Manual Cloro Merk
Manual Cloro Merk
June 2019
1.14978.0001 6. Preparation
• Analyze immediately after sampling!
• Check the chlorine content with the MQuant® Chlorine Test.
MQuant ®
Cl2
Samples containing more than 2.0 mg/l Cl2 must be diluted with distilled
water.
Chlorine Test • The pH must be within the range 4 - 8.
with liquid reagent Adjust, if necessary, with sodium hydroxide solution or sulfuric acid.
• Filter strongly turbid samples.
Sample material:
Groundwater and surface water Notes on the measurement:
Drinking water and mineral water
• The color of the measurement solution remains stable for only a short
Waters from aquaculture
time.
Wastewater
Electroplating wastewater • Turbidity in the measurement solution makes the color comparison more
Disinfectant solutions difficult.
This test is not suited for seawater. • If the color of the measurement solution is equal to or more intense
than the darkest color on the scale, repeat the measurement using
4. Influence of foreign substances fresh, diluted samples until a value of less than 2.0 mg/l Cl2 is obtained.
This was checked individually in solutions containing 1 mg/l Cl2. The con- • In the event of chlorine concentrations exceeding 25 mg/l, other reaction
centrations of foreign substances given in the table lie below the limit at products are formed and false-low readings are yielded. In such cases it
which the determination is interfered with. Cumulative effects were not is advisable to conduct a plausibility check of the measurement results
checked; such effects can, however, not be excluded. by diluting the sample (1:10, 1:100).
• Concerning the result of the analysis, the dilution (see also section 6)
Concentrations of foreign substances in mg/l or % must be taken into account:
Al3+ 250 Mn2+ 100 Br2 0.2 Result of analysis = measurement value x dilution factor
Ca2+ 1000 NO2- 0.1 ClO2 0.2
CN- 0.1 S2- 0.1 I2 0.4
CO32- 1000 H2O2 0.05
Cr3+ 250 O3 0.05 8. Method control
Cr2O7 0.1 NaCl
2-
10 % To check test reagents, measurement device, and handling:
Cu2+ 100 NaNO3 10 % Freshly prepare a chlorine standard solution containing 1.0 mg/l Cl2
Fe3+ 100 Na2SO4 10 % (application see the website) and immediately analyze as described in
section 7.
5. Reagents and auxiliaries Additional notes see under www.qa-test-kits.com.