Defibrillator Sync Testing
Defibrillator Sync Testing
Abstract—A defibrillator sync output signal connector The delivery of a therapeutic dose of electric current to
provides an ECG synchronization signal that can be used by the heart during the R-wave of the cardiac cycle is the key
some defibrillators for the purpose of performing to the successful conversion of supraventricular
synchronized cardioversion [1]. This process is used to stop
an abnormally fast heart rate or cardiac arrhythmia by the
tachyarrhythmias to normal sinus rhythm. It is critical to
delivery of a therapeutic dose of electric current to the heart have the patient monitor and the defibrillator synchronized
during the R-wave of the cardiac cycle. Timing the shock to properly during cardioversion. To ensure the proper
the R-wave prevents the delivery of the shock during the synchronization, the patient monitor’s defibrillator sync
vulnerable period of the cardiac cycle, which could induce and analog output function need to be checked. Note that
ventricular fibrillation [2]. GE patient monitors include a the synchronization is not required in the most frequent
selectable analog output feature, which provides an analog
ECG or arterial blood pressure signal. The blood pressure
uses of defibrillators: ventricular fibrillation and in AEDs.
signal can be used to synchronize balloon pumps to provide But whether the clinical use is frequent or infrequent is not
cardiac assist to post-MI patients with poor injection fraction. as important to us as much as the simple fact that
Proper operation requires the defibrillator sync and analog defibrillator sync testing is required. Checkouts are done
output function to be checked. Checkouts are typically done during planned maintenance and after major part
during planned maintenance and after major part replacements such as a patient monitor’s main CPU board.
replacements such as patient monitor’s main CPU board.
Checking out defibrillator sync signals could be done using a
The GE defibrillator sync tester can be used as an easy
GE defibrillator sync tester. The defibrillator sync tester to use, portable and low-cost solution when checking out
provides a loop back path for the defibrillator sync signals to the defibrillator sync/analog out signals on the patient
be displayed on the patient monitor screen and eliminates the monitors.
need for an external oscilloscope.
3013
Note: The patient monitor user manual can be D. BP Test
referenced to set BP scale. 1) Set the selector switch of the tester to BP.
6) Connect the analog output/defib sync cable to the 2) On the patient monitor observe that the V lead
analog output/defib sync port on the patient monitor. waveform matches the BP waveform (Figure 5).
7) Connect the bare wires from the analog output cable to
their corresponding connections on the tester. Connect
the tester ground to the analog out ground. Use the
patient monitor service manual to determine the
ground terminal and wire.
Note: “Mark out” from Analog output/ Defib sync
cable connects to “MARK” on the tester.
Note: The patient monitor’s service manuals can be
referenced for the test cable pinout and color codes.
8) Plug the RL ground cable from the tester into the
patient simulator’s RL terminal. Ensure both patient
lead wire and the tester ground cable are connected to
the RL terminal.
9) Disconnect the V lead of the patient cable from the
simulator and connect it to the “V LEAD” terminal of
Fig. 5. BP test results
the tester (Figure 3).
E. Marker Out Test
1) Set the selector switch of the tester to “MARK”.
2) On the patient monitor observe that the V lead
waveform has a marker pulse corresponding to the R
wave of the primary ECG waveform (Figure 6).
C. ECG Test
1) Set the selector switch of the tester to ECG.
2) On the patient monitor observe that the V lead
waveform matches the primary ECG waveform Fig. 6. Marker Out test results
(Figure 4).
F. Marker In Test
1) Set the selector switch of the tester to ECG.
2) Connect the analog output/ defib sync cable “Marker
in” and “Marker out” wires together by connecting
both to the “MARK” on the tester.
3) On the patient monitor verify markers are present on
the ECG waveform (Figure 9).
Note: Markers are shown below from the GE DASH
patient monitor. The marker display may vary
depending on the model of the monitor.
3014
[3] R. J. Myerburg, A. Castellanos, “Cardiac arrest and sudden cardiac
death,” In: Braunwald E, ed. Heart Disease: A Textbook of
Cardiovascular Medicine. Philadelphia, W. B. Saunders; pp. 890-
931, 2001.
[4] D. S. Pinto, et al. “Sudden cardiac death,” In: Fuster V, ed. Hurst’s
The Heart. New York: McGraw-Hill; pp. 1015-1048, 2001.
[5] R. F. Gillum, “Sudden cardiac death in Hispanic Americans and
African Americans,” Am J Public Health. 87, pp. 1461-1466,1997.
[6] R. F. Gillum, A Folsom, R. V. Luepker, et al. “Sudden death and
acute myocardial infarction in a metropolitan area,” 1970-1980: The
Minnesota Heart Survey. N Engl J Med. 309, pp. 1353-1380, 1983.
[7] R. F. Gillum “Sudden coronary death in the United States 1980-
1985”. Circulation; 79, pp. 756-765, 1989.
[8] M. S. Eisenberg, B. T. Horwood, and R. 0. Cummins et al, “Cardiac
arrest and resuscitation: A tale of 29 cities,” Annals of Emergency
Medicine, 19, pp. 179-186, 1990.
Fig. 7. Marker In test results [9] Cardiac monitors, heart rate meters, and alarms, American National
Standard ANSI/AAMI EC13, Arlington, VA, Association for the
G. Test complete Advancement of Medical Instrumentation, 2002.
If any of these tests fail, troubleshoot analog output or
remove the patient monitor from the service.
IV. CONCLUSION
In general we have shown that a simple tester that
leverages a patient monitor’s built in display capability can
be an effective and inexpensive alternative to an
oscilloscope for testing the defibrillator sync output of a
standard patient monitor.
There are some drawbacks to the tester. It is highly
specialized to this one test. It relies on a visual comparison
of the signals. At a typical ECG display time scale of
25mm/sec a difference of 35ms is slightly less then 1mm
and can be difficult to detect. However, visual comparison
has been judged to be sufficient for use in the previous
procedures. Even in the case when an oscilloscope is used,
the ability to quantify the timing difference in two
channels is not used. In the case where more general
purpose and/or more quantitative measurements are
required an oscilloscope is still the instrument of choice.
But these cases are expected to be infrequent.
As a result GE now manufactures and sells a standalone
defibrillator sync tester. We expect the tester to make it
easier to execute the defibrillator sync test procedure for
service providers. We also hope to reduce the possibility
that the procedure might be inadvertently missed. As a
result signals required to operate a defibrillator are
properly verified.
ACKNOWLEDGMENT
We would like to acknowledge Mike McLeod for design
and research support.
REFERENCES
[1] R. J. Fairbanks et al. “Usability Study of Two Common
Defibrillators Reveals Hazards,” Annals of Emergency Medicine, 50
(4), pp. 424-432, 2007.
[2] E. H. Shortliffe et al. “Medical Systems and Rehabilitation,” Human
Factors and Ergonomics Society, 2004.
3015