DAS Operations Manual
DAS Operations Manual
Edition 1.1
MREL does not warrant that the product will meet the Customer’s requirements, or that it will operate in the combinations which may be
selected by the Customer.
MREL does not and cannot warrant the performance or results that may be obtained by using the product. Accordingly, the product and its
documentation are sold “as is” without warranty as to their performance, merchantability or fitness for any particular purpose. The Customer
assumes the entire risk as to the results and performance of the product.
MREL is committed to product innovation; accordingly product may undergo specification improvements without notice. Copyright © 2017
MREL Group of Companies Limited. DAS™ Data Acquisition Suite, DAS™ Data Acquisition Suite Logo, HandiTrap II™ VOD Recorder,
HandiTrap II™ VOD Recorder Logo, MicroTrap™ VOD/Data Recorder, MicroTrap™ VOD/DATA Recorder Logo, DataTrap II™ Data/
VOD Recorder, DataTrap II™ Data/VOD Recorder Logo, and MREL Logo are trademarks or registered trademarks of MREL Group of
Companies Limited. Windows® is a registered trademark of Microsoft Corporation.
© Copyright 2017 MREL Group of Companies Limited. This Operations Manual supersedes any earlier editions. All Rights Reserved.
Reproduction or adaptation of any part of this documentation or Software without written permission of the Copyright owner is unlawful.
Publication Histor y
Edition 1.0, March 2016
This version of the Manual
1.1 Introduction 2
Chapter 2: Installation 3
Chapter 1
Introduction
1.1 Introduction
MREL’s DAS™ Data Acquisition Suite has been designed to be utilized with MREL VOD/Data recorders. This software suite will allow the
user to program (where applicable), download and analyze the data from each recorder. The DAS™ Data Acquisition Suite will allow the
user to manage all of the files captured from MREL VOD/Data recorders in one piece of software.
The Standard Edition will allow the user to open each data file from any of the three recorders. The user can perform all of the standard
features that the users have been accustom to. For a VOD channel, the measurements that a user can do are as follows:
For a Scope channel, the measurements that a user can do are as follows:
The Advanced Edition will allow the user to apply filters to the data to either clean the data up for ease of reading/presentation or to
apply a specific filter to allow the data to be compliant with a particular standard.
The Advanced Edition can perform all of the Standard Edition features plus the following to Scope channels:
See Section 8 for more details on the Advanced Edition. The software can be downloaded from http://www.mrel.com/das at any time
and registered online.
Chapter 2
Installation
Internet connection required for registration, upgrades, updates and support. The software can be used offline when not performing any
of the previous tasks.
Double click on the DAS™ Data Acquisition Suite setup package (.msi)
file or right click on the file and choose install. You need to be using
administrative account on the computer to perform the installation.
If setup program has detected that the .NET Framework 4.0 is not
installed, it will show the following message.
The package will uncompress and shows the initial Microsoft .NET
Framework 4 license Agreement dialog. Check Accept check box
and click Install button as shown to the right.
I M P O R TA N T
MREL recommends exiting all programs before
starting installation.
Click the Next button which results in the following User Account
Control (UAC) dialog for Windows 7® or 8®.
Installation process will begin and the image to the right will be
shown.
For DataTrap II™ USB Interface Cable driver installation, check the
driver’s manufacturer’s trust check box and click the Install button
to install the needed drivers as follows.
A program group MREL with the folder DAS will be created in your
program files menu. Also a shortcut named DAS™ Data Acquisition
Suite is created on the desktop as show to the right.
In Windows Vista®, the UAC dialog box will look like the image
shown to the right. Click Allow to proceed
Uninstall will show a progress dialog box similar to the image to the
right.
You can now close Programs and Features dialog and the Control
Panel itself.
This will allow the user to evaluate the software for 30 days. After 30 day time limit, the user will need to either register the software or
uninstall the DAS™ Data Acquisition Suite Software. Registration is free for the Standard Edition.
The user can at any time click on the Help tab and then on Register/Upgrade…. When the user selects to register the software, an
internet connection is required. The user will fill in the required fields of First Name, Last Name, Company Name, City, Country, Email and
Industry. The user will select the register tab and with an internet connection connected to the computer, the software will be registered.
At this point, the lower right corner of the software will display Evaluation Edition, Registered.
An email to the address provided in the registration form will include the Registration Key for the computer. When the email arrives,
the user will click on Help and then Register/Upgrade…. At the bottom of the form is a pull-down menu of Install/Upgrade Type. In the
window, select Standard Edition and a text window will appear for the input of the Registration Key. Copy and paste the key from the
email into the text box and click Upgrade. The lower right of the software will then change to Standard Edition, Registered. At this point,
the software will not expire on the computer.
Until the registration key is applied, the software will stop running 30 days after installation. To Upgrade to the Advanced Edition, refer to
Chapter 8.
Chapter 3
Starting the DAS™ Data
Acquisition Suite
Note: The box to the right is displayed for a short period of time upon starting
the software. The software is checking for any hardware attached to
the computer. The software periodically checks for changes in the
hardware and is not updated live.
Chapter 4
Programming Hardware
Click on the Hardware tab on the left panel of the screen (As
marked in the right image) and the image on the next page will be
displayed.
This screen will display many of the details of the individual unit.
This includes the serial number, calibration due date, if a Memory
Upgrade or Scope Upgrade has been installed. This screen will also
display the total number of tests that the memory has been divided
into and the number of tests remaining.
Also, the area highlighted by the box, is the check box that will
delete all of the saved data in the MicroTrap™ VOD/Data Recorder
when it changes the settings.
CAUTION
IF THERE IS IMPORTANT DATA IN THE UNIT, DO NOT CHECK THIS BOX.
If the box is checked, the total number of tests that the memory is divided into can be changed.
With the box checked, the software will tell the user how many tests will be erased.
With the standard MicroTrap™ VOD/Data Recorder, the VOD radio button can only be selected. Once the items have been selected on
this screen, click on the Next button.
This screen will display many of the details of the individual unit.
This includes the serial number, calibration due date, if a Memory
Upgrade or Scope Upgrade has been installed. This screen will also
display the total number of tests that the memory has been divided
into and the number of tests remaining.
CAUTION
IF THERE IS IMPORTANT DATA IN THE UNIT, DO NOT CHECK THIS BOX.
If the box is checked, the total number of tests that the memory is divided into can be changed.
With the box checked, the software will tell the user how many tests will be erased.
With the Scope Upgrade in the MicroTrap™ VOD/Data Recorder, the user can select any of the 3 different radio buttons. In this section,
the Scope radio button will be selected. Once the items have been selected on this screen, click on the Next button.
This screen will display many of the details of the individual unit.
This includes the serial number, calibration due date, if a Memory
Upgrade or Scope Upgrade has been installed. This screen will also
display the total number of tests that the memory has been divided
into and the number of tests remaining.
CAUTION
IF THERE IS IMPORTANT DATA IN THE UNIT, DO NOT CHECK THIS BOX.
If the box is checked, the total number of tests that the memory is divided into can be changed.
With the box checked, the software will tell the user how many tests will be erased.
With the Scope Upgrade in the MicroTrap™ VOD/Data Recorder, the user can select any of the 3 different radio buttons. In this section,
the VOD & Scope radio button will be selected. Once the items have been selected on this screen, click on the Next button.
This screen will also allow the user to configure the External Trigger.
Refer to the MicroTrap™ VOD/Data Recorder Operations Manual
for definitions of the settings. Once all of the settings have been
selected, the user will need to click on Change Settings.
Click on the Hardware tab on the left panel of the screen and the
following will be displayed.
In the Setup Mode window, there are four (4) options. The options
are: Clear All Tests, Standard Setup, Advanced Setup and Recall
Setup File.
When the user selects Clear All Tests and clicks Next, the following
window will appear.
This operation will delete all of the tests currentlyu saved in the
DataTrap II™ Data/VOD Recorder and will keep the settings.
The user will need to confirm the action of clearing all the test data in the unit by clicking
Clear All Tests button. If the user mis-selected, the Cancel button can be clicked.
When OK has been selected, the user will receive conformation that the data has been deleted. The user will
then need to click the Cancel button to return back to the main menu.
This section describes the details of the Standard Setup for the
DataTrap II™ Data/VOD Recorder. This will be the most popular
programming option by the majority of users.
When the user selects Standard Setup and clicks Next, the
following window will appear.
If the unit was last setup in Advanced Setup and the Standard
Setup is chosen, the
following message
will appear.
If the user wishes to continue in the Standard Setup click Yes, if not, click No or Cancel.
In the next section down labeled Change Number of Tests, if the radio button Keep Stored Tests is selected, the box to the right of
Total Number of Tests will display the number of tests the memory has been divided into, if Clear Stored Tests is selected, a slide bar is
displayed to the right of Total Number of Tests for the user to change the number of tests the total memory is divided into.
The third section is the details of each channel. With the DataTrap II™ Data/VOD Recorder, there are 8 channels for the user to configure.
The check box beside each number will enable (when checked) or disable (when not checked) each channel. The user can independently
change the voltage range of each channel to suit the signal that it is intended for. This is the setup for the standard scope unit.
Note: If the VOD Upgrade is installed in the unit, VOD will be able to be selected in the Input Range. The standard Trigger settings for
the internal Trigger or a VOD Channel is Low Level with the Trigger Level set to 90%.
The next line in the third section displays the status of the External Trigger. The External Trigger can be modified in the lower left corner
of the screen and will be described later in the manual. The next line displays the status of the Internal Trigger which can be modified in
the lower right corner of the screen and will be described later in the manual. The sample speed of each channel can be independently
changed to suit the input signal or the recording time required. In the bottom part of the third section is the display of total time for each
channel followed by the Pre-Trigger time. The Pre-Trigger time can be adjusted by the slider below and to the left labeled Pre-Trigger
Time. It is done in percentage of the total time available for the channel.
To the right of the Pre-Trigger Time slider is the check box for Internal Trigger to set the internal trigger for VOD. Since this unit does not
have the VOD option installed, this check box is not available and has been greyed out. If the VOD Upgrade was installed, the check
box would be available but this check box would only effect the channels set to VOD. Below the check box is the settings for the internal
trigger. There are six (6) choices: Low Level, High Level, Rising Edge, Falling Edge, Rising or Falling Edge and Off with a sider bar to
adjust the level for the trigger. To the left of the Internal Trigger settings is the External Trigger settings. There is six (6) choices: TTL Low,
TTL High, Make, Break, Make or Break and Off. For details of each of the twelve (12) settings, refer to the DataTrap II™ Data/VOD
Recorder Operations Manual.
The bottom left of the screen is a button that will change all of the
settings of Channels 2 through 8 to how Channel 1 has been setup.
This is a quick option for the user to configure Channel 1 and click
this button to automatically setup the following channels.
At any time the user wishes to exit the Standard Setup of the
DataTrap II™ Data/VOD Recorder, the user can select the Close
button in the lower right. If the operator uses multiple setups with
the unit, the user can save the individual setups in a setup file to
be recalled at a later date. To save the file, the user will need to
configure the unit as described above and once complete, the user
will click on the Save to File button. This will open a file structure
window from the local computer as shown to the right.
Once the user has configured the DataTrap II™ Data/VOD Recorder with the desired
settings, Apply Settings must be clicked to upload the settings into the unit. Once the
button has been clicked, the following box will appear.
The user will need to click OK if the settings are desired to be uploaded to the unit or
Cancel if a mistake has been made. Once OK has been selected, the file will upload to the unit and
the following message will be displayed.
Once OK has been selected, the software will return to the main screen.
I M P O R TA N T
This section should only be used by either
experienced DataTrap II™ Data/VOD Recorder users
or veteran data collectors from various different
systems.
With the Advanced Setup radio button selected and the Next
button clicked, the following screen will appear.
In this window, the user can select the groupings of the channels that are to be
triggered together. The user has the ability to have each and every channel trigger
separately. Once the user has selected the required groupings of the channels, click OK. If a change is not required, click Cancel.
To the right of the trigger group selection is the External Trigger. There is six (6) choices: TTL
Low, TTL High, Make, Break, Make or Break and Off. For details of each of the six (6) settings,
refer to the DataTrap II™ Data/VOD Recorder Operations Manual.
Below the trigger group selection is tabs labelled 1-8. These are the individual details and the
configuration abilities for each channel, beside the number is a check box. If this box is not
checked, the channel will be disabled. Within this section, the user can select the sampling
rate, voltage range, including VOD if the VOD Upgrade has been installed, can review and
change the triggering group (by clicking on Edit), review the total amount of time and change
and review the pre-trigger time with the slide bar. This is the setup for the standard scope
unit. If the VOD Upgrade is installed in the unit, the standard Trigger Level for the internal
Trigger are Low Level with a Trigger Level set at 90%. In this section, the user can click on Edit
Memory Allocation. It will display the following window.
The next section refers to the External Trigger. If in the upper section the External Trigger is set to OFF, then in this section the only choice
is OFF and the ON is labelled in red. If a selection in the upper section other than OFF is selected, than ON is in black and the radio
button has the ability of being selected.
When the ON radio button is selected, the
type of trigger that was selected above
is displayed to the right of the button. To
the right of the External Trigger OFF/ON area is a window displaying the current channel and the channels that can trigger the current
channel. This is a direct display of the trigger grouping window and to ensure the user is aware of the channels that are effected by the
decisions above.
I M P O R TA N T
This is the setup for each individual channel.
At any time the user wishes to exit the Advanced Setup of the
DataTrap II™ Data/VOD Recorder, the user can select the Close
button in the lower right. If the operator uses multiple setups with
the unit, the user can save the individual setups in a setup file to
be recalled at a later date. To save the file, the user will need to
configure the unit as described above and once complete, the user
will click on the Save to File button. This will open a file structure
window from the local computer as shown to the right.
Once the user has configured the DataTrap II™ Data/VOD Recorder with the desired
settings, Apply Settings must be clicked to upload the settings into the unit. Once the
button has been clicked, the following box will appear.
The user will need to click OK if the settings are desired to be uploaded to the unit or Cancel if
a mistake has been made. Once OK has been selected, the file will upload to the unit and the
following message will be displayed.
Once OK has been selected, the software will return to the main screen.
This section describes how the user can recall a saved setup file
generated from the previous section.
When the user selects Recall Setup File and clicks Next, the
following window will appear.
The next screen displays the same file structure window displayed
when the file was saved. The user MUST select a file or only the
Back and Cancel buttons are available. When the file has been
selected, the Finish button will become selectable.
This window to the right will be displayed as the file is being checked against the
configuration of the DataTrap II™ Data/VOD Recorder.
If the file is configured for a VOD setting and the unit does not have a VOD Upgrade
installed, the following window will be displayed.
If the Setup File is configured properly for the DataTrap II™ Data/VOD Recorder, the screen of either Standard or Advanced Setup will
be displayed. This will depend on which window the file was created from. The user would then need to click Apply Settings to save the
configuration to the unit.
Chapter 5
Downloading from the
Equipment
In this window is the file structure of the computer. The user will
need to select a directory on the hard drive to where the data is to
be saved. With the location selected, an appropriate file name that
is meaningful to the data will need to be typed into File name. In
the box below is a comments box where some details are captured
including the type of recorder and date with time of download. The
user can add additional notes regarding the test including but not
limited to: location, hole size, hole depth, explosive type, explosive
column length and designed booster location. The selection of the
VOD units should be selected next in order to display the Probe
values in the selected units. The user will need to select the correct
probe that was used in the test from the pull-down menu that was
used during the test. The user should type in the Serial Number of
the unit in the box to the right of the VOD Units. The user can click
Download to start the process or Cancel to exit. When Download
has been selected, the following progress bar will be displayed.
Once complete, the following box will be displayed informing the user that the retrieval has been completed.
The user must click OK.
With the Download All Tests radio button selected, the user will need
to click on Next.
When the Download button is clicked, the following progress bar is displayed.
When the download has completed, the DAS™ Data Acquisition Suite will confirm
the successful finish. The user will need to click on OK.
When the acceptance of the download has been completed, the following screen is displayed.
On the top of the screen is the file structure of the local computer.
On the right side, the user selected name followed by _01 for the
first test and _02 for the second test from memory. Each of the
MicroTrap™ VOD/Data Recorder downloaded tests comprise of two
files. A .mcr file for the configuration of the unit and a .raw is the
data file. If the files are moved from one directory to another, both
files are required to be moved. This example shows the download of
both tests.
Just below the file structure window shows the user which test the
following information is for. In the image to the right, this is the
text box for Test 1 of 2 downloaded in this session. The user can
enter any important and relevant information regarding the test. For
example: Location, hole size, hole depth, explosive type, gauge
type, gauge serial number, signal amplification, gauge location.
In the next section, the user should select the units prior to selecting the Probe to ensure the correct units are displayed in the pull-down
box. To the right of the VOD Units is the display of the Serial Number. Below is the display of the individual channel details. If the VOD
channel is OFF (on units with the Scope Upgrade), the Probe selection pull-down is unavailable. When all of the data has been entered
into Test 1, the user needs to click Accept Test 1. The next test will be displayed and the above process needs to be completed for each
individual test downloaded. The unit can store up to 16 individual tests.
Note: For VOD channels, the correct Probe MUST be selected. This cannot be changed later.
Once all of the data has been entered for all of the downloaded tests, the following window appears. At
any time the Cancel button can be pressed to exit the download process.
This section will detail the Download Selected Tests options of the
MicroTrap™ VOD/Data Recorder download feature.
When the user selects the Download Selected Tests and presses the
Next button, the following screen will appear.
This is the screen that the user will view showing the file structure
of the computer. The user will need to select a directory on the
hard drive to where the data is to be saved. With the directory
selected, an appropriate file name for the group of tests that are to
be download needs to be entered in the File Name text box. Below
is the test number to be downloaded. If all of the tests are checked,
the software will be exactly as if the Download All Tests button was
selected and is described in Section 5.2.1. The user will use this
option if the user desires that each test will have a unique file name.
The user will enter the name and only leave checked the tests that
is desired to have that name. In this example, only Test 1 will be
selected.
When the Download button is clicked, the following progress bar is displayed.
When the acceptance of the download has been completed, the following screen is displayed.
On the right side, the user selected name followed by _01 for the
first test and _02 for the second test from memory. Each of the
MicroTrap™ VOD/Data Recorder downloaded tests comprise of
two files. A .mcr file for the configuration of the unit and a .raw
is the actual data. If the files are moved from one directory to
another, both files are required to be moved. This example shows
the download of 1 test.
Just below the file structure window shows the user which test the
following information is for. In the image above, this is the text
box for Test 1 of 2. The user can enter any important and relevant
information regarding the test.
In the next section, the user should select the units prior to selecting
the Probe to ensure the correct units are displayed in the pull-down
box. To the right of the VOD Units is the display of the Serial Number. Below is the display of the individual channel details. If the VOD
channel is OFF (on units with the Scope Upgrade), the Probe selection pull-down is unavailable. When all of the data has been entered
into Test 1, the user needs to click Accept Test 1. If more than one test was selected for download, each test will cycle with the same
screen.
Note: For VOD channels, the correct Probe MUST be selected. This cannot be changed later.
If only one test was selected, as in this example, the user will click on Accept Test 1 and the following
window will be displayed. At any time the Cancel button can be pressed to exit the download process.
With the Download All Tests radio button selected, the user will
need to click on Next.
This is the screen that the user will view showing the file structure of
the computer. The user will need to select a directory on the hard
drive to where the data is to be saved. With the directory selected,
an appropriate file name for the group of tests that are to be
download needs to be entered in the File Name text box.
When the Download button is clicked, the following progress bar is displayed.
When the download has completed, the DAS™ Data Acquisition Suite will confirm the successful finish. The
user will need to click on OK.
On the top of the screen displays the file path on the local
computer where the file has been saved.
In the File Comment section, the user can enter any important and
relevant information regarding the test. For example: Location, hole
size, hole depth, explosive type, gauge type, gauge serial number,
signal amplification, gauge location.
The next button down is the Add New Sensor Device or Formula. Refer to Section 5.4 that describes this function separately.
In the table is the configuration of the DataTrap II™ Data/VOD Recorder referencing the channel number and the voltage settings (or
VOD) during the recording. In the Probe Values/Device Sensors, the user would use this menu to pre-select the calibration factors of
sensors that have been entered into the system using the Add New Sensor Device or Formula or VOD Probe. Otherwise, if No Device
Selected is inputted the data will be displayed in Volts.
Note: For VOD channels, the correct Probe MUST be selected. This cannot be changed later.
The Orientation column is designed for the user to input a specific orientation of the particular channel. This is generally used in
accelerometers or strain gauges. This information is kept for the purpose of complete data.
If the DataTrap II™ Data/VOD Recorder has the Strain Upgrade installed and the channel is connected to
a strain gauge then the user must check the box of the particular channel that is a strain channel. This will
allow the user to then enter the Pre-Amp Resistance and the Strain Gauge Factor.
At any time the Cancel button can be pressed to exit the download process.
When the user selects the Download Selected Tests and presses the
Next button, the following screen will appear.
This is the screen that the user will view showing the file structure
of the computer. The user will need to select a directory on the
hard drive to where the data is to be saved. With the directory
selected, an appropriate file name for the group of tests that are to
be download needs to be entered in the File Name text box. Below
is the test number to be downloaded. If all of the tests are checked,
the software will be exactly as if the Download All Tests button was
selected and is described in Section 5.2.1. The user will use this
option if the user desires that each test will have a unique file name.
The user will enter the name and only leave checked the tests that
is desired to have that name. In this example, only Test 1 will be
selected.
When the Download button is clicked, the following progress bar is displayed.
When the download has completed, the DAS™ Data Acquisition Suite will confirm the successful finish. The
user will need to click on OK.
On the top of the screen displays the file path on the local
computer where the file has been saved.
In the File Comment section, the user can enter any important and
relevant information regarding the test.
The next button down is the Add New Sensor Device or Formula. Refer to Section 5.4 that describes this function separately.
In the table is the configuration of the DataTrap II™ Data/VOD Recorder referencing the channel number and the voltage settings (or
VOD) during the recording. In the Probe Values/Device Sensors, the user would use this menu to pre-select the calibration factors of
sensors that have been entered into the system using the Add New Sensor Device or Formula or VOD Probe. Otherwise, if No Device
Selected is inputted the data will be displayed in Volts.
Note: For VOD channels, the correct Probe MUST be selected. This cannot be changed later.
The Orientation column is designed for the user to input a specific orientation of the particular channel. This is generally used in
accelerometers or strain gauges. If the DataTrap II™ Data/VOD Recorder has the Strain Upgrade installed and the channel is connected
to a strain gauge then the user must check the box of the particular channel that is a strain channel. This will allow the user to then enter
the Pre-Amp Resistance and the Strain Gauge Factor. The user will need to enter these values to allow
the DAS™ Data Acquisition Suite software to convert the recorded data into the proper microstrain (µϵ)
units. For hardware configurations for Strain Channel setup, refer to the DataTrap II™ Data/VOD Recorder
Operations Manual.
At any time the Cancel button can be pressed to exit the download process.
A new sensor, device or formula can be added into the database from the main menu under the Hardware
tab.
The names that are typed into the Quantity and Unit text boxes are the label for the Y-axis of the graph.
Various examples for Quantity text box are Distance, Length, Pressure, Acceleration, Temperature, Force
and so on. Various examples for the Units text box are metres, feet, kPa, psi, g’s, m/s^2, Celsius, Fahrenheit,
Kelvin, Rankine and so on. This is displayed in the Y-Axis box at the bottom. There are two options for
adding information for quick reference: Add Linear Sensor and Add Sensor Formula. The formula or
equation can be applied to the data in the Analyze section of the software under the tab filters and
formulas.
The user will use Add Linear Sensor when a sensor is used with a calibration value. This will allow a quick selection of the intended sensor
with the proper calibration to be applied to the collected data.
When the Add Linear Sensor button is clicked the following window is displayed.
The definition of the Y-Axis is shown at the top and the user is required to enter a
Device Name. This is a definition of the sensor for the user to quickly select from a list.
In this example, the serial number of the sensor has been selected. From the calibration
sheet from the manufacturer of the sensor will be the calibration value of the sensor.
Enter the numerical value of the calibration and select the correct units to the right of
the calibration. The Frequency Limits is a position for the user to enter information of
the sensor. This information can be referred to at a later date without looking up the
frequency limits on the technical specifications of the sensor. The user will then select
Save to add the device to the list or Back to return to the previous screen or Cancel to stop
the process. Once Save has been clicked, the following window will be displayed.
The user will then select Save to add the device to the list or Back to return to the previous
screen or Cancel to stop the process.
Once Save has been clicked, the following window will be displayed.
The user will select this button when straight math is need to be applied to the collected
data. When Add Sensor Formula button is clicked, the following window is displayed.
The user can enter any mathematical equation into the Y= text box. To the right is a
window describing all of the different mathematical expressions that can be used to
modify the voltage that has been recorded.
With the formula entered into the text box, the user can click Check Formula and confirm
that the format of the equation is correct by viewing a check mark in the check box
labelled Formula Validated.
At any time, the user can select Back to return to the previous screen and Cancel to exit.
Chapter 6
Opening a Data File
6.2 Browse
The Browse feature is the typical way to use the file browser to find
a file. By clicking the Browse radio button and selecting Next, the
following screen will be displayed.
This screen will allow the user to navigate through the file structure
of the computer to find the desired file. From the previous page,
only the selected file types will be displayed. The user can click
on Go To Home to return to the default directory defined in the
software. When the desired file is located, select the file by left
clicking on the file name once and then clicking Open. To stop
the process, click Cancel. When Open has been clicked, the main
window will return with the Analyze tab open. To analyze the data,
refer to the Analyze section.
The user will select Recently Opened Files radio button from the
Open Data File tab and select Next and the following window will
appear.
The DAS™ Data Acquisition Suite will display eight (8) previously
opened files. This includes all of the file types selected in the File
Type window in the previous section. The user will just click on
the blue link and the file will open and the file will show up in the
Analyze tab. If no files wish to be opened from the list, the user can
click Close.
The DAS™ Data Acquisition Suite will display eight (8) previously
opened files. This includes all of the file types selected in the File
Type window in the previous section. The user will just click on
the blue link and the file will open and the file will show up in the
Analyze tab. If no files wish to be opened from the list, the user can
click Close.
6.5 Calendar
This section is used to open a file if the download date is known.
With the Calendar radio button selected and selecting Next, the
following window will be displayed.
This option will display a calendar for the user to select the date
that the file was downloaded. A blue box is placed around the
current date and the date that is in bold has a file associated with
that date. In the example above, February 21, 2014 has a DataTrap
II™ Data/VOD Recorder file associated the date. In the box below
is the file location of the data with the file name. All of the files
downloaded on the selected date will show in the bottom window.
The user will select the file from the bottom box and click Open. The
file will open in the Analyze tab. If no files are to be opened, click
Cancel.
In the next example shown to the right, the file name of DT2010Oct004_02.dt2 shows 8 channels and the
display All Graphs. The User can select the All Graphs will overlay on the same screen all of the channels
the user wants to display for comparison. Also in this example, channel 1 is a VOD channel, channel 2 is
a Strain channel and channels 3 through 8 are scope. This example has no saved graphs on any of the
channels. When the user clicks on the channel number, the software will display a summary of the settings
and information typed in when the file was downloaded.
Chapter 7
Data Analysis
This section will detail the analysis of the data collected from the recorders. It will show how the software will work and not the correct or
incorrect way to analyze certain types of data. The data filtering portion of the software can be found in Section 8. Below is the view of
the software with a data file open. Across the top of the screen is a ribbon with six (6) different tabs.
In the Home tab are many of the general functions that are used in graphing. It includes the following functions:
• Full – Allows a single button click to zoom out to the maximum extents of the graph
Click on once for maximum zoomed out area
• Original – Allows a one button click to zoom out to the maximum extents of the saved graph
Click on once for the saved graph zoomed area
• Save As – Allows a graph to be saved with the zoomed area desired. Used with the Original zoom button
With the viewable area as desired, click on Save As and a text box will pop up to allow a name as desired by the user.
• Delete – Allows the generated graphs to be deleted. Will not delete the original graphs.
With a generated graph selected, the user can click this button and a conformation box will appear to confirm the deletion of the
graph.
• Point Cursor – Selects the X,Y coordinate of a data point on the graph.
Click on the button and follow the on screen directions. The Point Cursor can be moved with the mouse or the left and right arrow
keys.
To select an individual point, the user may need to zoom in closer to select the desired point. The user can have multiple Point
Cursors on the screen.
To remove, right click on the Point Cursor and select Delete. Also with the right click on the Point Cursor and selecting Modify, the
user can adjust all of the individual properties on the Point Cursor.
The Show/Hide Point Cursor will allow the user to copy the individual coordinates to the clipboard to be pasted into a different
application.
• Show/Hide - Will allow the user to copy the individual coordinates to the clipboard to be pasted into a different application.
Click on the button. Can also be found in the right click menu of the individual Point Cursor
• Annotation – Will allow user to enter text onto the graph. Follow the instructions on screen.
• Paste – Will allow the item that has been Cut or Copied to be pasted.
This button will allow the user to Paste the item that has been placed on the computer’s clipboard.
7.2 View
In the View tab are features that effect how the data is displayed.
• Plotting Styles – Allows the user to change the way the graph is drawn.
The default is Line. This will connect the data points with a line. Point will only display the data
points while Point + Line will display both.
• Line Thickness – Allows the user to change the line thickness of the graph.
The user can either click and drag the slider bar or use the + and – buttons to change the thickness
of the drawn line.
• Point Colour – Allows the user to change the colour of the data points.
The Point Colour button will change the colour of the data points while the pull down arrow will
allow the user to select the colour from a pallet of colours.
• Line Colour – Will allow the user to change the colour of the data line.
The Line Colour button will change the colour of the data line while the pull down arrow will allow
the user to select the colour from a pallet of colours.
A single click on the Graph Tree button will change back and
forth between the two screens.
• Grid Lines Major – By selecting the check box will display the major grid lines.
It can be turned on and off by the check box.
• Grid Lines Minor – By selecting the check box will display the minor grid lines.
It can be turned on and off by the check box.
• Start/Done – Allows the start of the VOD measurement. The button will change to Done after the process has been started.
Follow the on-screen directions and click and drag the vertical bars to the desired location. See the Quick Start Guide for the use of
the button.
• Linear Regression – Allows the user to use Simple Linear Regression to produce a line of best fit of the points between the two
end points.
The definition of Simple Linear Regression is a fit of a straight line through the set of n points in such a way that makes the sum of
squared residuals of the model (that is, vertical distances between the points of the data set and the fitted line) as small as possible.
• 2-Point – Allows the user to measure VOD with using only the start and end point.
The VOD is calculated by the following equation: (y2-y1)/(x2-x1), where (x1,y1) is the first point and (x2,y2) is the second point. This
is used when the data is ‘noisy’ and the points in the middle need to be ignored. As long as the line matches the straight edge of the
data, then the value obtained from the calculation is valid.
• Data Section
In this area is displayed the values when a VOD measurement has started. The VOD value will be displayed and will change if the
user selects Linear Regression or 2-Point.
The user will need to decide which type of VOD measurement that is desired: Linear Regression or 2-Point.
Please refer to the descriptions below to which type of VOD measurement is desired.
With the desired check box selected and the graph zoomed into the relevant
area, click Start. The following screen will be displayed.
Click OK to continue.
Reconfirm that the desired method has been selected and the Data
will display a quick reference to the two points and the VOD prior
to pressing Done.
Once Done has been pressed and the line is displayed, this is the
only time that the user can compare the VOD line to the data.
If the VOD line does not match the data, it is not a valid VOD
measurement.
As shown in the graph to the right, the selected line is not the
best fit for the data. The user can place as many different VOD
measurements as desired.
The line is shown above the data at the beginning, below in the
middle and above the measurement at the end. The image to the
right is a better measurement.
From the screen to the right, the two VOD lines have been fit to the
data where each line fits the data in the section of measurement.
This shows a change in velocity from the bottom of the hole to the
top.
• Start/Done – Allows the user to measure the amount of time (delta X) and the amount of units (m, V or engineering units)
(delta Y) between the two selected points.
The user will select start and will follow the on-screen instruction.
The screen to the right shows the two vertical bars and the two
(X,Y) points. In the top bar in the Data section, the values of the
two points and the difference is also displayed.
• Delta Time (ms) – This check box defines if the value is to be displayed. If the box is unchecked, the delta value will disappear
in the Data section. Also, if the box is unchecked, the value will not be displayed when Done is selected.
• Delta Distance (units of graph – m, V or engineering units) – This check box defines if the value is to be displayed. If the box
is unchecked, the delta value will disappear in the Data section. Also, if the box is unchecked, the value will not be displayed when
Done is selected.
• Apply Offset – Allows the user to shift all of the data on the vertical
axis of a single channel. Does not apply to VOD data. This cannot be
applied to the original data file. It only can be used on saved files. If it is
attempted to be performed on an original graph, the following pop up is
displayed.
Click on the button and follow the on-screen directions. On the pop up to
the right, click OK and then select the data point at the user wants to be a
zero value on the Y-Axis. When the desired point has been selected, the
following pop up is displayed.
• Move Data Pt – Allows the user to move a single point in the vertical
axis. This operation can only be performed on Saved Graphs and not on
original graphs.
This feature will allow the user to move a data point that has an error in
the collection. The user will select Move Data Pt and the following pop up
is displayed. Select OK to continue. Select the point of interest to move
and then the user will be able to move in the vertical axis. The user will
click again to place the point.
This section describes the use of Formulas. For discussion on Filters, which is only available in the Advanced Edition, refer to the
Advanced Edition Section.
The image above is the tab when a VOD data channel has been selected. There is no formula that can be applied to the graph.
In the image below is the tab when a scope data channel has been selected.
When the user selects Apply Formula, the following window will appear.
In the Current Formula text box is the current formula being applied to the
data.
In this example, Y=V. The user will enter the equation into this text box using V
as the current value of the data.
When the equation has been entered, the user can select the Units and
Quantities tab. The user will assign the Y Units (psi, g’s, etc) and the
Quantities (Pressure, Acceleration, etc). These units can be selected from
the pull down window (if already in the system) or typed in and saved by
clicking Add Formula & Units To Lists. As stated previously, the formula cannot
be applied to an original graph. If this work has been done on an original
graph, the Apply button will be greyed out. The user must select Apply To
New Graph check box and then the Apply button will appear. The user can
always select Cancel.
7.6 Help
• Support – Opens an email from your email client to send an email to [email protected]
• About – Opens the window showing the version number of the software and the contact details for MREL Group of Companies
Limited including website, email, address and phone and fax numbers.
Chapter 8
Advanced Edition
The request is sent to MREL Group of Companies Limited and if the order has
been paid, the Upgrade Key will be sent to the email address listed in the
Registration Box. With an order paid in full, the key may take up to 1 business
day to process the request and longer if the order has not been paid.
When the user clicks on the button, the following screen will
appear.
• Bessel
• Butterworth
• Chebyshev
• Elliptic
• Miscellaneous
• Median
• Moving Average
• Savitzky-Golay
• WindowedFir
Along the top of the window are 5 features that include: Zoom In, Zoom Out, Zoom Out to Original Extents, Pan, Filtered Data Graph
Colour and a text box for a number for the thickness of the line of the filtered data.
The Plot Settings will adjust the settings for both the top and bottom graph. The Plot Style defines the plotting method of the graphs. The
choices are: Line, Point or Point+Line. The check box below Show Original Range will zoom the graph to the maximum extents. If the
graph has been zoomed in, this is a quick way to see the overall data and uncheck to return to zoomed in view. The next check box
Original & Filtered will place both graphs on the same axis to allow the user to view the data to see the differences of the effect of the
filter.
Bessel filter is a type of linear filter with a maximally flat group delay (maximally linear phase response). Bessel filters are often used
in audio crossover systems. Analog Bessel filters are characterized by almost constant group delay across the entire passband, thus
preserving the wave shape of filtered signals in the passband.
The options for the Bessel filter are: Lowpass, Highpass, Bandpass and Bandstop
With the type of Bessel filter selected, the user will have the following options for Lowpass and Highpass.
• Filter Order
• Cutoff Frequency (Hz)
With the options selected by the user, the user needs to select Apply Filters to update the bottom graph and if the desired effect has been
generated, the Apply and Close button can be clicked. The user can apply different graphs and arrange the order, as the order of the
filters in the Current Added Filters is the order that they are applied to the data. The Cancel button will exit the Apply Filters window.
8.2.2 Butterworth
The Butterworth filter is a type of signal processing filter designed to have as flat a frequency response as possible in the passband. It is
also referred to as a maximally flat magnitude filter.
The frequency response of the Butterworth filter is maximally flat (i.e. has no ripples) in the passband and rolls off towards zero in the
stopband. When viewed on a logarithmic Bode plot the response slopes off linearly towards negative infinity. A first-order filter’s response
rolls off at −6 dB per octave (−20 dB per decade) (all first-order lowpass filters have the same normalized frequency response). A
second-order filter decreases at −12 dB per octave, a third-order at −18 dB and so on. Butterworth filters have a monotonically changing
magnitude function with ω, unlike other filter types that have non-monotonic ripple in the passband and/or the stopband.
The options for the Butterworth filter are: Lowpass, Highpass, Bandpass and Bandstop
With the type of Butterworth filter selected, the user will have the following options for Lowpass and Highpass.
• Filter Order
• Cutoff Frequency (Hz)
With the options selected by the user, the user needs to select Apply Filters to update the bottom graph and if the desired effect has been
generated, the Apply and Close button can be clicked. The user can apply different graphs and arrange the order, as the order of the
filters in the Current Added Filters is the order that they are applied to the data. The Cancel button will exit the Apply Filters window.
Chebyshev filters are analog or digital filters having a steeper roll-off and more passband ripple than Butterworth filters. Chebyshev filters
have the property that they minimize the error between the idealized and the actual filter characteristic over the range of the filter, but
with ripples in the passband.
The options for the Chebyshev filter are: Lowpass, Highpass, Bandpass and Bandstop
With the type of Chebyshev filter selected, the user will have the following options for Lowpass and Highpass.
• Filter Order
• Cutoff Frequency (Hz)
• Stopband Ripple (dB)
With the options selected by the user, the user needs to select Apply Filters to update the bottom graph and if the desired effect has been
generated, the Apply and Close button can be clicked. The user can apply different graphs and arrange the order, as the order of the
filters in the Current Added Filters is the order that they are applied to the data. The Cancel button will exit the Apply Filters window.
8.2.4 Elliptic
An Elliptic filter is a signal processing filter with equalized ripple (equiripple) behavior in both the passband and the stopband. The
amount of ripple in each band is independently adjustable, and no other filter of equal order can have a faster transition in gain between
the passband and the stopband, for the given values of ripple (whether the ripple is equalized or not). Alternatively, one may give up the
ability to independently adjust the passband and stopband ripple, and instead design a filter which is maximally insensitive to component
variations.
With the type of Elliptic filter selected, the user will have the following options for Lowpass and Highpass.
• Filter Order
• Cutoff Frequency (Hz)
• Stopband Ripple (dB)
• Stopband Attenuation (dB)
The Formula filter applies a formula to create the filtered plot. The unfiltered plot is not modified. The quantity/units remain the same.
Quick examples of formula filter is inversion (-1.0*V), Signal Offsetting (V + 2.58) or other possible formula equations. This is in contrast
to Apply Formula functionality in the main screen where formula is applied to unfiltered plot along with any quantity/units changes. Both
Formula filter and Apply Formula are applicable to Scope type of graphs only.
8.2.6 Median
The Median filter is a nonlinear digital filtering technique, often used to remove noise. The only parameter is the number of points which
needs to be an odd number between 1 (no change to data) and 99.
To demonstrate, using a window size of three with one entry immediately preceding and following each entry, a Median filter will be
applied to the following simple 1D signal with a window set of 3:
x = [2 80 6 3]
i.e. y = [2 6 6 3].
Moving Average is using a given series of numbers and a fixed subset size, the first element of the Moving Average is obtained by taking
the average of the initial fixed subset of the number series. Then the subset is modified by “shifting forward”; that is, excluding the first
number of the series and including the next number following the original subset in the series. This creates a new subset of numbers,
which is averaged. This process is repeated over the entire data series. The plot line connecting all the (fixed) averages is the moving
average. A Moving Average is a set of numbers, each of which is the average of the corresponding subset of a larger set of datum
points. A Moving Average may also use unequal weights for each datum value in the subset to emphasize particular values in the subset.
The only parameter is the number of points which needs to be an odd number between 1 (no change to data) and 99.
8.2.8 Savitzky-Golay
A Savitzky–Golay filter is a digital filter that can be applied to a set of digital data points for the purpose of smoothing the data, that is,
to increase the signal-to-noise ratio without greatly distorting the signal. This is achieved, in a process known as convolution, by fitting
successive sub-sets of adjacent data points with a low-degree polynomial by the method of linear least squares. Using this unweighted
smoothing filter, there are two options to be modified by the user. The Polynomial Order and the Number of Side Points. The number of
side points MUST be greater than or equal to the order of the polynomial.
A window function is a mathematical function that is zero-valued outside of some chosen interval. When another function or waveform/
Within the filter parameters, there are different types of Smoothing Windows that can be used. Below is a list of the available windows.
For more details on the individual windows, please refer to the following website: http://en.wikipedia.org/wiki/Window_function
1. Rectangular
2. Hanning
3. Hamming
4. BlackmanHarris
5. ExactBlackman
6. Blackman
7. FlatTop
8. BlackmanHarris4Term
9. BlackmanHarris7Term
10. LowSidelobe
11. BlackmanNuttall
12. Triangle
13. Kaiser
14. DolphChebyshev
15. Gaussian
For Lowpass and Highpass of the WindowedFir, the following Filter Parameters are available for the above windows 1 – 12.
• Number of Coefficients
• Cutoff Frequency (Hz)
• With a Smoothing Window for 1 – 12.
For Bandpass and Bandstop of the WindowedFir, the following Filter Parameters are available for the above windows 1-12.
• Number of Coefficients
• Lower Cutoff (Hz)
• Upper Cutoff (Hz)
• With a Smoothing Window for 1 – 12.
For Lowpass and Highpass of the WindowedFir, the following Filter Parameters are available for the above windows 13 – 15.
• Number of Coefficients
• Cutoff Frequency (Hz)
• Window Parameter
• With a Smoothing Window for 13 – 15.
For Bandpass and Bandstop of the WindowedFir, the following Filter Parameters are available for the above windows 13 – 15.
• Number of Coefficients
• Lower Cutoff (Hz)
• Upper Cutoff (Hz)
• Window Parameter
• With a Smoothing Window for 13 – 15.
Chapter 9
Contacting MREL for
Technical Support
Email: [email protected]
• Myfile.dt2
• Myfile.d2d
• Any related files with the same name such as Myfile.xml and Myfile.zip if applicable
• Myfile.mcr
• Myfile.raw or myfile.cmp
• Any related files with the same name such as Myfile.xml and Myfile.zip if applicable
• Myfile.ht2
• Myfile.raw
• Any related files with the same name such as Myfile.xml and Myfile.zip if applicable