M 5-D P M E F S S: 1.0 Scope and Application
M 5-D P M E F S S: 1.0 Scope and Application
While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the
official version. The most recent edits to this method were published here:
https://www.gpo.gov/fdsys/pkg/FR-2016-08-30/pdf/2016-19642.pdf. To see a complete version
including any recent edits, visit: https://www.ecfr.gov/cgi-bin/ECFR?page=browse and search under
Title 40, Protection of Environment.
1.2 Applicability. This method is applicable for the determination of PM emissions from
stationary sources.
1.3 Data Quality Objectives. Adherence to the requirements of this method will enhance the
quality of the data obtained from air pollutant sampling methods.
Particulate matter is withdrawn isokinetically from the source and collected on a glass fiber filter
maintained at a temperature of 120 ±14 °C (248 ±25 °F) or such other temperature as specified
by an applicable subpart of the standards or approved by the Administrator for a particular
application. The PM mass, which includes any material that condenses at or above the filtration
temperature, is determined gravimetrically after the removal of uncombined water.
5.1 Disclaimer. This method may involve hazardous materials, operations, and equipment. This
test method may not address all of the safety problems associated with its use. It is the
responsibility of the user of this test method to establish appropriate safety and health practices
and to determine the applicability of regulatory limitations prior to performing this test method.
6.1 Sample Collection. The following items are required for sample collection:
1
Method 5 8/2/2017
6.1.1 Sampling Train. A schematic of the sampling train used in this method is shown in Figure
5-1 in section 18.0. Complete construction details are given in APTD-0581 (Reference 2 in
section 17.0); commercial models of this train are also available. For changes from APTD-0581
and for allowable modifications of the train shown in Figure 5-1, see the following subsections.
NOTE: The operating and maintenance procedures for the sampling train are described in APTD-
0576 (Reference 3 in section 17.0). Since correct usage is important in obtaining valid results, all
users should read APTD-0576 and adopt the operating and maintenance procedures outlined in
it, unless otherwise specified herein.
6.1.1.1 Probe Nozzle. Stainless steel (316) or glass with a sharp, tapered leading edge. The angle
of taper shall be ≤30°, and the taper shall be on the outside to preserve a constant internal
diameter. The probe nozzle shall be of the button-hook or elbow design, unless otherwise
specified by the Administrator. If made of stainless steel, the nozzle shall be constructed from
seamless tubing. Other materials of construction may be used, subject to the approval of the
Administrator. A range of nozzle sizes suitable for isokinetic sampling should be available.
Typical nozzle sizes range from 0.32 to 1.27 cm (1⁄8 to 1⁄2in) inside diameter (ID) in increments
of 0.16 cm (1⁄16 in). Larger nozzles sizes are also available if higher volume sampling trains are
used. Each nozzle shall be calibrated, according to the procedures outlined in section 10.1.
6.1.1.2 Probe Liner. Borosilicate or quartz glass tubing with a heating system capable of
maintaining a probe gas temperature during sampling of 120 ±14 °C (248 ±25 °F), or such other
temperature as specified by an applicable subpart of the standards or as approved by the
Administrator for a particular application. Since the actual temperature at the outlet of the probe
is not usually monitored during sampling, probes constructed according to APTD-0581 and
utilizing the calibration curves of APTD-0576 (or calibrated according to the procedure outlined
in APTD-0576) will be considered acceptable. Either borosilicate or quartz glass probe liners
may be used for stack temperatures up to about 480 °C (900 °F); quartz glass liners shall be used
for temperatures between 480 and 900 °C (900 and 1,650 °F). Both types of liners may be used
at higher temperatures than specified for short periods of time, subject to the approval of the
Administrator. The softening temperature for borosilicate glass is 820 °C (1500 °F), and for
quartz glass it is 1500 °C (2700 °F). Whenever practical, every effort should be made to use
borosilicate or quartz glass probe liners. Alternatively, metal liners (e.g., 316 stainless steel,
Incoloy 825 or other corrosion resistant metals) made of seamless tubing may be used, subject to
the approval of the Administrator.
6.1.1.3 Pitot Tube. Type S, as described in section 6.1 of Method 2, or other device approved by
the Administrator. The pitot tube shall be attached to the probe (as shown in Figure 5-1) to allow
constant monitoring of the stack gas velocity. The impact (high pressure) opening plane of the
pitot tube shall be even with or above the nozzle entry plane (see Method 2, Figure 2-7) during
sampling. The Type S pitot tube assembly shall have a known coefficient, determined as outlined
in section 10.0 of Method 2.
6.1.1.4 Differential Pressure Gauge. Inclined manometer or equivalent device (two), as described
in section 6.2 of Method 2. One manometer shall be used for velocity head (Δp) readings, and
the other, for orifice differential pressure readings.
2
Method 5 8/2/2017
6.1.1.5 Filter Holder. Borosilicate glass, with a glass or Teflon frit filter support and a silicone
rubber gasket. Other materials of construction (e.g., stainless steel or Viton) may be used, subject
to the approval of the Administrator. The holder design shall provide a positive seal against
leakage from the outside or around the filter. The holder shall be attached immediately at the
outlet of the probe (or cyclone, if used).
6.1.1.6 Filter Heating System. Any heating system capable of monitoring and maintaining
temperature around the filter shall be used to ensure the sample gas temperature exiting the filter
of 120 ± 14 °C (248 ± 25 °F) during sampling or such other temperature as specified by an
applicable subpart of the standards or approved by the Administrator for a particular application.
The monitoring and regulation of the temperature around the filter may be done with the filter
temperature sensor or another temperature sensor.
6.1.1.8 Condenser. The following system shall be used to determine the stack gas moisture
content: Four impingers connected in series with leak-free ground glass fittings or any similar
leak-free noncontaminating fittings. The first, third, and fourth impingers shall be of the
Greenburg-Smith design, modified by replacing the tip with a 1.3 cm (1⁄2 in.) ID glass tube
extending to about 1.3 cm (1⁄2 in.) from the bottom of the flask. The second impinger shall be of
the Greenburg-Smith design with the standard tip. Modifications (e.g., using flexible connections
between the impingers, using materials other than glass, or using flexible vacuum lines to
connect the filter holder to the condenser) may be used, subject to the approval of the
Administrator. The first and second impingers shall contain known quantities of water (Section
8.3.1), the third shall be empty, and the fourth shall contain a known weight of silica gel, or
equivalent desiccant. A temperature sensor, capable of measuring temperature to within 1 °C (2
°F) shall be placed at the outlet of the fourth impinger for monitoring purposes. Alternatively,
any system that cools the sample gas stream and allows measurement of the water condensed and
moisture leaving the condenser, each to within 1 ml or 1 g may be used, subject to the approval
of the Administrator. An acceptable technique involves the measurement of condensed water
either gravimetrically or volumetrically and the determination of the moisture leaving the
condenser by: (1) monitoring the temperature and pressure at the exit of the condenser and using
Dalton's law of partial pressures; or (2) passing the sample gas stream through a tared silica gel
(or equivalent desiccant) trap with exit gases kept below 20 °C (68 °F) and determining the
weight gain. If means other than silica gel are used to determine the amount of moisture leaving
the condenser, it is recommended that silica gel (or equivalent) still be used between the
condenser system and pump to prevent moisture condensation in the pump and metering devices
and to avoid the need to make corrections for moisture in the metered volume.
3
Method 5 8/2/2017
6.1.1.9 Metering System. Vacuum gauge, leak-free pump, calibrated temperature sensors, dry
gas meter (DGM) capable of measuring volume to within 2 percent, and related equipment, as
shown in Figure 5-1. Other metering systems capable of maintaining sampling rates within 10
percent of isokinetic and of determining sample volumes to within 2 percent may be used,
subject to the approval of the Administrator. When the metering system is used in conjunction
with a pitot tube, the system shall allow periodic checks of isokinetic rates.
6.1.1.10 Sampling trains utilizing metering systems designed for higher flow rates than that
described in APTD-0581 or APTD-0576 may be used provided that the specifications of this
method are met.
NOTE: The barometric pressure reading may be obtained from a nearby National Weather
Service station. In this case, the station value (which is the absolute barometric pressure) shall be
requested and an adjustment for elevation differences between the weather station and sampling
point shall be made at a rate of minus 2.5 mm Hg (0.1 in.) per 30 m (100 ft) elevation increase or
plus 2.5 mm Hg (0.1 in) per 30 m (100 ft) elevation decrease.
6.1.3 Gas Density Determination Equipment. Temperature sensor and pressure gauge, as
described in sections 6.3 and 6.4 of Method 2, and gas analyzer, if necessary, as described in
Method 3. The temperature sensor shall, preferably, be permanently attached to the pitot tube or
sampling probe in a fixed configuration, such that the tip of the sensor extends beyond the
leading edge of the probe sheath and does not touch any metal. Alternatively, the sensor may be
attached just prior to use in the field. Note, however, that if the temperature sensor is attached in
the field, the sensor must be placed in an interference-free arrangement with respect to the Type
S pitot tube openings (see Method 2, Figure 2-4). As a second alternative, if a difference of not
more than 1 percent in the average velocity measurement is to be introduced, the temperature
sensor need not be attached to the probe or pitot tube. (This alternative is subject to the approval
of the Administrator.)
6.2 Sample Recovery. The following items are required for sample recovery:
6.2.1 Probe-Liner and Probe-Nozzle Brushes. Nylon bristle brushes with stainless steel wire
handles. The probe brush shall have extensions (at least as long as the probe) constructed of
stainless steel, Nylon, Teflon, or similarly inert material. The brushes shall be properly sized and
shaped to brush out the probe liner and nozzle.
4
Method 5 8/2/2017
6.2.2 Wash Bottles. Two Glass wash bottles are recommended. Alternatively, polyethylene wash
bottles may be used. It is recommended that acetone not be stored in polyethylene bottles for
longer than a month.
6.2.3 Glass Sample Storage Containers. Chemically resistant, borosilicate glass bottles, for
acetone washes, 500 ml or 1000 ml. Screw cap liners shall either be rubber-backed Teflon or
shall be constructed so as to be leak-free and resistant to chemical attack by acetone. (Narrow
mouth glass bottles have been found to be less prone to leakage.) Alternatively, polyethylene
bottles may be used.
6.2.4 Petri Dishes. For filter samples; glass or polyethylene, unless otherwise specified by the
Administrator.
6.2.5 Graduated Cylinder and/or Balance. To measure condensed water to within 1 ml or 0.5 g.
Graduated cylinders shall have subdivisions no greater than 2 ml.
6.2.7 Funnel and Rubber Policeman. To aid in transfer of silica gel to container; not necessary if
silica gel is weighed in the field.
6.3 Sample Analysis. The following equipment is required for sample analysis:
6.3.2 Desiccator.
7.1.1 Filters. Glass fiber filters, without organic binder, exhibiting at least 99.95 percent
efficiency (<0.05 percent penetration) on 0.3 micron dioctyl phthalate smoke particles. The filter
efficiency test shall be conducted in accordance with ASTM Method D 2986-71, 78, or 95a
(incorporated by reference—see §60.17). Test data from the supplier's quality control program
5
Method 5 8/2/2017
are sufficient for this purpose. In sources containing SO2 or SO3, the filter material must be of a
type that is unreactive to SO2 or SO3. Reference 10 in section 17.0 may be used to select the
appropriate filter.
7.1.2 Silica Gel. Indicating type, 6 to 16 mesh. If previously used, dry at 175 °C (350 °F) for 2
hours. New silica gel may be used as received. Alternatively, other types of desiccants
(equivalent or better) may be used, subject to the approval of the Administrator.
7.1.3 Water. When analysis of the material caught in the impingers is required, deionized
distilled water [to conform to ASTM D1193-77 or 91 Type 3 (incorporated by reference—see
§60.17)] with at least <0.001 percent residue shall be used or as specified in the applicable
method requiring analysis of the water. Run reagent blanks prior to field use to eliminate a high
blank on test samples.
7.2 Sample Recovery. Acetone, reagent grade, ≤0.001 percent residue, in glass bottles, is
required. Acetone from metal containers generally has a high residue blank and should not be
used. Sometimes, suppliers transfer acetone to glass bottles from metal containers; thus, acetone
blanks shall be run prior to field use and only acetone with low blank values (≤0.001 percent)
shall be used. In no case shall a blank value of greater than 0.001 percent of the weight of
acetone used be subtracted from the sample weight.
7.3 Sample Analysis. The following reagents are required for sample analysis:
7.3.2 Desiccant. Anhydrous calcium sulfate, indicating type. Alternatively, other types of
desiccants may be used, subject to the approval of the Administrator.
8.1 Pretest Preparation. It is suggested that sampling equipment be maintained according to the
procedures described in APTD-0576. Alternative mercury-free thermometers may be used if the
thermometers are at a minimum equivalent in terms of performance or suitably effective for the
specific temperature measurement application.
8.1.1 Place 200 to 300 g of silica gel in each of several air-tight containers. Weigh each
container, including silica gel, to the nearest 0.5 g, and record this weight. As an alternative, the
silica gel need not be preweighed, but may be weighed directly in its impinger or sampling
holder just prior to train assembly.
8.1.2 Check filters visually against light for irregularities, flaws, or pinhole leaks. Label filters of
the proper diameter on the back side near the edge using numbering machine ink. As an
alternative, label the shipping containers (glass or polyethylene petri dishes), and keep each filter
in its identified container at all times except during sampling.
6
Method 5 8/2/2017
8.1.3 Desiccate the filters at 20 ±5.6 °C (68 ±10 °F) and ambient pressure for at least 24 hours.
Weigh each filter (or filter and shipping container) at intervals of at least 6 hours to a constant
weight (i.e., ≤0.5 mg change from previous weighing). Record results to the nearest 0.1 mg.
During each weighing, the period for which the filter is exposed to the laboratory atmosphere
shall be less than 2 minutes. Alternatively (unless otherwise specified by the Administrator), the
filters may be oven dried at 105 °C (220 °F) for 2 to 3 hours, desiccated for 2 hours, and
weighed. Procedures other than those described, which account for relative humidity effects,
may be used, subject to the approval of the Administrator.
8.2.1 Select the sampling site and the minimum number of sampling points according to Method
1 or as specified by the Administrator. Determine the stack pressure, temperature, and the range
of velocity heads using Method 2; it is recommended that a leak check of the pitot lines (see
Method 2, section 8.1) be performed. Determine the moisture content using Approximation
Method 4 or its alternatives for the purpose of making isokinetic sampling rate settings.
Determine the stack gas dry molecular weight, as described in Method 2, section 8.6; if
integrated Method 3 sampling is used for molecular weight determination, the integrated bag
sample shall be taken simultaneously with, and for the same total length of time as, the
particulate sample run.
8.2.2 Select a nozzle size based on the range of velocity heads, such that it is not necessary to
change the nozzle size in order to maintain isokinetic sampling rates. During the run, do not
change the nozzle size. Ensure that the proper differential pressure gauge is chosen for the range
of velocity heads encountered (see section 8.3 of Method 2).
8.2.3 Select a suitable probe liner and probe length such that all traverse points can be sampled.
For large stacks, consider sampling from opposite sides of the stack to reduce the required probe
length.
8.2.4 Select a total sampling time greater than or equal to the minimum total sampling time
specified in the test procedures for the specific industry such that (l) the sampling time per point
is not less than 2 minutes (or some greater time interval as specified by the Administrator), and
(2) the sample volume taken (corrected to standard conditions) will exceed the required
minimum total gas sample volume. The latter is based on an approximate average sampling rate.
8.2.5 The sampling time at each point shall be the same. It is recommended that the number of
minutes sampled at each point be an integer or an integer plus one-half minute, in order to avoid
timekeeping errors.
8.2.6 In some circumstances (e.g., batch cycles) it may be necessary to sample for shorter times
at the traverse points and to obtain smaller gas sample volumes. In these cases, the
Administrator's approval must first be obtained.
7
Method 5 8/2/2017
8.3.1 During preparation and assembly of the sampling train, keep all openings where
contamination can occur covered until just prior to assembly or until sampling is about to begin.
Place 100 ml of water in each of the first two impingers, leave the third impinger empty, and
transfer approximately 200 to 300 g of preweighed silica gel from its container to the fourth
impinger. More silica gel may be used, but care should be taken to ensure that it is not entrained
and carried out from the impinger during sampling. Place the container in a clean place for later
use in the sample recovery. Alternatively, the weight of the silica gel plus impinger may be
determined to the nearest 0.5 g and recorded.
8.3.2 Using a tweezer or clean disposable surgical gloves, place a labeled (identified) and
weighed filter in the filter holder. Be sure that the filter is properly centered and the gasket
properly placed so as to prevent the sample gas stream from circumventing the filter. Check the
filter for tears after assembly is completed.
8.3.3 When glass probe liners are used, install the selected nozzle using a Viton A O-ring when
stack temperatures are less than 260 °C (500 °F) or a heat-resistant string gasket when
temperatures are higher. See APTD-0576 for details. Other connecting systems using either 316
stainless steel or Teflon ferrules may be used. When metal liners are used, install the nozzle as
discussed above or by a leak-free direct mechanical connection. Mark the probe with heat
resistant tape or by some other method to denote the proper distance into the stack or duct for
each sampling point.
8.3.4 Set up the train as shown in Figure 5-1 ensuring that the connections are leak-tight.
Subject to the approval of the Administrator, a glass cyclone may be used between the probe and
filter holder when the total particulate catch is expected to exceed 100 mg or when water droplets
are present in the stack gas.
8.4.1 Leak Check of Metering System Shown in Figure 5-1. That portion of the sampling train
from the pump to the orifice meter should be leak-checked prior to initial use and after each
shipment. Leakage after the pump will result in less volume being recorded than is actually
sampled. The following procedure is suggested (see Figure 5-2): Close the main valve on the
meter box. Insert a one-hole rubber stopper with rubber tubing attached into the orifice exhaust
pipe. Disconnect and vent the low side of the orifice manometer. Close off the low side orifice
tap. Pressurize the system to 13 to 18 cm (5 to 7 in.) water column by blowing into the rubber
tubing. Pinch off the tubing, and observe the manometer for one minute. A loss of pressure on
the manometer indicates a leak in the meter box; leaks, if present, must be corrected.
8.4.2 Pretest Leak Check. A pretest leak check of the sampling train is recommended, but not
required. If the pretest leak check is conducted, the following procedure should be used.
8.4.2.1 After the sampling train has been assembled, turn on and set the filter and probe heating
systems to the desired operating temperatures. Allow time for the temperatures to stabilize. If a
8
Method 5 8/2/2017
Viton A O-ring or other leak-free connection is used in assembling the probe nozzle to the probe
liner, leak-check the train at the sampling site by plugging the nozzle and pulling a 380 mm (15
in.) Hg vacuum.
NOTE: A lower vacuum may be used, provided that it is not exceeded during the test.
8.4.2.2 If a heat-resistant string is used, do not connect the probe to the train during the leak
check. Instead, leak-check the train by first plugging the inlet to the filter holder (cyclone, if
applicable) and pulling a 380 mm (15 in.) Hg vacuum (see note in section 8.4.2.1). Then connect
the probe to the train, and leak-check at approximately 25 mm (1 in.) Hg vacuum; alternatively,
the probe may be leak-checked with the rest of the sampling train, in one step, at 380 mm (15 in.)
Hg vacuum. Leakage rates in excess of 4 percent of the average sampling rate or 0.00057 m3/min
(0.020 cfm), whichever is less, are unacceptable.
8.4.2.3 The following leak-check instructions for the sampling train described in APTD-0576
and APTD-0581 may be helpful. Start the pump with the bypass valve fully open and the coarse
adjust valve completely closed. Partially open the coarse adjust valve, and slowly close the
bypass valve until the desired vacuum is reached. Do not reverse the direction of the bypass
valve, as this will cause water to back up into the filter holder. If the desired vacuum is exceeded,
either leak-check at this higher vacuum, or end the leak check and start over.
8.4.2.4 When the leak check is completed, first slowly remove the plug from the inlet to the
probe, filter holder, or cyclone (if applicable), and immediately turn off the vacuum pump. This
prevents the water in the impingers from being forced backward into the filter holder and the
silica gel from being entrained backward into the third impinger.
8.4.3 Leak Checks During Sample Run. If, during the sampling run, a component (e.g., filter
assembly or impinger) change becomes necessary, a leak check shall be conducted immediately
before the change is made. The leak check shall be done according to the procedure outlined in
section 8.4.2 above, except that it shall be done at a vacuum equal to or greater than the
maximum value recorded up to that point in the test. If the leakage rate is found to be no greater
than 0.00057 m3/min (0.020 cfm) or 4 percent of the average sampling rate (whichever is less),
the results are acceptable, and no correction will need to be applied to the total volume of dry gas
metered; if, however, a higher leakage rate is obtained, either record the leakage rate and plan to
correct the sample volume as shown in section 12.3 of this method, or void the sample run.
NOTE: Immediately after component changes, leak checks are optional. If such leak checks are
done, the procedure outlined in section 8.4.2 above should be used.
8.4.4 Post-Test Leak Check. A leak check of the sampling train is mandatory at the conclusion of
each sampling run. The leak check shall be performed in accordance with the procedures
outlined in section 8.4.2, except that it shall be conducted at a vacuum equal to or greater than
the maximum value reached during the sampling run. If the leakage rate is found to be no greater
than 0.00057 m3 min (0.020 cfm) or 4 percent of the average sampling rate (whichever is less),
the results are acceptable, and no correction need be applied to the total volume of dry gas
9
Method 5 8/2/2017
metered. If, however, a higher leakage rate is obtained, either record the leakage rate and correct
the sample volume as shown in section 12.3 of this method, or void the sampling run.
8.5 Sampling Train Operation. During the sampling run, maintain an isokinetic sampling rate
(within 10 percent of true isokinetic unless otherwise specified by the Administrator) and a
sample gas temperature through the filter of 120 ± 14 °C (248 ± 25 °F) or such other temperature
as specified by an applicable subpart of the standards or approved by the Administrator.
8.5.1 For each run, record the data required on a data sheet such as the one shown in Figure 5-3.
Be sure to record the initial DGM reading. Record the DGM readings at the beginning and end of
each sampling time increment, when changes in flow rates are made, before and after each leak
check, and when sampling is halted. Take other readings indicated by Figure 5-3 at least once at
each sample point during each time increment and additional readings when significant changes
(20 percent variation in velocity head readings) necessitate additional adjustments in flow rate.
Level and zero the manometer. Because the manometer level and zero may drift due to vibrations
and temperature changes, make periodic checks during the traverse.
8.5.2 Clean the portholes prior to the test run to minimize the chance of collecting deposited
material. To begin sampling, verify that the filter and probe heating systems are up to
temperature, remove the nozzle cap, verify that the pitot tube and probe are properly positioned.
Position the nozzle at the first traverse point with the tip pointing directly into the gas stream.
Immediately start the pump, and adjust the flow to isokinetic conditions. Nomographs are
available which aid in the rapid adjustment of the isokinetic sampling rate without excessive
computations. These nomographs are designed for use when the Type S pitot tube coefficient
(Cp) is 0.85 ±0.02, and the stack gas equivalent density [dry molecular weight (Md)] is equal to
29 ±4. APTD-0576 details the procedure for using the nomographs. If Cp and Md are outside the
above stated ranges, do not use the nomographs unless appropriate steps (see Reference 7 in
section 17.0) are taken to compensate for the deviations.
8.5.3 When the stack is under significant negative pressure (i.e., height of impinger stem), take
care to close the coarse adjust valve before inserting the probe into the stack to prevent water
from backing into the filter holder. If necessary, the pump may be turned on with the coarse
adjust valve closed.
8.5.4 When the probe is in position, block off the openings around the probe and porthole to
prevent unrepresentative dilution of the gas stream.
8.5.6 During the test run, make periodic adjustments to keep the temperature around the filter
holder at the proper level to maintain the sample gas temperature exiting the filter; add more ice
and, if necessary, salt to maintain a temperature of less than 20 °C (68 °F) at the condenser/silica
gel outlet. Also, periodically check the level and zero of the manometer.
10
Method 5 8/2/2017
8.5.7 If the pressure drop across the filter becomes too high, making isokinetic sampling difficult
to maintain, the filter may be replaced in the midst of the sample run. It is recommended that
another complete filter assembly be used rather than attempting to change the filter itself. Before
a new filter assembly is installed, conduct a leak check (see section 8.4.3). The total PM weight
shall include the summation of the filter assembly catches.
8.5.8 A single train shall be used for the entire sample run, except in cases where simultaneous
sampling is required in two or more separate ducts or at two or more different locations within
the same duct, or in cases where equipment failure necessitates a change of trains. In all other
situations, the use of two or more trains will be subject to the approval of the Administrator.
NOTE: When two or more trains are used, separate analyses of the front-half and (if applicable)
impinger catches from each train shall be performed, unless identical nozzle sizes were used on
all trains, in which case, the front-half catches from the individual trains may be combined (as
may the impinger catches) and one analysis of front-half catch and one analysis of impinger
catch may be performed. Consult with the Administrator for details concerning the calculation of
results when two or more trains are used.
8.5.9 At the end of the sample run, close the coarse adjust valve, remove the probe and nozzle
from the stack, turn off the pump, record the final DGM meter reading, and conduct a post-test
leak check, as outlined in section 8.4.4. Also, leak-check the pitot lines as described in Method 2,
section 8.1. The lines must pass this leak check, in order to validate the velocity head data.
8.6 Calculation of Percent Isokinetic. Calculate percent isokinetic (see Calculations, section
12.11) to determine whether the run was valid or another test run should be made. If there was
difficulty in maintaining isokinetic rates because of source conditions, consult with the
Administrator for possible variance on the isokinetic rates.
8.7.1 Proper cleanup procedure begins as soon as the probe is removed from the stack at the end
of the sampling period. Allow the probe to cool.
8.7.2 When the probe can be safely handled, wipe off all external PM near the tip of the probe
nozzle, and place a cap over it to prevent losing or gaining PM. Do not cap off the probe tip
tightly while the sampling train is cooling down. This would create a vacuum in the filter holder,
thereby drawing water from the impingers into the filter holder.
8.7.3 Before moving the sample train to the cleanup site, remove the probe from the sample train
and cap the open outlet of the probe. Be careful not to lose any condensate that might be present.
Cap the filter inlet where the probe was fastened. Remove the umbilical cord from the last
impinger, and cap the impinger. If a flexible line is used between the first impinger or condenser
and the filter holder, disconnect the line at the filter holder, and let any condensed water or liquid
drain into the impingers or condenser. Cap off the filter holder outlet and impinger inlet. Either
ground-glass stoppers, plastic caps, or serum caps may be used to close these openings.
11
Method 5 8/2/2017
8.7.4 Transfer the probe and filter-impinger assembly to the cleanup area. This area should be
clean and protected from the wind so that the chances of contaminating or losing the sample will
be minimized.
8.7.5 Save a portion of the acetone used for cleanup as a blank. From each storage container of
acetone used for cleanup, save 200 ml and place in a glass sample container labeled “acetone
blank.” To minimize any particulate contamination, rinse the wash bottle prior to filling from the
tested container.
8.7.6 Inspect the train prior to and during disassembly, and note any abnormal conditions. Treat
the samples as follows:
8.7.6.1 Container No. 1. Carefully remove the filter from the filter holder, and place it in its
identified petri dish container. Use a pair of tweezers and/or clean disposable surgical gloves to
handle the filter. If it is necessary to fold the filter, do so such that the PM cake is inside the fold.
Using a dry Nylon bristle brush and/or a sharp-edged blade, carefully transfer to the petri dish
any PM and/or filter fibers that adhere to the filter holder gasket. Seal the container.
8.7.6.2 Container No. 2. Taking care to see that dust on the outside of the probe or other exterior
surfaces does not get into the sample, quantitatively recover PM or any condensate from the
probe nozzle, probe fitting, probe liner, and front half of the filter holder by washing these
components with acetone and placing the wash in a glass container. Deionized distilled water
may be used instead of acetone when approved by the Administrator and shall be used when
specified by the Administrator. In these cases, save a water blank, and follow the Administrator's
directions on analysis. Perform the acetone rinse as follows:
8.7.6.2.1 Carefully remove the probe nozzle. Clean the inside surface by rinsing with acetone
from a wash bottle and brushing with a Nylon bristle brush. Brush until the acetone rinse shows
no visible particles, after which make a final rinse of the inside surface with acetone.
8.7.6.2.2 Brush and rinse the inside parts of the fitting with acetone in a similar way until no
visible particles remain.
8.7.6.2.3 Rinse the probe liner with acetone by tilting and rotating the probe while squirting
acetone into its upper end so that all inside surfaces will be wetted with acetone. Let the acetone
drain from the lower end into the sample container. A funnel (glass or polyethylene) may be used
to aid in transferring liquid washes to the container. Follow the acetone rinse with a probe brush.
Hold the probe in an inclined position, squirt acetone into the upper end as the probe brush is
being pushed with a twisting action through the probe; hold a sample container underneath the
lower end of the probe, and catch any acetone and particulate matter that is brushed from the
probe. Run the brush through the probe three times or more until no visible PM is carried out
with the acetone or until none remains in the probe liner on visual inspection. With stainless steel
or other metal probes, run the brush through in the above prescribed manner at least six times
since metal probes have small crevices in which particulate matter can be entrapped. Rinse the
brush with acetone, and quantitatively collect these washings in the sample container. After the
brushing, make a final acetone rinse of the probe.
12
Method 5 8/2/2017
8.7.6.2.4 It is recommended that two people clean the probe to minimize sample losses. Between
sampling runs, keep brushes clean and protected from contamination.
8.7.6.2.5 Clean the inside of the front half of the filter holder by rubbing the surfaces with a
Nylon bristle brush and rinsing with acetone. Rinse each surface three times or more if needed to
remove visible particulate. Make a final rinse of the brush and filter holder. Carefully rinse out
the glass cyclone, also (if applicable). After all acetone washings and particulate matter have
been collected in the sample container, tighten the lid on the sample container so that acetone
will not leak out when it is shipped to the laboratory. Mark the height of the fluid level to allow
determination of whether leakage occurred during transport. Label the container to clearly
identify its contents.
8.7.6.3 Container No. 3. Note the color of the indicating silica gel to determine whether it has
been completely spent, and make a notation of its condition. Transfer the silica gel from the
fourth impinger to its original container, and seal. A funnel may make it easier to pour the silica
gel without spilling. A rubber policeman may be used as an aid in removing the silica gel from
the impinger. It is not necessary to remove the small amount of dust particles that may adhere to
the impinger wall and are difficult to remove. Since the gain in weight is to be used for moisture
calculations, do not use any water or other liquids to transfer the silica gel. If a balance is
available in the field, follow the procedure for Container No. 3 in section 11.2.3.
8.7.6.4 Impinger Water. Treat the impingers as follows: Make a notation of any color or film in
the liquid catch. Measure the liquid that is in the first three impingers to within 1 ml by using a
graduated cylinder or by weighing it to within 0.5 g by using a balance. Record the volume or
weight of liquid present. This information is required to calculate the moisture content of the
effluent gas. Discard the liquid after measuring and recording the volume or weight, unless
analysis of the impinger catch is required (see NOTE, section 6.1.1.8). If a different type of
condenser is used, measure the amount of moisture condensed either volumetrically or
gravimetrically.
8.8 Sample Transport. Whenever possible, containers should be shipped in such a way that they
remain upright at all times.
9.2 Volume Metering System Checks. The following procedures are suggested to check the
volume metering system calibration values at the field test site prior to sample collection. These
procedures are optional.
13
Method 5 8/2/2017
9.2.1 Meter Orifice Check. Using the calibration data obtained during the calibration procedure
described in section 10.3, determine the ΔH@ for the metering system orifice. The ΔH@ is the
orifice pressure differential in units of in. H2O that correlates to 0.75 cfm of air at 528 °R and
29.92 in. Hg. The ΔH@ is calculated as follows:
Where:
9.2.1.1 Before beginning the field test (a set of three runs usually constitutes a field test), operate
the metering system (i.e., pump, volume meter, and orifice) at the ΔH@ pressure differential for
10 minutes. Record the volume collected, the DGM temperature, and the barometric pressure.
Calculate a DGM calibration check value, Yc, as follows:
where:
9.2.1.2 Compare the Yc value with the dry gas meter calibration factor Y to determine that:
0.97Y <Yc <1.03Y. If the Yc value is not within this range, the volume metering system should
be investigated before beginning the test.
9.2.2 Calibrated Critical Orifice. A critical orifice, calibrated against a wet test meter or
spirometer and designed to be inserted at the inlet of the sampling meter box, may be used as a
check by following the procedure of section 16.2.
14
Method 5 8/2/2017
10.1 Probe Nozzle. Probe nozzles shall be calibrated before their initial use in the field. Using a
micrometer, measure the ID of the nozzle to the nearest 0.025 mm (0.001 in.). Make three
separate measurements using different diameters each time, and obtain the average of the
measurements. The difference between the high and low numbers shall not exceed 0.1 mm
(0.004 in.). When nozzles become nicked, dented, or corroded, they shall be reshaped,
sharpened, and recalibrated before use. Each nozzle shall be permanently and uniquely
identified.
10.2 Pitot Tube Assembly. The Type S pitot tube assembly shall be calibrated according to the
procedure outlined in section 10.1 of Method 2.
10.3.1 Calibration Prior to Use. Before its initial use in the field, the metering system shall be
calibrated as follows: Connect the metering system inlet to the outlet of a wet test meter that is
accurate to within 1 percent. Refer to Figure 5-4. The wet test meter should have a capacity of 30
liters/rev (1 ft3/rev). A spirometer of 400 liters (14 ft3) or more capacity, or equivalent, may be
used for this calibration, although a wet test meter is usually more practical. The wet test meter
should be periodically calibrated with a spirometer or a liquid displacement meter to ensure the
accuracy of the wet test meter. Spirometers or wet test meters of other sizes may be used,
provided that the specified accuracies of the procedure are maintained. Run the metering system
pump for about 15 minutes with the orifice manometer indicating a median reading as expected
in field use to allow the pump to warm up and to permit the interior surface of the wet test meter
to be thoroughly wetted. Then, at each of a minimum of three orifice manometer settings, pass an
exact quantity of gas through the wet test meter and note the gas volume indicated by the DGM.
Also note the barometric pressure and the temperatures of the wet test meter, the inlet of the
DGM, and the outlet of the DGM. Select the highest and lowest orifice settings to bracket the
expected field operating range of the orifice. Use a minimum volume of 0.14 m3 (5 ft3) at all
orifice settings. Record all the data on a form similar to Figure 5-5 and calculate Y, the DGM
calibration factor, and ΔH , the orifice calibration factor, at each orifice setting as shown on
Figure 5-5. Allowable tolerances for individual Y and ΔH values are given in Figure 5-5. Use
the average of the Y values in the calculations in section 12.0.
10.3.1.1 Before calibrating the metering system, it is suggested that a leak check be conducted.
For metering systems having diaphragm pumps, the normal leak-check procedure will not detect
leakages within the pump. For these cases the following leak-check procedure is suggested:
make a 10-minute calibration run at 0.00057 m3/min (0.020 cfm). At the end of the run, take the
difference of the measured wet test meter and DGM volumes. Divide the difference by 10 to get
the leak rate. The leak rate should not exceed 0.00057 m3/min (0.020 cfm).
10.3.2 Calibration After Use. After each field use, the calibration of the metering system shall be
checked by performing three calibration runs at a single, intermediate orifice setting (based on
the previous field test), with the vacuum set at the maximum value reached during the test series.
To adjust the vacuum, insert a valve between the wet test meter and the inlet of the metering
15
Method 5 8/2/2017
system. Calculate the average value of the DGM calibration factor. If the value has changed by
more than 5 percent, recalibrate the meter over the full range of orifice settings, as detailed in
section 10.3.1.
NOTE: Alternative procedures (e.g., rechecking the orifice meter coefficient) may be used,
subject to the approval of the Administrator.
10.3.3 Acceptable Variation in Calibration Check. If the DGM coefficient values obtained before
and after a test series differ by more than 5 percent, the test series shall either be voided, or
calculations for the test series shall be performed using whichever meter coefficient value (i.e.,
before or after) gives the lower value of total sample volume.
10.4 Probe Heater Calibration. Use a heat source to generate air heated to selected temperatures
that approximate those expected to occur in the sources to be sampled. Pass this air through the
probe at a typical sample flow rate while measuring the probe inlet and outlet temperatures at
various probe heater settings. For each air temperature generated, construct a graph of probe
heating system setting versus probe outlet temperature. The procedure outlined in APTD-0576
can also be used. Probes constructed according to APTD-0581 need not be calibrated if the
calibration curves in APTD-0576 are used. Also, probes with outlet temperature monitoring
capabilities do not require calibration.
NOTE: The probe heating system shall be calibrated before its initial use in the field.
10.5 Temperature Sensors. Use the procedure in Section 10.3 of Method 2 to calibrate in-stack
temperature sensors. Dial thermometers, such as are used for the DGM and condenser outlet,
shall be calibrated against mercury-in-glass thermometers. An alternative mercury-free NIST-
traceable thermometer may be used if the thermometer is, at a minimum, equivalent in terms of
performance or suitably effective for the specific temperature measurement application. As an
alternative, the following single-point calibration procedure may be used. After each test run
series, check the accuracy (and, hence, the calibration) of each thermocouple system at ambient
temperature, or any other temperature, within the range specified by the manufacturer, using a
reference thermometer (either ASTM reference thermometer or a thermometer that has been
calibrated against an ASTM reference thermometer). The temperatures of the thermocouple and
reference thermometers shall agree to within ±2 °F.
10.6 Barometer. Calibrate against a mercury barometer or NIST-traceable barometer prior to the
field test. Alternatively, barometric pressure may be obtained from a weather report that has been
adjusted for the test point (on the stack) elevation.
10.7 Field Balance Calibration Check. Check the calibration of the balance used to weigh
impingers with a weight that is at least 500g or within 50g of a loaded impinger. The weight
must be ASTM E617-13 “Standard Specification for Laboratory Weights and Precision Mass
Standards” (incorporated by reference-see 40 CFR 60.17) Class 6 (or better). Daily before use,
the field balance must measure the weight within ± 0.5g of the certified mass. If the daily
balance calibration check fails, perform corrective measures and repeat the check before using
balance.
16
Method 5 8/2/2017
10.8 Analytical Balance Calibration. Perform a multipoint calibration (at least five points
spanning the operational range) of the analytical balance before the first use, and semiannually
thereafter. The calibration of the analytical balance must be conducted using ASTM E617-13
“Standard Specification for Laboratory Weights and Precision Mass Standards” (incorporated by
reference-see 40 CFR 60.17) Class 2 (or better) tolerance weights. Audit the balance each day it
is used for gravimetric measurements by weighing at least one ASTM E617-13 Class 2 tolerance
(or better) calibration weight that corresponds to 50 to 150 percent of the weight of one filter or
between 1g and 5g. If the scale cannot reproduce the value of the calibration weight to within 0.5
mg of the certified mass, perform corrective measures, and conduct the multipoint calibration
before use.
11.1 Record the data required on a sheet such as the one shown in Figure 5-6.
11.2.1 Container No. 1. Leave the contents in the shipping container or transfer the filter and any
loose PM from the sample container to a tared glass weighing dish. Desiccate for 24 hours in a
desiccator containing anhydrous calcium sulfate. Weigh to a constant weight, and report the
results to the nearest 0.1 mg. For the purposes of this section, the term “constant weight” means a
difference of no more than 0.5 mg or 1 percent of total weight less tare weight, whichever is
greater, between two consecutive weighings, with no less than 6 hours of desiccation time
between weighings. Alternatively, the sample may be oven dried at 104 °C (220 °F) for 2 to 3
hours, cooled in the desiccator, and weighed to a constant weight, unless otherwise specified by
the Administrator. The sample may be oven dried at 104 °C (220 °F) for 2 to 3 hours. Once the
sample has cooled, weigh the sample, and use this weight as a final weight.
11.2.2 Container No. 2. Note the level of liquid in the container, and confirm on the analysis
sheet whether leakage occurred during transport. If a noticeable amount of leakage has occurred,
either void the sample or use methods, subject to the approval of the Administrator, to correct the
final results. Measure the liquid in this container either volumetrically to ±1 ml or
gravimetrically to ±0.5 g. Transfer the contents to a tared 250 ml beaker, and evaporate to
dryness at ambient temperature and pressure. Desiccate for 24 hours, and weigh to a constant
weight. Report the results to the nearest 0.1 mg.
11.2.3 Container No. 3. Weigh the spent silica gel (or silica gel plus impinger) to the nearest 0.5
g using a balance. This step may be conducted in the field.
11.2.4 Acetone Blank Container. Measure the acetone in this container either volumetrically or
gravimetrically. Transfer the acetone to a tared 250 ml beaker, and evaporate to dryness at
ambient temperature and pressure. Desiccate for 24 hours, and weigh to a constant weight.
Report the results to the nearest 0.1 mg.
NOTE: The contents of Container No. 2 as well as the acetone blank container may be evaporated
at temperatures higher than ambient. If evaporation is done at an elevated temperature, the
17
Method 5 8/2/2017
temperature must be below the boiling point of the solvent; also, to prevent “bumping,” the
evaporation process must be closely supervised, and the contents of the beaker must be swirled
occasionally to maintain an even temperature. Use extreme care, as acetone is highly flammable
and has a low flash point.
Carry out calculations, retaining at least one extra significant figure beyond that of the acquired
data. Round off figures after the final calculation. Other forms of the equations may be used,
provided that they give equivalent results.
12.1 Nomenclature.
18
Method 5 8/2/2017
12.2 Average Dry Gas Meter Temperature and Average Orifice Pressure Drop. See data sheet
(Figure 5-3).
19
Method 5 8/2/2017
12.3 Dry Gas Volume. Correct the sample volume measured by the dry gas meter to standard
conditions (20 °C, 760 mm Hg or 68 °F, 29.92 in. Hg) by using Equation 5-1.
Where:
K1 = 0.3858 °K/mm Hg for metric units, = 17.64 °R/in. Hg for English units.
NOTE: Equation 5-1 can be used as written unless the leakage rate observed during any of the
mandatory leak checks (i.e., the post-test leak check or leak checks conducted prior to
component changes) exceeds La. If Lp or Li exceeds La, Equation 5-1 must be modified as
follows:
(a) Case I. No component changes made during sampling run. In this case, replace Vm in
Equation 5-1 with the expression:
(b) Case II. One or more component changes made during the sampling run. In this case, replace
Vm in Equation 5-1 by the expression:
and substitute only for those leakage rates (Li or Lp) which exceed La.
Where:
K2 = 0.001333 m3/ml for metric units, = 0.04706 ft3/ml for English units.
20
Method 5 8/2/2017
NOTE: In saturated or water droplet-laden gas streams, two calculations of the moisture content
of the stack gas shall be made, one from the impinger analysis (Equation 5-3), and a second from
the assumption of saturated conditions. The lower of the two values of Bws shall be considered
correct. The procedure for determining the moisture content based upon the assumption of
saturated conditions is given in section 4.0 of Method 4. For the purposes of this method, the
average stack gas temperature from Figure 5-3 may be used to make this determination, provided
that the accuracy of the in-stack temperature sensor is ±1 °C (2 °F).
12.8 Total Particulate Weight. Determine the total particulate matter catch from the sum of the
weights obtained from Containers 1 and 2 less the acetone blank (see Figure 5-6).
NOTE: In no case shall a blank value of greater than 0.001 percent of the weight of acetone used
be subtracted from the sample weight. Refer to section 8.5.8 to assist in calculation of results
involving two or more filter assemblies or two or more sampling trains.
Where:
21
Method 5 8/2/2017
From To Multiply by
ft3 m3 0.02832
gr mg 64.80004
gr/ft3 mg/m3 2288.4
mg g 0.001
gr lb 1.429 × 10−4
Where:
Where:
12.11.3 Acceptable Results. If 90 percent ≤I ≤110 percent, the results are acceptable. If the PM
results are low in comparison to the standard, and “I” is over 110 percent or less than 90 percent,
the Administrator may opt to accept the results. Reference 4 in section 17.0 may be used to make
acceptability judgments. If “I” is judged to be unacceptable, reject the results, and repeat the
sampling run.
22
Method 5 8/2/2017
12.12 Stack Gas Velocity and Volumetric Flow Rate. Calculate the average stack gas velocity
and volumetric flow rate, if needed, using data obtained in this method and the equations in
sections 12.3 and 12.4 of Method 2.
16.1 Dry Gas Meter as a Calibration Standard. A DGM may be used as a calibration standard for
volume measurements in place of the wet test meter specified in section 10.3, provided that it is
calibrated initially and recalibrated periodically as follows:
16.1.1.1. The DGM to be calibrated and used as a secondary reference meter should be of high
quality and have an appropriately sized capacity (e.g., 3 liters/rev (0.1 ft3/rev)). A spirometer
(400 liters (14 ft3) or more capacity), or equivalent, may be used for this calibration, although a
wet test meter is usually more practical. The wet test meter should have a capacity of 30
liters/rev (1 ft3/rev) and capable of measuring volume to within 1.0 percent. Wet test meters
should be checked against a spirometer or a liquid displacement meter to ensure the accuracy of
the wet test meter. Spirometers or wet test meters of other sizes may be used, provided that the
specified accuracies of the procedure are maintained.
16.1.1.2 Set up the components as shown in Figure 5-7. A spirometer, or equivalent, may be used
in place of the wet test meter in the system. Run the pump for at least 5 minutes at a flow rate of
about 10 liters/min (0.35 cfm) to condition the interior surface of the wet test meter. The pressure
drop indicated by the manometer at the inlet side of the DGM should be minimized (no greater
than 100 mm H2O (4 in. H2O) at a flow rate of 30 liters/min (1 cfm)). This can be accomplished
by using large diameter tubing connections and straight pipe fittings.
16.1.1.3 Collect the data as shown in the example data sheet (see Figure 5-8). Make triplicate
runs at each of the flow rates and at no less than five different flow rates. The range of flow rates
should be between 10 and 34 liters/min (0.35 and 1.2 cfm) or over the expected operating range.
16.1.1.4 Calculate flow rate, Q, for each run using the wet test meter volume, VW, and the run
time, θ. Calculate the DGM coefficient, Yds, for each run. These calculations are as follows:
23
Method 5 8/2/2017
Where:
16.1.1.5 Compare the three Yds values at each of the flow rates and determine the maximum and
minimum values. The difference between the maximum and minimum values at each flow rate
should be no greater than 0.030. Extra sets of triplicate runs may be made in order to complete
this requirement. In addition, the meter coefficients should be between 0.95 and 1.05. If these
specifications cannot be met in three sets of successive triplicate runs, the meter is not suitable as
a calibration standard and should not be used as such. If these specifications are met, average the
three Yds values at each flow rate resulting in no less than five average meter coefficients, Yds.
16.1.1.6 Prepare a curve of meter coefficient, Yds, versus flow rate, Q, for the DGM. This curve
shall be used as a reference when the meter is used to calibrate other DGMs and to determine
whether recalibration is required.
16.1.2.1 Recalibrate the standard DGM against a wet test meter or spirometer annually or after
every 200 hours of operation, whichever comes first. This requirement is valid provided the
standard DGM is kept in a laboratory and, if transported, cared for as any other laboratory
instrument. Abuse to the standard meter may cause a change in the calibration and will require
more frequent recalibrations.
24
Method 5 8/2/2017
calibration curve. If the two coefficients are within 1.5 percent of the calibration curve values at
the same flow rates, the meter need not be recalibrated until the next date for a recalibration
check.
16.2 Critical Orifices As Calibration Standards. Critical orifices may be used as calibration
standards in place of the wet test meter specified in section 16.1, provided that they are selected,
calibrated, and used as follows:
16.2.1.1 The procedure that follows describes the use of hypodermic needles or stainless steel
needle tubings which have been found suitable for use as critical orifices. Other materials and
critical orifice designs may be used provided the orifices act as true critical orifices (i.e., a
critical vacuum can be obtained, as described in section 16.2.2.2.3). Select five critical orifices
that are appropriately sized to cover the range of flow rates between 10 and 34 liters/min (0.35
and 1.2 cfm) or the expected operating range. Two of the critical orifices should bracket the
expected operating range. A minimum of three critical orifices will be needed to calibrate a
Method 5 DGM; the other two critical orifices can serve as spares and provide better selection
for bracketing the range of operating flow rates. The needle sizes and tubing lengths shown in
Table 5-1 in section 18.0 give the approximate flow rates.
16.2.1.2 These needles can be adapted to a Method 5 type sampling train as follows: Insert a
serum bottle stopper, 13 by 20 mm sleeve type, into a 1⁄2 -inch Swagelok (or equivalent) quick
connect. Insert the needle into the stopper as shown in Figure 5-9.
16.2.2 Critical Orifice Calibration. The procedure described in this section uses the Method 5
meter box configuration with a DGM as described in section 6.1.1.9 to calibrate the critical
orifices. Other schemes may be used, subject to the approval of the Administrator.
16.2.2.1 Calibration of Meter Box. The critical orifices must be calibrated in the same
configuration as they will be used (i.e., there should be no connections to the inlet of the orifice).
16.2.2.1.1 Before calibrating the meter box, leak check the system as follows: Fully open the
coarse adjust valve, and completely close the by-pass valve. Plug the inlet. Then turn on the
pump, and determine whether there is any leakage. The leakage rate shall be zero (i.e., no
detectable movement of the DGM dial shall be seen for 1 minute).
16.2.2.1.2 Check also for leakages in that portion of the sampling train between the pump and the
orifice meter. See section 8.4.1 for the procedure; make any corrections, if necessary. If leakage
is detected, check for cracked gaskets, loose fittings, worn O-rings, etc., and make the necessary
repairs.
16.2.2.1.3 After determining that the meter box is leakless, calibrate the meter box according to
the procedure given in section 10.3. Make sure that the wet test meter meets the requirements
stated in section 16.1.1.1. Check the water level in the wet test meter. Record the DGM
calibration factor, Y.
25
Method 5 8/2/2017
16.2.2.2 Calibration of Critical Orifices. Set up the apparatus as shown in Figure 5-10.
16.2.2.2.1 Allow a warm-up time of 15 minutes. This step is important to equilibrate the
temperature conditions through the DGM.
16.2.2.2.2 Leak check the system as in section 16.2.2.1.1. The leakage rate shall be zero.
16.2.2.2.3 Before calibrating the critical orifice, determine its suitability and the appropriate
operating vacuum as follows: Turn on the pump, fully open the coarse adjust valve, and adjust
the by-pass valve to give a vacuum reading corresponding to about half of atmospheric pressure.
Observe the meter box orifice manometer reading, ΔH. Slowly increase the vacuum reading until
a stable reading is obtained on the meter box orifice manometer. Record the critical vacuum for
each orifice. Orifices that do not reach a critical value shall not be used.
16.2.2.2.4 Obtain the barometric pressure using a barometer as described in section 6.1.2. Record
the barometric pressure, Pbar, in mm Hg (in. Hg).
Where:
Calculate the arithmetic mean of the K′ values. The individual K' values should not differ by
more than ±0.5 percent from the mean value.
26
Method 5 8/2/2017
16.2.3.2 Calibrate the metering system according to the procedure outlined in section 16.2.2.
Record the information listed in Figure 5-12.
16.2.3.3 Calculate the standard volumes of air passed through the DGM and the critical orifices,
and calculate the DGM calibration factor, Y, using the equations below:
Where:
Vcr(std) = Volume of gas sample passed through the critical orifice, corrected to standard
conditions, dscm (dscf).
K1 = 0.3858 K/mm Hg for metric units
= 17.64 °R/in. Hg for English units.
16.2.3.4 Average the DGM calibration values for each of the flow rates. The calibration factor,
Y, at each of the flow rates should not differ by more than ±2 percent from the average.
16.2.3.5 To determine the need for recalibrating the critical orifices, compare the DGM Y factors
obtained from two adjacent orifices each time a DGM is calibrated; for example, when checking
orifice 13/2.5, use orifices 12/10.2 and 13/5.1. If any critical orifice yields a DGM Y factor
differing by more than 2 percent from the others, recalibrate the critical orifice according to
section 16.2.2.
16.3 Alternative Post-Test Metering System Calibration. The following procedure may be used
as an alternative to the post-test calibration described in Section 10.3.2. This alternative
procedure does not detect leakages between the inlet of the metering system and the dry gas
meter. Therefore, two steps must be included to make it an equivalent alternative:
(1) The metering system must pass the post-test leak-check from either the inlet of the sampling
train or the inlet of the metering system. Therefore, if the train fails the former leak-check,
another leak-check from the inlet of the metering system must be conducted;
(2) The metering system must pass the leak-check of that portion of the train from the pump to
the orifice meter as described in Section 8.4.1.
27
Method 5 8/2/2017
16.3.1.1 Ensure that the metering system has passed the post-test leak-check. If not, conduct a
leak-check of the metering system from its inlet.
16.3.1.2 Conduct the leak-check of that portion of the train from the pump to the orifice meter as
described in Section 10.3.1.1.
16.3.1.3 Calculate Yqa for each test run using the following equation:
Where:
Yqa = Dry gas meter calibration check value, dimensionless.
0.0319 = (29.92/528) (0.75)2 (in. Hg/°R) cfm2.
ΔH@ = Orifice meter calibration coefficient, in. H2O.
Md = Dry molecular weight of stack gas, lb/lb-mole.
29 = Dry molecular weight of air, lb/lb-mole.
16.3.2.1 Average the three or more Yqa's obtained from the test run series and compare this
average Yqa with the dry gas meter calibration factor Y. The average Yqa must be within 5
percent of Y.
16.3.2.2 If the average Yqa does not meet the 5 percent criterion, recalibrate the meter over the
full range of orifice settings as detailed in Section 10.3.1. Then follow the procedure in Section
10.3.3.
17.0 References.
1. Addendum to Specifications for Incinerator Testing at Federal Facilities. PHS, NCAPC.
December 6, 1967.
2. Martin, Robert M. Construction Details of Isokinetic Source-Sampling Equipment.
Environmental Protection Agency. Research Triangle Park, NC. APTD-0581. April 1971.
3. Rom, Jerome J. Maintenance, Calibration, and Operation of Isokinetic Source Sampling
Equipment. Environmental Protection Agency. Research Triangle Park, NC. APTD-0576. March
1972.
28
Method 5 8/2/2017
4. Smith, W.S., R.T. Shigehara, and W.F. Todd. A Method of Interpreting Stack Sampling Data.
Paper Presented at the 63rd Annual Meeting of the Air Pollution Control Association, St. Louis,
MO. June 14-19, 1970.
5. Smith, W.S., et al. Stack Gas Sampling Improved and Simplified With New Equipment.
APCA Paper No. 67-119. 1967.
6. Specifications for Incinerator Testing at Federal Facilities. PHS, NCAPC. 1967.
7. Shigehara, R.T. Adjustment in the EPA Nomograph for Different Pitot Tube Coefficients and
Dry Molecular Weights. Stack Sampling News 2:4-11. October 1974.
8. Vollaro, R.F. A Survey of Commercially Available Instrumentation for the Measurement of
Low-Range Gas Velocities. U.S. Environmental Protection Agency, Emission Measurement
Branch. Research Triangle Park, NC. November 1976 (unpublished paper).
9. Annual Book of ASTM Standards. Part 26. Gaseous Fuels; Coal and Coke; Atmospheric
Analysis. American Society for Testing and Materials. Philadelphia, PA. 1974. pp. 617-622.
10. Felix, L.G., G.I. Clinard, G.E. Lacy, and J.D. McCain. Inertial Cascade Impactor Substrate
Media for Flue Gas Sampling. U.S. Environmental Protection Agency. Research Triangle Park,
NC 27711. Publication No. EPA-600/7-77-060. June 1977. 83 pp.
11. Westlin, P.R. and R.T. Shigehara. Procedure for Calibrating and Using Dry Gas Volume
Meters as Calibration Standards. Source Evaluation Society Newsletter. 3(1):17-30. February
1978.
12. Lodge, J.P., Jr., J.B. Pate, B.E. Ammons, and G.A. Swanson. The Use of Hypodermic
Needles as Critical Orifices in Air Sampling. J. Air Pollution Control Association. 16:197-200.
1966.
13. Shigehara, Roger T., P.G. Royals, and E.W. Steward. “Alternative Method 5 Post-Test
Calibration.” Entropy Incorporated, Research Triangle Park, NC 27709.
TABLE 5-1 FLOR RATES FOR VARIOUS NEEDLE SIZES AND TUBE LENGTHS
29
Method 5 8/2/2017
30
Method 5 8/2/2017
31
Method 5 8/2/2017
32
Method 5 8/2/2017
33
Method 5 8/2/2017
34
Method 5 8/2/2017
Plant__________________________________
Date __________________________________
Run No. _______________________________
Filter No. ______________________________
Amount liquid lost during transport __________________________________
Acetone blank volume, m1_________________________________________
Acetone blank concentration, mg/mg (Equation 5-4) ____________________
Acetone wash blank, mg (Equation 5-5) ______________________________
35
Method 5 8/2/2017
36
Method 5 8/2/2017
37
Method 5 8/2/2017
38
Method 5 8/2/2017
39
Method 5 8/2/2017
Date _____________________
Train ID __________________
DGM cal. Factor ___________
Critical orifice ID__________
Run No.
Dry gas meter 1 2
Final reading m3 (ft3)
Initial reading m3 (ft3)
Difference, Vm m3 (ft3)
Inlet/Outlet
Temperatures: °C (( °deg;F) / /
Initial °C (( °deg;F) / /
Final min/sec / /
Av. Temeperature, tm min
Time, θ
Orifice man. rdg., ΔH mm (in.) H2
Bar. pressure, Pbar mm (in.) Hg
Ambient temperature, tamb mm (in.) Hg
Pump vacuum
K′ factor
Average
40
Method 5 8/2/2017
Date _________________________
Train ID ______________________
Critical orifice ID _______________
Critical orifice K' factor __________
Run No.
Dry gas meter 1 2
Final reading m3 (ft3)
Initial reading m3 (ft3)
Difference, Vm m3 (ft3)
Inlet/outlet temperatures °C (( °deg;F) / /
Initial °C (( °deg;F) / /
Final °C (( °deg;F)
Avg. Temperature, tm min/sec / /
Time, θ min
Orifice man. rdg., ΔH min
Bar. pressure, Pbar mm (in.) H2O
Ambient temperature, tamb mm (in.) Hg
Pump vacuum °C (( °deg;F)
Vm(std) mm (in.) Hg
Vcr(std) m3 (ft3)
DGM cal. factor, Y m3 (ft3)
41