This document provides guidance for a lab session on understanding electrical measurement instruments. Students will use instruments like oscilloscopes, signal generators, and digital multimeters (DMMs) to measure voltages and currents. They will compare readings from low-cost DMMs to higher-end models to understand differences in accuracy and capabilities. The session aims to teach students to properly select instruments for measurements and correctly interpret readings.
This document provides guidance for a lab session on understanding electrical measurement instruments. Students will use instruments like oscilloscopes, signal generators, and digital multimeters (DMMs) to measure voltages and currents. They will compare readings from low-cost DMMs to higher-end models to understand differences in accuracy and capabilities. The session aims to teach students to properly select instruments for measurements and correctly interpret readings.
DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERINGS
Practical sessions at laboratories
Pre-Lab: Understanding instruments
Luis Torres, [email protected] Luis Lopez, [email protected] I. Objective • Digital Clamp Meter (Fluke 302+ or similar) The purpose of this lab session is to make practitioners aware of the capabilities of lab instruments, their limitations and comparison to low-cost devices. Engineers should learn how to appropriately select instruments and understand their readings, avoiding deception. This session is focused more on measuring voltage than current, but the intention applies to both types of measurements.
II. Devices needed
Get from our labs:
• Low impedance (Low-Z / Utility Multi- • Signal generator (BK Precision 4040A or meter) DMM (Fluke 113, or similar) similar)
• Digital Multi Meter (DMM) (Fluke 115,
17B, 179, or similar, capable of measuring “True RMS” for voltage and current AC and DC)
• Oscilloscope (Tektronix TDS1000,
TDS2000 or similar) • Are the instruments giving the right Get yourself: measurement? Are there any differences? Why? • One cheap DMM (any brand around $30.000 pesos, AC/DC) • Use the oscilloscope to evaluate the RMS value of the waveform you have. Try to understand the relation among average (or DC) value, AC RMS value and the total (overall) RMS value of a waveform. How do you get these from the oscilloscope? And how from the DMMs? • What can a DMM measure and what cannot? What is the main difference if you set it to AC with respect to DC? Take into account that the voltage you have has both, a non-zero average level and an III. Using signal generator. alternating part. Get the signal generator produce a 500 Hz sinusoidal voltage with a DC level. Choose both Change now the signal generator to produce a to be similar and as high as possible without square-wave with a DC level. causing distortion of the waveform. • Repeat the measurements with all instruments. Observe the signal with the oscilloscope. Set the • Compare and discuss. DC vs. AC coupling and see the difference. Vary the DC level and amplitude a little and see the IV. Instrument impedances and test voltages. effect. Also, set different gains (x1, x10) at the probe without changing the internal setting in the It is said that voltmeters have large impedance. oscilloscope and see the effect. This is in general a desirable property. Find out • How can you tell that there is a probe gain (measure) the impedance of the DMMs when mismatch just by looking at the screen? measuring voltage AC/DC. All you have to do is set the device under measurement in Volts DC • What coupling (AC or DC) do you prefer, and use another DMM set as ohmmeter and and why? Give examples when you measure resistance. Then do it for Volts AC. specifically need AC coupling? And DC coupling? Doing these tests you will have two observations: the resistance of the instrument playing as voltmeter (read at the ohmmeter) and the test With the signal generator at 500Hz and with voltage of the instrument playing as ohmmeter sinusoidal output and DC level (as high as (read at the voltmeter, if it is in DC). Also, notice possible) do the following: how there are perhaps different resistances when • Measure the output voltage in AC and DC selecting different volt scales (if this is possible). with the DMM Fluke-115, then with the • Do it for the Fluke 115, Fluke 302 (the clamp meter Fluke 302, the Fluke 113, and ones you commonly use), the Fluke 113 finally with the cheap DMM and see the (called utility meter) and your cheap differences. (Note that Fluke 113 does not DMM. Try different combinations for who have AC/DC selection. What do you think plays the role of the ohmmeter and who about it? Discuss) that of the voltmeter. In both AC and DC. (Notice that Fluke 113 has no option AC/DC). What do you think about the • Even good DMM don't know what you results? Compare, discuss, understand. want to measure: you have to tell them that you want to get the average voltage • Pay attention to Fluke 113! How can this (DC setting) or the RMS of the alternating DMM affect your measurements at a component (AC setting). Therefore, when protoboard in one of your projects? What using DMMs you don't get the RMS value is the intended use of this DMM? (Hint: of the full compound waveform (AC + ghost voltages) DC). When you set the DMM into AC V. Discussion of experiences measurement, they get capacitively coupled! This means that DC components After doing all tests, each group should discuss cannot get through! As a result, you get about the instruments: DMMs, oscilloscope, only the RMS value of the alternating part. signal generator. What can they do and what If you wish to get the total RMS value you cannot. The group should acknowledge that good need the average value (measured with the instruments, such as the Fluke DMM, are DC setting) and the RMS of the expensive, and that we must take care of them. alternating part (measured with the AC Cheap instruments cannot replace good ones, setting): because, among other reasons, they do not TotalRMS = √ Average 2+ Alternating2RMS measure the same thing and they do not usually meet international safety standards for instruments. • The most professional way to clarify the handling of an instrument starts by reading its manual and understanding it! Be sure of having understood the following: • Please watch the video from EEV blog • DC coupling in oscilloscope is the made by Dave Jones: EEVblog #75 - preferred way to start measuring. It will Digital Multimeter Buying Guide for give you the realistic view. Beginners: • DMM commonly have high resistances https://youtu.be/gh1n_ELmpFI when used as voltmeters. • Low-Z DMM have resistances of some VI. Contribution to the lab report and to oral kΩ that have advantages, but could also defense introduce wrong measurements in your circuits. For example, when measuring the The experience gained in this session need not be voltage at the output of a voltage divider directly mentioned in the written report of the first made of two 1MΩ resistors in series from lab. Therefore, the reviewers will not expect a a 5V source, you would expect 2.5V at the “section” devoted to this part. The purpose of this output. However, the Fluke 113 will show lab session is to help students understand the a much smaller value! instruments and to successfully carry out the labs. If you think there was something interesting you • Only good DMMs can measure RMS learned in this session and want to mention it in a values of the waveforms. They have the future report, your are free to do it. label “True RMS”. This feature makes them expensive, but it is worth the money. However, in the oral defenses of every lab, Cheap DMM assume you have a sinusoid reviewers can check whether you attained the and give wrong values for non-sinusoidal intended understanding of this lab session. waveforms.