v10x e Manual Txblock Transmitter Usb 4-20ma English A4
v10x e Manual Txblock Transmitter Usb 4-20ma English A4
During the setup, the transmitter is powered by the USB, not 4. Sensor Failure: It establishes the output behavior, when the
requiring an external power supply. transmitter indicates a failure:
The transmitter setup can also be made by connecting it to the loop, Minimum: output current goes to < 3.6 mA (down-scale), typically
using the loop power supply. There is no electrical insulation between used for refrigeration.
the transmitter and the communication port (interface), therefore it is Maximum: output current goes to > 22.0 mA (up-scale), typically
not recommended to configure it with the sensor inlet connected to the used for heating.
process. See Fig. 2.
5. Zero Correction: It corrects small deviations presented in the
transmitter output, for example, when the sensor is replaced.
6. Send Configuration: It applies the new setup. Once sent, the
setup will be immediately adopted by the transmitter.
7. Read Configuration: Reads the current setup in the transmitter
connected. The screen now presents the current setup that may
be changed by the user.
DIMENSIONS:
ELECTRICAL INSTALLATION
• Section of the cable used: 0.14 to 1.5 mm ²
Fig. 3 – TxConfig II software main screen
• Recommended torque in the terminal: 0.8 Nm.
The fields in the screen mean:
1. General Information: This field shows information that identifying RECOMMENDATIONS FOR INSTALLATION
the transmitter. This information should be sent to the
manufacturer in an eventual request for technical assistance. • Sensor signals conductors must go through the plant system
separate from power leads (loop), if possible in grounded
2. Sensor: Select the type of sensor to be used. See Table 1.
conduits.
3. Measuring Range: Sets de measurement range of the
transmitter. • The instruments must be powered from the instrumentation power
supply circuit.
Lower Range Limit: equivalent temperature for a current of 4
mA. • In control and monitoring applications is essential to consider
what can happen when any part of the system fails.
Upper Range Limit: equivalent temperature for a current of 20
mA. • It is recommended the use of suppressors in contact coils,
solenoids and any inductive load.
Sensor Range
The values chosen cannot exceed the range of sensor shown in
this field. See Table 1 of this manual.
Minimum Range
Do not set a lower band (span) that the Minimum Range
indicated below in this same field. See Table 1 of this manual.
ELECTRICAL CONNECTIONS wire resistance is 25 Ω per wire leg. Usage of a 3 or 4 wire with
conductors of equal length and gauge is recommended.
The figures below show the electrical connections required. The
terminals 3, 4, 5 and 6 are dedicated to the sensor connection.
LOAD represents the 4-20 mA current measuring device (indicator, THERMOCOUPLES
controller, recorder, etc.).
PT100 2-WIRE
Note: When the Pt100 2-wire the terminals 3 and 4 must be
interconnected, according to the figure below.
The Pt100 wire length should be less than 30 cm to maintain the
measurement error within specifications (electrical resistance).
PT100 3-WIRE
OPERATION
The sensor offset can be changed through the TxConfig II software.
The USB cable may be connected to the transmitter without causing
any measurement errors. See item Zero Correction in the chapter
CONFIGURATION of this manual.
Fig. 6 – Transmitter electrical connections (Pt100 3-wire)
The uer must choose the most suitable sensor and range to the
process. The chosen range must not exceed the maximum range of
PT100 4-WIRE measurement defined for the sensor and should not be smaller than
the minimum range for the same sensor.
It is important to note that the transmitter accuracy is based on the
maximum range of the sensor used, even when a narrower range is
programmed.
Example:
• The Pt100 sensor in the range 0 to 100 °C and accuracy of
0.12 %, the maximum error will be 1.02 °C (0.12 % de 850 °C).
• The Pt100 sensor in the range 500 to 600 °C and accuracy of
0.19 %, the maximum error will be 1.61 °C (0.19 % of the 850
°C).
Note: When measurements are made at the transmitter, see if the
Pt100 excitation current required by the calibrator is compatible with
the Pt100 excitation current used in the transmitter: 0.8 mA.
SAFETY INFORMATION
Any control system design should take into account that any part of
the system has the potential to fail. This product is not a protection or
safety device and its alarms are not intended to protect against
product failures. Independent safety devices should be always
provided if personnel or property are at risk.
Product performance and specifications may be affected by its
environment and installation. It’s user’s responsibility to assure
proper grounding, shielding, cable routing and electrical noise
filtering, in accordance with local regulations, EMC standards and
good installation practices.