Lactoscan Indi en
Lactoscan Indi en
Milkotronic Ltd
LACTOSCAN INDI
MILK ANALYZER
LCD display – 4 lines x 16 characters
(with option for manual cleaning with syringe)
Operation manual
Switching Adapter
Measurement modes
cow milk
buffalo milk
mixed milk
CAUTION!
Keep the switching adapter dry!
Please, read and follow strictly all the instructions in the manual.
SAFETY INSTRUCTIONS
1. Read this manual carefully and make sure that you understand all the
instructions.
2. For safety purposes the device is equipped with grounded power
cable. If there is no grounded electrical outlet where the device will
be used, please, install such before using the device.
3. Place the device on leveled and stable plate. In case it falls or is
severely shocked it may be damaged.
4. Connect to the electrical network in such a way that the power cable
to stay away from the side for accessing the device and not to be
stepped on.
5. Every time before cleaning the device switch it off and unplug it from
the electrical outlet. The device has to remain unplugged till the
cleaning completion.
6. Do not disassemble the unit in order to avoid possible electrical
shock. In case of malfunction contact your local dealer.
7. Handle the liquids the device works with carefully, following all the
instructions for their preparation.
8. Place the switching adaptor in such a way as to be protected from
overflow and spillage of liquids.
In the table below the standard delivery configuration of the milk analyzer is
listed:
№ Description Item № pcs
In the table below the milk analyzer spares and accessories, which are
delivered on customers request are listed:
1. FUNCTION
2. TECHNICAL PARAMETERS
These modes have been calibrated on customers’ request for 3 milk types
from the following: cow, sheep, UHT, buffalo, goat, camel milk, cream, ice
cream mixtures, whey, recovered milk, etc. before leaving the production
facilities and the text on the display will be for the corresponding types, as is
indicated on page 2 Measurement modes.
2.1.4. Cleaning
Note:
Pressing the button labeled A - start re-printing the results
Pressing the button labeled B - start of rinsing (2.1.4.)
These commands are possible when the unit is in Idle mode, i.e. expecting
command for measuring sample.
2.3. Accuracy:
Fat …..……………………………….. …...………………….…... 0.06%
SNF ………………………………….. …………...………….…... 0.15%
Density ……………………………… ……………………….. 0.3 kg/m3
Proteins ……………………………... ……...……………….…... 0.15%
Lactose ……………………………… .…………………………... 0.20%
Water content ………………………. .……………………………. 3.0%
Temperature of milk ……………….. .……………………………… 1oC
Freezing point……………………….. ..……...………………… 0.005oC
Salts ………………………………... …..…………...…………... 0.05%
The difference between two consequent measurements of one and the same
milk could not exceed the maximum permissible absolute error.
90-1801-0008
RS232 Interface Cable - Milk Analyser – Serial Printer/ IBM PC
90-1801-0009
DC 12V Power Supply Milk Analyzer Cable
1. GND
2. No connection
3. No connection
4. 12V DC
3.2.1.1. Put the analyzer on the working place, providing good ventilation and
not in the vicinity of heat providing devices or sources. The temperature in the
premises has to be in the boundaries 10-30oC.
3.2.1.2. Check if the power switch is in "0" position and that the outlet voltage
complies with the voltage indicated on the rating plate of the analyzer.
Connect the power supply cable to the electrical outlet.
Milkanalyzer xxx
LCD vers xx
MA vers yy
MA ser. N. xxxx
where:
MA ser. N. xxxx is the serial number – written on the rear panel of the
analyzer.
These data are called analyzer’s Identity
3.2.1.4. Till the analyzer is prepared for work (at about 5 minutes) the
following message is written on the display: “Getting ready”. Above pointed
time is in dependence of the environmental temperature and increases with
decreasing the temperature.
3.2.1.5. When the device is ready for work the display shows: “Ready to
start”.
3.2.1.6. If you want to pass to another mode press the button Enter and hold
it pressed. The following message appears on the display:
Release button to
start menu
Release the button Enter. The display shows the possible working modes:
Mode selector
Cal1 – Cow
Cal2 – Sheep
Cal3 – UHT
------------------------------
Rinsing
Final Clean
Printing
Using “up”▲and ”down”▼buttons, choose the working mode and press Enter
in order to start it.
Results:
F=ff.ff S=ss.ss
D=dd.dd P=pp.pp
L=ll.ll W=ww.ww
Where:
By pressing the button” Down” ▼ the display shows the second page,
containing the results:
Page 2 Results:
T=tt.tC pH=pp.pp
FP=-0.fff
S=0.sss A=aa.aa
Where:
tt.tC - sample’s temperature;
3.2.3.2. Write down the results in the form. The results remain on the display
till a new measurement is started. If the analyzer is connected to a computer
or a printer, it sends the data to the computer or prints them.
SCHEME:
PRINTOUT OF THE RESULTS-EXAMPLE
Time: xx:xx:xx
Date: xx:xx:xxxx
Milk analyser 60 SN: xxxxx
Calibration 1 – UHT
Results:
Temp.Sample………………..˚C
Fat………………………..xx.xx%
SNF………………………xx.xx%
Density…………………….xx.xx
Protein…………………...xx.xx%
Lactose…………………..xx.xx%
Added water…………….xx.xx%
pH……………………………..xx
Salts…………………….xx.xx%
Freezing point…………-x.xxx˚C
Deliverer No……………….xxxx
Liters…………………………xx.x
In case you missed to order these chemicals, the alternative is to use alkaline
and acidic cleaning solutions for dairy equipment by one the companies,
producing such chemicals, as for example:
http://www.diversey.com
http://www.ecolab.com
http://www.calvatis.com
Do not use chemicals not intended for usage in the milking systems or
vessels in the dairy sector. Pay special attention to the concentration of the
acidic chemical. Increased concentration may damage the measuring
sensor.
It is done in the process of routine work of the analyzer. Its aim is to prevent
drying up and adhesion of different milk components in the milk analyzer’s
measuring system.
It is easy to understand what is the period on which the rinsing could be done
as the analyzer reminds you when it is necessary. This is done by a sound
signal in 1-second cycle after the set time intervals elapse:
55 min. after switching on the power supply of the analyser, but idle
work;
Time to start
cleaning
4.1.2. Rinsing
After above message is received put in the recess of the analyzer a sample
holder with water.
Press Enter to start the rinsing mode.
In this mode the analyzer makes 8 cycles and stops.
Already used solution is poured out of the analyser. Now the device is ready for
the next measurement. In case of doubt that the analyzer is still not well
cleaned, the procedure may be executed repeatedly.
4.2.2. Cleaning
Please, pay attention that, when the analysers gives a signal for need of
cleaning 15 min after the last measurement of real milk samples or 55 min.
after being powered and not used, cleaning is made ONLY with alkaline
solution in concentration 1-3%.
During the basic/final/manual cleaning sequence is: alkaline solution – water
– acidic solution - water
Place a glass with water and press button Enter to continue. The following
message appears on the display:
Phase 1:
Cleaning
And the number of the current cleaning cycle. After rinsing is finished, the
display shows:
Place a glass filled with cleaning solution and press button Enter. Display
shows:
Phase 2
Detergent
Temp
Cleaning Fail
Low Deterg Temp
Press Enter
to Start Again
Cleaned
Ready to Start
Please, pay attention that the cleaning solution for manual cleaning needs to
be preliminary heated up to 50-60 degrees C.
- Place the sample holder filled with warm cleaning liquid in the recess of the
analyser;
- Press the piston;
Do the total cleaning with alkaline solution – water – acidic solution – water.
IMPORTANT
THE MAIN REASON FOR MALFUNCTIONING OF THE
DEVICE IS THE BAD CLEANING OF THE SYSTEM AFTER
MAKING ANALYSIS.
In case of malfunction due to the bad cleaning of the
analyser your guarantee is not valid anymore and any
repair has to be paid.
In the table below are described the possible malfunctions during the milk
analyzer’s exploitation and ways for their repair/remedy. If the problem
persists after all recommended measures are taken, please, connect the
nearest service center for help. Do not forget to tell the analyser’s identity.
In the process of work with the analyser there is a possibility the results to
start differing between the data for some of the measuring parameters when
measured with the milk analyzer and the corresponding reference method of
analyses (Gerber for fat, Kjeldhal for proteins etc). In order to establish the
possible discrepancy and to correct the readings of the milk analyser do the
following:
This is a basic moment for the correct checking the accuracy of the analyser
and for making correct and precise correction and calibration. It is
accomplished according Appendix Sampling and preparation of samples for
verification the accuracy of the milk analyzer, making corrections and
recalibration.
Make measurements with different samples (not less than 3) with known
values of a separate parameter (for example fat content), determined by the
known reference methods of analyses (for example Gerber's method for
determination of fat content). For more accuracy it is recommended among
these samples to be also such with values, close to the lowest and highest
bounds for the measured parameters.
Make 5-time measurement for each of the samples. Calculate the average
value for each sample parameter, without taking into consideration the first
measurement for each sample.
Make comparison between the values of the parameter from the reference
sample and measured with the analyser. Make analyses of the difference
received.
6.2.2.1. If the received differences are relatively constant value for samples
with different content of the analysed parameter, it is necessary to make
correction.
For example
М% of the reference samples: 2,20 3,00 3,80 4,60 5,20
М%average when measuring
with the analyser: 2,38 3,17 4,01 4,79 5,42
Difference: 0,18 0,17 0,21 0,19 0,22
For example.
М% of the reference samples: 2,20 3,00 3,80 4,60 5,20
М% when measured with the
analyser: 2,02 2,93 3,76 4,75 5,44
Difference: -0,18 -0,07 -0,04 0,15 0,24
Conclusion: It is obvious that the difference is variable value and
recalibration have to be done (See Recalibration, p.6.4).
Release button
to start setup
Setup Menu
Special modes
Corrections
Settings
------------------------------
Tests
pH & Co Meter
Accessories
Exit
Corrections:
Measurement
Temperature
Cond measure
Exit
Cal1 Cow
Param:Fat
Correct=00.00
Edit OK Next
Using the buttons “up”▲ and ”down”▼position on the action you want to take
(for example Edit) and press the button Enter.
Cal:….
Param:….
Correct= 00,00
- OK +
Using the buttons “up”▲ and ”down”▼is possible to increase or decrease the
value of the measured parameter in the above pointed limits. Leaving this
mode means saving the correction value and activating it.
6.4.1.1. Press the button Enter and without releasing it switch on the power
supply of the device, wait for the starting identification messages and release
the button after the following message appears on the display:
Release button
to start setup
Setup menu
Special modes
Corrections
Settings
------------------------------
Tests
pH & Co Meter
Accessories
Exit
Cal: 1 Cow
Prev OK Next
You can choose the type of milk to be calibrated. By pressing ▼ (Next) you can
switch between Cal: 1, Cal: 2 or Cal: 3.
By pressing the button ОК you are choosing the type of calibration.
Cal1 Cow
High
FAT=f.ff
Edit OK Next
For example:
FAT=05.29
Cal1 Cow
High
FAT=f.ff
- OK +
With buttons ▼(-),▲(+) set the needed value. With next pressing of Enter the
cursor is moved to the next number. After needed value entering completion for
FAT, press Enter (ОК) and you are going back to the previous menu:
Cal1 Cow
High
FAT=05.29
Edit OK Next
With the button ▲(Next), choose SNF and in the same as above
described procedure, enter the value for SNF. After it is finished, press
“Enter” (OK) and you are going back to the previous menu. With button
▲ (Next), choose DEN (density) and enter the value for density; the rest
of the parameters are entered in the same manner – LAC (lactose), SOL
(salts), PRO (protein)
Cal1 Cow
High
PRO=f.ff
Edit OK Next
If you miss to enter some of the parameters of milk, the following warning
message will appear:
Then you must press the button Enter (ОК) and enter the missed parameters.
After all the parameters are entered, press Enter (ОК).
You must enter values for all the measured milk parameters!!!
The screen for entering the results, received with the corresponding
reference methods (See Appendix Methods) for the low-fat milk is displayed:
Cal1 Cow
Low
FAT=f.ff
Edit OK Next
In the same way the values of milk with low fat sample are entered.
You must enter values for all the measured milk parameters!!!
Cal: Cow
Put sample High
5 times
Before each milk measurement stir 2-3 times the milk sample by pouring it
from one vessel to another. The needed quantity is poured in the sample-
holder and it is put in the recess of the analyser. Start the measurement by
pressing the button Enter. The sample is sucked. Appears the following
menu:
Cal: Cow
Put sample: High
5 times
Temp=….
Cow
High
N1=….. 2=…….
Cal meas=1/5
Cal: Cow
Put Sampl: Low
5 times
Stir 2-3 times the milk sample before each measurement by pouring it from
one vessel to another. The needed quantity is poured in the sample-holder
and it is put in the recess of the analyser. Start the measurement by pressing
the button Enter. The sample is sucked. Appears the following menu:
Make 5 times measurement of the low FAT sample.
After 5th measurement completion automatically appears the menu:
Cal: Cow
Put sample: Water
5 times
Recalibrated
Power Off-On
This means that the calibration was completed successfully and the analyzer
is recalibrated for cow milk, marked as “Cal: Cow”.
Switch off the power supply of the device and switch it on again.
The device is ready to work with the new calibration.
Next time when the analyser is switched on, it will be ready for work with
those milk types it was just calibrated with.
If calibration with another milk type is needed, do not forget to change the
calibration number for the new type of milk.
Calibration for Sheep milk will be saved as second calibration, UHT – as
third. This consequence may not be followed and calibrations can be saved in
whichever order is needed. Calibration can be done with different liquid dairy
products using 2 representative samples.
In order to start the Setup of the device the operator has to press the button
Enter and without releasing it to switch on the power supply of the device, to
wait for the starting identification messages and to release the button after
the following message appears on the display:
Release button
to start setup
Setup Menu
Special modes
Corrections
Settings
------------------------------
Tests
pH & Co Meter
Accessories
Exit
You may move in the menus by using buttons “up” ▲ and “down” ▼.
If by pressing the button Enter you choose a menu, each menu offers new
points/submenus. When Exit is chosen the device leaves the Setup mode
and returns to normal work.
Due to the continuous improvements made in the milk analyser or due to the
type of the ordered product, it is possible some of the options in the device to
be not active. In this case, if you try to enter the corresponding menu, the
following message will appear: Not available option.
Serve for choosing special (technological) working modes. After starting it the
following appears on the display:
Special modes
Calibration
Cycle
Exit
7.2.2. Corrections
7.2.3. Settings.
Net number.
Serves for assigning the device network number when connecting it in the
production network. The possible numbers are from 0 to 15 including.
After starting this function the display shows the following:
Net number
0
- OK +
By using the button “up”▲ the operator has the possibility to increase the
number, showing the channel’s number, and by button “down”▼, to decrease
it. Pressing the button Enter saves the chosen channel and exits the function.
When connected in the production network each device has to have a unique
number.
Recalibrate.
Serves for changing definite calibration. Methods are described in point 6.4.
Save Calibr
Through this menu you may save the new calibration
Current calibration content is not changed, the analyzer continues using it,
but there is a reserve copy in an internal buffer.
Restore Calibr.
Through this menu you may restore the old one (factory) calibration. in the
device or to restore the old one (factory) calibration. This is necessary in case
that you’ve calibrated the device for cow milk, but after that the device is not
measuring correctly and you decide to restore the factory calibration settings.
Position the cursor across “Restore calibration” and press “Enter”
Restore calibration – restores the chosen calibration from the internal buffer.
The current calibration is replaced with the calibration from the internal buffer
and the analyzer starts working with it. The content of the internal buffer is not
changed.
Settings Page 2.
After this menu is started the display shows the following:
Settings Page2
Set Calibr. Name
Select High Fat
HF Speed for Cal
Re Reslt Precision
PCB Main Identity
Larg Res En/Dis
Exit
Large Results
No
No OK Yes
and normal), numbers and punctuation marks and popular symbols. The
calibration name consists of 8 symbols.
Example:
When it is suitable to us this possibility of the analyser? For example if you have
a device factory calibrated for Cow milk, Sheep Milk and UHT milk, but yоu need
oftenly to measure camel milk. Using the methods, explained in details in
Appendix Methods you may make a new calibration without need to send the
analyser back to the producer for calibration. Using this procedure you may
make calibrations for most oftenly analysed milk and to write down the exact
calibration name, which will be shown on the display and printed on the printer.
After starting this menu the display shows:
Select Calibr
Cal1: Sheep
Cal:Sheep
If a name from the preliminary given names list is chosen, the display shows:
Cal1: Sheep
UHT
Cal1: Sheep
User Edited
Name:
Prev Set Next
After editing the last (eighth) name symbol, the display shows:
Cal1: Sheep
User Edited
Name:MilkShp
Exit Save
Cal:1 Cow
Prev OK Next
After choosing freezing point for the calibration, the following is displayed:
Cal: 1 Cow
-0.fff
- OK +
Where:
Cal: 1 - basic freezing point to be edited for chosen calibration.
-0.fff - basic freezing point current value.
7.2.4. Tests.
Start different tests. Possibilities:
Test pump.
Starts pump’s test. The number of the completed suction/display cycles is
indicated.
Ultrasound.
Test for the ultrasonic system. Used in production conditions.
Set Amplitude.
Serves for ultrasound amplitude adjustment. It is used under production
conditions or by the customer (after sensor change) according the
instructions in the document SetCell.pdf.
Please, use this menu only after reading the above pointed document
SetCell.pdf
RS232 COMPort.
Keypad
USB Flash
7.2.5. Exit
By pressing the button you may leave the program and pass towards another
menu.
Setup menu
Special
modes
Calibration
Cycle
Exir
Corrections
Measurement
Cal 1
Fat
SNF
Density
Lactose
Salts
Proteins
Water
Cal 2
Fat
SNF
Density
Lactose
Salts
Proteins
Water
Cal 3
Fat
SNF
Density
Lactose
Salts
Proteins
Water
Temperature
Cond measure
Settings
Net number
Recalibrate
Cal 1
Cal 2
Cal 3
Edit FrPoints
FrPoint Calibr1
FrPoint Calibr2
FrPoint Calibr3
Save/Rest Cal
Save Calibr 1
Save Calibr 2
Save Calibr 3
Rest Calibr 1
Rest Calibr 2
Rest Calibr 3
Fan Temp Offs
Settings Page2
Set Calibr Name
Select High Fat
HFSpeed for Cal
Reslt Precision
PCB Main Identi
Larg Res En/Dis
Set Base FrPnt
Tests
Test pump
Ultrasound
Set Amplitude
RS232 COMPort
KeyPad
USB Flash
Battery
Measure Bat U
Ctrl Enabl/Dis
RT Clock
Display Time
Adjust Time
Adjust Date
APPENDICES
APPENDIX 1: PREPARATION OF SAMPLES FOR
MILKANALYSERS’ CALIBRATION
For calibration are needed samples of cow milk with the following
parameters:
It is recommended the first cow milk sample with low fat content to be with
the following parameters:
2-2,3% FAT; 8.7-9% SNF; 3,3-3,5 % Protein; 4,8-4,9% Lactose; 0,75 Salts;
1030-1033 kg/m3 Density.
The second cow milk sample with high fat content to be with the following
parameters:
5-5,3% FAT; 8.4-8,79% SNF; 3,1-3,2% Protein; 4,6-4,7% Lactose; 0,7 Salts;
1028-1029 kg/m3 Density.
If, after milk’s separation you do not obtain samples in the requested range,
then, by adding milk with high fat value into the low fat milk sample you can
obtain necessary value-2,3%
Analogous to this, by adding low fat milk sample into a milk sample with high
fat value you may receive 5,3%
Samples with medium values are received by mixing low fat and high fat
samples in necessary proportion.
If there is a need of longer sample storing they have to be preserved; the
most commonly used preservative is potassium dichromate (K2Cr2O7) - 1 g
for 1 000 ml.
When using samples, stored shortly, preliminary pour the sample from one
vessel to another in order to distribute the milk components evenly paying
attention not to form foam in the sample.
When the samples are stored for a longer period it is recommended to warm
it up to 35-45 °С, and the vessel to be shaken carefully. In case that there is a
cream stuck on the vessel’s surfaces – remove it. The sample is poured from
vessel to vessel several times and is cooled down (advisable to 20 °С /.
If there is separated liquefied fat or white particles with irregular form on the
vessel’s walls reliable results could not be received.
Because it is very difficult both lactose and salts to be measured but they are
substantial and influence in great extend when determine added water. That’s
why it is better both lactose and salts to be calculated by using SNF results.
The milk must be for sure without added water.
If you are unable to make the analysis of milk in certified methods in a pinch
you can use the following formulas:
2. SNF determination.
B/ Known quantity of fat and density (most commonly used method when
maximum accuracy is needed).
We recommend the following formula:
This is a universal formula and actual for milk of almost all kind of cows and
sheep all over the world.
Where
SNF –solids-non-fat content in percentages (%),
0,45 – constant coefficient.
This is an actual coefficient for sheep breeds on the territory of the Balkan
Peninsula.
This is an actual coefficient for sheep breeds on the territory of the Balkan
Peninsula.
This is an actual coefficient for sheep breeds on the territory of the Balkan
Peninsula.
The milk analyzer determines the freezing point of each sample and the
quantity of added water. The milk analyser does not measure the freezing
point, but calculates it from the components it depends on. The basic
components in the milk are water, solids, lactose, FAT, proteins, minerals
(salts) and acids. The freezing point depends only on the diluted in the milk
components and quantity of the solvent (in the milk it is water). The ultrasonic
technology allows direct measurement of FAT, proteins, lactose + salts (the
soluble components, only influencing the freezing point), and the quantity of
the solvent in % is determined by 100 % – total solids %, total solids = lactose
% + FAT % + proteins % + salts % + acids %.
Without understanding the meaning of the freezing point – determined or
shown from the milk analyzer added water result easily may lead to a mistake
for the value of this parameter.
Milk freezes at lower temperature than water. The average freezing point of
the raw milk in the most regions is at about -0,540С. The average reading
for your region is called “basic” freezing point.
The freezing point of milk is a “physiological constant”. This does not mean
that it will not vary. In fact feed, breed, season, time of lactation, climate,
whether the sample is taken at the beginning, middle or end of lactation – all
these factors will have an effect on the freezing point of the individual sample.
This means that there is an average value of all these numbers. The more
samples used in obtaining this average, the more reliable it is as a base. Or
the basic freezing point is an average of freezing points of milk, taken from
many cows. When a laboratory checks a producer, it is only comparing the
average of the producer’s cows against a larger area average.
The Health authorities establish the basic freezing point or agriculture
departments in some regions, sometimes by universities, separate dairy
producers, or their associations. Frequently, tolerances have been
established on top of a basic freezing point to allow some variations in the
milk as well as device or operator variations.
Without mentioning the basic freezing point, the Association of Official
Analytical Chemists now recommends an upper limit freezing point at -
Where:
FrPointBase is the basic freezing point
FrPointCalc is measured freezing point
Note:
If the freezing point is not correctly determined, the result for the added water
is not valid. In this case results for FrPoint and AddWater are not shown on
the display and on the printout from the printer. If the density of the measured
sample is 0, the result for AddWater is not valid and is also not shown on the
display and the printouts.
Sample:
First variant
If you’ve entered for milk analyzer basic freezing point -0.520C (according
article 5.9 of the EU Milk Hygiene Directive 92/46/ЕЕС), measured freezing
point –0.540C, using the above pointed formula you’ll receive –3,8%.
Because it is not possible the added water to be negative value, the milk
analyzer indicates 0% added water. The reason for this is the tolerance in the
basic freezing point, reasons for which are described below.
If in the same milk we add 3,8% water, and the basic freezing point is the
same, the milk analyzer will measure freezing point –0.520C, and will
indicate again 0% added water.
Second variant
If you’ve entered for the device basic freezing point –0.540C, measured
freezing point –0.540C, the milk analyzer will indicate 0%. When you add
3,8% water, the device will indicate 3,8%-added water.
From the above mentioned follows that it is very important to enter correct
basic freezing point in the device.
The device’s results for added water may give information about doubt of
added water in the milk and the exact value of this added water may be
determined after a “cowshed sample” is taken and the result for the freezing
point, measured by the milk analyzer of the “cowshed sample” is entered as
basic freezing point in the formula for calculation of added water.
Then the result from this formula will give us the absolute value of the added
water for the corresponding milk supplier.
Contents:
1. FUNCTION ............................................................................................................................................................. 5
3. QUALIFICATION OF RAW MILK, THERMALLY TREATED MILK, OTHER DAIRY PRODUCTS AND
DERIVATIVES .............................................................................................................................................................. 13
6.1. TAKING SAMPLES AND PREPARATION OF SAMPLES FOR CHECKING THE ACCURRACY OF
THE MILK ANALYSER, MAKING CORRECTIONS AND RECALIBRATION ................................................ 27
GUARANTEE CARD
LACTOSCAN INDI
Standard model
Password:
Distributor:
Signature:
Stamp:
GUARANTEE CARD
Purchaser:
Service report:
Service Delivery
Damage Signature
entry date date
Covers:
Lactoscan INDI
HEAD OFFICE: