Instruction Manual: Multi Modality Electrical Stimulator
Instruction Manual: Multi Modality Electrical Stimulator
MULTI MODALITY
ELECTRICAL STIMULATOR
INTERFERENTIAL | TENS | EMS | MICROCURRENT
Table of Contents
1 . Safety information…..……………………………………….. . 4
1.1 General description
1.2 Medical background
1.3 Indication for use
1.4 Contraindications
1.5 Warnings, Cautions, Adverse Reactions
2 . Presentation ……………………………………………...… 13
2.1 Front and Rear panel
2.2 LCD display
3 . Specification………………………………………………… 15
3.1 Accessories
3.2 Technical information
3.3 The waveforms of the stimulation programs
4 . Instruction for use ………………………………………….. 23
4.1 Battery
4.2 Connect electrodes to lead wires
4.3 Connect lead wires to device
4.4 Electrode
4.5 Turn ON
4.6 Select the Therapeutic Mode
4.7 Steps to set a new program
4.8 Adjust Channel Intensity
4.9 Lock the button
4.10 Stop the treatment
4.11 Turn OFF
4.12 Low battery indicator.
5. Program………………………………………………………. 36
6. Cleaning and Care………………………………………….. 37
6.1 Tips for skin care
6.2 Cleaning the device
6.3 Electrodes
6.4 Cleaning the Electrodes cords
6.5 Maintenance
7. Troubleshooting…………………………………………......41
8. Storage……………………………………………………….. 42
9 .Disposal……………………………………………………….43
10. Electromagnetic Compatibility (EMC) Tables……….....43
11. Glossary symbols…………………………………........... 48
12. Warranty…………………………………………………..…49
3
1. Safety information
1.1 General
MT8000 Electrical Stimulator is a portable
electrotherapy device featuring four therapeutic
m o d e s : Tr a n s c u t a n e o u s Electrical Nerve
Stimulator, Electrical Muscle Stimulation
Interferential, and Microcurrent,which are used for
pain relief and electrical muscle stimulation. The
stimulator sends gentle electrical current to
underlying nerves and muscle group via electrodes
applied on the skin. The parameters of device are
controlled by the press buttons. Its intensity level is
adjustable according to the needs of patients.
EXPLANATION OF TENS
Transcutaneous Electrical Nerve Stimulation (TENS)
is a non-invasive, drugfree method of controlling
pain. TENS uses tiny electrical impulses sent through
the skin to nerves to modify your pain perception.
TENS does not cure any physiological problem; it
only helps control the pain. TENS does not work for
everyone; however, in most patients it is effective in
reducing or eliminating the pain, allowing for a return
to normal activity.
EXPLANATION OF EMS
Electrical Muscle Stimulation (EMS) is an
internationally accepted and proven way of treating
muscular injuries. It works by sending electronic
pulses to the muscle needing treatment; this causes
4
the muscle to exercise passively. It is a product
derived from the square waveform (ladder-sha
ped). Through the square wave pattern it is able to
work dire ctly on muscle motor neurons. This device
has low frequency and this in conjunction with the
square wave pattern allows direct work on muscle
groupings. This is being widely used in hospitals and
sports clinics for the reatment of muscular injuries
and for the re-education of paralyzed muscles, to
prevent atrophy in affected muscles and improving
muscle tone and blood circulation.
EXPLANATION OF IF
Interferential Stimulation (IF) is an anti-inflammatory
based treatment modality. Interferential stimulation is
characterized by two alternating-current sine waves
or square waves of differing frequencies that “work”
together to produce an interferential current that is
also known as a beat pulse or alternating modulation
frequency. One of the two currents is usually held at
4,000 Hz, and the other can be held constant or
varied over a range of 4,001 to 4,100 Hz. Because of
the frequency, the interferential wave meets low
impedance when crossing the skin to enter deep into
soft tissues. The interferential currents reportedly
can stimulate sensory, motor, and pain fibers. These
large impulse fibers interfere with the transmission of
pain messages at the spinal cord level. This deep
tissue penetration stimulates parasympathetic nerve
fibers for increased blood flow and edema reduction.
It utilizes the low electric-current to stimulate muscle
nerves to achieve the symptomatic relief of chronic
intractable pain, post-traumatic pain, and post-
surgical pain.
5
EXPLANATION OF MICROCURRENT
Microcurrent stimulation is a type of therapy in which
very low current is sent into the cells of the body.
Microcurrent is a very faint current that is so small it is
measured in millionths of an amp (Microamps).
Human cells generate a current that is in the
microamp range which is why you can't feel it-the
current is so low it doesn't stimulate the sensory
nerves.
Microcurrent is a physiological electric modality that
increases ATP (energy) production in the cells of your
body. This dramatically increases the tissue's healing
rate. The immediate response to the correct
microcurrent frequency suggests that other
mechanisms are involved as well. The exact effects
or changes in the tissue are unmistakable; scars will
often suddenly soften; trigger points often become
less painful within minutes when the “correct”
frequency is applied. In many situations the changes
seen seem to be long lasting and in many cases
permanent.
6
IMPORTANT SAFETY INFORMATION!
1.4 Contraindications
7
1.5 Warnings, Cautions and Adverse Reactions
WARNINGS:
8
11) Do not use the stimulator in the vicinity of
shortwave or microwave therapy equipment, since
this may affect the output power of the stimulator.
12) Never use in environments with high humidity such
as in the bathroom or when having a bath or shower.
13) Caution should be used in applying electrical
stimulation to patients suspected of having heart
disease. Further clinical data is needed to show
there are no adverse results.
14) Never use near the heart. Stimulation electrodes
should never be placed anywhere on the front of
the thorax (marked by ribs and breastbone), but
above all not on the two large pectoral muscles.
Here it can increase the risk of ventricular
fibrillation and lead to cardiac arrest.
15) Electrodes should not be placed over the eyes, in
the mouth, near the genitals or internally.
16) Never use on the areas of the skin which lack
normal sensation
17) Apply the electrodes to clean, dry, and unbroken
skin only.
18) Keep electrodes separate during treatment,
electrodes in contact with other could result in
improper stimulation or skin burns.
19) Keep the stimulator out of reach of children.
20) Consult your doctor if you are in any doubt
whatsoever.
CAUTIONS:
9
4) This stimulator not intended for unattended,
personal use by patients who have noncompliant,
emotionally disturbed, dementia, or low IQ.
5) Read, understand, and practice the warnings,
cautions and operating instructions. Know the
limitations and hazards associated with using any
device. Observe the precautionary and operational
decals placed on the unit. Always follow the
operating instructions prescr ibed by your healthcare
practitioner.
6) The instruction of use was listed; any improper use
may be dangerous.
7) Do not use this device for undiagnosed pain
syndromes until consulting a physician.
8) Patients with an implanted electronic device, such
as a cardiac pacemaker, implanted defibrillator, or
any other metallic or electronic device should not
use this device without first consulting a doctor.
9) Stimulation delivered by this device may be
sufficient to cause electrocution. Electrical current
of this magnitude must not flow through the thorax
or across the chest because it may cause a cardiac
arrhythmia.
10) Do not place electrodes on the front of the throat as
spasm of the Laryngeal and Pharyngeal muscle may
occur. Stimulation over the carotid sinus (neck
region) may close the airways, make breathing
difficult, and may have adverse effects on the heart
rhythm or blood pressure.
11) Do not place electrodes on your head or at any sites
that may cause the electrical current to flow
transcerebrally (through the head).
12) Patients with heart disease, epilepsy, cancer or any
other health condition should not use this device
without first consulting a physician.
10
13) Some patients may experience skin irritation or
hypersensitivity due to the electrical stimulation or
silicone rubber. If rash develops or pain persists,
discontinue use and consult a doctor.
14) Electrode placement and stimulation settings
should be based on the guidance of prescribing
practitioner.
15) Effectiveness is highly dependent upon patient
selection by a person qualified in the management
of pain afflicted patients.
16) Isolated cases of skin irritation may occur at the
site of the electrode placement following long-term
application. If this occurs, discontinue use and
consult your physician.
17) The electrodes are only to be placed on healthy skin.
Avoid skin irritation by ensuring that good contact is
achieved between electrodes and skin.
18) If the stimulation levels are uncomfortable or
become uncomfortable, reduce the stimulation
Intensity to a comfortable level and contact your
physician if problems persist.
19) This device should not be used while driving,
operating machinery, close to water, or during any
activity in which involuntary muscle contractions
may put the user at undue risk of injury.
20) Never use the device in rooms where aerosols
(sprays) are used or pure oxygen is being
administered.
21) Do not use it near any highly flammable substances,
gases or explosives.
22) Do not use this device at the same time as other
equipment which sends electrical pulses to your
body.
23) Do not confuse the electrode cables and contacts
with your headphones or other devices, and do no t
connect the electrodes to other devices .
11
24) Do not use sharp objects such as pencil point or
ballpoint pen to operate the buttons on the control
panel.
25) Inspect Applicator cables and associated
connectors before each use.
26) Turn the device off before applying or removing
electrodes.
27) Electrical stimulators should be used only with the
leads and electrodes recommended for use by the
manufacturer
28) This device has no AP/APG protection. Do not use
it in the presence of explosive atmosphere and
flammable mixture.
Adverse Reactions:
12
2. Presentation
13
7) Parameter Selection [S]: press the button to
enter setting state; you can select the difference
parameters in conjunction with [▲] and [▼].
8) Turn OFF/ON: press the [ ]button to turn on th e
device or keep [ ] button for approx.3 seconds
to turn off the device.
9) Output socket: electric signal output afte r
connection of the cable with adhesive electrodes
channel 1
10) Output socket: electric signal output after
connection of the cable with adhesive electrodes
channel 2
11) Belt Clip
12) The battery compartment cover for opening
13) Adapter Receptacle
3. Specification
3.1 Accessories
No DESCRIPTION Q’TY
1 Electrical stimulator device 1 piece
2 Electrodes Leads 2 pieces
3 40*40 mm Adhesive Electrodes 4 pieces
4 9V Battery, type 6LR61 1 piece
5 Instruction Manual 1 piece
6 Carrying case 1 piece
15
3.2 Technical information
Channel Dual, isolated between channels
9.0 V DC -1 *6LR61 battery
Power supply
Adapter output:9.0Vdc, 800mA
5°C to 40°C( 41℉ to 104℉) with a relative
Operating humidity of 30%-75%, atmospheric
conditions pressure from 700 to 1060 Hpa
16
The pulse rate and pulse width are
Normal (P2) adjustable. It generates continuous
stimulation based on the setting value.
The pulse width is automatically varied in
a cycle time. The pulse width is decreased
from its original setting to 60% in setting
Pulse Width
cycle time, and then increased from 60%
Modulation
to its original setting in nest setting cycle
(P3)
time. In this program, pulse rate (1 to
150Hz), pulse width (50 to 300us) and
cycle time (5 to 30 sec) are fully adjustable.
The pulse rate is automatically varied in a
cycle time. The pulse rate is decreased
from its original setting to 60% in setting
Pulse Rate
cycle time, and then increased from 60%
Modulation
to its original setting in nest setting cycle
(P4)
time. In this program, pulse rate (1 to
150Hz), pulse width(50 to 300us) and
cycle time (5 to 30 sec) are fully adjustable.
17
Stimulation of both channels occurs
synchronously. The “ON” time including
Synchronous
“Contraction”, “Ramp Up” and “Ramp
(P1)
Down” time.ON TIME=Contraction +
Ramp up + Ramp down
The Stimulation of the CH2 will occur after
the 1st working of CH1 is completed. In
Alternate (P2) this program, The “ON” time including
“Contraction”, “Ramp Up” and “Ramp
Down” time. The OFF Time should be
equal or more than the ON Time
ON TIME=Contraction + Ramp up + Ramp
down OFF TIME≥ON TIME
The Stimulation of the CH2 will occur after
the 1st working of CH1 is started+ Delay
Time. Delay time is adjustable form 1 to
Delay (P3) 10 sec. In this program, The “ON” time
including “Contraction”, “Ramp Up” and
“Ramp Down” time.
ON TIME=Contraction + Ramp up + Ramp
down
18
The pulse rate of the CH1 is fixed in
4000Hz; CH2 pulse rate is increased from
4001Hz to 4010Hz in a cycle time, and
then decreased from 4010Hz to 4001Hz
P1 in nest setting cycle time. In this program,
CH2 interference frequency is varied from
1Hz to 10Hz, cycle time (5 to 30 sec) is
fully adjustable. CH 2 pulse rate=4000Hz+
Interference frequency
The pulse rate of the CH1 is fixed in
4000Hz; CH2 pulse rate is increased from
4001Hz to 4150Hz in a cycle time, and
then decreased from 4150Hz to 4001Hz
P2 in nest setting cycle time. In this program,
CH2 interference frequency is varied from
1Hz to 150Hz, cycle time (5 to 30 sec) is
fully adjustable. CH 2 pulse rate=4000Hz+
Interference frequency
The pulse rate of the CH1 is fixed in
4000Hz; CH2 pulse rate is increased from
4080Hz to 4150Hz in a cycle time, and
then decreased from 4150Hz to 4080Hz
P3 in nest setting cycle time. In this program,
CH2 interference frequency is varied from
80Hz to 150Hz, cycle time (5 to 30 sec) is
fully adjustable. CH 2 pulse rate=4000Hz+
Interference frequency
The pulse rate of the CH1 is fixed in
4000Hz; CH2 pulse rate is automatically
varied in a cycle time. Interference
frequency is increased from its original
setting to 60% in setting cycle time, and t
P4 hen decreased from 60% to its original
setting in nest setting cycle time. In this
program, CH2 interference frequency (2
to 150Hz) and cycle time (5 to 30 sec) are
fully adjustable. CH 2 pulse rate=4000Hz+
Interferencefrequency
19
Technical specifications for Microcurrent (MIC)
mode
20
3.3 The waveforms of the stimulation programs
Burst
Normal
Synchronous
CH1
CH2
21
Alternate
CH1
CH2
Delay
CH1
CH2
Interferential
Microcurrent(Constant) (P1)
22
4. Instruction for use
4 .1 Battery
23
Caution:
1) Battery may be fatal if swallowed. Therefore,
keep the battery and the product out of the
range of children, if a battery was swallowed,
consult a physician immediately.
2) If a battery has leaked, avoid contact with skin,
eyes and mucus membranes, Rinse the affected
spots with lots of clear water immediately and
contact a physician right away.
3) Battery may not be charged, dismantled, thrown
into fire or short-circuited.
4) Protect battery from excess heat; Take the
battery out of the product if they are spent or in
case you no longer use the article. This prevents
damage caused by leaking battery.
5) Always replace the same type battery.
Caution:
Do not insert the plug of the patient lead wire into any AC
power supply socket.
4.4 Electrode
25
Make sure the electrodes are placed firmly to the skin
and make good contact between the skin and the
electrodes. Place the electrodes over the skin; attach
them properly, firmly, and evenly.
Caution:
1) Before applying the self-adhesive electrodes, it
is recommended to wash and degrease the skin,
and then dry it.
2) Do not turns on the device when the self-
adhesive electrodes are not positioned on the
body.
3) Never remove the self-adhesive electrodes from
the skin while the device is still turns on.
4) It is recommended that, at minimum, 4cm*4cm
self-adhering based, square electrodes are used
at the treatment area.
26
4.5 Turn on
Before using the device for the
first time, you are strongly
advised to take careful note of
the counter-indications and
safety measures detailed at
the beginning of this manual
(Safety information), as this
powerful equipment is neither
a toy nor a gadget!
In order to turn on the device, keep the [ ] button
pressed down until the operation page appears on the
screen.
Caution:
Consult your physician for your suitable
therapeutic mode
27
1) Set the Therapeutic Program
There are 4 programs in TENS
therapeutic mode available –
Burst (P1), Normal (P2), Pulse
Width Modulation (P3), and
Pulse Rate Modulation (P4). The
therapeutic program can be
selected by pressing the [▲ ] and [▼ ] button. When
you choose to [B] program, program [B] outside of the
box will be flashing.
3) Set Timer
Press [S] button cycle to enter this setting. The
treatment time is adjustable from 1 to 60 minutes or
Continue. Press [▲ ] or [▼ ] button control to adjust
setting. You can set the timer to “Continuous” mode
by pressing the [▲] control when it shows 60 minutes.
Its output will be shut off when time is up.
28
4.7.2 IF Setting
Press the [S] button to enter the setting state. The
settings can be adjusted according to the following
steps:
29
4) Set Cycle Timer
Cycle time is adjustable form 5
to 30 seconds. Press [S] button
cycle to enter this menu, and
then press the [▲ ] and [▼ ]
button to adjusting the setting.
30
4) Set Pulse Rate
Pulse rate is adjustable from 1 Hz to 150 Hz. Press
[S] button cycle to enter this menu, and then press [▲]
or [▼] button to adjust the setting.
2) Set Timer
Press [S] button cycle to enter this setting. The
treatment time is adjustable from 1 to 60 minutes or
Continue. Press [▲] or [▼] control to adjust setting.
You can set the timer to “Continuous” mode by
31
pressing the [▲ ] button when it shows 60 minutes.
Its output will be shut off when time is up.
32
7) Set Contract Time
The contract Time controls the
time of stimulation. The
contraction time can be
adjusted. Press [S] button
cycle to enter this menu, and
then press the [▲ ] and [▼ ]
button to adjusting the setting.
Both channels’ stimulation is cycled on and off by the
contraction and relaxation settings. The range is
adjustable from 1 to 30 seconds.
Caution:
Contract time has not including the ramp up and
ramp down time, ON time=Ramp up + Contract time
+ Ramp down.
33
Caution:
1) If the stimulation levels are uncomfortable or
become uncomfortable, reduce the stimulation
intensity to a comfortable level and contact
your medical practitioner if problems persist.
2) TENS, EMS and IF therapeutic mode. If the
electrodes no placed firmly on skin or the
device has not connected on the electrodes,
the stimulator's output intensity surpasses
12mA, the intensity will enulls automatically.
Caution:
Default state, if the button is locked, you can
press only one of the [ ▼ ]buttons to unlock, and
then press the [M] button or the [▼ ] button to
control stop the treatment.
4.11. Turn OFF
Press [ ] button and hold for approx.3 seconds to
Turn OFF the device.
34
Caution:
1) If there is no operation in the panel for 3
minutes in the waiting state, the device will be
turns off automatically.
2) In shutdown state, keep pressing the channel 2
[ ▼] first, and then press [ ]button at the same
to restore factory parameter settings
35
5. Program
Modulation Pulse
Mode Program Frequency Treat time
Method Width
Pulse width
TENS P3 1-150Hz 50-300us 1-60min,continous
modulation
Frequency
P4 1-150Hz 50-300us 1-60min,continous
modulation
Frequency 4kHz
P1 125us 1-60min,continous
modulation 4001-4010Hz
Frequency 4kHz
P2 125us 1-60min,continous
modulation 4001-4150Hz
IF
Frequency 4kHz
P3 125us 1-60min,continous
modulation 4080-4150Hz
Frequency 4kHz
P4 125us 1-60min,continous
modulation 4001-4150Hz
Pulse width
P2 1-150Hz 2-200ms 1-60min,continous
MIC modulation
Frequency
P3 1-150Hz 2-200ms 1-60min,continous
modulation
36
6. Cleaning and Care
37
6.2 Cleaning the device
1) Remove the battery from the device every time
when you clean.
2) Clean the device after use with a soft, slight
moistened cloth. In case of more extreme soiling
you can also moisten the cloth with mild soapy
water.
3) Do not use any chemical cleaners or abrasive
agents for cleaning.
6.3 electrodes
1) Use the device only with the leads and electrodes
provided by the manufacturer. Use only the
electrode placements and stimulation settings
prescribed by your physician or therapist.
2) It is recommended that, at minimum, 4cm*4cm self
-adhering based, square electrodes are used at the
treatment area.
3) Inspect your electrodes before every use. Replace
electrodes as needed. Reusable electrodes may
cause slight skin irritation, lose adhesion and
deliver less stimulation if overused.
38
To use these electrodes:
1) Attach the electrode to the lead wire.
2) Remove the protective backing from the electrode
surface. Do not throw away the protective backing
because it is reused after the treatment session
has been completed.
3) Place the tacky surface to the prescribed skin area
by pressing the electrode firmly against the skin.
Caution:
Do not pull on the electrode wire. Doing so may
damage the wire and electrode.
Do not apply to broken skin.
The electrodes should be discarded when they
are no longer adhering.
The electrodes are intended for single patient
use only.
If irritation occurs, discontinue use and consult
your clinician.
Read the instructions for use of self-adhesive
electrodes before application.
39
Always use the electrodes with the requirements
of the IEC/EN60601-1, ISO10993-1/-5/-10 and I
EC/ EN60601-1-2, such as with CE mark, or are
legally marketed in the US under 510(K)
procedure.
6.5 Maintenance
1) Maintenance and all repairs should only be carried
out by an authorized agency. The manufacturer will
not be held responsible for the results of
maintenance or repairs by unauthorized persons.
2) The user must not attempt any repairs to the device
or any of its accessories. Please contact the
retailer for repair.
3) Opening of the equipment by unauthorized
agencies is not allowed and will terminate any
claim to warranty.
4) Check the unit before each use for signs of wear
and/or damage. Replace wear items as required.
40
7. Troubleshooting
If your device does not seem to be operating correctly,
refer to the chart below to determine what may be
wrong. Should none of these measures correct the
problem, the device should be serviced.
Electrodes
Stimulation 1. Dried out or Replace and re-connect
weak contaminated
2. Placement
Lead wires Replace
1.Old/worn/
damaged
Intensity is too
high Decrease intensity.
Stimulation Electrodes are Reposition the electrodes.
is too close
uncomfortable. together Replace.
Damaged
or worn Replace electrodes with
electrodes or ones that have an active
lead wires area no less than 16.0cm
Electrode active 2(4cm*4cm).
area size is too
small.
41
Intermittent Lead wires 1. Verify connection is
output secure. Insure firmly.
2. Turn down the intensity.
Rotate lead wires in socket
90°. If still intermittent,
replace lead wire.
3. If still intermittent after
replacing lead wire, a
component may have failed.
Call the repair department.
Improper
Stimulation electrode and Reposition electrode
is applicator and applicator
ineffective. placement Contact clinician.
Unknown
8. Storage
1) For a prolonged pause in treatment, store the
device in a dry room and protect it against heat,
sunshine and moisture.
2) Store the device in a cool, well-ventilated place
3) Never place any heavy objects on the device.
42
9. Disposal
Used fully discharged batteries must be disposed of in
a specially labeled collection container, at toxic waste
collection points or through an electrical retailer. You
are under legal obligation to dispose of batteries
correctly.
43
RF
emissions
CISPR11 Class B
The device is suitable for use in
all establishments other than
domestic and those directly
Harmonic connected to the public low-
emissions Not voltage power supply network
lEC applicable that supplies buildings used for
61000-3-2 domestic purposes.
Voltage
fluctuations
Not
/flicker
applicable
emissions
lEC
61000-3-3
Compliance Electromagnetic
Immunity IEC 60601 environment -
test test level level
guidance
Floors should be
wood, concrete or
Electrostatic ± 6 kV ceramic tile. If floors
discharge ± 6 kV
contact are covered with
(ESD) EC contact
synthetic material,
61000-4-2 therelative humidity
±8 kV air ±8 kV air
should be at least
30 %.
44
Guidance and- manufacturer's declaration.
Electromagnetic immunity
The device is intended for use in. the electromagnetic
environment specified below. The customer or the user
should assure that it is used in such an environment.
, 800MHz
to 2,5MHz
, 80MHz
to 800MHz
45
Where P is the
maximum output
power rating of the
transmitter In watts
(W) according to the.
Transmitter
manufacturer and
d Is the recommended
separation distance
in meters (m).
46
theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the
location in which the device is used exceeds the
applicable RF compliance level above, should be
observed to verify normal operation. If abnormal
performance is observed, additional measures may
be necessary, such as reorienting or relocating the
device.
47
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not
listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable
to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts
(W) accordable to the transmitter manufacturer.
NOTE I At 80 MHz and 800 MHz. the separation distance
for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Batch code
Serial number
Electrical devices are recyclable material and
should not be disposed of with household
waste after their useful life! Help us to protect
the environment and save resources and take
this device to the appropriate collection points.
Please contact the organization which is
responsible for waste disposal in your area
if you have any questions.
Attention: Read the operating instruction
for use!
48
12. Warranty
Please contact your dealer or the device centre in
case of a claim under the warranty. If you have to send
in the unit, enclose a copy of your receipt and state
what the defect is.