Especificaciones Tomografía
Especificaciones Tomografía
16 Slice Edition
Product Data Sheet – Dec. 2008
Allows the operator to recon once (instead of twice – lung kernel and
std kernel separately), and can diagnose the Lung or Mediastinum area
by adjusting the WW/WL. The new Chest Kernel provides up to 2-times
productivity and 50% HDD space saving for Chest exams.
The BrightSpeed Elite Select brings you multi-detector CT • 3D Dose modulation. Before the scan, clinicians can select the
capabilities with maximum convenience in an ultra-compact desired Noise/IQ: CT then tailored automatically exposure
design. Built with LightSpeed VCT inside, the BrightSpeed Elite parameters, patient to patient and real-time x-y-z during each scan,
Select provides high quality image across a wide range of resulting in average of 40% dose reduction.
applications and allows you to perform new clinical applications.
Page 1 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
New Xtream Injector: Xtream Injector is a powerful integrated • 2xLCD monitor 19”inch, standard, for maximum comfort in images
injection option which starts the Injection process in review.
synchronization with “Start Scan” on CT system to simplify the
enhancement exam workflow. The enhanced Xtream Injector
supports injector parameters to be entered on CT console.
New Graphic Retro Graphic Retro allows users to prescribe retro • 1400 mm scannable range for full body
recon graphically on appropriate prospective image by mouse. scans.
Visual adjustment parameters such as DFOV, AP/RL center
improves retro recon productivity. Note : variable depends on table height
• Productivity features designed for the CT Technologist: In- Room
New Enhancement for SmartPrep Enhancement SmartPrep can Start, Remote Gantry Tilt, Breathing Lights with
automatically play the prerecorded messages after the operator countdown timer, Controls Panel mounted on
triggered the scan phase. With Auto Minimum Delay function two corners ( four corners *) of the Gantry.
enabled, the diagnostic delay time is set automatically to the Integrated IV Pole* and Table Tray* at the foot
minimum time. of the table.
New Flip/Rotate provides image orientation flips and rotates
prospectively for auto-filmed images and retrospectively through • 0.35mm isotropic microVoxelTM image resolution
Exam Rx Display and the Image Works viewer. These image
orientation operations do not produce images installed in the
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 2 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
• Recon algorithms, HyperplaneTM and CrossbeamTM, providing • Up to 16 contiguous axial planes acquired simultaneously with each
virtually artefact-free images and optimized slice profile at any pitch, 360° rotation, with the time between scans set by the user-selected
by solving the technical challenges of cone beam and high pitch interscan delay (ISD) or intergroup delay (IGD);
helical scanning.
• Scans may be easily clustered in groups to allow multiple scans in a
single breath hold;
Clinical Performance Examples • Minimum scan-to-scan cycle time of only 1.8 sec with table moves of
≤ 10 mm
With the BrightSpeed Elite Select, users can routinely use a 0.8
second scan speed in conjunction with high pitch helical up to Scout™
1.75:1,affording, for the same image quality, to reduce breath-hold,
• Single radiographic plane for scan localization and graphical
perform better thin-slice CT angiography exams, use thinner slices
prescription of prospective reconstruction;
for most exams, and perform longer helical exams without tube
cooling delays. • Extended range matches helical scannable range.
• Preset WW/WL for scout
Chest / Abdomen / Pelvis
Helical Scans
Coverage 500 mm
Slip ring technology has advanced axial scanning by enabling scans
Rotation 0.8 sec
with simultaneous table movement.
Mode 16 x 0.625 mm
Pitch 1.75:1
mA 260 Helical Multi-slice Prescription
mAs 208
Streamlined prescriptions and easy-to-use default protocols make the
Speed 22 mm/s
BrightSpeed Elite Select fast and efficient in patient set up.
Scan Time 23.8 seconds
Multi-slice acquisitions and short intergroup delays helps to reduce
potential mis-registration between scans by increasing the number of
Peripheral Run-Off scans possible in a patient breath hold.
Coverage 1,400 mm Helical protocols are almost identical to “classical” axial scan protocols.
Rotation 0.8 sec At the beginning of a study, the operator selects the type of exam with
the anatomical programmer, and indicates the desired scan range -
Mode 16 x 1.25 mm either manually, or from a Scout.
Pitch 1.75:1
mA 240 After completing the prescribed exam, the system remains ready to
mAs 192 continue with additional helical scans or a set of axial scans.
Speed 44 mm/s The operator may reconstruct helical scans prospectively, and
Scan Time 32.8 seconds retrospectively, at any arbitrary table location in 0.1 mm increments.
Axial
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 3 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
1.25 6.25 8.75 13.5 16.75 This effectively facilitates later, more detailed image analysis without
additional patient scans and subsequent dose and image registration
6.25 8.75 13.5 16.75 concerns.
2.5
12.5 17.5 27.0 33.5
6.25 8.75 13.5 16.75
Helical Scan Parameters
3.75
12.5 17.5 27.0 33.5
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 4 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
Preset massages are supported in 9 different languages: prospective image by mouse. Visual adjustment parameters such
Chinese, English (Male/Female), French, German, Italian, Japanese, as DFOV, AP/RL center improves retro recon productivity.
Korean, Spanish, and Mexican Spanish.
Trauma Patient: Allows patient scans and image display/analysis Helical Scan Protocols
without entering patient data before scanning.
All protocols assume 120 kVp scans under typical clinical
Biopsy: Simplified prescription for single or multiple scans around conditions.
an arbitrary table position aids biopsy studies.
Reconstruction Algorithms: Soft, Standard, Detail, Bone, Bone Plus, Volumetric Helical Scan Image Quality
Lung , Edge and chest.
With BrightSpeed Elite Select being a sub-millimeter isotropic CT
Reconstruction Matrix: 512 scanner, it is now possible to specify coronal and sagittal image
Display Matrix: 1024 quality. Reformatted resolution is visually demonstrated on the
Catphan High Contrast Resolution Insert Module CTP528.
Display FOV: Freely variable center/off-center,
prospective/retrospective target selection. Single Helical Scans:
Scan Time (sec) Maximum mA
CT Number Scale: -1024 to 3071 HU or extended -31743 to 31743 10 sec 350**
Helical Reconstruction Times: 20 310**-350**
• Up to 6 frames per second reconstruction time 30 260-335**
• Iterative bone processing, which is always enabled for adult 40 225-260/295**
head scanning, reduces image artifacts in head scans stemming 50 200-260/265**
from X-ray beam hardening effects. 60 175-240
Minimum DFOV: 9.6 cm 70 145-215
80 135-200
Minimum Pixel Size: 0.19mm 90 120-185
Queued Recon: Requests will be processed continuously and 100 110-175
simultaneously with other processes on the system including 110 100-165
scanning. Prospective recon will be prioritized over retrospective 120 90-140
recon. 120s(80KV) 140-235
Priority Recon Queuing: One touch selection marks most recent
Multiple Helical Scans:
rotation for next available recon. Available during or after scanning.
Images annotated to indicate continuous scan acquisition with (contiguous helical coverage with 5-second IGD):
table incrementation:
Scan Time IGD No. Scans Max mA
• HE (helical) + Pitch
10 sec 5 sec 3 275**-345**
• Table speed
4 235-305**
Prospective Multiple Reconstruction (PMR): Up to 3 sets of 5 210-250/270**
reconstructions can be pre-programmed as part of the scan 6 185-240
protocol prior to acquisition. The operator can select different
20 sec 5 sec 2 225-295
start/end location, slice thickness, interval, reconstruction
algorithms and display fields of view for each reconstruction. Series 3 180-235
descriptions can be entered for each set of reconstructions to 4 135-195
facilliate hanging protocols on PACS display. This frees the 30 sec 5 sec 2 180-235
operator from sitting at the console and directly contributes to 3 120-185
increased productivity.
4 95-140
Prospective Recon: Operator may initiate full recons at any table
location in increments of 1/10 the image thickness; image thickness Contiguous Helical Coverage
remains constant.
Multiple helical scans can be performed in succession with only 5-
Retrospective Recon: Operator may initiate full recons at any table second delays between helical scans, providing up to 2000
location in 0.05 mm increments; image thickness remains constant. contiguous rotations (with up to 1500 images in one series).
Retrospective Image Decomposition: The operator has the option to (The following chart assumes 120-sec helicals with 5-sec IGD)
retrospectively decompose the original raw data set and
reconstruct additional images at any of the defined nominal image Total Scan Total Elapsed Max
thickness available for a given table speed and scan mode. . The Time (sec) Time (sec) mA
operator also can prescribe retro recon graphically on appropriate
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 5 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
CTDI
Head 18.4mGy/100mAs
.33mm
Body 9.3mGy/100mAs
.36mm
Axial Scans
Multi-slice acquisitions and short interscan delays helps to reduce
potential misregistration between scans by increasing the number of
.38mm scans possible in a patient breath hold.
0% 8.5 2i Mode:
• Produces 2 images per rotation
Hi-Res Algorithm (Edge) - typical
• Nominal Thickness: 0.625, 2.5, 5, 7.5, 10 mm
X/Y – lp/cm
(2.5mm Thickness in 2i Mode is available RETRO recon only)
50% 8.5
4i Mode:
10% 13.0 • Produces 4 images per rotation
Page 6 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
Intergroup Delay (IGD): The following table illustrates the retrospective reconstruction
• Minimum IGD is the same as minimum ISD, also user-selectable. image thicknesses available for each acquisition thickness and
mode:
Scan-to-Scan Cycle:
• Minimum scan-to-scan cycle of 1.8 sec possible for 0.8 sec scan
speed with minimum ISD's. Scan Mode Slice Thickness Recon Slice Thickness
Scan Fields of View: 16i – 0.625mm
• 25 cm for adult head
• 25, 50 cm for body 8i – 1.25mm
• 25 cm for pediatric head 16 row 0.625 4i – 2.5mm
Scan with 0 table increment, contiguous image location, or skipped 2i – 5mm
image location are possible. Overlapped axial scans are not
possible. 1i – 10mm
16i – 1.25mm
Axial Image Reconstruction 8i – 2.5mm
16 row 1.25
4i – 5mm
Reconstruction Algorithms:
• Soft, Standard, Detail, Bone, Bone Plus, Lung Edge and chest. 2i – 10mm
Page 7 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
Scan Technique:
2i – 0.625mm • kVp: 80, 100,120, 140 kVp
2 row 0.625
1i – 1.25mm • mA: 10 to 260mA /350mA** 5mA increments with 120kV
• Power: 31.2KW/42 kW* with 120kV
Orientation: AP, RLAT, PA, LLAT (preset); or any angle from 0° - 359°
with 1°inclement (manually selected).
Axial Scan Image Quality
Axial scan prescription lines indicate scan location to nearest 1 mm
table position.
High Contrast Spatial Resolution
Scouts longer than 1000 mm are auto minified to fit the display.
On GE Performance phantom: Preset WW/WL for scout
Standard Algorithm
•
• 4.0 lp/cm @ 50% MTF
• 6.5 lp/cm @ 10% MTF SmartViewTM Fluoro*
• 8.5 lp/cm @ 0% MTF
Hi-Res Algorithm Offers real-time 12 frames per second CT fluoroscopy with image
latency of less than about 200ms. Detailed targeting is supported
•
by multiple acquisition modes, up to and including high-resolution
• 8.5 lp/cm @ 50% MTF acquisition at 1.25mm (16x0.625 mode only). High dose efficiency is
• 13.0 lp/cm @ 10% MTF made available with continuous and quick check scan modes. A
• 15.4 lp/cm @ 0% MTF simple and efficient user interface provides six user-selectable
display layouts, in-room image review and WW and WL
Low-Contrast Detectability control Flip, rotate, roam and zoom capabilities maintained during
acquisition. Features ceiling-mounded in-room LCD monitor and
On 8 inch (20 cm) CATPHAN phantom: full-featured handheld / cradle-mounted controller. 1CT image
generator(IG) is the prerequisite of SmartViewTM Fluoro.
• 5mm @0.3% at 13.3 mGy
• 3mm @0.3% at 37.2mGy
User Interface
Noise
The BrightSpeed Elite Select Operator Console utilizes a computer
workstation with the following user interface features:
On either an AAPM water phantom or GE Quality Assurance
phantom: • Two 19-inch LCD monitors
• 0.32% +/- 0.03% at 29.3 mGy • Scan/recon monitor for scan and recon control with no image
display
CTDI • Image monitor for image display, analysis, processing, and
management
On CTDI Head and Body Dose Reference Phantoms: • Each monitor provides a 1280 x 1024 high resolution, flicker-free
display
CTDIvol expressed in mGy/100 mAs:
• Scan control keyboard assembly with intercom speaker, microphone
Head 18.4 mGy/100mAs and volume controls
• Three button mouse with mouse pad
Body 9.3 mGy/100mAs
• BrightBox (trackball assembly) *
Scout Scans • Two wide work surfaces
ScoutView™ scans provide excellent detail for anatomical All these devices are free-standing and can be easily moved to
localization in conjunction with scan prescription. accommodate a large variety of working conditions and individual
operator preferences.
Scan locations may be prescribed at the operator console either
graphically (via mouse), or explicitly (keyboard entry) from a Scout Split table top allows unrestricted patient viewing while still supporting
scan. 2 monitors. Each work surface can be adjusted at installation to help
accommodate a variety of siting requirements.
Prescription of scans with multiple tube angles are also available
on a single Scout. Multi-language UIF capability (including Chinese and Japanese)
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 8 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
This feature uses the same interface as New Patient selection for
Desktop Overview simplified, consistent programming.
The user interface utilizes the paradigm of managed work Patient information can be easily recalled to set up an immediate
environments for a more intuitive clinical workflow. exam via List/Select Scheduled Patient on the scan/recon monitor. Pre-
programmed patient exams can also be recalled from the New Patient
Virtually all clinical operations are managed through three “virtual screen automatically by entering the patient ID number.
desktops” or applications managers: Exam Rx, ImageWorks and
Learning Solutions. Operators can effortlessly move back and forth
between these environments simply by clicking on an icon. XtreamTM Patient Data Entry
technology enhances multi-tasking architecture and maintains
simultaneously all processes so no work is lost or disrupted as desktops Patient data can be entered as part of New Patient set-up, or can be
are switched. recalled from the list of pre-scheduled patients.
The Exam Rx desktop environment provides the clinical tools necessary Trauma Patient ID allows patient scans and image display/analysis
for comfortable, efficient control of patient studies. without entering patient data before scanning.
These tools include patient scheduling and data entry, exam protocol
selection, protocol viewing and editing, scan data acquisition, image Exam Protocol Selection
reconstruction, image display and routine analysis, AutoFilm or manual
filming, AutoStore and AutoTransfer. One of the main contributions of the BrightSpeed Elite Select CT
Scanner System to department productivity is its streamlined exam set-
up.
• Exam parameter set-up has been streamlined through the exclusive
ImageWorks use of protocols
• Protocols can be easily selected in one of three convenient ways:
ImageWorks is a desktop environment designed to take advantage of • A large, graphical Anatomical Programmer located on the New
the BrightSpeed Elite Select CT Scanner System computer and image Patient screen
processor.
• A default list of the “top 10” most commonly used protocols
Standard features include archive, network and manual film control, as located near the anatomical programmer
well as some advanced image processing such as multi-planar • A numerical entry
reformatting (MPR), multi-projection volume rendering (MPVR), and MR
image display. It also has optional add-on packages for Volume • Two Anatomical Programmers - one for adults and one for pediatrics
Viewer*, CT Perfusion* and DentaScan*. - provide quick and easy access to 6,840 user-programmable
protocols (total). Each programmer has ten anatomical regions.
The ImageWorks desktop also provides a gateway for DICOM 3.0 image • Default protocols have been expanded through Protocol Pro - a
transactions, either through a local area network, or via DICOM- “behind the scenes” protocol manager - that allows preselection of
formatted MOD media. automated features like AutoVoice, AutoFilm, AutoStore and
AutoTransfer on a per-series basis.
Computer Based Training • Protocol Pro also provides preselection of different window/level
settings for AutoFilm and can automatically display the 1024
The BrightSpeed Elite Select provides an on-screen, on-line operator Localizer each time a new series is requested.
manual via a multi-media CD-ROM player integrated into the operator’s • Default protocols also include preset scan time, kVp, mA, slice
console. Learning Solutions is also viewable on a stand-alone PC thickness, scan mode, table speed, image interval, gantry tilt, scan
providing flexibility and productivity for on-demand learning of system field-of-view, display field-of-view and center, recon types, and
operation.. breath timing parameters.
• Any scan parameter can be edited for each scan or all scans either
before or during an exam. Scans can be easily added or removed
from the prescription.
Exam Rx
• Scan/recon control uses only 2 screens to set up first scan - New
Patient and Protocol View/Edit.
Patient Scheduling
Patient demographics and exam protocols can be pre-programmed in
Protocol View/Edit
advance of patient arrival by selecting Schedule Patient from the
• A single, full screen View/Edit table allows fast and easy examination
scan/recon monitor. This productivity enhancement allows entry of all
and modification of exam parameters before scanning begins
or some of a patient’s demographic data, as well as pre-selection of the
exam protocol. • Exam parameters can be changed for just one scan, or for all scans
in a series
This feature is available any time a patient exam is not currently
underway. • When used in conjunction with the 1024 Localizer, changes made in
the View/Edit table that affect the number of scans, image interval,
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 9 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
starting/ending locations, tilt, or display FOV are automatically shown exam values are saved. Saved values can be networked, filmed
on the 1024 Localizer and archived
• Any changes made directly on the 1024 Localizer display using the
mouse and the on-screen cursor controls are also reflected AutoView Layouts
automatically in the View/Edit table
• View/Edit Wizard intuitively adjusts dependent parameters • Eight powerful AutoView layouts provide exceptional flexibility in
automatically in response to operator-initiated changes and tailoring the 1,024 image display to the user or the application at
highlights them for quick review. It also alerts the operator to hand - without the complexity of free-form “windows.”
incompatible dependencies requiring operator intervention. • AutoView Layouts include:
• Tab card groupings for Timing, Recon and Filming help organize the - 1024 AutoView image
large number of parameters available within each protocol. - 768 AutoView image (matches the image size shown on the
HiSpeed Advantage 2.X Series OC monitor)
Scan Data Acquisition - 512 AutoView image + 512 Localizer Scout with cut lines
automatically showing the location of the AutoView image on
• Full-screen DynaPlan Plus illustrates scan status graphically, with the Scout
real-time feedback while the exam is underway. Scans, programmed - Two 512 AutoView images (same image but at different
delays (prep, breathing, inter-group), and even AutoVoice window/level settings) + 512 Localizer Scout with cut lines
announcements are clearly shown before and during scanning. automatically showing the location of the AutoView images on
the Scout
• AutoScan: Fully automates longitudinal table movement and the start
of each scan - 512 AutoView image + 512 AutoFilm image
- Last two 512 AutoView images
•AutoVoice: 3 preset in 17 user-defined messages automatically - Last four 512 AutoView images
deliver patient breathing instructions with a programmable delay; - AutoLink which links the current series to a view port
especially useful for multiple helical scanning and SmartPrep.
Preset massages are supported in 9 different languages: • Basic image review features such as window/level, magnification and
Chinese, English (Male/Female), French, German, Italian, flip/rotate are available for AutoView images.
Japanese, Korean, Spanish, and Mexican Spanish. • Any window not used for AutoView is available to independent,
simultaneous review of other exams.
•Full Simultaneity allows scan and recon to work concurrently with • Special BrightBox*, a three-button trackball device, provides
image display, processing and analysis (including computationally independent control of image next, prior, manual paging and
intensive features such as MPR, MPVR and 3D*/MIP) while still trackball window/level for any review images in focus. This helps
running image archival, filming and networking processes. make two person operation practical.
• Remote Gantry tilt allows the operator to adjust scan angle from the • Regardless of the AutoView Layout used, AutoFilm viewing is
control room and minimize trips between the scan room and available anytime via an on-image selection - without disrupting
operators workstation. other image processes in progress. Background filming allows full
use of the image display monitor for AutoView and image
review/processing without interruption during AutoFilm.
Dose Computation & Display
• Volume CTDIw (CTDIvol), DLP (Dose Length Product), and Dose Image Review Layouts
Efficiency computation and display during scan prescription provides
patient dose information to the operator. • Five flexible Image Review Layouts are provided for those
• Volume CTDIw (CTDIvol) is a dose index defined by IEC60601-2-44. This applications where greater than 512 image display may be desired
index is computed automatically by the BrightSpeed Elite Select CT and AutoView is not required.
System and reported on the Exam Rx screen. Volume CTDIw is a • Image Review Layouts include:
single number consisting of 2/3 of the CTDI100 peripheral dose plus
1/3 of the CTDI100 central dose that is adjusted by the helical or axial Note: uses short notation for screen options
pitch factor. - 1024 single image display
• CTDI100 is a dose index based upon CTDI dose measurements over a - 768 single image display
100 mm volume, as defined in IEC 60601-2-44. - Two 512 image display, horizontal format
• Dose Length Product (DLP) is given in mGy*cm and is computed and - Two 512 image display, vertical format
displayed for each group prior to the scan. Additionally, an - Four 512 image display
accumulated DLP is displayed for the entire exam, as the exam • Each image display window can be further subdivided into four more
prescription progresses. The final exam accumulated DLP provides a images, increasing the total number of images that can be displayed
convenient measure for maintaining patient or procedure dose at once to 16.
management statistics.
• BrightBox image control is also available for Image Review Layouts.
• Dose Efficiency is automatically computed and displayed on the
Exam Rx screen. The dose efficiency is a measure of how much of
the Z-axis X-ray beam is used by the system, as defined in IEC 60601- Image Access
2-44.
• Dose Reporting saves the CTDIvol and DLP in the patient record • Point and click interface along with a pictorial directory (browser)
allows for easy selection by exam, series or image
as a DICOM secondary screen capture. Series and cumulative
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 10 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
Routine Measurements To save further filming cost, the Operator Console can directly print to a
postscript printer such as the GE Color Printer available as an option.
• Image measurement features provided within Exam Rx:
- Box ROI ImageWorks
- Ellipse ROI
ImageWorks software is designed to take advantage of the
- Trace ROI BrightSpeed Elite Select CT Scanner’s computer and image processor.
- Measure Distance This desktop environment includes image management and
- Measure Angle networking.
- Grid On/Off Because some of the image analysis and display features of
- Hide Graphics ImageWorks replicate those in Exam Rx, the next section describes only
- Erase features that are incremental or significantly different.
- Screen Save
- MIROI (Multiple Image ROI) Image Analysis
- Report Pixels
• Multi-Projection Volume Reconstruction (MPVR): Quick and easy
Display Preferences way to generate volumetric images for CT angiography without
thresholding data or removing unwanted anatomy. An entire
• Display settings available to tailor the overall display (settings apply volume is used to generate images in any plane, creating real-time
to all images in all exams): frames of reference at the same time;
- Annotation Levels • Clinical utility is extended via two additional modes:
- Inverse Video - MIPS - enhances contrast and improves visualization of
- Next/Prior Each View Port calcifications
- Next/Prior Series Binding - Average - generates 2D radiographic images;
- Continuous Report Cursor
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 11 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
• Multi-planar Reformation (MPR): Provides real-time assessment of • Groupings allow application of window/level values,
anatomy in offaxis planes. Sagittal, coronal, oblique and curved magnification/minification, image scroll or flip and rotate to a user-
planar reformations available; defined image set.
• Batch reformatting can also be defined and executed for later • Save State stores user-selected image orientation and window/level
with each data set.
viewing if desired;
• Window/Level values may be:
• Image Addition and Subtraction: Includes image addition of more - Preset to provide six on-screen instant window/level settings
than two images at a time.
- Set independently for up to 16 images on the screen
- User-modified in discrete or variable steps
Volume Viewer* - Adjusted real-time on-image by holding down the middle mouse
button and moving the mouse
Volume Viewer is an innovative and powerfull suite of
productivity enhanvers (Volume Rendering, Volume Analysis • Cine mode provides paging in up to 4 view ports of up to 128
and Navigator) that includes : previously-stored CT or MR images at full selected display frame rate.
For more than 128 images, display frame rate may be reduced.
• Dynamic Volume Review™ for Fast Screening
• Cine mode also provides temporal, spatial or manual playback loops.
• Curved Volume Of Interest
• Text Page
• Protocol Management and Loading
• Review Layout Presets Image Annotation
• Multiple VR Objects Merge
• Image annotation and cursor are shadowed to permit ease in
• Pseudo Surface Shading Mode reading.
• Predefined Cut Planes • Large Font configuration doubles the size of the Patient Name,
• Volume Rendered Navigator views Patient ID and Accession Number for image display and filming.
• VR Preset save/recall
• 3D Rendered Lumen View
• Automatic Path Tracking
Image Management
• Path Bridging (in case of occlusions) • Images may be stored and retrieved via Magnetic Optical Drive
(MOD*) media using DICOM 3.0 format. This allows interchange with
• SmartCursor™ for Easy Navigation other imaging systems supporting DICOM 3.0 MOD media. Not all
• Synchronized Reformatted Views vendors implementation of DICOM 3.0 are identical, so please check
with the manufacturer for compatibility.
• Cut visualization mode
Off-line retrieval of all image files. Images may be viewed as soon as
they are restored from MOD*.
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 12 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
Visual readout is easy to read from the table side or from the operator
Industry Standards console.
• The BrightSpeed Elite Select CT Scanner System complies with a wide
Gantry tilt controls are located on the side of the gantry.
variety of industry standards to facilitate more rapid adoption of
features and performance improvements as the computing and
medical imaging industry evolves. Table
• Single table, cantilever design with wide height range
DICOM Conformance Standards • Vertical Range: 44.1 cm to 99.1 cm
• Vertical Scannable Range: 77.7 cm to 99.1cm
• DICOM 3.0 Storage Service Class • Horizontal Range: Up to 162cm
• Service Class User (SCU) for image send • Horizontal Scannable Range: Up to 134.6cm ,126cm (helical &
• Service Class Provider (SCP) for image receive Scout ) at 140mm below ISO center with 300m extender
• Table automatically re-centers on scan plane with changes in
• DICOM 3.0 Query/Retrieve Service Class vertical position (after setting internal landmark with alignment
• DICOM 3.0 MOD Media Service Class on 1.2- and 2.3-GB MOD media* lights on)
• DICOM 3.0 Storage Commitment Class Push • Table Load Capacity:
• DICOM 3.0 Modality Worklist (including Performed Procedure Step) • 205 kg(450Ib) with +/- 0.5 mm of position repeatability
*(through ConnectPro option) • Controls on gantry for elevation and cradle incrementation. Foot
• DICOM 3.0 Print pedals on both sides of table for elevation. Cradle position controlled
from OC for prescribed scans.
• DICOM Gray Scale Presentation state for image presentation
• IV Pole* integrated at the foot-end of the table helps prevent IV lines
from becoming crossed and tangled, and helps ensure that the lines
• DICOM Structured Dose Report stay securely in place on the patient.
HIPPA
Password protected User login and Authentification
•
X-Ray Tube
• Image annonimization tool
Solarix 350 Tube Unit. Design optimized for exams requiring a large
Product Network Filters restricts access to scanner system by IP
number of scans with less tube cooling.
address, services type ( IE ftp, telnet) and DICOM port number.
User configurable. • Heat Storage Capacity: 3.5 MHU
-
•
-
Filming Protocol • Dual Focal Spots:
• Small Focal Spot:
• 3M-952 Standard
0.9mm (W) x 0.62mm(L) ( Traditional methodology)
0.8mm(W) x 0.5mm(L) Nominal Focal Spot Size ( IEC
System Components 60336:2005)
Loading factors 120kv 125mA
Gantry • Large Focal Spot:
1.2mm(W) x1.2mm (L) ( Traditional methodology)
Advanced slip ring design continuously rotates generator, tube, 1.1mm(W) x 1.0mm(L) Nominal Focal Spot Size ( IEC
detector and data acquisition system around the patient. 60336:2005)
• Aperture: 70 cm
Loading factors 120kv 250 mA
• Tilt: ± 30°
• Tilt Speed: 1°/sec.
• Maximum SFOV: 50cm • Maximum Power: 42 kW
• Rotational Speeds: 360° in 0.8,1.0,1.5, 2.0,3.0,4.0sec.
• Fan Angle 56.8º minimum
• Remote Tilt from Operator’s Console.
• Integrated breathing lights and countdown timer. .
Scan plane toward front of gantry for improved positioning access.
A more streamlined gantry shroud, bilateral table/gantry controls High Voltage Generation
and gantry display maximize maneuverability while working next to • High-frequency on-board generator. Continuous operation during
the gantry. scans.
Laser Alignment Lights: • 42 kW output power.
• Define both internal and external scan planes to ± 1 mm • kVp: 80, 100,120, 140
accuracy.
• Operate over full range of gantry tilt; activated any time during • mA: 10 to 260 / 350**mA, 5mA increments with 120 kV
exam (with tube stationary).
• Coronal light remains perpendicular to axial light as gantry tilts. Maximum mA for each KVp selection
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 13 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 14 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
Compatible Options
• A DICOM Print Interface is standard on the system Siting Requirements.
• Chinese UIF & Keyboard
• English UIF &Keyboard
Recommended Suite Size
• French UIF & Keyboard Recommended Suite Size
• German UIF& Keyboard Exam Room w/o Limited access option:
• Scandinavian UIF & Keyboard
Minimum Size
• Asian Keyboard
3. 91m x 5.87 m (12’ 10”x 19` 3")
• Danish UIF & Keyboard
Recommended Size:
• Dutch UIF & Keyboard
3.96 m x 6.30 m (13’ x 20 ’ 8”).
• Italian UIF & Keyboard
Exam Room w/ Limited access option:
• Norwegian UIF & Keyboard
Minimum Size
• Spanish UIF & Keyboard
3.31 m x 5.92 m (10'10" x 19'5")
• Swedish UIF & Keyboard
• Portuguese, Euro UIF & Keyboard Control Room:
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 15 of 16
BrightSpeed Elite Select GE Healthcare
16 Slice Edition
Product Data Sheet – Dec. 2008
Regulatory Compliance
Cooling Requirements
This product is designed to comply with applicable standards under the
The cooling requirements do not include cooling for the room lighting, Radiation Control for Health and Safety Act of 1968.
personnel or non-CT equipment present. Cooling requirements are
listed by subsystem to allow planning for each room of the CT suite. Laser alignment devices contained within this product are
appropriately labeled according to the requirements of the Center for
Cooling requirements are given for minimum, recommended and Devices and Radiological Health.
growth allowance scenarios.
• The minimum cooling figures assume patient throughput of 3
patients per hour and 75 scan rotations per patient.
• The recommended cooling requirements assume patient throughput
limited by the tube cooling algorithm.
• The suite cooling can be sized for future developments by using the
growth allowance figures. This cooling will accommodate more
patients per hour and/or potential future system enhancements.
This product is a CE-compliant device which satisfies regulations
regarding Electro-Magnetic Compatibility (EMC) and Electro-Magnetic
Interference (EMI), pursuant to IEC-60601-1-2.
Minimum Allowance
Subsystem Watts BTU/hr
Gantry 5500 18766***
Table 200 682
PDU 1500 5118 © 2008General Electric Company, all rights reserved
General Electric Company doing business as GE Healthcare.
Operator Console 2400 8189
* Option for BrightSpeed Elite Select & BrightSpeed Elite Select Standard
**Option for BrightSpeed Elite Select Standard only
Page 16 of 16