EZ Read 2000 Service Manual v1
EZ Read 2000 Service Manual v1
Biochrom Ltd, 1020 Cambourne Business Park, Cambourne, Cambridge, CB23 6DW, England
Telephone: +44 (0)1223 423723 Fax: +44 (0)1223 420164
E-mail: [email protected] Support: [email protected]
Contents
2.0 INTRODUCTION 7
2.1 GENERAL DESCRIPTION 7
2.2 TECHNICAL SPECIFICATIONS 7
4.0 ELECTRONICS 10
4.1 ELECTRONIC BOARD LAYOUT 10
4.2 POWER SUPPLY 11
4.3 BASIC CPU BOARD OPERATION 11
4.4 LIGHT CONTROL BOARD DIAGRAM 12
4.5 PREAMPLIFIER BOARD 12
4.6 INTERFACE BOARD 12
5.0 ADJUSTMENTS 13
5.1 REPLACING THE LAMP 13
5.2 LAMP TRIM PROCEDURE 15
5.3 LAMP VOLTAGE CHECK 18
5.4 WAVELENGTH CALIBRATION CHECK 18
5.5 WAVELENGTH CALIBRATION SETTING 19
5.6 MONOCHROMATOR REPLACEMENT. 20
5.7 TRANSPORT CALIBRATION 22
5.8 POSITIONING SETUP 23
5.9 ADC ADJUSTMENT 30
5.10 TOP COVER ADJUSTMENT 34
Changes or modifications
Any changes or modifications made to the instrument could void the user’s authority to
operate the instrument.
If the instrument is operated in a manner not specified, then the protection provided by the
equipment may be impaired and the instrument warranty withdrawn.
Consumption 65VA
Power Requirements
100 to 240 VAC 50/60Hz 65VA
A declaration of decontamination form is given in Appendix 1 (at the end of this document). This
should be completed and signed by the person who decontaminates the equipment.
1. Switch off the reader and disconnect it from the mains power supply and the PC.
2. Carefully wipe off the entire reader with lint-free tissues that have been moistened in a protein
degrading mild detergent or a saline solution.
3. Carefully wipe off the entire reader with non-lint tissues that have been moistened in a 70%
ethanol or a 0.5% bleach solution.
4. Soak non-lint tissues in a 70% ethanol or a 0.5% bleach solution and place onto the plate
transport mechanism and let it soak for ± 30 minutes.
5. If a bleach solution has been used, carefully wipe off the entire reader with non-lint tissues that
have been soaked in water.
Please Note: Before the reader can be returned to base for service, it must be disinfected and a
Declaration of Decontamination Certificate must be completed. This document can be downloaded
from the Biochrom website: http://www.biochrom.co.uk/content/1/65/returns.html
The instrumentation will not be accepted for servicing or return until the above form is completed
fully. Instrumentation that has not been cleaned sufficiently or decontaminated may be subject to
additional charge.
The optical system consists of the lamp, the condenser lenses, the beam splitter, wavelength reference
LED, the monochromator, the measurement and reference channel optic fibres and focusing lenses
below and above the microplate and silicon photodetectors. The light energy from the Lamp is
condensed by the Condensing lens and passes through the Beam Splitter, entrance slit and onto the inlet
mirror. The mirror directs the light to the Collimating mirror which directs the light onto the Grating
where the Wavelength is selected; it is then directed back onto the Collimating mirror then the Exit
mirror onto the Measurement Fibre Optic. From here it passes through a Focusing lens, an aperture,
through the microplate well, another focusing lens and onto the Measuring Photo diode. The energy
from the Reference diode is reflected by the Beam Splitter and follows the same course as the Lamp
energy until it reaches the Reference Fibre Optic where it travels to the Reference Photo Diode.
RS232 / USB
Board
Power
Supply
Filtered Power
Input Socket
with ON / OFF
Lamp Control
switch and Fuse
Board
Holder
Interface
CPU Board Board
The power supply is a switched mode supply which supplies +5 and +24 volts
DC to the instrument with its own internal fuse.
Power Supply
Lamp Control
Board
The CPU Board is piggybacked on to the interface board and uses a 4MHz 8-bit High Performance PLCC
Microcontroller with a 512 byte on board EEPROM and 1024 bytes of on board RAM. This device and the
Flash memory device which contains the operating Firmware, control the operation of the reader.
Screws
4. Place the unit on its left hand side and remove the 5 Allen head screws. Unplug all the
cable attached to the top plate from the base plate noting their positions.
6. Remove the lamp power socket and push the lamp in the direction indicated. Insert
the lamp in the reverse procedure with the indentation in the lamp housing facing in
the direction indicated. The lamp must “click” into position.
7. Reassemble the unit and carry out the “Lamp Trim Procedure” below.
Warning: Do not touch the glass envelope of the lamp with your bare fingers, use lint free cloth
or optical lens paper.
Lamp
Indentation
1. UVMTest.exe -> a tool that checks the basic functions and generates a text file
"UVM.txt".
2. Start the program and a text file will be generated. The file should look like the one below.
This can be emailed to: [email protected] for evaluation.
3. U340Srv.exe -> a useful service tool. It’s easy to use and you can do calibration of the
transport and light system.
4. Start U340Srv.exe
5. Choose "Service" and then "Lamp Trim"
A value of
“0“indicates the
Live lamp is off
value
shown
here
6. The window above appears. In the upper left corner you'll find the "Diode Values" and
next the "Lamp- Energy-Level". (The Lamp-Energy-Level must never be higher than 200! If
it is higher you have a problem with the "Light Control Unit!) The Diode value -> in the
picture above, “243” - is the “live” counts of the ADC (the lamp is off -> look at the “0” in
the "Lamp” row, so the counts are low!)
7. Remove the upper case of the instrument. Watch the diode value!
Due to stray light the value will change to 16383, which indicates an "Overflow" of the
ADC!
9. Before we start with the alignment you will have to cover the optical sensor unit with a
black cloth and make sure that you have access to the variable resistor (it is sealed).
Ensure the diode value is as low as possible! -> around 300
10. Select the first wavelength -> 800nm; push “Bright F3” and then “Lamp F6”. The reference
LED must be “OFF”. Watch the "Current Diode Value". It must not be higher than 15300 (±
100) -> valid range: 9500 – 15300. If the value is above 15300 you'll have an overflow.
(You will find the hole on the left hand side in which is the “Bright Value” adjustment.) In
that case turn the variable resistor counter-clockwise and watch the diode value. Turn the
resistor until the value is in range.
11. Now select the 340nm wavelength / filter. The reading must be more than 9500 without
changing any settings.
Tip: If you haven't any cloth available to cover the optical sensor unit, I would recommend the
following: Turn the variable resistor 5 turns counter-clockwise and replace the upper case. Have
a look at the diode values! Repeat this procedure until the values are lower than 15300.
******************************************************
Open the UVM340 Service Program and click on: Service, Lamp Trim, F6 (switch on lamp) and
select the 340nm wavelength. Using a test meter, connect the probes to the lamp power socket.
The reading must be below 11.8vDC.
1. From the main screen select “Service” then “Holmium Filter” and the screen below will
be displayed.
ADC values
638nm
419nm
537nm
361nm
447nm
Wavelength
9. Undo the two bolts on the opposite side of the assembly that the monochromator is
located and withdraw the unit and the spacer.
10. Undo the retaining bolts on the gear wheel attached to the monochromator and
remove it.
11. With a blade or other suitable instrument, gently slide out the mask from the
monochromator housing.
Installing a new Monochromator
1. Attach the gear wheel to the new monochromator.
2. Slide the mask into the large hole in the side of the new monochromator. Fig.1
Fig.1 Fig.2
3. Ensure that you have placed the calibration certificate for the new monochromator in the
instrument file.
4. Insert monochromator back into the chassis and replace screws loosely.
5. Insert the alignment tool into the fibre optic chassis housing and up into the
monochromator. (See Fig.2)
6. Tighten the screws holding the monochromator to the chassis.
7. Remove the alignment tool.
8. Re-insert the fibre optic cable into the monochromator; push it fully up, then withdraw it
0.5 – 1mm before tightening the retaining bolt.
9. Re-insert the other two ends of the fibre optic loom into their sockets ensuring that they
do not cross over each other, Fig.3.
10. Re-attach the belt onto the gear wheel on the monochromator and the gear wheel on the
motor, Fig.4.
Fig.3 Fig.4
This calibrates the Plate transport (X and Y) and the Monochromator transport. Open the UVM
340 Service Program and click on: Service, Transport and F6, calibrate Y-Transport first then click
on F5 and calibrate X-Transport. Then click on F7 and calibrate the Monochromator transport.
You will require the use of our internal software called UVMProd.exe; this software is
recommended for use by trained Asys service staff only.
Tools required:
UVMProd.exe Setup Software
Asys QC Test Plate
2mm Hex Driver
2.5mm Hex Driver (if parallel adjustment required)
3mm Hex Driver
Small piece of card
Small piece of tracing (or similar) paper
1. Place the instrument sideways so that the front door is facing your left.
2. Undo the six 2mm bolts holding the top cover to the instrument
3. Remove the cover from the rear of the instrument as the LED wire is attached to the front
internally.
4. Place the cover on its front, to the left of the main assembly. Take care not to stretch the
LED wire.
5. Undo and remove the 3mm bolt on the upper lens block assembly.
6. Whilst leaving the diode board connected, carefully remove the upper lens block
assembly and place behind the instrument facing upwards as shown.
10. Select ‘HOME’ from the X-Transp. Window first, then select ‘HOME’ from the Y-Transp.
This ensures that the plate carrier moves to the home positions.
11. Now enter these default positions (1080/868); always start with X before Y and press the
return key on your keyboard after inputting the position:
This will send the plate carrier to where it believes hole A1 of the Asys Test plate is
located.
13. From UVMProd.exe, select Service > Lamp Trim, then select F6 button to switch on the
lamp. ‘Lamp is ON’ should appear.
15. Place some tracing paper (or similar thin paper) over hole A1 of the plate to see where
the light beam is in relation to the hole.
Increase Y values
to move plate this
way Y axis
Decrease Y values
to move plate this
way
X axis
In this example:-
Decrease X
Decrease Y
In this example:-
Increase X
Decrease Y
20. Now select Transport > Define X/Y as > A/1 of Plate 8*12. If you have done this
correctly, the screen above in step 19 should ‘blink/flash’ once.
21. To verify the parameters have been stored, click Service > List Setup. The bottom of the
setup page shows the stored plate parameters.
a) Now that you have obtained the correct position of A/1 of the Asys Test plate, we need to
check the alignment of the plate carrier.
b) From UVMProd.exe, select Service > Transport
c) Select ‘HOME’ from the X-Transp. Window first, then select ‘HOME’ from the Y-Transp.
This ensures that the plate carrier moves to the home positions.
d) Calculate the position of A/12 by using the normal X-position of A/1 – 990. For example,
normal X = 1080, 1080 – 990 = 90. 90 = A/12
e) Type the new A/12 position into the X-Transp. text box and press the return key. Now
enter the normal Y-Transp. Figure into the Y.Transp. text box and press the return key.
Replace upper lens block and fix with 3mm bolt removed in step 5
Re-fit instrument cover and fix with six 2mm bolts removed in step 2
3. The diode reading should be approx. 200 counts; it is likely to vary by approx 10 counts as
it is very sensitive. If the diode reading is not correct, remove the instrument top cover
and adjust the potentiometer on the edge of the AD converter board:
Please note this potentiometer is very sensitive, only make small adjustments. Replace
the top cover, and then re-check the diode reading with the UVMSrvc.exe application
until correct.
3. Ensure no plate exists on the plate carrier then press OK to begin a wavelength scan. The
diode counts read for all wavelengths should be between 10,000 – 15,500, for example:
4. If at any point the counts read out of the range specified in step 3, adjustment is required
via the potentiometer located in the circular cut-out on the AD converter board shield:
If the top cover of the UVM340 has become loose or incorrectly seated, this document
provides an engineer with instructions for correct fitment.
1. Loosen the four 2.5mm hex bolts on the sides of the top cover (2 on each side)
2. Lift the front of the top cover until the front end of the instrument begins to lift
away from the workspace below.
3. Whilst holding the top cover, re-tighten the front bolts, this ensures the top cover is
at its highest point possible.
4. Return the instrument to normal placement on your work surface and tighten the
rear bolts.
6.1 QC Plate:
To check the performance of the Instrument, we recommend to the ASYS Hitech QC plate
(SG010160) with the WinQC software. The QC plate checks the mechanical alignment of both the
plate transport and the optical path and tests the accuracy and precision.
The QC-plate consists of a precise milled aluminium body in the dimensions of a standard
microplate and three neutral density glass filters. In contrast to other neutral density filters the
sensitive coated area is protected by another glass layer, avoiding the known problems with
scratches and degradation caused by frequent cleaning. The neutral density filters cover all of the
eight measurement channels.
For instructions on how to use the QC-plate and WinQC - software see the manual which comes
with the QC-plate.
An alternative to the QC-plate is the so-called ‘Orange Test plate’ (S5450174). The results
obtained with this plate are not as accurate as with the QC-plate but usually sufficient for a basic
check.
The ‘Orange Test Plate’ is a simple tool to check the correct wavelength and function of an
interference filter. This is performed by comparing the enclosed reference OD and scan data
with data obtained by a measurement using the wavelength in question.
Furthermore, the results of the measurement can demonstrate the correct function of the
reader.
Note: The correlation of the reference data with the measured data does not prove that the
wavelength is correct when using this plate.
NOTE: For a complete performance check of the Instrument, use the QC plate with the
WinQC software.
10.1 Details
Institute: Date:
Department: Contract No:
City: Engineer:
Customer name: Instrument:
Telephone: Serial No:
Computer model: Software used:
Software version: Accessories:
Engineer name: Representative name:
Customer:
Address:
Complete section A / C and if applicable complete all of section B, providing further information as requested or appropriate.
A. Has this instrument and its accessories been used in any invasive procedure or been in contact with blood, other body
fluids, respired gases or pathological samples?
B. This instrument and its accessories have been exposed internally or externally to hazardous material as indicated below.
YES/NO Blood, body fluids, respired gases, pathological samples, other biohazards.
C. Has the instrument and its accessories been cleaned / decontaminated and suitably prepared for safe
handling/transportation?
YES/NO If YES please indicate the methods and materials used for decontamination.
If the instrument and its accessories could not be decontaminated please state the reasons why.
I declare that I have taken all reasonable steps to ensure the accuracy of the above information, in accordance with
MHRA DB 2006(05)
Authorised signature:……………………………………………………………………….
Name (printed):………………………………………………………………………………..
Position:……………………………………………………………………………………………
Date:……………………………………………………………………………………....
Please note that instrumentation will not be accepted for servicing or return until this form is completed fully.
Instrumentation that has not been cleaned sufficiently or decontaminated may be subject to additional charges.