SOFTWARE MANUAL
from version 1.1.17
Table of Contents
Table of contents
Important Notices........................................................................6
First Steps....................................................................................7
System Requirements ...................................................................7
Installing Software .........................................................................9
Connecting the Device with the Computer.....................................10
Starting TRENDset ......................................................................11
Elements of the User Interface...................................................12
Menu Bar ....................................................................................13
Symbol Bar..................................................................................13
Status Bar....................................................................................14
Function Bar................................................................................15
Working with Files......................................................................16
Opening Files...............................................................................16
Close File.....................................................................................16
Save Data ...................................................................................17
Quit TRENDset . ..........................................................................17
Print............................................................................................18
Pre-Settings...............................................................................19
Setting the Language for TRENDset .............................................19
Registration and Deregistration for Service Personnel ...................19
Setting Event Limits for Statistics..................................................20
2
Table of Contents
3
Table of Contents
4
Table of Contents
5
Important Notices
Important Notices
TRENDset is the analysis software for respiratory therapy decvises pro-
duced by the company HOFFRICHTER GmbH.
Its main functions are the analysis of therapeutic data and remote
adjustment of the device parameters.
This software is used in the sleep laboratory by medical personal and
by the service.
WARNING
Before you use TRENDset:
NOTE:
1. Read the User´s Manual for the device. We recommend that you always
2. Read the Software Manual for TRENDset. use the current version of
TRENDset . Keep up to date at
www.hoffrichter.de.
6
First Steps
First Steps
System Requirements
Hardware Requirements
• AMD Athlon/Pentium Class (x86), Operating Frequency 700 MHz
• min. 192 MB RAM for MS Windows® 98, NT4.0, ME
• min. 400 MB RAM for MS Windows® 2000, XP, Server 2003, Vista
• min. 150 MB free hard disc space (depending on the desired stor-
age volume for patient data)
• min. 280 MB free hard disc space for the Microsoft .NET Frame-
work 2.0 Redistributable (x86)
• 2 MB Graphics card with at least High Color (16 Bit)
• Mouse
• One available COM-interface for each device
• One available SD-interface (only when using the DATA box)
7
First Steps
Software Requirements
• Operating system:
–– MS-Windows® 98
–– MS-Windows® NT 4.0, Service Pack 6a
–– MS-Windows® ME
–– MS-Windows® 2000
–– MS-Windows® XP
–– MS-Windows® Server 2003
–– MS-Windows® Vista
• Installation of the newest Service Packs and critical updates for
the operating system used
• Windows Installer 3.0 (Windows 2000/NT and lower)
• Microsoft .NET 2.0 Frameworks
• Microsoft Internet Explorer 5.01 or later versions
Other requirements
• a serial cable (connectors: 9pin D-Sub plug and RJ11 plug)
8
First Steps
Installing Software
There are two options for installing TRENDset: NOTE:
1. Installation from CD TRENDset can only be installed
with administrator rights.
2. Download from the HOFFRICHTER-Homepage
www.hoffrichter.de
Installating from CD
Insert the CD in your computer´s optical drive. The Setup will start up
automatically if the Autorun-Function is active. Follow the installation
instructions.
If the .NET-Framework is not installed on your computer, you will be
asked to install the .NET-Framework. Follow the installation instructions.
9
First Steps
10
First Steps
Starting TRENDset
1. When you start TRENDset, the “Intro Window” appears.
2. TRENDset searches automatically for connected devices.
11
Elements of the User Interface
12
Elements of the User Interface
Menu Bar
The menu bar contains all menus and menu items needed to operate
TRENDset.
Symbol Bar
Using the symbol bar, you can call up all of the program’s most impor-
tant functions.
13
Elements of the User Interface
Status Bar
In the status bar you will find information about the connection status,
the selected function area and the type of the connected device.
14
Elements of the User Interface
Function Bar
TRENDset is subdivided into 6 function areas:
• Patient Data
• Device Settings
• Therapy Calendar
• Data and Events
• Alerts and Parameter changes
• Statistics
• Times and Counters
After the program starts, the function area for Patient Data is always
opened.
If you select another function area, the display in the work area will
change.
NOTE:
The number of function areas and the display in the work area depend
on the type of device.
15
Working with Files
Close File
Select File > Close or in the symbol bar.
Closing causes all displayed data from the connected device or file to
be lost. If changes have not already been sent to the device, you will
be prompted to save the changes.
After closing, the status will change from Online to Offline.
16
Working with Files
Save Data
Save
Select File > Save or in the symbol bar. Use Ctrl+S to save a file
quickly.
The files will be saved in the folder Own Files (Documents)/TRENDset
with the extension .csw. If a file is saved from an SD card, it is saved in
the folder My Files (Documents)/TRENDset/THERAPY SD_Archive with
the extension .csa.
Save as
With command Save as you can save the open file under another name
or in another directory.
Quit TRENDset
Select File > Exit.
Before quitting, TRENDset checks whether all data have been saved.
If you have not already sent your changes to the device, you will be
prompted to save.
17
Print
Select File > Print or in the symbol bar. Use Ctrl+P for printing a file
1. Select the printer. quickly.
2. Set the number of copies to print.
3. Mark the documents you want to print by activating the check
box. You can select the number of events to be printed under
Alerts and Parameter changes.
NOTE:
As long as TRENDset is open
the last printer settings used
are saved.
18
Pre-Settings
Pre-Settings
Setting the Language for TRENDset
Select Settings > Language.
You can choose from several languages. If you need an additional lan-
guage, contact our service. The selection of available languages can
be expanded at any time.
19
Pre-Settings
For these types of devices you can set the event limits.
• TREND II AUTO/CPAP
• VECTOR et AUTO/CPAP
• VECTOR ET AUTO/CPAP
• (i)CARAT et AUTO
• TREND 210
• AutoTREND
• VECTOR II FLEXLINE
• VECTOR II AUTO
• TREND II with SD cards
• VECTOR et with SD cards
• iCARAT et with SD cards
20
Communication with the Device
All founded devices are listed. Select the desired device by mouse
click. Afterwards starts the data import. This process will take at most
50 seconds. TRENDset changes to the status Online (see page 14). If no
device has been detected under Found devices, no devices will be dis-
played. TRENDset remains in the status Offline (see page 14). All program-
ming and analysis functions remain deactivated.
If TRENDset is connected
with a device, the status
message Online appears.
21
Communication with the Device
If TRENDset is connected
with a device, the status
message Online appears.
22
Communication with the Device
Troubleshooting
While TRENDset searches for a device, the following system mes-
sages can appear:
23
Communication with the Device
24
Communication with the Device
25
Communication with the Device
26
Communication with the Device
27
Communication with the Device
Using an SD card
There are two ways to store data on an SD card. NOTE:
• Respiratory therapy device with SD card (only in device models Before an SD card can be used
with SD card slot) for the DATA box or the respi-
ratory therapy device, it must
• DATA box with SD card be formatted with TRENDset.
28
Communication with the Device
29
Communication with the Device
DATA box
30
Communication with the Device
Read SD card
Select SD-Card > Read SD card.
The dialogue box Read SD card will display the patient data that are
saved on the SD card.
31
Communication with the Device
Format SD card
Before an SD card can be used with a DATA box or a repiratory therapy NOTE:
device, it must be formatted with TRENDset. It is irrelevant whether this The SD card will not be
accepted by the DATA box
is a new or used SD card.
or the respiratory therapy
device unless it is formatted
Select SD-Card > Format SD card. with TRENDset.
The dialogue box Format SD card will open and will list all removable
disks for your computer. Select the data storage device by activating
the check box.
To prevent accidental erasures of files on the other removable disks, it
will check whether there are still files on the selected drive. If yes, you
will be prompted to deleted these files manually.
32
Communication with the Device
33
Communication with the Device
Using Help
Help opens immediately if you
Select ? > Help TRENDset. press F1.
The Help file will open in PDF-Format.
NOTE:
Help opens with Acrobat
Reader. You can install the
program from the TRENDset-
CD if needed.
34
Patient Data
Patient Data
You manage patient data in the function area Patient Data, turn the ther-
apy on and off and get information about the most important charac-
teristics of the devices.
The workspace is sub-divided into Patient Data, Display and Device
Characteristics.
Patient Data
Device
Characteristics
Display
35
Patient Data
confirm the entries with Accept. Save the file (File > Save/Save as). VECTOR II FLEXLINE
VECTOR II AUTO
(as of device software 1.502)
36
Patient Data
Device Status
37
Patient Data
38
Device Settings
Device Settings
In the function area Device Settings you set the device parameters.
The workspace, depending on the device connected, is subdivided
into Pressure, Respiration, Softstart Ramp/Mask Test, Mask Alarm, Device
and Intro Text.
PIN code
Activation
39
Device Settings
You activate or deactivate the trigger using the On / Off button. The
active state is shown left of the button.
All changes will be highlighted in red.
To restore all entries to the initial condition, click on Reset.
Click on Send, to send the changed data to the device. Then the data
will be read again and displayed in TRENDset. The sent and received
data are compared with each other. This assures that the correct
data was written in the device.
40
Device Settings
Definition of Terms
I-FLEX The value of pressure increase depends on the value of the breathing flow, the set I-FLEX and the set
therapy pressure. The I-FLEX settings (0 – 3) are for individual adjustment to the patients breathing flow.
E-FLEX The value of pressure decrease depends on the value of the breathing flow, the set E-FLEX and
the set therapy pressure. The E-FLEX settings (0 – 3) are for individual adjustment to the patients
breathing flow.
Start Pressure Starting pressure
Min. Pressure Minimum pressure
Max. Pressure Maximum pressure
Pressure Steepness The parameter Pressure Steepness increase determines how quickly pressure increases in an event.
The speed of the pressure increase can be adjusted to the patient’s sensitivity.
Inspiration Pressure Inspiration pressure
Expiration Pressure Expiration pressure
I-Slope Inspiration slope
E-Slope Expiration slope
Trigger Activating or deactivating the inspiration trigger in T-Mode
I-Trigger Sensitivity of the inspiration trigger
E-Trigger Sensitivity of the expiration trigger
Trigger Setting the trigger sensitivity
Target Volume Value for the target volume (only in the S, ST and T modes)
Additional Pressure Adjustable additional pressure, with target volume activated (only in the S, ST and T modes)
Min. Inspiration Time TI min Minimum inspiration duration (only in the S and ST modes)
Max. Inspiration Time TI max Maximum inspiration duration (only in the S and ST modes)
41
Device Settings
42
Device Settings
43
Device Settings
Definition of Terms
Frequency Breaths per minute
Inspiration Duration of inspiration over entire breath
Delay Time Delay for switch between from S-Mode to T-Mode
I:E Ratio I:E-Ratio from frequency and duration of inspiration
Backup frequency Frequency with which the patient continues to be ventilated
when an apnea has occurred
44
Device Settings
45
Device Settings
Definition of Terms
Softstart Time Time during which the pressure gradually increases
to therapeutic pressure, beginning at the starting
pressure for soft start or at minimum pressure.
Softstart Start Pressure Starting pressure for soft start
Mask Test In a mask test the device is operated at maximum
therapeutic pressure for the time selected in order
to test the tightness of mask. The maximum ther-
apeutic pressure results from the settings in the
function area Device Settings.
46
Device Settings
TREND 200
TREND 210
AutoTREND
TREND 300
TREND 400
TREND 500
TRENDlife
Calypso
VECTOR
VECTOR II
VECTOR II
FLEXLINE
VECTOR II AUTO
VECTOR II BILEVEL
47
Device Settings
48
Device Settings
Definition of Terms
Mask Alarm The mask alarm is activated/deactivated using the
On / Off button.
Alarm Delay 1 This value defines the delay period after which the
alarm will sound.
Pressure Difference 1 This value defines the tolerated deviation from the
set pressure.
1
only when the Mask Alarm is activated
49
Device Settings
Select Device Settings > Alarm Low Minute Volume or Alarm Apnoa.
Set the desired values using the selection fields. All changes are high-
lighted in red.
To restore all entries to their original condition, click on Reset.
Click Send to send the changed data to the device. The data is sub-
sequently read out again and displayed in TRENDset. The sent and
received data is compared. This ensures that the correct data has
been saved on the device.
Definition of Terms
Alarm Low Minute Alarm if the respiratory minute volume is too low
Volume
If the respiratory minute volume falls below the value
set here, the device sounds an alarm.
Alarm Apnea Alarm if an apnea is detected
50
Device Settings
51
Device Settings
Automatic Function
To set the automatic pressure adjustment, activate the check box Auto-
matic function adjustable. This will activate the check box “Automatic”
with the following options.
You can find additional information about the automatic function on page 53.
52
Device Settings
Definition of Terms
53
Device Settings
Pressure Ramp
Menu Scrolling
Display Breath
Heading Step
Threshold Off
Pressure Unit
Threshold On
Display Effi-
ciency Data
Brightness
Menu Type
Mask Type
Automatic
Automatic
Automatic
Automatic
Language
PIN Code
Intro Text
Pressure
Pressure
Display
Status
Mode
Delay
Device
VECTOR et
(i)CARAT et
VECTOR ET
TREND II
point
TREND 200
TREND 210
AutoTREND
TREND 300
TREND 400
TREND 500
TRENDlife
Calypso
VECTOR
VECTOR II
VECTOR II FLEXLINE
VECTOR II AUTO
VECTOR II BILEVEL
as of device software 1.501 only for the respective AUTO-Devices as of device software 2.140
54
Device Settings
55
Therapy Calendar
Therapy Calendar
The function area Therapy Calendar includes a graphic display of device
The therapy data are evalu-
use. It shows the duration of therapy per day and in some devices the ated in the function area
pressures used. The Therapy Calendar gets this data from the therapy Data and Events.
memory, the compliance memory and the efficiency data memory. 1
56
Therapy Calendar
Therapy Memory
Approximately 30 days of therapy are stored in the Therapy Memory.
Compliance Memory
When the Therapy Memory is full, the older data are transferred to the
Compliance Memory. The Compliance Memory stores data for approxi-
mately 1 year.
The data from the Compliance Memory are only displayed in the Ther-
apy Calendar and in the protocol from the Therapy Calendar. There
is no display or evaluation of these data in the function area Data and
Events or Statistics.
57
Therapy Calendar
58
Therapy Calendar
59
Therapy Calendar
selected
(red symbol)
not selected
(gray symbol)
If you want to select all therapy days, activate the control box Select
all displayed Days.
Additionally, you can decide whether or not therapy-free days will be
displayed in the Therapy Calendar. Activate the check box Display Days
without Therapy in Addition.
After you have marked the therapy days, you can switch to the func-
tion area Data and Events or Statistics and evaluate the data for the
marked therapy days.
The Complete Using Time in h and the Average Using Time per Day in h
are seen in the lower part of the workspace.
60
Therapy Calendar
61
Data and Events
62
Data and Events
63
Data and Events
A snoring phase is indicated if the snoring sound is above a threshold value for the time of the
Snoring
set event limit.
Device State Adjustment The adjustment phase includes the learning phase at the begin-
ning of a therapy and the 30 s timeout after an increased leak-
age, a mode switch between APAP and CPAP as well as after a
pressure change during CPAP therapy. During this period, the
APAP-Algorithm is turned off.
Leakage A leakage is indicated, if the middle system flow exceeds a
threshold value for a maximum of 2 l/s or a shows a change of
more than 300 ml/s.
Mode Display of Active Mode (APAP, CPAP, FLEX, BILEVEL S, BILEVEL ST, BILEVEL T)
64
Data and Events
Zoom View
Below the diagram are the control elements for setting the zoom
You enlarge the Zoom Area
area – 1 min to 24 h is possible.
by double-clicking with the
Enlarge Zoom Area left mouse button; by double-
Direct Selection
clicking on the right mouse
of Zoom Area
button you reduce the Zoom
Area.
Reduce Zoom Area
66
Data and Events
To enlarge, click on .
To reduce, click on .
67
Data and Events
68
Data and Events
Airway Constriction
Snoring
Mode
Adjustment
Leakage
Hyperventilation
Mouth Exhalation
General Artefact
CPAP
Standard
Standard for VECTOR et, TREND II and iCARAT et as of device software 2.140
69
Alerts and Parameter changes
71
Statistics
Statistics
In the function area Statistics, the therapy data are visualized with bar NOTE:
diagrams. This allows you to precisely analyze therapy data. The dia- In Statistics, only the pure ther-
grams support you in titrating an effective therapeutic pressure. apy time without Mask Test or
Softstart is evaluated.
TREND II AUTO
VECTOR et AUTO
VECTOR ET AUTO
(i)CARAT et AUTO
AutoTREND
VECTOR II AUTO
72
Statistics
Number of therapy days that were activated 60 % of all therapy days were activated
in the Therapy Calendar for the statistics in the Therapy Calendar for the statistics
function (see page 59) function
73
Statistics
Number of therapy
days selected in the Number of therapy
selection box days selected using the
selection box in %
74
Statistics
75
Statistics
To select an event, click on one of the radio buttons. You can further
specify whether the total number or the average number per hour should
be displayed, and for snoring phases total minutes or minutes per hour.
76
Statistics
77
Statistics
78
Statistics
Statistical Analysis
Click on Statistical Analysis.
The overview includes along with the statis-
tical data, data on the recording time, dura-
tion of therapy, number of days selected
for the statistics and the most important
device settings.
As an aid to decision-making for the phy-
sician, the lower window area shows what
therapeutic pressure was used 90 % of the
time and at what therapeutic pressure 90 %
of the events occurred.
To close the window, click on the “x” at the
upper right.
79
Times and Counters
80
Times and Counters
81
Times and Counters
82
Times and Counters
Set Date
1. In the Date selection box, mark the day, month or year and enter
the desired value.
OR
Click on the arrow in the selection box. A calendar will open.
Select the date.
OR
It is not possible to set the
To import the date from the computer, just click on system time or to change the
Set Computer Time. date on the SD card. In order
to make these changes, the
If you want to remove you settings again, click on Reset. device must be connected.
2. Click on Send. The date will be sent to the device.
83
Times and Counters
Set Time
1. In the Time selection box, mark the hour, minutes or seconds and
enter the desired value. You have the option of using the arrows
to increase or decrease the marked values.
OR
It is not possible to set the
To import the time from the computer, click on Set Computer Time.
system time or to change the
If you want to remove you settings again, click on Reset. date on the SD card. In order
to make these changes, the
2. Click on Send. The time will be sent to the device. device must be connected.
84
Times and Counters
NOTE:
If you de-activate the Alarm Clock Function, it is no longer available in
the device until you re-activate it.
85
Times and Counters
Alarm is on
86
System Messages
System Messages
If data are transferred incorrectly or incompletely over the serial inter-
face, TRENDset will issue a system message.
If a system message from 1 to 255 is issued, the device has regis-
tered the fault and exported it. For values seen on the following table,
TRENDset has only registered the fault.