Yes
Yes
MEASUREMENTS
1
INSTRUMENTATION INSTRUMENT
2
THREE GENERAL TYPES OF
INSTRUMENTS
1. Indicating
2. Recording
3. Controlling
3
MEASUREMENT
4
MEASUREMENT STANDARDS
1. International Standards - maintained at the
International Bureau of Weight and Measures in Paris
and are periodically evaluated and checked by absolute
measurements in terms of the fundamental units of
physics.
2. Primary Standards - maintained at national standards
laboratories in different countries. These are not
available for use outside the national laboratories. The
function is the calibration and verification of secondary
standards.
5
MEASUREMENT STANDARDS
3. Secondary Standards - the basic reference standards
used by measurement and calibration in the industry.
Each industrial laboratory is responsible for its own
secondary standards, which are calibrated and certified
at national standards.
4. Working Standards - the principal tools of a
measurement laboratory. They are used to check and
calibrate general laboratory instruments or to make
comparison measurements in industrial application.
6
ERROR
7
CATEGORIES OF
MEASUREMENT ERRORS
1. Gross Errors - generally the fault of the person using
the instrument
2. Systematic Errors - caused by problems with the
instrument, environmental effects, or observational
errors (parallax and estimation)
3. Random Errors - the accumulation of a large number
of small effects which are neither classified as gross nor
systematic.
8
WAYS OF EXPRESSING
ERRORS
ABSOLUTE ERROR (E) PERCENTAGE ERROR (%E)
9
ACCURACY
10
PRECISION (P)
(MEASUREMENT)
• Is a quantitative or numerical indication of the closeness
with which a repeated set of measurements of the same
variable agrees with the average of the set of
measurements.
12
PERMANENT MAGNET
MOVING COIL (PMMC)
• This movement operates upon the interaction of the magnetic fields
between a permanent magnet and an electromagnet. The amount of
movement is determined by the size of the current. By attaching a
needle to the coil, an indication of the size of the current is obtained.
A needle deflection is obtained corresponding to the amount of
current through the ammeter. To reduce friction, the coil mechanism
is set in a jewel and pivot suspension system. Taut-band system is
another method of suspension method that provides more sensitive
but more expensive meter movement. Typical full-scale current is
50µA while for taut-band is 2µA. The most popular deflecting
movement used in instrumentation is the D' 'Arsonval movement.
13
PERMANENT MAGNET
MOVING COIL (PMMC)
• Construction
14
PMMC FEATURES:
• COUNTERWEIGHT - attached to the pointer tail to
provide mechanical balance and eliminate gravitational effect on
the pointer.
• SPIRAL SPRING - to control the force exerted on the coil
when current passes through it.
• MOVING COIL - pivoted so as to move freely when current
is introduced to the system.
• POLE SHOES - used to fix the position of the permanent
magnet.
• ZERO ADJUSTABLE CONTROL - used to adjust the
pointer to its initial position or zero position when no current
is flowing POINTER - provide visual indication of the amount
of current flow on the system.
15
FORCES ACTIVE IN THE
PMMC
1. Deflecting Force — provided by the moving coil
pivoted so as to move freely when current flows to the
system
2. Controlling Force — provided by the spiral spnng so
that the pointer will deflect back to its initial position
when no current flows through the system
3. Damping Force — provided by the coil pointer to
eliminate pointer oscillation
16
DC AMMETER
17
PROPERTIES OF PROPERTIES OF
IDEAL AMMETER REAL AMMETER
1. The internal 1. It has some
resistance is zero resistance but made
2. The needle as low as possible
deflection is directly 2. The needle
proportional to
deflection is not
current
exactly proportional
to the amount of
current
18
ERRORS INTRODUCED BY
REAL AMMETER:
• Calibration Error (Cep) - it arises from the fact that the meter
face may not be accurately marked.
– The specification for this error is in terms of full- scale current, the
greatest inaccuracy will occur for small deflection. For instance, for a 1-
mA movement the calibration error would typically be about -0.03 to
+0.03 mA
• Insertion Error (caused by its internal resistance) — the real
ammeter resistance will add the resistance to the branch that will
reduce the current in any actual circuit.
19
• How much reduction takes place if considering the circuit below?
20
Note: A reading which is 100
percent accurate means that
the insertion error is 0
percent. A 99 percent
accurate reading means that
the insertion error is 1
percent.
21
SAMPLE PROBLEMS
1. An ammeter with a full- scale deflection of 1 mA has a
calibration error of ±3 percent of the full - scale current.
What is the possible range in the true current through the
ammeter if the ammeter reads 0.35 mA?
2. Determine the percentage error of reading of an ammeter
due to ammeter insertion. Ammeter parameters include
100ꭥ meter resistance and a load resistor of 1.5 k ꭥ.
3. In the given figure, assuming no calibration error, what does
the ammeter reads if the ammeter has a resistance of 150 ꭥ?
If the ammeter is ideal, what will it read?
22
AMMETER SHUNT
• An ammeter with a full — scale current of IFS can be shunted by a
resistor in order to measure current greater than IFS.
Note: If the total current is greater than IFS by a given factor, the
resistance of the shunted ammeter will be decreased by that factor.
23
PROBLEM
24
MULTI RANGE AMMETER
1. SWITCHED SHUNT
25
MULTI RANGE AMMETER
26
MULTI RANGE AMMETER
2. AYRTON SHUNT
27
DC VOLTMETER
28
DC VOLTMETER
29
DC VOLTMETER
• Sensitivity (Sdc) is the reciprocal of the full - scale of the
ammeter used in the voltmeter.
30
PROBLEM
31
DC VOLTMETER
32
DC VOLTMETER
33
PROBLEM
34
MULTI-RANGE VOLTMETER
1. Switched Multiplier Resistance
35
MULTI-RANGE VOLTMETER
2. Series Connected Multiplier Resistance
36
OHMMETER
• An ohmmeter is an instrument used to measure the
resistance across two points in an electric circuit. A simple
ohmmeter can be made by using a battery, an ammeter
and a resistor.
37
OHMMETER
• To measure the unknown resistance (Ru), connect it to
the terminals X and Y
38
OHMMETER
• Deflection Factor (D) equation is used to mark the
ohmmeter face
39
PROBLEM
40
AC VOLTMETER
• It consists of a rectifier and DC voltmeter to measure AC
voltage
1. Using Half-Wave Rectifier
41
PROBLEM
42
AC VOLTMETER
2. Using Full-Wave Rectifier
A. Using Center-Tapped Transformer
43
AC VOLTMETER
2. Using Full-Wave Rectifier
B. Using Bridged Type
44
PEAK DETECTION
• Responds to the peak of the input signal.
45
PEAK TO PEAK DETECTOR
• Consists of a clamping circuit followed by a peak detector.
46
DC BRIDGE CIRCUIT
• generally refers to a circuit in which a load is connected
between two levels of potential.
48
DC BRIDGE CIRCUIT
INSTRUMENTS
WHEATSTONE BRIDGE
49
DC BRIDGE CIRCUIT
INSTRUMENTS
SLIDE WIRE
• is one of the applications of the Wheatstone bridge.
Operation: The Bridge is
balanced by moving the key
along the wire until the
galvanometer reads zero. The
resistance of the unknown
resistor is determined from
the physical position of the
key on the scale.
50
DC BRIDGE CIRCUIT
INSTRUMENTS
KELVIN BRIDGE
• The Kelvin Bridge is a modified version of the Wheatstone
bridge eliminating the effects of contact and lead
resistance. When measuring unknown low resistances that
are from 1ꭥ to about 1µꭥ thus providing a high degree of
accuracy.
DC BRIDGE CIRCUIT
INSTRUMENTS
THE VARLEY LOOP
• The Vartey Loop is an application of the Wheatstone
bridge used to find the distance of a ground fault in
transmission line cables.
DC BRIDGE CIRCUIT INSTRUMENTS
THE VARLEY LOOP
Operation
Step 1: The SPDT switch is moved to position 1 and the
calibrating rheostat is adjusted to obtain a balance. The
resistance (C1) is used to obtain the resistance of both
cables.
69
ATTENUATORS
• ATTENUATORS are used to reduce the signal level by
a given amount. The most basic form of attenuator is
voltage divider attenuator circuit.
70
ATTENUATORS
The problem of voltage divider attenuator circuit is the
presence of the stray capacitance (C2) across R2 that is
"frequency limiting effect".
72
IMPORTANT PARAMETERS OF ATTENUATOR
1. Attenuation factor (af) — the ratio between input
signal and output signal of the attenuator or reciproca:
only of gain of the amplifier.
73
PROBLEM
74
SYMMETRICAL T ATTENUATOR
SYMMETRICAL Π ATTENUATOR
PROBLEM
77
BRIDGED T ATTENUATOR
• In reference to symmetrical T attenuator, there is an
additional resistor that bridges from the input to the
output.
VARIABLE ATTENUATOR
• It allows selection of different values of attenuation, while
maintaining the characteristic resistance at the value of
R1.
THANK YOU!
GOOD LUCK AND GOD BLESS!
80