uniformity of x-ray photons absorbed per unit area of the screen. It is based on
Poisson statistics where the standard deviation is N where N is the average number of
x-ray quanta per unit area. Quantum noise varies as the square root of the number of x-
rays used to produce the image. As the x-ray exposure increases, the fractional
contribution of this noise decreases (Wilkinson and Heggie 1997). Increasing the
phosphor thickness reduces quantum mottle by increasing x-ray absorption but spatial
resolution will be adversely affected.
Conversion noise is the variation in signal detected per absorbed x-ray and is
inversely proportional to x-ray exposure. The conversion efficiency value for a
46
particular phosphor is determined by its composition and design. A high conversion
efficiency increases screen sensitivity and reduces patient exposure. Unfortunately, an
increase in conversion efficiency decreases the quantity of x-rays that must be
absorbed in the phosphor, and this increases quantum noise. The only way to increase
phosphor sensitivity without increasing quantum noise is to increase the absorption
efficiency. An increase in absorption efficiency does not change the amount of
radiation that must be absorbed to produce an image. The design of the phosphor is
therefore based on a balance between screen sensitivity and quantum noise.
Structure noise is the spatial gain variation in the screen (granularity) and is
independent of the x-ray exposure. This noise is proportional to the amplitude of the
derived image signal and is more apparent at very high x-ray exposures. It is reduced
at higher spatial frequencies due to pixel signal averaging caused by the laser light
spread to adjacent pixels Structure noise is relatively insignificant in most
radiographic applications.
Laser noise is due to the fluctuation in the output intensity of the stimulating laser and
is independent of x-ray exposure. This type of noise produces low frequency
components within the noise spectrum. The CR processor minimises this noise by
detecting the intensity fluctuations of the laser output and adjusting the image signal
accordingly.
47
by the digitisation process and can be minimised using an adequate bit depth system
i.e. range of quantisation steps.
All of the above noise components have a spatial frequency dependence such that
lower spatial frequencies contribute more to the overall noise power spectrum
especially at low exposure levels (Yaffe and Rowlands 1997).
CR is considered to be a noise limited system such that the SNR is largely determined
by the x-ray dose applied to the PSP, especially at extremely low doses. The SNR at
very low doses is poor, due to the base noise contributors such as Structure Noise and
Laser Noise, which are independent of the x-ray exposure. When the effects of noise
contributors such as quantum mottle become more apparent then the visualisation of a
clinical image can be compromised. Higher exposure levels produce clear crisp
images with good contrast that appear to have a better signal to noise ratio. However,
saturated areas due to excessive exposure levels may mask anatomical features.
A local experiment was conducted to derive the image decay characteristic of Kodak
CR PSPs using a Kodak CR 900 processor. The results have been plotted in
Figure 4.5. This graph shows that the EI decreases exponentially with increasing
processing time delay. Following an eight hour processing delay the exposure index
value has decreased by almost 10% compared to the value obtained when the image
was processed immediately. This EI value recorded at this time delay was
approximately 2730 which is approximately 300 EI values less than the value obtained
at the 60 s delay interval. Using the Kodak dose/EI formula an EI variation of 300
points is equivalent to a delivered dose value equal to half of the original exposure. It
can also be shown that a clinical image can be read out from an exposed IP even after
several days of processing delay.
Delays in processing a PSP following its exposure to x-rays have an impact on the
reported exposure index and on image quality. This factor is critical in obtaining
accurate results for all standard Medical Physics testing procedures and for the
assessment of AEC performance for this thesis. Kodak recommends a 15 minute
48
processing delay for physics testing procedures. After a 15 minute processing delay a
point in the image decay characteristic is reached where the rate of EI change with
time begins to decrease. The rate of EI change in the first five minutes of the image
decay curve is shown as approximately 8 EI/min. The decay rate between 15 minutes
and 20 minutes is approximately 2 EI/min.
3080
3060
3040
3020
3000
2980
2960
2940
Exposure Index
2920
2900
2880 5% change @1hr
2860
2840
2820
2800
2780
2760
2740 9.9% change@8hr
2720
2700
2680
0 30 60 90 120 150 180 210 240 270 300 330 360 390 420 450 480 510
Delay (min)
Due to the large number of test exposures and the machine time required to complete
this work, it was impracticable to conform to the recommended 15 minute processing
delay. EI Measurement accuracy depends on consistent processing timing. The
processing time is less critical at the 15 minute point on the processing delay curve but
its accuracy is still dependant on test consistency. The practical alternative was to
choose a much shorter delay and this was set for this thesis to be 60 ± 10s. This
follows a similar process to that used by Doyle and Martin (2006) for their AEC
calibration method (Doyle and Martin 2006). A Kodak CR 900 reader requires
approximately 60 ± 10 s to processes a CR IP, resulting in a 120 ± 20 s time interval
between exposure and EI readout. Typically, clinical images are processed
immediately following x-ray exposure to obtain the required patient anatomical
information as soon as possible and to determine if further imaging is required prior to
49
releasing the patient. Clinical images are therefore relatively unaffected by the
processing delay effects presented above.
The first point on the image decay curve in Figure 4.5 shows the EI result of a CR
image placed in the processor 60 s after exposure to x-rays. After the 60 s processing
delay and the 60 s process time itself have elapsed the EI value has decreased by
approximately 20 index points. It follows that the possible range of EI value
inaccuracy for images processed with a 60 ± 10 s processing delay is ± 20 index
points. The Kodak CR reader has a reporting resolution of 10 index points which
implies that the actual EI variation lies within the range 15.1 to 24.9. This error
represents a dose variation of 0.010 mR which is less than 2 % of the average dose
measured for each scenario of the AEC kVp compensation tests.
50
Early in the development of digital radiography, it was thought that its image quality
should match the demanding resolution performance of screen-film based imaging. An
alternate proposition is that a high value of limiting resolution is not as important as
the ability to provide high quality image contrast over a wide range of x-ray exposures
for all spatial frequencies up to the maximum possible limiting resolution (Yaffe and
Rowlands 1997).
The spatial resolution of radiographic film is good because the film emulsion is
usually no thicker than 10 µm (Heggie et al 2001). However its sensitivity to x-rays is
poor due to the granular nature of the silver halide suspension in the emulsion. The
low sensitivity of film was overcome when luminescent screens were introduced. In
this situation the film is sandwiched between luminescent screens inside the cassette.
The screens are more sensitive to x-rays than the film by itself and fluoresce to
produce visible light to which the film is very sensitive. However the gain in
sensitivity achieved by this is offset by a reduction in high spatial resolution due to
light scattering processes within the phosphor of the screens (Seggern 1999). The
negative effects of light scattering depend on the thickness of the screen’s phosphor
material. The problem is more significant on thicker screens but the inclusion of dye in
the phosphor can help overcome this. The use of screens using rare-earth phosphors
such as BaSrSO 4 with their increased absorption efficiency and conversion efficiency
has become very important in screen-film radiology (Heggie et al 2001).
The response of a CR PSP to x-rays has limitations in spatial resolution due to the
location of electron traps created within the PSP throughout the depth of the phosphor
material. Scattering of the stimulating laser light increases image blur as it penetrates
the phosphor and releases these traps over a wider area of the image than the size of
the incident laser beam. Figure 4.6 illustrates the effect of the laser on a PSP and
shows the scattering processes that occur. CR phosphor thicknesses vary between
manufacturers. Kodak uses a phosphor thickness of 300 µm for their standard GP
phosphors. The phosphor thickness determines the number of x-ray photons absorbed.
Thicker phosphors increases light scattering and therefore reduces image resolution.
While thinner phosphors produce sharper images, the amount of light released per
incident x-ray photon is significantly reduced (Bogucki et al 1995). The effect of a
thinner phosphor is similar to the statistical variation in the number of photons
detected from the laser stimulation of the phosphor and results in image variation
noise called photon or luminescent noise. Luminescent noise is described in
section 4.10.
51
Film imaging has limited exposure latitude due to its characteristic curve response
resulting in limited contrast resolution. A CR imaging system has a much larger
exposure latitude.
NOTE:
This figure is included on page 52 of the print copy of
the thesis held in the University of Adelaide Library.
A section relating to the cost analysis of a shift to CR imaging systems and image
storage was originally included here but was not considered core to the project and
therefore excluded. However should the reader wish to explore this subject area
further I have included a reference to an introductory paper by Orand (2004) relating
some of the issues of this cost analysis.
52
Chapter 5
AEC in a CR Environment
5.1 Introduction
Initially this thesis set out to explore the AEC testing protocol requirements when the
imaging process is CR compared to the requirements for screen – film imaging
systems. This exploratory stage determined that the only difference between the two
protocol requirements for AEC was the methods used to extract image exposure
information. The testing procedure requirements remain the same. However the most
significant finding during this stage was not any differences in protocol requirements,
but was an awareness that the AEC response over the clinical kVp range was not as
consistent as expected. The function of an AEC is to maintain image consistency over
the clinical kVp range. At this point the focus for the thesis was diverted towards the
exploration of AEC performance and to realise any benefits from attaining a higher
level of image consistency.
The AEC kVp compensation algorithm provided by Kodak for their CR systems and
installed into the x-ray AEC control device by the x-ray system manufacturers,
provides a generic relationship to the imaging medium. Medical Physics groups are
increasingly aware of the need for greater accuracy in the provision of AEC kVp
compensation, as evidenced by recent articles and presentations in the medical physics
field (Grieg et al 2005).
A series of standard AEC tests was carried out on several x-ray machines of the same
manufacturer at Christchurch Hospital, in order to determine their kVp response
characteristic with CR image processing systems.
53
exposure dose applied will be as low as possible while still providing satisfactory
image quality. The Medical Physics Monograph 30 (Goldman and Yester 2004)
suggests that the limits and limit parameters applicable to screen-film systems may not
be appropriate for CR. It proposes that EI limits could be based on just noticeable
differences in noise levels observed in images with low EI values and radiation dose
limits for images with high EI values. However this type of assessment is subjective
and is therefore inherently less accurate than a quantitative approach and its setting
cannot be absolute.
54
MAT G RA film used at the hospital, prior to the inception of CR, has an intrinsic
speed of between 250 and 300.
1.8 0
1.6 20
40
1.4
60
Speed class
1.2 80
Relative EI
1 100
Relative EI
120
0.8 Speed class
140
0.6 160
180
0.4
200
0.2 220
0 240
40 50 60 70 80 90 100 120 140
kVp
Once the dose set-point at 81 kVp for the AEC has been set up as explained in section
5.2.1, the exposure dose applied for all other kVp settings over the clinical kVp range
should be adjusted by the service engineer to provide the same image EI value as
obtained at 81 kVp. The engineer adjusts the AEC kVp response according to an
adjustable lookup table in the AEC control software. Figure 5.1 shows the AEC
control values applied to a Philips x-ray machine in two different ways. The speed
class is an index related to the speed of screen - film response. The definition of speed
class states that a speed class of 100 is accepted as being represented by an air kerma
of 11.1 µGy. Therefore an image plate requiring 2.8 µGy (11.1 µGy / 4) to achieve the
standard optical density (OD) = 1 above the base plus fog values has a speed class of
400 (Heggie et al 2001). CR is generally considered to have a speed class of 200 and
the response shown in Figure 5.1 can be compared to this value. A higher value for a
55
speed class implies a reduction in applied dose compared to a lower speed class value.
Philips use a reference to speed class values to determine the correct AEC calibration
adjustment for the type of x-ray machine used for this thesis. The relative EI is an
index based on the relative effect of kVp on the EI values recorded for test images and
is dependant on the AEC settings applied for each kVp.
The speed class curve shows data for the existing AEC kVp compensation algorithm.
The relative EI curve shows the data for the optimised AEC kVp compensation
algorithm. The relative difference between these two curves is a visual indication of
the differences in the AEC settings between the two test scenarios. The algorithm data
are stored in the AEC software for the Philips x-ray machines and also recorded by the
engineer on his laptop computer for each machine. Other types of machines may have
different arrangements to store the AEC setup values.
For this thesis a process of trial and error was used to obtain the optimal result for EI
consistency. The ideal result is a consistent EI response over the clinical kVp range
and, to a first order, a consistent noise response.
This translation process has already been explored and an EI variation of ± 70 index
values has been derived for Kodak CR systems (Wilkinson and Heggie 1997). Due to
CR’s wide exposure range and the complementary effect of digital image
enhancement techniques, a wider EI range for AECs using CR imaging could be
tolerated. The resulting ± 70 index point variation is less than 2% of the range of
possible EI values available in the Kodak system. However increasing this limit would
56
compromise the purpose of an AEC to minimise dose while maximising the control
over image consistency. The derived ± 70 index point limit should therefore be the
maximum variation allowed, consistent with the 0.2 OD variation allowed for screen-
film systems.
57
Chapter 6
Developing a test procedure to optimise AEC
performance with CR
6.1 Introduction
A range of tests was required to derive the AEC response characteristics of the
existing AEC setup for the x-ray machine. The test sequence was then repeated using
the same AEC which had been optimised for the kVp response of the image medium.
The two sets of response characteristics were analysed and compared to determine the
benefits of optimising AEC kVp compensation. The EI was recorded for each test
exposure of each AEC scenario. Each test exposure was repeated and the exposure
dose at the plane of a CR cassette, located in the Bucky, recorded. All exposures were
performed on a single AEC equipped x-ray machine to avoid inconsistencies due to
variations in operation and performance of different machines even of the same
manufacturer and type.
There are many types of ion chamber available to the medical physicist, each with its
own operating characteristics to match the requirements of the application. Two of
these types of ion chamber are available at Christchurch Hospital. The first of these is
the Capintec graphite lined, parallel plate and Omni directional cylindrical model
58
shown in Figure 6.1. It is ideal for Computed Tomography (CT) where the x-ray tube
rotates about the patient tunnel or “doughnut”. The design allows for the collection of
all x-ray radiation generated in the CT tube’s 360 degree rotation. The chamber’s
susceptibility to all scatter cannot be controlled due to its design requirement to collect
dose from any direction. It could however be accommodated within the confined space
of a Bucky tray for the purpose of the test procedure.
NOTE:
This figure is included on page 59 of the print copy of
the thesis held in the University of Adelaide Library.
Figure 6.1 A graphite lined parallel plate pencil chamber (Reproduced courtesy of
Radcal Corporation).
The second ion chamber available was the Inovision Model 96035B ‘pancake’ type
15 cc ion chamber. As discussed in section 6.3, this chamber satisfies both
requirements.
59
electrons and air ions. A positive 300 volt bias potential is applied to the collector
which separates the ion and electron clouds and sweeps the electrons out of the air
volume and into the electrometer before they recombine again. The guard electrode is
also connected to the 300 volt bias potential. The electrons captured by the collector
electrode form an electrical charge that is measured by an electrometer. The measured
result is displayed, in international units of radiation, on the display panel of the
electrometer.
This type of ion chamber, although designed to reduce the effects of scatter radiation
on dose measurements, is still affected by it.
NOTE:
This figure is included on page 60 of the print copy of
the thesis held in the University of Adelaide Library.
Figure 6.2 The Inovision Model 96035B 15cc Ion Chamber (courtesy of fluke
biomedical)
The Inovision instrument and ion chamber has an annual calibration check performed
by the National Radiation Laboratory (NRL) who issues a calibration certificate with
the test results.
cm30 cm film
cassette Cassette face
cut out
C to
accommodate
15cc pancake
chamber and
connector
65mm dia
Cassette centre
Cable to dosimeter
A test set-up was sought to measure the clinical AEC dose to a CR cassette by locating
the measuring ionisation chamber in the same plane as a cassette placed in a Bucky
tray. The test setup was required to minimise the effects of backscatter on
measurement accuracy. The concept of using a redundant film cassette to
accommodate and locate the ion chamber within the Bucky tray was developed by a
work colleague Stephen Rofe. He also constructed a prototype, based on the drawing
shown in Figure 6.3, using a redundant 24 cm × 30 cm Kodak film cassette which,
subsequent to operational tests to prove its effectiveness, was used in the recording of
dose measurements for this thesis.
The face of the cassette which faces the x-ray tube was cut out to accommodate the
ion chamber and its connector. The active face of the ion chamber is located in the
geometric centre of the cassette and lies flush with its surface. The cassette’s internal
structures including screens were removed from the test cassette. The 2 mm
aluminium rear panel was also cut out to the shape of the ion chamber to enable the
61
placement of various materials in order to test their effectiveness in reducing
backscatter as described in section 6.4.1.
The energy spectrum for this target material has a keV range of approximately 12 keV
– 100 keV. For clarity the characteristic radiation peaks of the tungsten spectrum have
been removed. Depending on the interactions of the primary beam photons with
objects in their path, the x-ray scatter produced may have energies anywhere below
the maximum energy. By definition the scatter energies cannot reach the maximum
energy of 100 keV.
A significant part of the scatter incident on any dose measuring instrument is called
backscatter. This backscatter is generated from x-ray photon interactions with
materials located behind the ion chamber. Backscatter can be generated in all
directions. Since there is no grid to minimise the effects of backscatter to the image
medium, it is detrimental to image quality.
In order to minimise the backscatter component of the dose to the ion chamber, it was
essential to determine the most effective backscatter attenuator to mount on the rear of
the test cassette containing the ion chamber. A series of tests was carried out using a
62
fixed exposure to determine this factor. The attenuating medium was attached over the
rear of the cassette, completely covering the rear of the ion chamber. For consistency
the test exposures were carried out with the same exposure parameters used for setting
up the dose set-point for the AEC given in section 5.4, namely a beam energy of
81 kVp, an SID of 1000 mm and a 150 mm Perspex PEP phantom. The central AEC
chamber only was selected on the x-ray machine’s operator panel.
Since the dose reading includes incident radiation and backscatter radiation, the
attenuating material which results in the lowest recorded dose reading has the best
backscatter reduction characteristic.
Table 6.1 shows the recorded dose results for exposures to the chamber with various
metal attenuators placed behind the ion chamber for backscatter reduction. The most
effective attenuator for backscatter under the test setup condition described was 1 mm
or 2 mm aluminium. The dose measured using this backscatter control medium was
5.2 ± 0.1 µGy. The value derived corresponds to the measured value of the AEC dose
set-point and is within the dose set-point limit of 4.5 ± 1 µGy.
Kodak uses 2 mm aluminium as the support medium for the PSP and to minimise the
effects of backscatter to PSP material. Within experimental error the difference
between the use of 1 mm aluminium and 2 mm aluminium for backscatter reduction
was insignificant. Since Kodak uses 2 mm aluminium for its CR cassettes, this option
was also chosen for the experimental test procedures for this work, both for its proven
reduction of backscatter and for its similarity to the physical design of Kodak CR
cassettes.
63
6.5 X-ray Bucky arrangement in test set-up
Figure 6.5 is a photograph of a typical Bucky assembly as used in the experimental
work. The anti scatter grid is located on the fixed part of the Bucky, shown on the left
of the photograph, just behind the front cover showing the layout of the AEC
chambers. The AEC chambers are also located on the fixed part of the Bucky and are
positioned just behind the anti-scatter grid. The film or CR cassette is clamped onto
the cassette tray shown on the right side of this photograph. The tray is slid into the
fixed Bucky assembly. This locates the CR cassette behind the AEC chambers and
anti scatter grid.
64
oriented along the direction of the x-rays emitted from the x-ray tube and designed to
be focussed at a fixed SID.
Focal spot
X-ray tube
Collimator (LBD)
X-ray
Patient Equivalent Phantom (PEP) beam
Table top
Scatter x-rays
Moving antiscatter
grid (focused)
Film Cassette tray AEC chambers
Figure 6.6 is a diagram of a Potter moving grid Bucky. It was developed in 1920 by
Hollis Potter and uses the principle of reciprocating (back and forth) grid motion not
only to reduce the effects of x-ray scatter but also to reduce the visibility of the grid
lines on an image. Moving grids remove scattered radiation very effectively while
blurring out the image of the grid lines themselves. However, they introduce several
disadvantages: they are more costly, subject to greater failure incidence due to their
mechanical construction and degrade image density. The latter problem requires that
the patient dose be increased to maintain image density and quality.
Essentially, only those x-ray photons emitted from the x-ray tube or exiting a patient
in the same focus direction as the grid structure lines are able to pass through the grid
and contribute to the formation of an x-ray image. This places a high degree of
importance on setting the SID correctly, to match the grids focus distance, to avoid
grid cut-off effects. Any scattered radiation that may otherwise have contributed to
65
image noise is absorbed in the lead strips of the grid and does not contribute to the
formation of the x-ray image. Prior to the development of the moving grid, the
benefits of the stationary version were well appreciated even though the grid lines
themselves may have been visible in the resulting images.
For consistency, a single x-ray machine was used in the collection of all data for these
tests. The x-ray machine was a standard Philips Optimus 50 equipped with a table
Bucky and an erect Bucky, both of which incorporate a typical moving grid
mechanism. All AEC doses were measured at the plane of a CR cassette, when placed
in the Bucky tray, using the test cassette described in section 6.4, with 2 mm
Aluminium in place as the cassette anti back-scatter material. A single Kodak CR 900
reader was used to process the test images. The EI of the homogenous images were
recorded from the ROP after each exposure.
The Inovision dosemeter accuracy has been certified by NRL as – 1 ± 0.1% at 60 kVp,
+ 1 ± 0.1% at 80 kVp and + 3 ± 0.1% at 100 kVp. A series of 20 exposures using the
same test parameters was carried out to determine the reproducibility of the x-ray
machine. The measured dose values were calculated to have a maximum deviation of
± 2.5% from the average. All EI values were calculated as having a maximum
deviation from the average value of less than 1%. These measurement uncertainties
are much smaller than the variations observed between mean values for the different
test settings.
6.7 Summary
Tests were conducted on a single Philips Optimus x-ray machine equipped with AEC.
The service engineer was requested to provide a switchable option which enabled the
66
AEC to perform using the generic algorithm or using the experimentally derived
algorithm which optimally matches the kVp response of the CR imaging medium.
This switching option was set according to which operating mode was required.
Chapter 7 provides the results of tests conducted using each of these scenarios.
The initial series of tests were performed with the generic AEC kVp compensation set
on the x-ray machine. They were then repeated with the x-ray machine set for optimal
AEC compensation designed to match the kVp response characteristics of the Kodak
CR imaging medium as described in section 5.2.2. All tests where recording of image
data was required used an image processing delay of 60 ± 10s between exposure to x-
rays and the placement of the CR cassette into the CR processor.
67
Chapter 7
Test results: the rationale for an optimised AEC
7.1 Introduction
The original idea that began this work was an investigation into the transferability of
current protocols for AEC functionality testing with screen- film imaging systems, to
CR imaging systems. This concept has evolved to become an investigation into the
optimisation of AEC control to ensure the highest level of image consistency over the
clinical kVp range. In order to perform this investigation an experimental procedure
was developed which provided a repeatable test setup to record the results for both the
current AEC control algorithm and for an optimised AEC. This chapter is the
culmination of this development, providing the experimental results with discussions
pertaining to the benefits of optimised AEC control.
The error bars shown on the following test results for dose measurements relate to the
dosemeter accuracy stated in section 6.6 from a report by NRL. The image EI
accuracy is derived from the results from experimentation with image processing
times given in section 4.12.
68
7.3 Test results for an AEC with generic kVp
compensation applied for CR
The histograms of Figure 7.1 and Figure 7.2 are the results of tests conducted on the x-
ray machine with the AEC set to generic kVp compensation. They show the dose
profile and corresponding IP exposure index (EI) response, respectively, over the
clinical kVp range 60 kVp to 125 kVp and PEP thicknesses of 75 mm to 300 mm
Perspex simulating exposures over the normal range of patient sizes.
The AEC dose response shown in Figure 7.1 exhibits a dose average of 0.65 mR with
a maximum variation from this average of 0.17 mR representing 26% dose variability.
0.9
0.8
0.7
0.6
Dose mR
0.5
0.4 75mm perspex
Figure 7.1 Generic AEC response to kVp and PEP: Dose response
The EI response to changes in kVp and PEP thickness plotted in Figure 7.2 shows a
similar response pattern. Scatter in the absorber material is perhaps the most important
factor which influences the EI pattern deviations from the dose pattern. As the
exposure kVp increases, the scatter in any absorber material in its path increases. The
dose to the IP decreases and the AEC prolongs the exposure to maintain the dose to its
sensors. This results in a higher dose being recorded by the measuring dose chamber
which is placed in the same relative position as the AEC’s centre chamber. The IP is
itself a dosemeter but its response is kVp dependant and the reported EI value is the
average pixel value over the entire IP. The overall effect is a reduction in the reported
EI values for increasing kVp and for the scattering effects due to increasing absorber
thickness.
69
1880
1860
1840
1820
1800
Exp. Index(EI)
1780
1760
1740
1720 75mm perspex
1700 150mm perspex
1680
1660 225mm perspex
1640 300mm perspex
1620
1600
1580
1560
60 70 81 102 125
kVp
Figure 7.2 Generic AEC response to kVp and PEP variability: EI response
0.90
0.85
0.80
0.75
±10%
0.65
0.60
0.55
0.50
0.45
0.40
1600 1620 1640 1660 1680 1700 1720 1740 1760 1 780 1800 1820 1840 1860 188 0
Exposure Index
Figure 7.3 Generic AEC response to kVp and PEP: Dose and EI distribution
The average EI value was 1740 with a maximum EI variation from this average of
100 index points. This variation exceeds the theoretical ± 70 EI limit, as described in
section 5.3, of a consistent response to changes in PEP thickness over the clinical kVp
range. While these results for this thesis were obtained from a single x-ray machine
they are typical of several AEC equipped x-ray machines at the hospital. They
70
represent the status of the generic AEC kVp compensation response before the
experimental kVp compensation was installed.
The graph in Figure 7.3 shows the dose and EI distribution of the data from Figure 7.1
and Figure 7.2. It illustrates that the generic AEC kVp compensation algorithm
produces a dose and EI profile which is inconsistent over the clinical kVp range. A
10% dose limit (shown by the dashed magenta lines) placed about the average dose
value of all the measurements highlights the fact that several measurements fall
outside this range depending on error values. The EI range over these measurements
was 190 index points with an average EI of 1740 resulting in a range of 1740 ± 95 EI.
Some values lie outside the theoretical limit of ± 70 EI referred to in section 5.3, even
allowing for the error values. The relationship highlighted by this set of results will be
used later to compare with the relationship produced when the AEC has been
optimised to match the CR response to kVp.
0.9
0.8
0.7
0.6
Dose (mR)
The histograms shown in Figure 7.4 and Figure 7.5 are the results of a repeat of the
tests conducted to derive the results for the histograms in Figure 7.1 and Figure 7.2 but
with optimised kVp compensation applied to the AEC. This is the process by which an
AEC is adjusted to optimally match the kVp response of the imaging medium as
described in section 5.2.2.
71
The dose average for the results shown in Figure 7.4 is 0.612 mR with a maximum
dose variation from this average of 0.051 mR representing 8.3% variability. The EI
response to changes in kVp and PEP shown in Figure 7.5 has an average EI of 1715
with a maximum variation from this of 35 index points.
1880
1860
1840
1820
Exposure Index (EI)
1800
1780
1760
1740 75mm perspex
1720
1700 150mm perspex
1680 225mm perspex
1660
1640 300mm perspex
1620
1600
1580
1560
60 70 81 102 125
kVp
0.90
0.85
0.80
0.75
0.70
Dose mR
Dose average
0.65
±10%
0.60
0.55
0.50
0.45
0.40
1600 1620 1640 1660 16 80 1700 1720 1740 1760 1780 1800 1820 1840 1860 1880
EI
The graph of Figure 7.6 shows the dose and EI distribution of the data from Figure 7.4
and Figure 7.5. It illustrates that the AEC, optimised to match the kVp response of the
image medium, produces a dose and EI profile that approaches the ideal situation of a
72
constant dose and EI over the clinical kVp and PEP thickness range. There are many
exposures around the ideal 1710-1720 EI range. Taking into account the error, most
measurements fall within a 10% dose range about the average dose of 0.612 mR
shown on Figure 7.5 by dashed magenta lines. The EI range, over all these
measurements, was 70 index points with an average EI of 1715 corresponding to a
variation of 1715 ± 35 EI.
73
Chapter 8
Image Quality comparison of AEC setups
8.1 Image Quality testing
Having established that better optimisation of AEC functionality with a CR imaging
medium can be achieved by applying appropriate AEC kVp compensation, it is
necessary to determine the effect of this compensation on the system image quality. In
any radiographic imaging system, image quality can be determined by assessing
contrast detail and spatial resolution performance. These in turn are dependent on:
System noise
Quality
74
Figure 8.1 Leeds TO. 20 contrast-detail test object radiographic layout
75
Window
width
Level adjust
Image data
Light Dark
contrast values. Each image pixel will have a contrast value attributed to it once it has
been scanned by the CR laser. Figure 8.2 shows the layout of the Kodak histogram.
The Kodak image display software package incorporates a histogram which shows the
pixel value on the 0- 4096 scale on the horizontal axis and the number of pixels on the
vertical scale. The process of applying the full 0- 4096 pixel range to different parts of
the available image data is called windowing and levelling. The dark diagonal line in
Figure 8.2 covers the range of pixel values available. When the slope of the black line
is increased the image component window is decreased such that less of the total
image is included in the pixel range. This expands and amplifies small changes in
pixel values of the selected image pixel range which has the effect of increasing the
image detail visibility. If the thick black line is moved to the left, without changing its
slope, the image window remains the same but the image content containing the lower
value pixels are highlighted. In this circumstance this means toward the lighter
components of the image content. The converse occurs when the line is shifted to the
right where the larger pixel values or darker components of the image are highlighted.
100
80
60 Generic A EC
40 Optimised AEC
20
0
60 81 125
kVp
Figure 8.3 Contrast detail comparison between the two AEC compensation scenarios
The variability of the observers’ subjective assessment is indicated by the error bars.
The reasons for this variability even for a single observer can be understood if slight
changes to window and level adjustments are considered in conjunction with the
decision process for the observation of each detail object between observations. The
differences between the results for each AEC scenario at each kVp setting are not
statistically significant.
77
Figure 8.4 X-ray image of the Nuclear Associates spatial frequency test phantom
The accurate representation in digital form of an analogue signal such as the sinusoidal
spatial frequencies of the test phantom is based on the Nyquist frequency. The Nyquist
frequency is explained in the Nyquist ‘Sampling Theorem’ (Nyquist 1928). According
to the Nyquist sampling theorem, an analogue signal must be sampled at twice the
frequency of its highest-frequency component in the signal to be converted into an
adequate representation of the signal in digital form. Thus, the "Nyquist frequency" is
the highest frequency that can be accurately sampled. It is one-half of the sampling
frequency. If sampling of an analogue signal is carried out according to the Nyquist
theorem then the reconstruction of the digital signal can restore the analogue signal
without error.
78
Kodak sets its limits for the visualisation of line pair patterns at 90% of the Nyquist
frequency as shown in Table 8.1 for each available screen size at a beam energy
setting of 75 kVp. This allows a margin from the theoretical maximum resolution
given by the Nyquist theorem.
The physical dimensions of a Kodak CR IP are related to its pixel size which is a
determinant of its spatial resolution characteristic. At Christchurch Hospital three
cassette (IP) sizes are in use. Their dimensions are 35 cm × 43 cm, 24 cm × 30 cm and
18 cm × 24 cm and have pixel sizes as displayed in Table 8.1. These exhibit line pair
resolution limits based on 90% of the Nyquist limit ranging from 2.7 lp / mm for the
largest pixel size to 4.6 lp / mm for the smallest pixel size.
The spatial resolution of any CR system also depends on such factors as the
dimensions of the scanning laser spot and the amount of light scattering within the
phosphor. The latter effect depends on the phosphor composition and its thickness.
The spatial resolution assessment uses comparative pattern visualisation from test
images to determine differences in spatial resolution between AECs with generic AEC
kVp compensation and AECs which have been optimally kVp compensated for a CR
imaging medium.
79
A single 24 cm × 30 cm CR cassette was used for all tests to maintain result
consistency. This was placed in the central position in the x-ray Bucky cassette tray,
with the same orientation for each exposure. The line pair test phantom was placed on
the front of the erect Bucky over the central AEC field area and oriented so that the
length of the phantom lay perpendicular to the long side of the cassette. In this
orientation, the phantom lies perpendicular to the laser scan direction for the cassette
type used. The CR reader scan directions for this cassette type are illustrated in
Figure 8.5. The raster scan lines in this image are slightly skewed, relative to the IP
edge in the fast or laser scan direction. This is because the plate undergoes
simultaneous laser beam scanning and linear plate translation (AAPM TG 10 2002).
NOTE:
This figure is included on page 80 of the print copy of
the thesis held in the University of Adelaide Library.
Figure 8.5 CR IP laser scan and plate movement orientations and typical resolutions
(courtesy of J.Anthony Seibert, PhD 2003)
80
This implies an error variation of approximately ± 0.4 lp/mm about the observed line
pair limit. This error variation is due to the observer’ window and level settings
between successive readings and the decision process for line pair visualisation.
The subjective response of the observer is influenced by the ambient light in the image
viewing area and the distance of the observer from the image surface.
5.0
4.5
Line pairs observed
4.0
3.5
3.0
2.5
Generic
compensation
2.0
1.5 Optimised
1.0
Compensation
0.5
0.0
60 81 125
kVp
The standard rules given by Kodak apply to the decision making process such that a
line pair is considered visible if any parts of the lines are visible throughout their
length. There is no statistically significant difference in the spatial resolution obtained
with the two AEC compensation scenarios; the maximum variation of 0.3 lp / mm is
smaller than the experimental error.
A comparison of the relative SNR for the two AEC test scenarios was made using the
standard test exposure parameters presented in section 6.6 to obtain a series of
exposures. For consistency, a single 35 cm × 43 cm CR cassette was used throughout
81
this test procedure. For each AEC test scenario, the flat field test images were saved to
PACS and reviewed using ‘e-film’ on a PACS workstation. No post processing of the
processed images was enabled.
The homogenous images produced were analysed by sampling five regions of interest
(ROI) of equal size and recording the pixel standard deviation (PSD) for each area.
The PSD is a function provided by the image analysis software and gives the variation
in pixel homogeneity due to noise factors within each sample area. The ROIs were
taken from each corner and the centre of these images. The PSD provides a measure of
image noise within each ROI such that a low PSD value indicates that pixel values are
uniform and noise is minimal. A high PSD value indicates that the variation between
pixel values in an area is large and that image noise is more significant.
23
22
Pixel standard deviation
21
20
AEC standard
19 compensation
AEC optimised
18 compensation
17
16
50 60 70 80 90 100 110 120 130
kVp
Figure 8.7 Image noise comparison: Generic AEC kVp compensation / AEC with
optimised kVp compensation
The average PSD of all five regions of the test images was calculated to provide the
data for the image noise results shown in Figure 8.7. The error bars shown on this
graph are given to indicate the range of values covered by the measurements at each of
the five sampled regions. Although quantum mottle effects increase with increasing
kVp, the effects of increased scatter components at higher energies is greater. The
combined effect causes an increase in PSD which is evident in the results for both
scenarios. The optimised kVp compensated AEC shows a more consistent noise
response which is directly attributable to the greater consistency of the optimised AEC
dose profile.
82
The results obtained follow a proposal by Christodoulou et al (2000) that an AEC
system is correctly set up when it produces a consistent PSD on homogenous images
over the clinical kVp range. The results of optimally setting AEC kVp response for an
image medium are consistent with Christodoulou’s proposal. Other similar
experimental work by Mazzocchi et al (2006) explains a simple AEC calibration
procedure for CR systems that functions by correlating the digital signal exposure
index with the SNR. The original purpose of AEC was to provide image consistency
over a wide range of exposure conditions. EI consistency is a measure of image
consistency over the clinical kVp range and is interrelated to improvements to image
noise consistency.
To assure some confidence in this statement, a limited clinical image assessment was
carried out. Its purpose was to ascertain if a clinician could discern any significant
changes between images produced from an AEC using generic kVp compensation and
those performed using an AEC which has been optimally compensated for the kVp
response of the imaging medium. A blind study of several randomly selected chest
images produced from both AEC test scenarios was presented to a radiologist for
assessment. All images were reported as being radiologically similar and acceptable as
diagnostic in both image quality and the level of noise present. This result is an
indication that any differences between similar images produced under either AEC
83
compensation scenario have no clinical implications when applied to a Kodak CR
system.
84
Chapter 9
A protocol to test AEC functionality by the
measurement of dose alone
9.1 The advantages of AEC optimisation
The AEC optimisation process described in section 6.6 and the experimental results
obtained from following this process presented in chapter 7 validate a simplified AEC
functionality test procedure using the measurement of dose alone to an image plate.
The results from section 7.3 show that the dose profile of an optimised AEC can be
constrained within close limits over the clinical kVp range whilst achieving consistent
EI results. Image quality from a clinical perspective has been shown to be
diagnostically acceptable under these conditions.
The AEC optimisation procedure for CR described in section 5.2 relies on the ability
to adjust the AEC control circuitry appropriately for the imaging medium. The AEC
circuitry of older x-ray machines may not be adjustable or may require hardware
changes to accommodate the requirements for CR. Some later AEC designs may be
software controlled but may not have the adjustment range to accommodate CR
response characteristics. Then there is a potential problem that the need for such a
calibration may not be fully recognised because good quality images will be produced
with CR, even when the AEC calibration is incorrect (Greig and Williams 2005).
When an AEC is optimally calibrated to match the response of an imaging medium the
result is greater EI and dose consistency. There is also a greater level of consistency in
image quality and noise response. However, while the experimental results provide the
evidence of quantitative improvements in image quality and noise, they do not
manifest themselves to a level which is quantifiable when viewing test images. This
86
has been confirmed by a blind study of clinical images randomly selected from a range
of images and derived from exposures produced with and without AEC kVp
optimisation. This result implies that no changes to clinical procedures or practices are
required due to the introduction of the proposed AEC assessment protocol.
9.2.1 Test time reduction for AEC assessment using a dose alone
protocol
A dose alone measurement protocol for AEC testing will substantially reduce the AEC
functionality testing time for the Medical Physicist. The actual test time reduction will
depend on the test mode currently employed by the Physicist and the combination of
equipment at a site. Test procedures employed by other medical physicists groups may
vary from those described for this thesis. Any variations from the described test
procedures can have a direct impact on the AEC functionality test time. For example,
the number of people involved in the test process and the choice of recording
exposures on individual films compared to the rotating window image capture method
will each have a direct influence on the time required to perform the test procedures.
Table 9.1 illustrates the test sequence and test times at Christchurch Hospital for a
conventional AEC functionality test sequence for a single Bucky. The testing
sequence is divided into several operations or tasks to which the total test time is
apportioned. Tasks marked with an asterisk are common to screen-film and CR based
AEC setups.
It should be noted that the CR related times are for a Kodak CR system. Other CR
systems such as Agfa allow a region of interest (ROI) to be selected on their image
monitor and an image mean pixel value can be recorded from this. For this type of CR,
the rotating lead window method of recording AEC exposures on a single IP referred
to in section 3.6 could be used to good advantage to improve test efficiency.
The times shown are based on my own experience with AEC testing at the Hospital.
Over three test sequences the average test time was 44 ± 10 minutes. The variability
factor allows for the operational nuances of individual x-ray machine, test set-up time
and unplanned delays that can arise in the normal run of a test sequence.
When a dose alone measurement AEC protocol is used the total test time for the
Medical Physicist is 22 minutes, a reduction of 50% when compared to the normal
AEC test time of 44 minutes. A single x-ray room can be equipped with two Buckys
which implies a 44 minute time reduction per room. Since no image processing is
87
required the CR reader is always available for clinical use. This has a positive
influence on the operational efficiency of the Radiology Department.
The practical time savings made using the dose alone protocol will depend on
individual test methods in current use which may differ from that described in this
work.
88
achieved greater consistency under these conditions but also that measured AEC dose
values were more consistent over the clinical kVp range.
The outcome of these tests was used to develop a new Medical Physics protocol for
testing AEC functionality within the CR environment using just the measurement of
AEC dose alone. The successful application of this protocol relies on the
implementation of and adherence to stringent AEC set up procedures for all new and
existing x-ray installations. The protocol requires optimisation of the AEC calibration
to match the kVp characteristic of the imaging medium for each x-ray machine.
Optimisation of AEC calibration implies consistent exposure index values over the
clinical kVp range 60 kVp – 120 kVp. The dose variation over this kVp range should
also be consistent around the set dose-point value of the AEC. The experimental
results have shown that an 8% variation of dose about the dose set-point is achievable
corresponding to range of ± 35 exposure index values. Accordingly a variation limit of
twice the experimentally attained value would provide a reasonable limit of
expectation for a test procedure to provide a method to assess the accuracy of AEC
functionality.
The use of a dose alone measurement protocol for AEC testing still requires a regular
qualitative assurance program to assess the accuracy of AEC functionality and to
assure consistent imaging quality.
Doyle and Martin (1996) proposed a very similar procedure for determining AEC
consistency. They are in concordance with this thesis that image receptor dose levels
must change to reflect the variable kVp response characteristic of the receptor to
obtain consistent image detector dose indices (equivalent to EI values used in this
thesis). They also determined that the relative response of the detector dose indices at
different tube potentials follows a similar pattern to the SNR and relates to the level of
image quality. The SNR was defined as the mean pixel value divided by the standard
deviation and taken from regions of interest of approximately 10 cm2 drawn at the
centre of each image. This concept was explored in section 8.4 of this thesis and
resulted in similar findings. The fundamental difference between the procedures used
by Doyle and Martin (1996) and this thesis is the analysis of image exposure indices to
determine AEC consistency compared to optimizing image receptor dose values.
This thesis derives a process to achieve optimised kVp compensation for any AEC
associated with any imaging medium which consequently enables the use of a test
procedure which measures the consistency of dose to a CR IP to assess AEC
functionality. There is no apparent significant clinical impact to normal radiographic
practice due to the implementation of this process however it may provide benefits in
operational efficiency with reductions in machine test times. In contrast to the
subjectively variable nature of the measurement of image quality, the effects of the
proposed AEC setup protocol on the measurements of EI, exposure dose and image
noise are based in physics and produce quantifiable results. The test results show the
benefits of achieving greater dose control consistency for AEC controlled exposures.
90
absolutely consistent as long as the image medium has a non linear response to kVp.
However it is the attainment of maximum consistency which makes it reasonable to
accept the measurement of dose alone to be sufficient to indicate the state of an AEC’s
calibration. The test time to assess AEC performance can be considerably reduced
when a dose alone measurement protocol is used. This is achieved because there is no
requirement to expose and process an IP for each exposure of a full AEC test sequence
of variable kVp, absorber thickness and density control.
END
91
References
AAPM (American Association of Physics in Medicine) 2002 Acceptance testing and
Quality control of photostimulable storage phosphor imaging systems report of
task group #10 Draft Document
Doyle, P. Gentle D. and Martin C.J. 2005 Optimising automatic exposure control in
computed radiography and the impact on patient dose. Radiation Protection
Dosimetry: 14:236-239
Doyle, P. and Martin C.J. 2006 Calibrating automatic exposure control devices for
digital radiography. Physics in Medicine and Biology 51: 5475-5485
Grieg, L. and Williams J. 2005 AEC Calibration with CR systems’ Oral presentation
British Institute of Radiology Scientific meeting, Department of Medical
Physics, Royal Infirmary of Edinburgh, NHS Lothian.
<www.mph.ed.ac.uk/nhs/downloads/CR0205Abs2.pdf> Accessed November
18, 2005
92
Hale, J. 1989 Routine calibration of automatic exposure control systems for diagnostic
x-ray machines: A 5yr report. Health Physics 57 (6): 1021-1024.
Heggie, J.C.P. Liddel N.A. and Maher K.P. 2001 Applied Imaging Technology
Fourth Edition. St Vincent’s Hospital.
Hunt, A. and Plain S. 1993 A simple solution to the problems of testing automatic
exposure control in diagnostic radiology. The British Journal of Radiology 66:
360-362
Mould R.F. 1995 The early history of x-ray diagnosis with emphasis on the
contributions of physics 1895-1915. Physics in Medicine and Biology 40: 1741-
1787
NIST [National Institute of Standards and Technology] x-ray mass attenuation tables
U.S. Department of Commerce
<http://physics.nist.gov/Divisions/Div846/div846.html> Accessed March 09,
2004
Newing, A. 1999 Light visible and invisible and its medical applications. British
Journal of Radiology 74: 212
NRL [National Radiation Laboratory] C5 1994 Code of safe practice for the use of x-
rays in medical diagnosis. Ministry of Health Christchurch New Zealand.
93
Orand, M. 2004 ‘Cost Analysis: Film vs. PACS’
<www.radiologytoday.net/archive/rt_051004p38.shtml> Accessed October 06,
2003
Rowlands, J.A 2002 The Physics of Computed Radiography. Physics in Medicine and
Biology 47:R123-R166.
Van Loon, R. and Van Tiggelen R.2004 History of dosimetry in radiology and early
radiotherapy
<www.radiology-museum.be/English/Collection/RadioprotectHist.asp>
Accessed 20 July 2005
Wilkinson, L.E. and Heggie J.C.P. 1997 Determination of Correct AEC Function with
Computed Radiography Cassettes. Australian Physical and Engineering
Sciences in Medicine 20 (3):186-191
Workman, A. and Cowan A. 1993 Signal, Noise and SNR transfer properties of
computed radiography. Physics in Medicine and Biology 38:1789-1808
Yaffe, M. and Rowlands J. 1997 X-ray detectors for digital radiography. Physics in
Medicine and Biology 42:1-39.
94
Index of Figures
Figure 2.1 The rotating anode x-ray tube with line focus .............................................. 14
Figure 2.2 Ernest Harnack (around 1896) using a portable military set-up, including a
Ruhmkorff coil and early vacuum tube. (Photograph courtesy of the
Radiology Museum Brussels, Belgium) ..................................................... 15
Figure 2.3 The Holzknecht chromoradiometer scale (Photograph courtesy of the
Radiology Museum Brussels, Belgium) ..................................................... 17
Figure 2.4 X-ray energy spectrum from a tungsten target ............................................. 18
Figure 2.5 AEC control schematic diagram................................................................... 21
Figure 2.6 X-ray machine erect Bucky (centre left of photo) showing the physical
arrangement of the three ionisation chambers used as x-ray detectors for
the AEC. ...................................................................................................... 22
Figure 2.7 Standard AEC 3 chamber thumbnail layout ................................................. 23
Figure 3.1 Typical screen-film characteristic curve (H&D curve) ................................ 26
Figure 3.2 Typical kVp response of a Kodak T-Mat-G-RA type screen-film system to
a fixed dose of 0.8 mR ................................................................................ 27
Figure 3.3 Physical layout of the AEC test cassette showing film window at position 132
Figure 4.1 Energy level diagram of PSP (Reproduced courtesy of Kenji Takahashi
with permission from Elsevier Limited) ..................................................... 36
Figure 4.2 X-ray absorption vs photon energy............................................................... 38
Figure 4.3 The kVp response of a Kodak CR PSP exposed to 1.0 mR ......................... 40
Figure 4.4 The exposure dose range of a Kodak CR system ......................................... 41
Figure 4.5 The IP response to image processing delay following exposure to 10 mR .. 49
Figure 4.6 The process of Photostimulated Luminescence (Reproduced with courtesy
from J.Anthony Seibert, Ph.D 2003)........................................................... 52
Figure 5.1 Philips AEC kVp compensation ................................................................... 55
Figure 6.1 A graphite lined parallel plate pencil chamber (Reproduced courtesy of
Radcal Corporation). ................................................................................... 59
Figure 6.2 The Inovision Model 96035B 15 cc Ion Chamber (courtesy of fluke
biomedical).................................................................................................. 60
Figure 6.3 Test cassette layout ....................................................................................... 61
Figure 6.4 X-ray spectrum for tungsten target ............................................................... 62
Figure 6.5 Bucky cassette tray ....................................................................................... 64
Figure 6.6 Potter Bucky / moving grid Bucky ............................................................... 65
Figure 7.1 Generic AEC response to kVp and PEP: Dose response.............................. 69
Figure 7.2 Generic AEC response to kVp and PEP variability: EI response................. 70
Figure 7.3 Generic AEC response to kVp and PEP: Dose and EI distribution.............. 70
Figure 7.4 AEC optimally kVp compensated: Dose response....................................... 71
Figure 7.5 AEC optimally kVp compensated: EI response ........................................... 72
Figure 7.6 AEC optimally kVp compensated: Dose and EI Distribution...................... 72
95
Figure 8.1 Leeds TO. 20 contrast-detail test object radiographic layout ....................... 75
Figure 8.2 The Kodak image histogram......................................................................... 76
Figure 8.3 Contrast detail comparison between the two AEC compensation scenarios 77
Figure 8.4 X-ray image of the Nuclear Associates spatial frequency test phantom ...... 78
Figure 8.5 CR IP laser scan and plate movement orientations and typical resolutions
(courtesy of J.Anthony Seibert, Ph.D 2003) ............................................... 80
Figure 8.6 The spatial resolution (line pairs) test result................................................. 81
Figure 8.7 Image noise comparison: Generic AEC kVp compensation / AEC with
optimised kVp compensation ...................................................................... 82
96
Index of Tables
Table 3.1 Radiographic comparison: PEP to human tissue. (Data reproduced from
National Institute of Standards and Technology ( NIST) x-ray mass
attenuation tables) ....................................................................................... 31
Table 4.1 Kodak screen and pixel sizes (Bogucki et al 1995) ....................................... 43
Table 6.1 The effectiveness of various attenuators to reduce the backscatter incident to
an ion chamber placed within a Bucky cassette tray................................... 63
Table 8.1 Kodak CR specifications and spatial resolution limits at 75 kVp.................. 79
Table 9.1 Breakdown of AEC test sequence times for a single Bucky.......................... 88
97
Acronyms
AEC Automatic exposure control
CR Computed Radiography
DR Digital Radiography
EI Exposure Index
IP Imaging plate
Kgs Kilograms
OD Optical Density
TFT Thin film transistor (technology used in flat panel imaging arrays)
99