BC 5380 Feature and Specification - 20071127 - PDF

The BC-5380 Auto Hematology Analyzer performs 5-part differentiation on 27 parameters with 1 scattergram and 3 histograms using semi-conductor laser scatter and chemical dye methods. It has a 30 tube autoloader, tests up to 60 samples per hour with a 20ul sample volume, and stores up to 40,000 sample results. The analyzer uses impedance for WBC, RBC, PLT counting and flow cytometry for independent basophil analysis.

Uploaded by

Nur Cholis
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
435 views

BC 5380 Feature and Specification - 20071127 - PDF

The BC-5380 Auto Hematology Analyzer performs 5-part differentiation on 27 parameters with 1 scattergram and 3 histograms using semi-conductor laser scatter and chemical dye methods. It has a 30 tube autoloader, tests up to 60 samples per hour with a 20ul sample volume, and stores up to 40,000 sample results. The analyzer uses impedance for WBC, RBC, PLT counting and flow cytometry for independent basophil analysis.

Uploaded by

Nur Cholis
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

BC-5380 Auto Hematology Analyzer

Features:

’ 5-part differentiation, 27 parameters, 1 scattergram and 3 histograms

’ Semi-conductor Laser scatter combined with chemical dye method,


advanced flow cytometry

’ Compact, powerful and affordable

’ Only 20ul sampling volume

’ 30 tubes autoloader with random access

’ Closed tube for STAT samples

’ Up to 60 samples per hour

’ 2 counting modes: whole blood and prediluted

’ Capability to flag abnormal samples

’ Large storage capacity: up to 40,000 samples


Technical Specifications:

Principles
Impedance method for WBC, RBC, PLT counting
Cyanide-free reagent for Hemoglobin test
Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method,
independent Basophil channel

Parameters
WBC, Lym%, Mon%, Neu%, Eos%, Bas%, Lym#, Mon#, Neu#, Eos#, Bas#, RBC,
HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, LIC%, LIC#,
ALY%, ALY#
3 histograms and 1 scatter gram

Sampler
30 tubes autoloader with random access
Closed tube compartment

Throughput
Up to 60 samples per hour

Sample Volume
Whole Blood: 20ul
Prediluted: 20ul

Test Mode
CBC CBC+DIFF

Performance
Carryover Precision Linearity
WBC ≤0.5% ≤2.0% (4-15×109/L) 0.00-99.99×109/L
RBC ≤0.5% ≤1.5% (3.5-6.0×1012/L) 0.00-8.00×1012/L
HGB ≤0.6% ≤1.5% (110-180g/L) 0-250g/L
PLT ≤1.0% ≤4.0% (150-500×109/L) 0-1000×109/L

Data Storage Capacity


Up to 40,000 results including numeric and graphical information

Communication
LAN port supports HL7 protocol

Operating Environment
Temperature: 15oC-30oC
Humidity: 30-85%
Air Pressure: 70-106 kPa

Power Requirement
A.C. 100-240V ≤300VA 50/60Hz
Dimension and Weight
570mm(L) × 590mm(W) × 525mm(H)
Weight: 56.1 kg