0% found this document useful (0 votes)
25 views

PDF Trend II Software Manual

This software manual provides instructions for using TRENDset software to analyze data from respiratory therapy devices. It covers system requirements, installation, connecting devices, using the user interface, working with files and data, device settings, statistics and more.

Uploaded by

ahmadba70
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
25 views

PDF Trend II Software Manual

This software manual provides instructions for using TRENDset software to analyze data from respiratory therapy devices. It covers system requirements, installation, connecting devices, using the user interface, working with files and data, device settings, statistics and more.

Uploaded by

ahmadba70
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 88

Quality makes the Difference

SOFTWARE MANUAL


from version 1.1.17
Table of Contents

Table of contents

Important Notices........................................................................6
First Steps....................................................................................7
System Requirements ...................................................................7
Installing Software .........................................................................9
Connecting the Device with the Computer.....................................10
Starting TRENDset ......................................................................11
Elements of the User Interface...................................................12
Menu Bar ....................................................................................13
Symbol Bar..................................................................................13
Status Bar....................................................................................14
Function Bar................................................................................15
Working with Files......................................................................16
Opening Files...............................................................................16
Close File.....................................................................................16
Save Data ...................................................................................17
Quit TRENDset . ..........................................................................17
Print............................................................................................18
Pre-Settings...............................................................................19
Setting the Language for TRENDset .............................................19
Registration and Deregistration for Service Personnel ...................19
Setting Event Limits for Statistics..................................................20

2
Table of Contents

Communication with the Device................................................21


Search the Device Automatically...................................................21
Search Device Manually................................................................22
Troubleshooting...........................................................................23
Read Device Data.........................................................................24
Delete Device Data.......................................................................26
Using an SD card........................................................................28
Using an SD card in a respiratory therapy device............................28
Use of the DATA box...................................................................30
Read SD card . ............................................................................31
Format SD card ...........................................................................32
Save Data from SD cards .............................................................33
Open Data from SD cards ............................................................33
Information about TRENDset........................................................34
Using Help ..................................................................................34
Patient Data . .............................................................................35
Entering, Changing and Managing Patient Data.............................36
Display current Therapeutic Pressure............................................37
Display Device Characteristics .....................................................38
Device Settings..........................................................................39
Set Therapeutic Pressure..............................................................40
Set Respiration Parameters..........................................................43
Set Softstart Ramp and Mask Test................................................45

3
Table of Contents

Set Alarm parameter....................................................................48


Setting Device Parameters...........................................................51
Working with the Parameter-Popup . ............................................55
Therapy Calendar.......................................................................56
Therapy Memory..........................................................................57
Compliance Memory....................................................................57
Efficiency Data Memory................................................................57
Therapy Calendar Areas...............................................................59
Select Therapy Days for Statistics.................................................60
Open Protocol on Duration of Therapy . ........................................61
Data and Events.........................................................................62
Analysis of Data and Events .........................................................63
Select Therapy Days.....................................................................66
Zoom View...................................................................................66
Enlarge and Reduce Diagrams......................................................67
Monitor Course of Therapy in Real Time........................................68
Alerts and Parameter Changes..................................................70
Display of the Events....................................................................71
Using the Filter Options.................................................................71
Statistics....................................................................................72
Select Therapy Days.....................................................................73
Analyze Distribution of Pressure Over Time....................................75
Analyze Frequency of Events . ......................................................76

4
Table of Contents

Effect of the Therapeutic Pressure.................................................77


Chronological Distribution of Events..............................................78
Statistical Analysis........................................................................79
Times and Counters...................................................................80
Run Times of the Device or Components.......................................81
Reset Filter Age............................................................................82
Set Date......................................................................................83
Set Time......................................................................................84
De-Activate Alarm Clock Function ................................................85
Set Alarm Time.............................................................................86
System Messages......................................................................87

5
Important Notices

Important Notices
TRENDset is the analysis software for respiratory therapy decvises pro-
duced by the company HOFFRICHTER GmbH.
Its main functions are the analysis of therapeutic data and remote
adjustment of the device parameters.
This software is used in the sleep laboratory by medical personal and
by the service.

WARNING
Before you use TRENDset:
NOTE:
1. Read the User´s Manual for the device. We recommend that you always
2. Read the Software Manual for TRENDset. use the current version of
TRENDset . Keep up to date at
www.hoffrichter.de.

6
First Steps

First Steps
System Requirements
Hardware Requirements
• AMD Athlon/Pentium Class (x86), Operating Frequency 700 MHz
• min. 192 MB RAM for MS Windows® 98, NT4.0, ME
• min. 400 MB RAM for MS Windows® 2000, XP, Server 2003, Vista
• min. 150 MB free hard disc space (depending on the desired stor-
age volume for patient data)
• min. 280 MB free hard disc space for the Microsoft .NET Frame-
work 2.0 Redistributable (x86)
• 2 MB Graphics card with at least High Color (16 Bit)
• Mouse
• One available COM-interface for each device
• One available SD-interface (only when using the DATA box)

7
First Steps

Software Requirements
• Operating system:
–– MS-Windows® 98
–– MS-Windows® NT 4.0, Service Pack 6a
–– MS-Windows® ME
–– MS-Windows® 2000
–– MS-Windows® XP
–– MS-Windows® Server 2003
–– MS-Windows® Vista
• Installation of the newest Service Packs and critical updates for
the operating system used
• Windows Installer 3.0 (Windows 2000/NT and lower)
• Microsoft .NET 2.0 Frameworks
• Microsoft Internet Explorer 5.01 or later versions

Other requirements
• a serial cable (connectors: 9pin D-Sub plug and RJ11 plug)

8
First Steps

Installing Software
There are two options for installing TRENDset: NOTE:
1. Installation from CD TRENDset can only be installed
with administrator rights.
2. Download from the HOFFRICHTER-Homepage
www.hoffrichter.de

Installating from CD
Insert the CD in your computer´s optical drive. The Setup will start up
automatically if the Autorun-Function is active. Follow the installation
instructions.
If the .NET-Framework is not installed on your computer, you will be
asked to install the .NET-Framework. Follow the installation instructions.

Installing the Download Version


Execute the TRENDset.zip.exe and unpack the files. The installation
will start automatically.
If the .NET-Framework is not installed on your computer, you will be
asked to install the .NET-Framework. Follow the installation instructions.

9
First Steps

Connecting the Device with the Computer


1. Connect the 9pin D-Sub plug1 of the serial cable to the COM interface
of your computer.
2. Connect the RJ11 plug (2) to the device.
3. Connect the device to a grounded contact outlet using the delivered
power cord.

1  The cable can be ordered from HOFFRICHTER under Article-No. 00005232.

10
First Steps

Starting TRENDset
1. When you start TRENDset, the “Intro Window” appears.
2. TRENDset searches automatically for connected devices.

Intro Window with TREND-


set version number

All founded devices are listed. Select the desired device by


mouse click. Afterwards starts the data import. This process
will take at most 50 seconds. TRENDset changes to the status
Online (see page 14). If no device has been detected under Found
devices no devices will be displayed and TRENDset remains in
the status Offline (see page 14). All programming and analysis func-
tions remain deactivated.
You can also open an existing file. Just click on File. Furthermore you
can read-in data from the SD card. Just click on Read SD card. If TRENDset is connected
with a device, the status
message Online appears.

11
Elements of the User Interface

Elements of the User Interface

Symbol bar Menu bar Status bar

Function bar Workspace

12
Elements of the User Interface

Menu Bar

The menu bar contains all menus and menu items needed to operate
TRENDset.

Symbol Bar

Using the symbol bar, you can call up all of the program’s most impor-
tant functions.

Close file Read SD card

Open file Parameter-Popup

Save file Print data

Search for device automatically

13
Elements of the User Interface

Status Bar
In the status bar you will find information about the connection status,
the selected function area and the type of the connected device.

Connection status Function area Device type

14
Elements of the User Interface

Function Bar
TRENDset is subdivided into 6 function areas:
• Patient Data
• Device Settings
• Therapy Calendar
• Data and Events
• Alerts and Parameter changes
• Statistics
• Times and Counters
After the program starts, the function area for Patient Data is always
opened.
If you select another function area, the display in the work area will
change.

NOTE:
The number of function areas and the display in the work area depend
on the type of device.

The active function area is high-


lighted in gray in the function bar.

15
Working with Files

Working with Files


Opening Files
Select File > Open or in the symbol bar.
Use Ctrl+O to open a file
The dialogue box Open will appear with the preset pathway Own Files quickly.
(Documents)/ TRENDset. The folder TRENDset will be generated the first
time the program is started and will be the standard folder for saving
all files belonging to TRENDset. It can only open files with the exten-
sion .csw or .csa.
Before a new file is opened, TRENDset first checks whether or not all
open data have been saved.
As soon as a file has been opened, TRENDset switches into the status
Offline. If you open a file that was saved to the hard drive from an SD
card, TRENDset switches to the status SD card archive.

Close File
Select File > Close or in the symbol bar.
Closing causes all displayed data from the connected device or file to
be lost. If changes have not already been sent to the device, you will
be prompted to save the changes.
After closing, the status will change from Online to Offline.

16
Working with Files

Save Data
Save
Select File > Save or in the symbol bar. Use Ctrl+S to save a file
quickly.
The files will be saved in the folder Own Files (Documents)/TRENDset
with the extension .csw. If a file is saved from an SD card, it is saved in
the folder My Files (Documents)/TRENDset/THERAPY SD_Archive with
the extension .csa.

Save as
With command Save as you can save the open file under another name
or in another directory.

Quit TRENDset
Select File > Exit.
Before quitting, TRENDset checks whether all data have been saved.
If you have not already sent your changes to the device, you will be
prompted to save.

17
Print

Print

Select File > Print or in the symbol bar. Use Ctrl+P for printing a file
1. Select the printer. quickly.
2. Set the number of copies to print.
3. Mark the documents you want to print by activating the check
box. You can select the number of events to be printed under
Alerts and Parameter changes.
NOTE:
As long as TRENDset is open
the last printer settings used
are saved.

4. Confirm with OK.

18
Pre-Settings

Pre-Settings
Setting the Language for TRENDset
Select Settings > Language.
You can choose from several languages. If you need an additional lan-
guage, contact our service. The selection of available languages can
be expanded at any time.

Registration and Deregistration for Service


Personnel
This menu items enables the access to the program’s service area. The
service area offers special actions and additional device data. The access
is password-protected so that only authorized personnel have access.
Registration
Select Settings > Service > Log in.
1. Enter your username.
2. Enter your password.
3. If the service area should be displayed each TRENDset is started,
activate the check box.
4. Confirm with OK.
Deregistration
Select Settings > Service > Log out.

19
Pre-Settings

Setting Event Limits for Statistics


Event limits establish if and after how much time the occurrence of
events should be displayed and recorded in the statistics.

Select Settings > Event Limits.


1. Activate an radio button.
The selection options are none, 5 s and 10 s.
2. Confirm with OK.

For these types of devices you can set the event limits.
• TREND II AUTO/CPAP
• VECTOR et AUTO/CPAP
• VECTOR ET AUTO/CPAP
• (i)CARAT et AUTO
• TREND 210
• AutoTREND
• VECTOR II FLEXLINE
• VECTOR II AUTO
• TREND II with SD cards
• VECTOR et with SD cards
• iCARAT et with SD cards

20
Communication with the Device

Communication with the Device


Search the Device Automatically
Select Communication > Automatic Device Search or in the symbol bar.

All founded devices are listed. Select the desired device by mouse
click. Afterwards starts the data import. This process will take at most
50 seconds. TRENDset changes to the status Online (see page 14). If no
device has been detected under Found devices, no devices will be dis-
played. TRENDset remains in the status Offline (see page 14). All program-
ming and analysis functions remain deactivated.

If TRENDset is connected
with a device, the status
message Online appears.

21
Communication with the Device

Search Device Manually


If multiple devices have been connected to your computer, apart of the
automatic device search function, you may have access to a particular
system by manual selecting of the respective COM-PORT.

Select Communication > Manual Device Search.


1. Select the COM-Port to which the device is connected.
2. TRENDset searches for a connected device.
a. When a device is detected, the data import starts. This pro-
cess will take at most 50 seconds.
b. If no device is detected, the message “No Connection to Ther-
apy Device” appears. Confirm this with OK. Select another
COM-Port and confirm again with OK. Repeat this procedure
until TRENDset has found the device.
3. TRENDset changes to the status Online (see page 14).

If no device is detected, TRENDset remains in the status Offline (see


page 14). All programming and analysis functions remain deactivated.

If TRENDset is connected
with a device, the status
message Online appears.

22
Communication with the Device

Troubleshooting
While TRENDset searches for a device, the following system mes-
sages can appear:

System message Cause


No Connection to Therapy A wrong COM-Port was selected.
Device.
The COM Port “x” is invalid The COM-Port is either unavailable or blocked
or is used by another ap- by another application.
plication.
No connection to therapy • The device is not turned on.
device at the COM Port “x”. • The serial cable is not properly con-
nected with the computer or the device.
• The device is connected with another
COM-Port.
• No device is connected.

23
Communication with the Device

Read Device Data


Select Communication > Read Device Data.
You will have the option of three sub-menus:
• Read Device Memory
• Read Device Parameters
• Read Patient Data (if supported)

Read Device Memory


Select Communication > Read Device Data > Read Device Memory.
The device memory consists, depending on device type, of a therapy
memory and a compliance memory. The therapy memory contains all
therapy data. In the compliance memory, only data on the use of the In these function areas you will
device are saved. You find the therapy data in the function areas Therapy find the imported therapy data.
Calendar, Data and Events as well as their statistical evaluation are found
in Statistics. The use of the device will displayed in the Therapy Calendar.

24
Communication with the Device

Reading the Device Parameters


Select Communication > Read Device Data > Read Device Paramaters.
The device parameters are displayed in the function areas Patient Data,
Device Settings as well as Times and Counters. In these function areas you
will find the imported device
parameters.

Reading Patient Data


Select Communication > Read Device Data > Read Patient data.
Patient data are imported and displayed in the function area Patient data.
In this function area you will
find the imported patient data.

25
Communication with the Device

Delete Device Data NOTE:


Select Communication > Delete Device Data. The menu item Delete
Device Data will only be dis-
You will have the option of four sub-menus: played when a device is rec-
ognized and TRENDset is in
• Delete Patient Data Online status.
• Delete Device Memory and Send Factory Settings
• Delete Device Memory
• Delete Factory Settings

Delete Patient Data WARNING!


The deletion process lasts
Select Communication > Delete Device Data > Delete Patient Data.
depending on device type
up to 60 sec. The device may
not be disconnected from the
Delete Device Memory and Send Factory Settings computer or the power supply
during this time.
Select Communication > Delete Device Data > Delete Device Memory and
Send Factory Settings.
This function is to be used if you wish to restore the factory settings on
the device and delete the device’s memory.

26
Communication with the Device

Delete Device Memory


Select Communication > Delete Device Data > Delete Device Memory. NOTE:
Before each action, you will be
The entire therapy memory, with all pressure and flow values and the asked if the command should
related events, will be deleted along with the compliance memory with actually be executed.
the data on device use.

Send Factory Settings


Select Communication > Delete Device Data > Send Factory Settings.
All parameters and properties of the device will be restored to their
factory-standard values. Additionally, all patient data will be deleted.

27
Communication with the Device

Using an SD card
There are two ways to store data on an SD card. NOTE:
• Respiratory therapy device with SD card (only in device models Before an SD card can be used
with SD card slot) for the DATA box or the respi-
ratory therapy device, it must
• DATA box with SD card be formatted with TRENDset.

Using an SD card in a respiratory therapy device


The last 100 events are saved on the SD card. Events are changes of
parameters such as therapy settings, resetting counters (e.g., filter
counters) etc. The parameters are saved with the date and time they
were changed, as well as the old and new value. All alarms that occurred
are also saved with date and time. The following therapy-related events
are also saved on the SD card:
Device type CPAP AUTO BILEVEL BILEVEL BILEVEL
Parameter ST20 ST30
Pressure     
Respiratory Flow     
Average System Flow
    
(Leakage Flow)
Hyperventilation     
Leakage     
Mode     
Frequency     

28
Communication with the Device

Device type CPAP AUTO BILEVEL BILEVEL BILEVEL


Parameter ST20 ST30
Relative Respiratory Tidal

Volume
Central Sleep Apnea 
Obstructive Sleep

Apnea
Mixed Sleep Apnea 
Hypopnea  
Apnea    
Airway Constriction 
Snoring 
General Artifact 
Adjustment     
Mouth Exhalation 
Tidal Volume    
Apnea alarm 
Low MV 

29
Communication with the Device

Use of the DATA box


The DATA box can store additional data such as oxygen partial pres-
sure and pulse frequency. These data are shown as a curve diagram
in the function area Data and Events.
The data can be read from an SD card on your computer and evaluated.
You will need a card reader device on your computer for this.

DATA box

30
Communication with the Device

Read SD card
Select SD-Card > Read SD card.
The dialogue box Read SD card will display the patient data that are
saved on the SD card.

Multiple patient data can


be saved on an SD card.

Activate a check box. Click on OK.


The patient data are imported and displayed in TRENDset.
When using a data box, TRENDset changes to the status DATA box
and when using a device with an SD card to the status SD card. You
can now change parameters in the function areas Device Settings or
Patient Data and save on the SD card via Send.
To load the files into the device, connect the DATA box to the device
and then plug the SD card into the DATA box or plug the SD card in the
SD card slot of the device.

31
Communication with the Device

Format SD card
Before an SD card can be used with a DATA box or a repiratory therapy NOTE:
device, it must be formatted with TRENDset. It is irrelevant whether this The SD card will not be
accepted by the DATA box
is a new or used SD card.
or the respiratory therapy
device unless it is formatted
Select SD-Card > Format SD card. with TRENDset.
The dialogue box Format SD card will open and will list all removable
disks for your computer. Select the data storage device by activating
the check box.
To prevent accidental erasures of files on the other removable disks, it
will check whether there are still files on the selected drive. If yes, you
will be prompted to deleted these files manually.

32
Communication with the Device

Save Data from SD cards


All of a patient’s data can be saved on the hard drive. Read the data Files from the SD card will be
from the SD card (SD-Card > Read SD card). Then save the file (File > stored with the extension .csa.
Save/Save as).
Unlike other TRENDset-files (*.csw) files from an SD card have the exten-
sion .csa. These file are saved in the sub-directory THERAPY SD_Archive.

Open Data from SD cards


Select File > Open or in the symbol bar.
Before a new file is opened TRENDset checks whether all open files NOTE:
have been saved. Files from the SD card (*.csa )
As soon as a file is opened, TRENDset switches to the THERAPY SD_ contain patient data. These
files are often much larger than
Archive status. In this status you cannot store any patient data or change
files with the extension .csw.
parameters on the SD card.

33
Communication with the Device

Information about TRENDset


Select ? > About.
The version number of TRENDset will be displayed along with informa-
tion about the manufacturer.

Using Help
Help opens immediately if you
Select ? > Help TRENDset. press F1.
The Help file will open in PDF-Format.
NOTE:
Help opens with Acrobat
Reader. You can install the
program from the TRENDset-
CD if needed.

34
Patient Data

Patient Data
You manage patient data in the function area Patient Data, turn the ther-
apy on and off and get information about the most important charac-
teristics of the devices.
The workspace is sub-divided into Patient Data, Display and Device
Characteristics.

Patient Data

Device
Characteristics

Display

35
Patient Data

Entering, Changing and Managing Patient Data


All text fields highlighted in white can be written, changed and deleted.
Changes will be highlighted in red.
To restore all entries to their original state, click on Reset.
VECTOR et
Click on Send, in order to send changed files. Then the data are read (i)CARAT et
again and displayed in TRENDset. The sent and received data are com- VECTOR ET
pared with each other. This assures that the correct data was written TREND II
in the device. point
TREND 210
Not all devices can save patient data. If there is a connection to devices AutoTREND
that cannot store any patient data, enter the patient data manually and (as of device software 1.502)

confirm the entries with Accept. Save the file (File > Save/Save as). VECTOR II FLEXLINE
VECTOR II AUTO
(as of device software 1.502)

These device types can


save patient data.

36
Patient Data

Display current Therapeutic Pressure


In the Display area the current pressure is displayed along with the
device status.
Click on the control button On / Off and the therapy starts.

Pressure Display (graphic) Turn Therapy on-/off

Device Status

Pressure Display (numeric)

37
Patient Data

Display Device Characteristics


In the area Device Characteristics you see the main features of the
devices. They are purely informative and cannot be changed.

38
Device Settings

Device Settings
In the function area Device Settings you set the device parameters.
The workspace, depending on the device connected, is subdivided
into Pressure, Respiration, Softstart Ramp/Mask Test, Mask Alarm, Device
and Intro Text.

Pressure (numeric) Soft start Ramp/Mask Test

Graphic Display of the


Selected Parameters

PIN code
Activation

Respiration Device Intro Text for


Parameters Parameter the Device

39
Device Settings

Set Therapeutic Pressure


Select Device Settings > Pressure.
You can change the therapeutic pressure in two ways:
a. using the check box
b. using the diagram
Hold the mouse down and drag it along the red points. The drag
direction is indicated by arrows. The settings in the check boxes
will change automatically.

You activate or deactivate the trigger using the On / Off button. The
active state is shown left of the button.
All changes will be highlighted in red.
To restore all entries to the initial condition, click on Reset.
Click on Send, to send the changed data to the device. Then the data
will be read again and displayed in TRENDset. The sent and received
data are compared with each other. This assures that the correct
data was written in the device.

40
Device Settings

Definition of Terms

Therapy Pressure Level of therapy pressure


FLEX Level The level of pressure increase or decrease is depending on the flow of breath, the FLEX Levels set
and the therapeutic pressure setting. The three FLEX Levels (1- 3) are for individual adjustment to
the patient.

I-FLEX The value of pressure increase depends on the value of the breathing flow, the set I-FLEX and the set
therapy pressure. The I-FLEX settings (0 – 3) are for individual adjustment to the patients breathing flow.
E-FLEX The value of pressure decrease depends on the value of the breathing flow, the set E-FLEX and
the set therapy pressure. The E-FLEX settings (0 – 3) are for individual adjustment to the patients
breathing flow.
Start Pressure Starting pressure
Min. Pressure Minimum pressure
Max. Pressure Maximum pressure
Pressure Steepness The parameter Pressure Steepness increase determines how quickly pressure increases in an event.
The speed of the pressure increase can be adjusted to the patient’s sensitivity.
Inspiration Pressure Inspiration pressure
Expiration Pressure Expiration pressure
I-Slope Inspiration slope
E-Slope Expiration slope
Trigger Activating or deactivating the inspiration trigger in T-Mode
I-Trigger Sensitivity of the inspiration trigger
E-Trigger Sensitivity of the expiration trigger
Trigger Setting the trigger sensitivity
Target Volume Value for the target volume (only in the S, ST and T modes)
Additional Pressure Adjustable additional pressure, with target volume activated (only in the S, ST and T modes)
Min. Inspiration Time TI min Minimum inspiration duration (only in the S and ST modes)
Max. Inspiration Time TI max Maximum inspiration duration (only in the S and ST modes)

41
Device Settings

Available Parameters Depending on Therapy Mode


Mode CPAP FLEX APAP BILEVEL BILEVEL BILEVEL
S ST T
Therapy Pressure  
FLEX-Level  The devices have
Start Pressure  different parameters.
Min. Pressure 
Max. Pressure 
Pressure

Steepness
Inspiration
  
Pressure
Expiration
  
Pressure
I-FLEX  
E-FLEX  
I-Slope     only for VECTOR et, VECTOR
ET, (i)CARAT et, TREND II
E-Slope   
 only for VECTOR et BILEVEL
Trigger (T-Mode) 
ST30, iCARAT et BILEVEL
I-Trigger   ST30, TREND II BILEVEL
E-Trigger   ST30

Trigger     only for VECTOR et BILEVEL


ST20 und ST30, iCARAT et
Target Volume    BILEVEL ST20 und ST30,
Additional TREND II BILEVEL ST20 and
  
Pressure ST30
TI min    only for VECTOR et, TREND II
und iCARAT et as of device soft-
TI max  
ware 2.140

42
Device Settings

Set Respiration Parameters


Select Device Settings > Respiration.
There are two ways to change the inspiration:
a. using the check box
b. using the diagram
Hold the mouse down and drag it along the red points. The drag
direction is indicated by arrows. The settings in the check boxes
will change automatically.

All changes will be highlighted in red.


To restore all entries to their original condition, click on Reset.
Click on Send, to send the changed data to the device. Then the data
will be read again and displayed in TRENDset. The sent and received
data are compared with each other. This assures that the correct data
was written in the device.

43
Device Settings

Definition of Terms
Frequency Breaths per minute
Inspiration Duration of inspiration over entire breath
Delay Time Delay for switch between from S-Mode to T-Mode
I:E Ratio I:E-Ratio from frequency and duration of inspiration
Backup frequency Frequency with which the patient continues to be ventilated
when an apnea has occurred

Available Parameters Depending on Device Type

Mode CPAP FLEX APAP BILEVEL BILEVEL BILEVEL


S ST T The devices have
different parameters.
Frequency  
Inspiration  
Delay Time 
Backup

frequency

44
Device Settings

Set Softstart Ramp and Mask Test


Select Device Settings > Softstart Ramp/Mask Test.
There are two ways to change the parameters for the Softstart Ramp
and the Mask Test:
a. using the check boxes
b. using the diagram
Hold the mouse down and drag it along the red points. The drag
direction is indicated by arrows. The settings in the check boxes
will change automatically.

All changes will be highlighted in red.
To restore all entries to their original condition, click on Reset.
Click on Send, to send the changed data to the device. Then the data
will be read again and displayed in TRENDset. The sent and received
data are compared with each other. This assures that the correct
data was written in the device.

45
Device Settings

Definition of Terms
Softstart Time Time during which the pressure gradually increases
to therapeutic pressure, beginning at the starting
pressure for soft start or at minimum pressure.
Softstart Start Pressure Starting pressure for soft start
Mask Test In a mask test the device is operated at maximum
therapeutic pressure for the time selected in order
to test the tightness of mask. The maximum ther-
apeutic pressure results from the settings in the
function area Device Settings.

46
Device Settings

Available Parameter Depending on Device Type


Device Softstart Time Softstart Mask Test
Start Pressure
VECTOR et   
(i)CARAT et   
VECTOR ET   
TREND II    The devices have
point    different parameters.

TREND 200  
TREND 210   
AutoTREND   
TREND 300  
TREND 400  
TREND 500  
TRENDlife  
Calypso  
VECTOR  
VECTOR II  
VECTOR II
  
FLEXLINE
VECTOR II AUTO   
VECTOR II BILEVEL  

47
Device Settings

Set Alarm parameter


Activate/Deactivate Mask Alarm
Select Device Settings > Mask Alarm.
You activate or de-activate the mask alarm using the check box. The
active state is displayed left of the button.
There are two ways to change the parameters for the Mask Alarm:
a. using the check boxes
b. using the diagram
Hold the mouse down and drag it along the red points. The drag
direction is indicated by arrows. The settings in the check boxes
will change automatically.
All changes will be highlighted in red.
To restore all entries to their original condition, click on Reset.
Click on Send, to send the changed data to the device. Then the data
will be read again and displayed in TRENDset. The sent and received
data are compared with each other. This assures that the correct
data was written in the device.

48
Device Settings

Definition of Terms
Mask Alarm The mask alarm is activated/deactivated using the
On / Off button.
Alarm Delay 1 This value defines the delay period after which the
alarm will sound.
Pressure Difference 1 This value defines the tolerated deviation from the
set pressure.
1
only when the Mask Alarm is activated

49
Device Settings

Set Alarms “Low MV” and “Apnea”

Select Device Settings > Alarm Low Minute Volume or Alarm Apnoa.
Set the desired values using the selection fields. All changes are high-
lighted in red.
To restore all entries to their original condition, click on Reset.
Click Send to send the changed data to the device. The data is sub-
sequently read out again and displayed in TRENDset. The sent and
received data is compared. This ensures that the correct data has
been saved on the device.

Definition of Terms
Alarm Low Minute Alarm if the respiratory minute volume is too low
Volume
If the respiratory minute volume falls below the value
set here, the device sounds an alarm.
Alarm Apnea Alarm if an apnea is detected

If an apnea is detected that is longer in duration


than the period of time set here, the device sounds
an alarm.

50
Device Settings

Setting Device Parameters


Select Device Setting > Device.
Use the check boxes to change the parameters.

All changes will be highlighted in red.


To restore all entries to their original condition, click on Reset.
Click on Send, to send the changed data to the device. Then the data
will be read again and displayed in TRENDset. The sent and received
data are compared with each other. This assures that the correct data
was written in the device.

51
Device Settings

Automatic Function
To set the automatic pressure adjustment, activate the check box Auto-
matic function adjustable. This will activate the check box “Automatic”
with the following options.

OFF no automatic pressure adjustment possible

Start/Stop • Automatic Pressure Threshold On


• Automatic Pressure Threshold Off
• Automatic Delay
adjustable
Start • Automatic Pressure Threshold On
adjustable

You can find additional information about the automatic function on page 53.

52
Device Settings

Definition of Terms

Mode Device mode


Language Device‘s menu language
Pressure Unit Indicates in what units the pressure should be displayed
Brightness Display Brightness of the display
Pressure Ramp Soft increase in pressure when pressure changes
Heating Step Selection of the heating level of the humidifier
Menu Type Menu, which displayed in the device
Activate PIN Code When you activate the check box, you can enter a four-digit PIN Code. Once you have sent the PIN Code
to the device, you can only change the device’s pressure parameters and mode after entering the correct
PIN Code.
Automatic • Automatic Threshold On
The first value indicates inspiration, the second value expiration.
These values must be exceeded for the therapy to start.
• Automatic Threshold Off
The pressure must fall below this value for the therapy to stop.
• Automatic Delay
Time after which the therapy is automatically stop when the pressure drops below the threshold
value.
Display Breath Here you can define whether the display of the tidal volume is activated or deactivated during running therapy.
Status
Display Efficiency Here you can define if the display of efficiency data, such as apnea index (AI) or the apnea-hypopnea index
Data (AHI in CPAP and AUTO devices) and the leakage (Leak) are displayed in the info menu.
Mask Type Selection of the mask type in use
Intro Text The intro text is shown when the device starts. It can be a maximum of 2 lines of 16 characters each. You
can choose from several pre-defined texts or enter your own text. Select the check box Other text.
Menu Scrolling Here you can set whether scrolling is allowed in the menu or not.

53
Device Settings

Available Parameters Depending on Device Type

Pressure Ramp

Menu Scrolling
Display Breath
Heading Step

Threshold Off
Pressure Unit

Threshold On

Display Effi-
ciency Data
Brightness

Menu Type

Mask Type
Automatic
Automatic

Automatic

Automatic
Language

PIN Code

Intro Text
Pressure

Pressure
Display

Status
Mode

Delay
Device
VECTOR et               
(i)CARAT et               
VECTOR ET              
TREND II               
point          
TREND 200       
TREND 210         
AutoTREND         
TREND 300      
TREND 400       
TREND 500       
TRENDlife      
Calypso      
VECTOR       
VECTOR II       
VECTOR II FLEXLINE         
VECTOR II AUTO         
VECTOR II BILEVEL       
 as of device software 1.501  only for the respective AUTO-Devices  as of device software 2.140
54
Device Settings

Working with the Parameter-Popup


The Parameter-Popup is a small dialogue box which contains the essen-
tial functions from Device Settings.

Select Settings > Parameter-Popup or in the symbol bar.


The user interface will close and the Parameter-Popup will open. The
advantage: You can use TRENDset and other applications on the screen
at the same time and continue to control the device.
You can only use the Parameter-Popup in the Status Online, SD-Card
and DATA box.
If necessary you can change the position and size of the Parameter-
Popup. Hold down the mouse and drag the edges of the window.
To close the Parameter-Popup and return to the usual user interface
panel, click on the “x” on the upper right.

Save space on your screen


with the Parameter-Popup.

After closing the Parameter-


Popup, its position and size
will be saved.

55
Therapy Calendar

Therapy Calendar
The function area Therapy Calendar includes a graphic display of device
The therapy data are evalu-
use. It shows the duration of therapy per day and in some devices the ated in the function area
pressures used. The Therapy Calendar gets this data from the therapy Data and Events.
memory, the compliance memory and the efficiency data memory. 1

The Therapy Calendar,


shown here with display
of pressures used.

1  only TREND II, VECTOR et and iCARAT et as of device software 2.140

56
Therapy Calendar

Therapy Memory
Approximately 30 days of therapy are stored in the Therapy Memory.

Compliance Memory
When the Therapy Memory is full, the older data are transferred to the
Compliance Memory. The Compliance Memory stores data for approxi-
mately 1 year.
The data from the Compliance Memory are only displayed in the Ther-
apy Calendar and in the protocol from the Therapy Calendar. There
is no display or evaluation of these data in the function area Data and
Events or Statistics.

Efficiency Data Memory


Devices from device software 2.140 have an efficiency data memory.
The following values are saved in the efficiency data memory:
• Apnea index (AI) or apnea-hypopnea index (AHI)
• increased leakage (Leak) in %
• average operating time (Use)
The values are saved for one day, one week, one month, 6 months and
one year, respectively.

57
Therapy Calendar

Available Memory Depending on Device Type


Device Therapy Memory Compliance Memory Efficiency Data Memory
VECTOR et CPAP and AUTO   
VECTOR et BILEVEL and BILEVEL ST 20 and ST30   
(i)CARAT et AUTO   
(i)CARAT et BILEVEL and BILEVEL ST 20 and ST30   
VECTOR ET CPAP and AUTO   
VECTOR ET BILEVEL and BILEVEL ST 20   
TREND II CPAP and AUTO   
TREND II BILEVEL and BILEVEL ST 20 and ST30   
point 
TREND 200 
TREND 210  
AutoTREND  
TREND 300 
TREND 400 
TREND 500 
TRENDlife
Calypso
VECTOR 
VECTOR II 
VECTOR II FLEXLINE  
VECTOR II AUTO  
VECTOR II BILEVEL 
 VECTOR et, TREND II and iCARAT et as of device software 2.140

58
Therapy Calendar

Therapy Calendar Areas

Pressure, numerical in hPa


Color definition of the
specific pressures
Day Display
Leakage in % 1

Date of the Evaluation and Therapy duration AI or AHI (CPAP and


therapy day Statistics (On/Off) with indication of AUTO devices) 1
the pressures used
Duration of therapy Therapy duration
without indication of with indication of the
the pressures used pressures used

A therapy day lasts from


12:00 noon to 12:00 noon.
When you move the mouse
across a curve or bar, a
Quick info is displayed in
overlay. 1
only TREND II, VECTOR et and iCARAT et as of device software 2.140

59
Therapy Calendar

Select Therapy Days for Statistics


To evaluate the therapy data in the function areas Data and Events as
well as in Statistics, mark the desired therapy days. Click on the sym-
bol to the right of the date.

selected
(red symbol)

not selected
(gray symbol)

If you want to select all therapy days, activate the control box Select
all displayed Days.
Additionally, you can decide whether or not therapy-free days will be
displayed in the Therapy Calendar. Activate the check box Display Days
without Therapy in Addition.
After you have marked the therapy days, you can switch to the func-
tion area Data and Events or Statistics and evaluate the data for the
marked therapy days.
The Complete Using Time in h and the Average Using Time per Day in h
are seen in the lower part of the workspace.

60
Therapy Calendar

Open Protocol on Duration of Therapy


Click on Protocol.
The protocol will include, along with
data on the daily duration of therapy, informa-
tion on the recording time, on the days used
for therapy, on the total duration of use, and
on the average duration of use per day.
The devices VECTOR et, TREND II and
iCARAT et, from device software 2.140, also
have an efficiency data memory.
To close the window, click on the “x” on the
upper right.

The protocol evaluates the


Therapy, Compliance and
Efficiency Data Memory.

61
Data and Events

Data and Events


The data can be analyzed in the function area Data and Events. Addition-
ally, you can monitor the course of therapy in real time. The diagrams
enable you to evaluate the data minute-by-minute.

When you move the


mouse across a curve or
bar, a Quick info is dis-
played in overlay.

62
Data and Events

Analysis of Data and Events


Definition of Terms

Pressure Pressure in patient’s mask


Respiratory Flow Patient’s flow
Average Flow which escapes from the exhalation valve. In a leakage, i.e. if the mask slips out of face, the
System Flow flow will increase.
Relative Respira- The relative volume of breath is calculated from the breath-flow.
tory Volume (%)
Oxygen Average oxygen saturation
Pulse Patient’s current pulse
Central Sleep A central sleep apnea is indicated if, in addition to a limitation of respiratory flow, little or no
Apnea change in the respiratory flow is detected for the time of the set event limit, without simultane-
ous airway constrictions.
Obstructive An obstructive apnea is indicated if, in addition to a limitation of respiratory flow, little or no change
Sleep Apnea in the respiratory flow is detected for the time of the set event limit, and if the airway constriction
during this time is above a threshold value.
Mixed Sleep A mixed apnea is indicated if the apnea consists of a central and an obstruction portion and
Apnea there is no normal breathing between the two portions.
Apnea An apnea is indicated if, in addition to a limitation of respiratory flow, little or no change in the
respiratory flow is detected for the time of the set event limit.
Hypopnea A hypopnea is indicated if a respiratory flow limitation lasts for the time of the set event limit.
Airway An airway constriction is indicated if the resistance of the upper airways is above a threshold
Constriction value for the time of the set event limit.

63
Data and Events

A snoring phase is indicated if the snoring sound is above a threshold value for the time of the
Snoring
set event limit.
Device State Adjustment The adjustment phase includes the learning phase at the begin-
ning of a therapy and the 30 s timeout after an increased leak-
age, a mode switch between APAP and CPAP as well as after a
pressure change during CPAP therapy. During this period, the
APAP-Algorithm is turned off.
Leakage A leakage is indicated, if the middle system flow exceeds a
threshold value for a maximum of 2 l/s or a shows a change of
more than 300 ml/s.

Mouth Exhalation Displaying an expiration by mouth

Hyperventilation Displaying a hyperventilation

General Artefact Displaying a general artefact

CPAP Displaying an interrupted FLEX-Function

Mode Display of Active Mode (APAP, CPAP, FLEX, BILEVEL S, BILEVEL ST, BILEVEL T)

64
Data and Events

Displaying Data and Events Depending on Device Type


Device VECTOR et/ET CPAP, VECTOR et/ET AUTO, VECTOR et/ET BILEVEL, point
Trend II CPAP, TREND II AUTO, BILEVEL ST20, BILEVEL ST30,
TREND 210, (i)CARAT et AUTO, TREND II BILEVEL,
VECTOR II FLEXLINE AutoTREND, VECTOR II AUTO BILEVEL ST20, BILEVEL ST30,
(i)CARAT et BILEVEL,
BILEVEL ST20, BILEVEL ST30
Mode CPAP FLEX CPAP/AUTO FLEX all Modes CPAP
Central Sleep Apnea 
Obstructive Sleep Apnea 
Mixed Sleep Apnea 
Hypopnea   
Apnea   
Airway Constriction  
Snoring  
Mode     , 

Adjustment    
Leakage     
Hyperventilation     
Mouth Exhalation 
General Artefact 
CPAP  
Pressure      
Respiratory Flow , 
 , 
 , 
 , 
 , 

Average System Flow     , 

Relative Respiratory Volume , 
 
Qxygen      
Pulse      
 Standard  only when connected to a DATA box  only when connected to a DATA box with Oxygen-Kit
 Standard for VECTOR et, TREND II and iCARAT et as of device software 2.140 65
Data and Events

Select Therapy Days


Below the diagram are the control elements for selecting the day, which
you want analyze.

Back One Day Last Therapy Day


First Therapy Day Forward One Day

Direct Day Selection

Zoom View
Below the diagram are the control elements for setting the zoom
You enlarge the Zoom Area
area – 1 min to 24 h is possible.
by double-clicking with the
Enlarge Zoom Area left mouse button; by double-
Direct Selection
clicking on the right mouse
of Zoom Area
button you reduce the Zoom
Area.
Reduce Zoom Area

Zoom Area 6 h

66
Data and Events

Enlarge and Reduce Diagrams


For a better display you can zoom the size of the diagrams.

To enlarge, click on .

To reduce, click on .

67
Data and Events

Monitor Course of Therapy in Real Time


Click on Run Real Time Monitoring.
The data will be updated every 500 ms. In contrast to the therapy data, NOTE:
all data are shown, including events, which are very short and otherwise The device must be in opera-
are not recorded. The event limits have no influence here. tion for data to be displayed.

To end the connection, click on Stop Real Time Monitoring.

68
Data and Events

Real Time-Monitoring Depending on Device Type


Device TREND II CPAP, TREND II AUTO, TREND II, all
VECTOR et/ET CPAP, VECTOR et/ET AUTO, VECTOR et/ET, other
(i)CARAT et CPAP, (i)CARAT et AUTO, (i)CARAT et BlLEVEL, devices
TREND 210, AutoTREND, BlLEVEL ST20,
VECTOR II FLEXLINE VECTOR II AUTO BlLEVEL ST30
Mode CPAP FLEX CPAP FLEX APAP
Pressure       
Respiratory Flow     
Average System Flow     
Relative Respiratory
Volume
 

Airway Constriction   
Snoring  
Mode     
Adjustment   
Leakage     
Hyperventilation      
Mouth Exhalation  
General Artefact   
CPAP  

 Standard
 Standard for VECTOR et, TREND II and iCARAT et as of device software 2.140

69
Alerts and Parameter changes

Alerts and Parameter Changes


In the function area Alerts and Parameter changes, you can view and
evaluate events such as
• alarms,
• parameter changes and
• the deletion of times and counters Device settings at the
time of the event
which are saved on an SD card.

Filter options Marked event


(gray highlight) 70
Alerts and Parameter changes

Display of the Events

Type Time of event Description


of event (date and time) of event

Using the Filter Options


The filter options enable you to filter out the events that are of inter-
est to you. Click on the relevant checkbox to activate or deactivate it.

Activated = events are displayed

Deactivated = events are not displayed

Date and time Parameter


changed change

End or start Counter reset


of an alarm

71
Statistics

Statistics
In the function area Statistics, the therapy data are visualized with bar NOTE:
diagrams. This allows you to precisely analyze therapy data. The dia- In Statistics, only the pure ther-
grams support you in titrating an effective therapeutic pressure. apy time without Mask Test or
Softstart is evaluated.

TREND II AUTO
VECTOR et AUTO
VECTOR ET AUTO
(i)CARAT et AUTO
AutoTREND
VECTOR II AUTO

The statistics function is avail-


able for these device types.

When you move the


mouse across a curve or
bar, a Quick info is dis-
played in overlay.

72
Statistics

Select Therapy Days


Select all activated days in the Therapy Calendar
In the selection box Therapy Day “See Therapy Calendar” is pre-selected.
This assures that all selected days in the Therapy Calendar are included
(see page 59).
The sum of events is shown in the diagrams.
The number of selected days can be seen under the check box. The
value in brackets shows the ratio of selected days to total therapy days
in percent.

Selection of all therapy First Therapy Day


days that were activated
for the statistics function Back One Day

Number of therapy days that were activated 60 % of all therapy days were activated
in the Therapy Calendar for the statistics in the Therapy Calendar for the statistics
function (see page 59) function

73
Statistics

Select Individual Therapy Days


Below the diagram are the control elements for selecting the day, which
you want evaluate.
You can select all days, regardless of whether they are activated in the
Therapy Calendar or not.
The number of selected days can be seen under the selection box.
The value in brackets shows the ratio of selected days to total therapy
days in percent.

Forward One Day First Therapy Day


Last Therapy Day Back One Day

Direct Date Selection

Number of therapy
days selected in the Number of therapy
selection box days selected using the
selection box in %

74
Statistics

Analyze Distribution of Pressure Over Time


With the help of the Temporary Pressure Distribution diagram, you ana-
lyze the percentage distribution of pressures over the selected therapy
period.

On bar width stands for


0.5 hPa.

When you move the


mouse across a curve or
bar, a Quick info is dis-
played in overlay.

75
Statistics

Analyze Frequency of Events


With the help of the Frequency of Events diagram, you analyze the num-
ber of central, obstructive and mixed apneas, hypopneas and airway
constrictions depending on pressures. The snoring phases are in con-
trast given in minutes.

One bar width stands for


0.5 hPa.
The selection in the radio but-
tons remains in effect until the
next change.

When you move the


mouse across a curve or
bar, a Quick info is dis-
played in overlay.

To select an event, click on one of the radio buttons. You can further
specify whether the total number or the average number per hour should
be displayed, and for snoring phases total minutes or minutes per hour.

76
Statistics

Effect of the Therapeutic Pressure


With the help of the Effect of Therapy Pressure diagram, you analyze the
percentage occurrence of events depending on the therapy pressure.
Snoring phases are not taken into account.

One bar width stands for


0.5 hPa.

When you move the


mouse across a curve or
bar, a Quick info is dis-
played in overlay.

77
Statistics

Chronological Distribution of Events


With the help of the Distribution of Events based on Time diagrams, you
analyze the percentage occurrence of events depending on the time
of day. The snoring phases are not taken into account.

One bar width stands


for 1 h.

When you move the


mouse across a curve or
bar, a Quick info is dis-
played in overlay.

78
Statistics

Statistical Analysis
Click on Statistical Analysis.
The overview includes along with the statis-
tical data, data on the recording time, dura-
tion of therapy, number of days selected
for the statistics and the most important
device settings.
As an aid to decision-making for the phy-
sician, the lower window area shows what
therapeutic pressure was used 90 % of the
time and at what therapeutic pressure 90 %
of the events occurred.
To close the window, click on the “x” at the
upper right.

79
Times and Counters

Times and Counters


In the function area Times and Counters you will find information on the
device’s counters, set the date and time and control the alarm clock Changed values that have
function. not yet been transferred to
The workspace is sub-divided into Counter, Date and Time and Alarm the device are shown in red.
Clock.

80
Times and Counters

Run Times of the Device or Components


In the Counter area, data on the device’s run times or those of the
device’s components are saved.
Filter Display Interval for filter change
(->FILTER<-)
Blower Display Interval for service checkup, based on compressor
(->SERVICE<-) run time
Filter Age Compressor run time since last filter change
Standby Time Time in which the device was on
Blower Running Time in which the compressor was on, incl. mask tests
Time and soft start
Therapy Time Time in which the compressor was on and the mask
was at therapeutic pressure

81
Times and Counters

Reset Filter Age


In the Counter > Filter Age area, press the Delete button. The command
will be sent to the device immediately and the counter will be reset to 0.

82
Times and Counters

Set Date
1. In the Date selection box, mark the day, month or year and enter
the desired value.
OR
Click on the arrow in the selection box. A calendar will open.
Select the date.
OR
It is not possible to set the
To import the date from the computer, just click on system time or to change the
Set Computer Time. date on the SD card. In order
to make these changes, the
If you want to remove you settings again, click on Reset. device must be connected.
2. Click on Send. The date will be sent to the device.

83
Times and Counters

Set Time
1. In the Time selection box, mark the hour, minutes or seconds and
enter the desired value. You have the option of using the arrows
to increase or decrease the marked values.
OR
It is not possible to set the
To import the time from the computer, click on Set Computer Time.
system time or to change the
If you want to remove you settings again, click on Reset. date on the SD card. In order
to make these changes, the
2. Click on Send. The time will be sent to the device. device must be connected.

84
Times and Counters

De-Activate Alarm Clock Function


The Alarm Clock Function can be de-activated on some device types.
• TREND II
• VECTOR et
• VECTOR ET
• (i)CARAT et
Remove the check in the check box Alarm Clock function adjustable
and click on Send.

NOTE:
If you de-activate the Alarm Clock Function, it is no longer available in
the device until you re-activate it.

85
Times and Counters

Set Alarm Time


1. Activate the alarm clock.
2. In the Time field, mark the hour or the minute and enter the
desired value. If you wish, you can use the arrows to increase or
decrease the marked value.
3. Set the snooze time in the selection box Slumber. You can select The snooze time (Slumber)
defines the time interval which
a time interval from 5 to 30 minutes. To de-activate the snooze passes after turning off the
time, select Off. alarm before the alarm sounds
again.
If you want to remove you settings again, click on Reset.
4. Click on Send. The alarm time will be sent to the device.

Alarm is on

86
System Messages

System Messages
If data are transferred incorrectly or incompletely over the serial inter-
face, TRENDset will issue a system message.
If a system message from 1 to 255 is issued, the device has regis-
tered the fault and exported it. For values seen on the following table,
TRENDset has only registered the fault.

System Message Description


1020 COM-Port could not be opened
1030 Incorrect data received
1040 Transfer not started or interrupted
1050 Checksum test failed
1060 Value field violated
1070 Data not received back after sending
1080 Sent data do not match imported value
1090 Too little data received
2000 TRENDset does not support the connected device

87
HOFFRICHTER GmbH
Mettenheimer Straße 12 / 14
19061 Schwerin
GERMANY
Phone: +49  385  39925 - 0
Fax: +49  385  39925 - 25
[email protected]
www.hoffrichter.de

TRENDset-eng-041111-03

You might also like