Shimadzu+TOC-V+CPH+Software+Users+Manual E PDF
Shimadzu+TOC-V+CPH+Software+Users+Manual E PDF
(M393-E189)
TOC-VCPH/CPN &
TOC-Control V Software
User Manual
SHIMADZU CORPORATION
ANALYTICAL & MEASURING
INSTRUMENTS DIVISION
KYOTO, JAPAN
TOC-VCPH/CPN
Copyright Shimadzu Corporation 2003. All rights reserved, including those to reproduce this publication or
parts thereof in any form without permission in writing from Shimadzu Corporation.
Read and understand this manual before operation. Save this manual.
Information in this publication is subject to change without notice and does not represent a guarantee on the part
of the vendor.
Any errors or omissions which may have occurred in this publication despite the utmost care taken in its produc-
tion will be corrected as soon as possible, but not necessarily immediately upon detection. We appreciate notifica-
tion of any errors or omissions.
Shimadzu does not have any responsibility for errors or injuries resulting from following the instructions in this
manual.
TOC-VCPH/CPN
0
Introduction
Thank you for purchasing the TOC-VCPH/TOC-VCPN Total Organic Carbon Analyzer and
the TOC-Control V software.
This instruction manual is a TOC-V user manual. It provides operation and maintenance
procedures for the TOC-V. The procedures related to TOC-Control V software system
administration are described in detail in the TOC-V Administrator's Manual.
Be sure to read this documentation carefully before using the instrument for the first time.
Read the Operational Precautions before operating the instrument. Follow the procedures in
the manual to avoid compromising user or instrument safety.
CAUTION
• Do not operate this instrument before understanding the contents of this manual.
• If this manual or the warning labels on the instrument become lost or damaged,
promptly obtain replacements from your Shimadzu representative.
• To ensure safe operation, read and follow the procedures in Section 1.2 "Opera-
tional Precautions" before operating the instrument.
Product Warranty and Post Sale Service
Product Warranty
If any problems occur with this instrument, inspect it and take appropriate corrective action
as described in Section 5.7 "Troubleshooting". If the problem persists or symptoms not
covered in the Troubleshooting section occur, contact your Shimadzu representative.
iv TOC-VCPH/CPN
Contents of This Manual
Chapter 1 Overview
This chapter provides an overview of the instrument and operational precautions for its use.
Chapter 4 Operation
This chapter describes pre-analysis preparations and instrument and software functions,
such as setting measurement parameters, starting up the instrument, generating calibration
curves, and conducting sample analysis. This chapter also includes a tutorial that covers
setup and measurement procedures.
Chapter 5 Maintenance
This chapter describes maintenance procedures and periodic checks required to ensure that
the instrument provides consistently high accuracy analyses. Troubleshooting procedures
and error messages are also included in this chapter.
TOC-VCPH/CPN v
Notation Conventions
Notation Conventions
Convention Description
Note: Provides additional information to ensure the correct use of the instru-
ment.
Supplementary Conventions
In addition to the precautionary conventions, the “Tips” and “Reference” conventions are
used to supply additional information.
Convention Description
Reference: Indicates a reference location in the manual where more detailed information can
be found.
vi TOC-VCPH/CPN
Warning Labels Used on this Instrument
The internal temperature of the ozone treatment unit reaches 50°C while the instrument is
running. To prevent burns, switch off the TN unit and wait at least 2 hours for the unit to
cool sufficiently before performing any maintenance.
TOC-VCPH/CPN vii
Warning Labels Used on this Instrument
Due to the danger of electric shock, do not remove this cover. In the event that the cover
must be removed, first disconnect the TNM-1 power cord.
viii TOC-VCPH/CPN
Warning Labels Used on this Instrument
Injury Warning
The sample injection components of this instrument move during operation. To prevent
injury, keep hands away from this mechanism.
During operation, the pressure in the CO2 absorber of the gas purifier is about 200kPa. To
prevent injuries, turn off the carrier gas and switch off power to the TOC main unit. Wait
until the pressure inside the container and the instrument flow lines return to normal
atmospheric pressure before performing any maintenance.
TOC-VCPH/CPN ix
Warning Labels Used on this Instrument
Corrosion Warning
Acid and other corrosive substances are present in the liquid waste drained from this
instrument. Be careful not to touch or spill the liquid waste.
The soda lime in the CO2 absorber of the carrier gas purifier is a corrosive chemical
substance. Be careful not to touch or spill this substance.
x TOC-VCPH/CPN
Warning Labels Used on this Instrument
The liquid discharged from the drain tube connected to the TC injector contains acid and
other corrosive substances. Verify that the tip of the drain tube is securely inserted into the
Y-shaped fitting when connecting the combustion tube and sample injector.
The IC reaction solution is acidic. Be careful not to touch or spill the IC reaction solution.
TOC-VCPH/CPN xi
Warning Labels Used on this Instrument
Hydrochloric acid is a corrosive chemical substance. Be careful not to touch or spill the
hydrochloric acid.
xii TOC-VCPH/CPN
Table Of Contents
Introduction............................................................................................ iii
Product Warranty and Post Sale Service .................................................... iv
Product Warranty ........................................................................................... iv
Post Sale Service ............................................................................................ iv
Contents of This Manual ............................................................................. v
Notation Conventions ................................................................................. vi
Precautionary Conventions Used in this Manual ........................................... vi
Other Conventions Used in this Manual ........................................................ vi
Warning Labels Used on this Instrument .................................................. vii
1 Overview ................................................................................................... 1
1.1 Overview ..................................................................................................... 2
1.2 Operational Precautions............................................................................... 3
2 System Description................................................................................... 7
2.1 TOC-VCPH/CPN Construction...................................................................... 8
2.1.1 Front View ...................................................................................................... 8
2.1.2 Right Side View .............................................................................................. 8
2.1.3 Left Side View ................................................................................................ 9
2.1.4 Rear View........................................................................................................ 9
2.1.5 Inside Front View.......................................................................................... 10
2.1.6 Inside Top View............................................................................................ 11
2.1.7 Flow Diagrams .............................................................................................. 11
2.2 Autosampler ASI-V (Option) .................................................................... 14
2.2.1 Front View .................................................................................................... 14
2.2.2 Rear View...................................................................................................... 14
2.3 8-Port Sampler OCT-1 (Option)................................................................ 15
2.3.1 Front View .................................................................................................... 15
2.3.2 Rear View...................................................................................................... 15
2.4 TN Unit TNM-1 (Option).......................................................................... 16
2.4.1 Front and Right Side Views .......................................................................... 16
2.4.2 Inside Front View.......................................................................................... 17
2.4.3 Rear View...................................................................................................... 18
TOC-VCPH/CPN xiii
Table Of Contents
3.2 Instrument Setup and System Properties ................................................... 23
3.2.1 System Information Window ....................................................................... 23
3.2.2 Options Window........................................................................................... 24
3.2.3 TOC Window ............................................................................................... 26
3.2.4 ASI Window................................................................................................. 28
3.2.5 SSM Window ............................................................................................... 29
3.2.6 Communication Window.............................................................................. 30
3.2.7 History Function Window ............................................................................ 31
3.2.8 Instrument Properties Dialog Box ................................................................ 32
3.2.9 Removing a Configured System................................................................... 33
4 Operation ............................................................................................... 35
4.1 Analysis Preparation .................................................................................. 36
4.1.1 Catalyst Filling ............................................................................................. 36
4.1.1.1 Catalyst for TOC Analysis ................................................................ 37
4.1.1.2 TOC/TN Catalyst (Option) ............................................................... 39
4.1.2 Connecting the Combustion Tube ................................................................ 40
4.1.3 Installing the Syringe.................................................................................... 45
4.1.3.1 Syringe Pump Zero Point Detection ................................................. 47
4.1.4 Water Supply to the Cooler Drain Container ............................................... 48
4.1.5 Filling the Humidifier with Water................................................................ 49
4.1.6 Preparation of Dilution Water ...................................................................... 50
4.1.7 Acid Preparation........................................................................................... 51
4.1.8 Installation of the CO2 Absorber.................................................................. 52
4.1.9 Adjusting Sparge Gas Flow Rate ................................................................. 53
4.1.10 Installing the Suspended Solids Kit.............................................................. 54
4.1.11 IC Analysis Preparations .............................................................................. 56
4.1.12 Preparation for POC analysis ....................................................................... 58
4.1.13 TN Analysis Preparations............................................................................. 59
4.1.14 Preparation and Storage of Standard Solutions ............................................ 60
4.1.14.1 Preparation of Standard Solutions..................................................... 60
4.1.14.2 Storage of Standard Solutions........................................................... 62
4.1.15 Sample Preparation....................................................................................... 63
4.1.15.1 High Sensitivity Analysis.................................................................. 63
4.1.15.2 TC Blank Check Analysis................................................................. 66
4.1.15.3 Pretreatment for IC Removal (Sparging) .......................................... 68
4.1.15.4 Analysis of Samples Containing Suspended Solids.......................... 69
4.1.15.5 Handling Samples Containing Acids, Bases or Salts........................ 71
4.1.16 Sample Preparation....................................................................................... 72
4.1.16.1 Vial Types ......................................................................................... 72
4.1.16.2 Filling Vials with Sample.................................................................. 72
4.1.16.3 Placing Vials in the Vial Rack .......................................................... 74
4.1.16.4 High Sensitivity Autosampler Analysis ............................................ 75
4.1.16.5 Acid Addition in NPOC Analysis ..................................................... 76
4.1.16.6 Sparging in NPOC Analysis.............................................................. 77
4.1.16.7 Stirrer Preparation ............................................................................. 78
xiv TOC-VCPH/CPN
Table Of Contents
4.1.17 8-Port Sampler Measurement........................................................................ 80
4.1.17.1 Type of Sampler Container................................................................ 80
4.1.17.2 High Sensitivity Using the 8-Port Sampler........................................ 82
4.2 Setting General Measurement Parameters................................................. 83
4.2.1 TOC Parameter Settings................................................................................ 83
4.2.2 ASI Parameter Settings ................................................................................. 84
4.2.3 Setting Default Measurement Parameters..................................................... 87
4.2.4 Maintenance History Settings ....................................................................... 88
4.3 Measurement Procedures and Tutorial ...................................................... 89
4.3.1 Starting Up the Instrument ............................................................................ 89
4.3.1.1 Turning on the Power ........................................................................ 89
4.3.1.2 Carrier Gas Pressure .......................................................................... 89
4.3.1.3 Carrier Gas Flow Rate ....................................................................... 90
4.3.1.4 Electric Furnace ................................................................................. 90
4.3.1.5 TN Unit .............................................................................................. 91
4.3.1.6 Ozone Source Airflow Rate for the TN Unit..................................... 91
4.3.2 Tutorial.......................................................................................................... 92
4.3.3 Ending Measurement .................................................................................. 120
4.4 Sample Table Editor ................................................................................ 122
4.4.1 Sample Table Window Overview ............................................................... 122
4.4.1.1 Toolbar Functions ............................................................................ 122
4.4.1.2 Status Bar and Notification Bar ....................................................... 124
4.4.2 File Menu .................................................................................................... 125
4.4.2.1 New.................................................................................................. 125
4.4.2.2 Calibration Curve............................................................................. 126
4.4.2.3 Method Wizard ................................................................................ 135
4.4.2.4 Open................................................................................................. 143
4.4.2.5 Close ................................................................................................ 144
4.4.2.6 Save.................................................................................................. 144
4.4.2.7 Save As ............................................................................................ 145
4.4.2.8 ASCII Export Options ..................................................................... 146
4.4.2.9 Database Export............................................................................... 148
4.4.2.10 Print>Table ...................................................................................... 149
4.4.2.11 Print>Sample Report ....................................................................... 150
4.4.2.12 Print Preview>Table ........................................................................ 150
4.4.2.13 Print Preview>Sample Report ......................................................... 151
4.4.2.14 Print Setup........................................................................................ 151
4.4.2.15 Page Setup........................................................................................ 152
4.4.2.16 Exit................................................................................................... 155
TOC-VCPH/CPN xv
Table Of Contents
4.4.3 Edit Menu ................................................................................................... 156
4.4.3.1 Undo................................................................................................ 156
4.4.3.2 Cut/Copy/Paste................................................................................ 156
4.4.3.3 Replace............................................................................................ 157
4.4.3.4 Find ................................................................................................. 157
4.4.3.5 Exclude............................................................................................ 158
4.4.3.6 Recalculate>All............................................................................... 158
4.4.3.7 Recalculate>Highlighted................................................................. 158
4.4.3.8 Delete Data>All .............................................................................. 158
4.4.3.9 Delete Data>Highlighted ................................................................ 159
4.4.3.10 Import.............................................................................................. 159
4.4.4 View Menu ................................................................................................. 160
4.4.4.1 Calibration Curve ............................................................................ 160
4.4.4.2 Method ............................................................................................ 161
4.4.4.3 Sample Window .............................................................................. 167
4.4.4.4 Outlier Test...................................................................................... 169
4.4.4.5 Properties......................................................................................... 170
4.4.4.6 Customize........................................................................................ 177
4.4.4.7 Toolbar ............................................................................................ 178
4.4.4.8 Status Bar ........................................................................................ 178
4.4.4.9 ASI / 8-Port Sampler Vials ............................................................. 179
4.4.4.10 Data Profile ..................................................................................... 180
4.4.5 Insert Menu................................................................................................. 181
4.4.5.1 Auto Generate ................................................................................. 181
4.4.5.2 Sample............................................................................................. 185
4.4.5.3 Control............................................................................................. 193
4.4.6 Instrument Menu ........................................................................................ 203
4.4.6.1 Background Monitor ....................................................................... 203
4.4.6.2 Connect ........................................................................................... 205
4.4.6.3 Standby............................................................................................ 206
4.4.6.4 Measurement Start .......................................................................... 207
4.4.6.5 Stop>Peak Stop ............................................................................... 207
4.4.6.6 Stop>Finish Current Sample ........................................................... 207
4.4.6.7 Stop>Halt ........................................................................................ 207
4.4.7 Maintenance Menu ..................................................................................... 208
4.4.8 Tools Menu................................................................................................. 208
4.4.8.1 Control Charts ................................................................................. 208
4.4.8.2 Method Validation .......................................................................... 208
4.4.8.3 Administration................................................................................. 208
4.4.8.4 Screen Lock..................................................................................... 209
4.4.9 Options Menu ............................................................................................. 210
4.4.9.1 General Information ........................................................................ 210
4.4.9.2 Default Measurement Parameters ................................................... 212
4.4.9.3 Display Settings>Display Font ....................................................... 213
4.4.9.4 Display Settings>Table Settings ..................................................... 214
4.4.9.5 Display Settings>Notification Bar Settings .................................... 215
4.4.9.6 Display Settings>Floating Point Number Formats ......................... 216
4.4.9.7 Directories ....................................................................................... 217
xvi TOC-VCPH/CPN
Table Of Contents
4.4.10 Window Menu............................................................................................. 218
4.4.10.1 Cascade ............................................................................................ 218
4.4.10.2 Tile ................................................................................................... 218
4.4.10.3 Arrange Icons................................................................................... 218
4.4.11 Help Menu................................................................................................... 218
4.5 Application Operations............................................................................ 219
4.5.1 Manual Injection Analysis .......................................................................... 219
4.5.1.1 Analysis Preparation ........................................................................ 219
4.5.1.2 Measurement.................................................................................... 221
4.5.1.3 Measurement Precautions for Low Concentration Samples............ 225
4.5.1.4 Maintenance..................................................................................... 226
4.5.2 Gas Sample Analysis................................................................................... 227
4.5.2.1 Analysis Preparation ........................................................................ 228
4.5.2.2 Measurement.................................................................................... 229
TOC-VCPH/CPN xvii
Table Of Contents
5.5 TN Unit (TNM-1) Maintenance............................................................... 256
5.5.1 Replacing the Catalyst in the Ozone Treatment Unit................................. 256
5.5.2 Chemiluminescence Detector ..................................................................... 259
5.5.3 Ozone Generator Replacement................................................................... 260
5.5.4 Replacing the NOx Absorber...................................................................... 260
5.6 Software-Controlled Maintenance Functions .......................................... 261
5.6.1 Zero Point Detection................................................................................... 261
5.6.2 Regeneration of the TC Catalyst ................................................................ 262
5.6.3 Regeneration of the IC Solution ................................................................. 262
5.6.4 Residue Removal........................................................................................ 263
5.6.5 Replace Flowline Content ......................................................................... 263
5.6.6 Flowline Washing....................................................................................... 264
5.6.7 Mechanical Check ...................................................................................... 265
5.6.8 ASI / 8-Port Sampler Initialization............................................................. 267
5.6.9 ASI Rack Change ....................................................................................... 267
5.6.10 ASI Needle Change .................................................................................... 268
5.6.11 Change Syringe .......................................................................................... 268
5.6.12 Blank Check ............................................................................................... 269
5.6.13 Maintenance Log ........................................................................................ 269
5.6.14 History ........................................................................................................ 269
5.6.15 Sparge Gas Valve ....................................................................................... 269
5.6.16 Adjust Sparging Flow Rate (Wet Chem.)................................................... 269
5.7 Troubleshooting ....................................................................................... 270
5.7.1 Error Messages ........................................................................................... 270
5.7.2 Troubleshooting.......................................................................................... 274
5.7.2.1 TOC-V............................................................................................. 274
5.7.2.2 ASI-V .............................................................................................. 278
5.7.2.4 Corrective Actions for Poor Reproducibility .................................. 280
xviii TOC-VCPH/CPN
Table Of Contents
6.3 Specifications........................................................................................... 294
6.3.1 TOC-VCPH/CPN ........................................................................................... 294
6.3.2 ASI-V .......................................................................................................... 295
6.3.3 OCT-1 8-Port Sampler ................................................................................ 295
6.3.4 TNM-1......................................................................................................... 295
6.3.5 POC Kit....................................................................................................... 296
6.3.6 Carrier Gas Purifier Kit............................................................................... 296
6.3.7 Manual Injection Kit (Gas Sample Injection Kit) ....................................... 296
6.3.8 PC Hardware Requirements........................................................................ 297
6.4 Standard Accessories............................................................................... 298
6.5 Special Accessories ................................................................................. 303
6.6 Consumable Parts List ............................................................................. 304
6.7 Maintenance Parts List ............................................................................ 308
6.8 Installation ............................................................................................... 310
6.8.1 Before Installation....................................................................................... 310
6.8.2 Installation Site............................................................................................ 310
6.8.2.1 Installation Site Selection ................................................................ 310
6.8.2.2 Installation Site Conditions.............................................................. 311
6.8.3 Installation Procedure.................................................................................. 312
6.8.3.1 Connecting Power Supply and Ground ........................................... 312
6.8.3.2 Changing the Power Supply Voltage............................................... 313
6.8.3.3 Connecting Gas................................................................................ 313
6.8.3.4 Connecting the Drain Tubing .......................................................... 316
6.8.3.5 PC Cable Connections ..................................................................... 317
6.8.4 Installing the Autosampler .......................................................................... 317
6.8.5 OCT-1 8-Port Sampler Installation ............................................................. 325
6.8.5.1 Installation Procedure ...................................................................... 325
6.8.6 Installing the TN Unit ................................................................................. 328
6.8.6.1 Connecting the Power and Installation ............................................ 328
6.8.6.2 Connecting Ozone Source Air ......................................................... 329
6.8.6.3 Connecting Gas Exhaust Tubing ..................................................... 332
6.9 Material Safety Data Sheets .................................................................... 333
6.9.1 Hydrochloric Acid (HCl) ............................................................................ 333
6.9.2 Phosphoric Acid .......................................................................................... 334
6.9.3 CO2 Absorber, CA(OH)2, KOH, NAOH: soda lime .................................. 335
6.9.4 Lithium Hydroxide (monohydrate) ............................................................. 336
6.9.5 Halogen Scrubber........................................................................................ 337
6.9.6 Quartz Wool ................................................................................................ 338
6.9.7 High Sensitivity Catalyst............................................................................. 339
6.9.8 Normal Sensitivity Catalyst ........................................................................ 340
6.9.9 Potassium Nitrate ........................................................................................ 341
6.9.10 Sodium Bicarbonate .................................................................................... 342
6.9.11 Sodium Carbonate....................................................................................... 343
6.9.12 Potassium Hydrogen Phthalate ................................................................... 344
6.9.13 Water ........................................................................................................... 345
TOC-VCPH/CPN xix
Table Of Contents
Appendix A Method Validation ................................................................. 347
A.1 Main Window .......................................................................................... 348
A.2 File Menu ................................................................................................. 349
A.2.1 New............................................................................................................. 349
A.2.2 Open ........................................................................................................... 350
A.2.3 Save ............................................................................................................ 351
A.2.4 Save As....................................................................................................... 351
A.2.5 Print ............................................................................................................ 351
A.2.6 Print Preview .............................................................................................. 352
A.2.7 Print Setup .................................................................................................. 353
A.2.8 Page Setup .................................................................................................. 353
A.2.9 Exit ............................................................................................................. 355
A.3 Edit Menu................................................................................................. 356
A.3.1 Undo ........................................................................................................... 356
A.3.2 Cut/Copy/Paste ........................................................................................... 356
A.3.3 Insert Row .................................................................................................. 356
A.3.4 Delete Row ................................................................................................. 357
A.3.5 Insert Column ............................................................................................. 357
A.3.6 Delete Column............................................................................................ 357
A.3.7 Recalculate ................................................................................................. 357
A.4 View Menu............................................................................................... 358
A.5 Options Menu........................................................................................... 359
A.5.1 General Parameter ...................................................................................... 359
A.5.2 Method Validation Parameter..................................................................... 360
A.5.3 Font............................................................................................................. 361
A.5.4 Floating Point Numbering Format.............................................................. 362
A.5.5 ASCII Export Options ................................................................................ 363
A.5.6 Validation Strictly Following DIN38402................................................... 364
A.6 Help Menu ............................................................................................... 365
xx TOC-VCPH/CPN
Table Of Contents
B.3 View Menu .............................................................................................. 378
B.3.1 Statistics ...................................................................................................... 378
B.3.2 Comparison ................................................................................................. 379
B.3.3 Out of Control Events ................................................................................. 380
B.3.4 Toolbar/Status Bar ...................................................................................... 380
B.4 Options Menu .......................................................................................... 381
B.4.1 Control Chart Options ................................................................................. 381
B.4.2 Font ............................................................................................................. 381
B.4.3 Floating Point Numbering Format .............................................................. 382
B.4.4 ASCII Export Options................................................................................. 383
B.4.5 Enable Exclusion......................................................................................... 383
B.5 Help Menu ............................................................................................... 384
TOC-VCPH/CPN xxi
Table Of Contents
xxii TOC-VCPH/CPN
1
1Overview
This chapter provides an overview of the instrument and operational precautions for its
use.
1.1 Overview
1.1 Overview
Moreover, when the ASI-V Autosampler or OCT-1 8-Port Sampler is used in combination
with the main unit, a fully automatic system is created, allowing automatic analysis of
multiple samples.
This instruction manual covers both the TOC-VCPH and TOC-VCPN models. These models
will be referred to in relevant descriptions as the H Type and N Type, respectively.
Read this user manual carefully before using the TOC-V instrument to ensure its proper
use. The manual should be stored in a convenient location for future reference.
CAUTION
Safety may be compromised if the instrument is used in a fashion other than that
indicated in this user manual. Pay particular attention to the CAUTION sections
to ensure safety when using the instrument.
1. JIS K-0102 "Industrial Waste Water Testing", JIS K-0551: "Total organic carbon (TOC) testing methods for
ultra-pure water", U.S. Pharmacopoeia, EPA 415.1, EPA 9060A, ASTM D2575, Standard Methods for
Examination of Water and Waste Water 5310B
2 TOC-VCPH/CPN
1.2 Operational Precautions
CAUTION
• Do not touch the electric furnace while it is heating. The center of the electric furnace
(near the combustion tube insertion opening) reaches very high temperatures, and burns
may result. <Burn prevention>
• To prevent high sensitivity catalyst from being forced upward in the combustion tube,
verify that carrier gas is flowing when heating the electric furnace. <Prevents the
floating of high sensitivity catalyst>
• Allow the electric furnace to cool to room temperature before removing or exchanging
the combustion tube. Burns may result if this procedure is attempted when the furnace
is at a high temperature. <Burn prevention>
• Do not heat the electric furnace without the combustion tube installed. Parts employed
by the TC injection port (slider block assembly) will be exposed to radiant heat from the
center of the electric furnace and may be deformed. If the furnace must be heated with-
out a combustion tube installed, first remove the TC injection port parts to prevent
deformation, or close off the furnace core opening with a heat-insulating material such
as quartz wool. <Part damage prevention>
• To reduce back pressure, verify that the external drain tubing connected to the waste
port on the right side of the instrument does not touch the surface of the liquid in the
waste container. The height of the external tube must always be lower than the height of
the drain port. Excessive back pressure will result in insufficient drainage and waste will
overflow inside the instrument. <Part corrosion prevention>
Drain container
(Supplied by user.)
• Hand tighten the various connectors to the 8-port valve. Use of a tool may cause
deformation and leakage in the valve. <Part damage prevention>
TOC-VCPH/CPN 3
1.2 Operational Precautions
• Stop operation of the instrument when performing maintenance on any of the drive
components, such as connecting lines to the 8-port valve, or connecting/disconnecting
lines on the sample injector. If these parts are manipulated or if the lines are
disconnected during operation, the drive component could move and cause injury or
discharge of liquid. <Injury prevention, corrosion prevention>
• To shut down the instrument, first select “Shut down instrument” in the TOC-Control V
Standby window. Refer to Section 4.3.3 "Ending Measurement" for more information.
The electric furnace turns off and, after 30 minutes, the power supply shuts off. This
shutdown process extends the life of the TC injection port parts by preventing
unnecessary wear. This procedure does not apply to emergency shutdown.
<Maintaining durability of TC injection port parts>
• The lever of the 3-way valve at the bottom of the IC reaction vessel should normally be
turned to the left position. Refer to Figure 1.2 "Lever Position of 3-Way Valve on the IC
Reaction Vessel". The lever should never be turned to the lower position. The lower
position allows carrier gas to flow at a pressure of 200kPa upstream from the 3-way
valve, causing possible disconnection of joints and damage to parts such as the
combustion tube.
• To use the ozone generator of the TN unit, verify that air is supplied at a rate of 0.5 L/
min to the ozone air source flow meter inside the instrument. Refer to Figure 2.5 "Inside
Front View". Turn ON the ozone generator. This prevents ozone leaks and line
corrosion. <Ozone leak prevention, line corrosion prevention>
• If ozone odor is detected during operation of the TN unit, stop the gas supply by turning
the ozone generator OFF, and contact your Shimadzu representative to request
maintenance. Ozone can cause respiratory organ injury and equipment damage.
<Prevention of injury to respiratory organs and equipment damage>
• To shut down the TN unit, turn the ozone generator OFF at least 3 minutes prior to
switching OFF the TN unit power to stop ozone generation. This prevents ozone leaks
and line corrosion. <Ozone leak prevention, line corrosion prevention>
4 TOC-VCPH/CPN
1.2 Operational Precautions
• The optional POC measurement circuit incorporates a carbon dioxide (CO2) absorber
packed with lithium hydroxide for removal of carbon dioxide gas. Misuse of lithium
hydroxide can cause bodily harm. During packing or replacing the CO2 absorber
container, exercise adequate precaution. After use, treat the material as industrial waste.
The properties and handling guidelines for lithium hydroxide are described in the mate-
rial safety data sheets (MSDS) in Section 6.9 "Material Safety Data Sheets".
• To clean the instrument, wipe the surface with a soft cloth moistened with water. Do not
use chemicals.
• Repairs to the inside of the instrument can be dangerous. Contact your Shimadzu
representative to have a trained service engineer perform repairs.
• Do not disassemble or modify the instrument beyond the scope of the maintenance
procedures described in this manual, as safety may be compromised.
• The construction of the instrument is not explosion-proof. It must not be used in
dangerous areas.
• Use the instrument with a power supply of AC 100-127V ±10% (100V systems) or AC
220-240V ±10% (200V systems).
• The instrument backup battery will last for approximately one year. Connection to a PC
may not be possible if the instrument has been idle for more than one year. If this is the
case, turn the instrument power ON for at least 30 seconds, then turn the instrument
power OFF for 10 seconds and back ON again.
If the instrument usually remains idle for extended periods, turn the instrument power
ON for 24 hours once every six months.
• Do not use the following characters when naming files or systems to be saved:
/ , . \ (back slash) : ; (these are all normal-width characters), space (normal width, em
width)
• Turn OFF the Power Management function of the PC to prevent shutdown of the PC
power while it is controlling the TOC analyzer. A shutdown of this nature would
adversely affect performance.
• Hand-tighten the connectors to the 8-port valve of the 8-port sampler. Over-tightening
with a tool may exert excessive force to the internal valve body, causing deformation
and internal leakage. <Part damage prevention<
• Shut down the instrument before connecting or disconnecting tubing to the 8-port valve
or performing maintenance to the drive parts. Liquid may be discharged from tubes
during operation. <Part corrosion prevention>)
• When utilizing the software's printing function, set the paper size to letter or A4 and
printer graphic resolution to 600 dpi.
TOC-VCPH/CPN 5
1.2 Operational Precautions
6 TOC-VCPH/CPN
2
2System Description
This chapter describes the components of the TOC-VCPH/CPN, autosampler (ASI-V), 8-
port sampler (OCT-1), and the TN unit (TNM-1).
Start Button
Front Door
Power Switch
Stop Button
Drain Tubing
8 TOC-VCPH/CPN
2.1 TOC-VCPH/CPN Construction
2.1.3 Left Side View
Sample Tubing
Hydrochloric Acid
Dilution Water
Bottle
IC Reagent
AC Cord Socket
1
2 4 5
3 6
TOC-VCPH/CPN 9
2.1 TOC-VCPH/CPN Construction
2.1.5 Inside Front View
1 2 4 5
3
16
8
15
14
10
13
Front Door
12 11
10 TOC-VCPH/CPN
2.1 TOC-VCPH/CPN Construction
2.1.6 Inside Top View
Slide Motor
Cooling Fan for
Sample Injection
Block
Slider
Sample Injection
Tubing
TOC-VCPH/CPN 11
2.1 TOC-VCPH/CPN Construction
2.1.7 Flow Diagrams
12 TOC-VCPH/CPN
2.1 TOC-VCPH/CPN Construction
2.1.7 Flow Diagrams
TOC-VCPH/CPN 13
2.2 Autosampler ASI-V (Option)
2.2.1 Front View
Sample Tubing
Stirrer Speed
Sample Cover Control (option)
Signal Cable
Terminal
Power Cable
(attached to ASI) Drain Tubing
14 TOC-VCPH/CPN
2.3 8-Port Sampler OCT-1 (Option)
2.3.1 Front View
8-Port Valve
TOC-VCPH/CPN 15
2.4 TN Unit TNM-1 (Option)
2.4.1 Front and Right Side Views
16 TOC-VCPH/CPN
2.4 TN Unit TNM-1 (Option)
2.4.2 Inside Front View
Reactor
Viton Joint
Flowmeter
TOC-VCPH/CPN 17
2.4 TN Unit TNM-1 (Option)
2.4.3 Rear View
AC Cable Socket
18 TOC-VCPH/CPN
3
This section describes the various system administration functions in the TOC-Control V
software that are available to the user. For a description of system administration functions
that can be performed by the system administrator, refer to the TOC-V Administrator's
Manual.
20 TOC-VCPH/CPN
3.1 Introduction to System Administration Functions
3.1.2 Change Password
TOC-VCPH/CPN 21
3.1 Introduction to System Administration Functions
3.1.3 Add Event Log
Select one of the items in the Select Item array, and then enter a comment in the
Comment text field. If the Comment text field is left blank, an error message is
displayed, and the comment is not added to the Event Log.
22 TOC-VCPH/CPN
3.2 Instrument Setup and System Properties
3.2.1 System Information Window
This section describes the Instrument Setup Wizard, which is used to configure new
systems. This section also describes procedures for viewing the properties of and removing
previously-configured systems.
The TOC-Control V Instrument Setup Wizard is used to enter properties for a new system.
Access the Wizard by double-clicking the “New System” icon in the TOC-Control V main
window.
Multiple instruments can be configured for the TOC-Control V software system by
entering information into the series of Wizard screens.
TIP » Use the Back and Next buttons to move among the various Instrument Setup Wizard
windows.
System
Enter a unique name for the system, using up to 32 characters. A name must be entered
before continuing to the next Wizard page. If a system with the entered name already exists,
an error message is displayed requesting entry of a unique name.
Serial Number
Enter the serial number of the TOC-V. The serial number of the unit is an 8-character
number (No. ********) that is printed on the plate attached to the lower right side of the
instrument.
TOC-VCPH/CPN 23
3.2 Instrument Setup and System Properties
3.2.2 Options Window
User
The name of the current user is automatically entered into the User field and cannot be
edited.
Date of Creation
The current date is automatically entered into the Date of Creation field and cannot be
edited.
Comment
Enter additional system information in the Comment text box. The Comment field is
optional and can contain up to 512 characters.
Oxidation
Select “Combustion” from the drop-down list to specify the oxidation method.
Options
Select the installed instrument components by clicking the appropriate check boxes. The
options include TOC, ASI, SSM (solid-sampler unit), POC, IC Unit, TN, Sparge Kit,
Manual Injection Kit, 8Port Sampler-1, 8Port Sampler-2 and ESU (external scrubbing
unit). The selected options determine which Instrument Setup Wizard screens are
displayed.
Select the IC Unit for the H-type. If Manual Injection Kit is selected, the ASI, SSM, and
POC options are disabled.
24 TOC-VCPH/CPN
3.2 Instrument Setup and System Properties
3.2.2 Options Window
Option Description
TOC Displays the TOC screen of the Instrument Setup Wizard to allow TOC
instrument parameters to be set.
ASI Displays the ASI screen of the Instrument Setup Wizard to allow
autosampler parameters to be set. This option is disabled if the Manual
Injection Kit option is selected.
SSM Displays the SSM screen of the Instrument Setup Wizard to allow solid
sampler unit parameters to be set. This option is disabled if the Manual
Injection Kit option is selected.
POC Specifies that the purgeable organic carbon accessory is installed.
IC Unit Specifies that the instrument will be used to measure inorganic carbon.
Select this option for only instrument model TOC-VCPH. Do not select
this option for instrument model TOC-VCPN. If the IC Unit is selected as
an option, the “Auto regeneration of IC solution” check box will be
enabled on the Instrument Setup Wizard TOC screen.
TN Specifies that the TN accessory is installed. If the TN Unit is selected as
an option, the “TN Power” check box will be enabled on the Instrument
Setup Wizard TOC screen.
Sparge Kit Specifies that the Sparge Kit accessory is installed and that external
sparging is desired. If the Sparge Kit is selected as an option, the “Nee-
dle” parameter will be enabled on the Instrument Setup Wizard ASI
screen.
Manual Inj. Kit Specifies that the Manual Injection Kit accessory is installed. The ASI,
SSM, and POC accessories cannot be used simultaneously with the Man-
ual Injection Kit. If the Manual Injection Kit is selected as an option, the
options for ASI, SSM, and POC are deselected and disabled.
8-Port Sampler 1 Select this option when one 8-port sampler accessory is installed. 8-Port
Sampler 1 cannot be selected if the ASI or SSM are selected.
8-Port Sampler 2 Select this option when two 8-port sampler accessories are installed. 8-
Port Sampler 2 can only be selected if 8-Port Sampler 1 is selected.
TOC-VCPH/CPN 25
3.2 Instrument Setup and System Properties
3.2.3 TOC Window
Catalyst Type
Select the type of catalyst to be used in TC or TC/TN measurement from the drop-down
list. Available options are:
• Regular: for TC measurements (concentrations of 500ppb or greater)
• High Sense: for TC measurements (concentrations below 500ppb)
• TC/TN: for combined TC and TN measurements
Note: The instrument does not automatically recognize the type of cata-
lyst that is installed. Be sure to select the catalyst type that is actu-
ally installed in the instrument.
The high sensitivity catalyst should be used when the expected sample concentrations are
below 500ppb or when detection of minute changes in concentration is necessary.
Tubing Diameter
Select the diameter size of the sample injection tubing from the drop-down list. Available
options are:
• Regular: 0.5mm diameter
• Suspended Particle: 0.8mm diameter
The typical tubing diameter is 0.5mm. Select a tubing diameter of 0.8mm if the Suspended
Solids kit is used.
Cell Length
Select the length of the NDIR detection cell from the drop-down list. Available options are:
• Long: for typical measurements
• Short: for measurements using the Suspended Solids kit
26 TOC-VCPH/CPN
3.2 Instrument Setup and System Properties
3.2.3 TOC Window
Buzzer
Enables the internal buzzer function. Select this option to allow the buzzer to sound during
measurement.
TN Power
This item is displayed if “TN” was selected as an option in the previous screen. Select this
item to perform TN analyses.
TC Furnace
Select the desired TC furnace temperature by clicking the appropriate radio button. Select
680 (°C) for regular or high sensitivity catalyst. Select 720 (°C) for TC/TN catalyst.
TOC-VCPH/CPN 27
3.2 Instrument Setup and System Properties
3.2.4 ASI Window
Tray Type
Select the ASI tray type from the drop-down list. Options are 24mL Vial and 40mL Vial.
Needle
Select the needle type from the drop-down list. Options are:
• Sample: for analysis using only a sample needle
• Sample+Sparge: for analysis using both a sample needle and a sparge needle
• Coaxial: for analysis using an integrated sample and sparge needle
• Coaxial+Sparge: for analysis using both the coaxial needle and a sparge needle,
which allows a sample to be measured while the next sample is sparged.
The Needle field is enabled only if Sparge Kit was selected in the Options window of the
Wizard.
Note: The instrument does not automatically recognize the type of needle installed. Be
sure to select the type of needle that is actually installed in the instrument.
28 TOC-VCPH/CPN
3.2 Instrument Setup and System Properties
3.2.5 SSM Window
Rinse
Select this option to rinse the sampling needle between each sample with water from the
dilution water bottle.
Stirrer On
Select this item to use the optional magnetic stirrer during measurement.
SSM TC Furnace On
Select this option for TC measurement.
SSM IC Furnace On
Select this option for IC measurement.
TOC-VCPH/CPN 29
3.2 Instrument Setup and System Properties
3.2.6 Communication Window
Com Port
Select the communication port from the drop-down list. Options include Com1, Com2,
Com3, and Com4.
Stop Bits
The Stop Bits value is determined by the hardware and is set automatically based on the
Com Port selection. This field is disabled and cannot be edited by the user.
Parity
The Parity value is determined by the hardware and is set automatically based on the Com
Port selection. This field is disabled and cannot be edited by the user.
30 TOC-VCPH/CPN
3.2 Instrument Setup and System Properties
3.2.7 History Function Window
Finish Button
The History Function window is the final screen in the Instrument Setup Wizard. Make the
appropriate selections, then click the Finish button. The new instrument is created and
saved, and its icon appears in the TOC-Control V main window.
TOC-VCPH/CPN 31
3.2 Instrument Setup and System Properties
3.2.8 Instrument Properties Dialog Box
System Tab
The System tab displays the System Information window for the selected system and
allows editing of the Comment field.
Options Tab
The Options tab displays the oxidation method and instrument components that were
selected in the Options window of the Instrument Setup Wizard. Information on this tab can
be viewed but not changed.
TOC Tab
The TOC tab is identical to the TOC window in the Instrument Setup Wizard. Some fields
on the TOC tab can be edited. If the History Log was enabled during setup, edits to the
TOC tab will be tracked and listed in the History tab.
ASI Tab
The ASI tab is identical to the ASI window in the Instrument Setup Wizard. Some fields on
the ASI tab can be edited. If the History Log was enabled during setup, edits to the ASI tab
will be tracked and listed in the History tab.
Communication Tab
The Communication tab is identical to the Communication window in the Instrument Setup
Wizard. Some fields on the Communication tab can be edited. If the History Log was
enabled during setup, edits to the Communication tab will be tracked and listed in the
History tab.
32 TOC-VCPH/CPN
3.2 Instrument Setup and System Properties
3.2.9 Removing a Configured System
SSM Tab
The SSM tab is identical to the SSM window in the Instrument Setup Wizard. Some fields
on the SSM tab can be edited. If the History Log was enabled during setup, edits to the
SSM tab will be tracked and listed in the History tab.
History Tab
If the History Log was not enabled during setup, the History tab is identical to the History
Function window in the Instrument Setup Wizard. Use this screen to enable the History log.
If the History Log was enabled during setup, the History tab lists all changes made to the
system parameters. The log lists modifications chronologically, with the most recent
modification at the top. If the mandatory comment option was selected during setup, all
modifications listed in the History tab window contain a comment describing the reason for
the change. View or edit previously entered comments by selecting the modification from
the list and clicking the Edit Comment button. If the History Log is not required, it need not
be enabled, as the software operates properly without it.
Note that if the mandatory comment option was selected in the History Function window
during instrument setup, the Comment dialog box (shown above) is displayed after any
change is made in the Instrument Properties dialog box.
TOC-VCPH/CPN 33
3.2 Instrument Setup and System Properties
3.2.9 Removing a Configured System
34 TOC-VCPH/CPN
4
4Operation
This chapter describes procedures for pre-analysis preparations, setting measurement
parameters, starting up the instrument, generating calibration curves, and conducting
sample analysis. This chapter also includes a tutorial that covers setup and measure-
ment procedures.
CAUTION
Exercise caution when handling the combustion tube to prevent breakage, as it is
made of quartz glass.
Note: When filling the combustion tube, verify that the catalyst does not become con-
taminated. Contamination will cause high blank values and abnormal analysis
values.
36 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.1 Catalyst Filling
Either standard or high sensitivity catalyst can be used for TOC analysis. The furnace
temperature is set to 680°C for TOC analysis.
Combustion Tube
100mm
Quartz Wool
Platinum Mesh
TOC-VCPH/CPN 37
4.1 Analysis Preparation
4.1.1 Catalyst Filling
Combustion Tube
TOC High-Sensitivity
Catalyst
130mm
Platinum Mesh
Figure 4.2 Filling of TOC Combustion Tube with High Sensitivity Catalyst
38 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.1 Catalyst Filling
TOC/TN catalyst is used for TN analysis and TOC/TN simultaneous analysis. The
detection rate is low for the IC component when it exceeds 10mg/L; as a result, TOC in
samples containing more than 10mg/L IC cannot be measured using the TC-IC method and
should be measured using the NPOC method. The furnace temperature is set to 720°C for
TOC/TN and TN analysis.
Combustion Tube
140mm
Ceramic Fiber 10mm
Quartz Wool
Platinum Mesh
Figure 4.3 Filling Combustion Tube with Catalyst for TOC/TN Analysis
TOC-VCPH/CPN 39
4.1 Analysis Preparation
4.1.2 Connecting the Combustion Tube
CAUTION
• Due to the potential for burns, maintenance of the combustion tube and catalyst
should be performed after the electric furnace temperature has cooled to room
temperature.
• When connecting the combustion tube, handle it carefully to avoid breakage. The
combustion tube is constructed of quartz glass, and injury is possible.
• Do not use a wrench or other tool. Injury could result if the combustion tube
breaks.
40 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.2 Connecting the Combustion Tube
TC Combustion Tube
Nut
TC Combustion Tube
Stainless Tubing
(toward detector)
Compression Fitting
CAUTION
• The instrument is sealed sufficiently if the combustion tube does not move when
pulled upwards.
• The quartz glass combustion tube will break if the compression fitting is over-
tightened.
• Use only the provided Teflon ferrules for the front and rear ferrules. Do not use
any other material, such as metal.
• Do not allow organic contaminants such as grease or oil to touch the inner or
outer surfaces on the tip of the combustion tube and gas connection parts of the
ferrules. This is especially important for high sensitivity analysis.
TOC-VCPH/CPN 41
4.1 Analysis Preparation
4.1.2 Connecting the Combustion Tube
Combustion
Tube
Knurled
Thumbscrew
6. To prevent heat escape from the furnace, pack the space around the capillary part of
the combustion tube with the provided ceramic fiber.
CAUTION
Use caution to avoid inhaling ceramic fibers when inserting the fibers into the
combustion tube.
42 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.2 Connecting the Combustion Tube
7. Verify that the gap between the adjustable base at the bottom of the instrument and the
bottom-most part of the compression fitting is 1mm or less (preferably no gap).
8. If the gap is too large, loosen the nut at the top of the compression fitting and pull the
combustion tube upwards. Since the TC combustion tube is inserted into the
compression fitting, pulling up on the combustion tube shortens the insertion length
and ultimately lowers the compression fitting assembly. When almost no gap remains,
firmly tighten the nut at the top of the compression fitting.
Note: The sample rapidly vaporizes in the combustion tube when injected
in large quantities with a gap present between the bottom-most part
of the compression fitting and the base of the instrument. The
pressure increases, resulting in a separation at the connection
between combustion tube and sample injector.
CAUTION
• The instrument is sealed sufficiently if the combustion tube does not move when
it is pulled upwards.
• The quartz glass combustion tube will break if the compression fitting is over-
tightened.
• Do not allow organic contaminants such as grease or oil to touch the inner or
outer surfaces on the tip of the combustion tube and gas connection parts of the
ferrules. This is especially important for high sensitivity analysis.
TOC-VCPH/CPN 43
4.1 Analysis Preparation
4.1.2 Connecting the Combustion Tube
Combustion
Tube
5. Secure the sample injector with the knurled thumbscrew, as shown in Figure 4.8
"Securing the Sample Injector".
Knurled
Thumbscrew
6. To prevent heat escape from the furnace, pack the space around the capillary part of
the combustion tube with the provided ceramic fiber.
CAUTION
Use caution to avoid inhaling ceramic fibers when inserting the fibers into the
combustion tube.
44 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.3 Installing the Syringe
CAUTION
• Due to the potential for injury, keep hands away from the sample injector during
operation of the syringe pump unit.
• The syringe barrel is made of glass and must be handled carefully to avoid break-
age.
Syringe Installation
1. Open the TOC-Control V Sample Table Editor and connect the instrument.
2. From the Instrument menu, select Maintenance>Change Syringe.
The Syringe Change dialog box is displayed.
3. Click the Preparation Start button to begin the procedure. A progress bar is active
while the syringe moves to the position where it can be replaced.
4. When the progress bar begins blinking, remove the screw at the bottom of the syringe
pump cover, and remove the cover.
5. Remove the sparge gas line from the syringe.
6. Remove the syringe from the syringe connection port of the 8-port valve.
7. Attach the new syringe to the syringe connection port of the 8-port valve.
TOC-VCPH/CPN 45
4.1 Analysis Preparation
4.1.3 Installing the Syringe
8. Connect the sparge gas line to the syringe using the bushing and spacer.
Note: Do not use tools when replacing the syringe; hand-tighten only.
Over-tightening can deform the components of the 8-port valve and
result in leaks.
Syringe
Plunger
Retaining
Screw
9. Verify that the syringe moves smoothly between the upper and lower positions by
clicking the directional arrows on the Syringe Change dialog box.
10. With the syringe in this position, hand-tighten the plunger retaining screw on the
plunger holder.
11. Click the Finish Replacement button to move the syringe to its initial position.
12. Click the Close button.
Note: Once the syringe is installed, perform the syringe pump zero point
detection operation.See Section 4.1.3.1 "Syringe Pump Zero Point
Detection" for details.
46 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.3 Installing the Syringe
This section describes the procedure for performing automatic detection of the syringe
pump zero point.
Note: Perform the zero point detection procedure whenever a new syringe is installed
or the syringe is replaced.
Procedure
1. Open the TOC-Control V Sample Table Editor and connect the instrument.
2. From the Instrument menu, select Maintenance>Zero Point Detection.
The Zero Point Detection dialog box is displayed.
TOC-VCPH/CPN 47
4.1 Analysis Preparation
4.1.4 Water Supply to the Cooler Drain Container
Procedure
1. Remove the rubber cap from the drain container.
2. Remove the drain container from the mounting clip.
3. Introduce pure water (deionized water is acceptable) into the cooler drain container up
to the level of the drain discharge tube on the side of the container.
4. Cap the drain container with the rubber cap.
5. Return the drain container to its position in the mounting clip.
6. Verify that the cooler drain tube reaches to the bottom of drain container.
CAUTION
Do not bend drain tubing, as this may prevent drain discharge.
Drain Tubing
Mounting Clip
Drain Container
Pure Water
48 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.5 Filling the Humidifier with Water
Procedure
1. Remove the cap from the water supply port at the top of the humidifier.
2. Introduce water through the water supply port to the “Hi” marking on the side of the
humidifier.
Note: Use pure water containing as little TC as possible for high
sensitivity analysis.
Water Supply
Port
Pure Water
Humidifier
“Hi” Marking
TOC-VCPH/CPN 49
4.1 Analysis Preparation
4.1.6 Preparation of Dilution Water
Procedure
1. Wash the provided dilution water bottle with laboratory glassware detergent, rinse
well with tap water, then wash a number of times with pure water.
2. Fill the container with pure water to the 2 liter level.
3. Place the container to the left of the instrument, toward the rear.
4. Pass the dilution water intake tubing, labeled DILUTION, through the hole in the cap.
5. Secure the tubing, using a wire clamp about 200mm from the end of the tubing, to
prevent the tubing and cap from separating.
6. Place the cap on the container.
Note: Verify that the tip of the intake tubing nearly reaches the bottom of
the container.
Dilution Water
Intake Tubing
Dilution Water
Bottle
50 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.7 Acid Preparation
Procedure
1. Prepare 2 M HCl by diluting 1 part commercially available concentrated HCL (12 M)
with 5 parts pure water. A final concentration accuracy of ±2% is acceptable.
Note: The pure water used as the diluent must have a TOC concentration
no greater than 0.5 mg/L. Since the TOC value of the diluent pure
water can have a large effect on the analysis result with high
sensitivity analysis (low TOC concentrations), it is necessary to use
pure water having a very low TOC concentration.
CAUTION
Hydrochloric acid is a hazardous chemical. Use precautions to avoid contact when handling
this substance. For details, refer to Section 6.9 "Material Safety Data Sheets".
2. Transfer the prepared acid to the 250 mL bottle provided as a standard accessory, and
place it to the left of the instrument, toward the front.
3. Pass the acid intake tubing through the hole in the cap.
4. Secure the tubing, using a wire clamp about 100mm from the end of the tubing, to
prevent the tubing and cap from separating.
5. Place the cap on the container.
Note: Verify that the tip of the acid intake tubing nearly reaches the
bottom of the container.
Cap
Hydrochloric Acid
Container
TOC-VCPH/CPN 51
4.1 Analysis Preparation
4.1.8 Installation of the CO2 Absorber
Procedure
1. Cut the tips of the 2 rigid tubes in the lid of the CO2 absorber with a knife.
2. Connect the flexible tubing that exits the rear panel of the instrument (near the top) as
follows:
Tubing labeled “L”: Connect to tube labeled “L” on CO2 absorber.
Tubing labeled “S”: Connect to tube labeled “S” on CO2 absorber.
Note: The flexible tubing should not be bent.
3. Place CO2 absorber in the holder behind the instrument.
Flexible Tubing
Rigid Tubes
CO2 Absorber
52 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.9 Adjusting Sparge Gas Flow Rate
Procedure
1. Open the TOC-Control V Sample Table Editor and connect the instrument.
2. From the Instrument menu, select Maintenance>Sparge Gas Valve.
The sparge gas solenoid valve opens and sparge gas begins to flow.
3. Open the front door of the instrument.
4. Adjust the flow rate to about 100mL/min using the sparge gas flow rate adjustment
knob.
Note: The flow rate should be adjusted according to the size and shape of
the sample container.
5. Close the front door of the instrument.
Halogen
Scrubber
TOC-VCPH/CPN 53
4.1 Analysis Preparation
4.1.10 Installing the Suspended Solids Kit
CAUTION
Switch OFF the TOC-V before attempting to install the Suspended Solids Kit.
Procedure
1. Remove the bushing from port 7 of the 8-port syringe pump valve, and disconnect the
TC injection tubing. For the H-type, remove the bushing from port 5, and disconnect
the IC injection tubing.
2. Remove the slider mounting screws from the TC and IC injection blocks, and detach
the sliders.
3. Remove the slider bushings and replace the sample injection tubing with the 0.8mm
diameter sample injection tubing provided in the kit.
4. Reinstall the sliders on the sample injection blocks.
5. Connect the TC injection tubing and the IC injection tubing to port 7 and port 5,
respectively.
Note: During analysis, verify that the sample flow is normal. Refer to
Section 5.7.2.4 "Corrective Actions for Poor Reproducibility"
If the sample stream is not normal, the tip of the Teflon tube may be
deformed or scratched. Use a knife to cut off a portion of the tube.
54 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.10 Installing the Suspended Solids Kit
TOC-VCPH/CPN 55
4.1 Analysis Preparation
4.1.11 IC Analysis Preparations
CAUTION
The IC reagent contains phosphoric acid. Use care while handling to prevent contact or
spills. For details, refer to Section 6.9 "Material Safety Data Sheets".
Procedure
1. Prepare the IC reagent by diluting 50 mL of commercial 85% phosphoric acid (ACS
Reagent grade) to a final volume of 250 mL with pure water.
Note: Since the specific gravity of 85% phosphoric acid is about 1.7, the
concentration of the phosphoric acid is about 25% (weight%) in the
prepared IC reagent. A final concentration accuracy of ±2% is
acceptable.
2. Transfer the prepared acid to the 250 mL bottle provided as a standard accessory, and
place it on the left side of instrument, toward the front.
3. Pass the tip of the IC reagent tubing through the hole in the cap.
4. Secure the tubing, using a wire clamp at a point about 100mm from the end of the
tubing tip, to prevent the tubing and cap from separating.
5. Replace the cap on the container.
56 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.11 IC Analysis Preparations
Procedure
1. Verify that the lever of the 3-way valve at the bottom of the IC reaction vessel is
positioned to the left as shown in Figure 4.18 "Lever Position of IC Reaction Vessel 3-
Way Valve".
2. Open the TOC-Control V Sample Table Editor and connect the instrument.
3. From the Instrument menu, select Maintenance>Regeneration of the IC solution.
The Regeneration of IC solution dialog box is displayed.
4. Click the Start button. The IC reagent will be delivered to the IC reagent vessel by the
liquid delivery pump. During the process, the Start button is disabled.
5. A progress bar is active until the IC reagent is completely regenerated. Click the Close
button.
TOC-VCPH/CPN 57
4.1 Analysis Preparation
4.1.12 Preparation for POC analysis
CAUTION
• Lithium hydroxide is a corrosive (strongly basic) powder. Use care when han-
dling to prevent contact or inhalation. Refer to Section 6.9 "Material Safety Data
Sheets" for details.
• Lithium hydroxide is available in anhydrous and hydrated forms. Since the abil-
ity of the hydrated form to absorb carbon dioxide is slightly inferior to that of the
anhydrous form, use anhydrous lithium to fill the CO2 absorber container.
• When introducing the quartz wool and lithium hydroxide into the container, do
not pack too tightly.
• The CO2 absorber container is constructed of glass. Handle it carefully to prevent
breakage.
Cap
Nipple
O-Ring
Lithium Hydroxide
Quartz Wool
58 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.13 TN Analysis Preparations
Procedure
1. Cut the tips of the rigid tubes (2) in the lid of the CO2 absorber with a knife.
2. Connect the Viton tubing to the 2 rigid tubes as follows:
Tubing labeled “L”: Connect to tube labeled “L” on the CO2 absorber.
Tubing labeled “S”: Connect to tube labeled “S” on the CO2 absorber.
3. Place the NOx absorber behind the instrument.
Viton Tubing
TN Unit
Rigid Tubing
TOC-VCPH/CPN 59
4.1 Analysis Preparation
4.1.14 Preparation and Storage of Standard Solutions
Zero water refers to a standard solution with zero concentration of TC, IC or TN; zero
water also refers to the water used to prepare these standard solutions. Water that is, in
theory, absolutely free of carbon (TC) or nitrogen (TN) is used as zero water. In practice, it
is difficult to obtain this type of water. Even ultra-pure water obtained using a high-
performance membrane technique and water that has been repeatedly distilled will have a
TC content of about 10µg/L. This value is measured directly after manufacture, and
increases when the water is stored in a container. Carbon dioxide is contained in the
atmosphere at concentrations ranging from 300 to 500mg/L. The CO2 content of ambient
air is higher in a room with many people or where combustion equipment is used. Carbon
dioxide in ambient air dissolves in stored water, taking the form of IC. The dissolved
amount is generally about 0.2mg/L, but depends on the water temperature and carbon
dioxide content of the air in contact with the water. Refer to Table 6.2 • "CO2 Content
(ppm) in Distilled Water Equilibrated with Atmosphere Temperature (°C)". Instruments
that prevent contact with air are not incorporated in water uptake, retention or storage
processes during common distillation and purification of water. For this reason, IC is
present in most types of distilled and purified water. In some cases, the TC content of
commercially-available distilled water stored in plastic or glass containers is 1ppm.
Note: The degree of purity required varies depending on the analysis range. Use of
commercially-available purified water is acceptable when measuring samples
with expected TC concentrations around 100ppm. The quality of the zero water
should be considered along with the analysis range.
60 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.14 Preparation and Storage of Standard Solutions
Preparation of POC Standard Solution (using the optional POC analysis kit)
IC standard solution is used as the POC standard solution with this instrument for the
following reasons.
• In POC analysis, the measured substance is an organic material that is volatile and
poorly soluble in water (for example, benzene or chloroform). Preparing a stan-
dard solution at an appropriate concentration using these types of organic sub-
stances is difficult.
• Volatile and poorly soluble organic substances tend to undergo changes in concen-
tration during storage or handling after preparation.
• Standard solutions that are produced using volatile and poorly soluble organic sub-
stances are difficult to use.
Consequently, the IC standard solution is used for this instrument. The preparation
method is listed above.
TOC-VCPH/CPN 61
4.1 Analysis Preparation
4.1.14 Preparation and Storage of Standard Solutions
Storage Time
The limitation on storage of standard solutions is about 2 months for 1000mgC/L standard
stock solutions and about 1 week for diluted standard solutions (for example, 100mgC/L).
The limitations are for cold storage in sealed containers.
Note: IC standard solution absorbs atmospheric carbon dioxide and undergoes concen-
tration changes. As a result, it is particularly important to store IC standard solu-
tion in a sealed container.
62 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.15 Sample Preparation
High sensitivity analysis refers to use of the high sensitivity analysis kit to measure trace
amounts (commonly 0.5mg/L or less) of TOC in samples such as ultra-pure water. The
precautions described in the sections below should be followed when conducting trace
analysis on the order of 10 - 100µg/L.
TOC-VCPH/CPN 63
4.1 Analysis Preparation
4.1.15 Sample Preparation
64 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.15 Sample Preparation
Instrument Preparation
The blank peaks present at the start of high sensitivity analysis will become progressively
smaller and gradually stabilize. As a result, calibration curves generated too soon after the
start of analysis may be inaccurate because they contain data collected during unstable
conditions. To prevent this problem, perform 5 or more measurements of purified water
containing 2-3 drops of 1M HCl per 100mL of water prior to analyzing the calibration
standards. Use the same measurement parameters (injection volume, sparge time) that will
be used for the calibration standards and samples.
Reference: Refer to Section 4.3.2 "Tutorial" and Section 4.4.2.2 "Calibration Curve" for detailed
descriptions of generating calibration curves using the TOC-Control V software.
Sample Analysis
Use the calibration curve generated in the previous section to measure NPOC in the sample.
Measure the sample using the same method used for the calibration standards. If the values
gradually decrease with each repeated analysis using the optional external sparge kit,
extend the sparge time to ensure sufficient sparging. The shape of the sample container, the
sample amount, the amount of sparge gas and the IC content of the sample all have a
bearing on the required sparge time. Determine the appropriate sparge time in accordance
with these parameters.
Reference: Refer to Section 4.3.2 "Tutorial" and Section 4.4.5.2 "Sample" for detailed descriptions of
conducting sample analysis using the TOC-Control V software.
TOC-VCPH/CPN 65
4.1 Analysis Preparation
4.1.15 Sample Preparation
A TOC system invariably contains substances in the instrument flow lines that produce
peaks. These peaks are referred to as blank values or system blank values, and are present
even in water samples containing absolutely no carbon. The magnitude of peaks in the
blanks differs based on various factors such as system construction and the catalyst used.
For the TOC-V, every effort has been made to minimize these peaks, which can affect the
accuracy of TC analyses. Accuracy is particularly affected when trace TOC is measured
(such as in high sensitivity analysis of ultra-pure water samples); for this type of analysis,
the blank peaks must be small and stable. To check and correct for the magnitude of the
system blank peaks, use the TOC-Control V Blank Check procedure described below.
Note: • The Blank Check procedure is available only when the high sensitivity
catalyst is used.
• The Blank Check procedure takes approximately 8 hours to complete and
does not require continuous monitoring by the user. As a result, performing
the procedure overnight is recommended.
• Blank values measured during the Blank Check procedure should rapidly
decrease in magnitude and stabilize. If the instrument flow lines, syringe
pump interior, or catalyst have been contaminated, the blank values will not
rapidly decrease. In such instances, the Blank Check procedure should be
repeated.
• Blank peaks are negligible in the IC analysis flow line. The IC value will be
zero if the water is acidified and sparged. This can easily be confirmed by
conducting analysis using the IC flow line; however, this is not normally
necessary. Consequently, there are no special IC flow line check procedures,
as with the TC blank check.
66 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.15 Sample Preparation
TOC-VCPH/CPN 67
4.1 Analysis Preparation
4.1.15 Sample Preparation
Errors for TC and IC analyses are additive. For samples that have a comparatively large IC
concentrations relative to TOC concentrations, the TOC analysis error is large if it is
determined by the difference between TC and IC results. Samples of this type include
samples from water treatment facilities or from the natural environment, such as rivers,
lakes, swamps, oceans and groundwater.
Using the difference between TC and IC to determine TOC is not appropriate for purified
water or ultra-pure water samples. The CO2 in air dissolves in the sample and is measured
as IC, causing the concentration to vary.
In such cases, the sample is pretreated with aeration or sparging before analysis to remove
the IC. The sample is then subjected to TC analysis, immediately followed by TOC
analysis using the NPOC method.
Sparge Sequence
There are two methods for sparging with this instrument.
(1) Sparging in the sample injector syringe
This is performed only when the instrument is set for automatic sparging before
analysis.
68 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.15 Sample Preparation
(2) Independent sparging of sample using the optional external sparge kit
Sample is introduced into a clean sample container, acidified, and then sparged.
A glass container with an external diameter of 24-25mm and a height of 200-250mm
is appropriate for use as the sample container. Test tubes, stopper-equipped test tubes
and chromatography tubes of suitable size are available commercially. If the outer
diameter of the container is too large, the air to water contact efficiency is decreased,
requiring a longer sparge time.
Sparge Procedure
1. Refer to Section 4.1.9 "Adjusting Sparge Gas Flow Rate" for the sparge gas delivery
procedure.
2. Add 2M hydrochloric acid to the sample to adjust the pH to 2-3.
3. Insert the sparge tubing into the sample.
To improve the sparging efficiency, the tip of the sparge tubing should be as close to
the bottom of the sample container as possible.
Note: In this method, sparge gas can be used independently. While the first sample is
being measured, another sample can be sparged, thereby improving analysis effi-
ciency.
Effect on Analysis
Large variations sometimes occur in TOC analyses of samples containing suspended solids
because TOC is often contained in the suspended material. This effect is evident when the
same sample is analyzed on two different TOC analyzers. Good results are not obtained if
the suspended solids are handled differently in the gathering of biochemical oxygen
demand (BOD) and TOC correlative data.
The suspended solids described here are primarily considered impurities in the sample that
will pass through a 50 mesh strainer (openings of approximately 0.28mm). When larger
solid particles are present, analysis should be performed without drawing water into the
instrument.
The procedure described below is for TOC analysis, but is also valid for TN analysis.
Analysis Restrictions
Since Teflon tubing with an inner diameter of about 0.5mm is used in the instrument for
injecting the sample, suspended solids that do not pass through cannot be measured.
Analysis reproducibility is poor for samples that do not uniformly contain such solids even
if they should pass through the tubing.
Solids that are heavy and readily form sediment, such as silt, can accumulate at the bottom
of the syringe. This can scratch the rotor of the sample injector 8-port valve, causing leaks.
Sedimentation and separation should be performed to as great an extent as possible prior to
analysis.
TOC-VCPH/CPN 69
4.1 Analysis Preparation
4.1.15 Sample Preparation
Countermeasures
The following types of processing operations are normally used for measuring samples
containing suspended solids.
(1) TOC Analysis of Supernatant
In situations when the suspended solids are of no great concern, sedimentation occurs
in the sample container, and the supernatant sample is used. Results depend on the
sedimentation and separation properties of the suspended solids, but usually the
dissolved TOC (DOC: dissolved organic carbon) is measured.
(2) TOC Analysis of Filtrate
The suspended solids are removed using glass fiber filter paper or a membrane filter,
and only the dissolved TOC is measured. Because the filter paper or filter itself is
likely to contain TOC, thorough washing with water or sample is performed prior to
use until no TOC is present in the eluate. Perform heat treatment on the filter for glass
fiber filter paper.
(3) TOC Analysis of Stirred Samples
To perform intentional analysis of suspended solids, the sample is stirred using a
magnetic stirrer or other instrument. Simple stirring can lead to extreme variability in
the results because the suspended solids contained in the sample may not be
distributed uniformly throughout the suspension.
(4) TOC Analysis after Homogenization of the Sample
If the application requires that the suspended solids be included in the analysis, it is
critical that the solids are as fine as possible and distributed uniformly throughout the
suspension. Accurate and reproducible TOC results can be obtained when the sample
is highly homogenized to form uniform micro-size particles.
Homogenizing systems may include a mixer to perform high-speed stirring or an
ultrasonic system. (These are homogenizers with outputs of 150-300W. Typical
washing sonicators do not have sufficient output for this use.) Ultrasonic systems are
generally more powerful, though performance varies with the type of suspended
solids.
Uniform suspension cannot always be achieved with all types of suspended solids.
Sludge-like materials are comparatively easy to process while fibrous materials are
more difficult.
70 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.15 Sample Preparation
Various problems can be encountered when measuring samples that contain acids, bases or
salts. Such problems include corrosion of the NDIR analysis cell as well as detrimental
effects on analysis values and the life of the combustion tube and catalyst. If corrosion of
the inner surface of the NDIR cell occurs, baseline stability will decline and noise will
increase. Samples containing acids, bases or salts should be handled as described below.
The concentration at which adverse effects are produced varies depending on the type of
acid or salt and the TOC or TN concentration. Prior to analysis, dilute the sample so that
acid, base and salt concentrations are reduced to 1000mg/L or less. Neutralization is also
required for bases. The TOC-V is equipped with an automatic dilution function to perform
these dilutions.
Acids
Acids should be diluted to a concentration of 1000mg/L or less. Concentrations up to
5000mg/L can be used if this level of dilution cannot be reached. The acidity of
hydrochloric acid or sulfuric acid can be reduced to a pH of 2-3 with sodium hydroxide or
potassium hydroxide. This process produces thermally stable sodium chloride or sodium
sulfate, which prevent the generation of acidic gas when combusting the sample.
Neutralization of the acid is stopped at a pH of about 2-3, and TOC direct analysis (NPOC
analysis) with sparging to remove the IC is performed. It is necessary to consider the TOC
associated with impurities contained in the alkali reagents used for neutralization. All salts
of nitric acid generate acidic gas under thermal decomposition; as a result, neutralization is
ineffective. When the sample contains hydrochloric or hypochlorous acid, hydrochloric
acid gas is generated from salts that are readily subject to thermal decomposition, such as
ammonium chloride, calcium chloride and magnesium chloride. The instrument is
equipped with a halogen scrubber; however, the scrubber’s main function is absorb
hydrochloric acid gas generated by corrosion of the gold plating on the internal surface of
the NDIR cell.
Alkaline samples
When performing TC analysis on alkaline samples, the following problems can arise:
• Rapid decrease in sensitivity and loss of reproducibility
• Rapid degradation of catalyst or wear of the combustion tube
• Likelihood that IC will be high due to absorption of carbon dioxide
For these reasons, the NPOC analysis is suitable for alkaline samples.
In NPOC analysis, dilute hydrochloric acid is added to adjust the solution to a pH of 2-3,
thereby forming a salt. For example, calcium chloride will be formed as a salt if the sample
contains calcium hydroxide as the base. Since it is preferable for this salt concentration to
be low, alkaline samples should be diluted as much as possible.
Salt accumulation
Samples should be diluted as much as possible to reduce the salt concentration and extend
the life of the combustion tube and catalyst. Conducting a large number of analyses on
samples containing salts will allow the salts to accumulate in the combustion tube,
gradually increasing resistance to the flow of carrier gas and resulting in decreased analysis
reproducibility. When decreased reproducibility is observed, perform the combustion tube
maintenance procedure described in Section 5.2.3 "Washing/Replacing the Combustion
and Carrier Gas Purification Tubes".
TOC-VCPH/CPN 71
4.1 Analysis Preparation
4.1.16 Sample Preparation
Two types of autosampler vials are available. The sample volume associated with each vial
type is shown below. Introduce the specified volume of sample into the vials; introducing
too little sample may result in insufficient sample for analysis.
In most cases, vials can be placed in the ASI-V turntable and analyzed without being
sealed. Vials must be sealed in the following situations:
(1) sparging with the autosampler for NPOC analysis
(2) measuring acidic samples
(3) IC analysis
(4) POC analysis
(5) low-concentration samples (as a guideline, less than 10mg/L)
(6) alkaline samples (pH 7 or higher)
(7) samples that contain volatile organic hydrocarbon components
(8) a large number of samples to be measured
(9) standard solutions
72 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.16 Sample Preparation
In situations (1) and (2) above, corrosion of the instrument may occur due to spattering of
sample, which contains acid. If the samples are left in the autosampler following
measurement, there is danger that the autosampler may be subjected to corrosion due to the
volatilization of acid.
In situations (3) through (9), results may be affected by either volatilization of the TOC
components in the sample or contamination from substances that absorb CO2 from the
atmosphere. If aluminum foil is used to cap the sample vial for NPOC or acidic sample
measurement, there is a danger of acid volatilization through the hole caused by insertion of
the needle. Do not use aluminum foil to cap the vials.
¬Using Caps
Procedure
1. Cut the Parafilm or other sealing material into approximately 25 x 25mm squares.
2. Place a square over the opening of the vial, and stretch it to seal to the edges of the
vials.
TOC-VCPH/CPN 73
4.1 Analysis Preparation
4.1.16 Sample Preparation
3. Press the edges down so that they wrap around the outside of the vial.
Be careful that
film edges do not
descend below
vial threads.
Remove the vial rack from the ASI-V before inserting the sample vials. There are three
different types of vial racks, each accommodating a different vial size.
The ASI-V permits random access of vials in the vial rack. Vials are measured according to
the sequence set in the Sample Table Editor, thus enabling analysis to start from any
position number in the vial rack, and for analysis to occur in non-sequential vial positions.
TIP » When samples are to be sparged using the optional external sparge kit, analysis is
performed more efficiently if the vials are placed in sequential rack positions without gaps.
This allows a vial to be sparged while an adjacent vial is being measured.
74 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.16 Sample Preparation
If samples with TOC concentrations of 100µg/L or less (such as purified water or ultra-pure
water) are measured using an autosampler, the following precautions associated with high
sensitivity analysis must be considered. Refer to Section 4.1.15.1 "High Sensitivity
Analysis" for additional information.
(1) NPOC analysis is recommended. TOC (TC-IC) analysis tends to have a greater
variation in measured values than the NPOC analysis because IC often accounts for a
significant portion of the TC, and the errors in the two values are additive, which
significantly affects TOC analysis.
(2) To generate a calibration curve, fill well-washed vials with standard solutions to about
5-10mm from the top of the vial, and immediately seal each vial.
Use a cap that has been washed with purified water prior to use.
When placing the vial in the rack, handle the vial carefully to avoid contact between
the standard solution and the cap.
Low-concentration standard solutions must be prepared each time they are used.
(3) To stabilize the peaks, perform 4 to 5 injections using the same analysis parameters
prior to analyzing the calibration standard solutions. For example, to generate a
400µg/L calibration curve, place a 0µg/L standard solution in vial position 1 and
measure it 4 to 5 times using the same analysis parameters that were selected for the
calibration curve. Refer to Section 4.4.4.1 "Calibration Curve". The objective at this
point is not to generate a calibration curve, but to inject water using the same injection
volume used for the calibration curve that is to be generated. This not only stabilizes
peaks but also thoroughly washes the sample needle and sparge needle with purified
water.
After the stabilizing injections are complete, place the 0µg/L and 400µg/L calibration
standard solutions in vial positions 2 and 3, insert the calibration standards into the
Sample Table Editor, and start measurement. Refer to Section 4.4.4.1 "Calibration
Curve".
(4) Acid addition for NPOC analysis can be performed automatically using the ASI-V. It
is best to perform manual addition of hydrochloric acid for NPOC analysis on the
order of 10µg/L. This is because, automatic acid addition pierces the vial seal and
allows CO2 contamination from the air.
(5) To prevent rinsing of the needle, deselect the “Rinse” option and set the “No. of
Needle Washes” field to zero in the Instrument Properties Dialog Box>ASI Window.
Refer to Section 3.2.4 "ASI Window".
Even if ultra-pure water is used as rinse water, the TOC concentration of the rinse
water will increase during the time it is exposed. For best results, do not rinse the
needle for analyses on the order of 100µg/L.
(6) Do not use the analysis value acquired from the first sample vial. In sample analysis,
particularly analysis on the order of 10µg/L, the first sample measured may yield high
results.
(7) When using the optional external sparge kit, perform sparging using a sparge gas flow
rate of 100mL/min.
As a guideline, use sparge times of 5 to 10 minutes for 24 and 40mL vials. Increase
the gas flow rate or sparge time if sparging is insufficient.
TOC-VCPH/CPN 75
4.1 Analysis Preparation
4.1.16 Sample Preparation
Automatic addition of acid is possible using the autosampler when the optional external
sparge kit is installed. Using the ASI-V, acidification and sparging can be performed on
samples during NPOC analysis. To perform acid addition using the autosampler, enter the
desired amount in the “Acid Addition” field on the Injection Parameters window of the
Sample Wizard (refer to Section 4.4.5.2 "Sample") or Method Wizard (refer to 4.4.2.3
"Method Wizard").
When analysis is started, acid is drawn from the acid bottle into the syringe, and the set
amount of acid is added from the sample needle into each of the vials that have been
designated for acid addition. Use 2M hydrochloric acid. The acid addition ratio (ratio of
acid with respect to sample) should be set to about 2-5%. However, a setting of about 0.5%
is used for purified water analysis.
The hydrochloric acid concentration or amount added should be adjusted in accordance
with the sample pH or buffering strength to bring the sample pH to 2-3. Prior to analysis,
verify that the pH is 2-3 using pH paper. The pH need not be checked each time if the same
type of sample is routinely measured.
CAUTION
Hydrochloric acid is a hazardous chemical. Exercise caution when handling to
prevent contact and spills. For details, refer to Section 6.9 "Material Safety Data
Sheets".
If the sample pH does not reach a value between 2 and 3, increase the acid addition ratio or
the acid concentration. Excessively increasing the acid addition ratio will result in dilution
of the sample. On the other hand, if the acid amount is too small, the size of the acid droplet
on the tip of the needle will not be large enough to drop from the needle.
Visually observe the process through the glass vial to determine if the acid droplet is
actually added to the sample.
The needle will be rinsed following each acid addition by selecting the “Rinse after acid
addition” option in the Instrument Properties Dialog Box>ASI Window. To save time
during the analysis sequence, rinsing of the needle is unnecessary if the variation in NPOC
concentration among the samples is not large.
76 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.16 Sample Preparation
Samples placed in the autosampler can be sparged by setting a sparge time if the optional
external sparge kit is installed.
If the coaxial needle is used (24mL or 40mL vials), sparging can be performed for the
specified time, followed immediately by drawing sample into the syringe and then by
analysis.
If the coaxial+sparge needle is used (24mL and 40mL vials), the next sample to be
measured can be sparged while the sample vial in the sampling position is being measured.
In this case, sparging is not performed again on the sample when it reaches the sampling
position, which does not pose a problem for ppm-level NPOC.
When measuring NPOC at the µg/L level, use the coaxial needle. This will prevent
atmospheric CO2 and other contaminants from entering the vial.
If the sparge gas flow rate is set too high, sample will be spattered onto the cap seal. Adjust
the sparge gas flow rate to avoid this undesirable spattering.
To verify that enough IC is removed during sparging using the autosampler to enable
unhindered NPOC analysis, perform IC analysis on the remaining sample following NPOC
analysis.
CAUTION
When the vials are capped and sparging is performed without automatic acid
addition via autosampler, pierce the caps before analysis by using the tip of a
provided needle or other instrument. This allows the sparge gas to vent. If sparge
gas is not vented from the vials, pressure will build inside the vials and prevent
effective sparging.
TOC-VCPH/CPN 77
4.1 Analysis Preparation
4.1.16 Sample Preparation
This section describes the procedure for enabling the optional stirrer.
Procedure
1. Open the TOC-Control V Sample Table Editor.
2. From the Instrument menu, select Properties.
The Instrument Properties dialog box is displayed.
3. Click the ASI tab, then click in the “Stirrer On” check box to enable the stirrer.
4. Click the OK button.
78 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.16 Sample Preparation
TOC-VCPH/CPN 79
4.1 Analysis Preparation
4.1.17 8-Port Sampler Measurement
There are no restrictions on the type of sample containers that can be used with the 8-Port
Sampler. It is only necessary that the containers be placed so the sample tubing from the
various ports can be inserted directly into the containers.
Note: Wash all sample containers with an appropriate laboratory detergent, and rinse
them thoroughly with water before using them for the first time.
Loading the Sample
Although it is possible to perform measurement of a sample using the 8-Port Sampler,
while the container open, in the following situations, it is necessary to seal the sample
container.
(1) When performing IC measurement
(2) When performing POC measurement
(3) When performing measurement of low concentration samples (e.g., 10 mg/L)
(4) When performing measurement of alkali samples of pH 7 or higher
(5) When performing measurement of samples containing volatile organic carbon
compounds
(6) When the number of samples is so great as to require a long measurement time
(7) When performing measurement of standard solutions
Note: If the sample container openings are not sealed in the situations listed above,
measurement values may be influenced by such factors as absorption of atmo-
spheric CO2, contamination with carbon containing substances, and evaporation
of TOC constituents from the samples.
Sealing the Sample Containers
Sample containers may be sealed using commercial sealing films such as Parafilm. It is also
possible to pierce the lid of the container and insert the sample tubing into the container
through this opening.
Note: In either case, wash the container with a laboratory detergent, and rinse it thor-
oughly with water before use.
80 TOC-VCPH/CPN
4.1 Analysis Preparation
4.1.17 8-Port Sampler Measurement
Parafilm
TOC-VCPH/CPN 81
4.1 Analysis Preparation
4.1.17 8-Port Sampler Measurement
82 TOC-VCPH/CPN
4.2 Setting General Measurement Parameters
4.2.1 TOC Parameter Settings
Setting Procedure
1. In the TOC-Control V main window, double-click the icon for the desired instrument.
The Instrument Properties dialog box is displayed.
TOC-VCPH/CPN 83
4.2 Setting General Measurement Parameters
4.2.2 ASI Parameter Settings
Buzzer
Enables the internal buzzer function. Select this option to allow the buzzer to sound
during measurement.
Auto regeneration of IC solution
Enables the automatic regeneration of IC solution. This option is enabled only when
IC measurement is being performed, and is typically selected. Refer to Section 3.2.3
"TOC Window" for a detailed description of this option.
Enable ready status check
Enables the Ready Status Check function. When the function is enabled, the software
always verifies that the instrument is in the Ready state prior to starting measurement.
Enabling the Ready Status Check function is recommended. When the function is
deselected, measurement can be performed when the instrument is not in the Ready
state, an option that is mainly for use by service personnel.
TN Power Source
Select this item if the optional TNM-1 unit is installed to perform TN analyses.
TC Furnace
Select the desired TC furnace temperature by clicking the appropriate radio button. If
“Combustion” is the selected oxidation method, select 680 or 720. If “Wet Chemical”
is the selected oxidation method, the TC Furnace field is set to Off and disabled.
Setting Procedure
1. In the TOC-Control V main window, double-click the icon for the desired instrument.
The Instrument Properties dialog box is displayed.
84 TOC-VCPH/CPN
4.2 Setting General Measurement Parameters
4.2.2 ASI Parameter Settings
TOC-VCPH/CPN 85
4.2 Setting General Measurement Parameters
4.2.2 ASI Parameter Settings
86 TOC-VCPH/CPN
4.2 Setting General Measurement Parameters
4.2.3 Setting Default Measurement Parameters
Procedure
1. From the Options menu in the Sample Table Editor, select Default Measurement
Parameters.
The Default Measurement Parameters window is displayed.
TOC-VCPH/CPN 87
4.2 Setting General Measurement Parameters
4.2.4 Maintenance History Settings
Procedure
1. From the Instrument menu in the Sample Table Editor window, select
Maintenance>History.
The Maintenance History dialog box is displayed.
2. Enter warning limits by placing the cursor in the appropriate fields and entering
numeric values.
3. Click the OK button.
TIP » Measurement will be interrupted if the dilution water or acid reservoirs run dry during
analysis. To prevent measurement interruption, set the dilution water and total acid volume
limits to values slightly less than the remaining volume of each respective liquid. A warning
message will be generated before each liquid is completely consumed.Click the Reset
button to the right of a parameter to clear its current value.
88 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.1 Starting Up the Instrument
Turn on the power to the instrument by pressing the power key on the bottom right of the
front panel.
Note: Press the power key again to turn off power to the instrument.
Carrier Gas
Pressure Regulator
TOC-VCPH/CPN 89
4.3 Measurement Procedures and Tutorial
4.3.1 Starting Up the Instrument
Note: • Do not change the carrier gas flow rate during analysis. The peak area
changes in an inversely proportional relationship and may cause errors in
analysis precision.
• With the N Type, although the flow rate is adjusted to indicate 130mL/min on
the flow rate gauge, the actual flow rate used is 230mL/min.
The electric furnace power must be turned on for TC, TOC, NPOC, POC and TN analyses.
The electric furnace does not need to be turned on if only IC analysis is performed. When
TN analysis is conducted, the TN unit must also be turned on.
90 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.1 Starting Up the Instrument
4.3.1.5 TN Unit
When TN analysis is to be performed using the optional TN unit, power to the TN unit and
ozone generator must be turned on.
Adjust the ozone source airflow knob on the TN unit to indicate 500mL/min on the flow
rate gauge.
TOC-VCPH/CPN 91
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
4.3.2 Tutorial
This tutorial gives step-by-step instructions on the use of the TOC-Control V software. The
tutorial assumes that the hardware and software have been installed in the manner
described in Chapters 2 and 3, and that users have a configured system that includes the
TOC-VCPH, ASI and TNM-1 unit.
Note: • If your system does not include the ASI and TNM-1 units, please adjust the
tutorial procedure as appropriate.
In the tutorial, a three-point TC/TN calibration curve and one unknown sample will be
measured. Pure water and the following calibration standards are needed for the tutorial:
• A calibration standard mixture of 0ppm potassium hydrogen phthalate and 0ppm
ammonium chloride
• A calibration standard mixture of 50ppm potassium hydrogen phthalate and 10ppm
ammonium chloride
• A calibration standard mixture of 100ppm potassium hydrogen phthalate and 20ppm
ammonium chloride
92 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
Procedure
1. Start the TOC-Control V software by double clicking the icon or selecting it from the
Start>Programs list.
The TOC-Control V main window is displayed.
2. The instrument configuration must be specified in the software before the instrument
can be connected to the PC for the first time. Only users with administrative privileges
can configure the instrument. If the system has already been configured, proceed to
Step 2 of the tutorial.
3. From the TOC-Control V main window, double-click the New System icon.
The User dialog box is displayed.
4. Enter your user name and password as shown below.
TOC-VCPH/CPN 93
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
5. In the System Information window of the Instrument Setup Wizard, enter the System
Name and Instrument Serial Number. Click the Next button.
The Options window of the Instrument Setup Wizard is displayed.
Reference: For details on the System Name and Instrument Serial Number, refer to Section 3.2.1
"System Information Window".
6. Specify the options that are configured on the system by clicking in the appropriate
check boxes, as shown in the above figure.
7. Click the Next button. The TOC window of the Instrument Setup Wizard is displayed.
94 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
Reference: For details on TOC parameters, refer to Section 4.2.1 "TOC Parameter Settings"
9. Click the Next button. The ASI window of the Instrument Setup Wizard is displayed.
Verify that the ASI parameters match those shown on the figure below.
10. Click the Next button. The Communication window of the Instrument Setup Wizard
is displayed.
11. Verify the communication information (COM Port, Stop Bits, and Parity).
Note: If a Shimadzu representative installed your TOC-Control V soft-
ware, the correct settings for your system are already configured.
Please do not change the settings. Refer to Section 3.2.6 "Commu-
nication Window" for more information on Communication set-
tings.
TOC-VCPH/CPN 95
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
12. Click the Next button. The History Function window of the Instrument Setup Wizard
is displayed. If desired, enable the History Log to track all changes to the method.
Refer to Section 3.2.7 "History Function Window" for more information about the
History Log.
Note: If Audit Trail was checked for the GLP/GMP 21 CFR item in
System Administration Tools of the TOC-Control V main menu,
the History Log function is automatically activated.
13. Click the Finish button.
An icon for the system now appears in the TOC-Control V main window.
TIP » To edit the system configuration, double-click the system icon in the TOC-Control V main
window or, when in the Sample Table Editor window, select Properties from the Instrument
menu. The Instrument Properties dialog box opens, and system configuration options are
displayed on tabs in the dialog box.
3. Open a new Sample Table by either selecting New from the File menu or clicking the
New toolbar button. The New dialog box is displayed.
Change Graphic
96 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
5. Click the System tab and select the desired system from the drop-down list.
6. Select Save File As in the File menu. Enter Tutorial and click the Save button.
TOC-VCPH/CPN 97
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
1. To establish communication between the software and the instrument, either select
Connect from the Instrument menu or click the Connect toolbar button.
The Parameter Configuration dialog box is displayed. Click the Use Settings on PC
button.
The Sequence dialog box is displayed.
2. The Sequence dialog box displays information about the attempt to establish
communication between the software and the instrument.
¬Creating a Sample Table (Step 3)
Procedure
Creating a Sample Table involves three steps. First, enter the default analysis parameters
for the Sample Table (.t32) file.
1. From the Options menu, select Default Measurement Parameters.
2. Three injections of each calibration standard and unknown sample will be made. On
the TC tab change the “No. of injections” value to 3.
The other default measurement parameters should not be changed.
3. On the TN tab change the “No. of injections” value to 3.
The other default measurement parameters should not be changed.
98 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
Reference: For details, refer to Section 4.2.3 "Setting Default Measurement Parameters".
Modify the Sample Table window display settings so that all items contained in the
Sample Table are shown on the screen. The default display settings may not include
all of the Sample Table parameters.
4. From the Options menu, select Display Settings>Table Settings.
5. Ensure that all items on both tabs in the Table Options window are selected. If not,
click the Select All button.
Note: Use this window in the future to create custom Sample Table win-
dows that display only the desired items.
6. Save the Sample Table (.t32) file.
7. From the File menu, select Save As.
The Save As dialog box is displayed. Enter “Tutorial” as the file name and click the
Save button.
Note: The name “Tutorial” appears at the top of the Sample Table win-
dow and for the directory name after the file is saved.
The basic instrument configuration and analysis parameters have been set up.
Reference: Refer to Section 4.4 "Sample Table Editor" for detailed descriptions of the Sample Table
Editor menus and commands.
TOC-VCPH/CPN 99
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
1. Open the Sample Table. From the File menu, select New.
The New dialog box is displayed. Click the Calibration Curve icon.
2. Click the OK button to open the Calibration Curve Wizard. Specific information
about the calibration will be entered in the wizard.
100 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
3. Select the system to be used for the analyses from the drop-down list on page 1 of the
Calibration Curve Wizard. Click the Next button.
Page 2 of the Calibration Curve Wizard, which lists options for calibration curve type,
is displayed. The four types of calibration curves available are:
• Calibration points are distributed uniformly over the concentration range
• Edit calibration points manually
• Calibration curve according to DIN38402/P-51
• Calibration curve according to USP/EP
4. Click the radio button for “Calibration points are distributed uniformly over the
concentration range.” In the future, other options may be more appropriate. Refer to
Section 4.4.2.2 "Calibration Curve" for a detailed description of the calibration curve
types.
Click the Next button.
TOC-VCPH/CPN 101
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
5. Page 3 of the Calibration Curve Wizard is displayed. The following options for
calibration analysis are displayed:
• Analysis
• Default Sample Name
• Default Sample ID
• Calculation Method: Linear Regression or Point to Point
• Check box for Zero Shift
• Check box for Multiple Injections
• Calibration File Name
Select the following options, as shown on the figure below:
• Analysis: TC
• Default Sample Name: Standard
• Default Sample ID: Tutorial
• Calculation Method: Linear Regression
• Check box for Zero Shift: Deselect this option
• Check box for Multiple Injections: Deselect this option
• Calibration File Name: TCTutorial.cal
6. Click the Next button. Page 4 of the Calibration Curve Wizard is displayed.
Notice that the No. of Injections displayed matches the number entered when setting
the default measurement parameters for the Sample Table in Step 3 of the tutorial.
102 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
TOC-VCPH/CPN 103
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
2. On page 4 of the Calibration Curve Wizard, enter 0 to 20ppm for the Concentration
Range and 3 for the number of calibration points.
104 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
3. Page 5 of the Calibration Curve Wizard should match the following figure.
4. Complete the TN calibration file by viewing but not changing pages 6 and 7 of the
Wizard and clicking the Finish button.
The TN calibration curve file is now complete and has been saved.
Reference: Refer to Section 4.4.2.2 "Calibration Curve" for a detailed description of the Calibration
Curve Wizard.
TOC-VCPH/CPN 105
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
4. Enter the vial positions (1-3) for the calibration standards directly into the Vial
column of the table. As vial positions are entered in the table, they are highlighted on
the tray diagram in the Sparging/Acid Addition window.
TIP » The vial position for a sample can also be entered by placing the cursor in the appropriate
row of the Sparging/Acid Addition window and then double-clicking the desired vial
position in the tray diagram.
106 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
6. Insert the TN calibration standard runs by placing the cursor in the second line of the
Sample Table.
From the Insert menu, select Calibration Curve. The Open dialog box is displayed.
Specify the name of the TN Calibration Curve (TNtutorial.cal) and click the Open
button.
TOC-VCPH/CPN 107
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
8. Click the OK button to display the Sample Table window. Notice that the second row
of the Sample Table now contains complete TN calibration standard measurement
information.
Take a moment to scroll through the completed Sample Table. Notice that the col-
umns in the left portion of the Sample Table contain the specified analysis parameters.
Actual measurement information will appear in the columns in the right portion of the
Sample Table once the measurements are completed.
9. Save the Sample Table by selecting Save from the File menu or by clicking the Save
button in the toolbar.
2. The New dialog box is displayed. Click the Method icon and then click the OK
button. The Method Wizard opens.
3. Select the system to be used for the analysis from the drop-down list on page 1 of the
Method Wizard. Click the Next button.
Page 2 of the Method Wizard lists options for the analysis.
4. Select TC/TN from the Analysis drop-down menu. Enter “Sample” in the Default
Sample Name field and “Tutorial” in the Default Sample ID field. In the File Name
field, enter “Tutorial.met” as the name of the Method File.
108 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
TOC-VCPH/CPN 109
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
Reference: Refer to Section 4.4.2.3 "Method Wizard" for a detailed description of the Method Wizard.
110 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
17. Insert the unknown sample by placing the cursor in the third line of the Sample Table.
From the Insert menu, select Sample.
Page 1 of the Sample Wizard (Parameter Source) is displayed.
TIP » To set up multiple analyses using a previously created method, select the Auto Generate
Table command from the Insert menu. Refer to Section 4.4.5.1 "Auto Generate".
18. Click the Method radio button. By selecting this option, the sample will be analyzed
using the measurement parameters specified in the method file we created previously.
Enter the name of the method file (Tutorial.met) or click the Browse button to search
for the file.
Click in the “Skip remaining Wizard pages and use measurement parameters from the
source” check box. By selecting this option, the analysis parameters for the unknown
sample are copied from the source (method) file. The parameters do not need to be
entered manually and the remaining pages of the Sample Wizard do not need to be
displayed.
19. Click the Finish button.
The Sparging/Acid Addition window is displayed.
20. Enter the vial position (2) for the unknown sample (the 50ppm TC/10ppm TN
calibration standard) directly into the Vial column of the table.
21. Verify the information, then click the OK button.
The third row of the Sample Table now contains the measurement parameters for the
unknown sample. (Note that a value will not be displayed in the Result column until
measurement is complete.)
22. Save the Sample Table by selecting Save from the File menu or by clicking the Save
button in the toolbar.
TOC-VCPH/CPN 111
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
Procedure
1. Prepare the three calibration standards (0ppm potassium hydrogen phthalate/0ppm
ammonium chloride, 50ppm potassium hydrogen phthalate/10ppm ammonium
chloride, and 100ppm potassium hydrogen phthalate/20ppm ammonium chloride) and
place them in autosampler tray vial positions 1, 2, and 3.
Note: Calibration standards and unknown samples can be placed in any
vial position when controlling the instrument with the TOC-Control
V software.
2. Check the status of the instrument detectors before starting analysis. From the
Instrument menu, select Background Monitor, or click the Background Monitor
button on the toolbar.
The Background Monitor window is displayed; this window is used to monitor the
real time signal from the ASI and the detectors. On the TOC tab, the status of the
baseline should be OK for each parameter (position, fluctuation, and noise).
3. Click the TN tab to check the status of the nitrogen detector.
5. Place the cursor in the first row of the Sample Table; this allows the measurement
sequence to begin with the TC calibration. From the Instrument menu, select Start, or
click the Start button on the toolbar.
The Standby window is displayed.
112 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
Reference: Refer to Section 4.4.6.3 "Standby" for detailed descriptions of the Standby window options.
7. Click the Keep Running radio button. Then Click the Standby button.
The Sparging/Acid Addition window is displayed.
8. Verify the vial positions, then click the OK button.
9. The Start ASI measurement window is displayed. Click the Start button to begin the
analysis.
10. The status of the measurements can be monitored during analysis by viewing the
Sample Window. To open the Sample Window, either select Sample Window from the
View menu or select the Sample Window button on the toolbar.
TOC-VCPH/CPN 113
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
Note: • The Sample Window is divided into 3 sections, as shown in the above figure.
The Graph is in the upper left corner, the Parameter Table is in the upper right
corner, and the Injection Table is the bottom half of the window. The current
injection is displayed with asterisks in the Injection Table. As each
measurement occurs, information in the appropriate Injection Table row is
completed. Previous injections are shown above the current row. Use the
scroll bar to see all of the information, and use the tabs at the bottom of the
table to toggle between analysis types (between TC and TN, in this case).
• To view the peak profile of the measurement in progress, click the asterisk of
Sample No. To view a previous measurement, click that row.
• When the Sample Window is opened, the text on the graph may be overly
enlarged. In that situation, close the Sample Window and then open it again.
• During measurement, even if “All Peaks” is selected, because it takes time for
the screen to update, the peak display may not change. In this case, select “All
Peaks” following measurement.
• When the measurement Type is TC/TN, NPOC/TN, TOC/TN or
POC+NPOC/TN, the peak profile of the selected measurement type may not
be displayed even after clicking the measurement type (for example, TC and
TN toggle) tab in the Sample Window. In that situation, click the data in the
Sample Window corresponding to the peak you wish to display.
11. Statistical information about the current set of injections is calculated and displayed in
the Sample Window. As a result, preliminary decisions about the measurements can
be made while the analysis is still in progress.
Note: Information about the current injection is shown in the status box in
the upper right portion of the Sample Window.
TIP » The drop-down list above the graph is used to set the graph display. When Current Peak is
selected, the peak profile of the current injection is drawn in the graph as the measurement
is performed. Use this window to monitor the analysis as it occurs. The blue peak denotes
carbon, while the red peak denotes nitrogen. To see the peak profile of a previous injection,
click its row in the Injection Table. To see the peak profile of all the injections for a sample,
select All Peaks from the drop-down list at the top of the graph. Use the graph right-mouse
menu to customize the peak profile graph.
Reference: Refer to Section 4.4.4.3 "Sample Window" for a detailed description of the Sample Window
options.
12. When the measurements are complete, the autosampler moves to the rinse well for a
flow line rinse and the instrument status tab displays a Ready status. Close the Sample
Window.
114 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
2. From the View menu, select Calibration Curve, or click the Calibration Curve button
on the toolbar.
The Calibration Curve Properties window for the TC calibration is displayed.
3. Select the Graph tab to view the TC calibration curve data in graph form.
TIP » A successful calibration can be used for future analyses by specifying the calibration file
name (*.cal) in the method.
5. Close the TC Calibration Curve Properties window when the evaluation is complete.
TOC-VCPH/CPN 115
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
1. View peak profiles by placing the cursor in the desired row of the Sample Table and
either selecting Sample Window from the View menu or clicking the Sample button
on the toolbar.
The Sample Window is displayed; this is the same window we used to observe the
real-time injections.
2. To view the peak profiles for all injections in a Sample Table row, select All Peaks
from the drop-down list at the top of the graph.
All peak profiles for the selected row are displayed in the graph.
TIP » When All Peaks is selected from the drop-down list at the top of the graph, clicking in
another row of the Sample Table will automatically update the Sample Window display with
the peak profiles for that row.
3. To view the peak profile for individual injections, click in the appropriate row of the
Injection Table and select Current Peak from the drop-down list at the top of the
graph.
TIP » Use the TC and TN tabs at the bottom of the Sample Information table to view the results
for both analyses.
116 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
Outlier Test
The Outlier Test is used to provide additional information about the measurement result.
1. Place the cursor in the third row of the Sample Table, which contains analysis
information for the unknown sample.
2. From the View menu, select Outlier Test.
Note: The Outlier Test is conducted to identify suspect results. The test
statistically compares all unknown samples that have the same
sample name. Because this tutorial includes only one measurement
of the unknown sample, the outlier test calculation is not
meaningful.
Reference: Refer to Section 4.4.2.15 "Page Setup" for detailed descriptions of the Page Setup options.
3. On the Instrument tab, ensure that the Print system information check box is selected,
then select the options shown in the figure below.
TOC-VCPH/CPN 117
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
4. Click the Calibrations tab and select the options shown in the figure below.
118 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.2 Tutorial
TIP » The report can also be previewed from outside the Page Setup window by selecting Print
Preview>Sample Report from the File menu.
The Preview feature displays each page of the report exactly as it will be printed.
Notice that each item that was selected in the Page Setup window appears in the
report.
TIP » Magnify the view by clicking the magnifying glass tool, or on the Zoom In button. Move
from page to page by clicking the Next Page and Previous Page buttons.
2. Print the report by clicking the Print button.
Reference: The tutorial provided instructions for setting instrument configuration and analysis
parameters, adding calibration standard and unknown sample information to the Sample
Table, creating calibration curves and method files, conducting measurements, and
checking and printing results. More details on all of these procedures are discussed in
Section 3.2 "Instrument Setup and System Properties", Section 4.2 "Setting General
Measurement Parameters", and Section 4.4 "Sample Table Editor".
TOC-VCPH/CPN 119
4.3 Measurement Procedures and Tutorial
4.3.3 Ending Measurement
Procedure
1. From the Instrument menu, select Standby.
120 TOC-VCPH/CPN
4.3 Measurement Procedures and Tutorial
4.3.3 Ending Measurement
Auto Restart
The Auto Restart function is used to shut down the instrument and then restart it at a
specified time. Selections are made for the date and time that measurement will next be
performed, as well as for the instrument restart settings. After shutting down the
instrument, the carrier gas flow and power to the furnace are automatically turned off.
Approximately one hour before the restart time, the furnace and the carrier gas will
automatically be turned on to allow the instrument to achieve a Ready state by the restart
time.
1. Select Auto Restart in the Standby window.
2. Select the appropriate instrument restart settings (TC furnace, carrier gas, ozone
generator) and set the restart date and time. If a check is placed in the box beside a
unit, the furnace power remains on and the carrier gas continues to flow. In order to
have the power turned off automatically, do not place a check mark in any of these
boxes.
3. Click the Standby button.
Power to the instrument is automatically switched off and will be automatically
switched on approximately one hour before the restart date and time. Using Auto
Restart, measurement can be performed immediately at the specified startup time.
Reference: Refer to Section 4.4.6.3 "Standby" for detailed descriptions of the available instrument
shutdown options.
TOC-VCPH/CPN 121
4.4 Sample Table Editor
4.4.1 Sample Table Window Overview
The TOC-Control V Sample Table Editor menus and commands are described in this
section.
If the Toolbar is not visible, select Toolbar from the View menu, then specify the desired
Toolbar options.
Note: Many of the Toolbar functions are disabled until a Sample Table is opened or
communication is established with an instrument.
TIP » A description of each Toolbar button’s function appears when the mouse pointer is
positioned over the button.
TIP » All of the Toolbar commands are also accessible through the Sample Table Editor menus.
New File
Opens a new, blank Sample Table.
Open File
Opens a saved Sample Table. Clicking the button displays the Open dialog box,
in which a *.t32 file can be selected.
Save File
Saves the data in the open Sample Table. If the open Sample Table is new and has
not yet been saved, the Save As dialog box is displayed.
Cut
Removes data from highlighted cells in a table and saves them to the clipboard.
Copy
Copies data from highlighted cells in a table and saves them to the clipboard.
Paste
Copies the contents of the clipboard to a selected location in a table.
Print
Sends the data from the currently displayed window to the printer. Clicking the
button starts printing immediately using the current printer settings.
Help - About
Provides information about the TOC-Control V software.
Background Monitor
Displays the Background Monitor window, which displays the status of the ASI
and the instrument detector(s).
122 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.1 Sample Table Window Overview
Connect
Opens or closes the electronic connection between the instrument and the PC.
Standby
Shuts down or reactivates the instrument.
Start/Continue
Starts measurement. Either the Manual Injection Parameter or Sparging/Acid
Addition window is displayed, depending on the instrument settings.
Stop
Stops the measurement sequence after all injections of the current sample are
complete.
Halt
Immediately stops the current measurement and discards data from the inter-
rupted measurement.
Peak Stop
Stops the measurement sequence and aborts analysis of the current sample. The
peak in progress is treated as if it were a completed analysis.
Properties
Displays the Sample/Method Properties dialog box.
Calibration Curve (View)
Displays the calibration curve data. Highlight the calibration standard row in a
Sample Table to select the calibration curve to examine.
Sample
Opens the Sample Window, which displays detailed information for each sample
injection and can be used to view measurement data in real time.
Method
Opens the Method Properties window, which contains the method parameters for
a selected sample.
Exclude
Excludes a selected sample or injection from the calculation. Toggles to include
an excluded sample.
Recalculate Highlighted
Recalculates the highlighted sample results after applying a new calibration
curve or excluding injections.
Recalculate All
Recalculates all sample results after applying a new calibration curve or exclud-
ing injections.
Delete Highlighted Data
Deletes the highlighted measurement data. This function does not change the
structure of the Sample Table.
Delete All Data
Deletes the measurement data for all samples. This function does not change the
structure of the Sample Table.
TOC-VCPH/CPN 123
4.4 Sample Table Editor
4.4.1 Sample Table Window Overview
Format
Displays the Floating Point Format dialog box, where options for numeric for-
mats can be modified.
Font
Displays the Font dialog box, where options for fonts can be modified.
Insert Sample
Opens the Sample Wizard, and allows a sample to be inserted in the active posi-
tion of the Sample Table.
Autogenerate
Opens the Sample Group Wizard, and allows a specified group of samples to be
inserted in the active position of the Sample Table.
Insert Control
Displays the Open dialog box, and allows a selected control sample template to
be added to the Sample Table.
Insert Calibration Curve
Displays the Open dialog box, and allows a selected calibration curve to be added
to the Sample Table.
The Status Bar displays the status of various software functions and is displayed at the
bottom of the open window for all applications. If the Status Bar is not visible, select Status
Bar from the View menu.
The Notification Bar displays information about the status of the TOC-Control V software,
the connection between the instrument and the PC, measurement events, and operation
errors. Use the Options>Display Settings>Notification Bar Settings window to enable,
disable, or select settings for the Notification Bar. Refer to Section 4.4.9.5 "Display
Settings>Notification Bar Settings".
Notification
Bar
Status Bar
124 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
4.4.2.1 New
The New command opens the New dialog box, which displays the options for all new file
types. The options are Sample Run, Calibration Curve, Method, Control Sample Template,
and Parameter File, as shown in the figure below and described in the text following the
figure.
TIP » A new file can also be created by clicking the New File button on the Toolbar and typing in
the new file name.
Sample Run
Displays the General Information window, which can also be accessed by select-
ing General Information from the Options menu. Refer to Section 4.4.9.1 "Gen-
eral Information" for a detailed description of the General Information window.
Calibration Curve
Opens the Calibration Curve Wizard, which can also be accessed by selecting
Calibration Curve from the Insert menu. Refer to Section 4.4.2.2 "Calibration
Curve" for a detailed description of the Calibration Curve Wizard.
TOC-VCPH/CPN 125
4.4 Sample Table Editor
4.4.2 File Menu
Method
Opens the Method Wizard. Refer to Section 4.4.2.3 "Method Wizard" for a
detailed description of the Method Wizard.
Control Sample Template
Opens the Control Sample Wizard. Refer to Section 4.4.5.3 "Control" for a
detailed description of the Control Sample Wizard. Refer to Appendix B for a
description of the Control Charts function.
Parameter File
Displays the Save As dialog box, which can be used to save a parameter file.
OK Button
Opens a new file for the selected icon, and closes the New dialog box.
Cancel Button
Press the Cancel button to close the New dialog box without making a new file.
The Calibration Curve command is used to insert a calibration curve into the Sample Table.
Selecting the command displays the File>Open dialog box.
Figure 4.75 Selecting Insert>Calibration Curve displays the File>Open dialog box
A table of calibration information is displayed in the window at the bottom of the Open
dialog box after a calibration curve is selected. Click the OK button to insert the selected
calibration curve into the Sample Table. Click the Cancel button to abort the operation.
Note: • The Insert>Calibration Curve command is not available during sample
analysis.
• An error message is displayed if the selected file does not have the correct
format or is not a calibration curve.
126 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
To create a new calibration curve for insertion into the Sample Table, click the New button
to open the Calibration Curve Wizard. The Calibration Curve Wizard consists of several
windows, each of which defines specific parameters for the calibration curve, as described
below.
TIP » The Calibration Curve Wizard can also be accessed by selecting File>New and double-
clicking the Calibration Curve icon in the New dialog box.
Calibration Curve Wizard (Page 1) System Information
Parameter Description
System Select the system to be used.
User Displays the user name. This field cannot be changed.
Date of Creation Displays the current system date and time. This field cannot
be changed.
Comment Enter a comment (512 characters maximum).
TOC-VCPH/CPN 127
4.4 Sample Table Editor
4.4.2 File Menu
128 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
Note: In the case of the TOC-VCPN, dilution measurement cannot be performed for IC
measurement. Therefore, since the dilution function will not work even if Use
Source Dilution Function or Dilution Factor Manual Setting are selected, select
Dilution Factor Automatic Setting. In IC measurement using the TOC-VCPH, the
dilution function can be used.
The Analysis Information window is used to specify the type of analysis and the standard
name for the calibration curve.
Parameter Description
Analysis Select the analysis type.
Default Sample Name Enter the sample name for the calibration standards (64 char-
acters maximum).
Default Sample ID Enter the default sample identification for the calibration
standards (64 characters maximum).
Calculation Method Select the calculation method to be used for the calibration
from the drop-down list. Available calculation methods are
point-to-point and linear regression. Refer to Section 4.4.4.5
for detailed descriptions of the calibration methods.
Zero Shift Select this option to shift the calibration curve through the
origin. Refer to Section 4.4.4.5 for a detailed description of
the Zero Shift option.
Multiple Injections Select to enable the syringe to draw up a single aliquot with a
volume equal to that required for all injections of the stan-
dard. When this option is not selected, the syringe will draw
up a separate aliquot for each injection; as a result, selecting
this option reduces analysis time.
File Name Enter the file name of the calibration curve.
TOC-VCPH/CPN 129
4.4 Sample Table Editor
4.4.2 File Menu
Parameter Description
Units Select the concentration units from the options on the drop-
down list.
Concentration Range Enter the maximum and minimum concentrations of the cali-
bration curve standard solutions.
No. of Cal. Points Enter the number of calibration standards that will be used to
create the calibration curve.
No. of Injections Enter the minimum/maximum number of injections for the
calibration standards.
No. of Washes Specifies the number of times the syringe is to be washed
with standard prior to the first injection. Note that if the Auto
Dilution option was selected, the standard will be diluted and
the syringe will then be washed with the diluted standard.
Enter an integer from 0 to 10.
SD Max and CV Max Specifies the maximum standard deviation and coefficient of
variation that is acceptable for the injections. Enter a value
ranging from 0 to 9999 for SD Max and from 0 to 100 for CV
Max. If either of these values are met, no additional injec-
tions are required. However, if both of these values are
exceeded, the calibration standards are automatically rein-
jected up to the maximum number of times specified in the
No. of Injections field. The SD Max and CV Max fields are
disabled if the No. of Injections is 1 or if the minimum is
equal to the maximum.
Sparge Time Enter the length of time sparging will occur. (0 - 20 minutes)
Acid Addition Enter the desired quantity of acid to be added. This option is
available only for NPOC or related analyses.
130 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
Parameter Description
No. Displays the calibration point number. Note that the standards
will be analyzed in the order shown. To add a calibration
point at a certain position, click the desired number in the
“No.” column, then click the Add button and enter the appro-
priate information. The new point will be added in the speci-
fied position, and the point that was previously in that
position will move down the list to the next position.
Inj. Volume Displays the injection volume for the calibration standards.
Calibration Points Displays a table of measurement parameters (such as those
entered in the previous page of the Calibration Curve Wizard)
for each calibration point. To edit the parameters for a cali-
bration point, highlight the point in the table and click the
Edit button. To add a calibration point, click the appropriate
number in the No. column to specify the insertion point, then
click the Add button. To delete a calibration point, highlight
the point in the table and click the Delete button. To delete all
calibration points, click the Delete All button.
TOC-VCPH/CPN 131
4.4 Sample Table Editor
4.4.2 File Menu
Click the Edit and Add buttons to display the Edit Calibration Point Parameters window.
This window is used to enter calibration points manually and to edit the parameters of
existing calibration points.
Parameter Description
Calibration Point No. Displays the calibration point number. This field will reflect
the number selected by clicking in the “No.” field in the Cali-
bration Points List. The number cannot be changed directly in
the window.
Standard Solution Conc. Enter the concentration of the calibration standard solution.
When creating a calibration curve from a single standard
solution by using the Auto Dilution function, enter the con-
centration of the original standard solution in this field.
Auto Dilution Enter the factor by which the instrument should automati-
cally dilute the calibration standard solution to obtain the
desired concentration for the calibration point.
Cal. Point Conc. Displays the actual concentration of the calibration point. The
concentration is calculated automatically based on the values
entered into the Standard Solution Conc. and Auto Dilution
fields.
No. of Injections Enter the minimum and maximum number of injections.
SD Max and CV Max Enter the maximum standard deviation or coefficient of vari-
ation values.
132 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
TOC-VCPH/CPN 133
4.4 Sample Table Editor
4.4.2 File Menu
134 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
The Method Wizard is used to set method parameters and is accessed by selecting
File>New>Method. The Method Wizard consists of several windows, each of which is
described below. The Back and Next buttons are used to navigate between windows. Note
that each set of parameters selected for a specific method applies only to that method.
Note: Selected windows in the Method Wizard will be repeated for combined analysis
types such as TOC, POC + NPOC. Each duplicate window represents one analy-
sis type. Parameters that are not applicable to an analysis type will not appear in
the Method Wizard window.
Method Wizard (Page 1) System Information
TOC-VCPH/CPN 135
4.4 Sample Table Editor
4.4.2 File Menu
136 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
TOC-VCPH/CPN 137
4.4 Sample Table Editor
4.4.2 File Menu
138 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
TOC-VCPH/CPN 139
4.4 Sample Table Editor
4.4.2 File Menu
140 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
TOC-VCPH/CPN 141
4.4 Sample Table Editor
4.4.2 File Menu
142 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
4.4.2.4 Open
The Open command is used to select and open a Sample Table, Method, Calibration Curve,
Parameter, or Control Sample Template file. Save the open Sample Table before executing
the Open command. More than one Sample Table can be opened at the same time.
Selecting the Open command displays the Open dialog box shown below.
TIP » The Open dialog box can also be displayed by clicking the Open File button on the Toolbar.
Table 4.2 displays the types of files that can be opened using this function. No other file
types should be opened using this menu command.
Table 4.2 • Acceptable File Types for Open Function
File Extension Description
.t32 Sample Table
.met Method
.cal Calibration Curve
.tpl Control Sample Template
.trd Data Profile
The opened file replaces the currently displayed table or function, depending on the type of
file selected.
Dialog Box Options Description
File name Enter the name of the file to be opened.
Look in Find the file to be opened by choosing the appropriate folder
or accessing other directories. The selected file will appear in
the File name text box.
Files of type Select the type of file from the drop-down list. Choose only
files of type *.t32 when opening a Sample Table.
Open button Opens the selected file.
Cancel button Aborts the Open operation.
TOC-VCPH/CPN 143
4.4 Sample Table Editor
4.4.2 File Menu
4.4.2.5 Close
The Close command is used to close the current window and is an option only when a file is
open. If the open file has not been saved and the Close command is selected, the Save As
dialog box is displayed. Refer to Section 4.4.2.7 "Save As". Enter a name for the file and
click the OK button to save the file. If the open file has been modified but not saved and the
Close command is selected, a dialog box is displayed as shown below. Click the Yes button
to save the file before closing. Click the No button to close the file without saving the
modified information. Click the Cancel button to abort the Close command.
4.4.2.6 Save
The Save command is used to save data in the current active file. If the file has not yet been
saved and the Save command is selected, the Save As dialog box is displayed. Refer to
Section 4.4.2.7 "Save As".
TIP » The Save command can also be executed by selecting the Save File button on the Toolbar.
144 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
4.4.2.7 Save As
The Save As command is used to save the current active file with a new file name.
Selecting the command opens the Save As dialog box shown below. Button options in the
dialog box are similar to those in the Open dialog box. Refer to Section 4.4.2.4 "Open".
Click the Save button to save the data under the specified file name. Click the Cancel
button to abort the Save As operation.
TOC-VCPH/CPN 145
4.4 Sample Table Editor
4.4.2 File Menu
The ASCII Export function is used to export Sample Table data to an ASCII file. Open a
Sample Table that contains completed measurement data. Select ASCII Export from the
File menu. The Save As dialog box is displayed. Refer to Section 4.4.2.7 "Save As". Enter
a file name for the ASCII file to be created and confirm that the “Save as type” field
contains the ASCII file type (*.txt). If the entered file name matches a file that already
exists, a confirmation message is displayed.
Click the Yes button to overwrite the existing file and close the dialog box. Click the No
button to redisplay the Save As dialog box and enter a new file name.
Select the ASCII Export Options command from the File menu. The ASCII Export Options
dialog box, which contains several tabs, is displayed. Each tab contains a list of export
options, as shown in the figures below. Note that on each tab, certain items are selected by
default. Individual options can be selected or deselected by clicking in the check box. Use
the Select All and Select None buttons to quickly select and deselect all options on a tab.
Use the Default button to select the default options on a tab. Click the Cancel button to exit
the ASCII Export Options dialog box without saving changes. Click the OK button to save
changes and exit the ASCII Export Options window.
Header Tab
146 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
System Tab
Data Tab
TOC-VCPH/CPN 147
4.4 Sample Table Editor
4.4.2 File Menu
Miscellaneous Tab
The Miscellaneous tab contains additional options for the format of exported data. To
change data format, open a Sample Table, select File>ASCII Export Options, then select
the file to which changes should be made. On the Miscellaneous tab, select the desired type
of Separator from the drop-down list. The Separator is the type of delimiter that will be
used to separate items of data. Select the “Export strings in quotation marks” option to
enclose all of the character string items in quotation marks.
The Database Export command is used to enable automatic export of data to an external
(Microsoft Access) database during analysis. Selecting the command displays the
Database Options window.
148 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
4.4.2.10 Print>Table
The Print>Table command is used to print the active Sample Table in a table format. Select
this command to open the Print dialog box, which may have a variable display depending
on the installed printer driver. Specify items to be included in the printout by using the Page
Setup command. Refer to Section 4.4.2.15 "Page Setup" for more information.
The Name text field identifies the printer that will be used to print the table. Specify a
different printer by selecting it from the Name drop-down list.
Setting Item Description
Properties Modify the printer settings using standard Windows proce-
dures.
Print range Click the All radio button to print all pages of the table. Click
the Pages radio button and enter the page numbers to be
printed if all pages are not desired.
Number of copies Enter the number of copies to be printed.
TOC-VCPH/CPN 149
4.4 Sample Table Editor
4.4.2 File Menu
The Print>Sample Report command is used to open the Print dialog box to print the Sample
Table in report format. The print dialog box is the same as that described in Section
4.4.2.10 "Print>Table".
The Print Preview>Table option is used to display the sample table before printing. Use the
buttons at the top of the Print Preview window to change the display or print the table, as
described below.
Button Description
Next Page and Previous Page Click the Next Page button to display the next page of the
report. Click the Previous page button to display the previous
page of the report. Note that the Next Page button is disabled
when the last page of the report is displayed, and the Previous
Page button is disabled when the first page of the report is
displayed.
One Page/Two Page Click the One Page button to display one page at a time in the
Print Preview window. After selection, the button toggles to
the Two Page button. Click the Two Page button to display
two pages side by side.
Zoom In and Zoom Out Click the Zoom In and Zoom Out buttons to change the mag-
nification of the displayed pages. Three different zoom fac-
tors are available. Repeated selection of either option will
continue to increase or decrease the magnification, as appli-
cable.
Close Click the Close button to close the Print Preview window and
return to the previous window. Note that the Print Preview
window must be closed to edit the report.
Print Click the Print button to print the report.
150 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
The Print Preview>Sample Report option is used to display the sample report before
printing. Use the buttons at the top of the Print Preview window to change the display or
print the report, as described in Section 4.4.2.12 "Print Preview>Table".
The Print Setup command is used to set print options. Select this command to open the
Print Setup dialog box, which may have a variable display depending on the installed
printer driver. Modifications to the Print Setup dialog box are saved until the TOC-Control
V software is exited.
TOC-VCPH/CPN 151
4.4 Sample Table Editor
4.4.2 File Menu
The Page Setup command is used to select items to be printed on a report or table and to
specify format parameters such as margins, headers, and footers. Selecting the command
displays the Page Setup window, which consists of several tabs. Each tab contains a list of
page setup options shown in the figures below. On each tab, certain options are selected by
default. Individual options can be selected or deselected by clicking in the check box. Use
the Select All and Select None buttons to quickly select and deselect all options on a tab.
Use the Default button to select the default options for a tab. Click the Cancel button to exit
the Page Setup window without saving changes. Click the OK button to save changes and
exit the Page Setup window.
Instrument Tab
The Instrument tab displays a list of instrument information items that can be printed on the
report.
Calibrations Tab
The Calibrations tab displays a list of calibration information items that can be printed on
the report.
152 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
Samples Tab
The Samples tab displays a list of sample information items that can be printed on the
report.
Header Tab
The Header tab displays a list of header items that can be printed on the report.
TOC-VCPH/CPN 153
4.4 Sample Table Editor
4.4.2 File Menu
Miscellaneous Tab
The Miscellaneous tab displays a list of format items for the report, as described below.
154 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.2 File Menu
4.4.2.16 Exit
The Exit command is used to exit the Sample Table Editor. If the active file has been
modified but not saved and the Exit command is selected, a dialog box is displayed as
shown below. Click the Yes button to save the file before exiting. Click the No button to
exit without saving the modified information. Click the Cancel button to abort the Exit
command.
If the software is connected to an instrument and the Exit command is selected, a dialog
box is displayed as shown below. Click the OK button to disconnect the instrument before
exiting the software. Click the Cancel button to abort the Exit command.
TOC-VCPH/CPN 155
4.4 Sample Table Editor
4.4.3 Edit Menu
4.4.3.1 Undo
The Undo function is used to reverse the last change made in the active Sample Table and is
only available after the user changes the content of a cell in the table. The Undo function is
not available during actual measurement or when the user has inserted or deleted completed
samples, calibration curves, or control samples.
4.4.3.2 Cut/Copy/Paste
The Cut, Copy, and Paste functions are used to edit selected Sample Table cells and rows.
The Cut command deletes selected information from rows or cells and stores it on the
clipboard. Note that the Cut command cannot be used to delete entire samples after
measurement is completed. The Copy command copies selected data from rows or cells and
stores it on the clipboard. The Paste command copies the contents of the clipboard to the
selected cells or rows of the table. When pasting an entire row, click the destination row
number to highlight the entire destination row, then execute the Paste command.
Note: • When pasting, do not select a destination row that currently contains data.
The data will be overwritten!
• The Cut, Copy, and Paste commands can also be accessed by clicking the
icons on the Toolbar.
156 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.3 Edit Menu
4.4.3.3 Replace
The Replace function is used to replace text in Sample Table cells with specified text.
Select the Replace function to open the Replace dialog box. The Replace command can be
used to replace data in the Sample Name, Sample ID, Dilution, and Comment fields of the
Sample Table. The command cannot be used to replace data for samples after measurement
is complete.
4.4.3.4 Find
The Find function is used to search for specified text. Selecting the Find command opens
the Find dialog box, which is similar to the Replace dialog box.and displays the options
described below. Refer to Section 4.4.3.3 "Replace".
Setting Item Description
Find what Enter the text to be searched for.
Match case Select to perform a case-sensitive text search.
Direction Click the Up radio button to search the table backwards from
the current location. Click the Down radio button to search
the table forward from the current location.
Find Next Select to begin the search and to continue a search that has
already begun.
Cancel Select to close the Find dialog box.
TOC-VCPH/CPN 157
4.4 Sample Table Editor
4.4.3 Edit Menu
4.4.3.5 Exclude
The Exclude function is used to exclude selected injections from the calculations. A user
must have access rights for updating data to use the Exclude function. To exclude data,
highlight the desired injection shown in the Sample Window Injection Table and select
Exclude from the Edit menu. A confirmation dialog box is displayed. Click the Yes button
to exclude the selected data. Click the No button to abort the operation.
4.4.3.6 Recalculate>All
The Recalculate>All function is used to recalculate the data for all measured samples in a
Sample Table after applying a new calibration curve or excluding injections. To execute the
command, open the appropriate Sample Table and select Recalculate All from the Edit
menu. A message is displayed in the Notify folder of the Notification Bar when the
recalculation is complete.
4.4.3.7 Recalculate>Highlighted
The Recalculate>Highlighted function is used to recalculate the data for selected samples
after applying a new calibration curve or excluding injections. The function is similar to the
Recalculate All function described in Section 4.4.3.6 "Recalculate>All". To execute the
command, open the appropriate Sample Table, highlight the samples to be recalculated, and
select Recalculate Highlighted from the Edit menu.
The Delete Data>All function can only be used if the GLP conformity was disabled during
system installation. The function is used to delete all measurement data in the Sample
Table. The Sample Table structure can then be used to enter new data. A user must have
access rights for updating data to use the Delete Data>All function.
To use the Delete Data>All function, open a Sample Table that contains measurement data
and select Delete Data>All from the Edit menu. A confirmation dialog box is displayed, as
shown in the figure below. Click the Yes button to delete the data. Click the No button to
abort the operation. A message is displayed on the Notifications tab in the Notification Bar
when the data has been deleted.
158 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.3 Edit Menu
The Delete Data>Highlighted function is used to delete the measurement data for selected
samples. To execute the command, open the appropriate Sample Table, highlight the
samples to be deleted, and select Delete Data>Highlighted from the Edit menu. A message
is displayed on the Notifications tab in the Notification Bar when the selected data has been
deleted.
4.4.3.10 Import
The Import function is used to import data from an ASCII file into a Sample Table. Note
that the Import function is not available during sample analysis. To import data, open the
destination Sample Table and place the cursor in the row of the table into which data will be
imported. Select the Import function from the Edit menu. The Open dialog box is
displayed. Select the ASCII file from which data will be imported. The data is inserted in
the current position of the active Sample Table.
TOC-VCPH/CPN 159
4.4 Sample Table Editor
4.4.4 View Menu
The View>Calibration Curve command is used to view the calibration curve information
associated with a sample. Place the cursor in the desired row of the Injection Table and
select the Calibration Curve command from the View menu. The Calibration Curve
Properties window is displayed. Refer to Section 4.4.4.5 "Properties".
TIP » The Calibration Curve Properties window can also be displayed by clicking the
Calibration Curve (View) button on the Toolbar.
160 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.4 View Menu
4.4.4.2 Method
The View>Method command is used to view the properties associated with a method.
Selecting the command displays the Sample/Method Properties window, which consists of
several tabs, including a separate tab for each applicable method. Each tab lists options for
method properties, as shown in the figures below. Individual options can be selected,
deselected, or edited by clicking in the check boxes or entering information into the text
fields. Note that certain options will be disabled if sample analysis is complete. Each tab
contains the same command buttons: OK, Cancel and Apply. Click the OK button to save
changes and exit the Sample/Method Properties window. Click the Cancel button to exit the
Sample/Method Properties window without saving changes. Click the Apply button to save
changes without exiting the Sample/Method properties window.
Common Tab
Setting Item Description
System Displays the system currently in use. Changes cannot be
made to this field.
User Displays the name of the user who created the method.
Changes cannot be made to this field.
Date of Creation Displays the date on which the method was created. Changes
cannot be made to this field.
Comment Enter a comment (512 characters maximum).
TOC-VCPH/CPN 161
4.4 Sample Table Editor
4.4.4 View Menu
Parameter Tab
162 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.4 View Menu
TC Tab
TOC-VCPH/CPN 163
4.4 Sample Table Editor
4.4.4 View Menu
IC Tab
The IC tab options are similar to those described above for the TC tab.
164 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.4 View Menu
TN Tab
The TN tab options are similar to those described above for the TC tab, except that some
options (such as Use area retrieved from blank check) are not available for TN
measurements.
TOC-VCPH/CPN 165
4.4 Sample Table Editor
4.4.4 View Menu
History Tab
If the History Log was enabled when the method was created, the History tab displays a list
of changes made to the method parameters. If the History Log was not enabled previously,
it can be enabled on the History tab.
166 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.4 View Menu
The Sample Window is used to display detailed measurement information. The information
displayed in the Sample Window varies based on the type of analysis performed. To view
the Sample Window, open a Sample Table, place the cursor in the row of the Sample Table
that contains the desired sample, and select Sample Window from the View menu. The
Sample Window is displayed.
TIP » To view measurement data in real time, place the cursor in the row of the Sample Table that
contains the current sample, then open the Sample Window.
The Sample Window is divided into 3 sections: the Graph, the Parameter Table, and the
Injection Table. Note that the sections can be resized by dragging the dividers between
sections.
Graph
The upper left section of the Sample Window displays the peak profile graph. Change the
graph type by choosing from the options on the drop-down list.
Setting Items Description
Current Peak Displays the peak profile of the highlighted injection.
All Peaks Displays all injection peaks for the selected sample.
Calibration curve Displays calibration curves (for calibration standards only).
This function is not available during real-time analysis
Note: To view the peak profile of the measurement in progress, click the row with the
asterisk in the Sample No. column. To view a previous measurement, click the
desired row.
Parameter Table
The upper right section of the Sample Window displays sample parameters.
Setting Item Description
Sample Name Displays the name of highlighted sample.
Sample ID Displays the sample identification from the Sample Table.
Analysis Displays the analysis method used for the sample.
Result Displays the sample concentration.
Dilution Factor Displays the dilution factor. The result will be multiplied by
this factor to calculate the final concentration. If the sample
was not diluted, do not change the default value of 1.
TOC-VCPH/CPN 167
4.4 Sample Table Editor
4.4.4 View Menu
Injection Table
The lower section of the Sample Window displays the Injection Table, which contains
detailed information about the injections performed for the selected sample. Asterisks are
displayed in the Injection Table for the injection that is currently being measured. Some
columns in the Injection Table use the Merge Cell function to display information that is
common to all the injections of a particular sample. If the sample has been analyzed by
more than one analysis method, the Injection Table contains several folders displayed as
tabs at the bottom of the window. The columns in the Injection Table are described below.
Setting Items Description
Spl. No. Displays the sample number.
Inj. No. Displays the injection number.
Area Displays the area value.
Mean area Displays the mean area value for all injections of the sample.
Conc. Displays the concentration obtained for the injection.
Mean Conc. Displays the mean concentration obtained for all injections of
the sample.
Result Displays the analysis result.
Notes Displays any analysis flags or notations associated with that
sample.
SD Area Displays the standard deviation calculated from the area.
CV Area Displays the coefficient of variation calculated from the area.
SD Conc. Displays the standard deviation calculated from the concen-
tration.
CV Conc. Displays the coefficient of variation calculated from the con-
centration.
Status Displays the analytical status, either “measuring” or “fin-
ished”.
Vial Displays the vial number (for ASI samples only).
Excluded Displays the exclusion flag, denoted as “E” for user exclusion
of the sample, “ES” for exclusion by the system, or “Not
Excluded”.
Modified Displays the modification flag of “M” if the sample was
modified.
Inj. Volume Displays the injection volume for each injection.
Dilution Factor Displays the dilution factor used in measurement.
Calibration curve Displays the calibration curve used.
168 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.4 View Menu
The Outlier Test (also known as the Grubb test) is used to determine whether a selected
injection is an outlier. The test also determines whether the injection with the highest
deviation from the mean value is still within an acceptable range. The Outlier Test can only
be performed if the sample contains at least three non-excluded injections.
To perform the Outlier Test, open the Sample Window and select the cells containing the
data to be tested. From the View menu, select Outlier Test. The Outlier Test window is
displayed. Select either Conc. or Area (CNV) for the data source. Exclude the injection that
is suspected of being an outlier, then select OK to close the dialog box.
If the sample does not contain enough injections for the test to be performed, an error
message is displayed and the Outlier Test window will not open.
The Outlier Test window contains the parameters listed below.
Setting Items Description
Result List Displays the result of the Outlier Test.
Conc. radio button Select to indicate that test calculations are to be performed
based on the concentration data.
Area (CNV) radio button Select to indicate that test calculations are to be performed
based on the area data.
Result Displays the result of the test.
Exclude button Click this button to exclude a suspect sample and recalculate
the test result.
Suspicious inj. Indicates the injection suspected of being an outlier, which is
identified as the sample with the highest deviation from the
mean value.
SD Displays the standard deviation for all of the injections.
CV Displays the coefficient of variation for all of the injections.
Grubb Value Displays the corresponding value from the Grubb table.
Test Value Displays the calculated Outlier Test value.
TOC-VCPH/CPN 169
4.4 Sample Table Editor
4.4.4 View Menu
4.4.4.5 Properties
The View>Properties function displays various windows. The display depends on the
content of the cell selected in the Sample Table. If a range of cells is selected, the upper left
cell is considered the active cell. The table below shows the possible selections and the
corresponding Properties windows.
Table 4.4 • Properties Windows
Content of Active Cell Properties Window
Calibration Curve Calibration Curve Properties window
Method Sample/Method Properties window
Any other cell for a sample Sample/Method Properties window
Any other cell of a control Control Sample Properties window
sample
The Properties window contains options for viewing and modifying the method, calibration
curve, and sample properties. Each Properties window consists of several tabs. The
Calibration Curve Properties window tabs are shown and described below.
Calibration Curve Properties Window
Common Tab
The Common tab of the Calibration Curve properties window contains the following
information about the calibration curve.
Setting Item Description
System Displays the system used to create the calibration curve. This
field cannot be edited.
User Displays the name of the user who created the calibration
curve.
Date of Creation Displays the date and time the calibration curve was created.
Comment A comment regarding the calibration curve may be entered
here.
Print Prints the calibration curve information.
170 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.4 View Menu
Parameter Tab
TOC-VCPH/CPN 171
4.4 Sample Table Editor
4.4.4 View Menu
Analysis Tab
The Analysis tab displays additional analysis parameters.
Analysis Parameters
Setting Item Description
No. of Washes Displays the number of times (0-10) the syringe is washed with standard
prior to the first injection. Note that if the Auto Dilution option was
selected, the standard will be diluted and the syringe will then be washed
with the diluted standard. This option is not available for SSM. This field
can be edited.
Acid Addition Displays the quantity of acid to be added to each standard prior to injec-
tion. This value is expressed as a percent of the sample volume and is
available only for NPOC or related analyses. This field can be edited.
Sparge Time Displays the pre-sparge time (0-20 minutes). This option is only avail-
able when the external sparge kit is installed and the analysis type is
NPOC. This field can be edited.
Density: Displays the sample density. This option is available only for SSM.
By Weight/Volume: Displays the basis (Weight or Volume) for sample calculations. This
option is available only for SSM.
172 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.4 View Menu
Data Tab
The Data tab displays information about each calibration standard used in the calibration
curve.
TOC-VCPH/CPN 173
4.4 Sample Table Editor
4.4.4 View Menu
Columns of data are displayed in a window below the Inj. Volume selection, as described
below.
Table 4.5 • Data Tab Columns
Column Name Description
Auto Dilution Displays the dilution factor for auto dilution.
Standard Solution Concentra- Displays the concentration of the standard solution used for
tion analysis.
No. Displays the calibration point number.
Conc. Displays the concentration of calibration point.
Mean Area Displays the mean area value from all non-excluded injec-
tions
No. of Injections Displays the minimum and maximum number of injections.
Excluded “E” indicates calibration point has been excluded.
SD Max Displays the maximum standard deviation.
CV Max Displays the maximum coefficient of variation.
SD Displays the standard deviation.
CV Displays the coefficient of variation.
Additional buttons are displayed, below the data columns and can be used to manipulate the
calibration parameters. The buttons are described below. Refer to Section 4.4.2.2
"Calibration Curve" for more detailed descriptions.
Button Function
Edit Edit the parameters for the selected calibration point.
Add Add a new calibration point.
Delete Delete the selected calibration point.
Delete All Delete all calibration points.
Exclude Exclude the selected calibration point.
174 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.4 View Menu
Graph Tab
The Graph tab displays the calibration curve in a graphic format.
TOC-VCPH/CPN 175
4.4 Sample Table Editor
4.4.4 View Menu
History Tab
If the History Log was enabled when the calibration curve was created, the History tab
displays a list of changes made to the calibration curve parameters. If the History Log was
not enabled previously, it can be enabled on the History tab.
176 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.4 View Menu
4.4.4.6 Customize
The View>Customize window contains two tabs, one that displays or hides the toolbars and
one that customizes the toolbar content.
Toolbars Tab
Select the desired toolbars from the list. All items are selected by default.
Option Description
Show Tooltips Displays tooltips when the mouse cursor is positioned over
the toolbar button.
Cool Look Select this option to display the toolbar buttons as flat but-
tons. This item is selected by default.
Large Buttons Select this option to display the toolbar buttons in a larger
size.
New button Click this button to create a user-defined toolbar button.
Reset/Delete button Click this button to reset the toolbar display to its original
state or delete a user-defined toolbar button.
TOC-VCPH/CPN 177
4.4 Sample Table Editor
4.4.4 View Menu
Command Tab
Option Description
Categories Displays the available toolbars.
Buttons Displays the buttons in the selected toolbar. The buttons can
be moved by dragging them to another toolbar.
Description Displays a description of the selected button.
4.4.4.7 Toolbar
The View>Toolbar function is used to select or deselect individual toolbars for display.
Toolbar display options (such as button size or appearance) can also be set using this
command.
The View>Status Bar function toggles the display state of the status bar between Show and
Hide. If the status bar is currently displayed, selecting this option will hide the status bar. If
the status bar is not displayed, selecting this option will display the status bar.
178 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.4 View Menu
This function displays the ASI or 8-Port Sampler dialog box, and allows the user to enter
vial numbers for measurement using the ASI or the 8-Port Sampler. To access this
command, either create or open a sample table, and select View > ASI / 8-Port Sampler
Vials. The Sparge / Acid Addition dialog box opens automatically when the Start command
is made. The OK button saves the changes and closes the dialog box.
Using the Autosampler
Figure 4.134 Sparge / Acid Addition Dialog Box for the ASI-V Autosampler
Figure 4.135 Sparge / Acid Addition Dialog Box for One 8-Port Sampler
Figure 4.136 Sparge / Acid Addition Dialog Box for Two 8-Port Samplers
TOC-VCPH/CPN 179
4.4 Sample Table Editor
4.4.4 View Menu
When the mouse is positioned over a vial in the drawing of the of ASI or 8-Port Sampler, a
tool-tip showing the vial number is displayed. Click the vial to display the following items.
Parameter Description
Vial No. Displays the currently selected vial number.
Acid Addition rate Displays the acid addition rate for each vial.
Sparge time Displays the sparge time for each vial.
Total injection volume Displays the total injection volume for each vial.
TIP » The desired vial position for a sample can be entered either directly into the Vial column or
by placing the cursor in the appropriate row of the Sparging/Acid Addition window and
then double-clicking the desired position in the tray diagram.
Displays the raw measurement data. The data profile is never modified or processed
following the initial collection. It includes the corresponding measurement parameters. The
files which were created and named using the object ID, each occupy a row in the sample
table, and are stored within the RawData folder in folders named using the date of creation.
The View>Data Profile function is used to view the data profile for a sample. Highlight the
desired row of the Sample Table, then select the command.
180 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.5 Insert Menu
The Auto Generate command opens the Sample Group Wizard, which is used to create a
group of unknown samples, control samples, and calibration standards, and then insert the
group into the active position of the Sample Table.
Sample Group Wizard (Page 1) Sample Source
The Sample Source window is used to specify the source of sample measurement
parameters.
TOC-VCPH/CPN 181
4.4 Sample Table Editor
4.4.5 Insert Menu
182 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.5 Insert Menu
TOC-VCPH/CPN 183
4.4 Sample Table Editor
4.4.5 Insert Menu
Control Templates
The Control Templates table displays a list of the control sample templates currently
assigned to the sample group. The templates specify measurement parameters, evaluation
criteria, and failure actions for the control samples. This table is empty by default. Use the
options listed below to revise the list of control sample templates associated with the
sample group.
Option Description
Add Select this option to add a control template to the sample
group.
Delete Select this option to delete a highlighted control template
from the list.
After all four pages of the Sample Group Wizard are complete, click the Finish button to
save changes and add the sample group to the Sample Table. If invalid parameter values
were entered, an error message will be displayed and the sample group will not be inserted
into the Sample Table until after the error is corrected.
184 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.5 Insert Menu
4.4.5.2 Sample
The Sample command opens the Sample Wizard, which is used to insert sample
measurement parameters into an existing Sample Table. The Sample Wizard is not
accessible during an analysis.
Sample Wizard (Page 1) Parameter Source
The Parameter Source window is used to specify the source of the sample measurement
parameters. The source can either be a method file or a calibration file. When one of these
sources is selected, the sample will be analyzed using the same measurement parameters as
those in the specified method or calibration file. Some of the source parameters can be
edited in subsequent pages of the Sample Wizard. Note that a sample can also be added to
the Sample Table without specifying a source, an option that is desirable when a sample is
to be analyzed using different parameters than those specified in the method or calibration
files. In this case, unique measurement parameters for the sample will be entered on
subsequent pages of the Sample Wizard.
Parameter Description
Method radio button Select this option to use the sample measurement parameters from a specified
method. This is the default option. Enter the method name in the text field, or
use the Browse button to select the file using the File>Open dialog box
Calibration Curve radio Select this option to use the measurement parameters from a specified cali-
button bration curve. Enter the calibration curve name in the text field, or use the
Browse button to select the file using the File>Open dialog box.
Edit parameters manually Select this option to insert a sample without defining a source. Sample mea-
radio button surement parameters will be entered in subsequent pages (1-12) of the Sam-
ple Wizard.
Skip remaining Wizard Select this option to accept all measurement parameters from the source file.
pages and use measure- The Next button will become a Finish button, and subsequent pages of the
ment parameters from the Sample Wizard will not be displayed. This option is disabled if Edit parame-
source ters manually is selected.
TOC-VCPH/CPN 185
4.4 Sample Table Editor
4.4.5 Insert Menu
Parameter Description
Analysis Select the analysis type from the drop-down list.
Default Sample Name Enter the sample name.
Default Sample ID Enter the sample identification.
Dilution Enter the preparation dilution factor.
No. of Determinations Use the spin controls to enter the number of times the sample
is to be measured. Note that determinations and injections are
different; each determination is made up of one or more
injections. Enter an integer from 1 to 10.
Note: For combined analyses such as TC/TN or POC+NPOC, subsequent Sample Wiz-
ard windows may appear more than once so that parameter modifications spe-
cific to each analysis type can be made. The analysis type that is associated with
each Sample Wizard window can be found in the Analysis box at the top of the
window.
186 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.5 Insert Menu
Parameter Description
Analysis Displays the analysis type. This field cannot be changed.
Calibration Curve 1/2/3 Enter the name of the calibration curve or click the Browse
button to select a file using the File>Open dialog box.
Skip the injection parameter Select this item to skip the Measurement Parameters window
pages of the wizard of the Sample Wizard. Use this option when the sample mea-
surement parameters will be the same as those used for cali-
bration curve measurements.
TOC-VCPH/CPN 187
4.4 Sample Table Editor
4.4.5 Insert Menu
Parameter Description
Analysis Displays the analysis type. This field cannot be changed.
Units Enter the concentration units.
Injection Volume Enter the injection volume.
Expected Conc. Range Enter the maximum expected sample concentration.
No. of Injections Enter the minimum and maximum number of injections for
the sample.
SD Max Enter the maximum standard deviation. If this value is met,
no additional injections are required. However, if both this
value and the CV Max value are exceeded, the sample is
automatically reinjected up to the maximum number of times
specified in the No. of Injections field. The SD Max field is
disabled if the No. of Injections is 1 or if the minimum is
equal to the maximum.
CV Max Enter the maximum coefficient of variation. If this value is
met, no additional injections are required. However, if both
this value and the SD Max value are exceeded, the sample is
automatically reinjected up to the maximum number of times
specified in the No. of Injections field. The CV Max field is
disabled if the No. of Injections is 1 or if the minimum is
equal to the maximum.
No. of Washes Enter the number of times the syringe is to be washed with
sample prior to the first injection. Note that if the Auto Dilu-
tion option was selected, the sample will be diluted and the
syringe will then be washed with the diluted sample.
188 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.5 Insert Menu
Parameter Description
Auto Dilution Enter a dilution factor for the sample. The instrument will
automatically dilute the sample by this factor, and the result
will be multiplied by the factor to obtain the final concentra-
tion. The default Auto Dilution value is 1.
Sparge Time Enter the sparge time.
Acid Addition Enter the desired quantity of acid to be added. This value is
expressed as a percent of the sample volume and is available
only for NPOC or related analyses.
Multiple Injections Select this option to enable the syringe to draw up a single
sample aliquot with a volume equal to that required for all
injections of the sample. When this option is not selected, the
syringe will draw up a separate sample aliquot for each injec-
tion. Selecting this option reduces analysis time.
Use blank check area Select this option to subtract the blank check area from the
measured area of the sample. This option is only available for
TC and NPOC analyses that use the high sensitivity catalyst.
Auto. correction of inj. vol. Select this option to enable the instrument to automatically
and dilution adjust the injection volume and dilution factor when the peak
height for the initial injection exceeds full scale. After the
adjustment, measurement will be repeated.
TOC-VCPH/CPN 189
4.4 Sample Table Editor
4.4.5 Insert Menu
Parameter Description
Analysis Displays the analysis type. This field cannot be changed.
Use default settings Select this option to use the default integration parameters.
When this option is selected, all other options in this window
are disabled.
Allowed peak start delay The allowed peak start delay is the amount of time (in min-
utes) that can elapse before the analysis is stopped because a
peak is not detected. Enter a number between 0 and 20:00.
Total integration time The total integration time is the amount of time (in minutes)
that analysis continues after the end of the peak is detected.
Enter a number between 4:50 and 20:00.
190 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.5 Insert Menu
Figure 4.147 Sample Wizard (Page 6) Calibration Curve Window (Second Analysis)
Figure 4.148 Sample Wizard (Page 7) Injection Parameters Window (Second Analysis)
TOC-VCPH/CPN 191
4.4 Sample Table Editor
4.4.5 Insert Menu
Figure 4.149 Sample Wizard (Page 8) Peak Time Parameters Window (Second Analysis)
192 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.5 Insert Menu
4.4.5.3 Control
The Control command is used to insert a control sample into an existing Sample Table.
Note: A control sample template must be created before a control sample can be
inserted into a Sample Table. A control sample template has a file extension of
*.tpl and specifies measurement parameters for the Control Sample.
The following subsections describe the procedures for creating a control sample template
and for inserting a control sample into a Sample Table.
Parameter Description
System Select the instrument.
Operator Displays the current operator name. This field cannot be
changed.
Date of Creation Displays the current date and time. This field cannot be
changed.
Comment Enter a comment (512 characters maximum).
TOC-VCPH/CPN 193
4.4 Sample Table Editor
4.4.5 Insert Menu
Parameter Description
Control Sample Select this option to define individual sample parameters.
This is the default option.
Control Chart Sample Select this option to add the control sample to a defined con-
trol chart. Enter the file name or use the Browse button to
select the control chart file from the File>Open dialog box.
System Suitability Test for Select this option to check the control sample using USP/EP
USP/EP criteria.
Save the data in a control Enter a file name for the control sample template.
sample template
194 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.5 Insert Menu
Parameter Description
Method radio button Select this option to use the measurement parameters from a
specified method. Enter the method name in the text field, or
use the Browse button to select the file using the File>Open
dialog box. This is the default option.
Calibration Curve radio but- Select this option to use the measurement parameters from a
ton specified calibration curve. Enter the calibration curve name
in the text field, or use the Browse button to select the file
using the File>Open dialog box.
Edit parameters manually Select this option to insert a control sample without defining
radio button a source. Control sample measurement parameters will be
entered in subsequent pages of the Control Sample Wizard.
Skip remaining Wizard pages Select this option to accept all measurement parameters from
and use measurement param- the source file. The Next button will become a Finish button,
eters from the source and subsequent pages of the Control Sample Wizard will not
be displayed. This option is disabled if Edit parameters man-
ually is selected.
TOC-VCPH/CPN 195
4.4 Sample Table Editor
4.4.5 Insert Menu
Parameter Description
Analysis Select the analysis type from the drop-down list. If a source
method or calibration file was specified on Page 3 of the
Control Sample Wizard, this field is disabled and the applica-
ble analysis type is displayed.
Default Sample Name Enter the sample name. If a source method or calibration file
was specified on Page 3 of the Control Sample Wizard, the
default sample name from the source is displayed.
Default Sample ID Enter the sample identification. If a source method or calibra-
tion file was specified on Page 3 of the Control Sample Wiz-
ard, the default sample identification from the source is
displayed.
Dilution Enter the dilution factor. If a source method or calibration file
was specified on Page 3 of the Control Sample Wizard, the
default dilution factor is displayed.
Calibration Curve Enter the calibration file name that is to be used to calculate
control sample concentration. If a source method or calibra-
tion file was specified on Page 3 of the Control Sample Wiz-
ard, the default calibration curve name from the source is
displayed.
196 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.5 Insert Menu
Parameter Description
Analysis Displays the analysis type. This field cannot be changed.
Units Select the concentration units.
Injection Volume Enter the injection volume.
Expected Conc. Range Enter the maximum expected concentration.
No. of Injections Enter the minimum/maximum number of injections for the control sample.
The setting range is from 1-20.
SD Max Enter the maximum standard deviation that is acceptable for the injections. If
this value is met, no additional injections are required. However, if both this
value and the CV Max value are exceeded, the control sample is automatically
reinjected up to the maximum number of times specified in the No. of Injec-
tions field. The SD Max field is disabled if the No. of Injections is 1 or if the
minimum is equal to the maximum.
CV Max Enter the maximum coefficient of variation that is acceptable for the injec-
tions. If this value is met, no additional injections are required. However, if
both this value and the SD Max value are exceeded, the control sample is
automatically reinjected up to the maximum number of times specified in the
No. of Injections field. The CV Max field is disabled if the No. of Injections is
1 or if the minimum is equal to the maximum.
No. of Washes Enter the number of times the syringe is to be washed with sample prior to the
first injection. Note that if the Auto Dilution option was selected, the sample
will be diluted and the syringe will then be washed with the diluted sample.
Auto Dilution Enter a dilution factor for the samples. The instrument will automatically
dilute the control sample by this factor, and the result will be multiplied by the
factor to obtain the final concentration. The default Auto Dilution value is 1.
Acid Addition Enter the desired quantity of acid to be added. This option is available only for
NPOC and related analyses.
Sparge Time Enter the sparge time.
TOC-VCPH/CPN 197
4.4 Sample Table Editor
4.4.5 Insert Menu
Parameter Description
Multiple Injections Select to enable the syringe to draw up a single sample aliquot with a volume
equal to that required for all injections of the sample. When this option is not
selected, the syringe will draw up a separate sample aliquot for each injection;
as a result, selecting this option reduces analysis time.
Use area retrieved from Select this option to subtract the blank check area from the measured area of
the blank check the control sample. This option is only available for TC and NPOC analyses
that use the high sensitivity catalyst.
198 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.5 Insert Menu
TOC-VCPH/CPN 199
4.4 Sample Table Editor
4.4.5 Insert Menu
200 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.5 Insert Menu
Save the control sample template and close the Control Sample Wizard by clicking the
Finish button.
TOC-VCPH/CPN 201
4.4 Sample Table Editor
4.4.5 Insert Menu
Procedure
1. Open the Sample Table and place the cursor in the row where the control sample will
be inserted.
2. From the Insert menu, select Control.
The Open dialog box displays the available control sample template files.
3. Select the appropriate control sample template and click the Open button.
The Open dialog box closes and the Sparging/Acid Addition window is displayed.
4. Enter the ASI vial position for the control sample into the appropriate cell in the
Sparging/Acid Addition window.
5. Click the OK button.
The Sparging/Acid Addition window closes.
6. Scroll through the Sample Table to verify that the control sample has been inserted at
the correct location.
202 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.6 Instrument Menu
The Background Monitor command is used to open the Background Monitor window,
which displays the status of the autosampler and all installed instrument detectors. The real
time signal and ready status for various detectors can be viewed by toggling between the
individual instrument tabs. The command is available only when an instrument is
connected to the software. Instrument and communication errors are displayed in the
Notification Bar. The Background Monitor window cannot be open during measurement.
TOC Tab
Parameter Description
Furnace Displays the temperature of the TC furnace.
Dehumidifier Displays the temperature of the dehumidifier.
NDIR Baseline Pos.: Displays the status of the NDIR detector baseline position.
Baseline Fluc.: Displays the status of the NDIR detector baseline fluctuation.
Baseline Noise: Displays the status of the NDIR detector baseline noise.
The bottom portion of the TOC tab window displays a graph of the selected items.
TOC-VCPH/CPN 203
4.4 Sample Table Editor
4.4.6 Instrument Menu
TN Tab
Parameter Description
TN Reactor Displays the temperature of the TN reactor.
Chemiluminescence Baseline Pos.: Displays the status of the chemiluminescence detector baseline
position.
Baseline Fluc.: Displays the status of the chemiluminescence detector baseline
fluctuation.
Baseline Noise: Displays the status of the chemiluminescence detector baseline
noise.
The bottom portion of the TN tab window displays a graph of the selected items.
ASI Tab
The ASI tab is available only if the ASI is connected to the instrument. The ASI Power
indicator displays the power status of the ASI. The ASI Cover indicator displays the status
of the ASI cover.
204 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.6 Instrument Menu
4.4.6.2 Connect
The Connect command is used to connect or disconnect the instrument and the software. If
the instrument is not connected, selecting the Connect command opens the communication
port, initializes the instrument, and downloads and checks the instrument parameters.
During this process, the Sequence window displays the progress of the connection. To abort
the Connect operation while it is in progress, click the Cancel button.
The bottom portion of the Sequence window displays a list of tasks that are performed
during the Connect process. The progress of each task and the result after completing the
task are shown in columns to the right of the task list. The Sequence window closes when
the connection process is complete. If an error occurs during the Connect process, an error
message is displayed in the Notification Bar and the result column, and the Sequence
window remains open.
If the instrument is already connected to the software, the Connect command is used to
disconnect the instrument. A confirmation request dialog box is displayed when the
command is selected, as shown below.
Click the OK button to disconnect the instrument, or the Cancel button to abort the
operation.
TOC-VCPH/CPN 205
4.4 Sample Table Editor
4.4.6 Instrument Menu
4.4.6.3 Standby
The Standby function is used to shut down the instrument. To access the Standby function,
open a Sample Table, then select Standby from the Instrument menu. The Standby window,
which lists options for the shut down operation, is displayed. Select the desired options and
click the Standby button. Click the Escape button to abort the Standby operation and
maintain the instrument in Ready mode.
CAUTION
Do not turn off the TOC-V main instrument switch for at least 30 minutes after
initiating the shut down procedure.
206 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.6 Instrument Menu
Click the Measurement Start command to open the Standby dialog box if the ASI is
included in the system configuration. If the ASI is not included, the Measurement Mode
dialog box is displayed. To start measurement, open the Sample Table, define a new
sample, and connect to the instrument.
The Stop>Peak Stop command is used to interrupt the current measurement. The peak
measured up to the stopping point is considered a complete peak. After the Stop>Peak Stop
command is executed, the next injection listed in the Sample Table is made and analysis
continues normally.
The Stop>Finish Current Sample command is used to interrupt measurement after all
injections of the current sample are complete. After the Stop>Finish Current command is
executed, the instrument remains in Ready mode.
4.4.6.7 Stop>Halt
The Stop>Halt command is used to immediately interrupt the current measurement, discard
data from the current injection, and return the instrument to Ready mode.
TOC-VCPH/CPN 207
4.4 Sample Table Editor
4.4.7 Maintenance Menu
The Control Charts function is used to create control charts. Refer to Appendix B for a
detailed description of the function.
The Method Validation function is used to validate analytical methods. Refer to Appendix
A for a detailed description of the function.
4.4.8.3 Administration
The Administration function is used to access the Administration program of the TOC-
Control V software. Refer to Section 3.1 "Introduction to System Administration
Functions" for a detailed description of the Administration program.
208 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.8 Tools Menu
If the user must temporarily leave the PC he/she is logged into, the screen can be locked by
using the password function to prevent another individual from performing operations in
TOC-Control V. This function can be used only if the User ID / Password item is checked
in the Security dialog box.
To lock the screen, select Screen Lock in the Tools menu. The screen will then be locked
and the Re-enter Password dialog box is displayed. To release the screen lock, enter the
user ID and password into the designated fields. If the user ID and password entered are not
the same as those used up to that point, the screen lock will not be released.
TOC-VCPH/CPN 209
4.4 Sample Table Editor
4.4.9 Options Menu
The General Information command is used to view information about the system. The
General Information window contains two tabs, one that displays System parameters and
one that displays History information. Most of the fields in the General Information
window are disabled because they were set previously. After changing an active field, click
the OK button to close the window and save the changes or click the Apply button to save
the changes without closing the window. Click the Cancel button to close the window
without saving changes.
System Tab
The System tab displays general parameters.
Parameter Description
System Displays the instrument identification.
File Version Displays the instrument file version.
User Displays the name of the user who set up the system.
Date of Creation Displays the date the system was set up.
Comment Enter a comment (512 characters maximum), if desired.
210 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.9 Options Menu
History Tab
The History tab is used to view the system’s History Log or to enable the log if it was not
previously selected.
Note: Once the History Log is enabled, it cannot be disabled.
TOC-VCPH/CPN 211
4.4 Sample Table Editor
4.4.9 Options Menu
The Default Measurement Parameters command is used to enter the default values for
samples, methods, and calibration curves. The Default Measurement Parameters window
contains a separate tab for each analysis type, into which the default values are entered. The
SSM-TC and SSM-IC tabs are items associated with the installation of SSM-5000A. For
details, refer to the SM-5000A User's Manual
The parameters shown on each of the tabs are similar and are described below.
Parameter Description
Units Select the default concentration units.
No. of Injections Enter the default minimum/maximum number of injections.
No. of Washes Enter the default number of times the syringe is to be washed
with sample prior to the first injection. Note that if the Auto
Dilution option was selected, the sample will be diluted and
the syringe will then be washed with the diluted sample.
212 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.9 Options Menu
Parameter Description
Max SD Enter the default maximum standard deviation that is accept-
able for the injections. If this value is met, no additional
injections are required. However, if both this value and the
Max CV value are exceeded, the samples are automatically
reinjected up to the maximum number of times specified in
the No. of Injections field. The SD Max field is disabled if
the No. of Injections is 1 or if the minimum is equal to the
maximum.
Max CV Enter the default maximum coefficient of variation that is
acceptable for the injections. If this value is met, no addi-
tional injections are required. However, if both this value and
the Max SD value are exceeded, the samples are automati-
cally reinjected up to the maximum number of times speci-
fied in the No. of Injections field. The CV Max field is
disabled if the No. of Injections is 1 or if the minimum is
equal to the maximum.
The Display Settings>Display Font function is used to select font options for both the
general screen display (such as in the Sample Table and Print Preview screens) and printed
reports. Select a font type, style, size, and color from the lists in the Font dialog box. A
preview of the font appearance is displayed in the Sample field.
TOC-VCPH/CPN 213
4.4 Sample Table Editor
4.4.9 Options Menu
The Display Settings>Table Settings function is used to select the parameters that will be
displayed in the Sample Table window. The Table Options window contains two tabs, one
that lists Sample Table parameters and one that lists Injection Table parameters. Note that
on each tab, certain options are selected by default. Individual options can be selected or
deselected by clicking in the check box. Use the All and None buttons to quickly select and
deselect all options on a tab. Use the Default button to select the default options.
Sample Table Tab
Select the parameters that will be displayed in the Sample Table.
Samples Tab
Select the parameters that will be displayed in the Injection Table.
214 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.9 Options Menu
The Display Settings>Notification Bar Settings function is used to set options for the
Notification Bar, where error messages and other notices are displayed. The Notification
Bar is located at the bottom of the Sample Table Editor window.
Four folders can be displayed in the Notification Bar: Notifications, System Folders,
Errors, and Events. When a folder is selected for display, a tab appears for that folder in the
Notification Bar. When more than one folder is selected for display, the contents of a
specific folder can be displayed by clicking the appropriate tab.
To select a folder for display in the Notification Bar, click the folder in the Folder List, then
select the Display Folder option. Set the Max. Number of Displayed Lines field to a value
greater than 0. To change the color or appearance of the text displayed in the folder, use the
Font option. Note that the Display Folder check box, Max. Number of Displayed Lines
field, and Font option apply only to the selected folder; as a result, these options must be set
separately for each folder.
Note: A folder will not be displayed in the Notification Bar unless the Display Folder
option is selected and the Max. Number of Displayed Lines field is set to a value
greater than 0.
TOC-VCPH/CPN 215
4.4 Sample Table Editor
4.4.9 Options Menu
The Display Settings>Floating Point Number Formats command is used to select display
options for numeric results in the Sample Table. Note that these options apply only to the
manner in which numbers are displayed; internal data retain their original values.
Select settings for both common numbers and percent values, as described below.
Common Floating Point Numbers
Setting Item Description
Number of significant digits Select this option to display common floating point numbers
using a specified number of significant digits. Use the spin
control to set the number of significant digits. The default
value is 4.
Number of digits after the Select this option to display common floating point numbers
decimal point using a specified number of digits after the decimal point.
Use the spin control to set the number of digits after the deci-
mal point. The default value is 2.
Percent Values
Setting Item Description
Number of significant digits Select this option to display percent values using a specified
number of significant digits. Use the spin control to set the
number of significant digits. The default value is 4.
Number of digits after the Select this option to display percent values using a specified
decimal point number of digits after the decimal point. Use the spin control
to set the number of digits after the decimal point. The default
value is 2.
216 TOC-VCPH/CPN
4.4 Sample Table Editor
4.4.9 Options Menu
4.4.9.7 Directories
Delete
Click the Delete button to delete a selected directory from the list.
Up
Click the Up button to move the selected directory up one line in the list. The order of the
directories in the list is the order in which the directories will be searched during instrument
operation.
Down
Click the Down button to move the selected directory down one line in the list. The order of
the directories in the list is the order in which the directories will be searched during
instrument operation.
TOC-VCPH/CPN 217
4.4 Sample Table Editor
4.4.10 Window Menu
4.4.10.1 Cascade
The Cascade command is used to stack all open windows so that the title bars are visible.
4.4.10.2 Tile
The Tile command is used to size all open windows so that each window occupies an equal
portion of the screen.
The Arrange Icons command is used to neatly organize windows that have been minimized
or reduced to icons.
About
The About command is used to view copyright, software registration, and software version
information.
218 TOC-VCPH/CPN
4.5 Application Operations
4.5.1 Manual Injection Analysis
Knurled Thumbscrew
TC/TN Block
TOC-VCPH/CPN 219
4.5 Application Operations
4.5.1 Manual Injection Analysis
3. If the IC reactor kit is being used, remove the IC injector slider, and install the IC
block provided in the manual injection kit.
IC Block
Note: When installing the block, be sure that it is aligned with the
injection port.
Software Settings
Select the Manual Injection Kit option in the Instrument Setup Wizard Options window
during System Setup. Refer to Section 3.2.2 "Options Window".
Note: Selecting the Manual Injection Kit option disables operation of the TOC-V injec-
tor syringe.
220 TOC-VCPH/CPN
4.5 Application Operations
4.5.1 Manual Injection Analysis
4.5.1.2 Measurement
Measurement procedures using the manual injection kit are similar to those described in
Section 4.3 "Measurement Procedures and Tutorial".
Procedure
1. From the File menu, select New. The New dialog box is displayed.
2. Double-click the Calibration Curve icon. The Calibration Curve Wizard opens. The
System Information window of the wizard is displayed.
3. From the System drop-down list, select the system that was configured with the
Manual Injection Kit during setup.
4. Select the desired options on pages 2 and 3 of the Calibration Curve Wizard, which
list calibration curve type and analysis information settings, respectively. Enter the
name of the new calibration curve in the text box at the bottom of page 3 of the
wizard.
Note: Select the settings described in Section 4.2 "Setting General Mea-
surement Parameters".
TOC-VCPH/CPN 221
4.5 Application Operations
4.5.1 Manual Injection Analysis
5. Enter the desired number of injections on page 4 of the Calibration Curve Wizard
(Calibration Measurement Parameters).
6. Enter the injection volume on page 5 of the Calibration Curve Wizard (Calibration
Points List).
Note: An injection volume is displayed automatically. The displayed
value is based on the volume of the last injection that was per-
formed using the sample injector. This value can be changed by the
user as appropriate.
7. Select the desired options on pages 6 and 7 of the Calibration Curve Wizard, which
list peak time parameter and history settings, respectively. Click the Finish button to
create the new calibration curve.
Note: Select the settings described in Section 4.2 "Setting General Mea-
surement Parameters".
8. Create a Sample Table for the calibration curve. From the Insert menu, select
Calibration Curve. Select the newly created calibration curve from the list of files in
the Open dialog box.
9. Connect the instrument by selecting Connect from the Instrument menu. Then select
Start from the Instrument menu. The Manual Injection Parameter dialog box is
displayed.
10. Draw an appropriate amount of sample into the syringe.
Note: Draw in and discharge the sample 3 or 4 times before filling the
syringe for measurement.
11. Enter the injection volume in the Manual Injection Parameter dialog box, and click
the Set button.
12. Verify that the instrument is ready for measurement, and click the Start button.
13. For TC or TN measurement, open the injection port valve and insert the syringe
needle into the injection port. For IC measurement, insert the syringe needle into the
injection port septum.
Note: • Insert the syringe needle as far as possible into the injection port.
• To obtain reproducible results, perform all injections at the same
speed.
14. Withdraw the syringe from the injection port. For TC or TN measurement, close the
injection port valve.
The results are displayed when measurement is complete.
15. Click Additional Measurements to perform repeat measurements. Repeat the
procedure. Click the Finish button when all measurements are complete.
222 TOC-VCPH/CPN
4.5 Application Operations
4.5.1 Manual Injection Analysis
¬Sample Measurement
Procedure
1. From the File menu, select New. The New dialog box is displayed.
2. Double-click the Method icon. The Method Wizard opens. The System Information
window of the wizard (page 1) is displayed.
3. Select the system that was configured with the Manual Injection Kit and used to
create the calibration curve. Advance to page 2 of the Method Wizard (Analysis
Information) and enter the new method name in the text box at the bottom of the
window.
4. Advance to page 3 of the Method Wizard (Calibration Curves). Select the appropriate
calibration curve file.
5. Enter “1” in the No. of Injections field on page 4 of the Method Wizard (Injection
Parameters).
TOC-VCPH/CPN 223
4.5 Application Operations
4.5.1 Manual Injection Analysis
6. Select the desired options on pages 5, 6, and 7 of the Method Wizard, which list peak
time parameter, USP/EP check, and history settings, respectively. Click the Finish
button to create the new method.
Note: Select the settings described in Section 4.2 "Setting General Mea-
surement Parameters".
7. From the Insert menu, select Sample to open the Sample Wizard. Select the newly
created method from the list of files in the Open dialog box. Complete the remaining
wizard pages.
8. From the Instrument menu, select Start. The Manual Injection Parameter dialog box is
displayed.
9. Draw an appropriate amount of sample into the syringe.
10. Draw in and discharge the sample 3 or 4 times before filling the syringe for
measurement.
11. Enter the injection volume in the Manual Injection Parameter dialog box, and click
the Set button.
12. Verify that the instrument is ready for measurement, and click the Start button.
13. For TC or TN measurement, open the injection port valve and insert the syringe
needle into the injection port. For IC measurement, insert the syringe needle into the
injection port septum.
Note: • Insert the syringe needle as far as possible into the injection port.
• To obtain reproducible results, perform all injections at the same
speed.
14. Remove the syringe from the injection port. For TC or TN measurement, close the
injection port valve.
15. Click Additional Measurements to perform repeat measurements. Repeat the
procedure. Click the Finish button when all measurements are complete.
224 TOC-VCPH/CPN
4.5 Application Operations
4.5.1 Manual Injection Analysis
TOC-VCPH/CPN 225
4.5 Application Operations
4.5.1 Manual Injection Analysis
4.5.1.4 Maintenance
Replacement Procedure
1. Loosen and remove the injection port bushing.
2. Remove the old septum, and insert a new one.
3. Replace the bushing and tighten it.
CAUTION
IC reaction solution may spatter close to the injection port septum and injection
port; however, this poses no problem during analysis. When replacing the
septum, wipe away any liquid, taking care to avoid contact.
Gas-tight Syringe
For analysis, use a needle length of 50mm, or a 2 inch gas-tight syringe. The following gas-
tight syringes are available from Shimadzu. Order replacement syringes as necessary.
226 TOC-VCPH/CPN
4.5 Application Operations
4.5.2 Gas Sample Analysis
CAUTION
• Do not inject a large quantity of sample at one time when measuring combustible
or explosive gases. The combustion tube could break. The injection volume
should not exceed 10mL.
• Use caution when measuring combustible or explosive gases.
• Do not analyze samples that can release toxic substances during combustion. The
exhaust gas released during analysis passes through the CO2 detector to the CO2
absorber, after which it is used as NDIR purge gas, and then is vented inside the
instrument. Take appropriate countermeasures if there is a possibility that toxic
substances could be included in the exhaust gas. If appropriate countermeasures
cannot be taken, do not perform such analyses.
• The sample analysis results are displayed according to the gas concentration used in cal-
ibration. For example, if CO2 is used to generate the calibration curve, the sample anal-
ysis result is displayed based on a conversion using the CO2 gas concentration.
• Gaseous samples have certain properties such as adsorption, condensation and reactiv-
ity, and some gaseous samples are partially liquid or contain a solid component.
Depending on the properties, certain conditions may occur that cannot be ignored. Some
of these conditions are residual substances and contamination in the flow line, which
may result in peak tailing, carryover, baseline fluctuation, or other analytical problems.
• High concentrations of NOx or SOx in the sample or in the post combustion-oxidized
sample can interfere with the NDIR activity. To evaluate the degree of the interference,
combine the analysis of the base gas without the TOC component together with the
sample analysis.
TOC-VCPH/CPN 227
4.5 Application Operations
4.5.2 Gas Sample Analysis
Installation Procedure
1. Remove the knurled thumbscrew that secures the slider then remove the TC/TN
injector slider, and install the TC/TN block provided in the kit.
2. If the IC reactor kit is being used, remove the IC injector slider, and install the IC
block provided in the manual injection kit.
Note: When installing the block, be sure that it is aligned with the
injection port.
Software Settings
Select the Manual Injection Kit option in the Instrument Setup Wizard Options window
during System Setup. Refer to Section 3.2.2 "Options Window".
Note: Selecting the Manual Injection Kit option disables operation of the
TOC-V injector syringe.
228 TOC-VCPH/CPN
4.5 Application Operations
4.5.2 Gas Sample Analysis
4.5.2.2 Measurement
The procedure for setting the various parameters is the same as described in Section 4.2
"Setting General Measurement Parameters", Section 4.3 "Measurement Procedures and
Tutorial" and Section 4.4.2.2 "Calibration Curve".
Procedure
1. Prepare standard gas containing a known concentration of carbon.
Note: For IC (carbon dioxide) analysis, prepare a standard gas with a
known carbon dioxide concentration. For TC analysis, another
organic gas (for example, methane) of known concentration can be
used.
2. From the File menu, select New. The New dialog box is displayed.
3. Double-click the Calibration Curve icon. The Calibration Curve Wizard opens. The
System Information window of the wizard (page 1) is displayed.
4. From the System drop-down list, select the system that was configured with the
Manual Injection Kit during setup.
TOC-VCPH/CPN 229
4.5 Application Operations
4.5.2 Gas Sample Analysis
5. Select the desired options on pages 2 and 3 of the Calibration Curve Wizard, which
list calibration curve type and analysis information settings, respectively. Enter the
name of the new calibration curve in the text box at the bottom of page 3 of the
wizard.
Note: Select the settings described in Section 4.2 "Setting General Mea-
surement Parameters".
6. Enter the desired number of injections on page 4 of the Calibration Curve Wizard
(Calibration Measurement Parameters).
7. Enter the injection volume on page 5 of the Calibration Curve Wizard (Calibration
Points List).
Note: An injection volume is displayed automatically. The displayed
value is based on the volume of the last injection that was per-
formed using the sample injector. This value can be changed by the
user as appropriate.
8. Select the desired options on pages 6 and 7 of the Calibration Curve Wizard, which
list peak time parameter and history settings, respectively. Click the Finish button to
create the new calibration curve.
Note: Select the settings described in Section 4.2 "Setting General Mea-
surement Parameters".
9. Create a Sample Table for the calibration curve. From the Insert menu, select
Calibration Curve. Select the newly created calibration curve from the list of files in
the Open dialog box.
10. Connect the instrument by selecting Connect from the Instrument menu. Then select
Start from the Instrument menu. The Manual Injection Parameter dialog box is
displayed.
11. Draw an appropriate amount of sample into the syringe.
Note: Draw in and discharge the sample 3 or 4 times before filling the
syringe for measurement.
12. Enter the injection volume in the Manual Injection Parameter dialog box, and click
the Set button.
13. Verify that the instrument is ready for measurement, and click the Start button.
14. For TC or TN measurement, open the injection port valve and insert the syringe
needle into the injection port. For IC measurement, insert the syringe needle into the
injection port septum.
Note: • Insert the syringe needle as far as possible into the injection port.
• To obtain reproducible results, perform all injections at the same
speed.
15. Withdraw the syringe from the injection port. For TC or TN measurement, close the
injection port valve.
The results are displayed when measurement is complete.
16. Click Additional Measurements to perform repeat measurements. Repeat the
procedure. Click the Finish button when all measurements are complete.
230 TOC-VCPH/CPN
4.5 Application Operations
4.5.2 Gas Sample Analysis
¬Sample Measurement
Procedure
1. From the File menu, select New. The New dialog box is displayed.
2. Double-click the Method icon. The Method Wizard opens. The System Information
window of the wizard (page 1) is displayed.
3. Select the system that was configured with the Manual Injection Kit and used to
create the calibration curve. Advance to page 2 of the Method Wizard (Analysis
Information) and enter the new method name in the text box at the bottom of the
window.
4. Advance to page 3 of the Method Wizard (Calibration Curves). Select the appropriate
calibration curve file.
5. Enter “1” in the No. of Injections field on page 4 of the Method Wizard (Injection
Parameters).
TOC-VCPH/CPN 231
4.5 Application Operations
4.5.2 Gas Sample Analysis
6. Select the desired options on pages 5, 6, and 7 of the Method Wizard, which list peak
time parameter, USP/EP check, and history settings, respectively. Click the Finish
button to create the new method.
Note: Select the settings described in Section 4.2 "Setting General Mea-
surement Parameters".
7. From the Insert menu, select Sample. Select the newly created method from the list of
files in the Open dialog box. Complete the remaining pages of the Sample Wizard.
8. From the Instrument menu, select Start. The Manual Injection Parameter dialog box is
displayed.
9. Draw an appropriate amount of sample into the syringe.
Note: Draw in and discharge the sample 3 or 4 times before filling the
syringe for measurement.
10. Enter the injection volume in the Manual Injection Parameter dialog box, and click
the Set button.
11. Verify that the instrument is ready for measurement, and click the Start button.
12. For TC or TN measurement, open the injection port valve and insert the syringe
needle into the injection port. For IC measurement, insert the syringe needle into the
injection port septum.
Note: • Insert the syringe needle as far as possible into the injection port.
• To obtain reproducible results, perform all injections at the same
speed.
13. Withdraw the syringe from the injection port. For TC or TN measurement, close the
injection port valve.
14. Click Additional Measurements to perform repeat measurements. Repeat the
procedure. Click the Finish button when all measurements are complete.
¬Maintenance
Replacing the IC Injection Port Septum (Cushion O-Ring)
The septum used with the IC injection port is a consumable item. The septum typically can
be used for 200 or more injections; however, it should be replaced if there is deterioration
in reproducibility or peak shapes.
Part number for replacement cushion O-ring (20 pcs.): P/N 630-01559-01
Replacement Procedure
1. Loosen and remove the injection port bushing.
2. Remove the old septum, and insert a new one.
3. Replace the bushing and tighten it.
232 TOC-VCPH/CPN
4.5 Application Operations
4.5.2 Gas Sample Analysis
CAUTION
IC reaction solution may spatter close to the injection port septum and injection
port; however, this poses no problem during analysis. When replacing the
septum, wipe away any liquid, taking care to avoid contact.
Gas-tight Syringe
For analysis, use a needle length of 50mm., or a 2-inch gas-tight syringe. The following
gas-tight syringes are available from Shimadzu. Order replacement syringes as necessary.
Table 4.9 • Relationship Between Appropriate Peak Height and (CO2) Gas
Concentration / Injection Volume *
Gas Concentration (CO2) Gas Injection Volume
TC 100ppm 10mL
1000ppm 1mL
1% 500uL
10% 300uL
100% 30uL
IC 100ppm 6uL
1000ppm 600uL
1% 300uL
10% 200uL
100% 20uL
*The values indicated in this table are reference values only. Data will vary slightly among
instruments.
TOC-VCPH/CPN 233
4.5 Application Operations
4.5.2 Gas Sample Analysis
234 TOC-VCPH/CPN
5
5 Maintenance
Inspection and maintenance of the TOC-V instrument are required to ensure normal
operation and reliable and accurate data acquisition.
5.7 Troubleshooting
Describes the error messages and troubleshooting procedures.
5.1 Daily Inspection
5.1.1 Checking Dilution Water
Check the levels of dilution water, acid, drain vessel water, and humidifier water each day
before starting the instrument.
Overflow Tube
Drain
Vessel
Pure Water
Verify that the water level in the drain vessel inside the instrument is within 10mm of the
overflow tube on the side of the drain vessel. Replenish as necessary with purified water
(ion-exchange water).
Reference: Refer toSection 4.1.4 "Water Supply to the Cooler Drain Container", for details on adding
purified water,.
Confirm that the carrier gas does not leak from the tip of the drain tubing when the flow
rate is 150mL/min.
236 TOC-VCPH/CPN
5.1 Daily Inspection
5.1.4 Checking Humidifier Water Level
Supply Port
Pure Water
“Lo” Mark
Verify that the water level in the humidifier is above the “Lo” mark. If the level is below the
“Lo” mark, replenish by adding purified water through the water supply port on top of the
vessel. Add water until the level reaches the “Hi” mark.
TOC-VCPH/CPN 237
5.2 Periodic Inspections
5.2.1 Catalyst Regeneration
Periodically inspect the items listed in this section to maintain instrument performance
quality.
Regeneration Procedure
The procedure for regenerating the TC catalyst is controlled by the TOC-Control V
software, as described below.
1. From the Instrument menu in the Sample Table Editor, select
Maintenance>Regeneration of the TC Catalyst.
The Regeneration of TC-Catalyst dialog box is displayed.
2. Click the Start button to begin the procedure. A progress bar is active until the TC
catalyst has been regenerated.
3. Click the Close button.
238 TOC-VCPH/CPN
5.2 Periodic Inspections
5.2.2 Washing or Replacing Catalyst
CAUTION
Dilute hydrochloric acid is the only type of acid that can be used to regenerate
TC catalyst. Other acids, particularly phosphoric acid, cause abnormally high
peaks and scattered peak values to occur. If this happens, the catalyst must be
replaced.
Standard catalyst can be washed and reused several times using the following procedure.
High sensitivity catalyst cannot be washed and reused.
WARNING
Due to the danger of burn injury, allow the oven temperature to cool to room
temperature before starting any maintenance procedure involving the
combustion tube or catalyst.
Procedure
1. Thoroughly wash the catalyst with tap water to remove any accumulated salts.
2. Neutralize the alkalis with dilute hydrochloric acid (5:1).
3. Thoroughly wash the catalyst with tap water to remove the hydrochloric acid.
4. Rinse with pure water (ion exchange water acceptable) and then dry.
CAUTION
• Dry the catalyst at a temperature lower than 700°C to avoid the degradation of
catalyst activity that occurs at higher temperatures.
• A large quantity of steam is produced when the regenerated catalyst is placed in
the oven and heated. Due to the danger of burn injury, leave the outlet of the
combustion tube open until no more steam is released.
TOC-VCPH/CPN 239
5.2 Periodic Inspections
5.2.2 Washing or Replacing Catalyst
If performance is not restored after washing the catalyst, it must be replaced. Replace the
catalyst with new platinum catalyst (alumina spherical support). Keep extra platinum
catalyst on hand for such a contingency. One bottle of platinum catalyst is enough for one
replacement.
Use the following procedure to replace the platinum catalyst.
Procedure
1. Wash the platinum mesh and the inside of the combustion tube with (1:2)
hydrochloric acid (1 part concentrated HCl and 2 parts water).
2. Rinse with water.
3. Dry, and reuse these items.
4. Fill the combustion tube as described in Section 4.1.1 "Catalyst Filling".
Disposal of Catalyst
Used catalyst should be disposed of according to local industrial waste disposal regulations.
It must be clearly disclosed using an industrial waste manifest that the used catalyst
contains platinum.
The carrier gas purification tube burns the carrier gas and oxidizes carbon-based impurities,
converting them to carbon dioxide. The carbon dioxide is absorbed and eliminated by the
CO2 absorber. The carrier gas purification tube is filled with oxidation catalyst (same as
TOC standard catalyst).
The procedure for replacing the purification tube and catalyst is as follows.
WARNING
Before attempting to remove the carrier gas purification tube, be sure to allow
the oven temperature to cool to room temperature. There is a danger of burn
injury if this procedure is performed while the oven is hot.
240 TOC-VCPH/CPN
5.2 Periodic Inspections
5.2.2 Washing or Replacing Catalyst
Removal Procedure
1. Remove the compression fitting joints at the top and bottom of the carrier gas
purification tube.
2. Detach the TC/TN sample injection block from the combustion tube. Refer toSection
4.1.2 "Connecting the Combustion Tube".
3. Extract the carrier gas purification tube from the electric oven by pulling upward.
Replacing Catalyst
CAUTION
The top nut of the compression fitting joint is hand-tightened. Do not use tools to
tighten or loosen this nut. Because the tube is constructed of quartz glass,
over-tightening or aggressive handling may result in damage.
Procedure
1. Remove the catalyst from the carrier gas purification tube through the opening at the
top.
2. Remove the quartz wool using a thin tube or other instrument.
3. Insert a small amount of new TOC standard catalyst into the combustion tube from
the bottom (straight end).
4. Pack a small amount of quartz wool into the tube from the bottom.
Note: The quartz wool is inserted to prevent the catalyst from falling out
of the bottom. Packing too much quartz wool into the tube will
greatly impede airflow, so do not use too much quartz wool. To
prevent contamination of the catalyst or the inclusion of foreign
particles, use the same precautions used when filling the
combustion tube with TOC platinum catalyst.
5. Insert the straight end of the carrier gas purification tube down into the front 8mm
diameter hole in the oven.
6. Using the compression fitting joints, restore the tubing connections to their original
condition.
TOC-VCPH/CPN 241
5.2 Periodic Inspections
5.2.3 Washing/Replacing the Combustion and Carrier Gas Purification Tubes
Reference: For details on replacing the CO2 absorbers, refer to Section 4.1.8 "Installation of the CO2
Absorber".
242 TOC-VCPH/CPN
5.2 Periodic Inspections
5.2.6 Replacing the Halogen Scrubber
Replacement Procedure
1. Unclip the upper and lower clips that secure the halogen scrubber.
2. Remove the upper and lower connection clamps.
3. Disconnect the Viton connectors and remove the halogen scrubber.
Dehumidifier To Sample
Cell Inlet
The baseline may fluctuate immediately after replacement. Allow the instrument to run for
a short time until the baseline stabilizes. Stabilization normally occurs within one hour.
Used halogen scrubbers should be disposed of according to local industrial waste disposal
regulations. It must be clearly disclosed using an industrial waste manifest that the halogen
scrubber contains copper.
Clamps
Clips
Viton Connectors
TOC-VCPH/CPN 243
5.2 Periodic Inspections
5.2.7 Replacing the Syringe Plunger Tip
Replacement Procedure
1. Remove the syringe referring to Section 4.1.3 "Installing the Syringe".
2. Use a sharp knife to make two or three incisions at the bottom of the plunger tip, as
shown in Figure 5.6 "Removing the Old Plunger Tip".
3. Grasp the plunger tip with a pair of pliers and remove it from the plunger. The O-ring
that is held in place by the plunger tip also comes off. DO NOT reuse this O-ring.
4. Pushing down on the plunger, press the leading edge of the new plunger tip straight
into the plunger tip seat, as shown in Figure 5.7 "Installing the New Plunger Tip
(cross-section)". Save the syringe for future use since the plunger tip seat is enclosed
inside the syringe.
Plunger Tip
Note: • Do not scratch the perimeter of the plunger tip. This could cause leaks.
• Use only Shimadzu-specified plunger tips. Use of unspecified parts may
result in malfunction and/or diminished performance.
244 TOC-VCPH/CPN
5.2 Periodic Inspections
5.2.8 Replacing Ferrules at the Combustion Tube Outlet Connection
Reference: For details on the replacement procedure, refer to Section 4.1.2 "Connecting the
Combustion Tube".
White O-ring
Black O-ring
If a gas leak develops between the injector block and the combustion tube, replace the
O-ring there as well.
CAUTION
Do not scratch the O-rings or the O-ring seat when replacing the O-rings.
TOC-VCPH/CPN 245
5.2 Periodic Inspections
5.2.10 Replacing the Syringe Pump 8-port Valve Rotor
CAUTION
Switch OFF the instrument before replacing the 8-port valve rotor.
Procedure
1. Shut down the instrument and turn OFF the power.
2. Verify that the O mark on the 8-port valve interrupt plate is directed toward the No. 1
port. If not, the instrument was not shut down using the appropriate procedure. In this
case, switch the instrument back ON, and follow the correct shutdown procedure.
Reference: For the correct shutdown procedure, refer to Section 4.3.3 "Ending Measurement".
3. Remove the nut using the provided wrench.
Nut
4. Pull up to remove the rotor drive component. Hang it on the screw on the right side of
the plunger.
246 TOC-VCPH/CPN
5.2 Periodic Inspections
5.2.10 Replacing the Syringe Pump 8-port Valve Rotor
Retaining Nut
6. Note the orientation (toward port No. 1) of the flat surface of the spring-equipped rod.
Remove the spring-equipped rod and retaining nut.
7. Note the orientation of the rotor hole position. Grasp the rotor with forceps and
remove by pulling up.
Rotor
8. Insert a new rotor, taking care that the rotor hole matches original orientation.
9. Reinsert the retaining nut and spring-equipped rod, with the flat surface directed
toward port No. 1. Tighten the retaining nut.
10. Replace the rotor drive component, and tighten the nut.
11. Switch the instrument ON.
Note: Perform the syringe pump zero detection procedure before analysis.
Refer to Section 4.1.3.1 "Syringe Pump Zero Point Detection".
TOC-VCPH/CPN 247
5.2 Periodic Inspections
5.2.11 Replacing the CO2 Absorber for the Optional Carrier Gas Purifier
5.2.11 Replacing the CO2 Absorber for the Optional Carrier Gas Purifier
The NDIR baseline becomes unstable with a decrease in the carbon dioxide absorption in
the carrier gas purifier CO2 absorber. The baseline position rises and reproducibility
diminishes, making analysis impossible. The soda lime must be replaced when this occurs.
The frequency with which soda lime must be replaced depends on the frequency of
analysis, quantity of organic impurities in the carrier gas, and the TOC content of the
samples. It is recommended that soda lime be replaced once every four months.
CAUTION
The carrier gas pressure applied to the carrier gas purifier CO2 absorber is
approximately 200kPa. It is dangerous to perform this maintenance procedure
under these conditions. Before replacing the soda lime, lower the carrier gas
pressure in the instrument to normal as follows:
• Turn the supply valve of carrier gas source OFF.
• Switch OFF the instrument.
Procedure
1. Remove the elbow joints of the carrier gas purifier CO2 absorber, which are behind
the instrument.
2. Invert the CO2 absorber and remove the screw cap.
3. Remove the used soda lime.
4. Refill the absorber container with about 500g of new soda lime until it reaches about
5mm from the interior screw threads.
CAUTION
Do not allow any soda lime to enter the tube in the center of the container when
refilling the CO2 absorber.
5. Fill the hollow in the screw cap with quartz wool and screw the cap back on the
container.
CAUTION
To maintain an airtight seal, a rubber O-ring is placed between the container and
the screw cap. Inspect the O-ring to verify that it is not scratched or twisted, then
attach the cap.
248 TOC-VCPH/CPN
5.2 Periodic Inspections
5.2.12 Replenishing IC Reaction Solution
6. Using the elbow joints, connect the carrier gas purifier CO2 absorber to the two tubes
exiting the back of the instrument as follows:
• Tubing labeled “L”: Affixed to the connection port of the CO2 absorber container
labeled “L”
• Tubing labeled “S”: Affixed to the connection port of the CO2 absorber container
labeled “S”
CAUTION
Tighten the nut by hand only. Do not use tools.
7. Place the carrier gas purifier CO2 absorber container on the stand behind the
instrument.
Used soda lime should be disposed of according to local industrial waste disposal reg-
ulations. It must be clearly disclosed using an industrial waste manifest that the mate-
rial contains soda lime.
CAUTION
Soda lime is a corrosive chemical substance. Be careful not to touch it with bare
hands. For details, refer to Section 6.9 "Material Safety Data Sheets".
CAUTION
The IC reaction solution is acidic. Use care not to spill or touch the fluid when
handling it. For details, refer to Section 6.9 "Material Safety Data Sheets".
TOC-VCPH/CPN 249
5.2 Periodic Inspections
5.2.13 Replacing the Pump Head of the IC Reaction Solution Pump
CAUTION
The pump head and rubber tubing inside the IC reagent pump are filled with IC
reagent solution. This solution is acidic. Use care to avoid touching or spilling
the solution when changing the pump heads. For details refer to Section 6.9
"Material Safety Data Sheets".
Replacement Procedure
Use the following procedure to replace the rubber tubing of the pump head.
1. Grasp the front and rear stops of the pump head by hand. Pull the white pump head
from the black pump body.
2. Disconnect both of the yellow rubber tubes on the left side of the pump head from the
Teflon tubes.
3. Connect the new rubber tubing to the Teflon tubes of the pump head.
4. Replace the pump head by following the above procedure in reverse.
Figure 5.12 Replacing the Pump Head of the IC Reaction Solution Pump
250 TOC-VCPH/CPN
5.2 Periodic Inspections
5.2.14 Replacing the Filler of the Optional POC Analysis CO2 Absorber
5.2.14 Replacing the Filler of the Optional POC Analysis CO2 Absorber
The CO2 absorber for POC analysis is filled with about 6 grams of lithium hydroxide,
which is sufficient to absorb carbon dioxide gas for several hundred analyses under normal
conditions. Airflow resistance gradually increases with use and may become excessive
irrespective of adequate carbon dioxide absorbency.
Replace the lithium hydroxide in the CO2 absorber with a new batch if analysis values
become unstable or carrier gas flow is impeded. For the procedure on filling the CO2
absorber with lithium hydroxide, refer to Section 4.1.12 "Preparation for POC analysis".
CAUTION
• Lithium hydroxide is a corrosive (strongly alkali) powder. Be carefuel not to spill
or touch it during handling. For details, refer to Section 6.9 "Material Safety
Data Sheets".
• Used lithium hydroxide should be disposed of by a certified industrial waste
handler according to local industrial waste disposal regulations. It must be clearly
disclosed using an industrial waste manifest that the waste is lithium hydroxide.
• Lithium hydroxide crystal is found in both anhydrous and hydrated forms.
Because the carbon dioxide gas absorbance properties of the hydrated form are
slightly inferior to those of the anhydrous form, use anhydrous lithium hydroxide
to fill the CO2 absorber.
• The life of the CO2 absorber will be shortened by performing POC analysis on
samples with high concentrations of IC.
TOC-VCPH/CPN 251
5.3 Autosampler (ASI-V) Maintenance
5.3.1 Rinse Bottle Inspection
Maintenance of the autosampler includes periodic inspection of the rinse bottle, sample
catcher and rinse pump.
Rinse Bottle
CAUTION
Corrosion of the instrument may occur if spilled sample is allowed to remain in
the sample catcher. Keep the sample catcher clean.
252 TOC-VCPH/CPN
5.3 Autosampler (ASI-V) Maintenance
5.3.3 Replacing the Pump Head of the Rinse Pump
Replacement Procedure
Use the following procedure to replace the pump head.
1. Remove the rectangular pump access cover on the back left side of the autosampler.
2. Grasp the latch on the pump head by hand and pull the white pump head from the
black pump body.
3. On the left side of the pump head, disconnect both of the yellow rubber tubes from the
Teflon tubes.
4. Connect the new rubber pump head tubing to the Teflon tubes.
5. Replace the pump head by following the above procedure in reverse and then replace
the pump access cover.
Pump
Head
TOC-VCPH/CPN 253
5.4 8-Port Sampler (OCT-1) Maintenance
5.3.3 Replacing the Pump Head of the Rinse Pump
CAUTION
Switch OFF the power before replacing to the 8-port valve rotor.
Replacement Procedure
1. Shut down the TOC-V unit and switch OFF the power.
Reference: Refer to Section 4.3.3 "Ending Measurement" for details on shutting down the TOC-V unit.
2. Remove the 4 screws at the top of the 8-Port Sampler, and take OFF the cover.
3. Verify that the O mark on the 8-port valve cover is directed toward port No.1.
Note: If the O mark is not directed toward port No.1, the instrument was not
shut down properly. If this is the case, turn the TOC-V on again, and
shut it down properly.
254 TOC-VCPH/CPN
5.4 8-Port Sampler (OCT-1) Maintenance
5.3.3 Replacing the Pump Head of the Rinse Pump
4. Remove the nut using the wrench supplied with the TOC-V.
Nut
7. Note the orientation (toward port No. 1) of the flat surface of the spring-equipped rod.
Remove the spring-equipped rod and retaining nut.
8. Note the orientation of the rotor hole position. Grasp the rotor with forceps, and
remove by pulling upwards.
Rotor
9. Insert a new rotor, positioning the rotor hole to the original position.
10. Reinsert the spring-equipped rod and retaining nut with the flat surface directed
toward port No. 1. Tighten the retaining nut.
11. Replace the 8-port valve in its original position, and secure it using the nut.
12. Replace the cover of the 8-Port Sampler and secure it with four screws.
13. Switch on the TOC-V unit.
TOC-VCPH/CPN 255
5.5 TN Unit (TNM-1) Maintenance
5.5.1 Replacing the Catalyst in the Ozone Treatment Unit
Maintenance of the TN unit includes periodic inspection of the ozone treatment unit
catalyst, chemiluminescence detector and ozone generator.
CAUTION
The interior of the ozone treatment unit is heated to 50°C. Perform maintenance
only after cooling for at least two hours.
Replacement Procedure
Use the following procedure to replace the catalyst in the ozone treatment unit.
1. Remove the knurled screw at the bottom of the unit, and remove the installation
bracket.
Ozone
Treatment Unit
Knurled Screw
Installation Bracket
Figure 5.18 Removing the Ozone Treatment Unit Screw and Bracket
256 TOC-VCPH/CPN
5.5 TN Unit (TNM-1) Maintenance
5.5.1 Replacing the Catalyst in the Ozone Treatment Unit
2. Remove the hose clamp at the bottom of the ozone treatment unit and pull off the
Viton tubing.
Hose Clamp
Viton Tubing
3. Remove the hose clamp at the top of the unit and remove it from the Viton coupling.
Hose Clamp
Viton Coupling
TOC-VCPH/CPN 257
5.5 TN Unit (TNM-1) Maintenance
5.5.1 Replacing the Catalyst in the Ozone Treatment Unit
4. Remove the cap from the ozone treatment unit and take out the used catalyst and
polypropylene packing material.
5. Pack a small amount of new polypropylene packing material into the bottom of the
ozone treatment unit, then fill with approximately 60 grams of catalyst.
6. Insert a small amount of new polypropylene packing material on top of the catalyst.
7. Place the cap back on the container.
8. Remove the Viton coupling connecting the detector exhaust opening to the inlet of the
ozone treatment unit.
Replace this Viton coupling with a new one whenever the catalyst is replaced. Inspect
the Viton tubing at the outlet of the ozone treatment unit and replace it if cracked or
damaged.
9. Reconnect the Viton tubing and Viton coupling, and reinstall the ozone treatment unit.
CAUTION
Used catalyst should be disposed of according to local industrial waste disposal
regulations. It must be clearly disclosed using an industrial waste manifest that
the catalyst contains manganese dioxide.
258 TOC-VCPH/CPN
5.5 TN Unit (TNM-1) Maintenance
5.5.2 Chemiluminescence Detector
Maintenance Procedure
Perform maintenance on the detector only after shutting down the instrument.
1. After removing the tubing from the reactor, as shown in the figure below, loosen the 4
mounting screws, and detach the reactor.
Detector
Reactor
Mounting
Screws
Gasket
2. Remove the gasket and confirm that it is not dirty or cracked. Replace it if necessary.
3. Confirm that the detector window is not dirty. If necessary, gently wipe the window
clean with a cloth.
4. Place the flat side of the gasket facing the detector window, sandwiching it between
the window and the reactor.
5. Secure the gasket in place with the 4 mounting screws, tightening them uniformly.
6. Reconnect the tubing.
CAUTION
Recheck the tubing after completing the maintenance work to make absolutely
certain that there are no mistakes in the line connections and that everything has
been properly tightened.
TOC-VCPH/CPN 259
5.5 TN Unit (TNM-1) Maintenance
5.5.3 Ozone Generator Replacement
CAUTION
If the ozone generator is used after 8000 hours of operation, ozone generation
may not be sufficient for analysis, or ozone may leak from inside the generator.
Replace the ozone generator without fail.
260 TOC-VCPH/CPN
5.6 Software-Controlled Maintenance Functions
5.6.1 Zero Point Detection
Maintenance functions that can be performed using the TOC-Control V software are
described below. These functions can be accessed through the Instrument>Maintenance
menu.
TOC-VCPH/CPN 261
5.6 Software-Controlled Maintenance Functions
5.6.2 Regeneration of the TC Catalyst
Click the Start button to begin the procedure. A progress bar is active until the TC catalyst
is regenerated. Click the Close button.
Click the Start button to begin the procedure. A progress bar is active until the IC solution
is regenerated. Click the Close button.
262 TOC-VCPH/CPN
5.6 Software-Controlled Maintenance Functions
5.6.4 Residue Removal
Click the Start button to begin the procedure. A progress bar is active until the residue
removal procedure is complete. Click the Close button.
Click the Start button to begin the procedure. A progress bar is active until the flow line
content has been replaced. Click the Close button.
TOC-VCPH/CPN 263
5.6 Software-Controlled Maintenance Functions
5.6.6 Flowline Washing
To perform flow line washing, click the Select Flow Lines button to open the Wash Flow
Lines dialog box. Select the flow lines to be washed, and click the Washing button. The
ports for the 8-Port Sampler 1 (2) can be selected if the 8-Port Sampler 1 (2) is being used.
While this process is being executed, the buttons in the Washing dialog box are disabled,
and progress bar activity is displayed.
264 TOC-VCPH/CPN
5.6 Software-Controlled Maintenance Functions
5.6.7 Mechanical Check
¬TOC Tab
Syringe Pump
Use the buttons to move the syringe pump to the Zero, Home, or Sparge posi-
tions. Click the Reset button to move the plunger to the highest position.
Solenoid Valves
Use the buttons to open or close the various solenoid valves. Different options
are available depending on the instrument configuration and oxidation method.
Port Select
Open the line between the syringe and an injection port by clicking the button for
the desired port.
TC Inj. Port
Specify whether the port should be set to drain or measure by clicking the appro-
priate button.
IC Inj. Port
Specify whether the port should be set to drain or measure by clicking the appro-
priate button.
Acid Pump
Use the buttons to open or close the IC acid pump lines.
TOC-VCPH/CPN 265
5.6 Software-Controlled Maintenance Functions
5.6.7 Mechanical Check
Figure 5.32 ASI / 8-Port Sampler Tab of Operation Verification Dialog Box
Rinse
Use the On and Off buttons to start and stop the ASI flow line rinsing option.
Move Tray
Use the directional arrows to move the ASI tray to the desired position. Click the
Reset button to return the tray to its home position.
Arm
Use the directional arrows to move the arm vertically. Use the drop-down list to
move the arm horizontally.
Vial Position
Use the spin controls to enter a vial position, then click the Move button to verify
that the arm moves to the specified vial position. This is the only option that is
enabled for systems using the OCT-1 8-port sampler.
Magnetic Stirrer
Use the On and Off buttons to start and stop the magnetic stirrer.
266 TOC-VCPH/CPN
5.6 Software-Controlled Maintenance Functions
5.6.8 ASI / 8-Port Sampler Initialization
Click the Start button to begin the procedure, during which a progress bar is active. The
progress bar becomes inactive after the ASI or 8-Port Sampler is initialized. Click the Close
button.
Click the Preparation Start button to begin the procedure. A progress bar is active while the
rack moves to the position where it can be replaced. When the progress bar begins blinking,
replace the rack. Click the Finish Replacement button to initialize the ASI and complete the
rack change process. Click the Close button.
TOC-VCPH/CPN 267
5.6 Software-Controlled Maintenance Functions
5.6.10 ASI Needle Change
Click the Preparation Start button to begin the procedure. A progress bar is active while the
needle moves to the position where it can be replaced. When the progress bar begins
blinking, replace the needle. Click the Finish Replacement button to initialize the ASI and
complete the needle change process. Click the Close button.
Click the Preparation Start button to begin the procedure. A progress bar is active while the
syringe moves to the position where it can be replaced. When the progress bar begins
blinking, replace the syringe. Verify that the syringe moves smoothly between the upper
and lower positions by clicking the directional arrows. Click the Finish Replacement button
to move the syringe to its initialization position, then click the Close button.
Note: Perform the zero point detection procedure whenever a new syringe is installed
or the syringe is replaced. Refer to Section 4.1.3.1 "Syringe Pump Zero Point
Detection".
268 TOC-VCPH/CPN
5.6 Software-Controlled Maintenance Functions
5.6.12 Blank Check
Note: Do not insert a sample into the Sample Table. The Blank Check procedure only
works in a blank Sample Table.
5.6.14 History
Use this option to view the operational history of various instrument components and to set
limits for various parameters of the components. When a limit is exceeded for an
instrument component, a warning message will be generated on the PC screen to remind the
user to perform maintenance on the component. Refer to Section 4.2.4 "Maintenance
History Settings" for a detailed description of this function.
TOC-VCPH/CPN 269
5.7 Troubleshooting
5.7.1 Error Messages
5.7 Troubleshooting
This section describes the error messages displayed by the TOC-Control V software and
the actions recommended to resolve the underlying causes.
270 TOC-VCPH/CPN
5.7 Troubleshooting
5.7.1 Error Messages
TOC-VCPH/CPN 271
5.7 Troubleshooting
5.7.1 Error Messages
The following messages indicate that an error has occurred that prevents measurement
from continuing. Take the appropriate corrective actions as indicated in the following table.
If the cause of the problem cannot be discovered, or if the same error message appears after
taking the indicated corrective actions, contact your Shimadzu representative.
Note: When the following errors are generated, power to the instrument oven is auto-
matically stopped. When power to the oven is switched on again, wait until the
temperature rises before starting measurement.
272 TOC-VCPH/CPN
5.7 Troubleshooting
5.7.1 Error Messages
TOC-VCPH/CPN 273
5.7 Troubleshooting
5.7.2 Troubleshooting
5.7.2 Troubleshooting
5.7.2.1 TOC-V
Does carrier
gas flow normally NO
to the analysis
cell?
Carrier gas leak.
Is dehumidifier
within the specified NO
temperature?
The dehumidifier temperature
setting is bad.
YES Contact your Shimadzu
representative.
The infrared gas
analyzer is bad.
Contact your Shimadzu
representative.
274 TOC-VCPH/CPN
5.7 Troubleshooting
5.7.2 Troubleshooting
Is carrier gas
NO
flowing normally?
Was
the catalyst NO
replaced?
Accumulated obstructing
substance.
YES
Replace the catalyst.
Is the
dehumidifier NO
operating
normally?
The dehumidifier is bad.
YES
Contact your Shimadzu
representative.
Are
external
conditions (vibration, NO
temperature, power
voltage)
normal? External conditions are
adverse.
YES
Eliminate the adverse
condition.
The infrared gas analyzer
is bad.
Contact your Shimadzu
representative.
TOC-VCPH/CPN 275
5.7 Troubleshooting
5.7.2 Troubleshooting
Is the NO
standard solution
normal?
Bad standard solution.
Is the
sample injected NO
normally from injection
tube?
Poor injection condition.
Do
bubbles remain
in syringe? (especially YES
in dilution
analysis)
Poor injection volume
repeatability.
NO Eliminate bubbles.
Does
carrier gas YES
sometimes leak from
the drain
bottle?
Low water level.
Does the
poor reproducibility IC
occur in the
TC or IC
circuit? IC reaction solution has
deteriorated.
Was the NO
TC catalyst
regenerated?
Insufficient regeneration of
theTC catalyst .
YES
Regenerate the TC catalyst.
TC catalyst deteriorated.
276 TOC-VCPH/CPN
5.7 Troubleshooting
5.7.2 Troubleshooting
Is
sample
injected into TC
NO
combustion tube or IC
reaction vessel?
(confirm
visually) Poor injection.
Is carrier gas
always flowing from YES
the drain
bottle?
Is the
level of the water NO
normal in the
drain bottle?
Low water level.
TOC-VCPH/CPN 277
5.7 Troubleshooting
5.7.2 Troubleshooting
5.7.2.2 ASI-V
Is
connection
cable and power cable NO
properly connected?
Improperly connected cable.
Is there
sufficient NO
rinse water in rinse
water bottle?
Insufficient rinse water.
Is the
rinse tubing NO
properly inserted into
the rinse water
bottle? The rinse tubing is not properly
inserted into rinse water bottle.
YES
Properly insert the tubing into
the rinse water bottle.
The rinse pump is bad.
278 TOC-VCPH/CPN
5.7 Troubleshooting
5.7.2 Troubleshooting
Is there
any material YES
adhering to the
sample
needle?
Poor reproducibility due to
foreign matter.
NO Remove the obstruction from
the opening at the tip of the
sample needle.
Is there
a leak at the
connection of sample YES
needle and ASI
sample
tubing? Air is drawn in at the
connection.
NO
Properly reconnect the tubing.
Poor reproducibility due to
a problem with the TOC-V.
Refer to Troubleshooting #1.
TOC-V item (5)
TOC-VCPH/CPN 279
5.7 Troubleshooting
5.7.2 Troubleshooting
Is the leakage
present at the 8-port valve YES
and sample tubing
connection?
Connection is leaking.
Ensure that the tip of the tubing does not protrude beyond the bottom surface of the slider.
The O-ring or other seal parts will be scratched if the tubing protrudes, resulting in gas
leaks. The sample is not properly injected unless the tip of the tubing extends into the
depression in the bottom of the slider.
Check the state of the injection for TC analysis by looking through the TC combustion tube
while it is attached to the instrument. The sample injection tubing can be detached from the
slider, and water can be injected for inspection using the functions available in the
Instrument>Maintenance>Mechanical Check screen.
280 TOC-VCPH/CPN
5.7 Troubleshooting
5.7.2 Troubleshooting
Salt or other substances adhering to the tip of the tubing after rinsing can allow large
droplets of sample to remain on the tip of the tubing after injection. A deformed tip on the
sample injection tubing (a beard-like protrusion, for example) may cause the sample to be
injected obliquely. In either case, reproducibility is adversely affected and the sample
injection tubing should be replaced. Another option available only if there is sufficient
length of tubing remaining is to use a sharp cutting tool at a 90° angle to cut off 2 - 3mm
from the tip of the currently used sample injection tubing.
Tube
Bushing
Packing (spacer,
washer, gasket)
Approximately
2 - 2.5mm
TOC-VCPH/CPN 281
5.7 Troubleshooting
5.7.2 Troubleshooting
282 TOC-VCPH/CPN
6
6 Reference Information
This chapter describes the principles of analysis and lists instrument specifications,
accessories, and parts. This chapter also includes installation procedures, for use in the
event that the instrument is relocated.
Two types of carbon are present in water: organic carbon and inorganic carbon. Organic
carbon (TOC) bonds with hydrogen or oxygen to form organic compounds. Inorganic
carbon (IC or TIC) is the structural basis for inorganic compounds such as gas carbonates
and carbonate ions. Collectively the two forms of carbon are referred to as total carbon
(TC) and the relationship between them is expressed TOC=TC-IC.
Nitrogen is also present in water in two types: organic and inorganic. The sum of these is
referred to as total nitrogen (TN).
The principles underlying TC and TN analysis are explained in the following sections.
284 TOC-VCPH/CPN
6.1 Principles of Analysis
6.1.2 Principles of IC (Inorganic Carbon) Analysis
Defining IC
The IC measured by TOC analysis consists of the carbon contained in carbonates and in
carbon dioxide dissolved in water. By acidifying the sample with a small amount of
hydrochloric acid to obtain a pH less than 3, all carbonates are converted to carbon dioxide
(CO2) by the following reactions:
Me2CO3 + 2HCl --> CO2 + 2MeCl + H2O
Carbon dioxide and dissolved carbon dioxide in the sample are volatilized by bubbling
(sparging) air or nitrogen gas that does not contain carbon dioxide through the sample.
TOC-VCPH/CPN 285
6.1 Principles of Analysis
6.1.3 Principles of NPOC (Non-Purgeable Organic Carbon) Analysis
Table 6.1 • Residual Rate of Volatile Organic Substances using Sparging (Nitrogen Gas)
Organic Substance Pre-Sparging Conc. Post-Sparging Conc. Residual Rate (%)
(ppm) (ppm)
Methanol 117. 5 116 98.6
Ethanol 106.5 105 98.5
Isopropyl alcohol 129 127 98.5
n-butyl alcohol 117 115 98.3
Acetone 106 101 95.3
Acetaldehyde 130 117 90.0
Ethyl acetate 102 88 86.3
Tyrosine 117 116.5 99.5
Benzene 85 2.5 2.9
Cyclohexane 79 2 2.5
286 TOC-VCPH/CPN
6.1 Principles of Analysis
6.1.4 Principles of POC (Purgeable Organic Carbon) Analysis
TOC-VCPH/CPN 287
6.1 Principles of Analysis
6.1.5 Principles of Measuring TOC
TC-IC Method
In the TC-IC method, TOC is measured as the difference between the TC and IC analysis
values. The TC-IC method is not recommended for samples that contain more IC than TOC
(samples where TC consists almost entirely of IC); the NPOC method is recommended for
such samples. The TOC value determined using the TC-IC method includes errors
associated with each of the individual TC and IC measurements, and can therefore result in
a large error in the TOC value.
Because detection accuracy decreases with increasing IC concentration, the NPOC method
is also recommended for samples containing IC concentrations exceeding 10ppm for TOC/
TN catalyst and 5ppm for high sensitivity catalyst.
NPOC Method
The NPOC method is the most widely used TOC analysis method. The NPOC method is
not recommended for samples that foam during sparging. Samples that become foamy
during sparging tend to form bubbles that flow out of the syringe, which removes the
concentrated TOC sample components and leaves a small amount of TOC in the syringe.
The TC-IC method is recommend in this situation.
POC+NPOC Method
The POC+NPOC method should be used when the amount of POC present in the sample
cannot be disregarded.
288 TOC-VCPH/CPN
6.2 Analysis-Related Technical Information
6.2.1 Peak Area Analysis
TOC-VCPH/CPN 289
6.2 Analysis-Related Technical Information
6.2.3 Calibration Curves
Shift to Origin
The “shift to origin” function is used when the amount of TC and IC or TN in the pure
water used for standard solution preparation cannot be ignored with respect to the standard
solution concentration. A correction must be made by shifting the calibration curve in a
parallel fashion so that it passes through the origin. This corrects for the TC and IC or TN
concentration in the water used for standard solution preparation. An example of a situation
in which the “shift to origin” function is useful is described below.
When water containing 0.5mg/L TC is used to prepare a 10mg/L TC standard solution, the
actual TC concentration in the standard solution will be 10.5mg/L (0.5mg/L from the water
+ 10mg/L from the standard solution). The 2-point calibration curve generated using such a
solution is shown by the solid line in Figure 6.2 "Shifting of Calibration Curve". If this
curve is used without correction, the sample analysis values will always be 0.5mg/L less
than the true value. By shifting the calibration curve so that it passes through the origin, as
shown by the broken line in Figure 6.2 "Shifting of Calibration Curve", the calibration
curve is effectively corrected.
290 TOC-VCPH/CPN
6.2 Analysis-Related Technical Information
6.2.3 Calibration Curves
A system blank is water that contains absolutely no carbon (zero water). When
measurement of a system blank produces a peak, the TC concentrations measured in
samples will be biased high by the amount of TC in the system blank. In most cases (with
the exception of high sensitivity analysis), this bias is not significant. Measurement error
resulting from the TC or TN content in the water used to prepare standard solutions is
typically much greater than measurement error resulting from the system blank value.
Correcting the calibration curve based on the TC or TN content in the standard solution
preparation water will produce satisfactory analyses. Using this correction method, the
value obtained for a system blank would be greater than zero by the amount of the blank
value.
The instrument allows shifting of calibration curves only to the extent of the system blank
value. The system blank peak height varies with such factors as the system configuration
and the type of catalyst used.
0 TC Concentration (mg/L) 10
(0.5) (10+0.5)
TOC-VCPH/CPN 291
6.2 Analysis-Related Technical Information
6.2.4 Sparging During Standard Solution Analysis
Table 6.2 • CO2 Content (ppm) in Distilled Water Equilibrated with Atmosphere
Temperature (°C)
Atmospheric CO2 0 5 10 15 20 25 30
(vol%)
0.030 1.00 0.83 0.79 0.59 0.51 0.44 0.38
0.033 1.10 0.91 0.76 0.65 0.56 0.48 0.42
0.044 1.47 1.22 1.02 0.87 0.74 0.64 0.56
The above CO2 content can be converted to IC using the following equation.
IC (ppm) = CO2 (mg/L) x 0.27
292 TOC-VCPH/CPN
6.2 Analysis-Related Technical Information
6.2.5 Automatic Selection of Optimum Calibration Curve
(2) If the measured value is less than the concentrations of Calibration Curve 1 and
Calibration Curve 2, and if the concentration of Calibration Curve 2 is less than that of
Calibration Curve 1, Calibration Curve 2 is used.
Conc. of Calibration Curve 1 > Conc. of Calibration Curve 2 > Measured Value →
Calibration Curve 2 is used
(3) If three calibration curves are specified, the calibration curve that is greater than and
closest to the measured value is used.
Conc. of Calibration Curve 1 > Conc. of Calibration Curve 3 > Measured Value > Conc. of Calibration
Curve 2 →
Calibration Curve 3 is used
(4) If the measured value is greater than the concentration of all curves and the
concentration of Calibration Curve 1 is greater than the concentration of Calibration
Curves 2 and 3, Calibration Curve 1 is used.
Measured Value > Conc. of Calibration Curve 1 > Conc. of Calibration Curve 2 > Conc. of Calibration
Curve 3 →
Calibration Curve 1 is used
TOC-VCPH/CPN 293
6.3 Specifications
6.3.1 TOC-VCPH/CPN
6.3 Specifications
6.3.1 TOC-VCPH/CPN
294 TOC-VCPH/CPN
6.3 Specifications
6.3.2 ASI-V
6.3.2 ASI-V
Sample containers 24mL vial rack: 24mL (External diameter 23 x Height 85mm), 100
vials, 100 septum-equipped caps
40mL vial rack: 40mL (External diameter 27 x Height 9mm), 72
vials, 72 septum-equipped caps
Vial rack capacity 24mL vial rack: 93nvials
40mL vial rack: 68 vials
Sample pretreatment • Acid addition and sparging possible for NPOC analysis
• Needle rinse possible
• Stirring with magnetic stirrer (with optional magnetic stirrer)
Ambient temperature 5 - 35°C
Approximate dimensions (W) 370 x (D) 540 x (H) 490mm (excluding projections)
Weight Approx. 13kg
6.3.4 TNM-1
TOC-VCPH/CPN 295
6.3 Specifications
6.3.5 POC Kit
Analyte POC
Measurement method Sparging / CO2 removal / combustion oxidation
CO2 removal method Absorption via lithium hydroxide, removal
Measuring range 0 - 500mg/L
Measurement time Approx. 4min
Repeatability CV within 5%
296 TOC-VCPH/CPN
6.3 Specifications
6.3.8 PC Hardware Requirements
TOC-VCPH/CPN 297
6.4 Standard Accessories
6.3.8 PC Hardware Requirements
298 TOC-VCPH/CPN
6.4 Standard Accessories
6.3.8 PC Hardware Requirements
TOC-VCPH/CPN 299
6.4 Standard Accessories
6.3.8 PC Hardware Requirements
* Flanged tubing in C sets have an inner diameter of 1.0 mm. This set is provided for use
with the combustion type TOC-V.
** Connect this tubing to the common (COM) port of the 8-Port Sampler and port No. 2 on
the TOC-V unit.
*** Flanged tubing in W sets have an inner diameter of 1.5 mm. This set is provided for use
with the wet chemical type TOC-V.
**** Connect this tubing to the common (COM) port of the 8-Port Sampler and port 1on the
TOC-V unit.
300 TOC-VCPH/CPN
6.4 Standard Accessories
6.3.8 PC Hardware Requirements
Table 6.7 • Standard Accessories List 5 (Manual [Gas Sample] Injection Kit)
Part Number Part Name Qty
638-10194 Block, M-INJ TC ASSY 1
638-10195 Block, M-INJ IC ASSY 1
631-43291 Syringe packing 10
TOC-VCPH/CPN 301
6.4 Standard Accessories
6.3.8 PC Hardware Requirements
302 TOC-VCPH/CPN
6.5 Special Accessories
6.3.8 PC Hardware Requirements
TOC-VCPH/CPN 303
6.6 Consumable Parts List
6.3.8 PC Hardware Requirements
304 TOC-VCPH/CPN
6.6 Consumable Parts List
6.3.8 PC Hardware Requirements
TOC-VCPH/CPN 305
6.6 Consumable Parts List
6.3.8 PC Hardware Requirements
306 TOC-VCPH/CPN
6.6 Consumable Parts List
6.3.8 PC Hardware Requirements
TOC-VCPH/CPN 307
6.7 Maintenance Parts List
6.3.8 PC Hardware Requirements
308 TOC-VCPH/CPN
6.7 Maintenance Parts List
6.3.8 PC Hardware Requirements
TOC-VCPH/CPN 309
6.8 Installation
6.8.1 Before Installation
6.8 Installation
Parts Inspection
Confirm that all of the components listed in Section 6.4 "Standard Accessories" have been
included in the shipping package.
100
370mm mm 440mm 160mm
310 TOC-VCPH/CPN
6.8 Installation
6.8.2 Installation Site
CAUTION
When lifting the instrument, lift from the bottom and at both sides of the
instrument.
Do not lift the instrument by the front access door, as this may cause damage.
CAUTION
Avoid locations where flames are prohibited. Some internal parts reach high
temperatures and could cause a fire hazard.
TOC-VCPH/CPN 311
6.8 Installation
6.8.3 Installation Procedure
CAUTION
The instrument may not operate properly if the power supply voltage exceeds
the range of AC100-127V ±10% or AC 220-240V ±10%. Unstable voltage may
cause problems during high sensitivity analysis.
Note: The instrument is compatible with both 50 Hz and 60 Hz power supplies and can
be used with either frequency.
¬Grounding
The included power cord is a three-conductor cable including a ground wire, and a 2P plug
(with ground) at the end. If the power outlet does not have ground receptacle, use the
included 2P (no ground) adapter. Connect the external ground terminal to the ground.
The voltage terminals (AC) and ground terminal (ACC) on the power plug of this
instrument are arranged as shown in Figure 6.5 "Power Plug". Connect them so that they
match the power outlet being used.
312 TOC-VCPH/CPN
6.8 Installation
6.8.3 Installation Procedure
CAUTION
• In addition to the consideration of safety, complete grounding is necessary to
prevent as much as possible the occurrence of signal noise. A Type-3 ground
(ground resistance: 100W or less, see Electrical Facilities Engineering Standards
for details) is recommended.
• Perform grounding separately and avoid contact with water supply pipes, gas
pipes, and lightning rods.
The instrument is set up before shipment to match the power supply voltage in the locale
where it will be used. In the event that it becomes necessary to change the power supply
voltage, please contact your Shimadzu representative. Depending on the voltage change,
an authorized Shimadzu service engineer may be required to rewire connections to the
terminal block to change the power supply voltage and/or perform changes to the main
board.
Use one of the following as the carrier gas supply for this instrument.
• Cylinders filled with high purity air
• Compressed air or instrumentation air (Requires special Carrier Gas Purifier Kit
accessory)
TOC-VCPH/CPN 313
6.8 Installation
6.8.3 Installation Procedure
¬Compressed Air or Instrument Air Purified with the Carrier Gas Purification
Kit
Using the Carrier Gas Purification Kit enables air from a compressor or instrumentation air,
or high-impurity pressurized air in a cylinder, to be used as the carrier gas.
Since there are limits to the capabilities of the Carrier Gas Purification Kit, the following
conditions must be met when supplying compressed air.
• Provide an air filter in the supply line to remove dust, oil mist, water drops, and other
contaminants. Place an air filter in the tubing near the gas connection inlet of the
instrument when a compressor is used.
• Select the Oil-less Carrier Gas Purification Kit when a compressor is used.
• There are cases where the organic gas concentration is too high to be removed
completely by the gas purifier. Analysis accuracy is poor in atmospheres containing
high concentrations of volatile gases from exposed organic solvents and gasoline, or
high concentrations of automobile and diesel engine exhaust. If the possibility of the
presence of these kinds of gases cannot be avoided, use a filter to remove organic
particles and/or gases.
314 TOC-VCPH/CPN
6.8 Installation
6.8.3 Installation Procedure
Procedure
Connection Inlet
Half Union
Nylon Tubing
1. Connect the half unions to the carrier gas inlet on the instrument and the connection
outlet (¼” NPT) of the gas supply source.
2. Connect the nylon tubing (4mm OD, 2.5mm ID) between the gas supply source and
the instrument.
Note: To connect the nylon tubing, firmly insert it into the half union. To
disconnect the tubing, pull the tubing off while pushing down
firmly on the blue ring of the half union with a screwdriver.
3. Verify that there are no gas leaks.
CAUTION
• Do not allow acute bends in the tubing.
• Clean the tubing and connectors to remove any oils or other contaminants on
their inner surfaces.
• Copper tubing or stainless steel tubing may also be used.
TOC-VCPH/CPN 315
6.8 Installation
6.8.3 Installation Procedure
Drain tubing is connected to provide for external discharge of measured samples and other
liquid waste. If a nearby drainage pit is not available, use a 10 to 20 liter polypropylene
container as the drain receptacle.
The drain tubing from the dehumidifier drain port is located inside the right panel. Connect
it to the external drain tubing using two of the provided wiring bands so that there are no
leaks.
CAUTION
• Drainage occurs on a gravity basis. Keep resistance to a minimum. Ensure that
the drain tubing is never higher than the drain discharge port. Do not allow the
tip of the drain tubing to come in contact with the surface of the liquid in the
container. If resistance is high, insufficient drainage could cause the drainage liq-
uid to overflow inside the instrument.
• Corrosive substances such as acids may be present in the drainage liquid. Avoid
touching or spilling these liquids.
316 TOC-VCPH/CPN
6.8 Installation
6.8.4 Installing the Autosampler
This section describes the procedure for connecting the TOC-V to the PC using the
supplied RS-232C cable.
Procedure
1. Insert the supplied RS-232C cable into the RS-232C-1 connector on the back of the
TOC-V, and tighten the retaining screws on both sides. Figure 2.4 "Rear View"
2. Insert the plug at the other end of the RS-232C cable into the RS-232C connector of
the PC, and tighten the retaining screws on both sides.
TOC-VCPH/CPN 317
6.8 Installation
6.8.4 Installing the Autosampler
CAUTION
To prevent instrument damage, verify that the ASI arm is firmly attached to the
retaining bracket with the shipping screw before shipping the instrument or
moving it to another location.
¬Connecting Cables
The TOC-V and ASI-V are connected using cables, as described below.
CAUTION
Verify that the TOC-V is switched OFF before performing the following
procedure.
Procedure
1. Connect the ASI-V power cord at the rear of the ASI-V to the ASI-V power connector
at the rear of the TOC-V.
2. Connect the provided signal cable to the ASI-V signal connectors at the rear of the
TOC-V and the ASI-V.
318 TOC-VCPH/CPN
6.8 Installation
6.8.4 Installing the Autosampler
¬Connecting the Sample Tubing and Sparge Tubing (External Sparge Kit)
The procedure for connecting the ASI-V sample tubing and sparge tubing for the External
Sparge Kit is as follows.
Procedure
1. Open the TOC-Control V Sample Table Editor and connect the instrument.
2. From the Instrument menu, select Maintenance>ASI Needle Change.
3. Click the Preparation Start button.
The autosampler arm moves to the front of the needle replacement window, allowing
easy access.
4. Remove the screw at the lower left side of the needle replacement window, and detach
the window cover by pulling downward.
5. Install the tubing clamp.
Sample Catcher
6. Install the bushing at the end of the sample tubing to the Number 2 port of the TOC-V
sample injector 8-port valve.
TOC-VCPH/CPN 319
6.8 Installation
6.8.4 Installing the Autosampler
7. Draw the sparge tubing out through the joint and connect the sparge tubing.
Sparge Tubing
(External Sparge Kit)
Sparge Tubing
Sparge Tubing
(External Sparge Kit)
ASI-V Cover
Note: Do not use the flow controller knob to turn off the sparge gas flow.
320 TOC-VCPH/CPN
6.8 Installation
6.8.4 Installing the Autosampler
Retaining Bracket
Needle Holder
Sample Needle
7. Securely fit the sample tubing or sparge tubing onto the upper end of the needle.
Note: Verify that the bushing is in position to secure the sample tubing
onto the sample needle.
8. Insert the lower tip of the needle into the needle guide opening, thrusting the needle
holder into the pin of the arm, and then tighten the screw.
TOC-VCPH/CPN 321
6.8 Installation
6.8.4 Installing the Autosampler
9. Reinstall the needle replacement window cover, and tighten the screw.
322 TOC-VCPH/CPN
6.8 Installation
6.8.4 Installing the Autosampler
Three Holes at
Turntable Center of Turntable
TOC-VCPH/CPN 323
6.8 Installation
6.8.4 Installing the Autosampler
CAUTION
• Verify that the drain tubing remains lower than the drain port of the ASI-V, and
that the end of the drain tubing remains above the surface of the liquid in the
drain tank.
• Corrosive substances such as acids may be present in the drainage liquid. Avoid
touching or spilling these liquids.
Preparation
• The rinse bottle is provided as a standard accessory.
• The drain bottle is a special accessory (P/N 638-52344). A suitable substitute may be
used as the drain container.
Procedure
1. Place the rinse bottle on the bench with the TOC-V and the autosampler.
2. Place the drain container on the floor.
3. To minimize resistance, keep the tip of the drain tubing above the surface of the drain
liquid and as close to the instrument as possible.
CAUTION
Do not allow any slack or bends in the tubing that could cause a buildup of water
pressure.
CAUTION
Perform the checks described above to ensure that rinse water will be delivered
and air will not be taken into the TOC-V injection pump.
324 TOC-VCPH/CPN
6.8 Installation
6.8.5 OCT-1 8-Port Sampler Installation
This section describes the procedure for installing the OCT-1 8-Port Sampler.
The unit is set up to the left of the TOC-V main body as shown in Figure 6.15“Installation
of the 8-Port Sampler”.
TIP » When 2 OCT-1 8-Port Samplers are to be used, place both units to the left of the TOC-V
main unit as shown in the overhead view of Figure 6.4“External Dimensions of Instrument
with OCT-1 8-Port Sampler”.
Cable Connections
The cable is connected between the main unit of the TOC-V and the 8-Port Sampler.
CAUTION
Confirm that the power to the TOC-V main unit is switched OFF before
performing this procedure.
Connection Procedure
1. Remove the 4 screws at the top of the 8-Port Sampler, and take off the cover.
2. Connect the power and signal cable to the circuit board of the 8-Port Sampler.
3. Connect the ground wire of the power cable to the grounding terminal on the circuit
board of the OCT-1.
4. Pass the cable through the rear panel, and replace the cover of the 8-Port Sampler.
TOC-VCPH/CPN 325
6.8 Installation
6.8.5 OCT-1 8-Port Sampler Installation
Rear Panel
Installation Procedure
8-Port Sampler 2
Signal Cable
8-Port Sampler 1
326 TOC-VCPH/CPN
6.8 Installation
6.8.5 OCT-1 8-Port Sampler Installation
3. Remove the port number label from the 8-Port Sampler 2 unit and replace it with the
port number label provided with the unit.
Note: The port that was originally labeled as port No. 1 should be labeled
as port No. 9.
4. Replace the covers on the 8-Port Samplers.
Sample Tubing
TOC-VCPH/CPN 327
6.8 Installation
6.8.6 Installing the TN Unit
CAUTION
The instrument may not operate properly if the power supply voltage exceeds
the range of AC100-127V ±10% or AC 220-240V ±10%. Unstable voltage may
cause problems during high sensitivity analysis.
Note: The instrument is compatible with both 50Hz and 60Hz power supplies and can
be used with either frequency.
¬Grounding
The included power cord is a three-conductor cable including a ground wire and a 2P plug
(with ground) at the end. If the power outlet does not have ground receptacle, use the
included 2P (no ground) adapter and ground the external ground tap terminal.
The voltage terminals (AC) and ground terminal (ACC) on the power plug of this
instrument are arranged as shown in Figure 6.19 "Power Plug". Connect them so that they
match the power outlet being used.
Voltage Terminal (AC) Ground Terminal (ACC)
328 TOC-VCPH/CPN
6.8 Installation
6.8.6 Installing the TN Unit
CAUTION
• In addition to the consideration of safety, complete grounding is necessary to
prevent as much as possible the occurrence of signal noise. A Type-3 ground
(ground resistance: 100W or less, see Electrical Facilities Engineering Standards
for details) is recommended.
• Perform grounding separately and avoid contact with water supply pipes, gas
pipes, and lightning rods.
The TNM-1 incorporates an internal ozone generator, requiring a source of air gas
(supplied at about 500mL/min) separate from the carrier gas supplied to the TOC-V. While
it is possible to use TOC-V high purity carrier gas by providing branch tubing, the
consumption of the combined-use carrier gas would increase to about 150 - 250mL/min.
Assuming operation at the rate of 8 hours/day, 5 days/week, the 47 L, 12MPa air cylinder
would require monthly replacement (as compared to every 3 months when used for carrier
gas alone).
Because air gas for ozone generation is not required to be the high purity air used as carrier
gas, it is recommended to use instrumentation air or compressor-supplied air for ozone
generation, separate from the carrier gas.
¬Using a High Purity Air-Filled Cylinder (Via Branch Tubing from Carrier Gas)
High purity air is synthesized air using pure nitrogen and pure oxygen. Ensure that
contaminants, including carbon dioxide, carbon monoxide and total hydrocarbons, are
guaranteed not to exceed 1ppm each.
Install a pressure regulator on the gas cylinder and supply gas to the instrument so that the
pressure is 300kPa at the instrument gas connection port.
Note: If there are excessive contaminants in the carrier gas, measurement results will be
adversely affected due to low measurement reproducibility, etc.
TOC-VCPH/CPN 329
6.8 Installation
6.8.6 Installing the TN Unit
Connection Inlet
Half Union
Nylon Tubing
330 TOC-VCPH/CPN
6.8 Installation
6.8.6 Installing the TN Unit
Procedure
1. Connect the half unions to the carrier gas inlet on the instrument and the connection
outlet (¼” NPT) of the gas supply source.
2. Connect the nylon tubing (4mm O.D., 2.5mm I.D.) between the gas supply source and
the instrument.
CAUTION
• Do not allow any acute bends in the tubing.
• When tubing and connectors are prepared by the user, be sure to use clean
components that are free of oils or other contaminants on their inner surfaces.
• Copper tubing or stainless steel tubing may also be used.
TOC-VCPH/CPN 331
6.8 Installation
6.8.6 Installing the TN Unit
Because the unit uses a chemiluminescence system for the detection of TN components, it
is equipped with an internal ozone generator. The ozone and NOx generated are removed
by the ozone treatment unit and NOx absorber, and are then exhausted. However, in the
event that the performance of the ozone treatment catalyst and NOx absorber deteriorate,
ozone and NOx gas may be expelled in the exhaust gas. Be sure to discharge exhaust gas to
the outdoors or to an exhaust pit using a 5mm I.D. flexible vinyl hose.
CAUTION
• To minimize resistance, leave the gas exhaust tubing open to the atmosphere.
• Be sure to connect the gas exhaust tubing to the instrument and discharge the
exhaust gas to the outdoors or an exhaust pit.
• If the gas exhaust tubing is not connected, ozone and NOx gas may be exhausted
from the instrument, causing harm to the respiratory organs of individuals in the
vicinity.
332 TOC-VCPH/CPN
6.9 Material Safety Data Sheets
6.9.1 Hydrochloric Acid (HCl)
TOC-VCPH/CPN 333
6.9 Material Safety Data Sheets
6.9.2 Phosphoric Acid
334 TOC-VCPH/CPN
6.9 Material Safety Data Sheets
6.9.3 CO2 Absorber, CA(OH)2, KOH, NAOH: soda lime
In case of contact: Immediately flush eyes or skin with copious amounts of water for at least 15 minutes while removing contaminated clothing and shoes. Assure
adequate flushing of the eyes by separating eyelids with fingers. If inhaled, remove to fresh air. If not breathing, give artificial respiration. If breathing is difficult,
give oxygen. If swallowed, wash out mouth with water provided person is conscious. CALL A PHYSICIAN. Wash contaminated clothing before reuse. Discard
contaminated shoes. For complete information, see actual entry in RTECS #VX965000.
SECTION IX SPILL OR LEAK PROCEDURES
Occupational spill: Evacuate area. Wear self-contained breathing apparatus, rubber boots and heavy rubber gloves. Absorb on sand or vermiculite and place in
closed containers for disposal. Ventilate area and wash spill site after material pickup is complete.
SECTION X SPECIAL PRECAUTIONS AND COMMENTS
The above information is believed to be correct but does not purport to be all inclusive and shall be used only as a guide. SHIMADZU shall not be held liable for
any damage resulting from handling or from contact with the above product. Users should make their own investigations to determine the suitability of the infor-
mation for their particular purposes.
TOC-VCPH/CPN 335
6.9 Material Safety Data Sheets
6.9.4 Lithium Hydroxide (monohydrate)
336 TOC-VCPH/CPN
6.9 Material Safety Data Sheets
6.9.5 Halogen Scrubber
Melt Boil Density Vapor Vapor Evaporation Odor Color Phase Water
Point Point Pressure Density Rate Solubility
1083°C 2595°C N/A N/A N/A N/A N/A Reddish Solid Malleable N/A
Metal
SECTION IV HAZARDS IDENTIFICATION
Dust, mist or fumes may cause eye and skin irritation. Avoid contact.
SECTION V TOXICITY DATA
RAT/MOUSE LD50 RTECS# OSHA PEL ACGIH TLV-TWA
N/A GL5325000 N/A N/A
SECTION VI HEALTH HAZARD DATA
Exposure to large amounts can cause gastrointestinal disturbances, nausea, vomiting, headaches, or dizziness.
SECTION VII FIRST-AID MEASURES
In case of contact, immediately flush eyes with copious amounts of water for at least 15 minutes. in case of skin contact, immediately wash skin with soap and
copious amounts of water. if ingested and subject is conscious, immediately give large amounts of water. get medical attention.
SECTION VIII FIRE FIGHTING MEASURES
EXTINGUISHING MEDIA: Appropriate to surrounding fire. Possible formation of toxic copper fumes. Use appropriate respiratory support.
SECTION IX ACCIDENTAL RELEASE MEASURES
N/A
SECTION X STABILITY AND REACTIVITY
INCOMPATIBILITIES: Strong acids, active halogens such as chlorine, fluorine, iodine, and bromine. ammonia.
SECTION XI SPECIAL PRECAUTIONS AND COMMENTS
The above information is believed to be correct but does not purport to be all inclusive. It shall serve only as a guide. Shimadzu shall not be held liable for any
damage resulting from handling or contact with the above product. Users should make their own determinations regarding the suitability of this information for
their particular purposes.
TOC-VCPH/CPN 337
6.9 Material Safety Data Sheets
6.9.6 Quartz Wool
338 TOC-VCPH/CPN
6.9 Material Safety Data Sheets
6.9.7 High Sensitivity Catalyst
TOC-VCPH/CPN 339
6.9 Material Safety Data Sheets
6.9.8 Normal Sensitivity Catalyst
ACUTE EFFECTS: May be harmful by inhalation or ingestion. May cause eye irritation. may cause skin irritation. Material may be irritating to
mucous membranes and upper respiratory tract. prolonged or repeated exposure may cause allergic reactions in certain
sensitive individuals.
EXPOSURE CAN CAUSE: Coughing, chest pains, difficulty in breathing, gastrointestinal disturbances.
CHRONIC EFFECTS: Prolonged exposure can cause damage to the lungs and bones. TO the best of our knowledge, the chemical, physical, and
toxicological properties have not been thoroughly investigated.
TARGET ORGAN DATA: Lungs, thorax or respiration (tumors) tumorigenic (neoplastic by RTECS criteria) tumorigenic (equivocal tumorigenic agent by
RTECS criteria) tumorigenic (tumors at site of application) only selected registry of toxic effects of chemical substances
(RTECS) data is presented here. see actual entry in RTECS.
SECTION VI HEALTH HAZARD DATA
OSHA PEL: 8H TWA 5MG/ME respirable fraction; 8h TWA 10MG/M3 total dust
SECTION VII FIRST-AID MEASURES
In case of contact, immediately flush eyes with copious amounts of water for at least 15 minutes while removing contaminated clothing and shoes. If inhaled,
remove to fresh air. If not breathing give artificial respiration. If breathing is difficult, give oxygen. If swallowed, wash out mouth with water provided person is
conscious. Call a physician. Remove and wash contaminated clothing promptly. In case of skin contact, immediately wash skin with soap and copious amounts of
water.
SECTION VIII FIRE FIGHTING MEASURES
EXTINGUISHING MEDIA: Carbon dioxide, dry chemical powder or appropriate foam.
UNUSUAL FIRE AND EXPLOSIONS HAZARDS: Material may have an exothermic reaction above 200 c with halocarbon vapors and may produce toxic
hydrochloric acid and phosgene.
SECTION IX ACCIDENTAL RELEASE MEASURES
Shut off all sources of ignition. Wear self-contained breathing apparatus, rubber boots and heavy rubber gloves. Sweep up, place in a bag and hold for waste dis-
posal. Ventilate area and wash spill site after material pickup is complete.
SECTION X STABILITY AND REACTIVITY
Stability: Stable.
Incompatibilities: Strong acids, strong bases, chlorine trifluoride, ethylene oxide, halocarbons, oxygen difluoride, sodium nitrate, vinyl acetate,
strong oxidizing agents,organic materials. Protect from moisture. Hazardous combustion or decomposition products. Nature of
decomposition products not known. Hazardous polymerization will not occur.
SECTION XI SPECIAL PRECAUTIONS AND COMMENTS
The above information is believed to be correct but does not purport to be all inclusive. It shall serve only as a guide. Shimadzu shall not be held liable for any
damage resulting from handling or contact with the above product. Users should make their own determinations regarding the suitability of this information for
their particular purposes.
340 TOC-VCPH/CPN
6.9 Material Safety Data Sheets
6.9.9 Potassium Nitrate
TOC-VCPH/CPN 341
6.9 Material Safety Data Sheets
6.9.10 Sodium Bicarbonate
342 TOC-VCPH/CPN
6.9 Material Safety Data Sheets
6.9.11 Sodium Carbonate
TOC-VCPH/CPN 343
6.9 Material Safety Data Sheets
6.9.12 Potassium Hydrogen Phthalate
344 TOC-VCPH/CPN
6.9 Material Safety Data Sheets
6.9.13 Water
6.9.13 Water
Shimadzu Scientific Instruments, Inc.
7102 Riverwood Drive Columbia, MD 21046 Control No.
Tel: 1(800) 477-1227 Date:
TOC-VCPH/CPN 345
6.9 Material Safety Data Sheets
6.9.13 Water
346 TOC-VCPH/CPN
Appendix A
Appendix A Method Validation
The Method Validation function tests the variance and the linearity of the data. The
function also determines performance characteristics according to the requirements of DIN
38402-Part 51. The Method Validation function displays the data used for executing
method validation, performs the required calculations and displays the results. Data can
either be imported directly from the TOC-Control V Sample Table, or entered manually
into the data table. Data can also be exported to ASCII files, stored and printed.
Method Validation
Main Window
The Main Window is separated into three sections. The upper left section of the window
displays the calibration curve, the upper right section displays the calculated validation
results and the lower portion displays the data table, which contains all relevant data for
performing the validation. The toolbar options (such as Open, Save, Cut, Copy, and Paste)
are similar to those described in Section 4.4.2 "File Menu" and Section 4.4.3 "Edit Menu".
The windows can be resized by dragging the dividers between the sections.
348 TOC-VCPH/CPN
Method Validation
File Menu
A.2.1 New
Use the New command to create a new, empty method validation file. A new file can also
be created by clicking the New toolbar button. If a file that contains modified and unsaved
data is open when the New command is selected, a prompt is displayed in a dialog box.
Select one of the options in the dialog box to determine if the unsaved data should be saved
before opening a new method validation file. Click the Yes button to save the document and
create a new method validation file. Click the No button to create a new document without
saving previous data.
TOC-VCPH/CPN 349
Method Validation
File Menu
A.2.2 Open
Use the Open command to open a saved method validation file (*.val). A file can also be
opened by clicking the Open toolbar button.
Start the method validation process by selecting a.cal file; application of DIN 38402 will
begin. The process can also be started by selecting the calibration curve row in the Sample
Table and opening the method validation function using the Tools>Method Validation
command.
When the Open command is selected, the File>Open dialog box is displayed. To open an
existing file, select the filename and choose the Open button. To create a new method
validation file, type the filename into the File name text field, then select the New button.
For further information on the standard functions of the Open dialog box, refer to Section
4.4.2.4 "Open".
Note: An error message is displayed if a file that is not in a correct, TOC-Control V for-
mat is selected.
350 TOC-VCPH/CPN
Method Validation
File Menu
A.2.3 Save
Use the Save command to save data in the current active file. If the file has not been saved,
selecting the Save command opens the Save As dialog box. The Save command can also be
executed by clicking the Save toolbar button.
A.2.4 Save As
The Save As command saves the active, current data to a new file name. Enter the desired
name in the File Name field. Two different file types are available: validation files (*.val)
used for storing the data in binary files and ASCII files (*.txt) used for exporting the data.
Select the desired file type from the Save As Type drop-down list. Use the Save In window
to select the location where the file should be saved. For details about the Save As
command, refer to Section 4.4.2.7 "Save As".
A.2.5 Print
Use the Print command to print data from a method validation file. Executing the Print
command opens the Print dialog box; displayed options vary based on the installed printer
driver. The Print command can also be executed by clicking the Print toolbar button.
Use the Print Range option to specify the range to be printed. The All selection prints all
pages in the file. Specific pages can be printed by selecting the Pages option and entering
the desired page numbers. Use the Number of Copies field to select the number of copies to
be printed.
TOC-VCPH/CPN 351
Method Validation
File Menu
Button Description
Next Page and Previous Page Click the Next Page button to display the next page of the
report. Click the Previous page button to display the previous
page of the report. The Next Page button is disabled when the
last page of the report is displayed, and the Previous Page
button is disabled when the first page of the report is dis-
played.
One Page/Two Page Click the One Page button to display one page at a time in the
Print Preview window. After selection, the button toggles to
the Two Page button. Click the Two Page button to display
two pages side by side.
Zoom In and Zoom Out Click the Zoom In and Zoom Out buttons to change the mag-
nification of the displayed pages. Three different zoom fac-
tors are available. Repeated selection of either option will
continue to increase or decrease the magnification, as appli-
cable.
Close Click the Close button to close the Print Preview window and
return to the previous window.The Print Preview window
must be closed to edit the report.
Print Click the Print button to print the report.
352 TOC-VCPH/CPN
Method Validation
File Menu
The print options available vary based on the printer installed. Select the paper size and
source from the drop-down lists, and select Portrait or Landscape to change the orientation
of the printed pages. Click the OK button to save the selections, or click the Cancel button
to close the dialog box without saving changes.
The Page Setup window consists of several tabs. Use the OK button to save changes and
exit the Page Setup window. Click the Cancel button to exit the Page Setup window without
saving changes. Click the Apply button to update the parameters without closing the Page
Setup window.
Note: On each tab in the Page Setup window, certain items are checked (selected) by
default. To deselect these options, click the item.
TOC-VCPH/CPN 353
Method Validation
File Menu
Data Tab
The Data tab displays a list of items that can be selected for printing. Select the desired
items and click the OK button.
Grid Tab
The Grid tab displays a list of format items for the report, as described below.
Option Description
Row Headers Prints row headers in the table.
Column Headers Prints column headers in the table.
Print Frame Prints a frame around the table.
Vertical Lines Prints vertical lines in the table.
Horizontal Lines Prints horizontal lines in the table.
Only Black and White Prints only in 2 colors, without gray backgrounds or other
colors.
First Rows, then Columns Prints rows first, followed by columns, when more than one
page is printed
First Columns, then Rows Prints columns first, followed by rows, when more than one
page is printed.
Vertical Prints the table on the vertical center of the page.
Horizontal Prints the table on the horizontal center of the page.
354 TOC-VCPH/CPN
Method Validation
File Menu
Miscellaneous Tab
The Miscellaneous tab displays a list of format items for the report, as described below.
Setting Item Description
Margins Use the spin controls to enter the page margins (in mm.).
Footer, Page No. Select this option to print the page number in the footer of
each page.
Header Organization: Select this option and enter the name of the
organization in the text box to print the organization name in
the page header.
Logo: Select this option to print a bitmap in the page header.
Use the Browse button to select the bitmap file.
Date/Time: Select this option to print the current system date
and time in the page header.
Operator: Select this option to print the current user name in
the page header.
A.2.9 Exit
Select Exit to close the Method Validation application. If an active file contains unsaved
information, a prompt to save changes is displayed. Select Yes or No, or select Cancel to
abort the exit function.
TOC-VCPH/CPN 355
Method Validation
Edit Menu
Options such as Cut, Copy, and Paste are available only if content is selected.
A.3.1 Undo
Use the Undo function to reverse the last change made in the data table. This function is
available only in documents that can be manually edited and when the user changes the
content of a cell in the data table.
A.3.2 Cut/Copy/Paste
Use the Cut, Copy, and Paste commands to edit highlighted table cells and rows. The Cut
command deletes information in rows or cells and stores it on the clipboard. The Copy
command copies data from rows or cells to the clipboard. The Paste command copies the
contents of the clipboard to the current row of the table. When pasting a whole row, select
the entire destination row first by clicking the row number.
Note: Do not select a row for pasting that currently contains data. The data will be over-
written!
TIP » The Cut, Copy and Paste commands have shortcut buttons on the toolbar.
356 TOC-VCPH/CPN
Method Validation
Edit Menu
A.3.7 Recalculate
Use the Recalculate command to recalculate all method validation data. The results and
calibration curve sections of the Method Validation window are also updated when the
Recalculate command is selected. Errors that occur during the recalculation process are
displayed in message boxes. For example, an error message may be displayed if the data in
the table does not contain sufficient information, if there are not enough data points defined
in the table, if less than 3 injections are available, and if numeric data is incorrect and
results in an illegal mathematical operation (such as division by zero).
TOC-VCPH/CPN 357
Method Validation
View Menu
Toolbar
The Toolbar function displays or hides the toolbar.
Status Bar
The Status Bar function displays or hides the status bar.
358 TOC-VCPH/CPN
Method Validation
Options Menu
The General Information dialog box is used to view and edit information about the method
validation file, such as the system used to create the file, the file version, the operator who
created the file and the date on which the file was created. Some information fields may be
disabled. Comments about the file are displayed in the Comment field.
TOC-VCPH/CPN 359
Method Validation
Options Menu
Item Description
Enable manual editing Select to allow data tables to be edited manually. This option
is disabled for files linked to TOC-Control V calibration
curves. If this option is deselected, it is disabled and cannot
be selected again. This option is selected by default.
Unit Displays the calibration curve units. This option is disabled if
the Enable manual editing option is not selected.
Analysis Type Displays the analysis used to create the calibration curve.
This option is disabled if the Enable manual editing option is
not selected.
Date/Time Displays the date and time the file was created.
Title Displays the title of the method validation file.
Calibration curve file name Displays the name of the TOC-Control V calibration curve.
360 TOC-VCPH/CPN
Method Validation
Options Menu
A.5.3 Font
Use the Font command to select fonts for printed reports and display screens. Selecting the
Font command opens the Font dialog box.
Choose a font for printed reports by selecting options from the Font, Font Style, Size, Color
and Script drop-down lists. A preview of the font appearance is visible in the Sample field
on the lower portion of the Font dialog box.
TOC-VCPH/CPN 361
Method Validation
Options Menu
Percent Values
Setting Item Description
Number of significant digits Select this option to display percent values using a specified
number of significant digits. Use the spin control to set the
number of significant digits. The default value is 4.
Number of digits after the Select this option to display percent values using a specified
decimal point number of digits after the decimal point. Use the spin control
to set the number of digits after the decimal point. The default
value is 2.
362 TOC-VCPH/CPN
Method Validation
Options Menu
To export a file, create or open a method validation file. Select the desired options from the
ASCII Export Options dialog box and select OK. The File>Save As dialog box is
displayed. Enter a file name for the ASCII file and select the *.txt file type.
TOC-VCPH/CPN 363
Method Validation
Options Menu
Figure A.22 Confirmation Message for Validation Strictly Following DIN 38402
Select Yes to execute recalculation. The calibration curve and result sections of the Method
Validation window are updated accordingly.
364 TOC-VCPH/CPN
Method Validation
Help Menu
TOC-VCPH/CPN 365
Method Validation
Help Menu
366 TOC-VCPH/CPN
Appendix B
Appendix B Control Charts
The Control Charts software monitors precision and accuracy of a defined data set. The
software supports five types of Control Charts:
•Mean Value Control Charts
•Recovery Control Charts
•Blind Value Control Charts
•Spanwidth Control Charts
•User Defined Control Charts
The Mean Value Control Chart checks the accuracy of an analytical method on an
instrument. Control value drifts may indicate a change in system performance.
The Recovery Control Chart tests an analytical method for matrix influences and provides a
restrictive control of accuracy by testing proportional system errors. Using certified
standards will test the accuracy of the system as well as the recovery rate.
The Blind Value Control Chart is a special form of the Mean Value Control Chart used for
quality assurance of the instrument and variable elements. Unlike the Mean Value Control
Chart, the control value is the information value (absorbance, peak height, peak area)
received from the analysis.
The Spanwidth Control Chart tests the precision of an analytical method and monitors the
accuracy under actual conditions (Drift Control). For this type of Control Chart, there are
several existing models for calculating the control and warning limits.
Each type of control chart can be used to create charts for different working range and
analysis methods. For each control chart, a file is created which contains general
information about the type, data structure and other global data.
Control Charts
File Menu
B.1.1 New
The New function is used to define new control charts and displays the Control Chart
Wizard. Follow the Control Chart Wizard process to enter a new control chart. The file
extension *.cc is used for all control charts.
Note: Control Chart parameters can be changed at a later time by selecting
Options>Control chart options from the main menu bar.
368 TOC-VCPH/CPN
Control Charts
File Menu
TOC-VCPH/CPN 369
Control Charts
File Menu
Figure B.5 Control Chart Wizard: Recovery Control Chart Parameter Window
370 TOC-VCPH/CPN
Control Charts
File Menu
Figure B.6 Control Chart Wizard: Spanwidth Control Chart Options Window
TOC-VCPH/CPN 371
Control Charts
File Menu
372 TOC-VCPH/CPN
Control Charts
File Menu
B.1.3 Open
The Open command is used to open files. Select a *.cc file to open a control chart file.
Refer to Section 4.4.2.4 "Open" for further information on the standard File>Open dialog
box.
B.1.4 Save
The Save command saves the current control chart.
B.1.5 Save As
Use the Save As command to save files for the first time, or to save a file using a new file
name. Files can be saved as file types *.cc or *.txt (ASCII format). For more information
on the Save As function, refer to Section 4.4.2.7 "Save As".
B.1.6 Print>Data
Use the Print>Data command to print data in spreadsheet format. For more information on
the Print function, refer to Section 4.4.2.10 "Print>Table".
B.1.7 Print>Graph
Use the Print>Graph command to print data in graphical format. For more information on
the Print function, refer to Section 4.4.2.10 "Print>Table".
TOC-VCPH/CPN 373
Control Charts
File Menu
The print options available vary based on the printer installed. Select the paper size and
source from the drop-down lists, and select Portrait or Landscape to change the orientation
of the printed file. Select OK to save the selections, or Cancel to close the dialog box
without saving changes.
The Page Setup window consists of several tabs. Use the OK button to save changes and
exit the Page Setup window. Click the Cancel button to exit the Page Setup window without
saving changes. Click the Apply button to update the parameters without closing the Page
Setup window.
Note: On each tab in the Page Setup window, certain items are checked (selected) by
default. To deselect these options, click the item.
374 TOC-VCPH/CPN
Control Charts
File Menu
Data Tab
The Data tab displays a list of items that can be selected for printing. Select the desired
items and click the OK button.
Grid Tab
The Grid tab displays a list of format items for the report.
Option Description
Row Headers Prints row headers in the table.
Column Headers Prints column headers in the table.
Print Frame Prints a frame around the table.
Vertical Lines Prints vertical lines in the table.
Horizontal Lines Prints horizontal lines in the table.
Only Black and White Prints only in 2 colors, without gray backgrounds or other
colors.
First Rows, then Columns Prints rows first, followed by columns, when more than one
page is printed
First Columns, then Rows Prints columns first, followed by rows, when more than one
page is printed.
Vertical Prints the table on the vertical center of the page.
Horizontal Prints the table on the horizontal center of the page.
TOC-VCPH/CPN 375
Control Charts
File Menu
Miscellaneous Tab
The Miscellaneous tab displays a list of format items for the report, as described below.
Setting Item Description
Margins Use the spin controls to enter the page margins (in mm).
Footer, Page No. Select this option to print the page number in the footer of
each page.
Header Organization: Select this option and enter the name of the
organization in the text box to print the organization name in
the page header.
Logo: Select this option to print a bitmap in the page header.
Use the Browse button to select the bitmap file.
Date/Time: Select this option to print the current system date
and time in the page header.
Operator: Select this option to print the current user name in
the page header.
B.1.11 Exit
Select Exit to close the Control Chart application.
376 TOC-VCPH/CPN
Control Charts
Edit Menu
Copy
Use the Copy command to copy the selected range from the data table to the clipboard.
Recalculate
Use the Recalculate command to recalculate the limits of the control chart. This function
should be performed after one or more control samples have been excluded or added to the
control chart.
Exclude
Use the Exclude command to exclude a selected range of control samples from the control
chart calculations. When Exclude is selected, the Exclude dialog box is displayed. Enter a
comment describing the reason for excluding the sample. The Exclude function is only
available when one or more control samples are selected. A dialog box displays a message
requesting confirmation that the sample be excluded. Select Yes to confirm and recalculate
the control limits, or No to abort recalculation.
TOC-VCPH/CPN 377
Control Charts
View Menu
B.3.1 Statistics
The Statistics command displays the Statistics window.
378 TOC-VCPH/CPN
Control Charts
View Menu
B.3.2 Comparison
After a control chart is completed, the results from the control period must be compared
with the results from the preparation period. For control charts without a preparation
period, the current values must be compared with the results of the preparation period or the
control period from the previous control chart. The View>Comparison command displays
the Comparison between control charts window.
TOC-VCPH/CPN 379
Control Charts
View Menu
Spl. No.
Displays the sample number for which the out of control event occurred.
Event
Describes the out of control event.
380 TOC-VCPH/CPN
Control Charts
Options Menu
B.4.2 Font
The Font command opens the standard Font dialog box. Refer to Section 4.4.9.3 "Display
Settings>Display Font" for detailed information about font options.
TOC-VCPH/CPN 381
Control Charts
Options Menu
Percent Values
Setting Item Description
Number of significant digits Select this option to display percent values using a specified
number of significant digits. Use the spin control to set the
number of significant digits. The default value is 4.
Number of digits after the Select this option to display percent values using a specified
decimal point number of digits after the decimal point. Use the spin control
to set the number of digits after the decimal point. The default
value is 2.
382 TOC-VCPH/CPN
Control Charts
Options Menu
To export a file, create or open a control chart. Select the desired options from the ASCII
Export Options dialog box, and select OK. The File>Save As dialog box is displayed. Enter
a file name for the ASCII file and select the *.txt file type.
TOC-VCPH/CPN 383
Control Charts
Help Menu
About ContrCharts...
Use the About command to view copyright and version information.
384 TOC-VCPH/CPN
Index
Numerics
8-Port Sampler Installation .......................................................325
Analysis Using ...................................................80 Maintenance............................................ 254–255
Connecting Two Units .....................................326 Rear View .......................................................... 15
Connection .......................................................327 Sealing the Sample Containers ..........................80
Front View .........................................................15 8-Port Valve
High Sensitivity..................................................82 Replacing Rotor (8-Port Sampler) ...................254
Initializing ........................................................267 Replacing Rotor (Syringe Pump).....................246
A
Accessories...................................................... 298–303 Application Operations ................................... 219–233
Special..............................................................303 ASCII Export Options
Standard ...........................................................298 Data Tab...........................................................147
Acid Header Tab.......................................................146
Addition in NPOC..................................... 76, 189 Miscellaneous Tab ...........................................148
Addition Volume..............................................139 System Tab ......................................................147
Checking Volume ............................................236 ASCII File, Exporting Control Charts .....................383
Effect on Instrument...........................................71 ASI-V, See Autosampler
Hydrochloric ....................................................333 Auto Dilution ..........................................132, 139, 189
Phosphoric........................................................334 Auto Generate ..........................................................181
Preparation .........................................................51 Auto Regeneration of IC Solution ............................. 27
Add Event Log ...........................................................22 Auto Restart .............................................................121
Administration .................................................... 19–33 Autosampler.............................................................295
Air, Low Pressure Safety .........................................331 Cables, Connecting ................................. 318, 319
Alkaline Samples .......................................................71 Cover Installation.............................................323
Analysis Drain Tank Installation .................................... 323
Blank Check.......................................................66 Front View .........................................................14
Calibration Parameters.....................................128 High Sensitivity Analysis ..................................75
Calibration Run ................................................100 Initializing........................................................ 267
Control Sample Parameters..............................196 Installation .......................................................317
Creating New Method......................................135 Maintenance............................................ 252–253
Ending ..............................................................120 Needle Change.................................................268
Gas Sample ......................................................227 Needle Types .............................................. 28, 85
Group Of Samples............................................181 Needle, Coaxial................................................322
High Sensitivity..................................................75 Needle, Sample ................................................321
IC........................................................................56 NPOC Analysis..................................................76
Individual Sample ............................................185 Positioning .......................................................317
Manual Injection ..............................................219 Rear View .......................................................... 14
Method Parameters ..........................................136 Rinse Bottle Inspection.................................... 252
New ..................................................................125 Rinse Bottle Installation...................................323
New Catalyst ......................................................36 Rinse Pump Head Replacement.......................253
POC....................................................................58 Sample Catcher Inspection ..............................252
Preparation .................................................. 36–82 Shipping Screw ................................................317
Principles................................................. 284–288 Software Settings ........................................ 28, 84
Procedures................................................. 89–121 Specifications...................................................295
Sample Parameters...........................................186 Tray Type...........................................................28
Sample Run ......................................................108 Troubleshooting ...................................... 278–280
Suspended Solids ...............................................69 Turntable Installation.......................................323
Technical Information............................. 289–293 Vial Positions...................................................179
TN ......................................................................59 Vial Rack Change ............................................267
Viewing Data in Real Time .............................167 Vial Rack Installation ......................................323
B
Background Monitor ................................................203 Blank, System ..........................................................291
Base, Effect on Instrument.........................................71 Buzzer, Setting in Software .......................................84
Blank Check...................................................... 68, 269
Analysis..............................................................66
Using Area in Calculations ............ 139, 189, 198
Index-1
Index
C
Cable Connections ................................................... 325 Setting Type ...................................................... 26
Cables, Connecting ......................................... 318, 319 Standard............................................................. 36
Calibration Curve .................................................... 290 TOC ................................................................... 37
1-Point and 2-Point .......................................... 290 TOC/TN............................................................. 39
Calculation Method ......................................... 171 Washing........................................................... 239
Creating .................................................. 126, 132 Ceramic Fiber ......................................................39, 42
Entering Calibration Standard Levels.............. 132 Change Password ...................................................... 21
Gas Sample Analysis ....................................... 229 Dialog Box ........................................................ 21
High Sensitivity Analysis .................................. 65 Chemiluminescence Detector.................................. 259
Inserting into Method ...................................... 137 CO2 Absorber .......................................................... 335
Inserting into Sample Table............................. 126 Disposal ........................................................... 242
Inserting Into Sample Wizard.......................... 187 Installation ...................................................52, 58
Manual Injection Kit........................................ 221 MSDS .............................................................. 335
Multi-Point....................................................... 290 Replacing........................................ 242, 248, 251
Optimum .......................................................... 292 Used as NOx Absorber...................................... 59
Options............................................................. 128 Coaxial Needle ........................................... 28, 85, 320
Points List ........................................................ 131 Com Port.................................................................... 30
Points Parameters ............................................ 132 Combustion Tube
Shifting To Blank Point ................................... 291 Connecting......................................................... 40
Shifting to Origin............................129, 172, 290 Filling with Catalyst ............................. 37, 38, 39
Types ...................................................... 128, 290 Installating H Type ............................................ 40
Viewing Data ................................................... 160 Installating N Type ............................................ 43
Viewing Properties .......................................... 170 Replacing......................................................... 242
Calibration Curve Wizard............................... 100, 126 Replacing Ferrules........................................... 245
Analysis Information Window ........................ 128 Washing........................................................... 242
Analysis Options.............................................. 102 Comment ................................................................... 22
Calibration Points List ..................................... 103 Communication Settings ........................................... 30
Calibration Points Window ............................. 131 Configured System, Deleting .................................... 33
Curve Type ...................................................... 101 Connecting Gas Tubing (TN Unit).......................... 330
Curve Type Window ....................................... 128 Connection
Default Measurement Parameters.................... 102 8-Port Sampler................................................. 327
History ............................................................. 104 Cables .............................................................. 325
History Window .............................................. 134 Two 8-Port Samplers....................................... 326
Measurement Parameters Window .................. 130 Consumable Parts ............................................304–307
Opening ........................................................... 125 Control Chart Wizard .............................................. 368
Peak Integration Parameters ............................ 103 Control Charts ........................................ 208, 367–384
Peak Time Parameters Window ...................... 133 Comparing ....................................................... 379
System Information Window........................... 127 Creating ........................................................... 368
Carrier Gas Editing .....................................................377, 381
Connecting....................................................... 313 Excluding Samples From ........................377, 383
Flow Rate........................................................... 90 Exporting To ASCII File ................................. 383
Pressure.............................................................. 89 Following Chart............................................... 372
Purification Tube ............................................. 240 Opening ........................................................... 373
Purifier ............................................................. 314 Out Of Control Events..................................... 380
Kit Specifications .................................... 296 Printing ............................................................ 373
Replacing CO2 Absorber ......................... 248 Recalculating Limits........................................ 377
Supply Pressure ............................................... 314 Saving .............................................................. 373
Tubing Connections......................................... 314 Setting Number Format ................................... 382
Catalyst Statistics........................................................... 378
Disposal ........................................................... 240 Types ............................................................... 367
Filling................................................................. 36 Control Checking .................................................... 200
Filling Rod .................................................. 37, 38 Control Sample
High Sensitivity ........................................ 37, 240 Identifying ....................................................... 184
New.................................................................... 36 Inserting Into Sample Table ............................ 202
Regenerating ........................................... 238, 262 Selecting Type ................................................. 194
Replacing ......................................................... 239 Template .......................................................... 193
Index-2
Index
Control Sample Wizard Corrosion Warning ......................................................x
Analysis Parameters Window ..........................196 Cover, Installation....................................................323
Control Checking Window ..............................200 Creating
Control Sample Type Window ........................194 Calibration Curve.............................................126
Creating A Template........................................193 Control Charts..................................................368
Failure Action Window....................................201 Control Sample Template ................................193
History Window...............................................199 Method .............................................................135
Injection Parameters Window..........................197 Method Validation File.................................... 349
Opening............................................................126 Sample Table ...................................................125
Parameter Source Window...............................195 TC Calibration File ..........................................100
Peak Time Parameters Window.......................198 CV Max
System Information Window ...........................193 Setting in Method Wizard................................138
Cooler Drain Container..............................................48 Setting In Sample Wizard ................................188
Corrective Action
Eliminate Syringe Bubbles...............................281
Poor Reproducibility ........................................280
Poor Sample Injection......................................280
D
Data Determinations, No. of............................136, 162, 186
Deleting............................................................158 Dialog Box
Exporting to Database......................................148 Change Password...............................................21
Importing from ASCII .....................................159 Dilution Water
Recalculating....................................................158 Checking ..........................................................236
Replacing .........................................................157 Preparing............................................................50
Database, Exporting Data to ....................................148 Display Settings
Default Measurement Parameters ..................... 87, 212 Floating Point Number Formats ......................216
Default Measurement Parameters Dialog Box...........98 Notification Bar Settings .................................215
Deleting Table Settings ..................................................214
All Sample Table Data.....................................158 Disposal
Configured System.............................................33 Catalyst ............................................................240
Selected Sample Table Data ............................159 CO2 Absorber ..................................................242
Description, System .............................................. 7–18 Drain Tank, Installation ...........................................323
Detector Drain Tubing, Connecting .......................................316
Chemiluminescence .........................................259 Drain Vessel Water, Checking.................................236
Viewing Status .................................................203
E
Edit Menu, Sample Table Editor..............................156 Excluding
Editing Data from Sample Table ..................................158
Calibration Point Parameters ...........................132 Injections from Calculations............................158
Control Charts ......................................... 377, 381 Samples From Control Charts................. 377, 383
Method Validation File ....................................356 Exporting
Sample Table....................................................156 Control Charts..................................................383
Electric Shock Warning ........................................... viii Data to ASCII File ...........................................146
Ending Measurement ...............................................120 Data to Database ..............................................148
Error Messages................................................ 270–274 Method Validation Data...................................363
F
File Menu, Sample Table Editor ..............................125 Flow Diagrams.................................................... 11–13
Flow charts, Troubleshooting ..................................274 Flow Line
Flow Diagram Replacing Content............................................263
TOC-VCPH .........................................................11 Replacing Parts and Slider.................................54
TOC-VCPH+POC OPTION ..................................12 Residue Removal .............................................263
TOC-VCPH+TNM-1...........................................12 Setting Washes in Software ...............................86
TOC-VCPN .........................................................13 Washing ...........................................................264
TOC-VCPN+TNM-1...........................................13
Index-3
Index
Flow Rate Front View
Carrier Gas......................................................... 90 8-Port Sampler................................................... 15
Ozone Source..................................................... 91 Autosampler ...................................................... 14
Sparge Gas ......................................................... 53 TOC-V................................................................. 8
Sparge Gas, Wet Chemical .............................. 269 Furnace ...................................................................... 90
Power................................................................. 90
Temperature.................................................27, 84
G
General Information Window.................................. 125 Ground, Connecting ................................................ 312
General Measurement Parameters ...................... 83–88 Grounding........................................................312, 328
Graph, in Sample Window ...................................... 167 Grubb Test, See Outlier Test
H
Halogen Scrubber History Log
MSDS .............................................................. 337 Calibration ....................................................... 134
Replacing ......................................................... 243 Control Sample................................................ 199
High Purity Air Cylinder ................................ 242, 313 Instrument.......................................................... 31
High Sensitivity Method............................................................. 142
Analysis ............................................................. 63 History Settings, Maintenance .................................. 88
Catalyst .............................................37, 240, 339 Humidifier
Combustion Tube .............................................. 40 Checking Water Level..................................... 237
Using 8-Port Sampler ........................................ 82 Filling with Water ............................................. 49
High Temperature Warning.......................................vii Hydrochloric acid .................................................... 333
I
IC ............................................................................. 285 Installation
Analysis Preparation.......................................... 56 8-Port Sampler................................................. 325
Injection Port Septum ...................................... 226 Autosampler .................................................... 317
Principles ......................................................... 285 Cover ....................................................... 323
Reaction Solution ..................................... 57, 249 Drain Tank............................................... 323
Reaction Solution Pump .................................. 250 Needle, Coaxial ....................................... 322
Reaction Solution Regeneration .........27, 84, 262 Needle, Sample........................................ 321
Removal Pretreatment ....................................... 68 Positioning............................................... 317
Standard Solution .............................................. 61 Rinse Bottle ............................................. 323
Injection Parameters Shipping Screw........................................ 317
Control Sample ................................................ 197 Turntable ................................................. 323
Method............................................................. 138 Vial Rack ...........................................74, 323
Sample ............................................................. 188 CO2 Absorber ..............................................52, 58
Injection Port Septum, Replacing ............................ 226 Combustion Tube
Injection Table, in Sample Window ........................ 168 High Sensitivity......................................... 40
Injection Volume, Setting............................... 138, 188 N Type....................................................... 43
Injury Warning ........................................................... ix Manual Injection Kit ....................................... 219
Inorganic Carbon, See IC Preparation....................................................... 310
Insert Menu, Sample Table Editor........................... 181 Procedure......................................................... 312
Inside Front View Site................................................................... 310
TN ...................................................................... 17 Suspended Solids Kit......................................... 54
TOC-V ............................................................... 10 Syringe............................................................... 45
Inside Top View TN Unit............................................................ 328
TOC-V ............................................................... 11 TOC-V.....................................................310–332
Inspection
Daily ....................................................... 236–237
Periodic ................................................... 238–251
Index-4
Index
Instrument Instrument Setup Wizard
Auto Restart .....................................................121 ASI Window ............................................... 28, 95
Background Monitor ........................................203 Auto Regeneration of IC Solution .....................27
Connecting And Disconnecting .......................205 Autosampler Parameters ....................................28
Detector Status .................................................203 Buzzer ................................................................27
Effects of Acid, Base, and Salt ..........................71 Cell Length ........................................................26
History Log ........................................................31 Communication Window...................................30
Maintenance .....................................................261 Configuration .....................................................24
Menu, Sample Table Editor .............................203 Furnace Temperature .........................................27
Operation................................................... 35–233 History Log ........................................................31
Overview......................................................... 1–5 Needle ................................................................28
Power .................................................................89 Options Window ......................................... 24, 94
Setup............................................................ 23–33 Oxidation Method ..............................................24
Shut Down Procedure ......................................120 Ready Status Check ...........................................27
Standby Window..............................................206 SSM Window.....................................................29
Starting ...................................................... 89, 207 System Information Window...................... 23, 94
Viewing Properties.............................................32 TOC Window.....................................................26
Instrument Properties Dialog Box..............................32 Tubing Diameter ................................................26
Introduction.................................................................iii
System Administration ...................................... 20
L
Left Side View Lithium hydroxide, MSDS ......................................336
TOC-V .................................................................9
M
Maintenance .................................................... 235–281 Measurment
8-Port Sampler ........................................ 254–255 8-Port Sampler ...................................................80
Autosampler ............................................ 252–253 Mechanical Check....................................................265
Chemiluminescence Detector ..........................259 Method
Daily Inspection ...................................... 236–237 Creating New ...................................................135
Gas Sample Analysis .......................................232 Viewing Properties ..........................................161
History..............................................................269 Method Properties
History Settings..................................................88 Common Tab ...................................................161
Log ...................................................................269 History Tab ......................................................166
Manual Injection Kit ........................................226 NPOC and POC Tabs ......................................165
Mechanical Check............................................265 Parameter Tab ..................................................162
Menu ................................................................261 TC Tab .............................................................163
Parts......................................................... 308–309 TN Tab.............................................................165
Periodic Inspection.................................. 238–251 Method Validation ..................................208, 347–365
Software-Controlled................................ 261–269 According to DIN ................................... 347, 364
TN ........................................................... 256–260 Creating New File............................................349
Warning Message Limits ...................................88 Editing File ......................................................356
Manual Contents Overview .........................................v Exporting Data To ASCII ................................363
Manual Injection Kit ................................................296 Main Window ..................................................348
Analysis............................................................219 Opening File ....................................................350
Calibration........................................................221 Printing File .....................................................351
Installation........................................................219 Saving File .......................................................351
Maintenance .....................................................226 Setting Number Display Format ......................362
Specifications ...................................................296 Viewing Parameters.........................................359
Material Safety Data Sheets, See MSDS
Index-5
Index
Method Wizard MSDS ..............................................................333–345
Acid Addition Volume .................................... 139 CO2 Absorber .................................................. 335
Analysis Information ....................................... 136 Halogen Scrubber ............................................ 337
Auto Dilution ................................................... 139 High Sensitivity Catalyst................................. 339
Calibration Curve ............................................ 137 Hydrochloric acid ............................................ 333
Creating New Method ..................................... 135 Lithium hydroxide........................................... 336
CV Max ........................................................... 138 Normal Sensitivity Catalyst............................. 340
Determinations................................................. 136 Phosphoric acid ............................................... 334
History Log...................................................... 142 Potassium hydrogen phthalate......................... 344
Injection Parameters ........................................ 138 Potassium nitrate ............................................. 341
Injection Volume ............................................. 138 Quartz Wool .................................................... 338
Multiple Injections........................................... 139 Sodium bicarbonate......................................... 342
Number of Injections ....................................... 138 Sodium carbonate ............................................ 343
Opening ........................................................... 126 Water ............................................................... 345
Peak Integration Parameters ............................ 140 Multiple Injections ................................. 129, 139, 189
SD Max............................................................ 138 Multi-Point Calibration Curve................................. 290
System Information ......................................... 135
Using Blank Check Area ................................. 139
USP/EP ............................................................ 141
N
Needle Notation Conventions................................................. vi
Change ............................................................. 268 Notification Bar ...............................................124, 215
Coaxial........................................................ 28, 85 NOx Absorber
Coaxial+Sparge .......................................... 28, 85 Connecting......................................................... 59
Installation, Coaxial......................................... 322 Replacement .................................................... 260
Installation, Sample ......................................... 321 Using a CO2 Absorber....................................... 59
Rinse .................................................................. 85 NPOC
Rinse after Acid Addition.................................. 86 Acid Addition .................................................... 76
Sample ........................................................ 28, 85 Analysis ............................................................. 76
Sample+Sparge........................................... 28, 85 Principles .................................................286, 288
Setting in Software ..................................... 28, 85 Number of Injections.......................................138, 188
Types ..................................................28, 85, 320
O
OCT-1/2, See 8-Port Sampler Options Menu, Sample Table Editor ....................... 210
Opening Outlier Test.............................................................. 169
Calibration Curve Wizard................................ 125 Oxidation Method, Selecting..................................... 24
Control Sample Wizard ................................... 126 Ozone
Method Validation File.................................... 350 Connecting Source .......................................... 329
Method Wizard ................................................ 126 Generator Replacement ................................... 260
Sample Group Wizard ..................................... 181 Generator, Low Air Supply Safety.................. 331
Sample Table ....................................96, 125, 143 Generator, Source Air Pressure ....................... 330
Sample Wizard ................................................ 185 Source Airflow Rate .......................................... 91
Sparge Gas Valve ............................................ 269 Treatment Unit ................................................ 256
P
Page Setup Parameter Source..................................................... 185
Calibrations Tab .............................................. 152 Parameter Table, in Sample Window...................... 167
Header Tab ...................................................... 153
Instrument Tab................................................. 152
Miscellaneous Tab ........................................... 154
Samples Tab .................................................... 153
Index-6
Index
Parts Precautions...................................................................3
Consumable Gas Sample Analysis .......................................227
ASI-V .......................................................306 High Sensitivity Analysis ..................................63
Carrier Gas Purifier Kit............................307 Preparation
TNM-1 .....................................................307 Acid....................................................................51
TOC-VCPH ...............................................304 Analysis ...................................................... 36–82
TOC-VCPN ...............................................305 Dilution Water ...................................................50
Maintenance High Sensitivity Analysis ........................... 63, 65
ASI-V .......................................................308 Manual Injection Analysis ...............................219
Manual Injection Kit ................................309 Sample ..................................................63, 72–79
TOC-VCPH ...............................................308 Standard Solution...............................................60
TOC-VCPN ...............................................308 TN Analysis .......................................................59
Peak Pressure
Area..................................................................289 Ozone Generator Source Air............................330
Integration Parameters ...........133, 140, 190, 198 Principles
Shape................................................................289 IC .....................................................................285
Start Delay........................................................140 NPOC...................................................... 286, 288
Total Integration Time .....................................140 Of Analysis ............................................. 284–288
Phosphoric acid ........................................................334 POC..................................................................287
Plunger Tip Backlash ...............................................281 POC+NPOC.....................................................288
POC TN ....................................................................288
Analysis Preparation ..........................................58 TOC ................................................................. 284
kit .....................................................................296 Print
Principles..........................................................287 Control Charts..................................................373
Standard Solution...............................................61 Method Validation File.................................... 351
POC+NPOC Principles ............................................288 Options.............................................................151
Post Sale Service.........................................................iv Page Setup........................................................ 152
Potassium hydrogen phthalate, MSDS.....................344 Preview, Report ...............................................151
Potassium nitrate, MSDS .........................................341 Preview, Table .................................................150
Power Sample Report..................................................150
Furnace...............................................................90 Sample Table ...................................................149
Instrument ..........................................................89 Product Warranty........................................................iv
TN ......................................................................84 Purgeable Organic Carbon, See POC
Power Supply
Changing Voltage ............................................313
Connecting .............................................. 312, 328
Q
Quartz Wool.........................................37, 39, 58, 338
R
Ready Status Check ............................................ 27, 84 Regenerating
Rear View Catalyst ................................................... 238, 262
8-Port Sampler ...................................................15 IC Solution................................................ 27, 262
Autosampler .......................................................14 Replacing Sample Table Data..................................157
TN ......................................................................18 Replacing the NOx Absorber ...................................260
TOC-V .................................................................9 Residue Removal .....................................................263
Recalculating Right Side View, TOC-V.............................................8
All Sample Table Data.....................................158 Rinse
Control Chart Limits ........................................377 Bottle....................................................... 252, 323
Selected Sample Table Data ............................158 Needle ................................................................85
Reference Information .................................... 283–345 Needle, After Acid Addition..............................86
Pump Head.......................................................253
Index-7
Index
S
Salt, Effect on Instrument .......................................... 71 Printing Sample Table ..................................... 149
Sample Recalculating Data .......................................... 158
Alkaline ............................................................. 71 Replacing Data ................................................ 157
Catcher............................................................. 252 Reversing Last Action ..................................... 156
Container ........................................................... 80 Sample Window ......................................114, 167
Containing Acid................................................. 71 Saving .............................................................. 144
Containing High IC Levels................................ 38 Saving As ........................................................ 145
Containing Salt .................................................. 71 Searching for Text ........................................... 157
Group ............................................................... 181 Sequence Window........................................... 205
High Sensitivity .......................................... 63, 65 Setting Vial Position................................106, 179
Loading .............................................................. 80 Sparging/Acid Addition Window.................... 106
Manual Injection.............................................. 223 Standby Window ............................................. 206
Needle .................................................28, 85, 320 Starting Instrument .......................................... 207
Preparation............................................63, 72–79 Status Bar ................................................124, 178
Report, Printing ............................................... 150 Stopping the Instrument .................................. 207
Suspended Solids ............................................... 69 Toolbar ....................................................122, 178
Sample Group Wizard Toolbar, Customize ......................................... 177
Calibration Curve ............................................ 183 View Menu ...................................................... 160
Controls ........................................................... 184 Viewing Calibration Curve .....................160, 170
Measurement Parameters................................. 182 Viewing Cell Properties .................................. 170
Opening ........................................................... 181 Viewing Data in Real Time............................. 167
Parameter Source ............................................. 181 Viewing Method Properties............................. 161
Sample Table Editor ....................................... 122–218 Viewing Vial Positions.................................... 179
Auto Restart ..................................................... 121 Window Menu ................................................. 218
Background Monitor........................................ 203 Sample Window ...................................................... 114
Closing a File................................................... 144 Graph ............................................................... 167
Connecting The Instrument ............................. 205 Injection Table................................................. 168
Creating New Sample Table..................... 98, 125 Parameter Table............................................... 167
Deleting All Data............................................. 158 Sample Wizard
Deleting Selected Data .................................... 159 Acid Addition .................................................. 189
Directories ....................................................... 217 Analysis Information Window ........................ 186
Edit Menu ........................................................ 156 Auto Dilution................................................... 189
Editing Rows and Cells ................................... 156 Calibration Curve ............................................ 191
Excluding Data ................................................ 158 Calibration Curve Window ............................. 187
Exiting ............................................................. 155 CV Max ........................................................... 188
Exporting Data to ASCII ................................. 146 Determinations ................................................ 186
Exporting Data to Database ............................. 148 Injection Parameters ........................................ 191
File Menu......................................................... 125 Injection Parameters Window ......................... 188
Finding Cell Content ....................................... 157 Injection Volume ............................................. 188
General Information Window.......................... 125 Multiple Injections .......................................... 189
Help Menu ....................................................... 218 Number of Injections....................................... 188
Importing Data from ASCII ............................ 159 Parameter Source Window .............................. 185
Insert Menu...................................................... 181 Peak Time Parameters Window ...................... 190
Inserting Calibration Curve ............................. 126 SD Max............................................................ 188
Inserting Control Samples ............................... 193 Units ................................................................ 188
Inserting Sample Group................................... 181 USP/EP Window ............................................. 192
Inserting Samples ............................................ 185 Screen Lock ............................................................... 20
Instrument Menu.............................................. 203 SD Max............................................................138, 188
Notification Bar ...................................... 124, 215 Sequence Window ................................................... 205
Opening File .................................................... 143 Shipping Screw........................................................ 317
Options Menu .................................................. 210 Sliding Sample Injector ........................................... 245
Outlier Test ...................................................... 169 Sodium bicarbonate, MSDS .................................... 342
Overview ......................................................... 122 Sodium carbonate, MSDS ....................................... 343
Page Setup Options.......................................... 152
Previewing Sample Report .............................. 151
Previewing Sample Table ................................ 150
Printing Options............................................... 151
Printing Sample Report ................................... 150
Index-8
Index
Software Standard Solution.......................................................60
Connecting To Instrument ...............................205 Analysis ...........................................................100
Creating New Method......................................135 High Sensitivity .................................................64
Maintenance Functions ....................................261 Preparing............................................................60
Maintenance History Settings ............................88 Sparging ...........................................................292
Module .............................................................122 Storing................................................................62
Overview..................................................... 19–33 Standby Window............................................. 120, 206
Ready Status Check ...........................................84 Starting The Instrument ...........................................207
Setting Measurement Parameters................ 83, 87 Starting the Instrument...............................................89
Setting up a Sequence ......................................125 Status Bar........................................................ 124, 178
Source Air (TN Unit) Stirrer .................................................................. 78, 86
Pressure ............................................................330 Stopping the Instrument
Sparge Halt...................................................................207
Gas Flow Rate....................................................53 Stop Finish Current..........................................207
Gas Flow Rate, Wet Chemical.........................269 Stop Peak Stop .................................................207
Gas Valve.........................................................269 Suspended Solids ................................................ 54, 69
Needle ....................................................... 28, 320 Syringe
NPOC Analysis ..................................................77 Changing................................................. 226, 268
Time, Setting....................................................139 Installation .........................................................45
Sparging/Acid Addition Window ............................106 Manual Injection Kit........................................226
Special Accessories..................................................303 Plunger Tip ......................................................244
Specifications .................................................. 294–297 Plunger Tip Backlash.......................................281
Autosampler .....................................................295 Replacing ...........................................................45
Carrier Gas Purifier Kit....................................296 Residue Removal .............................................263
Gas Sample Injection Kit .................................296 Washing with Sample .....................138, 188, 212
Manual Injection Kit ........................................296 Washing with Standard........................... 130, 172
POC Kit............................................................296 Syringe Pump
TN ....................................................................295 8-Port Valve Rotor...........................................246
TOC-V .............................................................294 Zero Point Detection..........................................47
Standard Accessories System Administration
ASI ...................................................................299 Add Event Log...................................................22
Manual Injection Kit ........................................301 Change Password...............................................21
POC Kit............................................................301 Introduction........................................................20
TN ....................................................................301 Screen Lock .......................................................20
TOC-VCPH/CPN ................................................298 System Blank ...........................................................291
System Description ............................................... 7–18
System Information Window
Instrument Setup Wizard ...................................94
T
TC TN
Analysis, New Catalyst ......................................36 Analysis Preparation ..........................................59
Blank Check.............................................. 66, 269 Analysis, New Catalyst...................................... 36
Principles..........................................................284 Connecting Ozone Source Air .........................329
Standard Solution...............................................60 Front and Right Side Views...............................16
TC/TN, Standard Solution .........................................61 Gas Exhaust Tubing.........................................332
TC-IC .......................................................................288 Grounding ........................................................ 328
Technical Information..................................... 289–293 Inside Front View ..............................................17
Time, Sparge ............................................................139 Installation .......................................................328
Maintenance............................................ 256–260
Ozone Source Airflow Rate...............................91
Power Source .....................................................84
Power Supply...................................................328
Principles .........................................................288
Rear View .......................................................... 18
Specifications...................................................295
Standard Solution...............................................61
Turning On.........................................................91
Unit ..................................................................295
Index-9
Index
TN Unit Toolbar, Sample Table Editor
Gas Tubing Connection ................................... 330 Customizing..................................................... 177
TOC Displaying ....................................................... 178
Principles ......................................................... 284 Functions ......................................................... 122
Setting Parameters ............................................. 83 Total Carbon, See TC
Setting Parameters in Software.......................... 26 Total Nitrogen, See TN
TOC window of the Instrument Setup Wizard .......... 94 Total Organic Carbon, See TOC
TOC/TN, Catalyst...................................................... 39 Troubleshooting...............................................270–281
TOC-Control V Autosampler ............................................278–280
Administration Tool .......................................... 21 Flow Charts ..................................................... 274
Administrator Window ...................................... 21 TOC-V.....................................................274–277
TOC-V Tubing Diameter........................................................ 83
Construction......................................................... 8 Turntable, Installation ............................................. 323
Front View ........................................................... 8 Tutorial ..............................................................92–119
Grounding ........................................................ 312 Calibration Runs.............................................. 100
High Sensitivity Catalyst ................................... 37 Configuring the Instrument ............................... 93
Inside Front View .............................................. 10 Connecting the Instrument ..........................96, 98
Inside Top View ................................................ 11 Creating a Sample Table ................................... 98
Installation .............................................. 310–332 Evaluating Results........................................... 114
Left Side View..................................................... 9 Measurement ................................................... 112
Rear View ............................................................ 9 Printing Report ................................................ 117
Right Side View................................................... 8 Unknown Run.................................................. 108
Specifications................................................... 294
Standard Catalyst ............................................... 37
Troubleshooting...................................... 274–277
U
Units....................... 87, 130, 138, 163, 171, 188, 212 USP/EP ............................................................141, 192
V
Vial View Menu, Sample Table Editor ........................... 160
Filling................................................................. 72 Viewing Measurement Data in Real Time .............. 167
Placing in Rack .................................................. 74
Rack Installation ....................................... 74, 323
Sealing ............................................................... 73
Setting Position....................................... 106, 179
Types ................................................................. 72
W
Warning Water
Corrosion ............................................................. x Dilution............................................................ 236
Electric Shock...................................................viii Drain Vessel .................................................... 236
High Temperature..............................................vii Humidifier ....................................................... 237
Injury .................................................................. ix MSDS .............................................................. 345
Warning Labels..........................................................vii Pure.................................................................... 50
Warning Message, Maintenance................................ 88 Supply to Cooler Drain Container..................... 48
Warranty ..................................................................... iv Zero ................................................................... 60
Washing
Catalyst ............................................................ 239
Combustion Tube ............................................ 242
Flow Lines ................................................ 86, 264
Syringe with Sample.......................138, 188, 212
Syringe with Standard ............................ 130, 172
Z
Zero Point Detection......................................... 47, 261 Zero Water................................................................. 60
Zero Shift ........................................................ 129, 172
Index-10