AJ30, AJ30ND Service Manual Rev1.0
AJ30, AJ30ND Service Manual Rev1.0
ND
Abacus junior 30
Hematology Analyzer
Service Manual
Rev 1.0
INTRODUCTION
TABLE OF CONTENTS
Abacus junior 30 .............................................................................................................................................. 1
Abacus junior 30ND ........................................................................................................................................... 1
Hematology Analyzer ................................................................................................................................ 1
1.1. Name and serial number ............................................................................................................. 5
1.2. Intended use .................................................................................................................................... 5
1.3. Integrated software....................................................................................................................... 5
2. FUNCTIONAL DESCRIPTION ............................................................................................................. 6
2.1. Main electronic parts of the analyzers................................................................................... 6
2.1.1. Counting chamber with electrodes and measuring aperture .............................. 8
2.1.2. HGB Measuring Head ........................................................................................................... 9
2.1.3. Cell Counter Amplifier Board ........................................................................................ 10
2.1.4. MAIN CPU Board................................................................................................................. 10
2.1.5. Dimm-PC* Module ............................................................................................................. 12
2.1.6. Opto sensors......................................................................................................................... 13
2.1.7. Valve boards ......................................................................................................................... 13
2.1.8. TFT Display and START button Board ....................................................................... 14
2.1.9. External Power Supply ..................................................................................................... 14
3. MAIN MECHANIC AND FLUIDIC PARTS ..................................................................................... 15
3.1.1. Sample/Horizontal and Vertical motors ................................................................... 16
3.1.2. Sampling needle.................................................................................................................. 17
3.1.3. Needle washing head ........................................................................................................ 17
3.1.4. Puffer reservoir ................................................................................................................... 18
3.1.5. Dilutor block......................................................................................................................... 18
3.1.6. Measuring block.................................................................................................................. 19
3.1.7. Pump ....................................................................................................................................... 19
3.2. Assembled Analyzer................................................................................................................... 20
3.2.1. Front Panel............................................................................................................................ 20
3.2.2. Rear Panel ............................................................................................................................. 20
3.2.3. Construction Front ............................................................................................................ 21
3.2.4. Construction Side ............................................................................................................... 21
4. OPERATION OF THE FLUIDIC SYSTEM ...................................................................................... 23
4.1. Initialization of the Fluidic System....................................................................................... 23
4.2. Operation of the fluidic system in Abacus Junior 30 ..................................................... 24
4.2.1. Flow diagram of measurement ..................................................................................... 24
INTRODUCTION
1. INTRODUCTION
Since Abacus junior 30 and Abacus junior 30ND differ in software we issue a common
Service Manual covering both instruments. Information herein applies for all
instruments unless otherwise noted.
To be well up in the instruments, please read this manual carefully to have the
knowledge for servicing the instruments perfectly and avoid extra costs and wasting
precious time.
In this manual, we are using the following conventions:
AJ30 – stands for Abacus junior 30
AJ30ND – stands for Abacus junior 30ND
This Abacus junior 30 Service Manual contain the functional descriptions of all
analyzers, operation of the fluidic systems, adjustments and settings, and very important
information for the Service Personnel about the service operations and possible
problems.
Every ABACUS JUNIOR 30 software version is upgradeable (using an USB flash drive) by
the latest program developed by Diatron, and it can be downloaded from:
http://www.diatron.com
FUNCTIONAL DESCRIPTION
2. FUNCTIONAL DESCRIPTION
2.1. Main electronic parts of the analyzers
ABACUS JUNIOR 30 contains the following electronic parts:
Pump driver
USB A 1
FPGA Pressure sensor
XCS30XL
USB A 2
External keyboard Measure control
DIGIO Display
USB A 3 Assembly
HVB(50V, 150V)
USB A 4
USB stick
DC/DC ±12V
320*240 (QVGA)
USB DIMMPC
TFT modul
HUB AMD Elan SC-520 Start
PIC24 w/ touchscreen
μController To PIC ADC button & w/ LED backlight
Status
LED board
USB B
LCD backlight driver
Sign Collection Board
External PC
BLTS v1.0
Touchscreen
interface
Washing
inlet
Chamber
extender
Counting
chamber Platinum
electrode
Draining
Opening for connection
measuring tube
Aperture
The aperture is made of ruby and it is molded into the end of the measuring tube.
FUNCTIONAL DESCRIPTION
LED TSL235
Connection to
CSA1 on Main
Connection Connection to
to HVB on DIGIO on Main
Main
Amplifier board includes its own voltage regulator, connection interfaces to HGB head
and to MAIN board. There is a current generator circuit on it, which works from 50V
measuring voltage (generated by MAIN) and the probe voltage (DC) is amplified with a
voltage follower (output: ELV). Nominal measuring current is 870 µA.
Amplifier board includes one input connector for the chamber (measuring electrode).
There are two opto switches (U1, U3) to connect high voltage to the probe with HSW
signal and isolate the input of the amplifier. Test circuit makes possible to generate test
pulses (with TEST and PLS signals through FETs) for checking the proper operation of
the amplifier channel.
Amplifier board includes a 3-stage main amplifier channel, which gains input signal to
the 0...3.3 V range (this is the input range of the A/D converter, which is placed on the
MAIN board). The RSW signal changes the gain (RBC, WBC) in the feedback of the
second amplifier stage with U2 (MAX319) analog switch. Amplifier gain and offset are
adjusted by software.
DHON signal switches on the LED and the MVON signal – which is active during counting
– switches off the Photo Detector in the HGB head, to prevent noise generated by the
HGB detector.
The other side of the amplifier board contains special connectors for the chamber and
the HGB head (JP4).
MAIN board is responsible to control the instrument: contains the main power
regulator circuits, valve and motor driver circuits and other connections for the fluidic
and pneumatic system’s parts, responsible for the specific measurement processing
functions.
The central micro-controller with a FPGA and with several other digital chips (buffers,
decoder, multiplexer) handles the pneumatic system, displaying, measurement and data
management.
Power system: filtering the +12V Input and generates +3.3V (FPGA), +5V (Digital
power), +7.5V (Printer power). Filtered +12V is used for the power of motors and
valves.
Motor drivers: 4 power drivers; Horizontal, Vertical/Sample rotor motors and dilutor
motors have separated ribbon cable connections.
Valve driver: consists two 8-bit, powered output shift registers (with built in protection
diodes) and there is one common ribbon cable connection for the valve boards. The
peristaltic pump has a separated power FET driver circuit for more reliable operation.
Measurement processing: the A/D conversion made by the micro-controller itself, but
several preprocessing steps (time limits, noise handling, pulse integration) taken by the
external analog circuitry.
Power Supply
for internal
printer 7.5VDC
USB HUB
TFT backlight driver
RS232/USB Converter
Speaker
FPGA Microcontroller
Power Supply
12VDC-> 5VDC, 3.3VDC
Pressure Sensor
Connector to TFT
Backlight and Start Button
Connectors to Amplifiers
Power Connector
Internal battery
AMD Elan
Edge connector SC520 CPU
Valves
Valve Board
Connection to MAIN
FUNCTIONAL DESCRIPTION
The valve boards are connected to controller and driver chips are located on the MAIN
BOARD.
Touchscreen
connector
TFT connector to
MAIN BOARD
TFT Backlight
connector
230V AC inlet
12V DC outlet
MAIN MECHANIC AND FLUIDIC PARTS
ABACUS JUNIOR 30 Analyzers consist of the following mechanic and fluidic parts:
1. Sample rotor
2. Sampling needle
3. Washing head
4. Sample/Horizontal moving unit
5. Micro Dilutor
6. Dilutor
7. Chamber
8. Cell-counter probe
9. Puffer reservoir
10. Pump
11. Valves
12. Tubing
MAIN MECHANIC AND FLUIDIC PARTS
V2 DilWash
1
3
V4 DilNeedle
2
1
2 1
V3 DilChamber
3
3
2
V5 LyseWbc DILUENT
1 2
3 2 1
Pressure V1 Cleaner
Meter
P
3
V9 DilAperture
1 2
CLEANER
Puffer
RBC
Reservoir
1
WBC
Macro Lyse
3
V8 Bubble
2
Micro dilutor
Dil M3
HGB
2
LYSE V7 DrainAperture
M2
1
Ver
2
V6 DrainChamber
2 1
1
3
V10 DrainPuffer
Sample/
M1 H motor
Pump
2 1 1 2 3-way
Stepper Motor 2-way Valve
M1 Valve
Closed = Off 3
WASTE 1-3 = Off
Open = On
2-3 = On
Vertical motor
Sample/Horizontal
motor (not visible)
Clean diluent
Pump to waste
Pistons
Diluent syringe
Micro dilutor
Lyse syringe
Positioners
In AJ30 this unit includes one dilutor stepper motor. The Micro dilutor syringe makes
the aspirating while the motor moves down. The syringes are mechanically connected
with a loose mechanism, so there is a phase along the track, where the micro dilutor
doesn’t move.
3.1.7. Pump
Pump generates regulated vacuum and drains the fluidic system. It is connected to the
MAIN board and it has its own driver circuit (Power FET).
The pump is maintenance free.
MAIN MECHANIC AND FLUIDIC PARTS
TFT display
Start button
Sample
holder/Sample rotor
USB socket
Power switch
Cover screws
USB A
inlets
USB B
inlet
12VDC power
IN
Reagent Grounding
inlets screw
MAIN MECHANIC AND FLUIDIC PARTS
Valves
MAIN Dilutor
board motor
Vertical
motor
Valves 1-5
Puffer
reservoir
Micro
dilutor
Lyse
Dilutor syrenge
Counting
chamber
Valves 6-10
MAIN MECHANIC AND FLUIDIC PARTS
MAIN board
Dimm-PC
OPERATION OF THE FLUIDIC SYSTEM
In the detailed process description figures, the active tube is filled with black color,
while an arrow ( ) shows the direction of the flow. Moving mechanic parts have another
arrow indicating direction of movement. Only opened (On) valves are mentioned in this
section while all the other valves are closed (Off).
ABACUS JUNIOR 30 employs software counters to estimate waste (and other reagent)
level. Software integrates volume of the reagents used, and gives a message when this
volume reaches the preset tank capacity.
Chamber draining
Sampling process
Needle washing
Diluting process
(1:160)
Sampling process
(from primary
dilution)
Needle washing
Lysing process
OPERATION OF THE FLUIDIC SYSTEM
Generating
measuring
vacuum
Dilution process
(1:25600)
HGB
measurement
Needle washing
Extensive cleaning
Generating
measuring
vacuum
Blank HGB
measurement
RBC
measurement
Draining
END
OPERATION OF THE FLUIDIC SYSTEM
V2 DilWash
1
3
V4 DilNeedle
2
1
2 1
V3 DilChamber
3
3
2
V5 LyseWbc DILUENT
1 2
3 2 1
Pressure V1 Cleaner
Meter
P
3
V9 DilAperture
1 2
CLEANER
Puffer
RBC
Reservoir
1
WBC
Macro Lyse
3
V8 Bubble
2
Micro dilutor
Dil M3
HGB
2
LYSE V7 DrainAperture
M2
1 Ver
2
V6 DrainChamber
2 1
1
V10 DrainPuffer
Sample/
M1 H motor
Pump
2 1 1 2 3-way
Stepper Motor 2-way Valve
M1 Valve
Closed = Off 3
WASTE 1-3 = Off
Open = On
2-3 = On
There is also another sampling process for the second (RBC) dilution, 25 µl of primary
dilution is aspirated by the sampling needle from the chamber but it is kept in the
sampling needle during the WBC measurement and the cleaning process.
OPERATION OF THE FLUIDIC SYSTEM
V2 DilWash
1
3
V4 DilNeedle
2
1
2 1
V3 DilChamber
3
3
2
V5 LyseWbc DILUENT
1 2
3 2 1
Pressure V1 Cleaner
Meter
P
3
V9 DilAperture
1 2
CLEANER
Puffer
RBC
Reservoir
1
WBC
Macro Lyse
3
V8 Bubble
2
Micro dilutor
Dil M3
HGB
2
LYSE V7 DrainAperture
M2
1
Ver
2
V6 DrainChamber
2 1
1
V10 DrainPuffer
Sample/
M1 H motor
Pump
2 1 1 2 3-way
Stepper Motor 2-way Valve
M1 Valve
Closed = Off 3
WASTE 1-3 = Off
Open = On
2-3 = On
The Macro syringe pushes the diluent through V4 (Off), V3 (Off), V2 (On). The Pump
aspirates the diluent from the washing head through V10(On), while the M2 Vertical
motor moves the sampling needle up.
OPERATION OF THE FLUIDIC SYSTEM
V2 DilWash
1
3
V4 DilNeedle
2
1
2 1
V3 DilChamber
3
3
2
V5 LyseWbc DILUENT
1 2
3 2 1
Pressure V1 Cleaner
Meter
P
3
V9 DilAperture
1 2
CLEANER
Puffer
RBC
Reservoir
1
WBC
Macro Lyse
3
V8 Bubble
2
Micro dilutor
Dil M3
HGB
2
LYSE V7 DrainAperture
M2
1
Ver
2
V6 DrainChamber
2 1
1
V10 DrainPuffer
Sample/
M1 H motor
Pump
2 1 1 2 3-way
Stepper Motor 2-way Valve
M1 Valve
Closed = Off 3
WASTE 1-3 = Off
Open = On
2-3 = On
The second sample – 25 µl of primary dilution – is stored in the sampling needle during
the WBC measurement and the cleaning process. The instrument makes the second
(RBC) dilution into the chamber after these processes.
OPERATION OF THE FLUIDIC SYSTEM
V2 DilWash
1
3
V4 DilNeedle
2
1
2 1
V3 DilChamber
3
3
2
V5 LyseWbc DILUENT
1 2
3 2 1
Pressure V1 Cleaner
Meter
P
3
V9 DilAperture
1 2
CLEANER
Puffer
RBC
Reservoir
1
WBC
Macro Lyse
3
V8 Bubble
2
Micro dilutor
Dil M3
HGB
2
LYSE V7 DrainAperture
M2
1
Ver
2
V6 DrainChamber
2 1
1
V10 DrainPuffer
Sample/
M1 H motor
Pump
2 1 1 2 3-way
Stepper Motor 2-way Valve
M1 Valve
Closed = Off 3
WASTE 1-3 = Off
Open = On
2-3 = On
For better mixing the macro syringe pushes some air bubbles (aspirated through the
washing inlet of the chamber and V8) after the lysing process through V4 (Off), V3 (On),
V9 (Off) V8 (On).
OPERATION OF THE FLUIDIC SYSTEM
V2 DilWash
1
3
V4 DilNeedle
2
1
2 1
V3 DilChamber
3
3
2
V5 LyseWbc DILUENT
1 2
3 2 1
Pressure V1 Cleaner
Meter
P
3
V9 DilAperture
1 2
CLEANER
Puffer
RBC
Reservoir
1
WBC
Macro Lyse
3
V8 Bubble
2
Micro dilutor
Dil M3
HGB
2
LYSE V7 DrainAperture
M2
1
Ver
2
V6 DrainChamber
2 1
1
V10 DrainPuffer
Sample/
M1 H motor
Pump
2 1 1 2 3-way
Stepper Motor 2-way Valve
M1 Valve
Closed = Off 3
WASTE 1-3 = Off
Open = On
2-3 = On
For noise prevention there is no mechanical or electronic activity during the counting
process and the door should be closed for better shielding.
OPERATION OF THE FLUIDIC SYSTEM
V2 DilWash
1
3
V4 DilNeedle
2
1
2 1
V3 DilChamber
3
3
2
V5 LyseWbc DILUENT
1 2
3 2 1
Pressure V1 Cleaner
Meter
P
3
V9 DilAperture
1 2
CLEANER
Puffer
RBC
Reservoir
1
WBC
Macro Lyse
3
V8 Bubble
2
Micro dilutor
Dil M3
HGB
2
LYSE V7 DrainAperture
M2
1
Ver
2
V6 DrainChamber
2 1
1
V10 DrainPuffer
Sample/
M1 H motor
Pump
2 1 1 2 3-way
Stepper Motor 2-way Valve
M1 Valve
Closed = Off 3
WASTE 1-3 = Off
Open = On
2-3 = On
5. ADJUSTMENTS
The adjustments below are made in the factory. Readjustment of following parts is
necessary if some components are replaced.
1 - flag
2 – opto-sensor
ADJUSTMENTS
6.2.2. Troubleshooting
With Troubleshooting options provide
tools to test mechanical components.
6.2.3. Stress
In Stress mode, the instrument performs measuring cycles without sample (blank
measurements) continuously. This can be used for burn-in tests, or to check pneumatic
system after changing any main fluidic parts.
CHECKING THE PROPER OPERATION
You can have information about stability, cleanliness, HGB operation, and counting time
stability. Results of the PLT and HGB blank results are displayed in table format.
You can detect any kind of noise, or bubbles in the system if the PLT is not stable low, or
HGB has big variation. To exit from this mode press the Abort button (at the end of a
normal cycle) until the Stress operation is finished.
7. SERVICE OPERATION
7.1. Possible Causes of Noise
Generally high count of any particle - even if you think it should be low, or near zero -
can be caused by NOISE, i.e. something interferes with measurement.
The most important thing in these cases to identify the source of NOISE, otherwise you
cannot protect the system against it.
NOISE can come from has several sources, and the different NOISE sources are added.
Sometimes we have to fight one of them, but sometimes more. Only one of them is
enough to make problem.
mode power supplies, computer monitors, since they are not shielded, centrifuges due
to high switching noise of rotor contacts, etc.), the power of the electromagnetic source,
because if high power is present, maybe relocation does not solve your problems,
sometimes the electric power supply makes the coupling, so UPS solves the problem.
Another source of coupling in external noise can be the reagent tanks and tubes.
Especially radio transmitters can cause problems of radiating so that even the reagents
(diluent) guides in the noise. A metal pack for the diluent tank, then a good earth
grounding of this metal box allows this coupling to disappear forever.
interference with CPU fan or other digital logic traces (CPU fan or other digital
signal radiates to chamber or to the shielded amplifier cable). Try keeping the ribbon
cables far from the chamber and shielded cable.
C. Bad components, or connections:
bad soldering, salt residuals or component failure on amplifier (especially if
some reagent could get in the amplifier section). Cleaning of PCB/electrode socket or
replacement of amplifier is required. Check for the correct soldering of reference
cable and its connector.
circuit board bad soldering or component failure. Check the shielded cable
connections as well. Sometimes inside out connection (hot electrode goes outside as
a shield) is the problem: both ends of amplifier signal cable must be reversed.
analog signal ribbon cable (it picks up noise). Check the ribbon cable between the
circuit board and the amplifier. Maybe it is pinched under some screws or
components. This may cause trouble and even noise.
D. Pneumatic failures, liquid paths that conduct noise into the chamber:
liquid remains under the chamber in drain tube (during measurement the
conducting liquid remains inside the drain tube making noise to appear there).
Check chamber draining path for clogging or salt crystals.
Check the pump operation. Since draining of the chamber goes under
pressure control, maybe a bad pressure sensor or connection can cause
trouble.
Clean the draining path. Do not use alcohol, but bleach. Replace chamber if
necessary.
liquid remains in the washing inlet at top of the chamber (during measurement
the conducting liquid remains inside the chamber wash tube making noise to
appear). The software is not compatible with the mechanics, or related valve is
bad/partly clogged, or the tubing is clogged/loose.
lyse path guides in noise (during counting, if the a liquid in the draining tube is
touching lyse reagent in T-fitting, noise can appear). Check the lyse path, and the lyse
valve as well.
MAINTENANCE
8. MAINTENANCE
8.1. Weekly User Maintenance
Perform weekly maintenance before turning on the power switch. The right side has a
side door giving access to the fluidic system and the mechanical parts easily.
Wash head
Measuring
chamber
1. Exit Measure menu. Open the side door after the needle has stopped moving.
2. Gently rub the lower surface of the washing head with a damp cloth or wiper
to remove the salt build-up.
3. Close the side door.
9. APPENDICES
9.1. Warning flags
Analyzer SW displays warning flags for each individual measurement to notify user
about status of results. The following table summarizes warning flags and gives
explanation of their possible cause and a few hints to overcome the problem.
Uppercase letters refer to WBC or HGB problems.
Flag Meaning Recommended user action
E No WBC 3-part * Possible lyse problem. May occur in pathological lymphocytosis.
differential
H * HGB blank is high, or no
* Repeat the blank measurement. If HGB blank is not stable there are
HGB blank probably bubbles in the WBC chamber: Run a cleaning and try blank
again. Close the side door if open during measurement.
B * WBC blank is high, or no
* Repeat the blank measurement, or run prime lyse and try blank again.
WBC blank * Possible lyse contamination, or noise problem.
M linearity range exceeded
* The analyzer found that the cell count is higher than the linearity range of
in WBC stage the analyzer. Make a pre-dilution, and run the same sample in prediluted
mode
R RBC cells found in * RBC cells were detected during the WBC measurement. Either the lyse
sample during WBC reagent is not effective enough (volume should be increased) or the
stage RBC’s in the sample are somewhat lyse resistive
W WBC 3-part warning Probably large PLTs or clumped PLTs are present in the sample. Usually
caused by the nature of the sample. cat and goat samples tend to clump.
Intensive, but careful mixing of the sample (e.g. Vortex) can help remove
the clumps. If the rerun sample gives the same results, consider that
WBC and NEU values seem higher because of the clumps. Lyse
modification can’t solve the problem.
L RBC-WBC limit warning Typically insufficiently lysed RBC’s interfere with the start of the WBC
histogram. Repeating the measurement with an increased lyse volume
should provide better separation. If the repeated run reports very similar
results then the MON and NEU results are VALID but the WBC and LYM
results may be higher because of interfering RBCs.
C WBC clogging * Aperture clogging. Perform cleaning and repeat the measurement. If it is
a general problem, please contact your Service Personnel.
* Low temperature reagents can cause it as well (mainly diluent), in this
case you will have to wait until they reach room temperature.
The last character of a record is always the single control character „End of
Transmission” (<EOT>, 4, 0x04). There is no terminating „NULL” (<NUL>, 0, 0x00)
character at the end. The next record can start right after the <EOT> character.
As mentioned above, after the last channel value in the PLT histogram the body of the record is closed with the
control character „End of Text” (<ETX>, 3, 0x03).
APPENDICES
MAIN Board
Pressure
Sensor PUFFER
V1 V2 V3 V4 V5
2 2 2 2 2
3 3 3 3
1 1 1 1 1
M D L
70u
Dilutor
1 1 1 3 1 1
3 3
2 2 2 2 2
V6 V7 V8 V9 V10
out in
Pump
D