Differential Pressure

Transmitter

PAD

OO Span: 0.75 … 15 mbar up to 4.137 … 413.7 bar


OO Static pressure: max. 310 bar
OO tmax: +120 °C
OOProcess connection: ¼" NPT, ½" NPT,
various diaphragm seals on request
OO Material: stainless steel, Hastelloy-C®, tantalum, Monel
OO Various output: 4 ... 20 mA, frequency output
OO Sensor input: differential-, gauge-, absolute pressure
OO Digital communication with HART® protocol
OO ATEX-approval

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

Features
Superior performance
OO High reference accuracy:
±0.075 % of calibrated span
(optional: ±0.04 % of calibrated span)
OO Long-term stability (0.125 % URL for 3 years)
OO High rangeability (100:1) for range 4-0

Flexibility
OO Data configuration with HART® configurator
OO Zero point adjustment

Reliability
OO Continuous self-diagnostic function
OO Automatic ambient temperature compensation
Description
OO EEPROM write protection
The Kobold Differential Pressure Transmitter model PAD is a
OO Fail-mode process function
micro processor-based high performance transmitter, which
has flexible pressure calibration and output, automatic com- OO CE  EMC conformity standards (EN 50081-2, EN 50082-2)
pensation of ambient temperature and process variable, con- Transmitter Description
figuration of various parameters, communication with HART® Electronics module
protocol. The application is very various, as measuring pres- The Electronics module consists of a circuit board sealed
sure, flow and level by application method. All data of sensor in an enclosure. There are a MCU module. an analog mo-
is to be input, modified and stored in EEPROM. dule, a LCD module and a terminal module in a transmitter.
As an option the Kobold Pressure Transmitter is also available The MCU module acquires the digital value from the ana-
as a flow meter. This flowmeter model PAD-F has added the log module and applies correction coefficients selected from
totalizing function in the PAD transmitter. So it is available to EEPROM. The output section of the MCU module converts
check the flow rate and totalizing flow. It measures the flow the digital signal to a 4…20 mA output. The MCU module
rate by using differential pressure without compensation of communicates with the HART®-based configurator or control
temperature and static pressure. The shape of the PAD-F is systems such as DCS. The Power section of MCU module
the same as the standard device and it is only the terminal has a DC-to-DC power conversion circuit and an input/out-
block which is different since there are two more terminals for put isolation circuit. The LCD module plugs into the MCU mo-
the read-out of the pulse output. dule and displays the digital output in a user-configured unit.

Functional Block Diagram

Sensor Part MCU Part 4~20mA

A/D Conversion Microprocessor


· Input Sensor Value DAC
EEPROM · Engineering Units
· Re–range (Zero / Span)
· Sensor Trimming
· Zero Point Adjustment
· DA Trimming
Capacitive · Damping / Filtering HART®
High Sensor Low · Transfer function
Pressure Pressure · LCD Engineering Mode HART® Protocol
· Diagnostics with Host
· Self Compensation
User-Selectable Input · Communications
· LCD Display
• - D /- G /- H /- A
· Configuration Data
• Measuring Range

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

Sensor Inputs Self-diagnosis and others


The models PAD - D, - G, and - H are available in a differen- • CPU & Analog Module Fault Detection
tial pressure sensor of a capacitance type. The capacitance • Communication Error
pressure sensor measures differential and gauge pressure
• Fail-mode handling
and is commonly used in flow and level applications. Both
sides in the capacitance sensor transmit process pressure • LCD indication
from the process isolators to the sensor. The model PAD-A is • Temperature measurement of sensor module
also available in an absolute pressure sensor of a piezoresisti-
ve type and measures absolute pressure. The sensor module
converts the capacitance or the resistance to the digital va-
lue. The MCU module calculates the process pressure based
on the digital value.
The sensor modules include the following features
• 0.075 % accuracy
• The software of the transmitter compensates thermal ef-
fects, improving performance.
• Precise Input Compensation during operation is achieved
with temperature and pressure correction coefficients that
are characterized over the range of the transmitter and
stored in the sensor module EEPROM memory.
• EEPROM stores sensor information and correction coe-
ficients separately from MCU module, allowing an easy re-
pair, reconfiguration and replacement.
Basic Setups
Following settings can be easily configured from any host that
support the HART® protocol:
• Operational parameters Multi Planar Process Connection
• 4-20 mA points (zero/span) Conventionally, in the case where the pressure transmitter
should be vertically installed irrespective of the orientation
• Engineering units
of the process connection lines, modified fIanges (as shown
• Damping time: 0.25...60 sec above) are required in addition to the basic flanges. Multi-
• Tag: 8 alphanumeric characters planar pressure transmitter have been made in an effort to
• Descriptor: 16 characters solve the problems occurring in the related art, and an object
• Message: 32 characters of this multi planar is to provide a pressure transmitter, ca-
pable of being vertically installed without separate adaptor
• Date: day/month/year or various types of brackets regardless of the position of the
Calibration and Trimming process connection lines.
• Lower/Upper range (zero/span)
• Sensor zero trimming Process Connection via Diaphragm Seals
• Zero point adjustment For the connection of the differential pressure transmitter
• DAC output trimming model PAD to all different process connections, diverse dia-
phragm seal versions are necessary.They can be connected
• Transfer function
to the differential pressure transmitter by direct mounting or
• Self-compensation via a capillary tube. Depending on the application different
combinations of diaphragm seals, capillary tubes and fill fluids
are possible. To clarify those possibilities, the special connec-
tions via diaphragm seals are always to be requested separa-
tely to the differential pressure transmitter.

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

Technical Details
Measuring principle: Capacitance sensor (PAD-D, -F, -G, -H) Wetted materials
Piezo-resistive (PAD-A) Isolating diaphragms: 1.4404 (316L st. st), MoneI, Tantalum,
Measuring span: 0.75 ... 15 mbar up to 4.137 ... 413.70 Hastelloy-C®
bar (depending on instrument version) Drain/Vent valves: 1.4401 (316 st.st)
Zero and span values can be set Flanges and adapters: 1.4401 (316 st.st)
anywhere within the range limits O-ring: FPM, PTFE as an option
Span must be greater than or equal
to the minimum span Non-wetted materials
Accuracy: • for range 2 Fill fluid: silicone oil or inert fill
± 0.25 % of span for Bolts: stainless steel
0.1 URL ≤  span ≤ URL Electronics housing: aluminum, or 316L st.st. (option)
± [0.24 + (0.008 x (URL/span)) % of flameproof (Ex d) and
span for 0.05 URL ≤ span ≤0.1 URL waterproof (lP67)
• for range 3 Cover O-ring: NBR
± 0.075 % of span for
Paint: epoxy-polyester or polyurethane
0.1 URL ≤ span ≤ URL
± [0.25 + (0.005 x (URL/span))] % of Mounting bracket: for 2-inch pipe, 1.4301 (304 sst),
span for 0.02 URL ≤ span ≤ 0.1 URL with 1.4301(304 sst) U-bolt
• for range 4 to 0 Nameplate: 1.4301 (304 sst)
± 0.075 % of span for Process connections: ¼" NPT with 54.0 mm centre
0.1 URL ≤ span ≤ URL distance for standard flanges
± [0.025+(0.005x(URL/span))] % of  ½" NPT with process adapter (option)
span for 0.01 URL ≤ span ≤ 0.1 URL Mounting position: upright (process connection more
Stability: ± 0.125 % URL for 36 month flexible by using multi-planar flange)
Static pressure effect: ± 0.1 % URL per 70 bar (zero error) Display: 5 Digit LCD
± 0.2 % of reading per 70 bar (span
Power supply: 12 ... 45 VDC -operation
error)
17.5 ... 45 VDC -HART® communication
Turndown ratio: ranges 4 ~ 0 = 100 : 1
range 3 = 50 : 1 Maximum load: 250 Ω at 17.5 VDC
range 2 = 20 : 1 550 Ω at 24 VDC
(U - 12 VDC)
Process temperature: -40 °C ... +120 °C max. loop resistance = 0.022 A
(Approval codes may effect limits. Loop load: 0...1500 Ω - operation
Max. ambiant temperature at 250...550 Ω - HART® communication
LCD = +80 C.) Failure mode: fail high: current ≥ 21.1 mA
Ambient temperature: -30 °C…+80 °C fail low: current ≤ 3.78 mA
Ambient temperature Electrical
effect: ± (0.019 % URL + 0.125 % span) / 28 °C connection: ½" NPT conduit with M4 screw
Storage temperature: -40 °C ... +85 °C (without condensing) terminals (G½ option)
Humidity limit: 5 % ... 100 % rH Output: • two wire 4 ... 20 mA, userconfigurable
Power supply effect: ± 0.005 % of span per Volt for linear or square root output, digital
Mounting position process value superimposed on
effect: zero shift up to 350 Pa 4 ... 20 mA signal, available to any host
no span effect that conforms to the HART® protocol
Pressure limits (with silicone oil) • frequency output for flowmeter
(valid for stand-alone intruments only without assembled dia- model PAD-F with pulse width of 10,
phragm seals) 50 or 100 ms (selectable, negative
Model D and G 0 ... 137.9 bar (for range 2 ... 8) going pulse)
Model G 0 ... 400 bar (for range 9) output type: open collector, 30 V,
500 mAapprovals..., max.
0 ... 750 bar (for range 0)
pulse rate: 49 pulses/sec max.
Model H 0 ... 310 bar (for range 4 ... 7)
Model A 0 ... 5 bar (for range 4) Turn-On time: 3 seconds
0 ... 30 bar (for range 5) Protection: IP 67 for Standard (code S)
0 ... 52 bar (for range 6) Weight: 3.9 kg (excluding options) standard
• Burst pressure 0 ... 750 bar (for range 0)
Model D, G and H 689 bar 5.35 kg (st. st. housing - excl. options)
800 bar (for model G, range 0) ATEX approval: II 2G Ex d IIC T6 ...T5 (option)
Model A 10 bar (for range 4)  II 1G or 2G Ex ia IIC T5 or T4 Ga or
40 bar (for range 5) Gb (option)
70 bar (for range 6)

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

Order Details (Example: PAD-D EE 2 S 2 N S0 0)

Material Calibrated span


Model Version Body/vent plug/ (Measuring range limits for PAD-D,
diaphragm -F, -G and -H in separate table)

Calibrated span for PAD-D, -F, -G, -H


23) = 0.75 ... 15 mbar
3 = 1.5 ... 75 mbar
D = differential pressure transmitter 4 = 3.73 ... 373 mbar
(static pressure 138 bar) EE = 316 st. steel / 316 st. steel /
316L st. steel 5 = 18.65 mbar ... 1.865 bar
F1) = differential pressure transmitter with
EH = 316 st. steel / 316 st. steel / 6 = 69 mbar ... 6.9 bar
pulse output and totalizer especially for
flow measurement Hastelloy-C® 7 = 206.8 mbar ... 20.68 bar
PAD-
H = differential pressure transmitter for high EM = 316 st. steel / 316 st. steel / 8 = 689.5 mbar ... 68.95 bar
3)

line pressure (static pressure 310 bar) Monel


93) 4) = 2.068 ... 206.80 bar
G = gauge pressure transmitter ET = 316 st. steel / 316 st. steel /
tantalum 03)4) = 4.137 ... 413.70 bar
A = absolute pressure transmitter Calibrated span for PAD-A
4 = 25 mbar ... 2.5 bar
5 = 150 mbar ... 15 bar
6 = 250 mbar ... 25 bar

Order Details (continued):

Approvals
Electrical
Filling liquid Process connection for hazardous Options
connection
applications

0 = without
C = engineering unit (must be chosen
when using the differential trans-
N = ½" NPT mitter as a flowmeter)
epoxy-
S0 = standard D = teflon o-ring (wetted part)
polyester
(waterproof IP67)
painted E = oil free finish
2 = ¼" NPT female aluminium F0 = ATEX, flameproof,
S = silicone (standard) Ex d F = side vent / drain bottom
G = G ½ epoxy-
X2) = special polyester E0 = ATEX,
G = side vent / drain top
painted instrinsically safe, H = multi-planar process connection
aluminium Ex i
M = housing in stainless steel
X2) = special
N5) = mounting of PAD onto diaphragm
seal
Y2) = special calibration range

1) 
Specify flow rate engineering unit, ∆ p and flow rate at URV (Upper Range Value), ∆ p and flow rate (generally '0') at LRV (Lower Range Value) pulse scale
(choose only one value from 0.001, 0.01, 0.1, 1, 10, 100, 1 000, 10 000 m³/pulse) and pulse width (choose only one value from 10 ms, 50 ms, 100 ms), while
ordering so that max. duty cycle is 49 pulses/sec
2)
Code X not available for ATEX. Order code X and Y must be specified in writing
3)
Not for PAD-H
4)
Not for PAD-D and PAD-F
5) 
Diaphragm seal model and application data to be speicified in clear text. Application Index on the last two pages of this data sheet to be filled out. For
summary of diaphragm seal models and possible ranges, see following page 11 onwards. For dimensional details see DRM data sheet.

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

Order Details Oval flange adapter

Dimensions [mm]

Technical Details
Material: 1.4401 (316 Stainless steel)
Seal: PTFE
Bolts: 2 x mounting screws UNF7/16-20
Process connection: ½" NPT female
Order code: ZUB-PAD-OVF

Order Details Mounting brackets


Description Order code
Angle type bracket for PAD/PAS
vertical pipe mounting for PAS
vertical pipe mounting for PAD
ZUB-PAD/PAS-K
incl. U-Clamp for 2“ pipe mounting bracket and 2 x mounting nuts/ washers
incl. 4 x mounting screws for PAS
incl. 4 x mounting screws for PAD
Flat type bracket for PAD/PAS
horizontal pipe mounting for PAS
vertical pipe mounting for PAD
ZUB-PAD/PAS-L
incl. U-Clamp for 2“ pipe mounting bracket and mounting nuts/ washers
incl. 4 x mounting bolts and washers for PAS
incl. 4 x mounting bolts for PAD

Order Details Manifold valves


Description Order code
3-way manifold valve, remote mount, machined ZUB-PAD-3WMR

5-way manifold valve, remote mount, machined ZUB-PAD-5WMR

2-way manifold valve, direct mount, machined ZUB-PAD-2WMD

2-way compact manifold valve, direct mount, machined V-2003CDADABAA

3-way compact manifold valve, direct mount, machined V-3151CHHHIBAA

5-way compact manifold valve, direct mount, machined V-5050CDAHIBAA

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

Measuring Range Limits for PAD-D, -F, -G and -H

Lower range limit (LRL)


Range code Calibrated span Upper range limit (URL)
PAD-D, -F PAD-G PAD-H
2 0.75...15 mbar - 15 mbar - 15 mbar - 15 mbar
3 1.5...75 mbar -75 mbar -75 mbar - 75 mbar
4 3.73...373 mbar -373 mbar -373 mbar -373 mbar 373 mbar
5 18.65 mbar...1.865 bar -1.865 bar -1 bar -1.865 bar 1.865 bar
6 69 mbar...6.9 bar -6.9 bar -1 bar -6.9 bar 6.9 bar
7 206.8 mbar...20.68 bar -20.68 bar -1 bar -20.68 bar 20.68 bar
8 689.5 mbar ...68.95 bar -68.95 bar -1 bar - 68.95 bar
9 2.068...206.80 bar - -1 bar - 206.80 bar
0 4.137...413.70 bar - -1 bar - 413.70 bar
* Special Measuring span with adequate lower and upper range limits on request

Unit Conversion
Range code bar kg/cm² KPa psi in H2O at 4 °C mm H2O at 4 °C in Hg at 0 °C
2 0.015 0.015 1.5 0.217 6 152 0.422
3 0.075 0.076 7.5 1.087 30 765 2.215
4 0.373 0.38 37.3 5.410 149 3804 11.014
5 1.865 1.902 186.5 27.049 749 19018 55.072
6 6.900 7.036 690 100.073 2773 70361 203.750
7 20.681 21.088 2068 299.930 8310 210878 610.660
8 68.950 70.309 6895 1000.009 27708 703097 2036.025
9 206.800 210.876 20680 2999.303 83105 2108781 6106.597
0 413.700 421.856 41370 6000.211 166085 4218566 12216.550

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

Dimensions [mm]
PAD standard1) PAD multi planar process connection1)
112*/119** 112*/119** 112*/119** 112*/119**
86 86 73*/80** 39 73*/80** 39 86 86 73*/80** 39 73*/80** 39

s s s s
U T I ON - U T I ON - U T I ON -
- CA - CA U T I ON - - CA
D
IN
D
IN - CA IN D
IN
D

E
E
E

EX
EX
EX

GI Z

GI Z
GI Z

EX
GI Z
PLO

PLO
PLO

PLO
E NE R

E NE R
E NE R

TERMINALS

TERMINALS
TERMINALS

E NE R

TERMINALS
S I VE A TM

S I VE A TM
S I VE A TM

FIELD

FIELD
FIELD

S I VE A TM

FIELD
ILE

ILE
ILE

ILE

113

113
113
WH

WH
WH

113
WH
OS

O
OS

SP
N

OS
P
P

E HE

N
E HE E HE
OP OP

P
OP RE RE E HE RE
D O N OT D O N OT OP RE D O N OT
LOCK LOCK LOCK LOCK LOCK LOCK
D O N OT
LOCK LOCK

192.5
192.5

202.5
202.5
7/16-20 UNF

79.5
XXXX CF8M
79.5

XXXX CF8M

89.5
89.5
42 42 42 42 316
316

54 54 54 54

* Ex d and standard * Ex d and standard


** For Ex i ** For Ex i

PAD with multi planar flange and angle type bracket1) PAD standard with flat type bracket (vertical mounted)1)

s s s
U T I ON - U T I ON -
- CA
U T I ON - - CA IN - CA IN
IN D D
D
E

E
EX

EX
GI Z
E

GI Z
EX
GI Z

PLO

PLO
PLO
E NE R

E NE R
TERMINALS

TERMINALS
E NE R

TERMINALS
S I VE A TM

S I VE A TM
FIELD

FIELD
S I VE A TM

FIELD
ILE

ILE
ILE

WH
WH
WH

OS
OS
OS

N
N

P
N

E HE E HE
P

E
OP
HE OP RE OP RE
RE D O N OT D O N OT
D O N OT
LOCK LOCKLOCK LOCK LOCK LOCK

XXXX XXXX
CF8M CF8M
323.5
324.5
324.5

316 316

210
122
122

47

94
110 110

PAD standard with flat type bracket (horizontal mounted)1)


210
120

s
U T I ON -
- CA IN
D
E

EX
GI Z

PLO
E NE R

TERMINALS
S I VE A TM

FIELD
ILE
WH

OS
N

E HE
OP RE
D O N OT
LOCK LOCK
205.5
47

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

PAD-G/A mounted with 2-way manifold valve1)


*/119**
*/119**

* Ex d and standard
** For Ex i

Ø 7.0 DIA
Mounting holes
2 places
side Block valve

Process side Instrument side


1)
For PAD-G/A, the low pressure port 'L' is always closed
228.0 (open, approx.)

Manifold valves (remote mount)


31.0
Block valve
24.0

54.0

86.0

Technical Details
Equalize valve
Material: 316SS body with PTFE packing Ø 7.0 DIA
Mounting holes
Connection and size: ½" NPT (F) Process side Block 2valve
placesSied
½" NPT
Pressure rating:
4 places 6 000 psig at 38 °C (≈410 bar) side Block valve
Temperature range: -54 °C ... ¼"+232 °C
NPT Flow schematic
Static vent
2 places
Process side Instrument side
3-way manifold valve Technical Specifications:
228.0 (open, approx.)

Material: 316SS
Block valve 44.0 Connection &Ø Size:
7.0 DIA½" NPT (F) Weight: 2 kg
Pressure rating: 6,000
Mounting psig at 38°C (≈410 bar)
holes
Block valve
Temperature2side
places
range: -54°C ... +232°C 31.0 Block valve
24.0

54.0

86.0

Weight: 2 kg
104.5 (open, approx.)

Instrument side
Equalize valve 86.0 Process side Instrument side Equalize valve
228.0 (open, approx.)

43.0
Instrument
Block valve ½" NPT Block valve Process side Block valve side
Sied
31.0
24.0

54.0

86.0

4 places
Instrument Instrument
33.3

Equalize Equalize ¼"Equalize


NPT valve Flow schematic
Static Static
(test) (test) Static vent
78.0 Block/ Vent Block/ 2 places
Process side Block valve Sied
Block valve ½" Isolate
NPT Isolate
4 places Technical Specifications:
24.0
Process Vent port Process Material: 316SS
Flow schematic
¼" NPT
228.0 (open, approx.) Instrument Block valve
sideFlow schematic
44.0
Static vent Connection & Size: ½" NPT (F)
Process side 86.0 2 places Pressure rating: 6,000 psig at 38°C (≈410 bar)
Technical Specifications:
Temperature range: -54°C ... +232°C
43.0 Specifications:
Technical Material: 316SS
5-way manifold valve BlockMaterial:
valve 316SS 44.0 Connection & Size: ½"Weight:
NPT (F) 2 kg
104.5 (open, approx.)

Connection & Size: ½" NPT (F) Pressure rating: 6,000 psig at 38°C (≈410 bar)
Pressure rating: Instrument
6,000 psig side Block
at 38°C (≈410valve
bar) Temperature range: -54°C ... +232°C
Equalize
Temperature range:valve
-54°C
86.0 ... +232°C Weight: 2.2 kg
Weight: 2 kg
Equalize valve Bleed valve Equalize valve Weight: 2.2 kg Instrument Instrument
104.5 (open, approx.)

43.0
Equalize valve
Block valve Equalize Equalize Static
85.0 (open, approx.)

Static
Instrument Instrument
(test) (test)
33.3

½" NPT
4 Places Block/ Block/
33.3

Equalize Equalize Static


Block valve Static Vent
Isolate
(test) Isolate
(test)
33.3

Block/ Block/
Block valve 24.0 Isolate
Vent
Isolate
78.0 24.0 78.0 Process Vent port Process
¼" NPT 54.0 ¼" NPT Process Vent port Process
Static port Vent port 228.0 (open, approx.) Flow schematic
2 places 228.0 (open, approx.) Flow schematic
Process side Process side
Technical Specifications:
Technical Specifications:
Material:
Material: 316SS
316SS
Connection & Size: ½" NPT (F)
Connection & Size:
Pressure rating: 6,000½" NPT
psig (F) (≈410 bar)
at 38°C
Pressure rating:
Temperature 6,000
range: psig
-54°C at 38°C (≈410 bar)
... +232°C
Equalize valve Bleed valve Equalize valve Weight: 2.2 range:
Temperature kg -54°C ... +232°C
Equalize valve Bleed valve Equalize valve Weight: 2.2 kg
x.)

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Instrument Process

1/4" NPT vent port 50.8


28.4 47.8
Differential Pressure Transmitter Model PAD

7.8
41.4
92.60 (open)

16

Vent
Manifold valves (Direct mount)
Bleed valve Bleed
Technical Details Block
Material: 316SS body with PTFE50.8
packing Pressure
Instrument rating: 3 000 psig at 232 °C (≈210 bar)
Process
Connection and size: ½" NPT (F) to flange Temperature range: -54 °C ... +232 °C

2-way Manifold valve Weight: 1.6 kg


45 1/4" NPT vent port
Block valve 50.8
28.4 47.8

Vent

7.8
63.5 (open)

1/2" NPT
2 places
Bleed

41.4
Block
92.60 (open)

16 Instrument Process
Process side
Instrument side
41.3
31

Ø 12.2 for Bleed valve


7/16" bolts 96.8 1/4" NPT vent port 50.8
50.8 28.4 47.8

7.8
45

41.4
Block valve

92.60 (open)

16
63.5 (open)

1/2" NPT
2 places

Process side Bleed valve


Instrument side
41.3
31

50.8
Ø 12.2 for
7/16" bolts 96.8

45
Block valve
63.5 (open)

1/2" NPT
2 places

Process side
Instrument side
41.3
31

Ø 12.2 for
7/16" bolts 96.8

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

Manifold valves (Direct mount, machined)


Technical Details
Material: AISI 316L
Pressure rating: 6 000 psi
Temperature range: -73 °C ... +210 °C (PTFE packing), standard
-54 °C ... +510 °C (GRAPHOIL packing), on request
Weight: 0.88 kg
2-way Manifold valve
(inlet: ½" NPT female / outlet: ½" NPT male)

Drain

VENT.
IN
Ø 30 Drain
148

VENT. ISOLATE
IN
Ø 30
148

ISOLATE

120

OUT IN

120

Drain 1/4" NPT-F with plug


OUT IN

Included accessories
Plug

Drain 1/4" NPT-F with plug

Order Code: V-2003CDADABAA (PTFE packing)

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

Manifold valves (Direct mount, machined) D


P
DRAIN DRAIN

Technical Details
Material: AISI 316L
Pressure rating: 6 000 psi IN IN D

Temperature range: -73 °C ... +210 °C (PTFE packing), standard


P
DRAIN DRAIN

-54 °C ... +510 °C (GRAPHOIL packing), on request


Weight:
224 2.17 kg
Flanged Type
IEC 61518/A
IN
3-way Manifold valve
IN

(inlet: flanged / outlet: flanged according to IEC 61518 Type B / Type A)


224
EQUALISE
EQU
AL IS
E Flanged Type
N°4 fori Ø12 IEC 61518/A
D
P
DRAIN DRAIN
OUT IN
41.4
ISOLATE
ISOLATE

EQUALISE N°4 holes EQU


AL IS
E
N°4 fori Ø12
Ø12
70
124

OUT IN
41.4
ISOLATE

IN
ISOLATE

IN
1/4" NPT-F 30
32 70
54
N°2 fixing
84 holes M10
DRAIN 1/4" NPT-F 30 224
106
54 32 Flanged Type
N°2 fixing
84 holes M10
IEC 61518/A
106

Included accessories: EQU


AL IS
E
4 carbon steel screws (stainless steel on request)
EQUALISE
N°4 fori Ø12
2 plugs
7/16" UNF
124

2 PTFE gaskets
1/4" NPT

2¾" OUT IN
7/16" UNF
41.4
ISOLATE
ISOLATE

1/4" NPT

70
2¾"
Order Code: V-3151CHHHIBAA (PTFE packing)

DRAIN
Possible ways 1/4" NPT-F3-way manifold30
of mounting valve with PAD
Description 54 Process connection
32 Illustration
N°2 fixing
84 holes M10

106
Valve mounted upside down on front side of PAD Flanged according to IEC 61518 type A

Valve mounted upside down on front side of PAD


½" NPT female
including oval flange adapter model ZUB-PAD-OVF
7/16" UNF
1/4" NPT

Valve mounted on front side of PAD, head of PAD


Flanged according to IEC 61518 type A
rotated by 90° clockwise
2¾"

Valve mounted on front side of PAD including oval


flange adapter model ZUB-PAD-OVF, head of PAD ½" NPT female
rotated by 90° clockwise

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

10.5

16
D
P

10.5

16
D
Manifold valves (Direct mount, machined) 30
P
DRAIN IN IN DRAIN

Technical Details Flanged Type


102 IEC 61518/A
Material: AISI 316L
Pressure rating: 6 000 psi 10.5

16
D
30
Temperature range: -73 °C ... +210 °C (PTFEFlanged
packing), standard DRAIN IN IN DRAINP
Type
-54 °C ... +510 °C
102 (GRAPHOIL packing), on request
IEC 61518/A
Weight: 2.80 kg
260
5-way Manifold valve 30 DRAIN IN IN DRAIN
(inlet: ½" NPT / outlet: flanged according to IEC 61518Flanged Type
Type A) VENT.

102
10.5

IEC 61518/A
16

260 D
P
VENT
.
IN
OUT
VENT. EQUALISE VENT.
124

70
260 IN
OUT
VENT. EQUALISE VENT.

41.4
124

VENT

ISOLATE
.
ISOLATE

70
30 DRAIN IN IN DRAIN
Flanged Type
41.4

102 IEC 61518/A


ISOLATE

IN
ISOLATE

OUT
VENT. EQUALISE VENT.

32
124

10.5
70

16
D
32 P

54 DRAIN 1/4" NPT-F


41.4

ISOLATE
ISOLATE

N°4 Ø12 holes DRAIN 1/4" NPT-F


54
N°4 Ø12 holes 142
260142 32
N°2 fixing holes M10 30 DRAIN IN IN DRAIN
N°2 fixing
54 holes M10 DRAIN 1/4" NPT-F VENT. Flanged Type
N°4 Ø12 holes 102 IEC 61518/A
Included accessories: 142
7/16" UNF
7/16" UNF

N°2 fixing holes M10


4 carbon steel screws (stainless steel on request) IN
OUT
VENT. EQUALISE VENT.

2 plugs
1/4" NPT
1/4" NPT
124

260
70

2 PTFE gaskets
7/16" UNF

1¾"
1¾" VENT
1/4" NPT

.
41.4

ISOLATE
ISOLATE

1¾" IN
OUT
VENT. EQUALISE VENT.

32
124

70
41.4

DRAIN 1/4" NPT-F


ISOLATE
ISOLATE

54
N°4 Ø12 holes
Order Code: V-5050CDAHIBAA (PTFE packing)
142 32
Possible ways of mounting 5-way manifold
N°2 fixing valve
holes M10with PAD DRAIN 1/4" NPT-F
54
N°4 Ø12 holes
Description Process connection 142 Illustration
N°2 fixing holes M10
7/16" UNF
1/4" NPT

7/16" UNF
1/4" NPT

1¾"
Valve mounted upside down on front side of PAD ½" NPT female
1¾"

Valve mounted on front side of PAD after rotating


½" NPT female
the head of PAD by 90° clockwise

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

Example of PAD direct assembed with (extended) diaphragm seal (not available as ATEX-Version)
(for dimensional details, see DRM data sheet)

welded

Δ L (min. 150 mm max. 500 mm)

Fig. 1

Dimensions [mm]: Examples for DN50 / DN 80/DN100 / 2" ANSI / 3" ANSI/4"ANSI


Flange type D k d2 b f d4 X d5 RL
DN50 PN16 165 125 18 18 2 4 48
102
DN50 PN40 165 125 18 20 2 4 48
welded
2" ANSI Cl. 150 152.4 120.6 19 19.1 2 4 48
92
2" ANSI Cl. 300 165.1 127 19 22.3 2 8 48 50 mm (2")/
DN80 PN16 200 160 18 20 2 8 76 100 mm (4")/
138
DN80 PN40 200 160 18 24 2 8 76 150 mm (6")/
3" ANSI Cl. 150 190.5 152.4 19 23.9 1.6 4 76 200 mm (8")/
127 (customer
3" ANSI Cl. 300 209.5 168.3 22 28.4 1.6 8 76
Δ L (min. 150 mm max. 500 mm) DN100 PN16 220 180 18 20 2 149 8 89 specified)
DN100 PN40 235 190 22 24 2 149 8 89
4" ANSI Cl. 150 228.6 190.5 19 24 1.6 157.2 8 89
4" ANSI Cl. 300 254 200 22 32 1.6 157.2 8 89

Example of PAD assembed with remote diaphragm seals and capillaries


(for dimensional details, see DRM data sheet)
Dimensions [mm]: Examples for DN50 / DN 80 /DN100 / 2" ANSI / 3" ANSI/4"ANSI
Flange type D k d2 b f d4 X
DN50 PN16 165 125 18 18 2 4
102
DN50 PN40 165 125 18 20 2 4
2" ANSI Cl. 150 152.4 120.6 19 19.1 2 4
92
2" ANSI Cl. 300 165.1 127 19 22.3 2 8
DN80 PN16 200 160 18 20 2 8
138
DN80 PN40 200 160 18 24 2 8
3" ANSI Cl. 150 190.5 152.4 19 23.9 1.6 4
127
3" ANSI Cl. 300 209.5 168.3 22 28.4 1.6 8
DN100 PN16 220 180 18 20 2 149 8
DN100 PN40 235 190 22 24 2 149 8
4" ANSI Cl. 150 228.6 190.5 19 24 1.6 157.2 8
4" ANSI Cl. 300 254 200 22 32 1.6 157.2 8

Fig. 2

[email protected]
976 587 019
https://basesistemas.com/kobold-elementos-de-instrumentacion-y-control-industrial/
Differential Pressure Transmitter Model PAD

Example of PAD-G remote assembed with (extended) diaphragm seal and capillary
(for dimensional details, see DRM data sheet)

Fig. 3

Dimensions [mm]: Examples for DN50 / DN 80 /DN100 / 2" ANSI / 3" ANSI/4"ANSI


Flange type D k d2 b f d4 X d5 RL
DN50 PN16 165 125 18 18 2 4 48
102
DN50 PN40 165 125 18 20 2 4 48
2" ANSI Cl. 150 152.4 120.6 19 19.1 2 4 48
92
2" ANSI Cl. 300 165.1 127 19 22.3 2 8 48
DN80 PN16 200 160 18 20 2 8 76 50 mm (2")/ 100 mm (4")/
138
DN80 PN40 200 160 18 24 2 8 76 150 mm (6")/
3" ANSI Cl. 150 190.5 152.4 19 23.9 1.6 4 76 200 mm (8")/
127 (customer specified)
3" ANSI Cl. 300 209.5 168.3 22 28.4 1.6 8 76
DN100 PN16 220 180 18 20 2 149 8 89
DN100 PN40 235 190 22 24 2 149 8 89
4" ANSI Cl. 150 228.6 190.5 19 24 1.6 157.2 8 89
4" ANSI Cl. 300 254 200 22 32 1.6 157.2 8 89